BC-5180&5380 - Installation Guide - V2.0 - EN
BC-5180&5380 - Installation Guide - V2.0 - EN
INSTALLATION GUIDE
Warning!
The manufacturer takes no legal obligation to inform the users before making any
change to this guide.
Symbols and Precautions
1. Symbols used in the guide
You may find the following symbols on the analyzer, reagents, controls or calibrators.
When you see... It means...
BIOLOGICAL RISK
HIGH VOLTAGE
I
EARTH (GROUND)
ALTERNATING CURRENT
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
Manufacturer
TEMPERATURE LIMITATION
II
3.Precautions
Be sure to observe the following precautions during installation for the safety of patients and
operators.
All the analyzer components and surfaces are potentially infectious, so take
proper protective measures for operation and maintenance.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
It is important for the hospital or organization that employs this equipment to
carry out a reasonable installation plan. Neglect of this may result in machine
breakdown or injury of human health.
Connect the analyzer to a socket having sole fuse and protective switch. Do
not use the same fuse and protective switch with other equipment (e.g. life
supporting equipment). Otherwise, the equipment failure, over current or
impulse current that occurs at the startup moment may lead to tripping.
To prevent personal injury during the maintenance, keep your clothes, hairs
and hands from the moving parts, such as sample probe, pincher and piercer.
III
Be sure to dispose of reagents, waste, samples, consumables, etc.
according to government regulations.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a doctor.
Improper installation may damage the analyzer. Install the analyzer strictly as
instructed by the installation guide and inspect the analyzer carefully after the
installation.
IV
Table of Contents
1 Preparations ................................................................................................................................. 3
1.1 Purpose .......................................................................................................................... 3
1.2 Tools ............................................................................................................................... 3
1.3 Accessories ................................................................................................................... 3
3 Unpacking ..................................................................................................................................... 7
3.1 Checking Before Unpacking ....................................................................................... 7
3.2 Unpacking the Main Unit ............................................................................................. 7
3.3 Checking Packing List ................................................................................................. 9
5 Connections ................................................................................................................................ 11
7 Start-up 20
7.1 Initial start-up ............................................................................................................... 20
7.2 Inspection after Startup ............................................................................................. 22
1
10 Calibration and Performance Test ......................................................................................... 31
10.1 Background Validation ............................................................................................... 31
10.2 Optical System Status Validation ............................................................................. 31
10.3 Validation and Calibration of Optical System Gain ................................................ 33
10.4 Calibration.................................................................................................................... 34
2
1 Preparations
1.1 Purpose
Following this installation guide ensures proper installation of the analyzer. This installation
guide is intended to ensure safe and effective operation of the analyzer at the client's.
1.2 Tools
1. Blade or clipper
2. USB
3. Pipette (200ul)
4. Cross-headed screwdriver
1.3 Accessories
1. Plastic test tube
2. Glass pipette
4. Gloves
5. Tissues
6. Calibrator
7. Control (normal)
Do prepare the 7μm standard particles and controls firstly to proceed with the later
confirmation of the optical gain.
3
2Installation Requirements
Operating
Storage Environment Running Environment
Environment
Requirements Requirements
Requirements
Ambient
15℃~30℃ -10℃~40℃ 10℃~40℃
Temperature
Atmospheric
70kPa~106kPa 50kPa~106kPa 70kPa~106kPa
pressure
c. Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close;
e. Do not place the analyzer in direct sunlight or in front of a source of heat or drafts;
Do not move the analyzer. Contact Mindray or your local distributor if
necessary.
4
If the ambient temperature is out of the specified operating range, the
analyzer will alarm you for abnormal ambient temperature and the analysis
results may be unreliable.
c. enough room on or below the countertop to accommodate the diluent, lyse and
waste containers.
Make sure all the requirements listed above are met. Otherwise, the
performance of the analyzer might be compromised.
Input
Voltage Frequency
Power
Before turning on the analyzer, make sure the input voltage meets the
requirements.
5
Use the original electrical wire shipped with the analyzer.
Tube outer
cap) (mm)
Vacutainer
BD 367835 Plastic EDTA-K2 3 13x75
Plus
Vacutainer
BD 367841 Plastic EDTA-K2 2 13x75
Plus
Anticoagulant
Guangzhou evacuated
80715 Glass EDTA-K2 2 12.6×75
Yangpu blood collection
tube
Anticoagulant
Jiangsu 12×75
tube for routine / Plastic EDTA K2 2
Changjian
blood test
Anticoagulant
Jiangsu 13×75
tube for routine / Plastic EDTA K3 3
Changjian
blood test
6
3 Unpacking
Binding
straps
Top
cover
Accesso
ry box
3. Lift the carton to expose the main unit, as shown in Figure 3-3.
7
Carton
4. Remove the protection foam from both sides of the main unit, as shown in Figure 3-4.
Protection
foam
8
Figure 3-5 Lift the main unit
When lifting the main unit, keep it as level as possible and avoid strong
impact. Never lift the autoloader to move the analyzer.
Always lift the left and right side of the main unit to move it, as lifting the
front of the analyzer may damage the front cover.
9
4 Removing the Fixing Components
Open the right side door of the main unit with the slot head screwdriver in the accessory box,
remove the plastic cable tie vertically fixing the sampling assembly and remove the clamp
fixing the synchronous belt. Pull the sampling assebly manually to make sure it can move
smoothly, as shown in Figure 4-1.
Clamp Plastic
cable tie
10
5 Connections
1. See Figure 5-1 for connections of reagents and the waste container. Note that the
M-53LEO(I) lyse, M-53LEO(II) lyse, M-53LH lyse and the analyzer should be placed in a
plane of the same level, and the M-53D diluent and waste containers should be placed under
the countertop.
Remove the plugs of the fluidic interfaces on the back of the analyzer.
Make sure the M-53LEO (I) lyse, M-53LEO (II) lyse and M-53LH lyse are placed
on the same level as the analyzer, and the cap assemblies are alright.
Install a supporting board on the cap assembly of the M-53D diluent container,
as shown in Figure 5-2.
The diluent container must not be placed1.0 meter higher or lower than the
analyzer.
Install the
supporting
board
11
Figure 5-2 Installing the supporting board
2. Connect the tubes according to their colors, and connect the sensor connectors (diluent
and waste sensor). See Figure 5-3, color of the tube and its connector must be the same.
Analyzer
network
port
M-53D
diluent
Power sensor
interface interface
Waste
sensor
interfa
ce
Waste
outlet
M-53D diluent
interface
M-53LH lyse
interface
Install the
supporting board
M-53LEO (II) lyse
interface M-53LEO (I) lyse interface
Figure 5-3Properly connected wires and lines
3. Connect the BC-5390analyzer to PC of the IPU; plug the power cords of the PC and
analyzer to the electrical outlets, as shown in Figure 5-4.
12
Figure 5-4 Connecting the PC and the main unit
13
6 Installation of IPU Software
2. The "User Account Control" dialog box displays, select "Yes" to install the application.
3. The installation program checks system environment, such as operation system version
and database, and then the following dialog box displays.
14
4. If there are modules missing for installation of the IPU, install the modules, and restart the
PC when the prompt message instructs so.
5. After all necessary modules are installed, the language selection screen displays, select
the desired language of the IPU here.
6. Click OK to enter the model selection dialog box to select the desired model of the
"Analyzer IPU Software" (select based on the actual model of your analyzer).
7. Click "Next"to enter the installation path selection dialog box. The installation program is
normally installed in non-system disk. To make sure the "Analyzer IPU Software" can run
15
properly, the residual space of the installation disk shall be more than 100M.
16
6.2 Connection Setup of the IPU and Analyzer
Note:
The default IP address of the analyzer is 10.0.0.7. To modify the IP address of the analyzer,
see section 7.1.
If the PC configured has dual network ports, select the right one when setting up PC of the
analyzer.
2. Click the button "Properties" and select "Internet Protocol Version 4 (TCP/IPv4)" in the
dialog box to set up IP address of the PC, as shown in Figure 6-9.
17
Figure 6-9 Property selection
3. Set the IP address to "10.0.0.1" (the last digit shall be different from that of the analyzer IP),
and the subnet mask to "255.0.0.0", and then click "OK" to exit. The IP address of PC and
analyzer shall be in the same network segment, as shown in Figure 6-10.
18
Figure 6-10 Setting up IP address
19
7 Start-up
An exclamation
mark appears on
the icon.
Figure 7-1 Network connected
2. Double click the icon of IPU on the PC desktop to enter the log-on screen. Enter
user name: service, password: Se s700 (Note: The password contains 7 characters,
there is a space between “e” and “s”.), and then click OK to enter the IPU software.
The analyzer starts to initialize automatically.
3. During the initialization, a dialog box "Skip fluidic initialization?" will pop up. Click
"No" to continue the initialization.See Figure 7-3.
20
Figure 7-5 Communication setup
5. If the analyzer is connected to LIS, before transmitting data, make sure PCs of the
LIS and analyzer are working, so that the IPU program could locate the destination
for transmitting.
6. Click "Setup" - "General Setup" - "Communication" to set up IP address and port
information as required by the LIS.
21
Figure 7-6 LIS communication setup
7. If the connection between PC of the IPU and analyzer is established, the “LIS” icon
in the IPU screen will turn colorized, as shown in Figure 7-7, indicating you can start
transmitting data now.
22
Figure 7-8Menu of Version and Config. Information
The specified version here means the PC end software should be complied with
that of the main unit end. Update the main unit end software to match the PC end
software if necessary.
23
8 Setup and Adjustment
24
3. If the analyzer is equipped with a built-in barcode scanner, set up the barcode
information by clicking "Barcode Info." and the code systems according to the
barcode digits used by the customer. Click "Ok" to save the settings, as shown in
Figure 8-3.
25
Figure 8-4 Maintenance setup
26
Figure 8-4 HGB blank voltage
Note: Make sure "Upload WAVE file" is selected for installation in evaluation or trial use
phase to collect data for analysis of possible errors.
27
9 Testing Other Functions
28
Figure 9-2 Connecting the external barcode scanner
To read barcodes with the barcode scanner, select"Use barcode scanner" in the reagent
setup screen, and point to the scanner to the barcode to read it and then click "Loading".
2. Set up barcode system in the "Setup" - "Barcode Info." screen. Check the barcode
system and digits of the client end. Select the barcode system used and fill in the
number of digits allowed, or select "Unlimited number of digits", as shown in Figure
9-5.
29
Figure 9-5 Setting the barcode information
3. Put the tubes in the tube rack. Click "Mode" on review screen and select the mode
"ALWB-CBC+DIFF” set rack No. and tube No. to 1 and sample ID to "ZDCS-01". For
analyzers configured with built-in barcode scanner, sample ID can be scanned
automatically.
4. Observe the fetching and placing motions of the mechanical arm. It is required that
the motions shall be smooth. Check position of the piercing hole in the tube cap, it is
required that the hole is within Φ4mm of the cap center, as shown in Figure 9-6.
Φ4mm
5. For analyzers configured with built-in barcode scanner, check if the tube rack No.
and sample ID are correct.
6. Select mode "CT-WB-CBC+DIFF", place a tube pasted with high temperature tape to
the tube position, and then start analysis to see if the piercing hole is within Φ4mm of
the cap center.
30
10 Calibration and Performance Test
Note: Before validating performance, please select"Upload WAVE file" as stated in 8.4.
31
Figure 10-1 Preparing 7um standard particles
3. Select "Standard Particle Mode", and click "Calculate". Check if the "FS Gravity Center
Position" and "SS Gravity Center Position" of particle 1 are within the ranges 17.6±0.5
and 105.3±1.0, if not, enter the target (FS 17.6, SS 105.3) of 7um standard particles to
the FS and SS peak targets of particle 1, then the analyzer will calculate the gain
automatically. Click the button "Gain Setup" to save the gain.
4. Run analysis with 7um standard particles in the sample analysis screen again, and
check if the "FS Gravity Center Position" and "SS Gravity Center Position" of particle 1
are within the ranges 17.6±0.5 and 105.3±1.0 in the optical debug screen. Repeat step
2-4 until the they falls in the ranges.
5. If the "FS Gravity Center Position" and "SS Gravity Center Position" of particle 1 meet
the requirements, check if the following requirements are met too.
FS 0.1max width (SD) ≤7.20
32
SS 0.1max width (SD) ≤15.21
And check if the FS and SS gain is within the following range in the screen "Setup→Analyzer
Setup→Gain".
FS gain ≤150
SS gain ≤220
6. If yes, then optical system status validation is completed, and result is Pass. If not, the
flow cell may be dirty, and flow cell maintenance shall be performed. After the
maintenance, repeat step 2-5, and see if the results meet the requirements. If the results
fail to meet the requirement, refer to the service manual for solution. And the analyzer
can be concluded as open-case defective
33
3. Enter the FS and SS gravity center target of the controls to the "Peak target of particle 2",
the analyzer will calculate the gain automatically, and display it in the "Gain" column. Click the
"Gain Setup" button, the current FS and SS gains will be refreshed to the calculated gain.
4. Run control analysis again, and enter the optical debug screen to check as per step 2. If
the values are still not in the target range, repeat step 1-3 until they fall into the target range.
10.4 Calibration
Introduction
Calibration is the process to calibrate the key parameters of the analyzer to a standard
comparator, making the results of the analyzer close to those of the comparator. You can
calibrate the analyzer under fresh blood or calibrator mode.
Procedure
A. Calibration under whole blood mode
1. Click "Menu→Calibration", and then click "Calibrator" to enter the "Calibrator" screen.
Select the calibration mode as "Whole Blood"; enter the lot No. and the expiration date
of the calibrator, and then enter the WBC, RBC, HGB, MCV and PLT targets of the
calibrator into the table. Click "Start". See Figure 10-12.
34
Press the
[Aspirate]
key
NOTE
Be sure to immerse the tip of the sample probe into the calibrator.
If the customer need to use the analysis results generated by the specified analyzer as
references, run the calibrator on the specified analyzer in the whole blood mode
consecutively for 3 times, and use the mean values of the results as targets. Type these
values into the target column at the "Calibration" screen and calibrate.
3. Repeat running the calibrator for 11 times as instructed by procedure 2. Then, the
setup of whole blood calibration factors is completed. When the calibration completes,
the calibration factor will be displayed at the bottom of the table automatically.
35
2. Click "Menu" - "Calibration" to enter the "Calibration Factors" screen. Click the
Mode button and select fresh blood calibration.
3. Click the "Fresh Blood" tab on the top of the "Calibration Factors" screen to enter
the "Fresh Blood" calibration screen.
4. Select the calibration sample ID of the current sample from the "Current sample ID"
pull-down list.
5. Select the parameters to be calibrated by clicking on the check boxes in the first line
of the table.
6. Enter the reference value of the parameter to be calibrated into the edit box "Ref.
value".
36
9. Press the aspirate key to start calibration count.
10. When valid calibration counts of each sample reach n (n≥6), the analyzer will
automatically calculate the Mean, CV% and calibration factor of all calibration data
marked with "√" (calibration data of the first count is not marked with "√", so it is not
included in the calculation).
11. Select other calibration samples from the "Current sample ID" pull-down list, and
perform calibration counts of the samples as instructed in step 8~12 to get their
calibration factors.
12. After getting 3 or more calibration factors of the fresh blood samples, click
"Calculate" to enter the calculate screen of fresh blood calibration. See Figure
10-18.
13. When there are 3 or more calibration factors marked with "√", the CV% of the
calibration factors has been re-calculated as per the calibration factors marked with
"√".
14. When calibration finishes, click "Yes" on the pop-up dialog box to save the new
factor.
37
11 Installation Checklist
38
bend
7. Supporting board Positions of Supporting board □Pass □Fail
installation supporting board is properly
and sensor installed
8. Height of diluent Relative height of No less than1m □Pass □Fail
diluent and the
analyzer
39
10. WBC [90, 125]% □Pass □Fail
11. RBC [90, 125]% □Pass □Fail
12. Calibration factor HGB [90, 125]% □Pass □Fail
13. MCV [95, 105]% □Pass □Fail
14. PLT [90, 125]% □Pass □Fail
15. WBC ≤2.0% □Pass □Fail
16. RBC ≤1.5% □Pass □Fail
17. Reproducibility HGB ≤1.0% □Pass □Fail
18. (whole blood) MCV ≤1.0% □Pass □Fail
19. PLT ≤4.0% □Pass □Fail
20. MPV ≤4.0% □Pass □Fail
21. WBC ≤4.0% □Pass □Fail
22. RBC ≤3.0% □Pass □Fail
23. Reproducibility HGB ≤3.0% □Pass □Fail
24. (predilute) MCV ≤2.0% □Pass □Fail
25. PLT ≤8.0% □Pass □Fail
26. MPV ≤4.0% □Pass □Fail
40
P/N: 046-000352-00 (2.0)