User Manual - Plus
User Manual - Plus
Microplate Spectrophotometer
Operator’s Manual
Part # 0112-0039
Rev. A
Molecular Devices Corporation
SPECTRAmax® PLUS Operator’s Manual
Copyright
Copyright 1997, Molecular Devices Corporation. All rights reserved. No part of this publication may
be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language
or computer language, in any form or by any means, electronic, mechanical, magnetic, optical, chem-
ical, manual, or otherwise, without the prior written permission of Molecular Devices Corporation, 1311
Orleans Drive, Sunnyvale, California, 94089, United States of America.
Patents
The SPECTRAmax PLUS and methods are protected by the following patents:
U.S. Patents 4,968,148 and 5,112,134; Canadian Patent 1,305,334.
Other U.S. and International patents pending.
Trademarks
SPECTRAplate, SPECTRAstrip, PathCheck and Automix are trademarks and SPECTRAmax and
SOFTmax are registered trademarks of Molecular Devices Corporation.
Apple and Macintosh are registered trademarks of Apple Computer, Incorporated.
All other company and product names are trademarks or registered trademarks of their respective
owners.
Disclaimer
Molecular Devices Corporation reserves the right to change its products and services at any time to
incorporate technological developments. This manual is subject to change without notice.
Although this manual has been prepared with every precaution to ensure accuracy, Molecular Devices
Corporation assumes no liability for any errors or omissions, nor for any damages resulting from the
application or use of this information.
Molecular Devices Corporation Instrument Warranty
Molecular Devices Corporation warrants this product against defects in material or workmanship as
follows:
1) All parts of the SPECTRAmax PLUS Microplate Spectrophotometer are warranted for a period of
one (1) year from the original date of delivery.
2) All labor charges to repair the product for a period of one (1) year from the original date of delivery
will be paid by Molecular Devices Corporation.
3) This warranty covers the SPECTRAmax PLUS instrument only, and does not extend to any com-
puter, printer, software, reagents, or disposables used with the instrument.
Glossary of Absorbance, A
Terms The amount of light absorbed by a solution. To measure absorbance accurately, it
is necessary to eliminate light scatter. In the absence of turbidity, absorbance =
optical density.
A = log (I0/I)
I0 = incident light
I = transmitted light
In this manual, we use the terms absorbance and optical density interchangeably.
Optical Density, OD
The amount of light passing through a sample to a detector relative to the total
amount of light available. Optical Density includes absorbance of the sample plus
light scatter from turbidity.
Transmittance, T
The ratio of transmitted light to the incident light.
T = I/I0
%T = 100 T
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
General Overview. . . . . . . . . . . . . . . . . . . . . . . . . 1-3
PathCheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Component Description . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Control Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Control Panel Display . . . . . . . . . . . . . . . . . . . . . 1-7
The Microplate Drawer . . . . . . . . . . . . . . . . . . . . 1-8
The Cuvette Chamber . . . . . . . . . . . . . . . . . . . . . 1-9
The Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Stand-Alone Operation . . . . . . . . . . . . . . . . . . . 1-11
Wavelength Selection . . . . . . . . . . . . . . . . . . . . . 1-11
Temperature Regulation . . . . . . . . . . . . . . . . . . 1-11
Data Collection . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Printed Data Output . . . . . . . . . . . . . . . . . . . . . 1-12
Computer Control . . . . . . . . . . . . . . . . . . . . . . . 1-13
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Chapter 1
An optional printer, for use with the cuvette chamber in a stand-alone mode, can
be purchased as an accessory.
PathCheck™
As predicted by the Beer Lambert law of light absorption, absorbance is propor-
tional to the distance that light travels through the sample—the longer the path-
length, the higher the absorbance.
Microplate readers use a vertical light path so the distance of the light through the
sample depends on the volume. This variable pathlength makes it difficult to per-
form extinction-based assays and confusing to compare results between micro-
plate readers and spectrophotometers.
reference measurements made with the cuvette. SOFTmax® PRO software from
Molecular Devices provides full instrument control and statistical data analysis.
Back Panel
Microplate
Drawer
Control Panel
and opening or closing the microplate drawer. Pressing a control panel key acti-
vates the function.
LCD
A 2-×-20-character liquid crystal display which shows the current instrument set-
tings.
Keys
TEMP
on/off (Incubator) Enables/disables the incubator function.
The left side of the display shows the temperature, both actual and set point, and
whether or not the temperature is at the set point (enunciator blinks if not at set
point). The middle of the display shows the wavelength. The right side of the dis-
play shows the data received from the reading as either absorbance or percent
transmission and indicates whether or not a reference measurement was made
(enunciator blinks if no reference reading was taken).
To change the contrast on the control panel, press [%T/A] and the [Temp] up
or down keys.
A1 Indicator
Power Switch
Wavelength Selection
The display on the control panel shows the currently selected measurement wave-
length (λ). You can change the wavelength by pressing the up or down arrows
until the display shows the desired wavelength. The wavelength will wrap around
to the top or bottom when the last selectable wavelength is reached.
Temperature Regulation
The SPECTRAmax PLUS has been designed to regulate the temperature of the
microplate chamber from 4°C above ambient to 45°C. Upon power up, when
the incubator is off, the temperature in the SPECTRAmax PLUS microplate
chamber is ambient and isothermal. Pressing the incubator [TEMP On/Off] key
will cause the SPECTRAmax to begin warming the microplate/cuvette chamber.
The temperature set point defaults to 37.0°C at start-up.
NOTE: Accuracy of the temperature set point is only guaranteed if the set point
is at least 4°C above ambient. If the temperature set point is lower than
the ambient temperature, the chamber temperature will remain at ambi-
ent. Temperature regulation is controlled by heaters only and, therefore,
cannot cool the temperature to a setting lower than ambient. Addition-
ally, the highest setting (45°C) can be achieved only if the ambient tem-
perature is >20°C.
You can change the temperature set point by pressing the up (▲) or the down
(▼) arrow keys above and below the word TEMP until the desired set point is
shown in the display. Typically, the microplate chamber will reach 37.0°C in 15
to 30 minutes.
The microplate chamber temperature is maintained at the set point until you
press the incubator [TEMP On/Off] key again, turning temperature regulation
off. The microplate drawer will open and the temperature within the chamber
will begin returning to ambient.
NOTE: Should you turn the incubator back on after a momentary shutdown,
allow about ten minutes after reaching temperature for the control algo-
rithm to fully stabilize the microplate chamber temperature.
Temperature regulation and control of the microplate/cuvette chamber are
achieved through electric heaters, a fan, efficient insulation, and temperature sen-
sors. The heaters are located in the chamber which is insulated to maintain the
temperature set point. The sensors, also mounted inside the chamber, measure
the air temperature. The temperature feedback closed-loop control algorithms
compare the measured air temperature inside the chamber against the tempera-
ture set point, and use the difference to calculate the heating cycles. This tech-
nique results in accurate, precise control of the chamber temperature with a
temperature variation of the air inside the chamber of less than 0.3°C. (The tem-
perature uniformity within the microplate or cuvette will depend upon the
design, materials, and/or configuration of that component.)
Data Collection
The SPECTRAmax PLUS stores only the most recent cuvette reading in a buffer
memory.
! CAUTION: Data in the buffer memory is lost when power to the
SPECTRAmax PLUS is turned off. This applies even to short power out-
ages. Do not turn the instrument off while important data remains in the
buffer memory.
Computer Control
The SPECTRAmax PLUS is equipped with an 8-pin DIN RS-232 serial port
through which a computer can communicate with and control the instrument.
SOFTmax PRO®
Molecular Devices’ SOFTmax PRO software is a highly integrated program that
can be used to control and collect data from the SPECTRAmax PLUS. SOFT-
max PRO is easy to use, yet is powerful and flexible, and makes it possible to
access the full capabilities of the SPECTRAmax PLUS.
SOFTmax PRO allows you to:
• Read microplates for endpoint, kinetic, or spectral scans
- Use up to six wavelengths for endpoint and kinetic readings
- Perform Absorbance or %Transmittance readings in the 190- to 1000-nm
range
- Extend kinetic run times up to 99 hours
Specifications Thermal specifications for microplates used in the SPECTRAmax PLUS apply to
flat-bottom microplates with isolated wells. All other microplate specifications
apply to standard 96-well polystyrene flat-bottom microplates. Performance spec-
ifications for cuvette readings apply only to aqueous solutions having solute molal
concentrations less than 0.4 M. When pathlength compensation is applied to
microplate absorbance measurements, agreement with cuvette absorbance mea-
surements for the same solution requires that the solution volume in the micro-
plate well be between 100 µL and 300 µL.
NOTE: Technical specifications are subject to change without notice.
Photometric Performance
Wavelength range 190–1000 nm
Wavelength selection Monochromator tunable in 1-nm increments
Wavelength bandwidth ≤ 2.0 nm FWHM (full width half maximum)
Wavelength accuracy ± 1.0 nm across wavelength range
Wavelength repeatability ± 0.2 nm across all optical channels
Photometric range -0.3 to 4.000 OD
Photometric resolution 0.001 OD
Photometric linearity (405 nm) 0.000 OD to 3.000 OD
Photometric accuracy (microplate) 0–2.0 OD: 190–1000 nm <± 1.0% and ± 0.006 OD
2.0–2.5 OD: 190–1000 nm <± 3.0% and ± 0.006 OD
Photometric accuracy (cuvette) 0–2.0 OD: 190–1000 nm <± 1.0% and ± 0.005 OD
2.0–2.5 OD: 190–1000 nm <± 3.0% and ± 0.005 OD
Photometric precision (repeatability) 0–2.0 OD: 190–1000 nm <± 1.0% and ± 0.003 OD
2.0–2.5 OD: 190–1000 nm <± 3.0% and ± 0.003 OD
Stray light ≤ 0.05% at 230 nm
Photometric stabilization Instantaneous
Photometric drift None—continuous referencing of monochromatic output
Calibration Automatic before first kinetic read and before every endpoint
reading
Optical alignment None required
Light source Xenon flash lamp (5 watts)
Average lamp lifetime 1 billion flashes
Illumination Top down (microplates); horizontal (cuvettes)
Photodetectors Silicon photodiode
Measurement Time
Microplate
Read time (endpoint) • 96 wells in 12 seconds (single wavelength)
• 96 wells in 12* N seconds (N wavelengths)
Kinetic read intervals • 96 wells, 9-second minimum interval between readings
(single wavelength)
• 1 column, 2-second minimum interval between readings
(single wavelength)
Cuvette
Read time (endpoint) • 1 second (single wavelength)
Kinetic read intervals • 2-second minimum interval between readings
(single wavelength)
Scan Speed
Cuvette: Normal scan 60*K nm/min (K = wavelength interval)
Cuvette: Speed scan 200*K nm/min
Microplate: Normal scan 33*K nm/min (8-well strip)
Microplate: Speed scan 135*K nm/min (8-well strip)
Temperature Regulation
Reading chamber Isothermal when temperature regulation is not enabled
Range 4°C above ambient to 45°C when temperature regulation en-
abled. The ambient temperature must be >20°C to achieve
temperature regulation at 45°C.
Resolution ± 0.1°C
Accuracy ± 1.0°C for microplate and cuvette chamber
Well-to-well uniformity at equilibrium ± 0.5°C at 37°C
Chamber warm-up time 15-30 minutes (measured on air) after initiation of tempera-
ture regulation
Temperature regulation 4 sensors
Variation < 0.3°C (regulated)
Temperature regulation diagnostics Temperature regulation system is continuously monitored
and updated
Evaporation Plate lid required to minimize evaporative cooling
Recommended microplate Flat-bottom microplates with isolated wells and lid
Compatibility
Microplate Standard and half-area 96-well flat-bottomed microplates
(0.3 mL). Polystyrene plates for wavelengths above 340 nm;
UV transparent plates above 220 nm; quartz plates above 190
nm.
Cuvette Standard height (45 mm) status cells with 10 mm pathlength
(12.5 mm x 12.5 mm outside) with minimum inside width
of 4 mm (typical for 3 mL volume cells).
Test Tubes 12 x 75 mm test tubes can be used in the cuvette chamber
with the test tube cover.
General Instrument
Display 2-×-20-character backlit LCD
Operating panel 9-key membrane keypad
Self-diagnosis Continuous on-board diagnostics
Spill control Drawer mechanism/reading chamber assembly is protected
from accidental spillage by drainage ports
Computer interface 8-pin DIN RS-232 serial (double shielding required)
Printer interface Parallel 25-pin to Centronics (double shielding required)
Microplates supported All 96-well and strip-well microplates, including lids
Environmental
Operating temperature 15 to 40°C
Operating humidity 0 to 85%, non-condensing
Storage temperature -20 to 65°C
Physical
Size (h × w × d) 8.6 in. (22 cm) × 22.8 in. (58 cm) × 15 in. (38 cm)
Weight 30 lb (13.6 kg)
Power consumption < 250 watts
Line voltage 90-250 VAC, auto-ranging
Line frequency 50/60 Hz
Installation 1) Always make sure the power switch on the instrument is in the OFF position
Warnings and remove the power cord from the back of the instrument prior to any instal-
lation or relocation of the instrument.
2) Do not operate the instrument in an environment where potentially damag-
ing liquids or gases are present.
Installation 1) Do not touch or loosen any screws or parts other than those specifically desig-
Cautions nated in the instructions. Doing so might cause misalignment and will void the
instrument warranty.
Unpacking The SPECTRAmax PLUS is packed in a specially designed carton. Please retain
the carton and the packing materials. If the unit should need to be returned for
repair, you must use the original packing materials and carton for shipping. If the
carton has been damaged in transit, it is particularly important that you retain it
for inspection by the carrier in case there has also been damage to the instrument.
WARNING: The SPECTRAmax PLUS weighs approximately 30 lb
(13.6 kg) and should be lifted with care. It is recommended that two per-
sons lift the instrument together, taking the proper precautions to avoid
injury.
After examining the carton, place it on a flat surface in the upright position.
Open the top of the box and lift the SPECTRAmax PLUS, along with the pack-
ing materials around the ends, up and out of the shipping box. Remove the pack-
ing material from both ends of the instrument and set the instrument down
carefully.
Setting Up 1) Place the SPECTRAmax PLUS on a level surface, away from direct sunlight,
for Stand- dust, drafts, vibration, and moisture.
Alone Use
2) Turn the instrument around so that the back of the instrument is facing you
as shown in Figure 2.1.
Power Switch
This chapter contains operating information for the SPECTRAmax PLUS Microplate/
Cuvette Spectrophotometer. If you are an experienced user of this instrument, you can
turn to the Operation Overview on page 3-4 for a quick review of the operating steps.
NOTE: Only temperature selection and single-wavelength cuvette readings are
available in stand-alone mode through front panel control. To achieve
full use of this instrument, connect it to a computer and use SOFTmax
PRO software.
Read the After setting the temperature (if desired), and choosing the wavelength for the
Cuvette reading, insert the cuvette into the chamber, making sure that the clear sides are
to the left and right (facing the instrument). Do not touch the clear surfaces of
the cuvette. Make sure the cuvette is completely seated in the chamber and close
the cuvette door. If the cuvette contains a blank (typically this solvent contains
everything that the samples contain except for analyte), press the [REF] key to
acquire the reference reading from the cuvette. The SPECTRAmax PLUS will
automatically calibrate in less than two seconds, send a header for the reading to
the printer, close the microplate drawer (if it was open), and read the cuvette
according to the selected instrument settings. If the cuvette contains a sample,
touch the [READ CUVETTE] key to acquire the sample reading from the
cuvette. When reading is complete, remove the cuvette.
Operation The following steps provide a quick reminder of the basic operating procedures
Overview required to read a cuvette using the SPECTRAmax PLUS.
1) Turn on the power switch of the SPECTRAmax PLUS (located on the back
panel).
2) If you wish to regulate the temperature inside the microplate reading chamber,
touch the [TEMP On/Off] key to bring the chamber to the default temperature
of 37.0°C. The microplate drawer will close and the indicator on the right of the
LCD will flash until the set temperature is reached.
3) If the incubator is on, the LCD will show the current temperature along with the
temperature set point. To change the set point (to any setting from ambient +4°
to 45°C), press either the up or down arrow above or below TEMP.
4) Select the desired measurement wavelength by pressing the up or down arrow
near λ. Scroll up or down through the list of wavelengths shown in the LCD
using the up or down arrows until the desired measurement wavelength is high-
lighted.
5) Load the prepared cuvette into the chamber, being sure that the clear sides are left
and right (when facing the instrument). Press the [REF] or [READ CUVETTE]
key.
WARNING: Always turn the instrument off and remove the power cord and
any computer/printer cables from the back prior to any maintenance or
installation operation
WARNING: Never perform any operation on the instrument in an environ-
ment where liquids or potentially damaging gases are present.
WARNING: Risk of electrical shock. Refer servicing to qualified personnel.
! CAUTION: Use of organic solvents (such as dichloromethane) may cause
harm to the optics in the SPECTRAmax PLUS. Extreme caution is
advised when using organic solvents. Always use a plate lid and avoid
placing a plate containing these materials in the reading chamber for pro-
longed periods of time. Damage caused by the use of incompatible or
aggressive solvents is NOT covered by the instrument warranty.
! CAUTION: Never touch any of the optic mirrors, filters, or cables or their
housing, or manifold. The optics are extremely delicate, and critical toe
of the instrument.
! CAUTION: Do not touch or loosen any screws or parts other than those spe-
cifically designated in the instructions. Doing so could cause misalign-
ment and possibly void warranty.
General Keep the drawer closed when the instrument is not in use. The drawer can be
opened by pressing the DRAWER button. Always close the drawer immediately
prior to switching the instrument off.
Cleaning BIOHAZARD: Wear gloves during any cleaning procedure that could
involve contact with either hazardous or biohazardous materials or flu-
ids.
Periodically, you should clean the outside surfaces of the SPECTRAmax PLUS
using a cloth or sponge that has been dampened with water. Do not use abrasive
cleaners. If required, clean the surfaces using a mild soap solution diluted with
water or a glass cleaner and then wipe with a damp cloth or sponge to remove any
residue. Do not spray cleaner onto the instrument.
If needed, clean the microplate drawer or cuvette chamber using a cloth or
sponge that has been dampened with water.
Should fluids spill in the drawer area (when the drawer is out) or into the cuvette
chamber (when the door is open), they will be directed to a tray at the bottom of
the instrument from which they will exit to the bench or counter beneath the
instrument. Wipe up any spills immediately. Clean only the exterior of the unit
(and the microplate drawer or inside the cuvette chamber if necessary). Never
clean the inside of the instrument. Do not allow excess water or other fluids to
drip inside the instrument.
Cleaning the The fan filter on the bottom of the instrument requires periodic cleaning. The
Fan Filter frequency of the cleaning depends on how dusty your particular lab is and could
range from once a month to once every six months.
1) Turn power to the instrument OFF and then remove the power cord and
cables from the back of the instrument.
2) Make sure no plate or cuvette is in the instrument. Tape the cuvette door
closed. Turn the instrument over so that it rests flat on the bench.
3) Pop the black fan cover off the remove the filter.
4) Clean the filter by blowing clean, canned air through it or by rinsing it—first
with water and then with alcohol—and allowing it to dry completely.
5) Place the clean, dry filter over the fan and replace the black cover.
6) Turn the instrument back over. Remove the tape from the cuvette door.
Reconnect the power cord and cables to the instrument.
Changing the Fuses burn out occasionally and must be replaced. If the instrument does not
Fuses seem to be getting power after switching it on (the LCD shows no display), first
check to see whether the power cord is securely plugged in to a functioning power
outlet and to the receptacle at the rear of the SPECTRAmax PLUS. If power
failed while the SPECTRAmax PLUS was already on, check that the power cord
is not loose or disconnected and that power to the power outlet is functioning
properly. If these checks fail to remedy the loss of power, follow the steps listed
below to replace the fuses. Spare fuses (two U.S. and two metric) are shipped with
the instrument in the original carton. The U.S. and metric fuses are identical
except for physical size. They may be taped to the back of the SPECTRAmax
PLUS.
If you no longer have spare fuses, you may obtain new ones from Molecular
Devices (part numbers: 4601-0013, U.S., 4601-0014, metric) or from a local
hardware store. Make sure fuses are rated SLOWBLOW (U.S.: 4-amp time-
delay; Metric: 4-amp, 5 × 20 mm, time-delay).
1) Switch power to the instrument off and then remove the power cord from the
outlet and from the SPECTRAmax PLUS power cord receptacle.
2) Remove the printer cable and computer cable (if connected) from the back of
the SPECTRAmax PLUS.
3) Turn the instrument around for easy access to the rear panel.
4) On the left-hand side of the rear panel (viewed from the back) is the power
cord receptacle. As shown in Figure 4.1, insert a small, flat-blade screwdriver
into the slot behind the tongue at the right of the black plastic cover. Gently
pry the cover open. It will begin to slide forward.
Figure 4.3: The Fuse Box and Holder with Fuses Removed
8) Insert new SLOWBLOW-rated fuses into the fuse holder. Either end of the
fuse may be forward.
9) Insert the fuse holder into the fuse box, making sure that the fuses face
toward the right (toward the tongue on the cover) as you insert it. Slide the
fuse holder all the way into the box.
10) Insert the fuse box into the opening in the instrument, making sure that the
fuses are on the right side (toward the power receptacle). Press the fuse box
into place, making sure the cover snaps closed.
11) Reconnect the power cord to the instrument and to the wall outlet and
reconnect other cables previously disconnected.
Error Codes If a problem occurs during operation that causes an unrecoverable error, the
and Probable instrument will stop and an error code number will be shown in the display on
the front panel. To correct the problem, call your local Molecular Devices repre-
Causes sentative for assistance.
Error Messages
The LCD will display Fatal Error codes when a situation arises that requires
attention. Any reading in progress will stop. Warning messages do not stop a read-
ing but are logged in the error buffer. Warning messages indicate a situation that
requires attention but is not sufficient to stop or prevent a reading. Examples of
situations that might cause warning messages are low memory, entries being out
of range, or operations that could result in loss of data. These messages are gener-
ally self explanatory. For assistance regarding warning messages, contact your
local Molecular Devices representative.
Error Code Classifications
Not all error messages are listed in this manual. The errors are grouped in rela-
tionship to possible causes as follows:
100 -199 Errors possibly caused by unrecognized commands being sent from the computer
to the instrument.
200-299 Errors probably due to a main board failure or a error in the firmware code. Most
of these errors require the assistance of Technical Support.
300-399 Instrument errors due to either a main board failure or other system failure. Most
of these errors require the assistance of Technical Support.
400-499 Errors due to failure or improper initialization of the instruments non-volatile
memory (NVRAM). All of these errors require the assistance of Technical Sup-
port.
Some errors are considered fatal in that if they are detected during power up (the
instrument will abort the power up sequence and display “FATAL ERROR ###”
on the LCD panel). Check the following list to see if there is something that you
can do to change the condition of the instrument to prevent the fatal error (e.g.,
closing the cuvette door during the power up sequence will prevent error num-
bers 111, 219, 302, and 310). Leave the instrument on for about five minutes,
turn it off and then back on. If you continue to get the fatal error message on
power up, record the error message number and contact Molecular Devices Tech-
nical Support or your local representative for assistance.
NOTE: If the instrument is functioning normally when using SOFTmax PRO,
no errors should be in the buffer (except error number 100).
Table 4.1: Error Codes, Messages, and Notes about the Errors
Error
Error Message Notes
Code
100 command not found Command string was not recognized by the instrument.
101 invalid argument Command Argument does not match specified parameters or
syntax.
102 invalid data Command argument is out of acceptable range.
103 too many arguments Too many arguments after a specific command.
104 not enough arguments Not enough arguments for a given command.
105 input line too long Too many characters in the input line.
106 command error, system busy Instrument could not perform the given command because it was
busy doing another task. Example: Request a wavelength while
the monochromator is in motion.
107 command error, measure- Instrument could not perform the command because a measure-
ment in progress ment was in progress.
108 command error, no data A request to the instrument to transfer data when there’s no data
available in the buffer.
109 data buffer full To many data sets in the buffer. Can be caused by setting up a
long kinetic run and disconnecting the computer or when Soft-
Max Pro is pre-empted by another application.
110 error buffer full More than 65 errors in the buffer.
111 stray light cuvette, door open Cuvette door open while doing a read. Note error’s # 219 and
302 and 310 can also be caused by this.
205 user generated bus error Result of improper use of certain commands.
209 receive buffer overflow Caused by an external device sending too much data over the
serial port and ignoring flow control.
210 serial port parity error Parity bit error detected with incoming serial data.
211 serial port overrun error Caused by host computer sending too much data and ignoring
the flow control signal.
219 Cuvette gain check fail Power up check of cuvette’s amplifier’s gain is out of tolerance.
Could be due to bad or disconnected pre-amp board or excessive
stray light or cuvette door being open on power up.
Table 4.1: Error Codes, Messages, and Notes about the Errors
Error
Error Message Notes
Code
300 temp sensor faulty Unable to read a reasonable thermistor value. Thermistor is
faulty or disconnected, Main board problem, or ambient temper-
ature is out of range.
301 safe temperature limit A temperature of over 50 deg C was detected on one or more of
exceeded the 4 thermistors. Temperature will be shut off and remain off
until power up reset is completed successfully.
302 low light measurement Not enough light detected to make an accurate measurement. If
reading a cuvette, make sure the cuvette door is closed.
310 Signal level saturation During a cuvette read, make sure the cuvette door is closed. Dur-
ing a plate read, make sure nothing is blocking the door from
shutting completely.
317 LCD time-out LCD front panel is bad or disconnected.
400 - Errors in Non-Volatile Mem- Contact Molecular Devices Technical Support.
499 ory
Opening the If an error occurs while the drawer is closed and you need to open the drawer,
Drawer press the [Drawer] key. If the drawer does not open, turn power to the instru-
ment off and then on again.
Manually
If the drawer still remains closed, turn the power off and using your thumb nail,
locate the groove in the upper left side wall of the door. Open the door, and with
your index finger, pull the microplate drawer out of the instrument (do not force
the drawer) and remove the microplate. This action will not harm the instrument,
but should only be taken if the first two options have failed to open the drawer.
If you are still unable to open the drawer, contact Molecular Devices Technical
Support or your local representative for assistance.
Compatible The SPECTRAmax PLUS can be used with an Epson-compatible strip printer.
Printers Compatibility may require changing the dip switch or pin settings on the printer
or insertion of an emulation cartridge. Please check your printer’s manual to
assess compatibility based on this criterion—some printers will not meet this
requirement.
Please contact your local Molecular Devices representative for information
regarding recommended printers.
Cables Molecular Devices recommends that you use high-quality, double-shielded cables
to connect the SPECTRAmax PLUS to other peripheral instruments, such as a
printer or computer. Choose cables that meet the following requirements:
Printer Cable (for Stand-alone operation only):
Centronics parallel, male 25D to Centronics.
Serial Interface Cable
(contact Molecular Devices for specific pin-out requirements)
Macintosh: Male DB8 to male DB8
IBM Compatible: Male DB8 to Female DB9 (custom cable made by
Molecular Devices)
Items
Available for
Use with the
SPECTRA-
max PLUS
Part Number
Cable, RS-232, Macintosh to SPECTRAmax PLUS,
8-pin DIN to 8-pin DIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9000-0091
Cable, RS-232, PC to SPECTRAmax PLUS,
9-pin to 8-pin DIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9000-0149
Cable, parallel printer, 25-pin to Centronics . . . . . . . . . . . . . . . . . . . 9000-0003
Cuvette Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9000-0161
Fuse, 4-amp Time Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4601-0013
Fuse, 4-amp (5 × 20 mm) Time Delay. . . . . . . . . . . . . . . . . . . . . . . . 4601-0014
Power Cord (US, Canada, Japan, Mexico, India) . . . . . . . . . . . . . . . . 4400-0002
Power Cord (Germany, France, Scandinavia, Italy, Korea) . . . . . . . . . 4400-0036
Power Cord (UK, Indonesia, Singapore, Malaysia). . . . . . . . . . . . . . . 4400-0037
Power Cord (Australia, Hong Kong, China) . . . . . . . . . . . . . . . . . . . 4400-0038
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9000-0189
SPECTRAplate—Disposable
UV-transparent disposable microplates (case of 50) . . . . . . . . . . . . . . R9013
SPECTRAplate—Quartz
UV-transparent quartz microplate . . . . . . . . . . . . . . . . . . . . . . . . . . . R8024
SPECTRAmax Mouse Pad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9000-0133
SPECTRAtest PLUS Validation Package . . . . . . . . . . . . . . . . . . . . . . 0200-2405
Test tube cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2300-0277
A I
Accessories B-1 Incubator 1-8, 1-11, 3-3
Installation 2-3
B Cautions and Warnings 2-3
Back Panel 1-10
Bandwidth 1-16 K
Keys (control panel) 1-5, 1-6, 3-4
C
Cables 2-4, A-3 L
connecting 2-4 Labels 1-10
serial 1-10 LCD (liquid crystal display) 1-6, 3-4, 4-7
Calibration 1-16, 1-17
Cleaning 4-4 M
Component description 1-5 Maintenance 4-1, 4-4
Computer control 1-13 Microplate 1-3
Connections 1-5, 1-10 96-well 1-8
cables 2-4, A-3 filter-bottom 1-8
printer 1-10, 2-4 strip wells 1-8
Control panel 1-5, 1-7 temperature uniformity 1-8
Cuvette 1-5, 1-9 Microplate drawer 1-5, 1-7, 1-8
closing 1-8
D incubator status 1-8
Data open manually 4-9
collection 1-12 Microplates 1-8
printed output 1-12
Drawer 1-5, 1-7, 1-8 P
closing 1-8 Pathlength correction 1-4
incubator status 1-8 Ports 2-4
open manually 4-9 printer 1-10
Drawer key 1-7, 1-8, 1-11 serial 1-10
Power cord 1-10, 2-4, B-3
E Power switch 1-5, 1-10, 3-3
Error messages 4-7 Print 1-12
Printed Data Output 1-12
F Printer 1-4, 1-12, 1-13, 2-4, 3-3, A-1
Fuses Printer port 1-10, 2-4
accessing 1-10
changing 4-4 R
Read mode 1-13, 3-4
Read the microplate 1-13, 1-14, 3-4
preparation 3-3
Reference reading 1-7, 3-4
Remote control 1-13
S
Serial port 1-10
Set point, temperature 1-6, 1-7, 1-12, 3-3
SOFTmax PRO (software) 1-3, 1-5, 1-13
Specifications 1-16
Spills 4-4
Stand-alone operation 1-11, 2-1, 2-4, 3-1
T
Technical service 4-3
Temperature 1-11, 1-12, 3-4
equilibration 1-12, 3-3
incubator 1-12, 3-3
regulation 1-11
set point 1-6, 1-7, 1-12, 3-3
setting 3-3
stabilization 1-12
U
Unpacking 2-3
W
Warranty ii
Wavelength 1-6, 3-3, 3-4
range 1-16
selection 1-11