BFT Ii Im 31 en
BFT Ii Im 31 en
Instruction Manual
We reserve the right to make changes in the course of technical development without previous notice.
© 2008 Siemens Healthcare Diagnostics Products GmbH (formerly Dade Behring). All rights reserved.
Emil-von-Behring-Str. 76
35041 Marburg
Germany
Neither this manual nor any parts of it may be duplicated or transmitted in any way without the written approval
of Siemens Healthcare Diagnostics Products GmbH.
SMN 10465104
Table of Contents
1 Introduction 1-1
2 Operation 2-1
1
not available in the US
4 Errors 4-1
1
not available in the US
5 Software 5-1
6 Appendix 6-1
6.7 Meanings of the symbols on the adhesive labels of the accessories 6-8
Index
1
not available in the US
1
not available in the US
1 Introduction
The measuring channels are addressed with the following keys: Incub for
the sample incubation timer, Reset for the adjustment and Start for a
manual start of measurements.
Since these keys are specific to the appropriate measuring channel, they
are placed adjacent to the measuring channel in the form of a membrane
keypad.
The arrow keys <- -> allow the operator to select the next menu from the
menu field. The numbers are used to enter method parameters. A selection
is confirmed by pressing Enter; a submenu will be displayed if appropriate.
The Esc-key cancels any operation in the parameter menu. The method
parameter menu can be accessed via the Cal-key.
By adding a reagent with a pipette into the sample cuvette the measure-
ment is automatically started. The light protection cap is equipped with a
guide, designed specially for Eppendorf pipette tips, for correct positioning
of the pipette.
The light protection cap has a red tongue that allows the pipette tip to be
introduced into the cuvette once the cuvette has been inserted. This tongue
prevents anything from being pipetted directly into the measuring channel
prior to a cuvette being inserted.
A power jack, is located on the back of the instrument. The power switch is
located next to the power supply
- I switch ON
- O switch OFF
together with the fuse holder in a housing. For data output a RS 232 C 6-
pin interface port is located on the back of the instrument.
1 2 3 Cal
< >
4 5 6 Incubation block 37 °C: The incubation block is
Esc
7 8 9 0
Enter
temperature-controlled by the software after the instru-
ment has been switched on.
- 3 x 10 positions for cuvettes
T
- 4 positions for reagent bottles, left position with
magnetic stir function
NOTE
The light protection caps have been specially constructed for Eppendorf
pipette tips. Ordering information for pipette tips that have been approved
for use on the BFT* II can be found in the current BFT* II Reference
Guide.
Made in Germany
Key Description
1 2 3 Cal
< >
4 5 6
Esc Enter
7 8 9 0
Cal - Calibration
Menu selection, Instrument setting and method parameter,
function to save data that has been entered.
1 Incub Incub
Incub. = Incubation:
Incub. - Start of sample incubation timer with respect to measuring
channel.
1.3 Installation
Remove BFT* II from its packaging and verify that the accessories kit is
complete. Please notify Dade Behring Service immediately in the event that 1
the shipment was incomplete. Refer to Chapter 6.4 for more details.
• Connect the power cable between the BFT* II and a power supply free
from interferences by large power users such as elevators and
centrifuges.
NOTE
Outside the US the BFT* II are delivered without main cables. The main
cables used must meet the local regulations (e.g. VDE, CSA-C22.2, No
21 and No 49). The cable has to be designed with NYLHY. The recom-
mended length is 1.5 m, and the minimum cross-section is 3 x 0.75 mm2.
A cold device plug is used at the BFT* II end.
Switch ON/OFF
• Turn on the BFT* II by switching the power supply switch to position I.
Refer to Figure 5.
• Use only original cuvettes and stir bars which will assure proper operation
of the instrument. Stir bars are mandatory for correct measurements.
Fuse housing
Main switch
230 V
The period from adding the starting reagent until clot formation is observed.
It then can be converted into the appropriate units (%, INR, mg/dl, g/l).
Once the start reagent has been added the measuring channel is
repeatedly adjusted, that is the lamp intensity automatically adjusts up or
down depending on the turbidity of the test sample. In this process the tur-
bidity of the sample plasma and the reagent are adjusted for.
A stir bar is located in each cuvette. During the measuring process the stir
bar provides homogeneity of the reagent-plasma medium. At the same time
a small whirl emerges through the stir bar movement which assures that
even the smallest fibrin clot is formed in front of the photodetector.
This stirring action combined with the optical measurement constitute the
basic features of the patented "turbodensitometric measuring principle".
test cuvette
Lamp Detector
permanent magnet
stirrer motor
ELECTRONICS
DISPLAY
1.5 Reagents
Various assays to determine PT, APTT, many factors1, etc. are available
from Dade Behring Service. 1
Informations about the Application Sheets of the reagents can be found in
the current BFT* II Reference Guide.
Contamination
When different reagents are applied and especially reagents containing
thrombin, the danger exists that reagents are spread.
When the reagent is added, the light protection cap will be moistened with
reagent and therefore constitutes a point of contamination.
CAUTION After each method change, this contamination point must be cleaned
with an appropriate thrombin inactivator (Washing Solution
for Coagulation Analyzers) and a cotton bud.
NOTE
- Use a pipette that is regularly checked.
- Use a tip that properly fits onto the pipette and has been approved for
use on the BFT* II.
- Prior to the use of reagents, read the Application Sheets and follow the
instructions.
- Only use cuvettes and stir bars from the original manufacturer. These
cuvettes are subject to strict quality control. Cuvettes and mixers are
for single use only!
2 Operation
User qualification
CAUTION It is absolutely necessary that only skilled personnel will access the
parameter menu with a secret number as improper handling of the
BFT* II might cause inaccurate measuring results. (see chapter 3.2.4
<secret number>)
NOTE
reading parameters
from internal
Should the message ERROR appear, you must contact Dade Behring
Service to have the fatal error corrected.
Printerpaper Esc: NO
inserted? Enter:Yes
If YES,
• press the Enter-key, the paper will advance 5 lines and the printer will
stay on.
If NO, press the Esc-key, the printer will be off.
Change Paper
To change printer paper refer to chapter 2.2 Change Paper.
The last 15 min. of the warm up phase will be supported with a timer count
down. After count down the BFT* II is ready for operation.
NOTE
Use the warm-up phase to load the BFT* II cuvettes and reagents for
the tests that will be performed. Each cuvette must contain a stir bar.
Follow the instructions by the reagent manufacturer. Verify the method
parameters with those that have been saved to the BFT* II.
For your own safety follow the instructions with respect to hygiene.
Once the operating temperature has been reached, the following request
will appear:
auto blanking
keep channels clear
2
The STANDBY mode will be displayed automatically.
The method selected last prior to restarting the analyzer, such as PT, is
displayed.
If you want to start with analytics, press the Enter-key to change to the
measurement mode (see chapter 2.1.3 Measurement).
Printer
If the printer is set to ON in the menu UTILITIES the current parameter
status will be printed out after the first measurement has been performed
and the Start-key has been pressed for conversion of the result.
The results from the left channel will be printed left justified and the results
from the right channel will be printed right justified.
Press the Enter-key for line feed. For more information and change of
paper refer to chapter 2.2 Measuring.
The temperature of the heating block and the selected method are
displayed.
incubat. incubat.
2 cuv in cuv in
2.1.3 Measuring
Both measuring channels can only operate in one selected method as for
example PT. It is not possible to select a method for a specific measuring
channel.
Cuvettes and reagents can be placed in the incubation block for one spe-
cific measuring channel.
The function keys Incub, Reset, and Start which are used to address the
measuring channels, are located to the side of the measuring channel.
Single determination
For all defined tests you can do single determination tests. A mean value
calculation from two single values is not supported by the software.
Sample incubation
• Pipette a plasma sample without air bubbles in a cuvette that has been
prewarmed in the Incubation block to 37 °C.
• Press the Incub-key; the incubation time as selected in the setup will run
and begins counting down. At 5 sec. remaining incubation time a signal is
heard.
NOTE
For methods you can enter a second sample incubation phase. See
chapter 3.1.5.1. By the display of incubat 1 and incubat 2 the incubation
phases will be shown.
5 s incubat.
cuv in cuv in
• Once the incubation has been completed a long beep can be heard, then
place the cuvette in the appropriate measuring channel. Please make
sure that knob in cuvette fits into notch in measuring channel. Close light
NOTE
If the cuvette has not been inserted into the measuring channel and
liquid is inadvertently pipetted on the red tongue of the light protection
cap, the cap should be removed by unscrewing the two front screws
2
from the thermoblock.
To clean the light protection cap, move the slider of the cap back and
forth to allow the liquid collected for protection to drain from the red
tongue. Then, disinfect the light protection cap, rinse it with distilled
water and let it dry.
After cleaning, fasten the light protection cap back on the thermoblock
by tightening both screws evenly. Ensure that the light protection caps
are aligned correctly.
CAUTION When cleaning the light protection cap, do not push the red tongue too
far back inside the cap, otherwise, the spring with the red tongue may
come off.
cuvette.
Due to the change in turbidity the measuring process will be started auto-
matically.
2 As soon as the clot formation has been recognized the measuring timer
displays the measured clotting time.
incubat. incubat.
t= 12.4 s t= 11.8 s
% %
93.0 101.1
The results will be printed out automatically. Measuring time, errors and all
conversions will be printed out. An identification can be entered manually
under Patient: _______
Sequential numbers will be assigned in the order of print-outs and not in the
order of started measurements.
INR INR
1.05 0.98
time time
12.4 s 11.8 s
• Press the Reset-key to clear the memory for the next measurement.
incubat. incubat.
cuv out cuv out
You can select a method from the method list by pressing the:
<1 PT >
<2 APTT >
<3 Thrombin Time1 >
<4 Batroxobin Time1 >
<5 Fib. g/l >
<6 Fib. mg/dl >
<7 Extr. Factor1 >
<8 Intr. Factor1 >
<9 PT-Innovin >
< 10 ProC G PCAT1 >
< 11 ProC G PCAT/01 >
< 12 ProC FV PCAT1 >
< 13 ProC FV PCAT/01 >
< 14 none >
< 15 none >
< UTILITIES >
writing parameters
to internal
The newly selected method has been initialized. Incubation for the first
measurements can be performed.
incubat. incubat.
cuv in cuv in
<none>
The method memory positions 14 and 15 -none- are free positions where
new methods can be entered manually or installed and modified by copying
existing methods.
See chapter 3.1 Method Parameter.
UTILITIES
The menu UTILITIES is not a method but a group of menus. As soon as a
secret number. has been entered, selected global settings can be accessed
more quickly.
incubat incubat
cuv in cuv in
• Remove the old paper tube by slightly bending the brackets to the outside.
• Add the new paper roll. Make sure that the paper roll fits properly onto the
brackets and that the paper can be inserted into the printer paper channel
below.
• Cut the end of the paper with scissors and do not tear! Then guide the
paper into the printer paper channel until you feel some resistance.
3 BFT* II Parametrization
The accessible method parameters are only for the selected method in
STANDBY. 3
The Secret number is necessary to access the method
parameters.
User qualification
Cal
Enter >
3
cuv in cuv in <1 PT >
1) method parameter
< measurement >
Menu <general>
1) method parameter method parameter group
< general> selection
1) method parameter
< 2nd conversion>
Menu <measurement>
CAUTION Parameters in this menu marked with an number symbol "#" may be
changed in accordance with Dade Behring only!
1) method parameter
< measurement>
Save parameters
After finishing parametrization following display appears:
3 1) checking parameters
please wait
You can either select ENTER = yes or ESC= no. Press the Enter-key to
save the new parameters or press the Esc-key to keep the
former parameters. In both cases a higher menu level can be accessed.
1)method parameter
< general >
3
• Press the Enter-key to confirm.
• Select <methodname> in menu general.
1) general
< methodname >
1) methodname:
PT >
NOTE
If you keep the arrow-key pressed the characters will be displayed
faster and you can speed up your selection.
If you keep the Enter-key pressed the cursor will move faster.
1) methodname:
PT .
Delete character
Use the Esc-key to set a blank character. The cursor will go forward one
step.
Follow the instructions below to load a method from the preset to a method
memory position. For example, method <6 Fibr. mg/dl> shall be loaded to
3
method memory position <14 none>
14) general
< load preset >
14) overwrite?
ENTER: Yes,Esc: No
• Press the Enter-key to save the desired method or press the Esc-key to
leave the menu without saving. With overwrite the new parameters will
be saved.
14 general
< copy method>
• Press the Esc-key to reach the upper menu to change parameter under
<1st- or 2nd conversion> or <measurement>.
• When in STANDBY mode select the method memory position (14 none).
• Press the Cal-key to access the parameter menu after having entered the
secret number.
14) overwrite?
ENTER= Yes,Esc= No
• Press the Enter-key to save the desired method or press the Esc-key to
leave the menu without saving. With overwrite the new parameters will
be saved.
1 not available in the US
14 general
<exchange method>
• Press the Enter-key to save parameters or press Esc to leave the menu
without saving.
• When in STANDBY mode select the method memory position <3 Throm-
bin Time>1.
• Press the Cal-key to access the parameter menu after having entered the
secret number.
3) exchange with:
< 3 Thrombin Time >1
• Press the arrow-key several times to select the desired method <9 PT-
Innovin>1.
3) exchange with:
< 9 PT-Innovin >
3) exchange?
ENTER: Yes,Esc: No
• Press the Enter-key to save the desired method or press the Esc-key to
leave the menu without saving.
14 general
3
< delete method>
• Press the Enter-key to save parameters or press Esc to leave the menu
without saving.
• Press the Cal-key to access the parameter menu after having entered the
secret number.
x) delete method?
ENTER: Yes,Esc: No
• Press the Enter-key to delete the method memory position or press the
Esc-key to leave the menu without deleting.
14 general
< methodname>
3
save new parameters?
ENTER=yes,Esc=no
• Press the Enter-key to save parameters or press Esc to leave the menu
without saving.
- press the Cal-key from STANDBY-mode of a method and enter the secret
number.
- press the Cal-key from "incubat/cuv in" directly without entering the secret
number.
Display Meaning
<reference curve> entering a 9-point reference curve
<rising>/<falling> rising or falling reference curve
<unit> unit of reference curve to be edited by the
text generator
<decimal place> number of digits after decimal point (format
xxx.x)
<min/max value> ranges of activity/concentration
<time interpolation> linear - reciprocal- logarithmic
<value interpolation> linear - reciprocal- logarithmic
<quick with factor> reference value and factor of curve gradi-
ent
<none> no conversion
NOTE
- press the Cal-key from STANDBY-mode of a method and enter the secret
number.
- press the Cal-key from "incubat/cuv in" directly without entering the secret
number.
NOTE
If <none> has been selected under first conversion an input field for the
reference value will be displayed automatically for the INR conversion
under <2nd conversion>
CAUTION All parameters marked with an number symbol "#" except the reagent lot
number may be modified only after consultation with Dade Behring.
Any modification will cause different measuring times.
Selection meaning
<mixer> stirrer starting speed - period of reduction until
final speed is reached - stirrer final speed
<min value> minimum measuring value in digits before clotting
<adjust> adjustment time / learning time / lag time
<time interpolat> evaluation reciprocal - linear - logarithmic
<incubation> first and second incubation time
<Start Reag> volume of starting reagent
<methodname> connection with parameters if the method-name
changed
Selection Meaning
3.1.5.1 PT Parameter
Method parameters in the BFT* II are preset by the manufacturer. Before
you start with your analysis the method parameters must be updated for the
respective reagent.
• Press the Cal-key; the system will request an up to 5-digit secret no. (No.
secret no.:
_ _ _ _ _
3
The following display will appear, provided the correct secret number was
entered:
1) method parameter
< general >
<1st conversion>
• Select the 1st conversion
• Press Enter; the following display appears:
1) 1st conversion
< reference curve >
<reference curve>
• Press Enter; the following display appears:
rising/falling
1) reference curve:#
< falling >
% %
time time
rising falling
3 Figure 9 Curve rising / falling
Calibration curves must be entered in such a way that you start with the
highest % values and shortest measuring times in seconds for falling
curves (coagulation factors). For rising curves (inhibitors or INR-Cal) you
have to start with the lowest % / INR values and the longest measuring
times in seconds. The software will support this procedure only.
unit
• Select the characters for the unit and confirm them by pressing the
Enter-key.
decimal place
The display to enter the digits (decimal place:) appears:
1) decimal place:
< format xxx.x >
conversion limits
You can enter the minimum and maximum values for the calibration curve
and conversion now:
1) min/max value:
5.0 /150.0 %
NOTE
Use extrapolated sample results only if accepted by local authorities.
3
An error message appears if the values are outside the ranges. These
parameters show the number of digits after the decimal point and the unit.
calibration points
A 9-point reference curve can be entered with this menu.
Points on the reference curve that are not to be used, must have the entries
0.0 % and 0.0 s. A minimum of 2 points must be defined. We recommened
to define at least 3 points. Before you leave this menu, enter all calibration
points to control or change the settings. All following calibration points are
for example.
1) 1. point:
97.0 % 11.5 s
The field for the next point on the reference curve appears.
1) 2.point:
43.0 % 20.1 s
interpolation
As soon as the last point on the reference curve has been confirmed
the following display will appear:
#
1) time interpolat.
< linear >
For the interpolation of the time axis the following options can be selected:
<linear> <reciprocal> <logarithmic>.
In the following display you can select interpolation modes for the activity/
concentration: <linear><reciprocal><logarithmic>.
1) value interpolat#
< reciprocal >
select:ESC= work
ENTER= more param.
conversion limits
You can enter the minimum and maximum values for the
calibration curve and conversion now:
1) min/max value:
5.0 / 150.0 %
NOTE
An error message will be displayed if the values are outside the range.
These parameters show the number of digits after the decimal point and the
unit as described under < reference curve >.
reference value/factor
The value factor is currently not supported for the BFT* II.
The factor has two digits. Internally the value will be multiplied by 10 (see
6.8 Mathematics).
NOTE
If complete entries have been made for <reference curve> and <quick
with factor>, the type of reference curve selected last will be active for
the conversions in the measuring mode.
<none>
No conversion under 1st conversion.
<2nd conversion>
Following menus can be selected if <reference curve>, <quick with factor>
or <none> was selected under 1st conversion:
Only the last selection under 2nd conversion is active for 2nd conversion.
1) method parameters
< 2nd conversion >
<INR>
• Press Enter; the following display will appear:
1) 2nd conversion
< INR >
3
1) 100% = 11.8 s Example
• Enter the ISI-value for the respective reagent as indicated on the table of
values.
select:ESC= work
ENTER= more param.
<none>
No conversion under 2nd conversion.
<measurement>
1) method parameter
< measurement >
• Enter the volume for the starting reagent, for example 100 µl.
1 not available in the US
3 18 BFT* II Analyzer - Instruction Manual - Version 3.1
3 BFT* II Parametrization
Incubation time
3
1) incubat. (0=off) 0= no incubation
1st= 0 s# 2nd= 60 s# max.: 600 s
• Enter the appropriate 1st. sample incubation time for 1st=xxx s and
2nd=xxx s.
0s 0s no incubation
>0 s 0s incubation
0s >0 s incubation
>0 s >0 s 1st + 2nd incubation
Print-out number
mixer function
As explained the measurement will start with 500 rpm mixer revolutions.
The final revolutions of 250 rpm will be reached after 40 seconds until the
clot has been found. Thus the sample will be blended well and even
unstable clots will be detected.
min value
3 1) min value:
150 digits#
1) adjust: 1 s #
learn: 3 s#,lag: 1 s #
Here you can enter parameters for the optical detection of the sample.
Before modification consult Dade Behring.
select:Esc= work
ENTER= more param.
checking parameters
please wait
• Press the Enter-key to save parameters or press Esc to leave the menu
incubat incubat
civ in cuv in
Continue as described.
3
3.1.5.2 APTT Parameter
Before you start with your analysis, the method parameters must be
updated for the respective reagent.
The preset method on the method memory position <2APTT> has the
same structure as the method described in 3.1.5.1 PT Parameter.
The preset method on the method memory position <7 Extr. Factor>1 =
extrinsic factors = exogenous coagulation factors has the same structure as
the method described in 3.1.5.1 PT Parameter.
The preset method on the method memory position <8 Intr. Factor> = intrin-
sic factors = endogenous coagulation factors has the same structure as the
method described in 3.1.5.1 PT Parameter.
3.1.5.14 none
The method memory position can be defined.
3.1.5.15 none
The method memory position can be defined.
3.2 Utilities
The menu UTILITIES comprises a group of submenus which allow quick
access to the settings in the following menus:
<printer>, <beeper>, <reagent-stirrer>, <secret number > and
< cuvette test >.
After a selection has been made in one of the listed menus, the following
display will appear:
select:ESC= work
ENTER= more param.
Press Esc to access the measuring mode or press Enter to access more
menu options.
checking parameters
please wait
You have the choice to press the Enter-key to save parameters or press
Esc to leave the menu without saving. In both cases an upper menu
appears.
- Error messages
3
3.2.2 Menu <beeper>
The beeper acoustically confirms the key functions and beeps in case of
error.
In this menu you can select between <ON>, <OFF>, and <CLICK>.
After <secret number> has been selected, the following display appears:
- Press Enter to confirm the entry. A print out will follow (printer mode: ON).
-----------------------------------------------------------
BFT II
ser. no. B1770711 Example
secret no=xxxxx x= count
-----------------------------------------------------------
If a zero is entered and saved, the secret number will not be asked for when
selecting the parameter menu. It will be printed out as “secret no=00000“.
After <cuvette test> has been selected, the following display appears:
123/622 122/597
3 --- cuv
If a cuvette was placed in the measuring channel, the display reads cuv.
When removing and inserting the cuvette, the display must show the
appropriate status.
The upper display shows lamp values. Lamp values inform
about the liquid turbidity. The information is for Dade Behring Service only.
4 Errors
Possible errors can be categorized in:
break bot lim: Sample preparation too turbid; the sample has been
adjusted but cannot be measured!
Print-out: channel x break bot lim
break top lim: Sample preparation very light; the sample has been
adjusted but cannot be measured!
Print-out: channel x break top lim
break
jump: When detecting a jump within a measuring signal the
measurement will be stopped and "break jump" will
be displayed. Air bubbles in the sample might cause
this.
Monitoring of rpms:
Message:
break motor: Speed difference is >50 rpm from predetermined-
stirrer-speed
Print-out: t=8.0s, s/i=500/446rpm
channel x: break motor
t=meas. time;
s/i=predetermined- / real-stirrer-speed
x=measuring channel 1 or 2
break
readjust Message is displayed if the light value is too dark
during the adjustment phase.
Print-out: channel x: break readjust
Measured
value: If liquid is too turbid the message "too dark" will be
displayed.
4
Consequence: Sample cannot be measured and may possibly be
too turbid.
Temperature fluctuations
If temperature fluctuates during the measurement, the following information
will be included in the print-out: WARN: temp. unstable
Temperature error
Internal defects are continuously monitored. Hardware and software are
controlled.
If the temperature is too low, the information WARM UP will be displayed in
the STANDBY mode.
In case the temperature is significantly higher, switch off the BFT* II.
Message:
Display: ERROR | xxx(xxx=error number)
4 Print-out: >>>ERROR<<< | xxx
The error message is sent via the serial interface. The beeper will send a
signal every second.
4.4 Troubleshooting
Please check BFT* II, reagents, and accessories as described below prior
to contacting Dade Behring Service.
Have the samples been carefully pipetted onto the bottom of the cuvette?
Has the reagent been properly mixed prior to application? This is especially
important for reagents containing Kaolin.
Does the applied calcium chloride have the correct molar concentration for
the type of analysis?
Has the reagent been sitting for a longer time period at 37 °C without being
4 closed so that its concentration may have changed due to evaporation? See
package insert.
Was the correct reference curve applied for the reagent lot?
Has the patient plasma been tested within four hours after taking the blood
(as recommended in the package insert)?
Was the measurement carried out immediately after the display of xxx ul
GO?
Have the reference curve points been correctly defined and entered?
You may need to change the fuses if the BFT* II does not operate even
though it is properly connected.
NOTE
• To check for blown fuses remove the fuses from the fuse holder and see if
wires have melted. Replace fuses accordingly.
NOTE
Contact Dade Behring Service in case the exchange of fuses has been
unsuccessful.
pull
1 pull
2
a b
4
c
5 Software
The software for the BFT* II is stored in a flash memory and is activated as
soon as the analyzer is switched on.
It controls the analyzer via start functions for the analytic program.
Visual communication between the BFT* II and the user is accomplished via
a Liquid Crystal Display with two rows and 20 characters.
The display will lead you in a dialog through all measurement steps. The
user will confirm these steps either by key strokes or by pipetting start
reagent. Thus correct handling of the system will be guaranteed.
An overview of the BFT* II Software can be seen in Figure 12, chapter 5.1
Software Overview
Once the secret code number has been entered, the user can
- change the method menu
- access all method parameters
- access all parameters in the menu UTILITIES
The preset methods in ROM memory cannot be changed by the user. They
are fixed together with the operating program.
BFT II
Power ON - Initialising
WARM UP
Software Version 2.01
remove all cuvettes
then press any key
auto blanking
keep channels clear
Cal
Enter >
5 Method parameter
secret no.:
Test run Method list
Cal
incubat incubat STANDBY 37.4 deg
..... cuv in cuv in <1 PT >
Direct access to special parameter
secret no.:
STANDBY 37.4 deg
.....
< 11 ProC G PCAT/0 > 1
UTILITIES select:
STANDBY 37.4 deg
< printer >
< 12 ProC FV PCAT > 1
Before routine tests can be performed, the user must change certain
reagent specific parameters such as lot number and reference curve
parameter.
For a full list of available method parameters contact Dade Behring Service.
6 Appendix
Danger!
6
NOTE
Introduces rules to be observed.
A Electrical Safety
DANGER Check the operating voltage before you connect the device to the main
power supply.
To connect the device to the power supply, use only sockets which are
Danger! grounded in order to keep the risk of an electrical shock as low as
possible. Use only grounded extension cables.
Other than the light protection cap, never remove screwed-in casing
parts, protective guards or secured components since this could
expose electrically live parts.
Surfaces (floors, work table) must not be moist when you are working
with any electrical device.
Use only replacement fuses of the stated type and with the stated
nominal current.
Never use fuses which have been "repaired".
Never short-circuit the fuse holder.
C Mechanical Safety
(BFT* II in operation)
CAUTION Never open screw-attached housing parts while the instrument is ON.
There is a risk of injury due to moving parts (fan, motor drives).
D Samples
Avoid any direct contact with samples which are potentially infectious or
which may generate other risks to the human body.
If sample material is spilled onto the BFT* II, wipe it off immediately and
decontaminate the analyzer (refer to chapter 6.2 Maintenance and
Hygiene).
E Reagents
CAUTION Observe the suggestions in the package inserts for a correct use of the
reagents. Wear gloves.
6
F Waste Solution
H Operator Qualification
I Room Temperature
6
Keep the instrument clean and do not spill liquids onto the analyzer.
To protect the instrument from dust, cover it with the supplied dust cover or
store instrument in a cabinet when not in use.
In case liquids were spilled onto the instrument, immediately absorb liquid
with an appropriate cloth.
If liquid has accidentally run or was pipetted onto the red tongue of the light
protection cap or into the measuring channel, remove the liquid with a
pipette and clean the light protection cap and/or the measuring channel with
a lint-free cloth (refer to chapter 2.1.3 Measuring).
NOTE
6.3 Disposables
NOTE
Cuvettes
5 x 500 cuvettes and 500 x mixers
or
500 cuvettes with mixers („Dispo System“)
Thermo paper
Printer paper for BFT* II
Operation manual
Light protection cap for protection against foreign light plus guide for
Eppendorf pipette tips.
The following features should be observed when disposing of the BFT* II:
• Make sure that the BFT* II has been decontaminated before disposal.
If the instrument bears no IVD symbol on the typeplate it complies with the
following directives:
- 73/23/EEC Low-voltage directive
- 89/336/EEC-EMC directive
Interference suppression
The instrument was checked according to EN 50081-1, EN 50082-1 and EN
61000-4-2 to EN 61000-4-5 and meets the requirements of limit value class
B. The use of screened data cables is a precondition for compliance with
the relevant regulations. The user is responsible for ensuring that screened
data cables are used.
Manufacturer
Batch code
6
CE mark
Catalogue number
Do not reuse
Best before
Temperature limitation
6.8 Mathematics
Computations for coagulation methods:
Computation of Factor:
( 25 % Value in sec ) - ( 100 % Value in sec )
———————————————————— = Factor
(1/25) - (1/100 %)
or
Conversion limits:
In general the tolerances for the conversion can be programmed. See chap-
ter 3.1.5.1 PT Parameter conversion limits min/max values.
Interpolation
Linear from reference point to reference point.
Extrapolation
PT: >100 % extrapolation over the last two higher points.
Extrapolation factor: 1.5
PT: < 10 % extrapolation over the last two lower points.
Extrapolation factor: 0.5
Conversion to INR:
INR = RATIO ISI
(International Normalized Ratio)
6.9 Print-outs
General print-outs
After a method has been selected the programmed reference curve param-
eters are printed prior to printing the results.
NOTE
Each method has its own counter. As soon as the BFT* II has been
switched on, the counter starts with "1". This feature can be set in menu
method parameter / measurement - printout number.
Print-out for PT
PT documentation
Example: conversions via a 4 point reference curve
(max 9 point) in %, INR as an example
start-reagent:
Lot 505400
reagent = 100 ul
(0=off)
incub.1= 0s
(0=off)
incub.2= 60 s
(0=off)
------------------------------
results
left channel
results
right channel
PT . Point = diff. to preset method
patient _____________ patient name entry
No. = 1 print-out number 1 .....
time = 24.5 s measured time in seconds
INR = 2.20 conversion in INR
% = 36.4% conversion in PT %
Additional:
Hardware
Mini-DIN round connector on the back of the BFT* II.
Assignment:
-Pin 1= do not use, for manufacturer only !
-Pin 2= GND
-Pin 3= CTS
-Pin 4= TXD
-Pin 5= do not use, for manufacturer only !
-Pin 6= RXD
Send and receive with 9600 Baud, 8 data bits, 1 stop bit, without parity bit.
Handshake possible with CTS-input, but the connected instrument should
be able to receive without handshake with 9600 Baud.
6 6 5
4 3
2 1
Software
Transmission format:
The BFT* II transmits each message in the following format:
STX <code><data> ETX <BCC><CR><LF>
whereby:
- STX byte $02
- <code> byte to label the type of message
- <data> a sequence from 1 - 256 bytes
- ETX byte $03
- <BCC> 2 bytes from the character set ‘0’..’9’, ‘A’..’F’.
If a message has been confirmed with NAK or the BFT* II did not receive a
confirmation within approximately 300 ms, the BFT* II will send the
message again. The same message can be sent up to three times.
Depending on the number and priority of other messages to be sent, fewer
repeats or even no repeats may be possible.
All other data is separated from the channel number by a space. The data
consists of:
-measured time in seconds with one decimal
-first conversion
-second conversion
-alarm code
If conversions are set to off, '0' will be shown in the respective places.
Procedure
A program that wants to receive measured results from the BFT* II can
ignore all messages whose code is not ‘E’ but must confirm them
immediately with ACK.
A
auto blanking automatic adjustment
activation activation period
B
beeper acoustic signal
break too dark refer to chapter 4.2
break bot lim refer to chapter 4.2
break top lim refer to chapter 4.2
break jump refer to chapter 4.2
break motor refer to chapter 4.2
break readjust refer to chapter 4.2
break range refer to chapter 4.2
break drift refer to chapter 4.2
C
6 Cal
cuv in
calibration
insert cuvette into measuring channel
cuv out remove cuvette from measuring channel
calibrate temp temperature adjustment
cool down Temperature is > 37°C, refer to 4.3
D
deg degree, °C
E
Esc escape
extr. extrinsic system (Factor II, V, VII, X)1
ERROR I xxx Technical error, see chapter 4.3
F
factor factor for PT reference curve
G
Go continue, next step
g/l grams / liter
I
incub incubating
incubat. incubation time
INR (inr) International Normalized Ratio
ISI International Sensitivity Index
intr. Intrinsic System (Factor VIII, IX, XI,XII)1
K
Keep channels-
clear keep channels empty
L
lot no. Lot number
M
more param. more parameters
mixer stir bar speed
O
on active
off not active
P
point point on the reference curve
parameter error write/read error, refer to chapter 4.2
R
ROM read-only-memory
RAM random access memory
S
secret no. secret number, secret code
s (60 s in display) seconds
StartReag starting reagent
6
T
t time
timeout refer to chapter 4.2
top lim / bot lim refer to chapter 4.2
too dark refer to chapter 4.2
U
ul µl = Microliter
W
writing para-
meters to internal saving parameters
work measuring mode
WARN:
temp. unstable warning, temperature unstable
refer to chapter 4.2
other:
% activity in percent
Cuvettes 6-4
activity 3-10 D
Additional requirements and recommendations 1-7 decimal place 3-14
Bullets 2-1
concentration 3-10
Contamination 1-7 I
conversion example 6-10
INR 6-9
conversion limits 3-14, 3-16
Installation 1-5
interpolation 3-15 Q
ISI-value 3-18
quality control 1-7
ranges 3-10
M
ready for operation 2-2
Temperature
error 4-3
fluctuations 4-3
turbidity 1-6
unit 3-14
UTILITIES 2-8
WARM UP 2-2