9700 User Guide
9700 User Guide
User Guide
© Copyright 2003, Applied Biosystems. All rights reserved.
For Research Use Only. Not for use in diagnostic procedures.
Information in this document is subject to change without notice. Applied Biosystems assumes no responsibility for any errors that may appear in this document. This
document is believed to be complete and accurate at the time of publication. In no event shall Applied Biosystems be liable for incidental, special, multiple, or
consequential damages in connection with or arising from the use of this document.
The combination of the Dual 96 and Auto-Lid Dual 96 Interchangeable Sample Block Modules with a GeneAmp® PCR System 9700 Base Unit is licensed under the
thermal cycler apparatus claims of U.S. Patent No. 5,333,675 and counterpart patent claims in other countries.
Patents are pending on the Auto-Lid mechanism.
TRADEMARKS:
ABI PRISM, Applied Biosystems, and MicroAmp are registered trademarks of Applera Corporation or its subsidiaries in the U.S. and/or certain other countries.
AB (Design) and Applera are trademarks of Applera Corporation or its subsidiaries in the U.S. and/or certain other countries.
GeneAmp is a registered trademark of Roche Molecular Systems, Inc.
All other trademarks are the sole property of their respective owners.
DRAFT
June 24, 2003 12:59 pm, 44343363 B 9700D96_UG_Title.fm
Contents
Preface
How to Use This Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
How to Obtain More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
How to Obtain Services and Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT iii
iv
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Audience This guide is intended for novice and experienced GeneAmp® PCR System 9700
users who incorporate either the Auto-Lid Dual 96 Sample Block Module or the
manual Dual 96 Sample Block module.
User Attention Two user attention words appear in Applied Biosystems user documentation. Each
Words word implies a particular level of observation or action as described below:
Note: Provides information that may be of interest or help but is not critical to the
use of the product.
IMPORTANT! Provides information that is necessary for proper instrument
operation, accurate chemistry kit use, or safe use of a chemical.
Examples of the user attention words appear below:
Note: The size of the column affects the run time.
Note: The Calibrate function is also available in the Control Console.
IMPORTANT! To verify your client connection to the database, you need a valid
Oracle user ID and password.
IMPORTANT! You must create a separate Sample Entry Spreadsheet for each 96-well
microtiter plate.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT v
Safety Alert Safety alert words also appear in user documentation. For more information, see
Words “Safety Alert Words” on page viii.
Send Us Your Applied Biosystems welcomes your comments and suggestions for improving its
Comments user documents. You can e-mail your comments to:
[email protected]
vi
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT vii
Definitions
IMPORTANT! – Indicates information that is necessary for proper instrument
operation, accurate chemistry kit use, or safe use of a chemical.
Examples
The following examples show the use of safety alert words:
IMPORTANT! You must create a separate a Sample Entry Spreadsheet for each
96-well microtiter plate.
The lamp is extremely hot. Do not touch the lamp until it has
cooled to room temperature.
viii
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Symbols on Instruments
Electrical The following table describes the electrical symbols that may be displayed on
Symbols on Applied Biosystems instruments.
Instruments
Symbol Description
Safety Symbols The following table describes the safety symbols that may be displayed on
Applied Biosystems instruments. Each symbol may appear by itself or in
combination with text that explains the relevant hazard (see “Safety Labels on
Instruments” on page x). These safety symbols may also appear next to DANGERS,
WARNINGS, and CAUTIONS that occur in the text of this and other
product-support documents.
Symbol Description
Indicates that you should consult the manual for further information and to
proceed with appropriate caution.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT ix
Symbol Description
Indicates the presence of a laser inside the instrument and to proceed with
appropriate caution.
English Francais
CAUTION Hazardous waste. Read the ATTENTION Déchets dangereux. Lire les
waste profile (if any) in the site preparation renseignements sur les déchets avant de
guide for this instrument before handling or les manipuler ou de les éliminer.
disposal.
x
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Operating the Ensure that anyone who operates the instrument has:
Instrument • Received instructions in both general safety practices for laboratories and
specific safety practices for the instrument.
• Read and understood all applicable Material Safety Data Sheets (MSDSs). See
“About MSDSs” on page xi.
Chemical Safety
Chemical Hazard CHEMICAL HAZARD. Before handling any chemicals, refer
Warning to the Material Safety Data Sheet (MSDS) provided by the manufacturer, and
observe all relevant precautions.
About MSDSs Chemical manufacturers supply current Material Safety Data Sheets (MSDSs) with
shipments of hazardous chemicals to new customers. They also provide MSDSs with
the first shipment of a hazardous chemical to a customer after an MSDS has been
updated. MSDSs provide the safety information you need to store, handle, transport,
and dispose of the chemicals safely.
Each time you receive a new MSDS packaged with a hazardous chemical, be sure to
replace the appropriate MSDS in your files.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT xi
Obtaining You can obtain from Applied Biosystems the MSDS for any chemical supplied by
MSDSs Applied Biosystems. This service is free and available 24 hours a day.
To obtain MSDSs:
1. Go to https://docs.appliedbiosystems.com/msdssearch.html
2. In the Search field, type in the chemical name, part number, or other
information that appears in the MSDS of interest. Select the language of your
choice, then click Search.
3. Find the document of interest, right-click the document title, then select any of
the following:
• Open – To view the document
• Print Target – To print the document
• Save Target As – To download a PDF version of the document to a
destination that you choose
4. To have a copy of a document sent by fax or e-mail, select Fax or Email to the
left of the document title in the Search Results page, then click RETRIEVE
DOCUMENTS at the end of the document list.
5. After you enter the required information, click View/Deliver Selected
Documents Now.
xii
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Waste Disposal If potentially hazardous waste is generated when you operate the instrument, you
must:
• Characterize (by analysis if necessary) the waste generated by the particular
applications, reagents, and substrates used in your laboratory.
• Ensure the health and safety of all personnel in your laboratory.
• Ensure that the instrument waste is stored, transferred, transported, and disposed
of according to all local, state/provincial, and/or national regulations.
IMPORTANT! Radioactive or biohazardous materials may require special handling,
and disposal limitations may apply.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT xiii
Electrical Safety
Fuses ELECTRICAL SHOCK HAZARD. Improper fuses or high-
voltage supply can damage the instrument wiring system and cause a fire. Before
turning on the instrument, verify that the fuses are properly installed and that the
instrument voltage matches the power supply in your laboratory.
Overvoltage The GeneAmp PCR System® 9700 system has an installation (overvoltage) category
Rating of II, and is classified as portable equipment
xiv
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Workstation Safety
Correct ergonomic configuration of your workstation can reduce or prevent effects
such as fatigue, pain, and strain. Minimize or eliminate these effects by configuring
your workstation to promote neutral or relaxed working positions.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT xv
U.S. and This instrument has been tested to and complies with standard UL 3101-1, “Safety
Canadian Safety Requirements for Electrical Equipment for Laboratory Use, Part 1: General
Standards Requirements.”
This instrument has been tested to and complies with standard CSA 1010.1, “Safety
Requirements for Electrical Equipment for Measurement, Control, and Laboratory
Use, Part 1: General Requirements.”
Canadian EMC This instrument has been tested to and complies with ICES-001, Issue 3: Industrial,
Standard Scientific, and Medical Radio Frequency Generators.
EMC
This instrument meets European requirements for emission and immunity (EMC
Directive 89/336/EEC). This instrument has been tested to and complies with
standard EN 61326 (Group 1, Class B), “Electrical Equipment for Measurement,
Control and Laboratory Use – EMC Requirements.”
Australian EMC This instrument has been tested to and complies with standard AS/NZS 2064,
Standards “Limits and Methods Measurement of Electromagnetic Disturbance Characteristics
of Industrial, Scientific, and Medical (ISM) Radio-frequency Equipment.”
xvi
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 1-1
Compatible Interchanging sample block modules may require a firmware change. If the sample
Software block module is placed on a base module running an earlier software version, the
instrument will not operate correctly. To upgrade the base module software, use the
floppy disk that is shipped with the Dual 96 Sample Block Module. For instructions
on upgrading the software, see the GeneAmp PCR System 9700 Base Module User’s
Manual (P/N 4303481) or see “How to Obtain Services and Support”on page vi.
Lid Choices The Dual 96 Sample Block Module comes with one of the following:
• An automatic lid
• A manual lid
The major difference between the two lids is that with the manual lid, you must
physically slide the lid back and forth into position; with the Auto-Lid, you simply
press the red button to move the lid.
Both lids are heated.
Heated Cover The heated cover slides over the sample blocks and performs two functions:
• It raises the temperature of the upper part of the tube (the part above the sample
blocks) above the temperature of the reaction mixture. This reduces
condensation, which may have a negative effect on your chemistry.
• It puts pressure on the reaction plates to seat them firmly and precisely in the
sample blocks. This enhances heat transfer.
1-2
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Auto-Lid The heated Auto-Lid automatically moves over to cover the reaction plates at the
start of a PCR run, and then moves to the back of the module and opens when the run
is complete.
IMPORTANT! If you do not want the lid to open after a run, you need to apply a 4 ° C
hold at the end of the program.
The Auto-Lid Sample Block Module (Figure 1-1) attaches to the top of the
GeneAmp® PCR System 9700 Base Module and allows you to perform up to 192
reactions, 0.2 mL each. The Auto-Lid Sample Block Module also adds automation to
the System 9700, which can increase throughput.
Heated Auto-Lid
(open)
GR1759b
GeneAmp® PCR
System 9700
Dual 96 sample blocks
Base Module
Move the Auto-Lid by pressing the red button
Figure 1-1 Auto-Lid Dual 96 Sample Block Module on the GeneAmp® PCR
System 9700 Base Module
Manual Lid The manual lid Dual 96 Sample Block Module (Figure 1-2) is used throughout this
manual as the primary example.
STOP F2
F3
1 F4
Gene 2
PCR Sy Amp ®
F5
4
stem 97 3
00 5
7 POWER
6
8
ENTER
9
0
CE
Figure 1-2 Manual Lid Dual 96 Sample Block Module on the GeneAmp® PCR
System 9700 Base Module
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 1-3
Sample Blocks There are two sample blocks in the Dual 96 Sample Block Module. Each holds one
96-well reaction plate.
IMPORTANT! The right sample block is the primary control block. If you are using
only one reaction plate:
• Place the reaction plate you are using in the right sample block.
• Place an empty reaction plate in the left sample block.
GR2249
GR2249
Using reaction plates on both sides If you use just one reaction plate, place it on
the right. Place an empty plate on the left.
1-4
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
GR2249
2. Place a second reaction plate in the left sample block.
IMPORTANT! You must insert two plates into the sample block for the
instrument to run properly.
3. Close the heated lid:
• For an Auto-Lid, press the red button located near the right sample block.
• For a manual lid, use the handle on top of the lid.
4. Process the samples.
Peltier Heating/ The internal Peltier heating/cooling unit is housed in the sample block module. The
Cooling Unit sample block module is made of aluminum to provide the optimal thermal transfer
rate.
• Platinum sensors in the sample block module provide:
• Wide temperature range: 4 to 99.9 ° C
• Accuracy: ±0.25 ° C from 35 to 100 ° C
• Uniformity of each sample block: ±0.5 ° C (measured after 30 sec at 95 ° C)
• Uniformity of combined sample blocks: ±0.75 ° C (measured after 30 sec at
95 ° C)
• Long-term stability and high reliability
Modules, The modules listed below are available for the GeneAmp PCR System 9700 Base
Accessories, and Module. The accessories and disposables are for use with the Dual 96 Sample Block
Disposables Module.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 1-5
Ordering Order modules, accessories, and disposables from Applied Biosystems by part
number. See Tables 1-1 through 1-3.
1-6
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 1-7
GR2249
Lever
GR1762
1-8
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Installing the To install the sample block module (Auto-Lid and manual lid) into the GeneAmp
Sample Block PCR System 9700 Base Module:
Module 1. Pull out the lever from the sample block module.
2. Place the sample block module onto the base module, then push the sample
block module back to seat the electrical connections.
3. Push the lever into the base module to secure the sample block module.
Note: If the sample block module is not seated correctly, the instrument cannot
be turned on.
Installing the To install the communication card and cable into the GeneAmp PCR System 9700
Auto-Lid Module Base Module:
Communication 1. Insert the communication card into the slot at the rear of the right side of the
Card and Cable instrument (the label on the card face should be facing you).
2. Insert the connector end of the communication cable (very short cable) into the
communication card.
3. Insert the other end of the cable (plug type) into the receptacle on the sample
block module labeled “I/O Port.”
Installing the To install the power line cords into the sample block module and the GeneAmp
Auto-Lid Module PCR System 9700 Base Module:
Power Line Cords 1. Plug one end of the power line cord into the sample block module receptacle and
and Powering Up the other end into an outlet.
2. Plug the very short jumper cable into the sample block module and the other
into the base module.
3. Flip the power switch located on the sample block module near the plug
receptacle and press the power button located on the front of the instrument.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 1-9
With this
Use... As Shown
vessel...
MicroAmp 96-Well
Support Base
MicroAmp 96-Well
MicroAmp 96-Well
Full Plate Cover
Full Plate Cover
MicroAmp
Optical 96-Well
Reaction Plate
MicroAmp 96-Well
Support Base
GR2250
MicroAmp Optical
96-Well Reaction
Plate
GR2251
MicroAmp 96-Well
Support Base
1-10
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
With this
Use... As Shown
vessel...
MicroAmp 96-Well
Tray for
Tubes with Caps
GR2252
MicroAmp 96-Well
Support Base
MicroAmp
96-Well Tray
MicroAmp 96-Well
GR2253
Support Base
MicroAmp
96-Well
Full Plate MicroAmp 96-Well
Cover Full Plate Cover
MicroAmp
96-Well Retainer
MicroAmp
8-Strip Tubes or
Single Tubes
MicroAmp
96-Well Tray
MicroAmp 96-Well
Support Base
GR2254
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 1-11
Loading Samples
The following procedures describe how to load samples for:
• Tubes with attached caps
• 96-Well reaction plate
• 96-Well Tray Retainer Assembly
• Sample volume range: 5 to 100 µL in “D96” mode
IMPORTANT! Do not use mineral oil or glycerine in the sample block or as a vapor
barrier over the PCR reaction mixture in the tubes. The MicroAmp Reaction Tubes
fit tightly in the wells and a heated cover exerts even pressure on all tubes, reducing
condensation on the tubes when you are using a tray or tray/retainer set.
1-12
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
H12
A
1
2
3
4
5
6
7
8
9
10
11
12
GR1070
Figure 1-5 Align well H12 with upper right corner of the tray
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 1-13
1-14
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 2-1
Cleaning the Clean the heated cover at least once a month or more frequently, if needed.
Heated Cover
During instrument operation, the temperature of the heated cover
can be as high as 108 ° C, and the temperature of the sample block module can be as
high as 100 ° C. Before performing this procedure, wait until the heated cover and
sample block module reach room temperature.
2-2
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Slot
Slot
F1
GR2255
STOP F2
F3
1 F4
Gene 2
PCR Sy Amp ®
F5
4
stem 97 3
00 5
7 POWE
6 R
8
ENTE
R 9
0
CE
5. Soak a cotton swab or clean cloth with pure isopropanol and gently wipe the
bottom of the heated cover.
Cleaning the Clean the sample block wells once a month or more frequently, if needed.
Sample Block
Wells To clean the sample block wells:
1. Let the system 9700 cool to the idle setpoint (25 °C).
2. Put the heated cover into the cleaning position. See step 4 on page 2-2 and
Figure 2-1 above.
3. Power off the instrument.
4. Unplug the instrument.
5. Remove the reaction plates from the sample block module and set them aside.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 2-3
6. Soak a cotton swab in pure isopropanol and clean the sample block wells
thoroughly.
7. Remove any remaining isopropanol from the sample block before reloading the
reaction plates.
8. If the sample block wells become contaminated with amplified DNA:
a. Clean the wells thoroughly with a cotton swab soaked in 10% bleach.
b. Wipe with a clean, damp cloth.
c.
CHEMICAL HAZARD. Sodium hypochlorite (bleach) is
a liquid disinfectant that can be corrosive to the skin and can cause skin
depigmentation. Please read the MSDS, and follow the handling instructions.
Wear appropriate protective eyewear, clothing, and gloves.
2-4
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-1
In this procedure, you use the digital thermometer to take temperature readings of
one sample well connected to the 0.2-mL Probe Assembly. You take a reading at two
setpoint temperatures, 85 °C and 45 °C.
Required This test requires the Temperature Verification System, Dual 0.2 mL (P/N 4346117),
Materials which includes:
• Light mineral oil
• Cotton swabs
• Temperature Verification Plates
• 0.2-mL Probe Assembly
• Digital thermometer with 9 V battery installed
Note: This system is also used for the Auto-Lid Sample Block Module.
3-2
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Configuring the Before you begin configuring the System 9700, perform the procedure described in
System 9700 “Setting Up the 0.2-mL Probe Assembly” on page 3-3.
Temperature Verification
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-3
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Place probe in well R3
Press Run
Run Cancel
F1 F2 F3 F4 F5
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Setpoint is 85°C
Cover must be within 1°C of 105°C
Cancel
F1 F2 F3 F4 F5
The System 9700 ramps up the lid temperature (cover temp) to 105 ° C (±1 ° C).
It may take several minutes for the System 9700 to ramp up.
When the lid temperature is reached, the “Stabilizing at setpoint” screen is
displayed. The “Stabilizing at setpoint” value begins to count down.
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Cancel
.
F1 F2 F3 F4 F5
2. When the “Stabilizing at setpoint” value decrements to zero, read the digital
thermometer.
Refer to the instructions included with the Temperature Verification System for
a detailed description on operating the digital thermometer.
3-4
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
3. Using the numeric keys, type the value displayed on the digital thermometer in
the “Enter actual block temperature” field.
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Cancel
F1 F2 F3 F4 F5
Note: The digital thermometer displays a four-digit value; round off the value
to three digits before typing it in the Calibration Verification screen.
Record the value on the Calibration Verification Test Data Sheet (page 3-7).
4. Press Enter.
The System 9700 automatically begins the second reading. The Calibration
Verification screen is displayed for the second setpoint, 45 °C. If necessary, the
System 9700 ramps up the lid temperature (cover temp) to 105 ° C (±1 ° C)
before initiating the “Stabilizing at setpoint” value countdown.
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Setpoint is 45°C
Cover must be within 1°C of 105°C
Cancel
F1 F2 F3 F4 F5
5. When the “Stabilizing at setpoint” value decrements to zero, read the digital
thermometer and enter the value for the second reading as described in steps 2
and 3. Record the value.
The System 9700 evaluates the calibration of the sample block temperature for
the setpoint values you entered and displays the results. A summary screen is
displayed at the conclusion of the test.
Calibration Verification
Accept Cancel
F1 F2 F3 F4 F5
Compare the values displayed with the actual test results you recorded. If the
values differ, repeat the procedure to make sure you entered the values
accurately.
6. If the values agree, press F1 (Accept).
7. Continue with “Evaluating the Results” on page 3-6 and “Ending the Test” on
page 3-6.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-5
Evaluating the When the System 9700 completes the Calibration Verification Test, one of two
Results screens is displayed (Table 3-1) to evaluate the results.
Is properly calibrated
Calibration Verification
Calibration is good
Exit
F1 F2 F3 F4 F5
3-6
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Data Sheet: When you run the Calibration Verification Test, record the setpoint values for
Calibration well R3 on a copy of the data sheet. At the end of the Calibration Verification Test,
Verification Test check the values displayed on the System 9700 against the values recorded to ensure
accurate test records.
IMPORTANT! Use a photocopy of this page to record values, and keep the original
intact.
.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-7
The Temperature Non-Uniformity Test uses the 0.2-mL Probe Assembly to test the
temperature uniformity of all six wells in the left and right sample blocks. You test
each well at two setpoints, 95 ° C and 55 ° C.
3-8
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
7. Slide the heated cover forward and pull the lever down.
IMPORTANT! Seat the probe properly and close the heated cover carefully. If
the probe wire is crushed when the heated cover is closed, the probe may be
damaged.
8. Power on the digital thermometer.
Note: Refer to the instructions included with the Temperature Verification
System for a detailed description on operating the digital thermometer.
9. Continue with “Configuring the System 9700” on page 3-9.
Configuring the Before you begin configuring the System 9700, perform the procedures described in
System 9700 “Setting Up the 0.2-mL Probe Assembly” on page 3-8.
Temperature Verification
5. Press F2 (TNU). The TNU Performance setpoint screen is displayed and the
system is automatically configured for the Temperature Non-Uniformity Test.
When the configuration is complete, the TNU Performance probe position
screen is displayed.
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-9
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Setpoint is 95°C
Sample must be within 1.0°C of setpoint
Cancel
F1 F2 F3 F4 F5
If necessary, the System 9700 ramps up the sample block (sample temp) and lid
(cover temp) temperatures to the specified values. It may take several minutes
for the System 9700 to stabilize at these temperature.
When the sample block and lid temperatures are reached, the “Stabilizing at
setpoint” screen is displayed. The TNU Performance screen counts down the
time until the setpoint is stabilized.
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Stabilizing at setpoint x:xx
Cancel
F1 F2 F3 F4 F5
2. When the “Stabilizing at setpoint” value decrements to zero, read the digital
thermometer.
Refer to the instructions included with the Temperature Verification System for
a detailed description on operating the digital thermometer.
3-10
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
3. Using the numeric keys, type the value displayed on the digital thermometer in
the “Enter actual block temperature” field.
TNU Performance
Note: The digital thermometer displays a four-digit value; round off the value
to three digits before typing it in the Calibration Verification screen.
Record the value and well number on the Temperature Non-Uniformity Test
Data Sheet (page 3-14).
4. Press Enter.
The System 9700 automatically begins the second reading.The TNU
Performance screen is displayed for the second setpoint, 55 °C. If necessary, the
System 9700 ramps up the sample and lid (cover temp) temperatures to the
specified values before initiating the “Stabilizing at setpoint” value countdown.
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Setpoint is 55°C
Sample must be within 1.0°C of setpoint
Cancel
F1 F2 F3 F4 F5
5. When the “Stabilizing at setpoint” value decrements to zero, read the digital
thermometer and enter the value for the second reading as described in steps 2
and 3. Record the value.
6. Press Enter.
The TNU Performance screen is displayed with the following message:
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Place probe in well xx
Press Run
Run Cancel
F1 F2 F3 F4 F5
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-11
Compare the values displayed with the actual test results you recorded on the
Temperature Non-Uniformity Test Data Sheet. If the values differ, repeat the
procedure to make sure the values were accurately entered.
11. If the values agree, press F1 (Accept).
12. Continue with “Evaluating the Results” on page 3-13 and “Ending the Test” on
page 3-13.
3-12
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Evaluating the When the System 9700 completes the Temperature Non-Uniformity Test, the TNU
Results Performance screen is displayed. See the table below to evaluate the results.
Ending the Test To end the test, follow the procedures in “Ending the Test” on page 3-6.
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-13
Data Sheet: When you run the Temperature Non-Uniformity Test, record the setpoint values for
Temperature wells L1–L6 and R1–R6 on a copy of this data sheet. At the end of the Temperature
Non-Uniformity Non-Uniformity Test, check the values displayed on the System 9700 against the
Test values recorded to ensure accurate test records.
IMPORTANT! Use a photocopy of this page to record values, and keep the original
intact.
Date
Tested By
Setpoint Value 95 °C 55 °C
L1
L2
L3
L4
L5
L6
R1
R2
R3
R4
R5
R6
3-14
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Accessing the To access the tests, press System from the Diagnostics screen. The System
Tests Performance screen is displayed.
System Performance
Performing the Use the Rate Test to verify that the Peltier units are operating correctly. The test takes
Rate Test approximately two minutes to run.
WARNING!!!
Install the appropriate empty
Consumables into the Sample Block.
Refer to System Performance Section
of the Block User Manual.
Cont Cancel
F1 F2 F3 F4 F5
Print Cancel
F1 F2 F3 F4 F5
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT 3-15
Performing the Use the Cycle Test to verify that the PCR cycling function operates properly. This
Cycle Test test takes approximately 15 minutes to run.
WARNING!!!
Install the appropriate empty
Consumables into the Sample Block.
Refer to System Performance Section
of the Block User Manual.
Cont Cancel
F1 F2 F3 F4 F5
The following table lists the passing ranges for the Cycle Test.
3-16
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Numerics conventions
bold text v
0.2-µL Probe Assembly, setting up IMPORTANTS! v
Temperature Non-Uniformity Test 3-8 in this guide v
96-Well Sample Block Module, part number 1-6 italic text v
96-well sample plate menu commands v
loading 1-12 Notes v
sample tray configurations 1-10, 1-11 user attention words v
conventions, safety viii
A cooling
information 1-5
Accessories, ordering 1-6 verifying cooling system 3-15
Applied Biosystems customer feedback, on Applied Biosystems
contacting vi documents vi
customer feedback on documentation vi
Services and Support vi Cycle Test
Technical Communications vi accessing 3-15
Technical Support vi running 3-16
Australian EMC standards xvi
Auto 1-3 D
Auto-Lid 1-3 DANGER
description viii
example viii
B
data sheet
biohazardous waste, handling xiii Calibration Verification Test 3-7
bold text, when to use v Temperature Non-Uniformity Test 3-14
diagnostics 3-15, 3-16
C accessing the tests 3-15
Cycle Test, running 3-16
Calibration Verification Test 3-2 to 3-6 Rate Test, running 3-15, 3-16
data sheet 3-7
diagrams, sample block module
discontinuing 3-2
cleaning position 2-2
exiting 3-2
top and bottom views 1-8
required materials 3-2
digital thermometer
Canadian safety standards xvi
Calibration Verification Test 3-2
CAUTION
disposables, ordering 1-6
description viii
example viii DNA contamination
heated cover 2-3
chemical safety xi
sample block wells 2-4
chemical safety guidelines xii
documentation
chemical waste feedback vi
hazards xiii related to this guide vi
safety xiii
Dual 384-Well Sample Block Module, part
safety guidelines xiii
number 1-6
chemical waste disposal, guidelines xiii
Dual 96 Sample Block Module Temperature Verification
cleaning the sample block module 2-2, 2-4 Kit, part number 1-6
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT Index-1
G L
labels, instrument safety x
general hazard symbol ix
laser hazard, symbol x
glycerine, warning 1-12
lid types 1-2
guidelines
chemical safety xii loading samples 1-12
chemical waste disposal xiii
chemical waste safety xiii M
waste disposal xiii
manual lid 1-3
menu commands, conventions for describing v
H MicroAmp Caps 1-10
hazard icons MicroAmp Full Plate Cover, part number 1-6
accompanying safety alert words viii
mineral oil, warning 1-12
components viii
described viii modules, ordering 1-6
in documents viii moving and lifting instrument, safety xi
on instruments viii, ix moving parts
See also hazard symbols hazard xiv
See also safety symbols hazard symbol x
hazard symbols safety xiv
electrical ix symbol xiv
general ix MSDSs
hot surface ix description xi
in documents viii obtaining vi, xii
laser hazard x referring to xii, xiii
moving parts x when to review xi
on instruments ix
See also hazard icons
See also safety symbols
N
hazards Notes, description v
chemical waste xiii
moving parts xiv O
moving/lifting instrument xi
physical xiv ordering parts 1-6
repetitive motion xv overvoltage category (rating) xiv
heated cover overvoltage rating xiv
cleaning 2-2
DNA contamination 2-3
heated lid 1-2
heating and cooling, information 1-5
Index-2
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
DRAFT Index-3
U
unloading samples, procedures for loading
samples 1-12
US safety standards xvi
user attention words, defined v
W
WARNING, description viii
waste disposal, guidelines xiii
workstation safety xv
Index-4
DRAFT GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide