7.4 SCBA Good Write Up 2
7.4 SCBA Good Write Up 2
Page 1 of 8, c:/Mukul/CTO/SCBA
If the facemask is dislodged whilst the wearer is in a smoky or toxic atmosphere,
it is vital that the wearer should hold his breath and immediately replace the
dislodged mask. If for any reason, there is likelihood of delay in doing so, the
wearer should be evacuated to fresh air at the earliest, even though he may not
feel any ill effects from the few breaths of contaminated air, which he has been
forced to inhale.
Please note in the above table the total full duration is given. Reduce the safety margin
from the above table to get the working duration.
Here we have taken an average consumption of 40 litres of air per minute being required
by human being when doing heavy work.
Pre-Operational Checks:
High Pressure Leak Test:
1. On the lung demand valves, press reset lever to switch ‘Off’ positive pressure
mechanism.
DO NOT press center of rubber cover while pressing reset lever, or force lever against
stop. See Figure below.
Page 2 of 8, c:/Mukul/CTO/SCBA
CORRECT WAY
WRONG WAY
Should whistle not sound at required pressure, whistle can be reset using manufacturers
instructions AND with Master or Chief Officer help.
DO NOT ATTEMPT TO ADJUST WHISTLE ON YOUR OWN.
Putting on equipment:
1. Extend shoulder straps and waist belt. Put on equipment.
2. Fasten buckle (See figure below). Pull ends of waist belt away from buckle until
secure and comfortable on hip. Tuck ends in belt loops.
Page 3 of 8, c:/Mukul/CTO/SCBA
3. Pull down shoulder straps (See figure below) until equipment is secure and
comfortable. Tuck strap ends under waist belt.
Page 4 of 8, c:/Mukul/CTO/SCBA
Connecting Demand Valve to Face piece:
1. Check face piece port and demand valve O-ring, are clean and undamaged.
2. Push demand valve into face piece port until it clicks into position.
3. Check attachment is secure. Pull demand valve away from face piece - there should
be no axial movement.
4. On achieving face seal positive pressure demand valves will automatically activate on
first inhalation.
5. Breathe normally.
6. Before use perform function checks.
Function Checks:
1. Close cylinder valve.
2. Breathe normally to vent system.
3. During venting observe gauge - whistle alarm should sound at preset pressure of
55bar +/-5bar.
4. When gauge indicates zero, hold breath. Face piece should hold onto face indicating
positive seal.
5. Open cylinder valve slowly, but fully to pressurize system.
6. Inhale and hold breath. Unit must balance, i.e. no audible leak.
7. Continue breathing. Expired air should easily flow out of exhalation valve.
8. Press center of rubber cover on demand valve checking supplementary supply.
Note: If leak detected, open cylinder valve, readjust head harness and retest.
During Use:
1. Cylinder pressure must not be less than 80% full.
2. Check gauge reading regularly. Whistle will sound at 55 bar +/- 5 bar.
3. When whistle sounds, exit and go to safe area, by shortest and safest route.
4. Do Not remove equipment until in safe area, clear of hazard.
After Use:
1. Press reset lever to switch ‘Off’ positive pressure.
2. Press and hold down the button and remove the demand valve.
3. Removing mask: Insert index finger behind the tabs on the neck straps and press the
buckles forwards with thumb, pulling the mask forward off your face and then up and
back over your head.
Page 5 of 8, c:/Mukul/CTO/SCBA
4. Close cylinder valve and remove face piece.
5. Unbuckle waist belt, lift shoulder strap buckles to loosen, remove equipment.
Maintenance on facemask:
Cleaning: The mask must be cleaned after every use. Do not use any organic solvents
such as acetone, alcohol etc.
Clean the mask with a cloth and lukewarm water mixed with all-purpose detergent.
Rinse thoroughly under running water.
Disinfecting: Disinfect mask after every use. Place mask in disinfectant bath. Only use
approved disinfectants - e.g. DAlSYdes. (R54 125)2). Excessive disinfectant
concentrations and over-long disinfecting times can damage the mask. Rinse thoroughly
under running water.
Drying: Leave to dry in air. Maximum temperature 60°C. Keep out of direct sunlight.
Testing: The mask must be tested every time it has been used, repaired or serviced as
per manufacturers instructions.
Visual inspection of the speech diaphragm: The speech diaphragm and O-ring must be
clean and undamaged. If not, clean or replace.
Visual inspection of inhalation valve: The inhalation valve is visible in the mask
connector. Check by blowing into it that both wings of the valve disc move freely.
Replace if seized.
Visual inspection of exhalation valve: Remove the protective cap from the connector.
Press one arm of the spring clamp inwards and remove the clamp. Hold valve disc by
the nipple and lift out of guide. The valve disc and seat must be clean and undamaged.
Clean or replace if necessary. Moisten valve disc with water and refit. When the seat is
Page 6 of 8, c:/Mukul/CTO/SCBA
horizontal, the valve disc must slide into the guide under its own weight. Fit spring clamp
and spring so that both clamp arms engage at the sides. The clamp is marked “L”= left
and “R” right and fits at an angle.
Page 7 of 8, c:/Mukul/CTO/SCBA
withdrawal, and the breathing mask shall be examined and checked in accordance with
the manufacturer's manual.
2. Every year all valves, sealing and speech membranes shall be checked.
3. Every second year the exhalation disc must be replaced. This should be done even if
the mask has not been used.
4. Every five years the complete breathing apparatus shall be submitted to a major
service check, which should include a major overhaul of the reduction valves in an
authorized workshop and a pressure test of the air bottles.
5. Pressure test of the air bottles shall be carried out by an approved service company.
6. If an air bottle shows any sign of corrosion or if it has been exposed to damage or
overheating it shall be replaced instantly.
Charging Cylinders:
Charge cylinder to pressure stamped on neck, or shoulder of cylinder.
Refer notes on Breathing Air Compressor.
Page 8 of 8, c:/Mukul/CTO/SCBA