YOR LAB

AN 1S/ISo/|EC 17025-NABL ACCREDITED LABORATORY


Plot No. 42/2, Survey No. 83, Opp. Gondal Bypass BRTS Bus Ston
CC-2342 150Feet Ring Road (West), Rajkot -360 004
E-mail :info @yorlab.com Web www.yorlab.com
Mo. +91 94266 04335 /94266 34335

Food Testng
CalibrationCertiicate
YUCCM 22/18
ULR: CC234223000006661 F
Certificate No. Date of Calibration Date of Issue Suggested Due Date
28/07/2023 27/07/2024
CC/2306661 28/07/2023
ISSUE/REV: 01/00
MetallurName & Address of Customer: SILICON PUMPS INDUSTRIES
Testing RS No. 723, URJA 2A, Plot No. 3,4,5, Near krishna Ind, B/H Goldcoin Foam, Sapar,
KotdáSangani, Rajkot -360024
Details of Unit Under Calibration (UUC):
Date of Receipt: 28/07/2023
CSLaIestina Condition on Receipt: Satisfactory
Name of Instrument: Pressure Gauge Range: 0 to 75 kg/cm²
Make: Pioneer Least Count: 0.6 kg/cm²
Instrun
Calibr Model: Location of Instrument:

Sr. No. I IDNo.: 1263 Location of Calibration: Permanent Laboratory


Any Other Details: -

Type:

Environmental Condition:
Temperature (°C ) Relative Humidity (%RH)
23 ± 1.5 50 + 10
Calibration Procedure No. :YUCP/18 Based on DKD R-6-1:2016
Environ Discipline : Mechanical, Group: Pressure Indicating Device
Testind
Master Equipment used for Calibration (National Traceability):
Instrument Des. Sr.No. / L.D.No. Certificate No. Due Date Cal. Agency
Digital Pressure Gauge -1 to 350 Bar VET05065 / YL/PGM/01 ETPL/N/2303007/01. 06/03/2024 CC-3065

CALIBRATION RESULTS:
UUC gauge UUC gauge MIW Deviation Hysteresi
Std. up Std. Down Std. up (Pactual Repeatability
Water Tesing reading reading
(M1) (M2) (M3) (M1+M3) (M3-M1) s(M2-M1)
Pactual Pactual 2+M2)/2 MIW) (bar)
(bar) (bar) (bar) (bar) (bar) (bar)
(kg/cm) (bar)
0.0 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00
|0
Building 10.0 9.80 9.81 9.82 9.84 9.82 -0.02 0.03 0.01
Material Tefung 20.0 19.61 19.63 19.59 19.64 19.61 0.00 0.01 -0.04
30.0 29.42 29.46 29.42 29.45 29.44 -0.02 -0.01 -0.04

40.0 39.23 39.26 39.27 39.28 39.27 -0.04 0.02 0.01

50.0 49.03 49.01 49.04 49.06 49.04 -0.01 0.05 0.03


70.0 68.64 68.61 68.63 68.66 68.63 0.01 0.05 0.02

Soil, Miner Maximum Expanded Uncertainty: +0.84 bar, Calculated At confidence level of approximately 95.45 %and k= 2
Agro Test Conformity Statement-NA
UUC = Unit Under Calibration.
Note-(1) Certiffed copies of traceability can be enclosed, on customers request if required. (2) The certificate pertains only to the items calibrated. (3) This
Certifcate should not be reproduced except in full without a written permission of the laboratory. (4) The reported results are valid at the time of performance
of calibration under specified conditions only.
Calibrated by Reviewed &Authorized by

Plastc, Rubber &MitshGajar **End of Certificate**


Haresh Kaneriya
Wooden Testing
Page 1 of 1