8928 Operation Manual
8928 Operation Manual
Model:8928
INTRODUCTION
Your digital sound level meter
provides automatic or manual ranging
4 measurement ranges from 40 to
130dB, and features 0.1dB resolution.
BUTTON DESCRIPTION
ON/OFF: Power on or power off the
meter.
RANGE: Auto range/ Manual range
select key.
RECORD: Record Maximum and
Minimum sound level
measurement.
WEIGHTING A/C:
A weighting and C
weighting select
key.
FAST/SLOW:
Response select key.
MAX HOLD:
Freeze Maximum
sound level reading.
1
MEASURING SOUND LEVELS
Sound levels are displayed both
digitally and in a bar graph.
The digital display is updated every
160ms, while the bar graph is updated
every 40ms.
1. Press the ON/OFF key to turn the
meter on. The unit will first display
the full screen and "18:18.8" then
count down from 99.9 to zero.
The meter will now begin measuring
the current sound levels.
SPL (Sound Pressure Levels)
appears on the left side."A'',"dB"on
the right side of screen.
Point the microphone toward the
source of the sound to be
measured.
REC MAX HOLD MIN
AUTO FAST
MANU SLOW
SPL A C
Diagram A BAT dB
40
+0 +10 +20 +30
Diagram B SPL
dB
90
+0 +10 +20 +30
AUTO FAST
Diagram C SPL A
dB
40
+0 +10 +20 +30
AUTO
SLOW
Diagram D SPL A
dB
40
+0 +10 +20 +30
Diagram E SPL C
dB
40
+0 +10 +20 +30
4
MAX HOLD
AUTO FAST
Diagram F SPL A
dB
40
+0 +10 +20 +30
SPL A
Diagram G dB
40
+0 +10 +20 +30
REC MIN
AUTO FAST
Diagram H SPL A
dB
40
+0 +10 +20 +30
5
4. Press the RECORD key again.
MAX will appear on screen side by
the "REC" and the maximum sound
level measurement will be
displayed. The unit is not recording
at this time, but th bar graph will
continue to show the current
reading.
REC MAX
AUTO FAST
Diagram I SPL A C
dB
40
+0 +10 +20 +30
AUTO
SLOW
Diagram J SPL C
dB
40
+0 +10 +20 +30
FAST
MANU
Diagram K SPL A
dB
40
+0 +10 +20 +30
7
If the meter is operating in manual
range and "LO" is displayed, the
sound is too low or the range.
If "HI" is displayed, the sound is too
loud.
In either case, you must adjust the
measurement range or your readings
will be inaccurate.
FAST
MANU
Diagram L SPL A
dB
100
+0 +10 +20 +30
FAST
MANU
Diagram M SPL A
dB
60
+0 +10 +20 +30
AUTOMATIC SHUTOFF
The meter will turn off automatically
after 20 minutes to preserve the
battery.
To override this feature:
1. Make sure the unit is turned off.
2. Press the ON/OFF and MAX HOLD
buttons simultaneously.
3. When the full display appears,
release the MAX HOLD button first,
" " will replaced full display.
8
4. Release the ON/OFF key.
The meter will remain on until the
ON/OFF button is pressed again.
The automatic shutoff feature will
resume the next time the meter is
turned on.
Diagram N
CALIBRATION
Using a standard Acoustic Calibrator
(Recommend B&K type 4231 model)
which generate "94dB" output.
1) Set whichever time weighting :
Fast or Slow.
2) Set sound level range at:80~110
dB.
3) Select weighting A or C .
4) Max.Hold function measurement
mode unenabled
5) Better to calibrate under 60dB
sound environment.
9
REPLACING THE BATTERY
When the screen display " BAT "down
at the left corner , the 9V battery has
fallen to a critically low voltage level
and should be replaced as soon as
possible. Use a screwdriver to
unscrew the back battery
compartment cover. Insert a fresh 9V
battery and screw the cover.
AUTO FAST
1
SPL A
BAT dB 2
40 3
+0 +10 +20 +30
Diagram O 4
Hygrometer/Psychrometer
Thermometer
Anemometer
Sound Level Meter
Air Flow meter
Infrared Thermometer
K type Thermometer
K.J.T. type Thermometer
K.J.T.R.S.E. type Thermometer
pH Meter
Conductivity Meter
T.D.S. Meter
D.O. Meter
Saccharimeter
Manometer
Tacho Meter
Lux / Light Meter
Moisture Meter
Data logger
Temp./RH transmitter
Wireless Transmitter ..........