MINDRAY MEDICAL INDONESIA

Ventilator Machine
SV 800 Standard without turbine
Technical Specification
Dimensions (HxWxD) : 1425mmX530mmX647mm
: (Including the trolley,including backup air supply)
Weight : Approximately 48kg
: (including the trolley,including backup air supply)
Screen : 18.5” Color active matrix TFT touch screen
: Brightness Adjustable
Trolley :
Dimensions : 687mm(L) X534mm(W) X 980mm(H)
Weight : 16.3 kg
Communication interface RS-232, Nurse call connector, VGA connector, USB PortX4,
Communication interface :
Ethernet

Ventilation Specifications :
Patient Type : Adult, Pediatric, Neonate
V-A/C, P-A/C, V-SIMV, P-SIMV, Duolevel, VS, APRV, PRVC, PRVC-SIMV, AMV, CPRV,
Ventilation Mode :
PSV-S/T, nCPAP, NIV, Apnea Ventilation
Controlled Parameters
O2% : 21 to 100 vol.%
TV (Tidal Volume) : 2 ml - 4000 ml
I:E : 1:10 to 4:1
Thigh : 0.10 to 30.00 s
Tlow : 0.20 to 30.00 s
Tpause : OFF, 5% to 60%
Flow : 2 to 180 l/m
∆Pinsp : 1 to 100 cmH2O
∆Psupp : 0 to 100 cmH2O
Phigh : 0 to 100 cmH2O
Plow : 0 to 50 cmH2O
PEEP : 0 to 50 cmH2O
Flow trigger : OFF,
: Adult/Pediatric: 0.5 to 20.0 L/min;
: Neonate: 0.1 to 5.0 L/min
Pressure trigger : OFF, -20.0 to -0.5 cmH2O
: Exp% (Expiration termination level)Auto, 5% to 85%
Sigh :
sigh : Sigh SwitchON, OFF
Interval : Interval20 s to 180 min
Cycle Sigh : Cycles Sigh1 to 20
Automatic Tube Resistance Compensation
Tube Type : ET Tube, Trach Tube, Disable ATRC
Tube I.D : .Adult: 5.0 to 12.0 mm Pediatric: 2.5 to 8.0 mm
: Neonate: 2.5 to 5.0 mm
Compensate : 1 to 100 % Expiration Compensation Switch
: ON, Off
O2 Therapy
O2% : 21 to 100 vol.%
Flow : Adult: 2 to 60 L/min
: Pediatric: 2 to 25 L/min
Monitored parameters
Airway pressure range : Ppeak, Pplat, Pmean,
Tidal volume range : TVi, TVe, TVe spn
Frequency range : fTotal, fmand, fspn
Minute volume range : MVi, MVe, MVspn, MVleak,
Leak% : 0 to 100%
Resistance : Rinsp, Rexp, (Range 0 to 600 cmH₂O/L/s)
Compliance : Cstat, Cdyn, (Range 0 to 300 mL/cmH₂O) Inspired Oxygen (FiO2)15 to 100 vol.%
RSBI : 0 to 9999 1/(min*L)
WOB : WOBtot,WOBvent, WOBimp, WOBpat
P0.1 : (-20 to 0 cmH₂O )
NIF : (-45 to 0 cmH₂O)
PEEPi : 0 to 120 cmH₂O
Vtrap : 0 to 4000 mL
RCexp : 0 to 10 s
TVe/IBW : 0 to 50 mL/kg
I :E : 150:1 to1:150
Tinsp : 0.00 to 60.00s
Waveforms : Airway pressure-time, Flow-time, Volume- time, CO2-time , Pleth-time
Loop : Paw-Volume, Flow-Volume, Paw-Flow,
: Volume-CO2
Alarm settings
Tidal volume High : Neo: Off, 3 to 200 mL Ped: Off, 25 to 600 mL Adu: Off,110 to 6000 mL
Low : Neo: Off, 1 to 195 mL Ped: Off, 10 to 595 mL, Adu: Off, 50 to 5995 ml

Minute Volume High : Neo: 0.02 to 30.0 L/min


: (can be set to Off in nCPAP)
: Ped: 0.2 to 60.0 L/min
: Adu: 0.2 to 100.0 L/min
Low : Neo: 0.01 to 15 L/min
: Ped: 0.1 to 30.0 L/min
: Adu: 0.1 to 50.0 L/min (can be set to Off in NIV)
Airway pressure : High 10 to 105 cmH2O
: Low OFF, 1 to 100 cmH2O
Frequency : High OFF, 2 to 160 /min
: Low OFF, 1 to 159 /min
Inspired Oxygen (FiO2)
High : FiO2 exceeds the alarm limit for at least
30 s, internal alarm limit: set value+max (7 vol.% or set value X10% ) or 100 vol.%,
:
whichever is lower.
Low : FiO2 lower than the alarm limit for at least 30 s, internal alarm limit: set
: value-max (7 vol.% or set valueX10%) or 18%, whichever is greater.
Apnea alarm time : Low5 to 60 s (can be set to Off in nCPAP)
Low battery voltage Gas supply pressure low Airway obstruction Tube disconnected
Other Alarms :
PEEP too high
Trend
Type : Tabular, Graphic
Length : 96 hours
Content : Monitor Parameters, Setting Parameters
Log
Type : Alarm, Operation
Max number : 5000

Ventilator components
O2 sensor : Type paramagnetic sensor
MainStream CO2 Module : Displayed numericsEtCO₂, VeCO2, ViCO2, MVCO2, Vtalv, MValv,
: VDaw, VDaw/TVe, SlopeCO2, VDalv, VDphy, VDphy/TVe, OI, P/F, VCO2
SpO2 module : Displayed numericSpO2, PR, PI

Gas supply : Gas typeO2 and Air


: Pipe ConnectorNIST, DISS
Power and Battery Backup : 180 min (Powered by two new fully-charged battery in standard working condition)

Special Functions and procedures : 100% O2


: Suction, Nebulization, Manual breath, Inspiratory hold, Expiratory hold, PEEPi
: P0.1, NIF, RSBI
: PV-Tool, SBT
: Auxiliary Pressure Monitor
: Lung Recruitment Tool (SI)
Accessories
1 pcs : Humidifier
1 pcs : Troly
1 pcs : Breathing circuit reus child
1 pcs : Breathing circuit reus adult
1 pcs : Breathing circuit disposible adult
1 pcs : Breathing circuit disposible child
1 pcs : Arm
1 pcs : O2 hose
5 pcs : Bactertial filter
1 pcs : NIV Mask size L
1 pcs : NIV Mask size M
1 pcs : NIV Mask size S
1 pcs : Test lung adult
1 pcs : Test lung infant
1 pcs : Nebuliser acc
1 pcs : Oxygen Therapi acc S, M, L
1 pcs : SPO2 accessory kit
1 pcs : CO2 accessory Kit
1 pcs : Ncpap accessories kit