Awa A56010 Voltohmyst
Awa A56010 Voltohmyst
47 York Street,
SYDNEY
1-56010R (i)
INDEX
1. DESCRIPTION:
1.1 Application 1
1.2 Mechanical Construction 3
1.3 Design Characteristics and Circuit Description 3
1.4 Summary of Controls, Facilities and Electrical
Characteristics 6
1.4.1 Facilities 6
1.4.2 Controls and Input Connections 7
1.4.3 Electrical Characteristics 7
1.5 Probes and Cables used with the Voltohmyst A56010 9
1.5.1 Cables and Probes Supplied with the Instrument 9
1.5.2 Probes Not Normally Supplied with the Instrument. 10
2. METHOD OF USE:
2.1 Warning Notes 13
2.2 Setting up Procedures 14
2.3 D.C. Voltage Measurement 14
(using D.C. and Direct Probes)
2.4 Centre-zero Readings 15
2.5 A.C. Voltage Measurement
(With Direct probe alone) 15
2.6 Resistance Measurement 16
2.6.1 Normal Measurement of Resistance
up to 1,000 M.ohm 16
2.6.2 High resistance (insulation) Measurement 16
2.7 Current Measurement 17
2.8 Measurement and Calculation of dbm 17
3. MAINTENANCE:
3.1 Valve Replacement 19
3.2 Checking the Condition of the 1.5V cell 19
3.3 Electrical Balance Check 19
3.4 Calibration 20
3.4.1 D.C. Calibration 20
3.4.2 A.C. Calibration 21
3.4.3 A.C. Compensation Adjustment (C2) 21
4. COMPONENT SCHEDULE: 23
5. DIAGRAMS:
Voltohmyst A56010 Drg. No. Fig. No.
Circuit 5601002 7
Probes and Cables used with Voltohmyst 56010D7 8
1.
DESCRIPTION
Application.
A.W.A. Voltohmyst type A56010 is a versatile instrument which has application in the
measurement of positive and negative D.C. voltages, D.C. current, resistance and A.C. voltages.
The instrument features a circuit which measures directly the peak-to-peak values of
sinusoidal and complex waveforms, the scales being calibrated directly in peak-to-peak values
and also in R.M.S. for sinusoidal waveforms.
The instrument has a rated input for non-symmetrical wave-forms of 2,100V, but
sinusoidal and symmetrical waveforms of up to 4,200V can be measured. (Refer to figure 1
below for waveforms measurable with the instrument.)
When measuring voltages, with a source impedance of approximately 100 ohms, the A.C.
ranges up to and including 0-500V are frequency compensated to provide a flat characteristic
from 30 cycles to 2.5 Mc. When used in conjunction with a crystal diode probe such as A.W.A.
type 2R56020, the input capacitance is reduced to a low value and the A.C. voltage ranges are
made accurate to within 10% from 50 kc. to 250 Mc. (Refer to subsection 1.4.3 for frequency
response figures.)
When using an associated D.C. probe, on all D.C. ranges, the instrument has an input
resistance of 11 M.ohms, thus offering considerable advantages over conventional test
equipment in a number of applications such as those listed below:-
2.
The instrument, when connected directly to the grid, measures this bias, with
negligible effect on the conditions of oscillation.
The peak-to-peak values of pulse waveforms may be measured, using the A.C.
peak-to-peak scale of the instrument. A long time-constant is incorporated in
the circuits to allow for measurement of signals having low repetition rates.
The instrument is housed in a compact metal case, having meter, four controls and
three terminals mounted symmetrically on the front. An engraved front panel is fitted,
and a carrying handle provided at the top of the case.
The circuit is designed to give a high degree of accuracy in all applications, making
the instrument especially adaptable for work in television and industrial equipment.
In the case of sine waves, the instrument is calibrated directly for R.M.S. and peak-
to-peak values (a feature made practicable by the constant ratio of 2.83:1 of these
values). The peak-to-peak-reading feature has particular application in servicing
television receivers when taking measurements of wave-forms as indicated in Fig.
It should be noted that in this instrument, as in all voltage measuring devices, certain
precautions are necessary in interpreting the readings, and inaccuracies may be due to
external conditions such as high impedance of the circuit being measured, high-
frequency voltage source, or if the voltage waveform consists of short pulses with a
low repetition rate. The error introduced increases with the ratio of the source
4
impedance to the impedance of the instrument, and varies with frequency. Refer to para. 1.4.3
(c). The error in short-pulse measurement when the repetition rate is low is due to the time
constants involving the capacitors in the peak-to-peak rectifier in the instrument. These
capacitors do not have time to become fully charged during the pulse, and in the relatively long
period between pulses, are unable to retain the charge they have received. The range or
conditions allowing satisfactory measurement of essentially rectangular pulses derived from a 50
ohms source indicated in figure 2 below. For sources of higher impedance error will be greater
than 2%.
The instrument is frequency-compensated for A.C. voltage ranges up to and including 0-500V.
On the higher range the response varies with frequency.
All resistors in the divider networks associated with the voltage and ohms applications are high
stability carbon types, having a tolerance of 1%, and in conjunction with the D.C. amplifier,
ensure reliable service for a considerable period.
The fundamental element of the circuit is a D.C. bridge involving type l2AU7 valve (V3) and its
associated network. - Refer to the simplified schematic in fig. 3 below.
5.
Fig. 3
Assuming the instrument is switched on, power is supplied from the power transformer (T1),
rectified by one half of a type 6AL5 valve (V2) and applied to the bridge (V3).
Providing no readings are being taken, and the circuit has been previously balanced by
adjustment to the balance resistor (R25), the cathode currents of the two triode sections of the
valve (V3) are equal, and the microammeter connected between the anodes reads zero.
When the facility selector (s2) is set in any of the voltage or ohms positions, and a reading is
taken, a voltage is applied via range selectors, etc., to the grid of the left-hand triode (as drawn
on the simplified schematic) section of V3, this altering the current through this section and in
turn changing the potential across the common cathode resistor (R25). The change in the latter
voltage then acts as a reverse change in the bias on the second section, and the meter responds to
the resulting changes in potential at the anodes due to the voltage drops across anode resistors
R20 and R22, respectively.
When used as a milliammeter, the facility selector (S2) disconnects the meter from the electronic
circuit, and connects the "+" terminal of the meter to the "+" mA-OHMS input terminal, and the
"-" terminal to the "-" mA. EARTH input, which is then disconnected from
6
earth. Operation is then as for a conventional milliammeter, but it should be noted that all
electronic protection for the meter is now out of circuit and care must be taken not to overload.
1.4.1 Facilities
Ranges
"+" D.C. Volts ) 0 – 1.5V. (On separate LO Scale for A.C.)
)
"-" D.C. Volts ) 0 – 5V.
) 0 –15V.
A.C. (R.M.S. volts ) 0 – 50V.
) 0 – 150V.
0 – 500V.
0 – 1,500V.
0 – 1.5 mA..
0 – 5 mA.
0 – 15 mA.
0 – 50 mA.
0 – 150 mA.
0 – 500mA.
0 – 1,500mA.
.
Ohms Measurement:- Ranges
(meter calibrated Rxl
0-1,000 ohms) Rx 10
Rx 100
Rx 1000
Rx 10,000
Rx 100,000
Rx 1 M,ohm
7.
(i) Facility Selector: Selects desired facility, i.e., OHMS, A.C., etc.
(ii) Range Selector: Selects desired range, i.e., l.5V., 5V., etc.
(iii) ZERO ADJ. Positions the meter needle to "0" or "-0+" (the
latter when the facility selector is set to "+"
VOLTS position, and the instrument is used for
centre-zero indication).
(iv) OHMS ADJ. Positions the meter needle at the extreme right-
hand mark of the ohms scale when the facility
selector is set to OHMS position.
(v) VOLTS, A.C.-D.C. Positions the meter needle at the extreme right-
hand mark of the ohms scale when the facility
selector is set to OHMS position.
(vi) "+" mA. – OHMS Positive input terminal for current readings.
Also associated with ohms facility.
(vii) -mA. EARTH. "-" terminal for current readings; earthed for
other facilities. Second terminal for ohms
measurement.
(b) Milliammeter
Accuracy on 3% of F.S.D.
all ranges
8.
Frequency Response
(using direct probe and cable):-
Frequency Response
Using crystal Within 1db from 50 kc. To 250 Mc.
Diode probe type
Accuracy:-
Peak-to-peak
Complex waveforms 2,l00V.
When measuring
A.C. component:- 2,100V.
D.C. component:- 1,500V
The instrument is normally supplied fitted with three probes (ohms probe,
direct probe type R56020, and D.C. probe type 1R56020), together with an earth
cable which is fitted with an alligator clip.
10.
The ohms probe and direct probe are each connected to individual cables, and are
thus used alone for the measurement of resistance and A.C. voltages respectively; the
circuit is completed in each case via the earth cable and clip.
The D.C. probe, however, consists of an extension piece which fits over the direct
probe and clamps into place on the latter. The whole is then used for measurement of
D.C. voltages. An alligator clip to fit onto the end of the direct probe is supplied.
1.5.2 Probes Not Normally Supplied with the Instrument ( available on separate order)
This probe consists of an extension piece similar to the D.C. probe, with a clip attached
to the side, and contains a small diode rectifier element, coupling capacitor and isolating
resistor. (Refer to the insert on the circuit diagram Drg. 5601002).
The probe may be used in the presence of D.C. voltages as high as 250V.
Input Resistance
Frequency Resistance
50 kc. 370,000 ohms
100 kc. 360,000 ohms
1 Mc. 312,000 ohms
10 Mc. 283,000 ohms
100. Mc. 90,000 ohms
220 Mc. 6,000 ohms
11.
Method of Use
(i) With the probe (type 2R56020 over the tip of the direct probe. Set the facility
selector on the Voltohmyst to "-" VOLTS position and the range selector to an
appropriate scale.
(ji) Connect the clip to the earth side of the voltage source and apply the probe to the
other side.
This probe consists of cable and hollow-molded piece containing a dropping resistor
capable of extending the D.C range of the instrument to 0-30,000V., and is indispensable in
television and heavy-duty industrial work.
The maximum input voltage usable with the probe is 30,000V., and since the probe
effectively provides the D.C. ranges of the instrument with a multiplying factor of 100, the range
selector may be set as required and a direct reading obtained. Note that the total input resistance
of the Voltohmyst when using the high voltage probe in the vicinity of 1100 M. ohms, providing
a valuable facility when working in high impedance circuits with poor regulation.
12.
Method of Use
Remove the direct probe and cable from the instrument and fit the high voltage probe and cable
in its place. Set the facility selector to "+ or "-" VOLTS as required, and the range selector to a
position considerably higher than the estimated value of voltage source. More accurate
readings may be made with successive settings of the range selector after taking the first trial
measurement.
IMPORTANT;
The probe is provided with an earthed guard ring in front of the hand-piece, and it is
important that the operator's hand be behind this guard ring at all times, so that in the
event of a flash-over occurring, current will pass only to the latter.
13.
2. METHOD OF USE
(The frame of the instrument is earthed via the 3-core mains cable, and
application of the test prods in incorrect polarity would therefore short-circuit
the supply.)
(ii) Before taking any voltage measurements, refer to the maximum input voltage
ratings given in para. 1.4.3. On no account should these values be exceeded.
(iii) Do not connect the EARTH cable to a point of high potential. (The connector to
which this cable is attached is connected to the instrument case when measuring
volts.) Reversal of D.C. polarity is obtained by switching between ~'+" and "-"
VOLTS positions on the facility switch.
(iv) When measuring the resistance of certain devices such as thermocouples, meters,
etc., a higher range than R x 1 or R x l0 should be used, because on these 1ower
ranges the instrument applies up to l.5V. (depending on the resistance of the
device being measured) in series with the external circuit, and could cause
damage to the thermocouple or meter.
(v) Exercise care when using the instrument as a milliammeter. The electronic
protection devices incorporated for voltage applications are switched out of
circuit when the facility selector is set to "mA." position.
(vi) Exercise extreme care when working on equipment in which high voltages
appear. Where possible it is advisable to follow the precautions listed below.
(a) Remove the power from the equipment under test before connecting the
instrument.
(f) Work in the presence of a second person who can render assistance in the
case of accident.
(ii) Identify the EARTH and OHMS cable and the Direct Probe (refer to Fig. 8) and
connect respectively to EARTH, OHMS and VOLTS (A.C.-D.C.) connectors on
the front panel.
(iii) Check the mechanical zero of the meter, and if necessary, re-adjust. Plug into
the mains and switch on. Allow a short warm-up period. Set facility selector to
"+" VOLTS position.
(v) Turn the ZERO ADJ. control until the meter needle coincides with the left-hand
"0" on the scale. Check the adjustment by now setting the facility switch to "-"
VOLTS position. If the meter needle moves from "0" again make an adjustment
to. the ZERO ADJ. control. If necessary, make alternate adjustments to the
facility switch and ZERO ADJ. control until a setting is obtained which allows
the facility switch to be set to either "+" or "-" VOLTS without affecting the
needle.
(v) Set the facility selector to the OHMS POSITION. The needle should be
deflected to approximately full scale. Turn the OHMS ADJ. control until the
needle coincides with the right hand edge of the OHMS scale.
Use the D.C. probe (type IR56020, indicated on Fig. 8) in conjunction with the direct
probe R56020 for all D.C. voltage measurements. Fit the probe piece over the tip of
the direct probe and clip into place.
(i) Set the facility selector to "+" VOLTS or "-" VOLTS as required (refer to
warning note (iii) in 2.l).
15.
(ii) Connect the EARTH cable to earth or low side of the voltage source under
measurement.
(iii) Set the range selector to a position considerably higher than that at first required.
Use the probe as a conventional multimeter test prod. Switch to progressively
lower settings of the range selector to obtain more readable deflections.
(ii) Turn the ZERO ADJ. control until the meter needle coincides
with the centre "0" of the scale.
(iii) Set the range selector to a position at least double the voltage to
be measured.
(iv) Take several test readings, re-setting the range selector for more
convenient readings.
Use the direct probe (type R56020) alone. Low values of R.M.S. (below 1.5V.) and
peak-to-peak (below 4. 2V.) should be read on the separate LO-A.C. scales.
Higher voltages should be read on the "D.C. or R. M. S." scale graduated to 1.5 or 5
respectively, in association with the appropriate multiplying factor given by the range
selector. Peak-to-peak values can be read simultaneously with the R.M.S. values.
(i) Set the facility selector to A.C. VOLTS position and if necessary, adjust
the ZERO ADJ. control to make the needle coincide with the left-hand
"0" of the scale.
(ii) Set the range selector to a position considerably higher than that at first
required. Use the probe as a conventional test prod. Switch to
progressively lower settings of the range selector to obtain more
convenient readings. Read direct from the appropriate scales.
Note that a delay will occur between the removal of the test prods from the source
voltage and the return of the meter needle to zero when measuring A.C. voltage.
16.
This delay is caused by the effects of circuit constants within the instrument,
such constants having been chosen to enable more accurate measurements of
pulse waveforms having low repetition rate.
(vi) Multiply the reading on the ohms scale by the factor against the
position of the range selector.
The instrument has application as a high resistance (above 1,000 M.ohms) test
unit when used in conjunction with an external D.C. voltage (between 20V.
and 500V.), in a circuit as indicated in Fig. 4 below.
17.
(i) Set the facility selector to "+" VOLTS, and measure the volts
at point "A". Let this applied voltage be "V".
(ii) Measure the voltage at point "B". Let this voltage be "v".
Connect the OHMS cable in the "+ side. Set the range selector to a considerably
higher position than that at first required, then make progressive adjustments for
more convenient readings. Read direct from the appropriate scale.
The chart given in Fig. 5 can be used in conjunction with the instrument to determine
dbm values corresponding to R.M..S. A.C. voltage values across a 600-ohm load.
The dbm is defined as the number of decibels above or below a reference level of 1
milliwatt in 600 ohms at 1000 cycles. Accordingly, 0 dbm indicates a power of
1mW.; 10 dbm, 10 mW.; 20 dbm, 100 mW. etc.
Since dbm are defined with respect to a 600-ohm circuit, the power levels in
these circuits correspond to voltage values, thus allowing dbm to be measured in
terms of R.M.S. voltages. The chart given in Fig. 5 is drawn for application to 600
ohm circuits, but may be used to convert R.M.S. to dbm values in load resistances
other than 600 ohm, by the algebraic addition of a suitable increment (ΔDBM).
Several values of this increment have been worked out resistances commonly
encountered, and are given in the table below. For other values of ΔDBM, use
the formula given in equation (i).
18.
Resistive load at ΔDBM (to be added to
1000 cycles the dbm value given on
the chart for 600 ohms)
600 ohms 0
500 ohms +0.8
300 ohms +3.0
250 ohms +3.8
150 ohms +6.0
50 ohms +10.8
15 ohms +16.0
8 ohms +18.8
3.2 ohms +22.7
Note that for load values greater than 600 ohms, ΔDBM will be a negative quantity, as
indicated in the formula:-
19.
3. MAINTENANCE
3.1 Valve Replacement
Run the valve with heater voltage only applied for a period of 48 hours before
placing it in the instrument.
Connect together the grids and cathode, apply the normal heater voltage and,
in addition, an A.C. voltage of approximately 100V., 50-cycle between anode
and cathode. Allow to run for 24 hours.
(iii) Rotate the OHMS ADJ. control for full-scale deflection on the meter.
(IV) Short together the ohms and earth cables for about ten seconds.
(v) Open the above short-circuit and observe the meter needle. If there is
appreciable deviation from F.S.D., a weak cell is indicated and should be
replaced.
(ii) Rotate the ZERO ADJ. control and check that it is possible to set the meter
needle at 60% of F.S.D. on any range.
(iii) Set the facility selector to "-" VOLTS, and again turn the ZERO ADJ. control.
It should be possible to set the meter needle at 10% of F.S.D. on any range.
20.
3.4 Calibration
Fig. 6 indicates the relevant positions of the adjustable controls used in the
calibration.
Note that the accuracy of calibration can only be comparable with the
accuracy of the voltage standard used in the procedure.
(i) Check the mechanical zero of the meter, and if necessary, re-adjust
(ii) Set the facility selector to "+" VOLTS. Switch on the instrument and allow a
warm-up period. Check the mains voltage. The instrument should be
calibrated when operating from a steady 240V, 50-cycle supply.
21.
(iii) Using the ZERO ADJ. control, set the meter needle to "0", and leave the
control set at this position throughout the remainder of the procedure.
(iv) Set the range selector to the 0-50V range.
(v) Connect the earth cable to the negative terminal of an accurately adjusted
50V. D.C. source, and the D.C. probe (fitted over the direct probe) to the
positive terminal of this source.
(vi) Adjust the "+" D.C. Cal. Control (R27) to bring the pointer exactly to the
50V. mark on the "D.C. or R.M.S." scale.
(vii) Leaving the other conditions set, reverse the connections to the 50V. source
and set the facility selector to the "-" VOLTS position.
(viii) Check the other D.C. ranges against reliable voltage sources, making
allowances for the tolerance given in the specification. If necessary re-
calibrate on a range other than 0-50V., and check all components associated
with the facility.
(ii) Remove the D.C. probe from the direct probe. Set the facility selector to A.C.
VOLTS, and range selector to 0-1.5V.
(iii) Short together the direct probe and earth cable, and adjust the A.C. ZERO
control (R37) for a zero on the meter. If it is not possible to zero the needle,
interchange the two type 6AL5 valves, and if a zero is still unobtainable,
replace one or both of the latter valves.
(v) Connect the earth cable and tip of the direct probe to an accurate 50V. , 50
cycle source, and check that the reading is within the tolerance stated in the
electrical specification. If outside this tolerance, check the components
associated with the facility.
(i) Set the facility selector to A.C. VOLTS. Check that the meter is positioned at
zero, firstly by setting the mechanical zero, and later (after switching on) by
adjusting the ZERO ADJ. control.
22.
(ii) Set the range selector to 0-500V and apply an accurate 150V., 100kc. Signal
to the instrument using the direct probe (alone) and earth cable.
(iii) Adjust the variable air trimmer (C2) at the rear of the instrument for a 150V.
reading on the "D.C. or R.M.S." scale of the meter.
23.
4. COMPONENT SCHEDULE.
When ordering replacement parts please quote ALL details given below for a
particular component.
(a) Capacitors
(b) Connectors
(c) Meter
(d) Resistors
24.
(f) Switches
(g) Transformer