A08!80!034 O2000 Operator Manual
A08!80!034 O2000 Operator Manual
To calibrate the unit, pass a gas of known concentration through the sensor. O2000R
If a known gas is not available, the unit can be calibrated to the Oxygen %
value of Air, which is 20.9%.
PORTABLE
With the known value of gas present at the sensor, press the Calibration button OXYGEN MONITOR
( Cal ), the mnemonic ( Cal ) will be displayed followed by the existing OPERATOR MANUAL
calibration value. Press and hold the Cal button until the programming
mnemonic is displayed ( pro9 ), alter the calibration value using the up (▲) or
down (▼) arrow buttons ( if a change is required ) when the correct calibration
value is displayed press the Cal button again to set this value into memory, the
done mnemonic will again be displayed, to indicate the setting has been
accepted into memory.
Sensor Replacement.
S.A.T. SYSTEMS.
The oxygen sensor, is located inside the monitor case. As this is a galvanic KIRK MICHAEL.ISLE OF MAN,IM6 1SS
type sensor it will require periodic replacement, which is dependant upon its Email : [email protected]
exposure to oxygen in %hrs. The replacement sensor part number is D-05 Tel : 01624 878673
To replace the sensor, unscrew the external flow adapter (and retain), remove Fax : 01624 878674
the connector from the rear of the sensor and unscrew the sensor from the
case. Fit the replacement sensor to the case and replace the flow through
adapter. After replacement of a sensor the monitor will require re-calibration. Feb 05 O2000 Mk 2. Instruction Manual A08-80-03 Issue 4.
MODEL O2000 PORTABLE OXYGEN MONITOR SPECIFICATIONS
The SAT Systems O2000 Portable Oxygen monitor, is a compact, battery Range : 0 - 100 %
powered unit with user set audible / visual alarms. The unit uses a galvanic Resolution : 0.1 %
type sensor which outputs a mV signal in direct proportion to the oxygen Battery : 9 V Rechargeable NiMH battery pack
concentration within the sample gas. Size : 120 high x 80 wide x 80 deep
Weight : 580 grams (Inc. Batteries and Sensor ).
On / Off.
The unit is powered by pressing the On / Off button on the keypad once,
KEY PAD, BUTTON FUNCTIONS this activates a self test facility, whilst the unit is being tested the display
shows the mnemonic 02000.
USER LEGEND AREA
Alarm Function.
The unit features an audible / visual alarm that is active when the display
reading either exceeds the HiAL setting or is less than the LoAL setting.
NUMERIC
INCREASE
SET / READ When an alarm condition is active the audible warning may be silenced by
HIGH ALARM
pressing the Mute button. The visual alarm, which is displayed as either HiAl
O2 000 % Oxygen
or LoAl will alternate with the current reading until the gas sample is back
within the range of the alarm set points.
SET / READ
ON / OFF l CALIBRATION Alarm Settings.
VALUE