Philips Power Distribution Unit Level 0 Documentation
Philips Power Distribution Unit Level 0 Documentation
Level 0 Documentation
333H97
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems (”Philips”) and
may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written
permission of the Philips Legal Department. This document is intended to be (a.) used by customers and is licensed to them as part of
their Philips equipment purchase or (b.) used to meet regulatory commitments as required by the FDA under 21 CFR 1020.30 (and any
amendments to it) and other local regulatory requirements. Use of this document by unauthorized persons is strictly prohibited.
File: PDU_OPTIMUS_RF_03241_AA
List of pages and drawings (LOPAD) Manual Order No: 4512 988 03241 AA
released: 02/2008
1
3 ...4 LOPAD
5 ...6 Contents
7...11 (08.0)
Blank page
Contents
TEXT
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1. Application / General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3. Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4. Installation (only when the PDU is retrofitted) . . . . . . . . . . . . . . . . . . . . . . . . 5
5. Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
6. Setting-to-work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
6.1. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
DRAWINGS
Power distribution unit: Dimensions and weight . . . . . . . . . . . . . . . . . . . . . . . . . . Z−1.4
Connection of generator with Power distribution unit . . . . . . . . . . . . . . . . . . . . . . Z−6.2
Labelling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Z−10
Blank page
1. Application / General
The power distribution unit (PDU) is needed for mains adaptation of various X-ray systems in combination with the
OPTIMUS R/F generator.
The output voltages of the PDU are needed for the X-ray system and for parts of the generator. The voltage supply
of the converter part of the generator does not depend on the PDU.
In addition to the current supply elements the PDU also contains the electrical function of the surge arrester.
2. Compatibility
3. Technical data
Usually the PDU is placed against the wall on the left of generator cabinet E. The distance between PDU and
generator cabinet E is limited to a maximum of 1.5m by the present cable tails.
Though not required for stability, the PDU may be fixed to the floor with 2 screws if desired by the customer. Also
see Z−1.4.
In seismic areas the PDU can additionally be fixed to the wall with screws through the holes in the frame.
5. Electrical connections
Also see
− Z1−1 Power distribution unit with OPTIMUS R/F
− Z1−2.2 Power supply with mains transformer
Note:
The power distribution unit (PDU) must be integrated in the wiring of the power part. In case the generator has not
yet been prepared for the installation of the PDU, the following wiring modifications must be carried out before the
PDU can be installed:
Connection of the PDU after generator E has been prepared for the installation
PDU Wall junction box Function Remark
MEX
Mains input / primary end of PDU:
L1 L1 Input phase L1 Cable tail of PDU
L2 L2 Input phase L2 Cable tail of PDU
L3 L3 Input phase L3 Cable tail of PDU
Output 4 of the PDU for the geometry segment, 220 / 127V, 16A (already connected in the PDU):
K1:L1 L1 / X3:01 Output geometry Cable tail of PDU
K1:L2 L2 / X3:02 Output geometry Cable tail of PDU
K1:L3 L3 / X3:03 Output geometry Cable tail of PDU
N N / X3:04 Null Cable tail of PDU
PDU Generator E
Mains cable for the generator:
X2 : 01 ENF1 : L1 Input phase L1 for converter Cable tail of generator E
X2 : 02 ENF1 : L2 Input phase L2 for converter Cable tail of generator E
X2 : 03 ENF1 : L3 Input phase L3 for converter Cable tail of generator E
PE PE Protective earth from generator Cable tail of PDU
Output of the generator 1 − 3, 400 / 220V, 10A / secondary end of the PDU:
L1 ENF12 : L1 Output 1 to 3 the generator Cable tail of PDU
L2 ENF12 : L2 Output 1 to 3 the generator Cable tail of PDU
L3 ENF12 : L3 Output 1 to 3 the generator Cable tail of PDU
X2 : 04 ENX 3201 Zero of PDU going to generator Cable tail of generator E
Generator E MEX
PE PE Protective earth for the generator Cable tail of generator E
6. Setting-to-work
6.1. Programming
PDU
Connect the programming links of the mains Programming links
primary winding
transformer according to the rated voltage of the L1
mains (default 480V).
Connect 415V mains systems up to the 400V 400
440
terminal. 460
480
Modify EMC filter EQ 200 in the converter 480V default
L2
assemblies EQ/E2Q if the generator is operated
on a grounded delta mains. 400
440
460
480
L3
400
440
460
480
303H96
L2
L2
L3
L3 EN
N
N
PE
PE
3 phase DELTA,
balanced or floating earth:
Mains transformer is requiered!
47K
13 14 R11
T 3300µ 3300µ 480 ... 750V
L2 X2:02 L2 2 4 3 L2 X1102
2
C2 C12
47K
L2 R1
L2
L3 T
23 24
5
3300µ 3300µ
-
L3 X2:03 L3 3 6 5 L3 X1103
1 6 Discharging of
N L3 33 34
DC circuit after
switch-off of
X1 X2 X3 X21 X22 X23 X12 X14 generator
PE L3 L2 L1 C1 C3 C2 R2
PE PE
C4 C5 C6 2K7
X13
42
K3
WT R1 ENK2
1M
1M
1M
X11
Phase superversion 2K7 41
Power distribution unit R3 R4 R5 C7 C8 C9 C10
X2402
T1
400V
440V
MEX 460V Y- condenser
480V
Thermo
fuse see also Z1-2 xx
"Power supply with mains transformer"
EQ 200
Link Y-capacitor C7:
X Default link: X34 - X44 EMC filter
3 F1 16A
01 L1 K1 T2
X2403
2 1 T1 L1
02 L2
X2401
EQ
4 3 T2 L2 K1 A1
A2 X3201
03 L3 N
6 5 T3 L3 N
L1
fuses L1 Output 5
02
Output 2
03
L2
04...06 II/TV
L2 N GECO
L3 07...09 SYSCO
L3
N
ENX21
PE 01
PE
02
Output 1
03
Connect the link of Y-capacitor (EQ200)
to the grounded phase. 04 Geometry
N
See unit manual "Converter R/F"
05
Schr.
ENF2 10A
L1 T1
L2 T2
L3 T3
K1
2 1
EZ102 low voltage supply
Power distribution unit (PDU)
4 3
EZ119 mA control with OPTIMUS R/F
6 5
WG Tube extension (option)
F1 16A Thermo
fuse X3:3 T1
K1
2 1 T1 L1 H1
230V~ gr
T2 X2403
Output 4 4 3 T2 L2
N N
X3:2
X3201
N
Geometry 6 5 T3 L3
low voltage
control K1
127/220V 16A PE
K1 A2
N
N 8 8 9
A1 K2
K1 K1
X23 F12 10A F2 10A 5 5 12
2x127V 15A X2401 12 9
01
T1 L1 L1 T1
S1
02
EN100
EN
T2 L2 L2 T2
Output 3 Power on circuit 2 3
03
T3 L3 L3 T3
V1
X1:6 X1:7 X1:2 X1:1 X2:1 X2:2 X1:9 X1:5 X1:4
Imaging 04...06
N 1
07...09 F5
6 4 2
K2 2
K11 14
X22 5 3 1
01 K3
11
02 T2 X2403
N
Output 2 Thermo
03 fuse
trafo
X2402
II/TV 04...06 T1
A1 K1 A1
GECO N
SYSCO 07...09 A2 K2 A2
N N
K11
X21 A1 EZ EZ EZ EZ EZ EZ EZ EZ EZ EZ
X13:3 X13:2 X13:1 X47:6 X47:7 X47:2 X47:1 X47:9 X47:5 X47:4
01
A2
AC AC AC EZ
230V 0V 230V X4:1
02 L2 XG L1
L3 L2 L1
Output 1 F3 EM
03 5A off EZX44:14
T3 T2 T1
Geometry 04
EZ
EZX44:6
X4:2
N L3 L2 L1
05 F11 GND RDMNON/ MNON/ PO12V
Reset
PWONNG SIPHIO PWONPO
5A 12V SW/
T3 T2 T1 EZ EZ EZ EZ EZ EZ EZ EZ
WG WG X14:1 X14:2 X15:1 X15:2 X46:15 X46:14 X46:6 X46:7
X1102
X3102 X3101 X3101 51 57
Schr.
L3 N L3 L3 N L3 L2 L1 N L3 N L2 N L1 L1 N L1 N
K1 K1
on
08-01-16
EZ139
YM GS1:1 G200 X2 RoCo LS WG EZ119 EZ102 EWN Adaptation off Central unit EZ130 kV control
M1 RoCo Oil circulation RoCo HS Tube mA low second 24V-/230V~ C
pump extension control voltage low voltage
N H.V. generator (option) supply supply
Desk
A2/A3
M cooler
H.V. generator
(option)
Power supply
see Z1 : H.V. generator R/F with mains transformer
4512 983 05931 CSIP Level 0 (e/08.0) Z1-2.2
c Koninklijke Philips Electronics N.V.
7
3
2
8
2
PHILIPS
PHILIPS
MADE IN GERMANY
3~50/60 Hz
400V 11A
480V 9A
7
C US
type Produktname
8 s/n Fabr. Nr.
code Code Nr.
Schr.
A4 08-03-04
Labelling