Manual Mini-Centrifuga Placas
Manual Mini-Centrifuga Placas
Instruction Manual
Catalog Numbers:
C1000
C1000-100V
C1000-230V
Table of Contents
1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
4.0 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
5.0 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
6.0 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
7.0 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
The electrical warning indicates the presence of a potential hazard which could result in electrical shock.
CAUTION: This symbol refers you to important operating and maintenance (servicing) instructions within the product
Instruction Manual. Failure to heed this information may present a risk of damage or injury to persons or equipment.
This symbol identifies a Protective Earth (PE) terminal, which is provided for connection of the supply system’s protective
earth (green or green/yellow) conductor.
WARNING: This unit is intended for short run usage only, requiring 15-20 seconds to spin down samples. Inadvertent
continuous use could cause sample temperature to rise and the thermal switch to open, which will shut down the motor.
4.0 Specifications
Capacity 2 PCR plates
Dimensions 7.5 x 8.2 x 7.2 in. (19 x 21 x 18.3 cm)
Electrical 100, 120 or 230V, 50/60 Hz
Max Speed 2,500 rpm ±10%
Max RCF 500 x g
Electrical 100V, 50/60Hz, 0.6A (Fuse 1.0 AT)
120V, 50/60Hz, 0.6A (Fuse 1.0 AT)
230V, 50/60Hz, 0.3A (Fuse 0.5 AT)
Instruction Manual | 1
5.0 Installation
The Labnet MPS-1000 Mini Plate Spinner should be placed on a clean, flat surface with sufficient space around it for ventilation. Be
sure that both the bench and the unit’s rubber feet are clean. Plug the power cord into the back of the unit and turn the power switch
knob (on the back of the unit) to the On position.
6.0 Operation
Open button
Proper loading
of a PCR plate
To avoid spillage of samples, only plates that have been heat or tape sealed should be placed in the centrifuge. Plates are loaded into
the rotor in a vertical position.
1. Lift the clear centrifuge cover to gain access to the rotor (press the Open button to release the lid if it is locked down).
2. Using the knob on the top of the unit, turn the rotor until the rotor slot lines up with the opening. Carefully, slide the first plate into
position so that the sealed side faces the center of the rotor and the bottom of the plate faces out. Be sure that the plate is pushed
all the way down into the slot.
3. Line up the second slot like the first, using the knob on the top of the unit.
4. Load the second plate in the same manner as the first, making sure that it is all the way down in the rotor slot. If spinning only
one plate, it is extremely important to load a dummy plate of the same brand, style and sample volumes (water can be used) as a
balance.
5. To begin a run, close the lid of the centrifuge. The centrifuge will quickly ramp up to speed. Allow the unit to run for 15-20 seconds.
This is sufficient time to spin the samples down to the bottoms of the wells.
6. To stop, press the Open button. Once the rotor has stopped, fully open the lid to gain access to your samples. To remove the plates,
turn the knob on the top of the unit until the rotor slot lines up with the opening.
NOTE: A slight vibration may be felt when the centrifuge reaches full speed. Excessive vibration indicates that the rotor is not
balanced. Do not run the centrifuge unbalanced. Always use a balance plate of the same brand, type, and sample volumes.
7.0 Cleaning
The outside of the centrifuge may be wiped down with a clean, soft cloth. The rotor slots may be cleaned with a sponge. Do not pour
liquids in or over the centrifuge. Do not immerse the centrifuge in liquid.
Instruction Manual | 3
To request certificates, please contact us at www.labnetlink.com.
Warranty/Disclaimer: Unless otherwise specified, all products are for research use or general laboratory use only.* Not intended for
use in diagnostic or therapeutic procedures. Not for use in humans. These products are not intended to mitigate the presence of
microorganisms on surfaces or in the environment, where such organisms can be deleterious to humans or the environment. Corning
Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications. *For a listing of US
medical devices, regulatory classifications or specific information on claims, visit www.corning.com/resources.
For a listing of trademarks, visit www.corning.com/clstrademarks. All other trademarks are the property of their respective owners.
© 2022 Corning Incorporated. All rights reserved. 12/22 CLSLN-AN-1017DOC REV1 www.labnetlink.com