Axcent PAVO SERVICE MANUAL
Axcent PAVO SERVICE MANUAL
Notice
Before maintaining, please read this Manual carefully to ensure proper use.
Please keep this Manual properly for future reference.
Version: 1.0
Date of Issue: August, 2018
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
The intellectual property rights of this Instruction for Maintenance (hereinafter referred
to as ‘Manual’) and the corresponding product belong to aXcent medical GmbH
(hereinafter referred to as “aXcent”).
This Manual contains proprietary information protected by copyright law. Without
written permission of aXcent, any organization or individual is prohibited to photograph,
copy, modify or print any part of this Manual, or translate into other languages.
However, aXcent is not liable for any incidental or consequential damages to the actual
performance and use due to errors in this Manual or the provision of this Manual.
aXcent does not provide license conferred by patent law to any other parties. aXcent
does not assume legal responsibility for the legal consequences resulting from violating
the patent law and the rights of any third party.
1
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Statement
aXcent reserves the right of final interpretation to this Manual.
The content of this Manual is subject to change without prior notice.
Only if the following requirements are met, aXcent will be responsible for the safety,
reliability and performance of the product, i.e.:
◼ Assembly, expansion, re-adjustment, improvement and repair should only be
performed by qualified personnel approved by aXcent;
◼ All the replaced parts and supported accessories and consumables relate to the
maintenance shall be originally from aXcent or others approved by aXcent;
◼ The electrical equipment complies with relevant standards and the requirements of
this Manual;
◼ The product is operated in accordance with this Manual.
2
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
3
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
4
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Preface
Notice
This Manual is the necessary instructions for the safe use of this product. This Manual
introduces the use, properties, method of operation, safety information and intended
use of the PAVO Multi-Parameter Patient Monitor in details. Before using the product,
please carefully read and understand the contents of this Manual and abide by the
method of operation stated in this Manual in order to ensure the safety of patients and
operators.
This Manual is a major component of the product and should always be placed near the
product for easy reference.
Object of Application
This Manual is intended for professional clinical staff or personnel with experience in
the use of monitoring equipment. The readers should have knowledge and work
experience in medical procedures, practices and terminology of patient monitoring.
Illustration
All the illustrations in this Manual are for reference only. The menus, settings and
parameters of the illustrations may be different from the monitor.
Convention
◼ [Character]: Used to represent the string in the software or characters on the
interface.
◼ →: This symbol is used to indicate operation step.
◼
Revision History
This manual has a revision number. This revision number changes whenever the manual
is updated due to software or technical change. Contents of this manual are subject to
change without prior notice.
◼ Revision number: 1.0
◼ Release time : August, 2018
5
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
CONTENTS
6
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
7
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
8
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
DANGER: DO NOT use flammable gases (e.g. anesthetics) or flammable liquids (e.g.
ethanol) near this product in order to avoid the risk of explosion.
WARNING: Please connect the device to a socket with a separate fuse box as its
protection switch. If other equipment (e.g. life support equipment) uses the same fuse box
as the protection switch, malfunction or overcurrent of the device or instant surge current
upon startup may result in tripping.
9
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
10
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
11
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
12
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
1. Equipotential terminal: When other devices are used in conjunction with the
monitor, you should connect the equipotential terminals of other devices
and the monitor with wires to eliminate the ground potential difference
between different devices and ensure safety.
2. AC adaptor socket: connect to AC adaptor: 16.8V, 1A, charge the battery of
the monitor.
3. USB interfaces (2): Connect an external storage device to upgrade the
monitor software.
4. Network interface
5. Signal input and output interface: standard DB9 interface, which can be
connected to the PC to output data or connected with a compatible device.
6. Speaker holes
7. Spiracle
13
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
J72
J4
J1
J19
J8
J7
J23
BAT71
J13
J20 J9
Fig. 2-4 Main Control Panel
14
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
J2 J7
J1
J4
J5
J4
15
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
J3 J4
J5
J6
J2 J1
Fig. 2-6 Integrate Parameter Board
16
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
J2
J1
17
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
J4 J5 J3
J1
No. Port Name Function Description
1 J1 Rotary Knob
2 J3 J3 port connects power board
3 J4 J4 port connects alarm indicator board J1
4 J5 J5 port connects and communicates with
main control panel J19
18
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Sensor
Analog circuit
processing
Data sorting
19
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
20
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Medical worker
21
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
22
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
4.1.2. Start up
After startup, check if the monitor works normally and if the screen displays
information normally without showing any error message. Check all of the four
waveforms (three waveforms if the lead type in the “Factory Setup” menu is 3-lead) to
see if they can be scanned normally; check if the parameters are refreshed every
second and if the clock works normally.
4.1.5. Recorder
The recorder should record according to the conditions and contents stated in the
Instructions for Use, including
1. Record correct information that is consistent with the on-screen waveforms or other
recorded contents.
2. Stop immediately when the Stop key is pressed during recording.
3. There will be corresponding prompt message on the screen in abnormal situations
such as “Out of Paper”; it will continue working normally after paper loading.
4. Correct prompt messages will be given in case of hardware failure of the recorder.
5. It can work normally for a long period of time without paper jam problem.
6. During paper loading, the auto feed function of the recorder should be normal.
23
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
4.1.7. Clock
Correct indication, accurate timing, no abrupt change, and no stop.
24
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
25
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Use a NIBP simulator with calibration function. Carry out calibration according to the
calibration method specified in the Instructions for Use and judge the measurement
accuracy of BP pump. Perform the following test if the calibration result is OK:
A. Select Adult mode on both the simulator and the monitor.
B. Select a set of BP values within the measurement range on the NIBP simulator.
For example:
NS=90
NM=70
ND=60
C. Check if the result measured by the PAVO patient monitor is identical to that set
on the simulator.
D. Change the BP value set on the simulator, and carry out measurement again.
E. Check if the measured result is identical to the set value.
26
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Parts List:
No. Name or Specification No. Name or Specification
1 Front shell of PAVO 9 LCD transfer board
2 Touch screen 10 Alarm lens
3 LCD screen pressing plate 11 Alarm indicator board
4 Function key 12 Article waterproof silicone(vertical)
5 LCD screen 13 Article waterproof
silicone( horizontal)
6 Key board 14 LED dustproof trim strip(vertical)
7 Main board 15 LED dustproof trim strip(horizontal)
8 Rotary cap 16
27
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Parts List:
No. Name or Specification No. Name or Specification
1 Rear shell of PAVO 14 Recorder Cover
2 Metal plate 15 CO2 label
3 Monitoring plug-in assembly 16 Recorder cover fixing bracket
4 Network port assembly 17 ECG/NIBP Board fixing bracket
5 Speaker trim strip 18 ECG Board
6 Earth terminal 19 Battery box
7 Screw plug (horizontal) 20 Battery with four core
8 Earth terminal fixed leg 21 SD Card hole label
9 Battery Cover 22 Power board
10 Data port label 23 NIBP Board
11 Silicon floor mat 24 SpO2 board
28
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
There is no display upon startup Fault of main control board or Refer to the “Display
or black screen occurs during display Failures” section
operation; but the power
indicator is on, and the fan works
normally
Character display is normal, but Error of data communication Replace the main
the waveform is shown between main control board control board, adapter
intermittently and parameter module plate or parameter
module according to
the error message, so
as to confirm the
failure.
Some operation or measurement Damage of main control board Further check the main
function is invalid or corresponding component control board and the
corresponding
component
Occasional system crash Strong instant interference of Check the power
power grid supply and earthing
Poor performance of power system
board Replace the power
Poor performance of main board
control board Replace the main
Connection problem of power control board
supply and main control board Replace or repair the
connector
29
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
component is connected
The power indicator and the Damage of +12V DC power Replace the power supply
main control board indicator supply
come on, but the fan does
not work and the adapter
plate indicator is not on
The power indicator and the Damage of +5V DC power Replace the power supply
main control board indicator supply
are not on, but the fan works
normally and the adapter
plate indicator is on
30
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
31
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Chapter6 Installation
6.1. Check Device Packing
Unpack the device and take out the packing list. Check the names, quantities and
specifications of the items inside the packing box one by one to see if they are
consistent with those listed in the packing list. Note the following points:
1. Optional parts or other accessories (if any) purchased by the user should also be
checked item by item.
2. If items in the packing box are not consistent with those listed in the packing list,
please contact the supplier immediately.
3. In case of any damage during transportation, please properly keep the package for
further check and meanwhile contact the supplier immediately.
6.2. Preparation
Before connecting the three-core power cable to the power socket on the main unit of
PAVO patient monitor, please first conduct the following checks:
1. Check if the grid voltage meets the device requirement.
2. In order to protect patients and medical workers, a three-core power cable should
be used for the monitor, and the power socket should be a matching three-pin
socket, so as to ensure good earthing condition of the monitor. DO NOT use two-
core AC power supply.
3. When the monitor is used simultaneously with other medical device, the
equipotential terminal at the rear of the monitor should be reliably connected to the
equipotential terminal on the said other device in the hospital.
4. DO NOT place the monitor under an infusion bag or at a position with liquid leakage.
Care should also be taken to prevent any type of liquid from flowing into the
monitor.
32
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
7.2. Keys
There are three types of keys on the monitor:
◼ Soft keys: Positions in the interface where the cursor can stay, enabling you to
quickly and conveniently access a menu or execute an operation. These keys mainly
include:
◆ Parameter hotkeys: Select a parameter area to access the corresponding
parameter setup menu.
◆ Waveform hotkeys: Select a waveform area to access the corresponding
parameter setup menu.
◆ Smart hotkeys: These are shortcut keys located at the bottom of the main screen,
which enable quick operations by the user. See the “Smart Hotkeys” section.
◼ Hard keys: They are tangible keys on the monitor, such as the Alarm and Mute keys
on the front panel.
◼ Popup keys: They are menu keys relating to tasks, which will appear on the screen of
the monitor when necessary. For example, the popup confirmation keys when you
need to confirm a change.
7.5. Menu
Select [Menu] in the monitor interface by turning the rotary knob or directly
tapping on the touch screen, and open the [Main Menu] as shown in the figure below.
You can complete most operations and settings on the Main Menu.
33
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Menu Close
title
Main display
area of menu
The form of other menus is basically similar to that of [Main Menu]. A menu generally
consists of the following parts:
◼ Menu title: This summarizes the current menu.
◼ Close: It is used to close the current menu; you can exit the current menu or close
the current menu and go back to the previous menu.
◼ Main display area: This area displays options, buttons, prompt messages, etc.; “>>”
means that you can access the corresponding submenu by selecting this option.
◼ Confirmation button area: There is a confirmation button area in some menus. It is
used to confirm menu operations; there are mainly two buttons: OK and Cancel.
◆ Select [Demo>>] → Enter and confirm the password, and the monitor will
enter the Demo mode.
◼ Exit the Demo mode:
34
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
◆ Select [ExitDemo>>], and the monitor will exit the Demo mode.
WARNING
⚫ The Demo function is only used for demonstration of device performance and for
user training. In actual clinical applications, use of the Demo function should be
forbidden in order to prevent medical workers from mistaking the information
displayed on the monitor for waveform and parameters of the patient under
monitoring; such mistake will affect patient monitoring and delay disease treatment.
7.7.1. Definition
When the monitor is installed or when the monitor is used for a new application, the
monitor should be defined as follows:
35
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
36
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
◼ Select [User Maintenance>>] → Enter and confirm the current password → [User
Maintenance] menu;
◼ Select [Unit Setup>>] → [Unit Setup] menu;
◆ Select [Height]; select the unit [cm]/[inch] as needed.
◆ Select [Weight Unit]; select the unit [kg]/[lb] as needed.
◆ Select [Press.Unit]; select the unit [mmHg]/[kPa] as needed.
◆ Select [Temp Unit]; select the unit [℃]/[℉] as needed.
37
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Chapter8 Cleaning
The PAVO series patient multi-parameter monitor should be cleaned according to
specified method. In order to avoid contamination and damage of the device, please
use recommended materials for cleaning. For the sake of patients’ safety, DO NOT use
any degraded or damaged cables, sensors or accessories.
8.1. Cleaning
1. Use soft cloth for wiping and cleaning;
2. Soap, amino or ethanol material can be used to clean the main unit and power box;
3. Soap or amino material can be used to clean ECG cables;
4. Soap, amino or ethanol material should be used to clean reusable cuffs;
5. Soap or amino material should be used to clean reusable IBP sensors;
6. Soap, amino or ethanol material should be used to clean reusable SpO2 probes;
7. Soap or ethanol material should be used to clean reusable TEMP probes.
Remarks:
1. Ammonia water: Diluted ammonia water <3%;
2. Ethanol: Ethanol 70%, isopropanol 70%;
3. DO NOT use abrasive materials such as wire brush or silver polish;
4. NEVER allow cleaning fluid to flow into the device or soak the component to be
cleaned.
8.2. Sterilization
1. After cleaning, sterilization should be carried out when necessary.
2. Ethanol or aldehyde cleaning agent should be used to clean the main unit and power
box;
3. Ethanol cleaning agent should be used to sterilize ECG cables;
4. Ethanol or aldehyde cleaning agent should be used to sterilize reusable cuffs;
5. Ethanol or aldehyde cleaning agent should be used to sterilize reusable SpO2 probes;
6. Bleaching agent should be used to sterilize reusable TEMP probes.
Remarks:
1. Ethanol cleaning agent: Ethanol 70%;
2. Bleaching agent: Sodium hypochlorite diluent (beaching agent for washing),
concentration 500ppm~5000ppm. NEVER allow the sterilizing liquid to flow into or
soak the device or its accessories. DO NOT leave any residual sterilizing liquid on any
surface.
8.3. Disinfection
After cleaning, disinfection should be carried out when necessary. Since the disinfection
process is complex, you must observe applicable hospital regulations and pay special
attention to safety. Generally, EtO gas is used for disinfection.
38
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Chapter9 Maintenance
The PAVO series multi-parameter patient monitor is a precise and sophisticated device.
Proper servicing and maintenance of the monitor not only can bring the device performance
into full play, but also can ensure the device accuracy during long-time work and avoid
various measurement errors. In order to avoid cross contamination, please confirm that the
device has been cleaned and disinfected before maintenance.
Frequently check the device and the cables, sensors and wires used;
Clean the device from time to time according to actual needs;
Perform the safety test once a year;
Calibrate the NIBP parameter once a year;
Calibrate the TEMP parameter once a year;
Test the device functions once a year;
Perform the safety test every time after the device is opened for repair;
In case of any problem found during maintenance, please contact the supplier immediately.
39
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Chapter10 Accessories
10.1 Accessories
Standard accessories are as follows (depending on configuration):
Standard
No. Name Specifications & characteristics
quantity
1 3 or 5-lead ECG cable Adult 3 or 5-lead all-in-one clip 1
2 Electrode slice 50mm×55mm, adult, disposable 1
3 SpO2 probe Adult finger clip 1
Blood pressure cuff Adult repetitive single-tube cuff 1
Blood pressure gas nozzle Metal 1
4
Blood pressure extension 3M gray
1
tube
2.25K-type temperature Adult body surface
5 1
probe
Detachable three-cell lithium battery
6 2.600mAh lithium battery 1
(11.1V 2.600mAh)
7 Ground wire V0020A 1
8 Power cord 3×0.75mm2 10A 250V L=2M 1
40
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
Risks by electric shock can be minimized by using tested and approved accessories and
regulary electrical safety testings by authorized staff after installation, before using and
during lifetime of the product.
41
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
42
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
43
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
44
aXcent medical GmbH PAVO | Instructions for Maintenance Ver.: 1.0
NOTES:
45
aXcent medical GmbH
Josef-Görres-Platz 2
56068 Koblenz / Germany
Tel.: +49 261 3011 117
Fax: +49 261 3011 111
[email protected]
www.axcentmedical.com