VAL 1000 Qualitative Validation Guidelines
VAL 1000 Qualitative Validation Guidelines
It must be revised to reflect your lab’s specific processes and/or specific protocol requirements.
Each laboratory should first design a validation plan describing how they will satisfy each of
these requirements. The validation plan must also detail the acceptability criteria for each
element. After completing all of the validation experiments, results should be compiled and filed
in an organized manner. All validation records should be retained for the life of the instrument.
A validation summary should be prepared that contains a place for the Laboratory Director to
sign, indicating the validation has been reviewed and approved.
The following are the required components of validation for most qualitative methodologies
• The manufacturer’s package insert describes precision specifications for the assay
a. Sample Criteria
• Long-term: Run each level of control at least once but not more than 5
times a day for a total of 20 runs. (For example, run each sample
once per day for 20 days, OR twice a day for 10 day, but no more
than 5 times per day for 4 days).
c. Acceptability Criteria
2. Accuracy is the true value of a substance being measured. Verification of accuracy is the
process of determining that the test system is producing true, valid results.
b. Sample Criteria
• For HIV rapid tests: if using patient samples for accuracy testing, results must
be confirmed positive or negative by an FDA approved, validated method.
• For urine hCG rapid tests: Ideally, in order to verify the manufacturer’s stated
cut-off, the 10 positive samples should include low as well as high positives,
and at least 1 sample should be close to the manufacturer’s stated cut-off.
• Two levels of quality control must be run each day that validation testing is
performed, not including controls internal to the kit cartridge/testing device
• Calculate the diagnostic sensitivity and specificity rate (true positive and true
negative percentage)
d. Acceptability Criteria
3. Linearity, Analytical Measurement Range (AMR) and Clinical Reportable Range are not
applicable for qualitative methods.
5. Reference Ranges can be determined by the laboratory with laboratory director approval.
Verification of manufacturer’s stated reference range is not required.
6. Method Approval
• There must be an approval with a signature from the Medical and/or Laboratory
Director and preparer of validation documents with dates.
References
1. CLSI. Evaluation of Qualitative, Binary Output Examination Performance. 3rd ed.
CLSI guideline EP12. Clinical and Laboratory Standards Institute; 2023.
2. EP Evaluator Release 12.0, David G. Rhoads Associates Inc.,
datainnovations.com/ep-evaluator-resources.
3. Westgard, James O., Basic Method Validation: Training in Analytical Quality
Management for Healthcare Laboratories, 4th edition, 2020 Madison, WI 53717.