APP Serial-Dilution Workflow On TurboMatrix MultiPrep 013991 01
APP Serial-Dilution Workflow On TurboMatrix MultiPrep 013991 01
Gas Chromatography/
Mass Spectrometry
Author:
Adam J. Patkin, Ph.D.
PerkinElmer, Inc.
Shelton, CT
2
Below are recommended MultiPrep+ configurations for Methods
Serial Dilution and a typical pre-run checklist. The TurboMatrix Sample Manager Calibration Dilution method is
shown in Table 3. The GC injection method is shown in Table 4.
Minimum Recommended MultiPrep+ Configuration Most parameters were used at their default values.
The Minimum Configuration Required to Perform Calibration
The PerkinElmer Clarus 690 GC used a 1 µL injection split 100:1
Dilution Is:
into a capillary split/splitless injector at 250 °C onto a PerkinElmer
• Robotic tool change syringe park station
Elite™ 5MS, 30 m x 0.25 mm ID x 0.25 μm column. Carrier gas
• Sample Tray holder
• 2 mL VT-54 rack was 1.5 mL/min helium (99.999 + % purity). The oven program
• Standard Wash station was 90 °C for 1.5 min, then ramp to 157 °C at 10 °C/min. The
• 1000 µL (57 mm needle, 23 Gauge) syringe and D8/57 tool (N6496004) split flow was reduced to 10:1 at 2 min to reduce gas
• 100 µL (57 mm needle, 23S Gauge) syringe and D7/57 tool (N6496002) consumption. The Clarus SQ 8T mass spectrometer operated
• 2 mL vials and caps under UltraTune conditions with a 250 °C transferline and ion
source, monitoring m/z 57 for the alkanes.
Suggested MultiPrep+ Configuration Enhancements
For homogeneous sample mixing and lower internal Table 3. Sample Manager Serial Dilution Method.
standard usage: Configuration
• Vortexer (strongly recommended) 100 μL Sample 1000 μL Solvent
P/N N6556086 P/N N6556089
• Magnetic metal crimp caps for 2 mL vials (required for Vortex mixing) Syringe Syringe
• L arge Wash station (N6496024), replacing Standard Wash for dilution Dilution Solvent
Dilution Vial Rack Rack 1 Standard Wash 1
solvent washing Station
• Solvent Module (N6496020), replacing Standard Wash for dilution solvent Wash Station Large Wash 1 Vortex Mixer Vortex Mixer 1
• 57 mm, 23 Gauge needles (required for bottom sensing vials) Dilute
• 1 0 µL (85 mm needle, 23S Gauge) syringe and D7/85 tool (N6496003) Dilution Steps 8 Repetitions 1
for GC injection
Sample Fill Rate Solvent Fill Rate
25 μL/s 50 μL/s
100 μL Syringe 1000 μL Syringe
Pre-Run Checklist Rinsing
1. Verify 1000 µL sample prep syringe is mounted. Stock Solution Dilution Sample
2 2
2. Verify 10 µL sample prep syringe is mounted. Rinse Cycles Rinse Cycles
3. (Optional) Verify 10 µL GC injection syringe is mounted. Wash 100 μL
Wash Cycles for
4. Place stock solution in Vial #54. Syringe After On 2
100 μL syringe
5. R eplace Standard Wash vials #1-#3 (depending on number of replicates) Each Sample
Wash Cycles for Rinse Volume for
with fresh 10 mL vials (to avoid carry-over) and dilution solvent. 2 10 μL
1000 μL Syringe 100 μL Syringe
6. R efill Large Wash station vials with syringe cleaning solvent.
Rinse Volume for
(#1 for internal standard, #2 for diluent). 20 μL
1000 μL Syringe
7. Put new, capped sample vials in tray positions being used (Figure 1).
Mixing
Mixing Speed 1200 rpm Mixing Time 6s
Stock Standards Preparation Advanced
Pesticide grade isooctane was used as the diluent and cleaning Height From
Bottom Sense
Off Bottom of 0.5 mm
solvent, and analytical standard grade n-decane, n-undecane, Stock Solution
Stock Solution
n-dodecane, n-tridecane, and n-tetradecane as the analytes.
Syringe Overfill 5% Wash Vial Depth 44 mm
The stock solution was prepared gravimetrically at 1.92, 2.34, Waste Port Sample Vial
1.73, 2.16, and 1.86 mg/mL, respectively. The naphthalene 12 mm 50 mm/s
Depth - Solvent Penetration Speed
internal standard was prepared at 1.01 mg/mL. Fast Expel for
50 μL Target Vial Depth 10 mm
1000 μL Syringe
Dilution Solvent Solvent Vial
44 mm 44 mm
Station Depth Depth
Stock Solution
30 mm Sample Pumps 5
Penetration Depth
Sample Vial
30 mm
Penetration Depth
3
Table 4. Sample Manager Injection Method. Excellent linearity was observed over a wide linear dynamic
Configuration range. Examples are shown in Figure 2 and Figure 3.
Gas Chromatograph GC2 Syringe P/N N6556084
Pre-injection
Verify Barcodes Off Standard Wash 1
Wash Station
Post-injection Standard
Peltier Stack 1 none
Wash Station Wash 1
Peltier Stack 1
20 °C Peltier Stack 2 none
Temperature
Peltier Stack 2
20 °C
Temperature
General Syringe
Sample Vial Depth 30 mm Sample Pumps 6
Fill Volume 3 μL
Pre-injection Wash
Figure 2. n-Undecane Calibration Curve.
Wash Cycles 6 Wash Solvent 1 2
Wash Solvent 2 3 Wash Solvent 3 0
Wash Solvent 4 0 Syringe Fill Volume 3 μL
Syringe Flow Rate 2 μL/s
Sample Rinse
Sample Rinses 1 Rinse Volume 3 μL
Delay After Pull Up 2 s Delay After Aspirate 1s
Sampling
Sample Viscosity
2s Sample Fill Rate 2 μL/s
Delay
Prep Ahead Disabled
Inject Sample
Figure 3. n-Tridecane Calibration Curve.
Injection Mode Normal Injection Flow Rate 50 μL/s
Post-injection The mean R2 coefficient of determination was better than 0.999
0s
Dwell Time
over five orders-of-magnitude dilution, limited on the low end by
Post-injection Wash
trace analyte contaminants in the solvent.
Wash Cycles 6 Wash Solvent 1 2
Wash Solvent 2 3 Wash Solvent 3 0 Table 5. Coefficients of Determination for analyte linearity.
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
Copyright ©2018, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
013991_01 PKI