Recorder/Controller

Specification Sheet
MODEL

• Secure data recording


The nanodac™ recorder/controller offers the ultimate in graphical
• 2 PID control loops recording combined with PID control for a box of its size. The compact ¼
DIN panel mount unit offers four high accuracy universal inputs for data
• High accuracy universal inputs recording and PID control. This secure data recording device with
• USB removable data storage accurate control is enhanced by a full colour, ¼ VGA display to bring a
facility crystal clear operator interface to even the smallest of machines.

• Compact design Crystal clear, colour display


The 3.5” TFT display offers incredibly clear visualisation of process
• 50MB flash memory
parameters with a wide selection of configurable views to best suit the
• Ethernet communications application. Views include: Horizontal and vertical trends; Horizontal and
vertical bar graphs; Numeric; Alarm status, and control loops. The unit
• ¼ VGA crystal clear display also provides user wiring from the front of the product for detailed
• 14 virtual channels configuration without the need to connect to a PC.

• Steriliser Application Block Data Acquisition and Recording


The nanodac recording functionality utilises the secure strategies and
• Relative Humidity Application UHH format developed by Eurotherm through years of recording
Block expertise. As well as multiple real-time views and historical review on the
• Multiple I/O options product, multiple data archiving strategies are provided utilising the
50MB onboard Flash memory, removable USB and data transfer via FTP
• Multi-language support to a specified server.
(French, German, Italian and The four universal input channels provide high accuracy (suitable for use
Spanish) in Nadcap applications) and 125ms parallel sampling. An additional 14
virtual channels can be utilised to provide maths, counter, slave
communications and totaliser functionality within the instrument.
Review software can also be utilised to automatically initialise pulling of
data to a Review database over the Ethernet network. Both Review and
Dream Report™ software can be used for report generation that meets
the needs of your process.

PID Control Loops


The nanodac instrument can also provide two independent control
loops (optional). This control functionality utilises the advanced
Eurotherm PID algorithm providing high performance and reliability to
your process. Functionality includes one of the best autotune facilities
available along with overshoot inhibition (cutbacks); compensation for
power fluctuations using power feedforward; linear, fan, oil and water
cooling.

imagine bigger better smaller


Specification Cable type: Category 5
Maximum length: 100metres (110 yards)
General Termination: RJ45.
General Green LED illuminated = link connected;
I/O types Analogue i/p: Four Amber LED flashing shows link activity
Digital i/p: Two USB port
Digital (logic) o/p: Two max (see order code) Number of ports: One at rear of instrument
Relay o/p: Four max (see order code) Standard: USB1.1
DC output: Two max (see order code) Transmission speeds: 1.5MBit/sec (low speed device)
Features: CSV archive format Maximum current: <100mA
File transfer protocol (FTP) Peripherals supported: Memory stick (8GB max), Bar code reader,
Messages QWERTY keyboard
Modbus TCP (Slave)
uhh (history file) archiving Update/Archive rates
USB port at rear of instrument Sample rate (input/output): 8Hz
User linearisation tables (four) Trend update: 8Hz max.
Two control loops (optional) Archive sample value: Latest value at archive time
Zirconia probe support (optional) Display value: Latest value at display update time
14 Virtual channels (each configurable as
counter, maths, totaliser or comms input)
Steriliser (optional)
Relative humidity (optional) Analogue Input
Customised start up screen
General
Environmental performance Number of Inputs: Four
Ambient temperature range Input types: dc Volts, dc mV, dc mA (external shunt
Operating: 0 to 55°C required), Thermocouple, RTD (2-wire and
Storage: –20 to +70°C 3-wire), Digital (Contact closure)
Humidity range Operating: 5% to 85% RH non condensing Input type mix: Freely configurable
Storage: 5% to 85% RH non condensing Sample rate: 8Hz (125ms)
Protection Front panel IP65, NEMA4X (International) Conversion method: 16 bit delta sigma
Behind panel: IP10 (International) Input ranges: See Table 1 and Table 2
Shock/Vibration: To BS EN61131-2 (5 to 150 Hz. at 1g; 1 Mains rejection (48 to 62Hz)
octave per min.) Series mode: > 95dB
Altitude: <2000 metres Common mode: >179dB
Atmosphere: Not suitable for use in explosive or corrosive Common mode voltage: 250V ac max.
atmospheres Series mode voltage: 280mV at lowest range; 5V peak to peak at
Electrical safety: BS EN61010-1 (Installation category II; highest range
Pollution degree 2) Input Impedance: 40mV, 80mV, 2V ranges > 100MΩ;
Electromagnetic compatibility 62.5kΩ for input voltages > 5.6V
Emmissions (Standard units): BS EN61326 Class B – Light industrial 667kΩ for input ranges < 5.6V
(Low voltage option): BS EN61326 Class A – Heavy industrial Overvoltage protection
Immunity: BS EN61326 Industrial Continuous: ±30V RMS
Transient (<1ms): ±200V pk-pk between terminals
Other approvals and compliance details Sensor break detection Type: ac sensor break on each input giving quick
General: CE and cUL, EN61010 response with no associated dc errors
PV input: AMS2750D compliant Recognition time: <3 seconds
RoHS EU; China Minimum break resistance: 40mV, 80mV ranges: 5kΩ; other ranges: 12.5kΩ
Packaging: BS61131-2 section 2.1.3.3. Shunt (mA inputs only): 1Ω to 1KΩ mounted externally
additional error due to shunt: 0.1% of Input
Physical Isolation:
Panel mounting: 1/4 DIN Channel to Channel: 300V RMS or dc (Double insulation)
Weight: Instrument only: 0.44kg (15.52ozs) Channel to common
Panel cutout dimension: 92 mm x 92 mm (both -0.0 +0.8) electronics: 300V RMS or dc (Double insulation)
or 3.62 in x 3.62 in (both -0.00 +0.03 in) Channel to ground: 300V RMS or dc (Double insulation)
Depth behind panel: 90 mm (3.54 in) excluding wiring Dielectric strength Test: BS EN61010, 1 minute type test
Channel to Channel: 2500V ac
Operator interface Channel to Ground: 1500V ac
Display: 3.5” TFT colour display
(320 pixels wide x 240 pixels high) Low High Resolution Maximum error Temperature
Controls: Four navigation pushbuttons below the Range Range (Instrument at 25°C) Performance
display screen (Page, Scroll, Lower and Raise) –40mV 40mV 1.9μV 4.6μV + 0.053% of reading 13ppm of input per °C
–80mV 80mV 3.2μV 7.5μV + 0.052% of reading 13ppm of input per °C
Power requirements
–2V 2V 82μV 420μV + 0.044% or reading 13ppm of input per °C
Supply voltage: Standard: 100 to 230V ac ±15% at 48 to 62Hz
Low voltage: 24V ac (+20% –15%) at 48 to 62Hz, or –3V 10V 500μV 1.5mV + 0.063% or reading 45ppm of input per °C
24V dc (+10% –20%) Table 1 Voltage input ranges
Power dissipation: 9W (max.)
Resistance input ranges
Fuse type: None
Temperature scale: ITS90
Interrupt protection: Standard: Holdup >10ms at 85V RMS supply voltage
Types, ranges and accuracies: See Table 3
Low voltage: Holdup >10ms at 20.4V RMS supply voltage
Maximum source current: 200μA
Battery backup Pt100 figures Range: 0 to 400Ω (–200 to +850°C)
Stored data: Time, date Resolution: 0.05°C
Support time Calibration error: ±0.31°C ±0.023% of measurement in °C
(for real-time clock): Minimum of 1 year with unit unpowered at 25°C ambient
Replacement period: Three years typical Temperaure coefficent: ±0.01°C/°C ±25ppm/°C measurement in °C
Type: Poly-carbonmonofluoride/lithium (BR2330) from 25°C ambient
(PA260195) Measurement noise: 0.05°C peak-peak with 1.6s input filter
Linearity error: 0.0033% (best fit straight line)
Replace battery with Panasonic BR2330/BE only. Use of another battery may Lead resistance: 0 to 22Ω matched lead resistances
present a risk of fire or explosion. See owners manual for safety instructions. Bulb current: 200μA nominal
Caution Battery may explode if mistreated. Do not recharge, disassemble or
dispose of in fire
Low High Resolution Maximum error Temperature
Range Range (Instrument at 25°C) Performance
Ethernet communications
Type: 10/100baseT Ethernet (IEEE802.3) 0Ω 400Ω 20mΩ 120mΩ + 0.023% of reading 25ppm of input per °C
Protocols: Modbus TCP/IP slave, FTP, DHCP Table 2 Ohms (RTD) input ranges

Eurotherm Part No. HA030686 Issue 4 March 11 2 nanodac Specification Sheet


RTD Type Overall range Standard Max. linearisation DC Output (option)
(°C) error
Cu10 -20 to +400 General Electric Co. 0.02°C DC analogue outputs
Cu53 -70 to +200 RC21-4-1966 <0.01°C Current outputs
JPT100 -220 to +630 JIS C1604:1989 0.01°C (O/P2 and O/P3)
Ni100 -60 to + 250 DIN43760:1987 0.01°C Output ranges: Configurable within 0 to 20mA
Ni120 -50 to +170 DIN43760:1987 0.01°C Load resistance: 500Ω max.
Pt100 -200 to + 850 IEC751 0.01°C Calibration accuracy: <±100μA ±1% of reading
Pt100A -200 to + 600 Eurotherm Recorders SA 0.09°C
Voltage outputs
Table 3 RTD type details O/P3 only)
Output ranges: Configurable within 0-10V
Thermocouple data Load resistance: 500Ω min.
Temperature scale: ITS90 Calibration accuracy: <±50mV ±1% of reading
CJC Types:Off, internal, external, remote.
Remote CJC source: Any input channel General
Internal CJC error:<1°C max., with instrument at 25 °C Isolation: 300V ac double insulated from instrument
Internal CJC rejection ratio:
40:1 from 25°C and other I/O
Upscale/downscale drive: High, low or none independently Resolution: >11 bits
configurable for each channel’s sensor break Thermal drift: <100ppm/°C
detection
Types, ranges and accuracies: See Table 4

T/C Type Overall range Standard Max. linearisation Installation


(°C) error
B 0 to +1820 IEC584.1 0 to 400°C = 1.7°C Side View
400 to 1820°C = 0.03°C 96mm (3.78in) 90mm (3.54in)
C 0 to +2300 Hoskins 0.12°C
D 0 to +2495 Hoskins 0.08°C

96mm (3.78in)
E –270 to +1000 IEC584.1 0.03°C
G2 0 to + 2315 Hoskins 0.07°C
J –210 to +1200 IEC584.1 0.02°C
K –270 to +1372 IEC584.1 0.04°C
L –200 to +900 DIN43710:1985 0.02°C
(to IPTS68)
N –270 to +1300 IEC584.1 0.04°C
R –50 to +1768 IEC584.1 0.04°C
S –50 to +1768 IEC584.1 0.04°C Panel cutout:
T –270 to +400 IEC584.1 0.02°C 92mm (3.62in) x 92mm (3.62in)

91mm (3.58in)
(both -0 + 0.8mm (0.03in))
U -200 to + 600 DIN43710:1985 0.08°C
NiMo/NiCo -50 to + 1410 ASTM E1751-95 0.06°C
Minimum inter-unit spacing:
Platinel 0 to + 1370 Engelhard 0.02°C Horizontal (‘x’) = 10mm (0.4in)
Mi/NiMo 0 to + 1406 Ipsen 0.14°C Vertical (‘y’) = 38mm (1.5in)
Pt20%Rh/Pt40%/Rh 0 to + 1888 ASTM E1751-95 0.07°C
Table 4 Thermocouple types, ranges and accuracies Top View

Relay and Logic I/O


O/P1, O/P2 and O/P3 logic I/O and relay specification Isolation
Single
Sin e (50V ac)
Active (current on) current sourcing logic output
(O/P1 or O/P2 only) Ethernet
Voltage o/p across terminals: +11V min.; +13V max. comms
Supply voltage
Short circuit output current: 6mA min. (steady state); 44mA max. 100 to 240V ac Rectification Double Double
Do ble (300V ac)
(switch current) or (300V ac)
24V ac or dc Relays/
Inactive (current off) current sourcing logic output Power DC output
(O/P1 or O/P2 only) Supply
Voltage output across terminals: 0V (min.); 300mV (max.) Circuits Double
Do ble (300V ac)
Output source leakage
current into short circuit: 0μA (min.); 100μA (max.) PV1*

Active (current on) contact closure sourcing logic input


(O/P1 only) Double
Do ble (300V ac)
Input current Input at 12V: 0mA (min.); 44mA (max.)
μP & PV2*
Input at 0V: 6mA min. (steady state); 44mA max.
System
(switch current) * Note: Each ‘PV’ is double Circuits
Open circuit input voltage: 11V (min.); 13V (max.) isolated (300V RMS) from Double
Do ble (300V ac)
Open circuit (inactive) resistance: 500Ω (min.); ∞ (max.) all other ‘PV’s SN1
Closed circuit (active) resistance: 0Ω (min.); 150Ω (max.) PV3*

Relay Contacts Double


Do ble (300V ac)
Contact switching power
(resistive): Max. 2A at 230V RMS ±15% PV4*
Min. 100mA at 12V
Current through terminals: 2A

USB comms
Digital Inputs
Dig InA and Dig InB contact closure logic input
PSU Logic/
Contact closure 2MΩ
Digital I/O
Short circuit sensing current
(source): 5.5mA (min.); 6.5mA (max.)
Open circuit (inactive) resistance: 600Ω (min.); ∞ (max.)
Closed circuit (active) resistance: 0Ω (min.); 300Ω (max.) Safety earth (Protective conductor terminal)

Eurotherm Part No. HA030686 Issue 4 March 11 3 nanodac Specification Sheet


Rear Terminals
O/P1
Mains (Line) supply
Low voltage option 1A 1A 1A
voltage wiring
supply voltage wiring 1B 1B 1B
20.4 to 28.8V ac; com no
3A LB R>500R = inactive 2A 2A 2A
48 to 62Hz or 3A LB
3B LC R<150R = active
100 to 230V ac 19.2 to 26.4V dc 3B LC
± 15%; 4A Contact closure Logic o/p (active high) Relay output
(polarity 4A
50/60Hz C
irrelevant) C
L Brown 5A O/P2
24V 5A
N Blue 24V 1B 1B 1B
2A 2A 2A
2B 2B 2B
com no
Each wire connected to LA LA LA
LA, LB and LC must be less
than 30 metres in length Isolated DC o/p (mA) Logic o/p (active high) Relay output
O/P3 Dig InA Dig InB O/P4; O/P5

3A 3A LB 4A
2B
1A 1I 3B 3B LC C
com no LA R>600R = inactive com no
O/P1 USB 4A 5A
1B 1+ An In 1 LC R<300R = active
2A 1– Internal Link (0V) E com no
O/P2
2B 8 2I
Relay output Isolated DC o/p (mA/ V) Contact closure Contact closure Relay output
LA Ethernet 2+ An In 2
Dig In A (RJ45) 1
LC 2– An In1; An In2; An In3; An In4
3A LB 3I I I I I I I
O/P3 Dig In B T/C
3B LC 3+ An In 3 + mV + + + + +
O/P4 4A 3– – 0 to 1V – – – – –
0 to 10V 1R0≤R≤1k0 R>350R = inactive
C 4I
O/P5 R<200R = active
Supply 5A 4+ An In 4
Voltage T/C, Volts, Milliamps RTD RTD Ohms Digital
Safety Earth 4–
millivolts (three wire) (two wire) inputs

Order codes
1 2 3 4 5 6 7 8 9

NANODAC

Basic Product 2 Controller 5 Application Blocks 8 Toolkit Blocks

NANODAC Graphical Recorder/ X None (default) XX None XXXXX None


Controller C 2 Control loops ZC Zirconia BASIC Basic toolkit blocks
RH Humidity
ST Steriliser
3 Programmer
1 Supply Voltage 9 Operating Language
X None (default) 6 Communications Protocol
VH 100-230V ac ±15% ENG English (default)
at 48-62Hz TS Modbus TCP/IP slave FRA French
VL 24V ac (+20% –15%) 4 Output Options 1-2-3 (default) GER German
at 48-62Hz, or ITA Italian
24V dc (+10% –20%) LRR Logic/Relay/Relay (default) SPA Spanish
LRD Logic/Relay/Iso DC output 7 Bezel
LLR Logic/Logic/Relay
RDD Relay/Iso DC/Iso DC SV Silver (standard)

Eurotherm: International sales and service www.eurotherm.com


AUSTRALIA Melbourne Beijing Office ITALY Como SWEDEN Malmo
T (+61 0) 8562 9800 T (+86 10) 5909 5700 T (+39 031) 975111 T (+46 40) 384500
E [email protected] E [email protected] E [email protected] E [email protected]
AUSTRIA Vienna FRANCE Lyon KOREA Seoul SWITZERLAND Wollerau
T (+43 1) 7987601 T (+33 478) 664500 T (+82 2) 2090 0900 T (+41 44) 7871040
E [email protected] E [email protected] E [email protected] E [email protected]
BELGIUM & LUXEMBOURG Moha GERMANY Limburg NETHERLANDS Alphen a/d Rijn UNITED KINGDOM Worthing
T (+32) 85 274080 T (+49 6431) 2980 T (+31 172) 411752 T (+44 1903) 268500
E [email protected] E [email protected] E [email protected] E [email protected]
BRAZIL Campinas-SP INDIA Mumbai POLAND Katowice U.S.A. Ashburn VA
T (+5519) 3707 5333 T (+91 22) 67579800 T (+48 32) 78395000 T (+1 703) 724 7300
E [email protected] E [email protected] E [email protected] E [email protected]
CHINA IRELAND Dublin SPAIN Madrid ED63
T (+86 21) 61451188 T (+353 1) 4691800 T (+34 91) 6616001
E [email protected] E [email protected] E [email protected]

© Copyright Eurotherm Limited 2011


Invensys, Eurotherm, the Eurotherm logo, Chessell, EurothermSuite, Mini8, Eycon, Eyris, EPower nanodac and Wonderware are trademarks of
Invensys plc, its subsidiaries and affiliates. All other brands may be trademarks of their respective owners.
All rights are strictly reserved. No part of this document may be reproduced, modified, or transmitted in any form by any means, nor may it be
stored in a retrieval system other than for the purpose to act as an aid in operating the equipment to which the document relates, without the prior
written permission of Eurotherm limited.
Eurotherm Limited pursues a policy of continuous development and product improvement. The specifications in this document may therefore be
changed without notice. The information in this document is given in good faith, but is intended for guidance only.
Eurotherm Limited will accept no responsibility for any losses arising from errors in this document.

Eurotherm Part No. HA030686 Issue 4 March 11 4 nanodac Specification Sheet

You might also like