Service Manual

Weighing Indicator
KPZ 52-18
KPZ 52-19

© Klaus-Peter Zander GmbH 2014. All rights reserved.

The information contained herein is the property of Klaus-Peter Zander GmbH and is supplied without liability for
errors or omissions. No part may be reproduced or used except as authorized by contract or other written
permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the
information may be embodied. The information remains our intellectual property. Due to the legal requirements it’s
not allowed, neither to copy them nor to acquaint or handle them to third persons without our written agreement
(Copyright law).

KPZ 52-18_52-19_DEALER_servicemanual_ENG
Table of Contents
INSTRUCTIONS FOR USE ................................................................................................. 3
PREPARING TO USE THE SCALE ..................................................................................... 3
QUICK SETUP CALIBRATION ............................................................................................ 5
CHAPTER 1 INTRODUCTION ............................................................................................ 6
1-1 FEATURES .................................................................................................................... 6
1-2 SPECIFICATIONS ........................................................................................................... 6
1-3 DISPLAY ....................................................................................................................... 6
1-4 KEYPAD FUNCTIONS DESCRIPTION ................................................................................. 7
1-5 POWER SUPPLY ............................................................................................................ 7
1-6 ERROR MASSAGE ......................................................................................................... 7
CHAPTER 2 OPERATING INTRODUCTION ...................................................................... 8
2-1 KEYPAD OPERATING...................................................................................................... 8
2-2 SELF TEST MODE ....................................................................................................... 10
2-3 GENERAL FUNCTION OPERATING ................................................................................. 11
2-3-1 Simple Counting Function ................................................................................. 12
2-3-2 Check Weighing Function .................................................................................. 13
2-3-3 Accumulation Function ....................................................................................... 14
2-3-4 Hold Function..................................................................................................... 15
2-4 FURTHER FUNCTION OPERATING.................................................................................. 16
2-4-1 General Setting ................................................................................................. 16
2-4-2 Check Voltage ................................................................................................... 16
2-4-3 External Weight Calibration ............................................................................... 17
2-4-4 Transmission Setting ......................................................................................... 18
CHAPTER 3 SERVICE MODE ACCESS ........................................................................... 20
3-1 SPECIFICATION SETTING ...................................................................... 21
3-1-1 Weight Unit Setting .................................................................. 22
3-1-2 Customized Capacity Setting ................................................... 25
3-1-3 Multi-range Setting ................................................................... 30
3-2 W EIGHT CALIBRATION .......................................................................... 31
3-2-1 Weight Calibration Setting ....................................................... 32
3-2-2 Manufacturing Location G Value Setting ................................ 33
3-3 LINEARITY CALIBRATION ....................................................................... 34
3-4 FUNCTION SETTING ............................................................................. 37
3-4-1 Environment Parameters Setting ............................................. 38
3-4-2 Zero Tracking Setting ............................................................... 39
3-5 LOCAL G VALUE SETTING ................................................................... 39
APPENDIX 1 7-SEGMENT DISPLAY CHARACTERS....................................................... 40
APPENDIX 2 INSTALLATION ............................................................................................ 41

2
Thank for your purchasing of our Weighing Scale. To
guide you to use our product correctly, please read
this User Manual carefully to extend the life of
machine and to avoid error.

Instructions for Use


1. Please keep scale in a cool and dry place. Do not store under high temperatures.
2. Please keep the scale clean and free from insect infestation.
3. Avoid impacting with other items or overloaded with excessively heavy weights.
(The load must not exceed the maximum capacity of the scale).
4. If the scale is not going to be used for some time, please clean it and store it in a plastic bag
in dry condition. A desiccant sachet may be included to prevent moisture from building up.
5. Any suggestion is warmly welcome.

Preparing to Use the Scale


1. Locate the scale on a firm level surface free from vibrations for accurate weight readings.
Adjust the four leveling feet to centre the leveling bubble on the scale.
2. Avoid hot sunshine directly on the scale or near the exhaust port of ventilating system.
3. Please use a separate power source plug, to avoid the disturbance of other electric
appliance.
4. There should be no weight on the scale when power is turned on.
5. Commodity should be placed at the centre of platter when being weighed, and its size
should not exceed the dimension of the platter.
6. We suggest to warm the scale for 15 ~ 20 minutes before using.。
7. Please note that when symbol shows up on the display, which means the indicator
needs to be recharged. The indicator could only operate for another 6~8 hours. Then
“ . . ” will show on the display and the indicator must be fully charged before operating
again.
8. Introduction of Storage Battery

3
Due to the storage battery adopt the advanced free-maintaining technique, customers need
not to replenish electrolyte.
The scale should be recharged every 3 months to prevent failure of the internal rechargeable
battery.
1. The battery should be charged for 8~10 hours.
2. The temperature of battery should below 45℃.
Maintaining
1. Please do not discharge with over-current when using the battery. Please charge the
battery after discharging current.
2. Please take down the battery when the scale is not used for a long time or break the
connection of cathode.
3. Do not short the battery terminals to check whether there is current. Please check whether
the connection point is firm to guarantee good connection.
4. The battery should be replaced by specialized person. No reverse-battery or the product
will be damaged.
a) Anode of battery should be connected with Anode of product battery(usually red
cable)
b) Cathode of battery should be connected with Cathode of product battery (usually
brown cable or black cable)
c) See the picture following

Brown cable(or black cable)


connected with Anode of
battery

Red cable connected with


Cathode of battery

Safety warnings
1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.
2. The hydrogen will be resolved when using or charging the battery and it will cause
explosion when approaches fire.

No burning Caution Corrosion Warning explosion Children faraway

4
Quick Setup Calibration
This page is to quickly initiate the scale, for the other functions configuration, you can refer the
chapters below.

Instructions:
Step 1 :
 Power off the scale and open the case, find the mini-jumper SWA1 on the main
board.
 Switch SWA1 to the ADJ position and then turn the power on. The display will
show .

Step 2 :
 Refer to the chapter 3-1 at page 20 to complete Specification Setting.

Step 3 :
 Refer to the chapter 3-3 at page 33 to complete Linearity Calibration.

Step 4 :
 Refer to the chapter 3-2 at page 30 to complete Weight Calibration.

Step 5 :
 When done the initiation, switch the jumper SWA1 back to the LOCK position.

 If the jumper SWA1 is switched to the LOCK position during calibration, the machine
will exit the service mode automatically.

5
Chapter 1 Introduction
1-1 Features
 The interface is simple and easy to operate
 Up to 1/15,000 display resolution
 High Speed 24bits AD
 AC / DC switching function (Built-in rechargeable battery)
 Low power indication
 LCD display (Weight-6 digits); Full range tare; Auto zero tracking
 Simple counting; Gross/Net indication; Hold function
 Check mode Lo / Hi / OK with 3 selectable beeping sounds
 Lo / Hi / OK indication shows on LED lights
 Selectable unit: Kilogram (kg) and pound (lb)
 Easily viewing single weight / total weight within 1 button press
Option:
 RS232: Bi-direction transmission, easy data reading or printing
 Sleeve connecting stand

1-2 Specifications
 Display LCD 6 digits, Height 25mm, LED Backlight
 Indicator Dimensions: 235 x 163 x 133 mm (W x H x D)

1-3 Display

● HI
● OK
● LO

1 2 3 4 5 6
● 7
Icon Introduction
→O← : “Zero” indication
: When the weight is negative, this icon appears.
Net : Put container on the platter, after stable and press tare key and the “Net” appears.
: “Low battery power” indication
kg : “kg” unit
lb : “lb” unit
斤 : “Taiwan Tael“ unit
Pcs : Counting mode
% : Percent indication
HI : The weight is higher than setting value, “HI “LED on
OK : When the weight is between HI limit values and LO limits value, “OK “LED on
LO : The weight is lower than setting value, “LO “LED on

6
▼ 1 : “STABLE” indication
: “GROSS” indication. When the object is in net mode and press F key, this icon
▼ 2 appears
▼ 3 : “M+” indication
▼ 4 : “PT” indication
▼ 5 : Range 2
▼ 6 : Range 1
● 7 : When this LED is on, the indicator is charging.

1-4 Keypad Functions Description


key Function Hold the key for 1 sec. Hold the key for 3 sec.
Zero key Check Voltage
 Backlight setting
Tare/Pre-tare key  Zero display setting
 Local G value setting
Enter password to enter
Unit key
external weight calibration
Function key

Print / transmit key RS-232 transmission setting

1-5 Power Supply


Power Selection
6V 4Ah / 4V 4Ah rechargeable battery
110V/220V AC/DC Adaptor
Power Consumption
About DC 37.7 mA (System + Load cell+ LED backlight) about 106 hours
LED charging icon
When “ . . ” shown on the display, the indicator must be charged. When the charging icon
becomes red, it means the indicator is charging; when the charging icon becomes green, it
means the indicator is fully charged.
The indicator needs to be charged for about 8~9 hours when“ . . ” shown on the display.

1-6 Error Massage


 The EEPROM is not working correctly.
(The EEPROM is not set yet, or the circuit on PCB is broken.)
 Zero is higher than the zero range when switching the indicator on.
 Zero is lower than the zero range when switching the indicator on.
 A/D value is unstable
 ADIC value is over the maximum range. (2097151 ~ -2097152)
 The weight of the object is over 9 divisions of the maximum capacity.
 The weight of the object is under -1/6 maximum capacity.

7
Chapter 2 Operating Introduction
2-1 Keypad Operating
: Zero key
If the weighing value is within the range of zero balance, it can be re-zeroed and tare
cancelled.

: Tare key
It can be tare except when the weight is negative and the maximum capacity.

Tare Function
 Place the container onto the scale, until the weight value is stable, press key for zero
return and the NET indication is shown on the display.
 Place the object into the container and the display shows the net weight value of the object.
 Remove both object and container, and negative value of the container will show on the
display. Press again to clear “tare value”. The scale returns to zero and NET indication
goes off.
 Tare can be continuously done until tare value=maximum capacity
 Continuous Tare  Press key for continuous weight increase/decrease on platter.
 If there is Tare, the pre-tare cannot be done. If there is pre-tare first, and the tare weight
Is more than pre-tare weight, Tare can be done.
 No Tare can be done under gross weight display mode.

Pre-Tare Function
When there is nothing on the platter, press key and use keypad to input pre-tare weight.
When the cursor flashes on the last digit of display, press key again to complete.
Method of clearing Pre-tare value: the same as the method of clearing tare value.
Pre-tare mode, keypad function as followed:
⇒ upward key ⇒ move cursor leftward

⇒ downward key ⇒ move cursor rightward

: Function key
 In weighing mode, use this key to shift among simple counting function, check weighing
function, accumulation function and hold function.
 In tare mode, use this key to shift between the “Net value” and the “Gross value”.

Shift between Net value and Gross value


 In the Tare mode, the “Net” icon shows up. Press F key, the screen displays the “gross
weight ” and when the ”GROSS” icon  shows up ,“Net” icon goes off. Press
1F key again the screen displays the “net weight ” and when the ”Net” icon shows up ,
the ”GROSS” icon  goes off .
When the screen displays the “gross weight” (the ”GROSS” icon  shows up), keys except
the F key, have no response.

8
:Unit key
Use key to select a unit as the display indicated.
 The selected unit will be memorized when you turn the indicator off. And the memorized
unit will appear after you turn on the indicator next time.

:Print key
In accumulation mode, if the parameter set as or transmission, it
transmits the format of simple mode or complete mode.
This key is the combination key. while total counts is shown and weight returns to net zero,
press key to clear data and RS-232 transmits the MC printing format.
 If there is new weight added on platter, a new data will be added to accumulation data. If
this weight is not taken off, nothing can be added to accumulation data. Display will show
the total counts for one second, then show net weight for one second, then the display
returns to the current weight.
 Clear Accumulation data: Press key and the total counts shows on the display. Press
key again to clear accumulation data.
 Weight must return to net zero if to perform clear function

9
2-2 Self Test Mode
Press key and not release. Then power on the indicator, Wait till display shows
to enter “Self-Test Mode”.

Press key

Press key Press key

Press key Press key

Press key Press key

Press key

INTERNAL VALUE MODE (must hook up full-bridge Load Cell to test)


 Press to enter, and the display shows internal value
 Please check the internal value is within normal range(0~400000 no load)。
 Check whether the backlight is on
 Press key to back to the last screen, the display shows
KEYPAD TEST MODE
 Press to enter, display shows
 Keypad’s internal code: key = 01、 key =31、 key=20、 key=30
 Press key to back to the last screen, the display shows
FIRMWARE VERSION DISPLAY MODE
 Press to enter, display shows the firmware version
 Press key again, the display shows maintenance number for 2 seconds
 Press key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press key to exit self-test mode, the scale will re-power on automatically

10
2-3 General Function Operating
Power on the indicator and enter into the setting that you set before or weighing mode.. Then
press to shift functions.
→0← kg
. Weighing Mode
▼1 ▼6

Press key
→0←

Pcs Simple Counting Function


▼1

Press key

Check Weighing Function

Press key

Accumulation Function

Press key

Hold Function
Press key

11
2-3-1 Simple Counting Function
1. In weighing mode, press key to enter into simple counting function.
2. Press key to select sample quantity (10, 20, 50, 100, 200)
LCD shows as right picture: →0←

▼1

Press key
→0←
.
▼1

Press key

→0←
.
Press key ▼1

3. Select sample quantity and then place samples on the platter. After
stable, press key and display shows ------. After stable, the scale
enters into counting mode and the display shows sample.
→0←

Pcs
▼1
Place samples
on the platter

. Pcs After stable, ▼1


▼1 shows on
Press key

Pcs
Enter into counting
▼1 function

Press key
kg
. Display total weight
▼1

After 1 second
kg
. Display Unit weight
Return to counting ▼1
mode after 1 second

4. Press key to enter into next function.

12
2-3-2 Check Weighing Function
1. In weighing mode, press key twice to enter into simple counting function.
2. Press key to enter into beeper setting.
LCD shows as right picture:
3. Press key to select beep sound.
Press key
⇒No beep
⇒OK (when the weight is
Beeper setting
between Low limit & High
limit.), the beeper beeps. Press key
⇒When the weight is under
or equal to Low limit & over
or equal to High limit, the
beeper beeps. Press key

Press key

4. Press key to enter into Hi limit value setting


5. key⇒move cursor to right
(When the cursor arrive to the last
Digit, it returns to the first digit.)
key⇒increase/decrease number Press key
key⇒confirm and enter into Lo kg

limit value setting . . Hi limit value setting


6. In Lo limit value setting
Press key
key⇒confirm and return to
Beeper setting kg
. Lo limit value setting
Press key to
confirm and return
to Beeper setting

7. In beeper setting, press key to enter


into check weighing function. When
Press key
the weight is more than 10d and it is
→0← kg
unstable, check weighing function open Check weighing
. function
8. Press key to exit the setting and ▼1

enter into next function. Shift “Hi,OK,LO “ LED light


Put weight (>10d) according Hi-Lo limit value
Repeat above steps to set this function again. setting
 Units can not be shift in check weighing kg

function. Please enter into weighing mode . ● OK

to shift units ▼1

13
2-3-3 Accumulation Function
1. In weighing mode, press key three times to enter into accumulation function.
LCD shows as following picture:
Press key to shift

Press key Press key


→0← kg →0← kg
Accumulation . Accumulation
. function
▼1 ▼6
function ▼1 ▼6
Put weight Put weight
On platter On platter
kg kg

. .
▼1 ▼6 ▼1 ▼6
Accumulate
Press key automatically
kg
kg Accumulation
Accumulation
counts counts
▼1 ▼3 ▼6 ▼1 ▼3 ▼6

After 0.5 second After 0.5 second


kg kg
Accumulation . Accumulation
Return to . weight
weight
Return to ▼1 ▼3 ▼6
accumulation ▼1 ▼3 ▼6
function after accumulation
0.5 second function after
0.5 second
⇒Manual accumulation: After the weight is stable, press key to accumulate and
“M+” icon  shows up. It does the second accumulation, after the weight returns
to zero.
⇒Automatic accumulation: After the weight is stable, it accumulates automatically,
and “M+” icon  shows up. It does the second accumulation, after the weight
returns to zero.
 If it set as , or transmission mode, RS-232 outputs printing
format while accumulating.
 The Minimum weight of accumulation: > 10d
→0← kg In accumulation
. function, there is
2. Press key to see accumulation counts accumulation
▼1 ▼3 ▼6
and accumulation weight. data.
Press key
→0← kg
Accumulation
counts
▼1 ▼3 ▼6
after 0.5 second
→0← kg
Accumulation
. weight
Return to accumulation
function after 0.5 second ▼1 ▼3 ▼6

14
3. Clear accumulation data: →0← kg No weight on
When there is no weight on the platter, . the platter
▼1 ▼3 ▼6
press key to display accumulation counts.
Then press key again to clear the data. Press key
“M+” icon  goes off. →0← kg
Accumulation
 If it set as or transmission mode, counts
▼1 ▼3 ▼6
RS-232 outputs MC printing format after data clear.
 Press key for 2 seconds and the beeper beeps Press key
3 times. The accumulation data is cleared and RS-232 →0← kg
Clear accumulation
does not output MC printing format. “M+” icon  goes off. data
▼1 ▼6
4. Press key to exit the setting and enter into next function.
The original accumulation data still exist.

2-3-4 Hold Function


1. In weighing mode, press key four times to enter into hold function.
2. Use key to select , or
Then press key to enter into Hold function.
When the weight is fixed, the beeper beeps.
Press key Press key
The weight is unlocked only after taking off
the weight and the weight returns to zero.
The weight is unlocked, when increase or
decrease the weight that is beyond the
tolerance d. Press key Press key
 Calculate the average weight and fixed the
display. To unlock it, press key to re-read
the weight.

Press key Press key

Press key Press key Press key


→0← kg
Weighing d value: 2~50d
.
mode defaults: 5d
▼1 ▼6

:Move to right :save calculate Press


average weight key
:adjust number
→0← kg Press →0← kg
key to enter .
.
▼1 ▼6
into weighing ▼1
mode
3. Press key to exit the setting
and enter into next function.
 Repeat above steps to set this function again.

15
2-4 Further Function Operating
2-4-1 General Setting For entering values:
 Upward key (0~9 digit entry)
→0← kg
Press key to save. Press  Downward key (0~9 digit entry)
key to exit without saving and return
. to upper level or weighing mode.  Move cursor rightward
▼1
 Move cursor leftward
Press key for 1 sec.

Backlight setting Zero display setting Local G value setting Backlight setting

Press key Press key Press key

Auto. Backlight
Use or Press
Press key Press key to enter parameter keys in order

. .

Backlight on Enter local G


Press key Press key
value

. .

Backlight off
After selection Press key
Press key

Restart automatically Fail

 Zero display setting: 0~9, when the weight is within +/- of the set values, it display as zero.
 Local G value conditions: The new Local G values must be within ±10% of the Local G
value set in the factory.
 If the external weight calibration has been used, the Local G value is disable.
 To enable it, please switch the jumper SWA1 to ADJ position first. Then do Weight
Calibration Setting (See 3-2-1).
→0← kg
2-4-2 Check Voltage .
Weighing mode

In weighing mode and the weight returns to zero, ▼1 ▼6


Press key
Press key for 1 second to check voltage. for 1 second
It returns to weighing mode after 3 second.
. Voltage
Returns to weighing
mode after 3 second

16
2-4-3 External Weight Calibration
In weighing mode and the weight returns to zero, Press key for 3 second to enter into
password display. Press keys in order and enter into External Weight Calibration.
Otherwise it will returns to weighing mode.

→0← kg
.
▼1 ▼6

Press key
for 3 seconds

Press
keys in order Press key to
move the cursor to
The display shows the kg
the left. kg
calibration value and unit. .
.
The right digit keeps
flashing.
Press key while Press key to
there is no load on abort zero calibration
platter.
kg
The scale is managing
the data of zero point
When it becomes stable, 2 seconds later
the display shows the
calibration weight.
kg
.
While no load, press key
Put the weight on the and then press key to
platter and press key. abort full capacity calibration
The scale is managing
the data of full capacity.
When it becomes stable,
the beeper beeps 3 times. 2 seconds later
Weight calibration
finished and then
take off the weight

Press key return


to the weighing mode After aborting zero calibration, please put full
weight on the platter and wait till the scale
→0← kg
stable. Then it will calibrate automatically.
.
▼1 ▼6
 Weight calibration conditions:
The calibration weight value placed on the platter must be over 100e, and the standard
deviation of the weight must be within ±10%.

17
2-4-4 Transmission Setting
1. In weighing mode, press key for 1 second to enter into transmission setting.
2. Press key to select transmission →0← kg
Weighing mode
No transmission .

Stable transmission ▼1 ▼6

Continuous transmission Press for 1 sec Press key


Press key to transmit
(Simple mode)
Press key to transmit
(Complete mode) Press key Press key
 Stable transmission
The format is as same
as

3. Press key to set baud rate Press key


Press key
(1200, 2400, 4800, 9600 or
19200.)

Press key Press key

4. Press key to save the setting and


return to weighing mode
Baud rate setting

Press key
→0← kg
Save the setting and
.
return to weighing mode
▼1 ▼6

 RS-232 Output Format


The output format of and
GROSS S T , G S , + 1 2 3 4 5 6 7 SP SP o z
Net S T , N T , + . 2 3 . 4 5 6 t l . g
Plus OL O L , G S , + SP SP SP SP SP SP SP SP SP SP SP CR LF
Minus OL O L , G S , - SP SP SP SP SP SP SP SP SP SP SP
Unstable U S , G S , + 1 2 3 4 . 5 6 SP SP l b

Stable transmission:when the weight is more than 10d and the weight is stable, it
transmit above format. It transmits next data format only when take off the weight
on the platter and put it on again.

Continuous transmission:It always transmit above format.


Output rate: 4counts / second

18
Press key to transmit (simple mode):use in accumulation mode.
Steps: 1. Press F key and select . Press key to enter into weighing mode.
2. Select and necessary baud rate.
3. Put the objects on the platter. After the weight stable, press key to
accumulate the data and print out the order and the net value of the data as
following example.
4. Repeat above step to accumulate and print continuously.
5. Press key twice to print TOTAL as following example and clear the data.
S/N WT
---------------------------------------
0001 2.500  Press key
0002 2.500  Press key
0003 2.500  Press key
---------------------------------------
0003 7.500  Press key twice to print TOTAL

Press key to transmit (complete mode):use in accumulation mode.


Steps: 1. Press F key and select . Press key to enter into weighing mode.
2. Select and necessary baud rate.
3. Put the objects on the platter. After the weight stable, press key to
accumulate the data and print out the data as following example.
TICKET NO. 0001
GS 2.500kg
TR 0.000kg
NT 2.500kg  Press key
4. Repeat above step to accumulate and record.
5. Press key twice to print TOTAL as following example and clear the data.
TOTAL NUMBER
OF TICKETS 0003
TOTAL WEIGHT
2.500kg  Press key twice to print TOTAL

Stable transmission (the format is as same as ) in accumulation mode.


Steps: 1. Press F key and select . Press key to enter into weighing mode.
2. Select and necessary baud rate.
3. Put the objects on the platter. After the weight stable, it automatically
accumulates the data and print out the order and the net value of the data as
following example.
4. Repeat above step to record and print continuously.
5. Press key twice to print TOTAL as following example and clear the data.

19
Chapter 3 Service Mode Access
 Open the case, and then switch the mini-jumper SWA1 on the main board to the ADJ
position. Turn the power on. The display shows .
 When finished, set the jumper SWA1 back to the LOCK position.
 If the jumper SWA1 is returned to the LOCK position during calibration, the scale exits the
service mode automatically.
The Service Mode Access contains 5 settings as below:

Press key

Press key Press key

Press key Press key

Press key Press key

Press key Press key

Press key

 Specification Setting
 Weighing Calibration
 Linearity Calibration
 Function Setting
 Local G Value Setting

20
3-1 Specification Setting

Press key

Input a parameter
“00~03”

 Upward key (0~9 digit entry)


After inputting 00, press
 Downward key (0~9 digit entry)
key to return to the
 Move cursor rightward
previous step.
 Move cursor leftward

 Return to the previous step


 Weight Units Setting
 Customised Weight Units Setting
 Multi-segment Setting

21
3-1-1 Weight Unit Setting
 There is no resolution limitation when the weight units “kg, g, lb, lb/oz” are selected.
 The weight units”oz, GN, dwt, and ct” are only available on indicators with less 1/10,000
external resolution.

The display shows previous setting


Press key is max capacity of 15kg with
kg 0.01kg/division for 2 seconds.
.

Example:
Two weight UNITS “kg” & “lb”, with
“kg” as the UNITS used for weight
Refer to Note 1
calibration.

Press key
To set up for max capacity of 15kg with
0.001kg / division
Enter 015000  15.000 kg
Refer to Note 2

Enter division (= m * 10-n*q) and segment setting


1  base value (m) = 1
Press key 3  n = 3, 3 decimal point
2  2 segment range
0  multi-intervals
1 multiplication factor (q) = 0,
Refer to Note 3 division = 0.001 kg

The display shows new setting is max


capacity of 15kg with 0.001 kg/division
Press key for 2 seconds. And back to the upper
kg level.
.
 Upward key (0~9 digit entry)
back to previous setting
 Downward keyfunction.
(0~9 digit entry)
 Move cursor rightward
 Move cursor leftward
Input parameter “00”
Press key
to complete the
setting

22
Note 1
The users can set up the different weight units in various orders according to their preference,
and the amount of the chosen weight units can be up to 5

(a)(b)(c) (d)(e) (f)


(a)  The first weight unit (only “kg”, “g”, or “lb” are available to choose from. Please select one of
the parameters 0, 1, or 2)
(b)  The second weight unit (select one of the parameters described below)
(c)  The third weight unit (select one of the parameters described below)
(d)  The fourth weight unit (select one of the parameters described below)
(e)  The fifth weight unit (select one of the parameters described below)
(f)  The amount of the weight units selected (select one of parameters 1 ~ 2)
The description of the parameters
0 kg (Decimal system) 5  Reserved
1 g (Decimal system) 6  Reserved
2 lb (Decimal system) 7  Reserved
3  Reserved 8  Reserved
4  Reserved 9  Reserved
For example:
Choose “kg” & “lb” (two weight units). NOTE the scale is calibrated using “kg” weights and key
in 020002

Note 2 Enter the maximum capacity of the scale, total 6 digits (not including 9d)

(g) (h) (i) (j) (k) (l)


For example:
15.000 kg  key in 015000
1500.0 g  key in 015000
6.000 lb  key in 006000

Note 3 Set the minimum division and decimal point position to determine the display resolution

(m) (n) .(o) (p) .(q)


Division = m * 10-n*q, m = base value, n = numbers of decimal point, q = multiplication factor
(m)  Division base value, select 1, 2, or 5
(n)  The number of decimal places (0 ~ 5)
For example:
15.000 kg  enter 3, 1500.0 g  enter 1, 6.000 lb  enter 3
(o)  range setting (select one of parameters 0, 1, 2, or 3)
For example:
0, 1  full segment range,
2  2 segment range (divided at 1/2 of the full scale),
3  3 segment range (divided at 1/6 of the full scale & 2/3 of the full scale)
(p)  0 : multi-interval 1 : multi-range

23
Multi interval: Multiple segment range and each segment with its own minimum and
maximum capacity and scale interval. The selection of the appropriate
weighing segment is determined automatically according to the load applied,
both on increasing and decreasing loads.

Multi range: Similar to Multi-interval, but the scale interval unchanged when unloading until
weight return to zero
2 Segment range:

Multi interval Multi range

3 Segment range:

Multi interval Multi range

(q)  Division multiplication factor: (Only one Weight UNITS Model is available)
0  no factor 1  base value X 10
Division Table for various m and q values:
(m) = 1 (m) = 2 (m) = 5 (m) = 1 (m) = 2 (m) = 5
(q) = 0 (q) = 1
1 2 5 10 20 50

 If 2 segment range and multi-interval is set, tare automatically cancels out after weight is
cleared.

24
3-1-2 Customized Capacity Setting

The display shows current setting


is max capacity of 6 lb with 0.001
Press key lb/division for 2 seconds.
kg
.
The display shows the number of weighing
UNITS and the UNITS for the weight
calibration set previously

Refer to Note 4 For example:


5  5 weight UNITS
2  “lb” is the weight UNITS for calibration

Press key
Show the code of weight UNITS for weight
calibration for 2 seconds
 “ lb ”
“g”
 “ kg ”

Refer to Note 5 For example:


012000  12.000 lb

Key in the mass value


Press key 012000  12.000 lb

Refer to Note 5 For example:


5  division base value = 5
3  3 decimal points  division = 0.005
2  “ lb ”
Press key 0  full segment range
1  save as the first selected weight unit
0  multi-interval

Refer to Note 6 The display shows new setting is max


capacity of 12 lb with 0.005 lb/division
for 2 seconds. And back to the upper
level.

Press key

lb Cont
.

25
Cont

The display shows current setting


is max capacity of 12 lb with 0.005
lb/division for 2 seconds.

Press key
To 2nd and
setup+1d weight unit weight
for maxfor
apacity UNITS
lb capacity of 6000g with 1g/division
. 2 seconds.
Enter 006000  6000 g

Enter equivalent mass value


Refer to Note 5 12 lb = 5443 g
005443  5443 g

For example:
Press key
1  division base value = 1
0  0 decimal points  division = 1
1“g”
Refer to Note 5 to input 0  full segment range
Weight value 2  save as the second selected weight unit
0  multi-interval

Press key
The display shows the 2nd weight unit
has max capacity of 6000g with 1
g/division for 2 seconds.
Refer to Note 6 to input
parameter
apacity +1d and UNITS weight for 2
 Upward key (0~9 digit entry)
seconds.
 Downward key (0~9 digit entry)
Press key , then 
back to upper
Move cursorlevel.
rightward
g
 Move cursor leftward
Set all weight units
according to preference After all weight units desired have been
set

Input 00 Press key


to complete the
setting

26
Note 4

(a) (b)
(a)  The number of the weight units (Max: 5, key in 1 ~ 5)
(b)  The weight unit for weight calibration (choose from “kg”, “g”, “lb”, key in 0, 1, or 2)
Parameter description:
0  kg , 1  g , 2  lb

Note 5

(c) (d) (e) (f) (g) (h) (i) (j) (k) (l) (m) (n)
c ~ h set the maximum capacity (6 digits)
i ~ n set the mass value for weight calibration (6 digits)
The maximum capacity needs to be presented based on the decimal system, and the first unit
must be the calibration unit.
For example:
How to calculate the maximum capacity and the mass value based on the different types of
weight unit.
A. Choose “kg” as the weight unit for the weight calibration
 The first weight unit setting: 6.000 kg / 0.002 kg
 Enter the maximum capacity 006000 at (c) ~ (h)
 Enter the mass value 006000 at (i) ~ (n)
 Unit “lb” (hexadecimal notation system)
Calibration weight is 6kg. 0.001 kg = 0.002204623 lb
6 kg = 62.204623 (lb) = 13.227738 lb. Take 12 lb as the max capacity
12.0.00 lb / 0.08 oz (minimum division=8, decimal point position=2)
12 lb 16 (oz) = 192.00 oz  Enter the maximum capacity 019200 at (c) ~ (h)
13.227738 lb 16 (oz) = 211.64 oz  Enter the mass value 021164 at (i) ~ (n)

B. Choose “lb” as the weight unit for the weight calibration


 The first weight unit setting: 12.000 lb / 0.005 lb (the maximum Capacity / division)
 Enter the maximum capacity 012000 at (c) ~ (h)
 Enter the mass value 012000 at (i) ~ (n)
 The second weight unit setting “g”
0.002204623 lb = 1 g
12 lb = 5443 g. Take 6000 g as the maximum capacity
6000 g / 2 g (the maximum Capacity / division),
 Enter the maximum capacity 006000 at (c) ~ (h)
 Enter the mass value 005443 at (i) ~ (n)

27
Note 6

(o) (p) (q) .(r) (s) (t)


(o)  Minimum division setting
Parameter description:
Decimal system: Input 1, 2, or 5 as the minimum division for the weight value
Hexadecimal system: Input 1, 2, 4, or 8 as the minimum division for the weight value
(p)  Decimal point position
Parameter description:
Decimal system: Hexadecimal notation system:
00 1  0.0 0  0.0 1  0.0.0
2  0.00 3  0.000 2  0.0.00 3  0.0.000
4  0.0000 5  0.00000
(q)  Weight unit displayed
Parameter 0 1 2
Unit kg g lb
Notation
10 10 10
system
symbol kg g lb

(r)  Scale change point (Input the parameter 0, 1, 2, or 3)


Parameter: 0  full range 1  full range 2  dual range (changes at 1/2 of full scale)
3  triple range (changes at 1/6 of full scale and 2/3 of full scale)
(s)  Save the weight units at preferred slots (no more than the number of set weight units)
Parameter: 1  the first slot (the weight calibration unit) 2  the second slot
3  the third slot 4  the fourth slot 5  the fifth slot
(t)  0: multi-interval 1: multi-range

28
Customized Capacity Setting- How to Revise

The display shows current setting


is max capacity of 12 lb with 0.005
lb/division for 2 seconds.
Press key
Display shows parameter settings
lb from last time.
.

For example:
5  5 weight UNITS are used
2  “lb” as weight calibration unit
Refer to Note 4 input the Display shows weight calibration UNITS
parameters or for 2 seconds:
do nothing  “ lb ” calibration
 “ g ” calibration
Press key  “ kg ” calibration

For example:
019200  192.00 oz

Please enter weight calibration value in


proper UNITS value. For example:
12 lb = 192 oz
Refer to Note 5 to input 019200  192.00 oz
parameters
For example:
5  division base value = 5
2  2 decimal points  division = 0.05 and
Press key display as 0.0.05
4  “lb-oz “
0  full segment range
3  save into third UNITS slot
Refer to Note 5 to input
0  multi-interval
Weight value
The display shows the 3rd weight
unit has max capacity of 12 lb with
Press key 0.05 oz/division for 2 seconds, and
it returns to last screen

 Upward key (0~9 digit entry)


Refer to Note 6 to input
parameters  Downward key (0~9 digit entry)
 Move cursor rightward
 Move cursor leftward
Press key
lb Continue to do another weight
. . unit revision, or exit
customized capacity

29
3-1-3 Multi-range Setting

Select the desired unit for setting


All the units which set at ,
Press key could be selected to set
the interval point.
Refer to Note 1 for unit setting.

Input the unit parameter


The display shows the first point setting value
2 segments model  the point is at 1/2 full scale
3 segments model  1st point is at 1/6 full scale
Press key 2nd point is at 2/3 full scale

Input the desired “first point”


Input the desired “first The first point could be set for the 2 or 3
point” segments model.

The display shows the second point setting value


3 segments model  1st point is at 1/6 full scale
Press key 2nd point is at 2/3 full scale

Input the desired “second point”


Input the desired “second
The second point only could be set for
point”
the 3 segments model.

Press key

Follow the steps


to input the other When all the points are entered,
points. please input “00” to finish the setting.

 Upward key (0~9 digit entry)


Press key  Downward key (0~9 digit entry)
to finish the setting  Move cursor rightward
 Move cursor leftward

 By the specification setting of , , the default separation points for the 2


segments model is 1/2 full scale, and 1/6 full scale and 2/3 full scale for the 3 segments
model. To change the interval point setting, please use the function.
 When the specification settings of and have been changed, the
separation points would be reset to the default setting.

30
3-2 Weight Calibration

Press key

Input a parameter “00~02”

After inputting “00”,,press  Upward key (0~9 digit entry)


key return to previous  Downward key (0~9 digit entry)
step.  Move cursor rightward
 Move cursor leftward

 Return to Previous Step


 Weight Calibration
 Local G Value Setting

31
3-2-1 Weight Calibration Setting

Press key, the display shows


the calibration value (the max
capacity of 1st unit setting) and
unit. The right digit keeps flashing. Press key to move
The calibration value kg the cursor to the left. kg
could be changed. . .

put the platter on the scale


and press key while Press key to
there is no load on platter. abort zero calibration
kg
The scale is managing
the data of zero point
When it becomes stable, 2 seconds later
the display shows the
calibration weight.
kg
.
While no load, press key
put the weight on the and then press key to abort
platter and press full capacity calibration
kg
.

The scale is managing the 2 seconds later


data of full capacity. When Then take off the weight
it becomes stable, the and press any key to
beeper makes 3 “beep” exit the calibration
sounds and the display mode.
keep flashing.

Input“00”
 Upward key (0~9 digit entry)
 Downward key (0~9 digit entry)
 Move cursor rightward
Press key to finish
 Move cursor leftward
the setting.

After aborting zero calibration, please put


full weight on the platter and wait till the
scale stable. Then it will calibrate
automatically.

32
3-2-2 Manufacturing Location G Value Setting

Press key

Manufacturer: Taipei Manufacturer: Shanghai Manufacturer: Other


Note 1 Note 1

Note 1: Press key


Use key or
to select the .
Press key manufacture location.
The display shows the
Press key last setting. And then
input the G value of
manufacture location.

Press key

Input“00”

Press key to
finish setting

 Upward key (0~9 digit entry)


 Downward key (0~9 digit entry)
 Move cursor rightward
 Move cursor leftward

33
3-3 Linearity Calibration
 After linearity calibration, it is necessary to re-calibrate the weight.
 Weight calibration should be included in linearity calibration. The last point of linearity
The first point of linearity calibration with calibration with plastic
AD plastic platter or aluminum support. platter or aluminum
support.

The zero point with steel platter or


0 plastic platter. 524288

Remove all weight from the pan,


Press key enter linearity
calibration mode.

Remove the steel weigh pan, and


leave the plastic one, ensure the pan
is empty. Press key to record *Weight factor
the “first point”. Refer to note below

Apply the 1st weight for the


second point, enter the weight
factor, press key to
record “second point”

Apply the 2nd weight, enter the


weight factor, press key to
record “third point”

Apply the 3nd weight, enter the


weight factor, press key to
record “fourth point”
 Upward key (0~9 digit entry)
 Downward key (0~9 digit entry)
Add 8th weight, enter weight factor,  ENTER
press key to record the ”ninth
 ESC
point”

34
 Weight factor

The weight factor is a single hexadecimal number which represents the value of the next
weight compared to the size of the first weight applied to the scale.
The weight factor is arranged as follows:-
1 = The weights are equal.
2 = The next weight is twice as big as the first weight
3 = 3 times as big as the first weight
4 = 4 times…
5 = 5 times …
6 = 6 times..
7 = 7 times…
8 = 8 times…
9 = 9 times….
A = 10 times…
B = 11 times…
C = 12 times…
D = 13 times…
E = 14 times…
F = 15 times as big as the first weight

Exampes: 30kg scale to be linearitied with the weight values shown in brackets:-
Ex1: 30kg (10kg ,10kg ,10kg)
Display Key Press Note
03 CLn Into linearity calibration
L0 First point (zero), remove weigh pan and press the
L1 1 Put 10kg on and press the key
L2 1 Put 10kg on and press the key
L3 1 Put 10kg on and press the key
L4 1 To finish linearity adjustment (4 points linearity calibration)
03 CLn

Ex2: 30kg (5kg, 10kg ,10kg, 5kg)


Display Key Press Note
03 CLn Into linearity calibration
L0 First point (zero), remove weigh pan and press the
L1 1 Put 5kg on and press the key
L2 2 Put 10kg on and press the key, 2 is the rate of L1
(10kg is 2 x 5kg, which was used in L1)
L3 2 Put 10kg on and press the key
L4 1 Put 5kg on and press the key
L5 1 To finish linearity adjustment (5 points linearity calibration)
03 CLn

35
Ex3: 30kg (5kg, 5kg ,10kg, 10kg)
Display Key Press Note
03 CLn Into linearity calibration
L0 First point (zero), remove weigh pan and press the key
L1 1 Put 5kg on and press the key
L2 1 Put 5kg on and press the key
L3 2 Put 10kg on and press the key
L4 2 Put 10kg on and press the key
L5 1 To finish linearity adjustment (5 points linearity calibration)
03 CLn

Ex4: 30kg (1kg, 2kg ,5kg, 10kg, 2kg, 10kg)


Display Key Press Note
03 CLn Into linearity calibration
L0 First point (zero), remove weigh pan and press the
L1 1 Put 1kg on and press the key
L2 2 Put 2kg on and press the key
L3 5 Put 5kg on and press the key
L4 A Put 10kg on and press the key
L5 2 Put 2kg on and press the key
L6 A Put 10kg on and press the key
L7 1 To finish linearity calibration (7 points linearity calibration)
03 CLn

 In the process of , , press key to abort the linearity calibration.


 In the process of , , , , , , or press key to finish and
save the 2, 3, 4, 5, 6, 7, or 8 points calibration.
 In the process of , press key to finish and save the 9 points calibration.

36
3-4 Function Setting

Press key

Input a parameter 00、01、06

 Upward key (0~9 digit entry)


 Downward key (0~9 digit entry)
Input “00” and press
 Move cursor rightward
key to return to the
 Move cursor leftward
previous step

 Return to the previous step


 Environment parameters setting
 Zero Tracing Setting

37
3-4-1 Environment Parameters Setting

Return to zero
The display shows the last setting

Press key Return to zero point


Using or to enter
 Default setting = 0
0  show all 5  within 5 d
Use or keys 1  within 1 d 6  within 6 d
to enter the parameter 2  within 2 d 7  within 7 d
3  within 3 d 8  within 8 d
4  within 4 d 9  within 9 d
 Weight value must over 1/3 full scale
Press key
Stabilization range
Display shows the last setting

Use or keys Stabilization range


to enter the parameter Use or keys input the parameters.
 Default setting = 0
Parameter 0 ~ 9, the larger the number the
more stable the weight.
Press key
Filter setting
Display shows the last setting.

Use or keys to Weighing Filter setting


enter the parameter
Use or keys input the parameters.
 Default setting = 5
Range 0 ~ 9, the larger the number, the faster
the filter response. Fast response could lead
Press key
to display weight instability.
Parameter 9  the AD value is not filtered.
Input AD value = Output AD value

Input“00”
 Upward key (0~9 digit entry)
 Downward key (0~9 digit entry)
Press key to  Move cursor rightward
finish the setting  Move cursor leftward

38
3-4-2 Zero Tracking Setting

Press key Switch on/off zero tracking


The screen shows last parameter setting

use or keys to
enter the parameter Switch on/off zero tracking
Use or keys to input parameter
 Default setting = 0
0  Zero tracking is off
Press key 1  Zero tracking is on

Input“00”  Upward key (0~9 digit entry)


 Downward key (0~9 digit entry)
 Move cursor rightward
 Move cursor leftward
Press key to
finish the setting

3-5 Local G Value Setting


 The G value should be among the
value of Equator and Polar.
Acceleration of gravity at the Equator:
Press key GE = 9.7803184558 m/sec2
Acceleration of gravity at the Poles:
.
GP = 9.8321772792 m/sec2
Taipei ≒ 9.78914 m/sec2
Enter the G value of the
customer’s location Shanghai ≒ 9.79423 m/sec2

.
 Upward key (0~9 digit entry)
Press key  Downward key (0~9 digit entry)
 Move cursor rightward
 Move cursor leftward

39
Appendix 1 7-Segment Display Characters
Number Display Letter Display Letter Display

0 A N

1 B O

2 C P

3 D Q

4 E R

5 F S

6 G T

7 H U

8 I V

9 J W

K X

L Y

M Z

40
Appendix 2 Installation
1. Please connect the Load Cell and Indicator as following
5PIN Connector
PIN 1 E+
PIN 2 E-
3 4
PIN 3 S+ 2 5
PIN 4 S- 1

PIN 5 GND

2. Connecting way of a 9 pin RS-232 connector as following.


PIN 2:Data inputting (TXD), position 2 in following picture
PIN 3:Data outputting (TXD), position 3 in following picture
PIN 5:Internal grounded (GND), position 5 in following picture
2 3 5

Relay Output
1. Principle of operation
Use check weighing function to set the ranges of High, OK, LOW.
If the weight is between the ranges of LOW, the data will output in LOW port;
If the weight is between the ranges of OK, the data will output in OK port;
If the weight is between the ranges of High, the data will output in High port;

 Please refer to check weighing function to set the ranges of High, OK, LOW.
2. Connecting ways and function
Terminal Block Connector’s PIN Function
A1 PIN 1 High Output
B1 PIN 2 OK Output
C1 PIN 3 Low Output
COM PIN 4 COMMON

41
42
43