JSMT 200
JSMT 200
dEDL
JSM-T200
SCANNING MICROSCOPE
No.ISM·T200-1
(EP165001)
7912003KP
T200
CONTENTS
Page
l o INTRODUCTION · ....... . .................. . ....................... 1-1
1. 1 General ..................................•........•....... 1-1
1. 2 Specifications ...................•..•....•....•........... 1-3
1. 3 Layout , dimension s and weight ...........•..••..........•.. 1-6
7907001 T200
T200
CHAPTER 1 INTRODUCTION
1.1 GENERAL
The scanning electron microscope is a comparatively recent addition to
the microscope family and is pr oving to be very popular along with the long
established light and transmission type electron microscopes .
In spite of their merits and demerits , each type of microscope has its
role to play depending on the field of application . For example , if a high
resolving power and large depth of focus is not called for, a light micro-
scope would be ideal. Where a very high resolving power is required , a
transmission electron microscope would be necessary . However, when using a
transmission electron microscope, the specimen must be very thin (less than
1 ~m (10,000 A) in thickness) , a factor which requires a great deal of skill
on the part of the user in order to prepare such specimens.
The scanning electron microscope, on the other hand, offers a fairly
high resolution and moreover, since it is possible to use bulk specimens
(specimen thickness being of no consequence), specimen preparation is easy .
In addition to which, the depth of focus is large, thereby enabling 3-D
observation.
The operational principle of the scanning electron microscope is
illustrated in Fig. 1.1. A finely focused electron probe is made to scan
the specimen, resultant upon which, secondary and backscattered electrons,
etc. (Fig . 1.2) are emitted from the surface of the specimen. These signals
are then detected by a detector and outputted via an amplifier to a synchro-
nously scanned CRT as an intensity variation signal . The CRT raster width
divided by the electron probe scanning width gives the image magnification.
By adding an appropriate detector (optional) to the standard T200,
transmitted electron images, cathodoluminescence images, etc. can be
observed in addition to secondary and backscattered electron images.
Further, by incorporating an X- ray detector, X-ray analysis becomes
possible.
Condenser
Objective
Beam deflector
coils ----~~
Amplifier
Incident electron
beam CRT
Secondary electrons,
Specimen other signals
T200
Specimen 1-------,
~
Electromotive force Absorbed electrons
Q)
~ ~ Transmitted electrons
Q)~
+.J
+.J ~
'g +.J2
CIJ
~ ()
CIlQ)
~..-I
H Q)
T200
1.2 Specifications
1. Technical data
Performance
o
Resolution: 10 nm (100 A) at 25 kV and 20 rom working
distance.
Magnification: l5 x to 100,000x (15 x available at work-
ing distance 48 rom only).
Electron optical system
Accelerating voltage: 2, 5, 10, 15, 25 kV.
Electron gun filament: Precentered cartridge tungsten filament.
Lens system: 3-stage demagnifying system (2-stage
condenser lens and objective lens).
Alignment: Mechanical.
Stigma tor: 8-pole electromagnetic type.
Image fine shift: Up to ±10 ~m (25 kV) in any direction;
electromagnetic, joystick control •
• Specimen stage (twin stage)
Specimen Up to 76 . 5 dia. x
Up to 10 dia. x 10 thick*(rom)
accomodation 25.5 thick** (rom)
Specimen
By drawing out the stage
ex change
Signal
Optional (max . 48 pins)
terminal
T200
T200
. Vacuum system
Operation: Fully automatic.
Ultimate pressure: 7 x 10- 4 Pa (5 x 10- 6 Torr).
Vacuum gauge: Pirani gauge.
Pump-down time: About 30 minutes (from cold).
Specimen exchange: About 2.5 minutes.
Oil rotary pump: 100 .R. /min 1 unit.
Oil diffusion pump: 420 .R. /sec 1 unit.
. Safety devices: Devices for power failure, water
failure and vacuum deterioration:
built-in.
. Miscellaneous: Service outlet: built-in (100 V, 2 A;
used for operating optional attach-
ments) .
The instrument can be wheeled on its
own casters.
2. Installation requirements
. Power and water
Power: 100 V, 50/60 Hz, single phase, 2 kVA
(basic instrument: 1.2 kVA;
attachments: 0 . 8 kVA).
Starting current 60 A (0.2 sec.)
Fluctuation: less than ±10% (that also
includes initial operation).
Ground terminal: 1 terminal, less than 100 Q.
Cooling water: Flow rate ... 2 .R./min at 0.05 - 0.2 MPa
(0.5 - 2 kg/cm 2 ).
Temperature ..• 20 ±5 °c (water
temperature at the
outlet: not greater
than 35°C).
Faucet .. .•.. . 1, 12 mm O.D.
Drain .....•.• 1•
• Installation room
Room temperature: 20 ±5 °C.
Relative humidity: Less than 80%.
Floor vibration: Less than 2 ~m p-p (5 Hz) in the X, Y
and Z directions, less than 3 ~m p-p
(10 Hz) in the X, Y and Z directions,
and less than 8 ~m p-p (50 Hz) in the X,
Y and Z directions (with the instrument
installed).
T200
2,000
c,...,
o
o
t
'"
Faucet
Water hose
Y~ Drain
Line cable
Line cable
terminal
I I
I I
o
o
~~
o
N '"
~ Table
(height: 750 mm)
r-----..., \
Specimen stage II I
\ J
,....... )( Operator
r---,
'---_.I
Column
(height: 1250 mm from floor) Unit: mm
I +
Doorway (800 wide x 1800 high)
T200
Control panel
Power switch
T200
Lar
Twin s
Eucentric
ecimen chamber
T200
CD CRT
~ Blank plate (to be removed when mounting an optional CRT for photo-
graphy )
The display panel lamps indicate the operating state, etc. of the
microscope as follows:
T200
® BRT I
Lights up when the BRIGHTNESS I IAUTO/MAN I
button is depressed.
@ FOCUS I
Lights up when the FOCUS I IAUTO /MAN I
button is depressed.
T200
2.3 Checker
This checker enables the operator to pinpoint a faulty circuit or unit
by monitoring 21 voltages and currents.
1. Auto checker
The auto checker automatically makes known a failure in any of the six
main power supplies (check points 6, 7, 8, 9, 10 and 15 in Fig. 2 . 4).
That is, if any of the voltages deviate by ±10% or more from the value
listed, the POWER lamp goes off to indicate an abnormality. The actual
power supply at fault is identified by positioning the POSITION selector
at the above check points and noting the meter reading at each position.
2. The POSITION selector enables the operator to monitor 21 voltages and
currents, including the above six main power supply voltages and various
other operational parameters such as 0 V, mains power voltage, vacuum,
load current, CRT and PMT high tensions, and CL and OL exciting currents.
Check points 22 to 24 have been allotted for attachments.
Note: Wh en calling in yo ur nearest JEOL Service Center in connection
with th e above, please indicat e the check point/points at fault
and the voltage or current reading/readings in question.
lamp selector
al select button
SEI: For observation of secondary electron images
BEl: For observation of backscattered electron images
T200
1 ZERO CHECK
2 POWER AC 200 V
3 VACUUM 6 V
4 LOAD CURR 200 JlA
5 CRT HT 20 kV
6 78 V
7 -78 V
8 30 V
9 -30 V
10 10 V
11 60 V
12 -60 V
13 60 V
14 -60 V
15 48 V
16 1100 V
17 CL CURR 2 A
18 OL CURR 2 A
19 15 V
20 -15 V
21 PMT HT -2200 V
22
23
24
} For attachments
T200
CD IFILAMENT I control
Controls the electron gun filament heating current.
Caution: To prevent overhea tin g the filament which curtails the fila ment
service life, avoid turn ing th e control excessively.
When the filament burns out, the monitor lamp goes out.
o IIMAGEIIMODEI buttons
LSP For observation of signal waveforms (slow scan only).
YMD For observation of Y-modulated images (slow scan only).
PIC
For observation of slow-scan and TV-scan images.
TV
CD IIMAGEIISIZEI buttons
EXP: For image focusing, astigmatism correction and exposure determi-
nation (by Rapid Exposure Marker).
PHOTO : For checking the photographing area.
FULL: For selecting the field of view.
(0 ISTIGMATORI controls
Used to correct astigmatism (that is, to obtain a perfectly round electron
probe) • When the astigmatism is corrected, the defocused image does not
exhibi t directional blurring. It is recommended that correction be
carried out at a magnification of 10,000 x or higher.
G) IIMAGE SHIFTI joystick
The joystick allows the image to be shifted electromagnetically up to f lO
f.Jm.
T200
For magnifications of
less than 5000 x ~
r;0
SPOT SIZE
For backscattered
electron images
- -+--,.-----_-1 Contrast
~----~ Brightness
® IFOCUS I controls
These controls are for focusing the image on the CRT screen. By depres-
sing the center and right IAUTO I buttons, the automatic focusing mode is
established.
@ IMAGNIFICATION ! cotnrol
. Selects the magnification of the CRT screen image. The magnification is
displayed on the display panel.
T200
CHAPTER 3 OPERATION
3.1.2 Shutdown
1. Depress the IpOWERlloFFI switch.
Notes: 1. If the column is not under high vacuum, evacuate the column
by pushing the IpUMP DOWN I switch before depressing the
IPOWER II OFF I switch.
2. If the microscope is going to be out of use for a prolonged
period of time, be sure to evacuate the column to prevent
corrosion.
2. Turn off the mains power switch, wait 10 to 15 minutes for the diffu-
sion pump to cool down to room temperature, then close the water
faucet.
Note: Although it is preferable to wait 10 to 15 minutes for the
diffusion pump to cool down to room temperature, shutting off the
cooling water immediately after turning off the power switch has
no adverse effect on the microscope.
3. 1 .4 Water failure
The microscope stops automatically. As in the case of power failure,
when water is restored, it is necessary to manually reactivate the micro-
scope by depressing the IpOWERlIoNI switch.
Note: Even when the cooling water is restored, depressing the IPOWER Ii ON I
switch 'will not reactivate the microscope unless the diffusion pump
thermostat has cooled down to its normal operating temperature.
T200
"
~ ';)1\ Front Specimen holder
T200
T2.00
• Specimen holder
x
CRT screen The directions of X and Y
About 15 ° arrows correspond to the
respective rotated directions
y of the X and Y controls
(Fig . 3.4).
T200
Y control 5 mID .
T200
Observation
1. Depress (ON) the IACCELERATING VOLTAGE I button. The raster now appears
over the entire CRT screen.
2. Depress the IIMAGEllMODEllLSpl and IIMAGE II SIZEI IEXPI buttons.
A line should now appear on the lower part of the CRT screen. If not,
gradually turn the IBRIGHTNESSI control clockwise until the line
appears.
D
A line appears.
D [ ~ !
J Note the height.
----.
No electron beam With electron beam
5. Turn the IFILAMENT I control beyond the 2 o'clock position until the
waveform remains stationary regardless of how much further the control
is turned.
Note: As the control is turned beyond the 2 o'clock position, it will
be noted that ~he waveform first moves down and then moves up
before reaching the stationary state.
6. Once the stationary position has been established, slowly turn the
IFILAMENT I control counterclockwise until the waveform starts to move
down abruptly. Stop turning the control counterclockwise at this point
and turn the control clockwise (just slightly) so as to position the
control at the point just prior to where the waveform starts to fall
(for details on the IFILAMENT I control setting procedure, see Sect.
3.6.1).
T200
7. Depress the IIMAGE I MODEl IPlci button. The rapid exposure marker
together with a raster should now be visible on the CRT.
Note: If only the rapid exposure marker is visible, turn the IBRIGHT-
NESsi and/or CONTRAST controls clockwise until the raster
appears .
[i]
No raster
~----~~----~Contrast
Brightness
10. Increase the magnification to the desired value and focus the image
with the IMAGNIFICATION I and IFOCusl controls.
Note: When the magnification is considerably changed, the rapid
exposure marker changes. In this case, readjust the related
controls.
T200
--
I
bar appears beyond the standard I
contrast control
After first obtaining optimum (Dark)
image brightness and contrast at a
magnification of about 1000x, b. Low
contrast ~~-~~---~---\~--------~---
establish the automatic control (Bright)
mode by depressing the AUTO/MAN
buttons on the control panel . By • •
. ,. -
so doing, the ICONTI and IBRTllamps on c. Excessive brightness
the display panel light up.
To set the image contrast and/or
d.
- _ ---.-----~---....\ . .
Ir..
Insufficient brightness
brightness as you desire, release
the IAUTO/MAN I button on the
ICONTRASTland/o r IBRI GHTNESS lpanel(s)
• - - -
and adjust the related control(s).
Fig. 3.7
T200
~o
a b c d
T200
T200
T200
Focusing knob
Shutter button
Latch
Connector
Fig. 3 . 9 CS1-1
Hinge pin
Latch
Dark slide
Fig. 3. 1 0 CS I - 2
T200
Camera body
Viewfinder
Latch
Automatic take-up
device
~------Release cord
Fi g. 3.11 CSI-3
Hinge pin
Shutter button
Polaroid If 545
Aperture adjust sheet film holder
Latch
Lens
Manual shutter
T200
T200
Film speed
f-number
ASA Dru
50 18 5.6 - 8
75, 100 21 8 - 11
200 24 11 - 16
400 27 16 - 22
C~
3.5.4 Use of ultrahigh speed Polaroid Land sheet film (type 107,57 , etc.)
( ,
_____________
-
~#I ~\ ~
T200
T200
• Hagnification
(Length of D (rum))
(Magnification)
(Calibrated length of micron bar (~m)
15 rum 15 rum
Ex. 15
1000 ~m 1 rum
Note: The central micron bar should be used to ascertain the length
of the mi cron bar.
T200
Electron gun
~
\
Cathode (Wehnelt unit)
Electron beam----'
Anode
T200
o
upward to such an
extent that the trace
disappears off the
screen, adjust the
T200
o
control so as to normal-
ize the situation.
0\-0\-0
approaches 3
o'clock.
2. If th e waveform
fails to saturate
and lose height
when the IFILA-
fE-G-fB
MENTI control is
turned beyond the
saturation posi-
tion, repeat
Steps 6 - 8. Set IFILAMENT I
control at
10. Re-adjust the axis align- ~he saturation position.
ment screws so as to
maximally heighten the
waveform.
11 . Check that the waveform ..
level follows the pattern
as shown in Fig. 3.15
by ret urning th e Ir::F:-:::I:=-L--;-A~ME:;::;N~T=1
control to the fully
counterclockwise position
and turning it clockwise
through the 11 o'clock, 2
o'clock and 3 o'clock
positions . Then, if all
---
is in order, turn the Maximally heighten the waveform with
the alignment screws
T200
OJ 0
caution: Operating the gun
filament in the
over-sa tura ted ..
state may curtail
the service life
of the filament
or cause it to
warp.
Notes: 1 . The first and
second peak
heights vary
t
~
] •
[ "f'NWWV\
]
slightly from
filament to
filament .
2 . With some new
filaments, the
first peak is
higher than the
second peak. Oversaturated region
2nd
,-
I
e
0
4-i
<ll
:>
(Ij
I
~
I
4-i L-- l .
, _____ -.l I L ____ --,
0 I I
I I
+J I I I
,..c;
bl)
'M
<ll
:::c:
I I I
I L _____ , I I
I , ____ .J
L----------l I
•
I
I
Fully counter- 11 0 ' clock I 11 o'clock I 3 o'clock
clockwise
I 12 o'clock 2 o'clock ~Operating
position
FILAMENT monitor
lamp lights up .
IFILAMENTI control position - - - t....
T200
T200
T200
If the filament monitor lamp remains unlit when the IFILAMENT I control
is turned beyond the 11 o'clock position, a burntout filament is indicated.
In this case, replace the filament as follows:
1. Admit air into the column by pushing the IVENTI switch.
2. Loosen the alignment screws (4 pes.), remove the electron gun by lifting
it straight up, turn it upside down and then place it on the column (see
Figs. 5.1 and 5.2).
Cautions: 1. Allow a few minutes for the Wehnelt unit to cool down
before handling it.
2. When carrying out the following steps, be sure to observe
the precautions described in Chap. 4. Furthermore, make it
a rule not to leave the column exposed to the atmosphere
for longer than necessary.
T200
3. Place the Wehnelt cap removing tool (hereafter re'f erred to as the
Wehnelt tool) over ' the Wehnelt cap, tighten the screws (see Figs. 5. 2
and 5.4), and remove the Wehnelt cap from the Wehnelt unit base by pull-
ing the tool.
Fig. 5.2
4. Loosen the screws and remove the cap from the Wehnelt tool.
5. Remove the filament by loosening the filament fixing screws (3 pcs.)
with the hexagonal key wrench and the setscrew with the screwdriver
(Fig. 5.3).
Guide Filament
groove ~
1.5 mm hexagonal ""'~ Spacer
key wrench.~~ ~lt cap
, Socket
Wa tchmaker' s
screw!~
/8
...~ ~, .. ----
•
1
/
v.
?OCket screw
____
Wehnelt cap
removing tool
screwdriver
a b
Fig. 5.3
T200
Wehnelt cap
removing tool 2.0 nun
(two lines)
Spacers
Socket screw
Fig. 5.4
9. Orientate the Wehnelt cap so as to align the cap guide groove with the
Wehnelt unit base guide pin and push the cap down onto the Wehnelt unit
base.
10. After applying the handblower to the Wehnelt unit and environs so as to
remove all traces of lint, dust, etc. , return the electron gun to the
column.
Caution: When returning the electron gu~, be sure not to twist the
electron gun cable.
11. Re-evacuate the column by pushing the !PUMP DOWN I switch.
12. Wait for the lMAGNIFICATIONI indicator to display a reading, then turn
!oNI the accelerating voltage by pressing the !ACCELERATING VOLTAGE!
button and carry out axis alignment (for details, see Sect. 3.6).
T200
T200
Pole piece
T200
Specimen chamber
Aluminum foil------)
If it proves difficult to
remove the cap by hand,
use tweezers.
@- Cap
®-- Spacer
®-- Aperture (0 .D.: 2 nun)
Aperture holder
T200
Coil spring
Pipe hOlder ~
Lead wires
T200
5. Apply a thin coat of silicone oil to the exposed end of the light pipe.
Caution: When cleaning the light pipe, use gauze lightly soaked in
alcohol. Do not use solvents other than alcohol because the
light pipe is extremely sensitive to nonalcoholic solvents.
6. Mount a new scintillator on the exposed end of the light pipe (excercis-
ing great care not to damage the scintillator), then attach the corona
ring and secure it with screws C.
Caution: Do not scratch or permit dust to settle on the scintillator or
light pipe. NEVER touch the surface of the scintillator.
7. Replace the collector and s e cure it with screws B.
8. Replace the detector and secure it with the four screws as provided.
9. Push the IpUMP DOWN I switch to re-evacuate the column.
T200
PMT cover
Flange
a-ring
Lead wire A .
Do not remove { a-ring
these parts. Light pipe
A
Lead wire B
Scintillator
Screw C
Collector
T200
.....
(f) ® Q) 9).
, t"
"
~ Cove,r
Q) Adap't ·e.r ~
-;'"
..•.
® Heater
G) Lead w'ires ..
® Socket
'-.
Q) Nuts "
T200
Rotary
Motor
Fig u 5.14
T200
FlA, FIB, FlC, F2A, F2B, F2C, F3A, F3B (from left to right)
T200
6. 1 Column
1. Turn off the mains power supply switch on the distribution board and
wait at least 30 minutes for the instrument to cool down.
2. Remove the alignment screws and remove the electron gun from the
condenser lens by lifting the gun straight up.
Caution: When handling internal parts which are exposed to vacuum, be
sure to wear clean, thin gloves. However, exercise care so
that traces of lint, etc. from the gloves do not remain in the
column as this is a cause of image deterioration.
3. Remove and disassemble the condenser lens pole pieces, and remove the
apertures (see Sect.5.2).
4. Remove the evacuation pipe and column magnetic shield.
5. Loosen off the four condenser lens holding screws and remove the
condenser lens by lifting it straight up.
Caution: When lifting the condenser lens, be careful as i t is very
heavy.
6. Remove the metal pipe from the beam deflector (see Sect. 5.4).
Caution: Take care not to drop the metal pipe.
7. Remove the beam deflector from the objective lens (see Sect. 5.4) and
remove the electrode.
8. Clean the parts specified in Table under Sect. 6.2.2 and then reassem-
ble them.
Note: When reassembling the specified parts, observe the following
precautions.
• Dust removal
As each component part is being reassembled, use a hand blower to
remove any traces of dust or lint which may have readhered to the
part. Also, if the reassembly operation is temporarily suspended
prior to completion, be sure to cover the exposed parts with saran
film or aluminum foil to prevent contamination. If these precautions
are neglected, the entire cleanup procedure will have been to no
avail. In some cases, high voltage discharge or image quality
deterioration may occur.
• a-ring and mating surface check
Inspect the a-rings and their mating surfaces for scratches, lint,
etc.
T200
6.2 Cleaning
6.2.1 Cleaning materials, tools, etc .
• Cleaning liquid (organic solvent such as trichloroethylene and alcohol)
The cleaning liquid should be of high cleaning quality, high purity,
preferably non-inflammable and volatile, and should have a high safety
factor.
Ensure adequate ventilation when using the cleaning liquid and do not
allow prolonged contact with the skin.
• Fine grain metal polish
The polish should be paste-like and easy to remove with organic solvent.
It is used for removing persistent contamination.
• Gauze or rayon paper (crepe or gauze type)
Should be of high quality and not release impurities when moistened with
organic solvent.
• Absorbent cotton
Should be of high quality.
• Toothpicks, cotton swabs (about 5 mm dia.) or cotton buds.
• Brushes
Saran fiber brushes or brushes whose bristles are unaffected by organic
solvent.
• Tweezers (supplied as accessory)
Used to handle small parts. When using the tweezers, take care not to
scratch or bend the parts •
• Screwdrivers and standard tools (supplied as accessory)
• Beaker
This should be made of stainless steel or aluminum, or enamel-coated.
Glass beakers are not recommended as they are easily broken.
• Work gloves
Any commercially available cotton, nylon, or polyethylene gloves are
suitable. Work gloves should always be worn when handling internal
parts exposed to ,vacuum.
• Ultrasonic cleaner
• Mini-drill
T200
T200
T200
Condense:r. lens
pole piece
Wehnelt cap
grG40
(f) M 3 x 4
)
! -"',J (2
pcs.)
Column cylinder C>-G40
Magnetic shield
I pipe
Condenser
~-- O~jective
~ lens <!:>-G40
Objective lens • I
pcs. )
Cover
Eucentric
stage
~,~
~~O
.
\~f:so
.
\ M 4x12
(4 pcs.)
Specimen chamber
Base plate
T200
• Specimen coating
In the case of non-conductive specimens, coat the specimens by vacuum
evaporation or sputtering in order to avoid any buildup of surface charge
and reduce damage by thermal effects of the electron beam and further to
improve the secondary electron yield ratio.
• Vacuum evaporation
Coating materials: Carbon (continuous; not in aggregate form; uniform;
highly adhesive),
Gold (non-oxidization; extremely fine particles;
low melting point; high secondary electron
emission),
Platinum-palladium alloy, aluminum, etc.
T200
T200
Container 1
Screwdriver, 1 set
Phillips-headed (3 pes.)
Screwdriver, 1 set
watchmaker's (6 keys)
Hexagonal key
wrench LLLLll 1 set
(6 pes.)
Pole-piece 1
removing tool
Pole-piece 1
assembling tool
Wehnelt cap
removing tool 1
T200
Hand blower 1
Magnifying lens 1
Fuse 1 set
(9 pcs.)
Lamp 1
. ,,' .
Specimen stub . - .,..J- r- 1-, t- .
......- ......
. . . ~ ~-,. t,. ",. +-;.f- ,~' 20
(10 dia.
~
x 5 thick rom)
.~.;.;'~,~
20
(10 dia.
x 10 thick rom)
Vacuum grease
Q 1 vial
(5 g)
T200
T200