Xi’an Desun Uniwill Electronic Technology Co., Ltd.

Digital pH Sensor User Manual

Xi’an Desun Uniwill Electronic Technology Co., Ltd.

Please Read This Manual Carefully Before Use.


Xi’an Desun Uniwill Electronic Technology Co., Ltd.

Preface
Thank you very much for choosing our products and services! Please read this
manual carefully before using this product, and save it for reference.
Please observe the operating procedures and precautions in this manual.
To ensure that the after-sale protection provided by this instrument is effective,
Please do not use and maintain the instrument in any way other than specified
in this manual.
As the failure to observe the precautions specified in this operation manual,
any failures and losses caused are not covered by the manufacturer's warranty,
and the manufacturer does not assume any related responsibility. Please keep
all documents in a safe place. If in doubt, please contact our after-sales service
department.
When receiving the instrument, please carefully open the package and check
whether the instrument and accessories are damaged due to shipping. If
damage is found, please contact ouri after-sales service department and keep
the package for returning. When the instrument doesn’t work, please do not
repair it by yourself, please contact our after-sales service department.We will
help you solve your problem in time.

CONTENT
I. Overview ...................................................................................................... 1
II. Technical Specifications ........................................................................... 1
III. Dimension ..................................................................................................2
IV. Installation ................................................................................................. 2
4.1 Installation Tips ..................................................................................2
4.2 Customer Installation Cases .............................................................3
V. Maintenance of Sensor ..............................................................................3
5.1 Maintenance Schedule ...................................................................... 3
5.2 Maintenance Method ......................................................................... 4
VI. Frequently Asked Questions ................................................................... 4
VII. Cable Connection .................................................................................... 5
VIII. pH Calibration 3 Points Instructions .....................................................5
IX. Quality Assurance .................................................................................... 7
X. Warranty Limitation ................................................................................... 7
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

I. Overview

Online Digital pH Sensor

The digital pH sensor uses an industrial wire electrode with a flat


front end for easy cleaning. Built-in temperature sensor for
automatic temperature compensation, suitable for online
long-term monitoring environment. The sensor adopts RS485
output and supports Modbus, which can realize networking and
system integration without controller.

Features
 Digital sensor, RS-485 output, support MODBUS
 Unaffected by transmission distance for better stability

 Built-in temperature sensor, automatic temperature compensation

RS485 Output Signal


 It is equipped with an RS-485 interface that
allows easy and fast sensor configuration via
Modbus. It can be easily connected to monitoring
system.

 DSMC5100 can be equipped with it to read


real-time monitoring data, and the data storage
function provides users with a basis for analyzing the data

II. Technical Specifications

Item Parameters
Model DS510
Range 0~14pH
Precision 0.01ph
Resolution 0.01
Deepest depth Underwater 30m
Protection level IP68
Maximum operating pressure 1bar
Storage temperature 0~ 50 °C
Temperature range 0~ 50°C
Output signal Support RS-485, MODBUS protocol

1
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

Assembly Input type


Power supply DC 8~12V, current <50mA
Size Φ 26 * 167.5 mm
Cable length 5meters (default), can be customized
Housing material POM
Calibration Three-point calibration

III. Dimension

IV. Installation
4.1 Installation Tips
Before installation and operation, please check the appearance of the
cable for damage, and spread the cable.
1. Input sensors should be used to avoid water flow caused by the sensor hit
the wall or other facilities, if the water flow rate is too fast need to fix the sensor.
2. The measurement of the front should not be any shade, such as
fluorescence sensors should be avoided light.
3. The installation of the sensor should not exceed the depth of water from the
level of 2 metres, and according to the fluctuation of water flow, the installation
position should not be less than 30 cm deep. (According to the measured
water body can be adjusted)
4. The sensor should be installed in the position of no bubble generation, in
order to prevent air bubbles on the sensor's measurement data.
5.The sensor should not be used to lift the sensor's own cable, the installation
of cable protection devices to prevent excessive taut tension between the
cable and the sensor, resulting in cable breakage.
6. The sensor should not be installed measuring surface facing up, can be
installed vertically down or horizontal installation.
7. Such as the use of this product in the pilot tank, please consult the
company's sales staff to confirm the installation of specific requirements.

2
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

4.2 Customer Installation Cases

V. Maintenance of Sensor
5.1 Maintenance Schedule
It is recommended to periodically calibrate or clean the sensor surface
according to the actual application scenario. The calibration sequence for the
three-point calibration is that the first point calibration is at pH 4.0, the second

3
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

point calibration is at pH 6.86, and the third point calibration is at pH 9.18 in a


buffer solution. Refer to the MODBUS document for specific calibration
methods.
Maintenance tasks Recommended maintenance frequency
Calibrate the sensor (if required According to the maintenance schedule required by
by the competent authority) the
competent authority
5.2 Maintenance Method

Sensor maintenance
1) External surface of the sensor:

Wash the external surface of the sensor with tap water. If debris remains,
wipe it with a moist soft cloth.

2) Check the cable of the sensor:

The cable should not be tight during normal work, otherwise the internal
wires of the cable will be easily broken and the sensor will not work properly.
3) Sensor storage:
When the sensor is not in use, you should fasten the black plastic cap and
check whether the sponge inside is wet. If it is not wet enough, please add
3mol / L potassium chloride solution and let the electrode be stored in the
solution with potassium chloride.

Precautions

The probe contains sensitive optical and electronic components. Ensure


that the probe is not subject to severe mechanical impact. Damage to the
sensor due to impact or manpower will not be covered by the warranty. Water
ingress to the sensor due to replacement of the pH electrode is not covered by
the warranty.

VI. Frequently Asked Questions


Error Possible causes Solutions

4
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

The operation interface cannot be Controller and cable Reconnect the


connected or connection error controller and cables
the measurement results are not Cable failure Please contact us
displayed

Measured values are too high, too Measurement shifted Recalibration


low, or the values remain unstable
Water has entered the product
the product Please contact us
The pH electrolyte is Replace the pH
exhausted electrochemical part

VII. Cable Connection


Cable information
 pH probe size
 26x167.5 mm(Φ xL)
 Power information
 Power supply must be DC 6-12v + / - 5%, current < 50mA
Cable information
4 wire awg-24 or awg-26 shielding wire. OD= 5mm

 Red line----power supply (VCC)


 White line---- 485 data -B (485-B)
 Yellow line---- 485 data -A (485-A)
 Black line----ground (GND)
 Bare line---- shielding layer

VIII. pH Calibration 3 Points Instructions


1. Open the software Select COM and connect

5
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

2. Test available on the right, calibration available on the bottom left.

Recommendation: Before calibration, complete a test to ensure that there is


communication between the sensor and the software and that the sensor is
working properly.

3. Before entering calibration, you need to click on the Stop button to stop the
sensor from working on the test. Then click on the three point calibration and
follow the instructions to calibrate.

4. Just follow the prompts to complete the calibration at points 4, 6.86 and
9.18.

6
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

During calibration, please stay in each standard solution for as long as


possible to ensure a more perfect calibration.

IX. Quality Assurance


Our company assures its first-hand buyers that there will not be any
product defects caused by substandard materials or factory manufacturing in
the 3 months after shipment. If a defect is found within the 3-month warranty
period, Desun Uniwill Electronics promises to repair or replace the defective
product or return the payment in addition to the first shipping and related
formalities. Any product repaired or replaced during the warranty period will
only enjoy the remaining warranty period of the original product.
This warranty does not apply to consumables such as consumable parts
(including but not limited to lamps, pipes, etc.). Contact Desun Uniwill
Electronics or your agent to begin technical support during the warranty period.
After receiving the customer's questions about product quality, our
company will confirm whether the product needs to be repaired within two
weeks; products that have not been approved for repair cannot be returned.

X. Warranty Limitation
This warranty does not cover the following:
● Damage caused by force majeure, natural disasters, social
unrest, war (published or
unpublished), terrorism, civil war, or any government coercion
● Damage caused by improper use, negligence, accident or

7
Xi’an Desun Uniwill Electronic Technology Co., Ltd.

improper application and


installation
● Freight charges for shipping the goods back to Desun Uniwill Sensing
Technology
● Expedited or express shipping for parts or products covered by the warranty
● Travel expenses for local warranty repairs
This warranty covers all aspects of our company's warranty regarding its
products.
This warranty constitutes the final, complete and exclusive statement of
the terms of the warranty, and no one or agent is authorized to formulate other
warranties in the name of Desun UniwillElectronics. The remedies such as
repair, replacement, or refund of the payment as described above are special
cases that do not violate this quality guarantee. The remedies such as
replacement or refund of the payment are targeted at our products. Based on
strict liability or other legal theories, Desun Uniwill Electronics is not
responsible for any other damage caused by product defects or negligent
operation, including the subsequent damage caused by causality with these
situations.

You might also like