C l i p electronics

Data Sheet
Industrial Amplifier

Special features

− Amplifier and additional units


for strain gauge full bridges and
inductive half and full bridges
− Modules for mounting onto sup-
port rails to DIN EN 50022
− Accuracy class 0.1
− Adjustment via DIP switches
and potentiometers
− Clip IG industrial amplifier (alu-
minium-die-cast enclosure) IP65
− Intrinsically safe [EEx ia]IIC
measurement circuit with safety
barriers

Dimensions (in mm; 1 mm= 0.03937 inches)

Clip Electronics
77 42
AE101
AE301 Clip IG
AE501/AE511 Industrial amplifier
EM201
15
EM201K2
111.3

110.1
35

Hole pattern for screws Ğ6


GR201 330
112.9

DA101 160
TS101
140

50.9
10
only with DA 101
10.5 37
69
NT101A
NT102A
100 with cover
102
113.5

56

350

53

B0615-1.7 en
Specifications
Clip IG Industrial amplifier in aluminium-die-cast enclosure
Degree of protection IP65
Weight approx. kg 4.3
Weight (empty) kg 3.2
Ambient temperature °C [°F] −20...+50 [−4...+122]
Operating voltage
with 101A Power supply unit (Type WG 010) V 230 "10 %
with NT 102A Power supply unit (Type WG 011) V 115 "10 %
Mechanical stress
(test similar to DIN IEC 68)
Vibration (30 min each direction) m/s2 50 (5...65 Hz)
Impact (3 times each direction, impact duration 6ms) m/s2 350

AE101, AE 301, AE501 Measuring amplifiers for support rail mounting


Type AE101 AE301 AE501
Accuracy class 0.1 0.1 0.1
Transducers that may be connected
Strain gauge full bridge
VE = 10 V Ω 340...5000 − −
VE = 5 V Ω 170...5000 170...5000 −
VE = 2.5 V Ω 85...5000 85...5000 −
Inductive half/full bridge
VE = 2.5 V mH − − 2.5...20
VE = 1 V mH − − 6...19
Bridge excitation voltage VE (symmetrical to earth) V 10 5 2.5
V 5 2.5 1
V 2.5 − −
Permissible cable length
between transducer and amplifier m 500
Carrier frequency (crystal-stabilised) Hz − 600 4800
DC
Bridge zero balance
coarse approx. mV/V "2 " 21) " 80
fine approx. mV/V " 0.08 " 0.09 "3.2
Measuring ranges
VE =10 V mV/V 0.1...2 − −
VE = 5 V mV/V 0.2...4 0.2...42) −
VE = 2.5 V mV/V 0.4...8 0.4...83) 8...160
VE = 1 V mV/V − − 20...400
Calibration signal, in addition to the meas. signal mV/V + 0.24) " 1 % +8"1%
Input impedance MΩ >10 / 2 nF > 1 / 3 nF > 1 / 2nF
Common mode voltage, max. perm. Vpp " 10 V
Common mode rejection
0... 300 Hz dB > 100 > 100
> 300 Hz dB > 85 −
Linearity deviation % full scale < 0.05 typ 0.03
Output voltage V " 10 " 10
Rise rate, max. V/µs 0.4 −
Load resistance kΩ ≥4
Internal resistance Ω <2
1) AE301S6 and AE301S7: "1 coarse, "0.05 fine
2) AE301S6 and AE301S7: 0.1...2
3) AE301S6 and AE301S7: 0.2...4
4) AE301S6 and AE301S7: 0,1

HBM 2 B0615-1.7 en
Specifications
AE101, AE 301, AE501 Industrial amplifiers
Type AE101 AE301 AE501
Measuring frequency range
Bessel 3rd-order low-pass filter. Hz 0...10 − −
changeover (-1 dB) kHz 0...6 − −
Bessel 3rd-order low-pass filter (-1 dB) Hz − 0...104) 0...10
Phase transit time
with 0...10 Hz filter ms < 18 < 175) < 17
with 0...6 kHz filter µs < 20 − −
Rise time with 0...10 Hz filter ms 256)
Overshoot in the case of voltage surge
with 0...10 Hz filter % 0 <2
with 0...6 kHz filter % < 10 −
Noise voltage
measuring range 0.2 mV/V (10 Hz) mVrms <4 <4 −
measuring range 2 mV/V (10 Hz) mVrms <4 <4 −
measuring range 8 mV/V (10 Hz) mVrms − − <4
measuring range 80 mV/V (10 Hz) mVrms − − <4
measuring range 0.2 mV/V (6 kHz) mVrms < 30 − −
measuring range 2 mV/V (6 kHz) mVrms <6 − −
Long term drift over 48 hours (after 1 h warm-up time) µV/V < 0.2 < 0.1 < 0.8
Influence of a 10 K-change in ambient temperature
on sensitivity % full scale < 0.1 typ 0.05
on zero point
measuring range 0.2 mV/V mV < 60 < 10 −
measuring range 2 mV/V mV < 10 < 4 −
measuring range 8 mV/V (1 mV/V) mV − − < 10
measuring range 10 mV/V mV − − < 4
measuring range 80 mV/V (10 mV/V) mV − − −
Influence of a +15...30 V change in operating voltage
on sensitivity mV <1
on zero point (350 Ω bridge resistance) mV <1
5V-synchronisation (square wave) kHz − 76.8
Residual carrier voltage mV − <5
Operating voltage (DC) VDC +15...30
Power consumption mA ≤ 125 ≤ 100
Nominal temperature range °C [°F] −20...+60 [−4...+140]

Service temperature range °C [°F] −20...+60 [−4...+140]

Storage temperature range °C [°F] −25...+70 [−13...+158]


Degree of protection IP10
Weight g 200
4) AE301S6: 0...2 (−1 dB)
AE301S7: 0...60 (−1 dB)
5) AE301S6: <80 (filter frequency 2 Hz)
AE301S7: <2.8 (filter frequency 60 Hz)
6) Rise time with AE301S7 6 ms
Rise time with AE301S6 200 ms

TS101 Tare and store unit


Type TS101
Accuracy class 0.1
Input voltage V " 10
Input impedance kΩ 100
Output voltage V "10

B0615-1.7 en 3 HBM
Specifications
TS101 Tare and store unit
Permissible load resistance kΩ ≥5
Linearity deviation % < 0.04 of full scale
Influence of a 10 K-change of the ambient temperature % < 0.1 of full scale
Influence of a 15...30 V-change of the operating voltage % < 0.01 of full scale
Long-term drift over 48h (after 1 hour warm-up time) % < 0.02 of full scale
Noise voltage of the output mVpp < 20
Control inputs (floating)
High signal level V 11...30 (24 V nominal)
Low signal level V 0...5
Control output
High signal level V Vb −2
Low signal level V <1
Output current mA < 500
Tare unit
Output ms Net value (alternatively pos. peak val.)
Net-value amplification 1, 2, 5, 10-fold, selectable in steps, for
taring of major initial loads
Tare error (with v=1) mV <4
Settling time for the output voltage after taring ms 40 (to 99.9 %)
Low-pass filter (before taring) Hz 0.1...12.5; adjustable
Transmission bandwidth kHz > 10
Storage time for tare value Unlimited as long as Vb is present
(alternatively, storage in EEPROM)
Control input Taring with rising edge
Delay time for taring ms <1
Control output Taring valid
Peak-value store unit
Output Peak value (alternatively, pos./neg. peak, peak/peak 0.5 x
peak/peak or instantaneous value or envelope-curve value,
tared and amplified (1, 2, 5, 10-fold))
Peak-value store update-rate ms < 1.3
Accuracy % 0.25 (in 6 ms)
% 0.05 (in 20 ms)
Transmission bandwidth Hz 15 (−1 dB)
Settling time for the output voltage ms 40 (to 99.9 %)
Discharge rate for envelope curve mV/s 5...1000, adjustable
Control inputs Run/Hold;
(clear/inst.value)
Delay time for the control signals ms <8
Connection 12 series terminals for wire∅ 0.13...1.5 mm2; 10 mm end
sleeves for strands
Operating voltage Vb VDC 15...30, unstabilized
Power consumption mA < 90
Nominal temperature range °C [°F] −20 to + 60 [−4...+140]
Service temperature range °C [°F] −20 to + 60 [−4...+140]
Storage temperature range °C [°F] −25 to + 70 [−13...+158]
Weight g ca. 200
Degree of protection to EN 60 529 IP10
Mounting On support rails to EN 50 022

HBM 4 B0615-1.7 en
EM201 Output stage module (with one EM002 module)
EM201K2 Output stage module (with two EM002 modules)
Accuracy class 0.1
Input
Voltage V " 10 (0...+ 10 V)
Impedance kΩ > 11.5
Operating voltage VDC +15...30
Power consumption
(fully assembled with 2 x EM002) mA < 180
Nominal temperature range °C [°F] − 20...+ 60 [−4...+140]

Service temperature range °C [°F] − 20...+ 60 [−4...+140]

Storage temperature range °C [°F] − 25...+ 75 [−13...+158]

Weight g 200

EM002

Output signal selectable mA " 20 4...20

Output current
with VE =10 V mA 20 "0.02 20 "0.5
with VE = 0 V mA < " 0.04 4 " 0.2
Output current limit − − > 3 (switchable)
Permissible load resistance Ω < 500
Linearity deviation % < 0.05 full scale
Internal resistance kΩ > 100
Measuring frequency range kHz 3 (−1 dB)
Degree of protection IP10

GR201 Limit value switch


Accuracy class 0.1
Differential input
Voltage V " 10
Impedance kΩ > 50
Reference voltage
coarse approx. V " 10
fine approx. V " 0.5
Switching hysteresis mV 220
Factory setting: R43, R48 kΩ 3.01
to be changed by R43 and R48 kΩ 670 mV / VHyst.
Influence of a 10K-change in ambient temperature
on the switching point % < 0.05 full scale
Switching-point error % < 0.05 full scale
Relay capacity
max. voltage V 45 (separated extra low voltage)
max. current A 1
max. power W 30 (25 VA)

B0615-1.7 en 5 HBM
Specifications
GR201 Limit value switch
Switching times (Factory setting)
Response time ms <5
Decay time ms < 25
Operating voltage VDC +15...26
Power consumption mA < 100
Nominal temperature range °C [°F] − 20...+ 60 [−4...+140]
Service temperature range °C [°F] − 20...+ 60 [−4...+140]
Storage temperature range °C [°F] − 25...+ 70 [−13...+158]
Degree of protection IP10
Weight g 200

NT 101A, NT 102A*) Power supply


Type NT101A NT102A
Input voltage V 230 " 10 % 115 "10 %
Permissible frequency range Hz 47...63
Output voltage VDC 15.3
Output current In at >25°...+60° ADC 0.45
Output power W 9.75
Efficiency approx. % 60
Current limiter (protected against sustained short circuit) 1.2 x In (permanently adjusted)
Residual ripple mVpp < 10
Ambient temperature °C [°F] − 20...+ 60 [−4...+140]

Excess-temperature protection °C [°F] typ.105 [221] (trafo temperature)

Test voltage kVeff 3.75 (prim/sec and prim/housing)


Degree of protection IP10
Weight g 420
*) Version to DIN −VDE0551, EN60742 Protection class 1
The maximum permissible continuous current is 450 mA.

Clip accessories:
Covering angle 3-6450.0001
Clip IG accessories:
Bag with accessories 2-9278.0339 anti-buckling sockets, earth sleeves and end sleeves for strands for connection
of one cable. End sleeves for strands (0,5 mm2, 10 mm long).

 Hottinger Baldwin Messtechnik GmbH. All rights reserved. Hottinger Baldwin Messtechnik GmbH
All details describe our products in general form only.
They are not to be understood as express warranty and do
not constitute any liability whatsoever.
Im Tiefen See 45 ⋅ 64293 Darmstadt ⋅ Germany
Tel. +49 6151 803-0 ⋅ Fax: +49 6151 803-9100
Email: [email protected] ⋅ www.hbm.com
B0615-1.7 en