Respiratory 24P PIS.115 Stab Mic PC Strip - SY-MX24L
Respiratory 24P PIS.115 Stab Mic PC Strip - SY-MX24L
115
For in vitro diagnostic use only.
For professional use only.
W AR N IN G : The RT-qPCR program file should be downloaded from the QR code on the left or
from the link below.
https://www.bioeksen.com.tr/files/L_TD_43P/
All default analysis options (e.g., auto-calculated threshold) in the PCR instrument software should not be changed to calculate Cq values.
The shape of the amplification curves should be examined for all reaction wells returning with Cq values. All the sigmoidal curves above the threshold should be recorded as
“positive” and their Cq values should be recorded. Non-sigmoidal curves should be recorded as “negative”.
If any control does not work as described above, the run is reported as follows:
1. Contamination: If Cq≤30 in any NTC test channel.
Recommended action: Repeat the analysis paying attention to the “Warnings and Limitations” section.
2. Reagent Problem: In case a sigmoidal curve with a Cq≤30 cannot be obtained for any of all the samples tested in the run, including the controls.
Recommended action: If all the tested samples show negative results for the target pathogens and controls, the run is considered invalid. In this scenario, it is
essential to conduct testing on the "Positive Control(s)" included in the kit. A negative Positive Control test result suggests a potential "Reagent Problem." If
encountered, please reach out to the manufacturer for further assistance.
3. Invalid: If the IC (Internal Control) and all other targets are “Not detected”.
Recommended action: Sampling isn’t successfully done, or there is a problem during the sample transportation. A new sample from the same patient should be
collected and tested again.
If all the controls are valid, the results are interpreted as follows:
Table 8. Interpretation of Patient Results
Target Internal Control (IC) Result Interpretation
If 26<Cq ≤30 “Low Positive”
Results are valid
Positive (+) Positive (+) or Negative (-) If 16<Cq≤26 “Positive”
Target is detected
If Cq≤16 “High Positive”
Results are valid
Negative (-) Positive (+)
Target is not detected
The results generated by the PCR instruments can be reported manually, as explained earlier, or automatically using the "Sigmoida" software. To obtain the "Sigmoida" software
installer, please send an email to [email protected].
4. Explanation of Symbols
Symbol Title of Symbol Symbol Title of Symbol Symbol Title of Symbol
Control