Manual CO IFU Leonardo-E Fuser Guide
Manual CO IFU Leonardo-E Fuser Guide
LEONARDO®
INSTRUCTION MANUAL
Distribution by:
CeramOptec GmbH
Siemensstr. 44, 53121 Bonn
Germany
Tel: +49 (228) 97 967 – 0
Fax: +49 (228) 97 967 - 99
LEONARDO®
INSTRUCTION MANUAL
TABLE OF CONTENTS
Caution: Direct eye exposure to the laser light leads to irreversible eye damage!
The laser system is not suitable for ophthalmological applications.
To avoid damage to the eye and retina, doctors, surgical personnel, patients and any other
persons present in the room during treatment must wear appropriate protective eyewear.
Only use protective eyewear provided or approved by CeramOptec. Suitable protective eyewear
can also be obtained from CeramOptec (see also the recommendations in section 5.2).
Caution: Do not look directly at the laser beam or a laser beam that is in use with optical devices
or instruments. Doing so may result in permanent damage to the eyes or to the instruments.
Avoid placing reflective material, such as metal and glass, into the path of the beam.
Caution: Accidental irradiation to tissue not intended as the target tissue may result in laser burn.
Attention: The LEONARDO® laser may only be used with the accompanying footswitch and the
specified application and light delivery systems.
When operating the device, ensure that it is evenly balanced on a stable surface and that a
distance of at least 25 cm is maintained between the ventilation fan and the walls. Position the
device so that there are no cables or optical fibers suspended in the air between the wall socket,
the device and the patient.
To avoid risk of electric shock, do not open the housing. Service and maintenance may only be
carried out by CeramOptec or by qualified personnel authorized by CeramOptec.
The equipment must be routinely inspected and maintained in accordance with the instructions
provided in the maintenance section of this manual. A technical safety check must be performed
on a two year basis (see section 8.4).
Noncompliance with the safety and operating instructions provided in this manual will in all cases
invalidate warranty and liability on the part of CeramOptec.
Note: Ensure that the switch on the back of the unit is in the off position to prevent unauthorized
use of laser devices.
Note: Note also the manufacturer’s specifications regarding electromagnetic compatibility and the
relevant requirements (see section 9.4).
2 Product description
LEONARDO® is a laser system with functions and ergonomics specially developed for medical
applications. A touchscreen is used to set treatment parameters, such as laser power. User-
friendly menu navigation and microprocessor-supported control ensure reliable operation while
allowing physicians to concentrate on the essential aspects of treatment.
The fiber-coupled semiconductor laser diodes convert electrical energy to coherent laser
radiation with the wavelength of 980 / 1470 nm +/-30 nm (aiming beam 635 nm +/-10 nm and 532
nm +/-10 nm). Depending on the model, your LEONARDO® laser has a maximum laser output
power of 15 W (1470 nm) or 30 W (980 nm).
LEONARDO® is available as a device with two wavelengths (980 nm and 1470 nm) or as a
single-wavelength device with 1470 nm.
All LEONARDO® lasers can be operated in two basic modes, CONTINUOUS or PULSE MODE.
Additional special treatment modes for specific treatment procedures or in combination with
corresponding application fibers are available, depending on the configuration of the unit. Special
treatment modes are available in combination with the corresponding application fibers for
dermatology ("Derma mode") and phlebology ("ELVeS signal mode" and "ELVeS segment
mode").
For safety reasons, the LEONARDO® laser is equipped with a system for automatic recognition of
the optical fibers used. Application fibers from CeramOptec have coding for communicating with
the laser device.
For delivery details, refer to section 5.1.
• Endovenous ablation of surface veins (saphenous veins, magna and parva), tributaries,
perforant veins, recurrences, venous leg ulcers)
Dermatology
Endovenous
Pulmono-
ENT / Otorhinolaryngology Colorectal surgery ablation of
logy
coagulation & vaporization of endobronchial tumors surface veins
condylomata accuminata
and/or villous adenoma
Indication
polyps (polyposis coli
anal fistulas (FiLaC)
soft tissue tumors
coccygeal fistulas
perforant veins
turbinectomy
anal stenoses
hemangioma
paracentesis
tonsillotomy
reccurences
tributaries
adhesions
mariscae
stenoses
epistaxis
septum
fissures
LAUPP
DCR
Laser model
®
LEONARDO
X X X X X X X X X - X Xx X X X X X X X X X X X X X -
1470
®
LEONARDO
X X X X X X X X X X X X X X X X X X X X X X X X X X X
Dual 45
1. Info area
2. Laser status button
3. Status messages
4. Treatment mode options
5. Function options
6. Main area
Screen [1] The display is subdivided into four main key fields and shows:
- Status of the device (Enable/Standby)
- Treatment mode (Continuous mode (Continuous Wave CW),
Pulse mode and additional programs (optional))
- Treatment parameters (duration, energy, pulse parameter)
- Power settings
- Fiber information
- Aiming beam settings
Settings can be changed via the key fields and the cursor on the
screen
screen.
Key fields on the screen - Switch between enable and standby status
[1] - Change treatment mode (Continuous mode, Pulse mode and
additional programs (optional))
- Select the parameter to be configured
- Aiming beam (Off, On, CW mode or Pulse mode with intensity
setting)
- Save changed treatment modes
- Select function
Info screen
Instructions for use and additional information about using lasers
can be retrieved and read on screen.
Video area
Treatment videos provided by biolitec biomedical technology
GmbH can be viewed here (coming soon).
User settings
Device version number, language version, aiming beam settings.
Feature settings
These settings are only to be used by the service staff of
CeramOptec or other authorized personnel.
Pulse mode
Header
PILOT ON / OFF
The pilot beam can turned on or off and indicates if it’s turned on.
Furthermore the color of the pilot beam is displayed.
PILOT beam settings
Pilotbeam color, intensity and operatinn mode (CW/Pulse) can be
adjusted
SYSTEM STATUS
In standby mode, the power connection is plugged in and the main
switch is in the On position. The device does not emit any laser
radiation in the standby mode even if the footswitch is pressed.
Pressing this button switches the device from standby mode to
ready mode.
If the status is switched from standby to ready, the laser device
prepares for use. Duration of the preparing phase is about 3
seconds.
Indication that the laser is now ready. The laser device now emits
laser radiation as soon as the footswitch is pressed.
By pressing this button, the status switches from ready to standby.
Laser warning signal [4] When laser radiation is emitted, the LED warning light comes on
and a warning signal is sounded.
If the message "Not a valid biolitec® fiber" appears, this means that the device did not recognize
the coding on the fiber or identified the fiber as unsuitable for the laser unit based on the data
defined for the device. Only approved application fibers may be used with this device (see
section 6). Please contact our service department if further information is required.
Caution: All operating steps after preparation of the laser unit may only be performed when all
persons in the room are wearing appropriate laser safety goggles.
Use of the operating controls or configuration options of the LEONARDO® in a manner other than
that described here in the instructions manual may result in radiation hazards.
Ensure that the treatment room is clearly identified and that only persons wearing the appropriate
laser protective eyewear may enter the room during treatment. If this is not possible, you must
install a door switch as per section 4.1.2 that switches off the laser output when the door is
opened. Note that switching off the laser during treatment may result in unintended
complications.
The footswitch is included in the delivery and is to be attached to the footswitch connector on the
rear of the device.
Warning: Do not touch the pins of the footswitch connector. Turn the device OFF before
connecting or disconnecting the footswitch.
+5V
electrically signal lamp with separate
R electric circuit
5V separated
5mA
IC
S1
2 PIN
Interlock
S2
LASER for example:
S1: door
S2: window
S3 S3: …
Warning: Do not touch the pins of the door interlock connector. Turn the device OFF before
connecting or disconnecting the door interlock connector.
Recommendation: It is recomm mended that all staff receive an explanation and training in ESD
procedures. The ESD precautions procedure training shall include at least the safe connection
and disconnection of the footswitch and door interlock connectors.
The language used by the user interface needs to be entered the first time the device is switched
on. This setting is requested just once, and only if the device does not have a saved preset
language.
The access code is entered using the numeric keypad. By pressing OK the code is being
submitted and the device switches to standby mode. The code will be provided upon delivery of
the laser and the related introduction by a service assistant or by personnel
personne authorized by
CeramOptec.
Figure 10: An example of the monitor display when CW mode was the last active mode (for dual-
dual
wavelength devices)
The device starts with the aiming beam switched off. You have to press the PILOT button to
switch on the aiming beam. The screen will then display whether the aiming beam is switched on
and what color it is.
The color of the aiming beam can be selected using the PILOT box in the top bar.
You can configure a number of different settings for the aiming beam (see illustration 14). There
are options for a continuously active aiming beam and a flashing aiming beam, or it can be
deactivated. Moreover, it is possible to interrupt the aiming beam as soon as the therapeutic
beam is active. To change the intensity of the aiming beam, move the green bar or press the -/+
button. A numeric display also indicates the intensity selected. The operation panel can be used
to set the aiming beam to green or red.
The help windows are marked by the violet frame and are intended to deliver quick information to
the user that can be helpful while using the application.
Figure 17: Help for 2D Power Control™ (not available in single-wavelength devices)
In case the brief instruction guide is not sufficient, by clicking the question mark you can find the
respective section in the user manual that is saved in the device. It can be found via the info
screen as well. A hard copy user manual is included in the delivery and should therefore also be
available.
By clicking the -/+ button you can scroll to the previous/next pages.
By pressing << you can directly jump to the first page of the document, by >> to the last one.
It is possible to add, edit and remove users. These options appear in the pop-up
pop window that
appears after the ADMIN password has been confirmed.
Pressing the and buttons in the lower right of the screen closes user administration.
To create a new user, press the “Add” button. This opens the fields for defining the new user:
When entering the parameters “Setting number”, “Password” and “Repeat password”, no letters
may be used, so the letter keys on the touchscreen keyboard will appear in gray to show that
they have been deactivated.
To save a new user, press the “disk” button in the upper right of the screen. The button
in the upper right can be used to close the “Add” option without saving. After it is pressed,
pressed the
option menu opens.
Choosing “Edit” from the option menu opens fields for making changes to existing user accounts.
The user account you want to edit must be selected from the drop-down
drop down list in the upper left of
the screen. Then the parameters “Comment” and “Password” (together with password
confirmation) can be
e changed for the user you have selected.
When entering the parameters “Password” and “Repeat password”, no letters may be used, so
the letter keys on the touchscreen keyboard will appear in gray to show that they have been
deactivated.
To save changes to the user account, press the “disk” button in the upper right of the
screen. The button in the upper right can be used to close the “Edit” option without saving.
After it is pressed, the option menu opens.
To delete a user account, select “Remove” from the option menu. The user account you want to
delete must be selected from the drop-down
drop down list in the upper left of the screen.
To remove the selected user account, press the “disk” button in the upper right of the
screen. The button in the upper right can be used to close the “Remove” option without
saving. After it is pressed, the option menu opens.
Method 1
Method 2
Method 1: The setting of the laser output can be configured by tapping the white operation panel
for laser power.
wer. To increase the power level currently set, move the bar to the right.
ri To decrease
the power level,, move the bar to the left.
In devices that emit two wavelengths, laser output can be adjusted simultaneously for both
wavelengths with a constant mixing ratio between the laser power output of the two wavelengths
(figure 27).
Figure 27: Setting the laser output for devices that emit one or two wavelengths
Method 2: In devices that emit two wavelengths, laser power output can be set separately for
each wavelength by using the pencil field. If 2D Power Control™ is NOT enabled, a window with
4 scroll bars will pop up (figure 29). In this window, the total output, the combined factor or the
output for each wavelength can be set separately, either by moving the scroll bars or by pressing
the or buttons.
Figure 28: Laser power setting for dual wavelength without 2D Power Control™
If 2D Power Control™ is enabled, two different therapy lasers can be set by using this function.
The therapy laser at a wavelength of 1470 nm is controlled on the horizontal axis and the therapy
laser at a wavelength of 980 nm is controlled on the vertical axis.
Figure 30: Setting output with constant total laser power output
By pressing the FIX field next to the guide surface, a green guide line pops up that helps you to
adjust the mixing ratio between the two wavelengths while the total output is held constant.
By pressing the FIX field at the bottom left of the window, a green guideline pops up that helps
you to adjust the total output of the laser.
You can also fix the output of the laser for one wavelength on the respective axis while the output
for the other wavelength remains adjustable.
Figure 32: Fixing the power for the wavelength 1470 nm on the horizontal
2D Power ControlTM also makes it possible for you to save four different power settings.
settings
For power settings that have been saved, just press one of the four buttons in the lower part of
the screen to retrieve the setting (e.g.
(e.g ).
To save a power setting, first make the selection using the control panel.
panel Keeping one of the four
save buttons for longer than 1 second stores the setting.
setting
This opens another pop-up window for naming the button.
button The button needs to be pushed briefly
so that a blank (_____) appears. Then you can use the screen keypad to enter and confirm any
name you wish to give the setting.
setting
Note: The previously selected power setting
set will appear below the button (the
the power for the 1470
nm wavelength on the left and the power for the 980 nm wavelength on the right).
right
Tapping the button in the lower right of the screen saves the settings.
settings Use the button to
close the menu without saving any changes.
changes
In CW mode, the laser continuously emits radiation at the selected power level as long as the
footswitch is pressed.
In Pulse mode, the laser emits radiation at the selected power level with the specified number of
pulses and pulse format (pulse duration/pulse pause) as long as the footswitch is pressed.
• Defined number of pulses → between 1 and max. 99 pulses (depending on configuration)
• A continuous series of pulses as long as the footswitch is pressed
This pulse procedure is repeated as long as the footswitch is pressed or until the defined number
of pulses is reached. If the footswitch interrupts a pulse or a pulse pause, the entire pulse
procedure is repeated when the footswitch is pressed again.
The ELVeS signal mode is derived from the continuous mode. The device indicates the pulling
speed of the fiber from the vein via auditory signals. This speed is calculated from a combination
of output and energy per signal that have been selected by the user.
The ELVeS segment mode is derived from the continuous mode and has an additional visual
support for the user during vein treatment.
The Derma mode for dermatology is derived from the pulse mode. Here, the laser sets the
required laser output for the selected power density in relation to the selected handpiece.
Figure 39:: Screen display of Continuous mode CW (for devices that emit one wavelength)
Laser power can be set in continuous mode as shown in figures 26 - 27. As soon as you have
changed all settings (operating mode, power, aiming beam), you can enable the laser unit for
laser emission by tapping the "Standby" button in the upper right corner of the screen and putting
it into "Ready" mode.
When the "Standby" button is activated, the laser is put into the preparation phase for about 3
seconds. After that, the laser is ready for emission.
The laser is now active and the laser hazard area must be secured. The yellow LED next n to the
display flashes and the message "Press Footswitch to Fire Laser" appears in the screen footer.
During laser emission, the message "LASER FIRING" appears on the screen. The treatment time
and the emitted treatment energy are simultaneously added up and displayed. The treatment
time and treatment energy can be reset when the laser is enabled or when it is in standby mode.
You can use the "Ready" button in the upper right of the screen to switch to secure standby at
any time. You should always switch to standby after a treatment is completed or when
temporarily putting down the application fiber.
You can end laser emission at any time by
• releasing the footswitch
• pressing the "STOP LASER" button
• pressing the "Standby" button
Laser emission can also be interrupted (usually unintentionally) by safety circuits. During normal
use, these circuits should not be triggered intentionally to switch off laser emission:
• by opening the door in the case of a connected door interlock
• by disconnecting the application fibers
In these cases, laser emission stops immediately, the audible signal stops and the laser emission
symbol is turned off. The laser unit switches to secure standby. Messages indicating the reason
for shutdown appear on the screen.
The settings for Pulse mode can be set on the display (as described on the following page). The
intervals are displayed graphically on the left.
left
For fine adjustments of the pulse duration and the pulse pause, there is a zoom function for
precisely adjusting single segments.
In pulse mode, you can set the following parameters by touching the appropriate bar on the
screen:
Pulse repetition Pulse duration Pulse pause
(Number of pulses) (Pulse length (Interval between pulses
max value configurable by max value configurable by
service assistant) service assistant)
1 – 99 pulses (depending on 0.01 – 60.0 s 0.01 – 60.0 s
configuration)
Continuous series of pulses as long as 0.01 – 60.0 s 0.01 – 60.0 s
the footswitch is pressed
To change the currently selected settings, touch the relevant field and move the green bar or
to the right or press the button. If the values are to be decreased, move the bar to the left or
Figure 49:: Start screen of ELVeS® segment mode for setting parameters (for devices that emit
one wavelength)
In the upper screen section of the ELVeS® segment mode, the laser output can be defined. In the
bottom section, pressing the white field opens a new window where the following parameters
must be configured:
To change the parameters, tap the or buttons or move the green bar. LEONARDO®
then automatically calculates the length of the vein segment in proportion to the selected quantity
of energy and displays it as a progress bar.
Once you have checked all parameters again, you can enable the laser unit for laser emission by
tapping the "Ready" button in the upper right of the screen. When the "Ready" button is activated,
the laser switches to the preparation stage. This is indicated by PREPARATION lighting up in the
top right next to SYSTEM STATUS. The laser emits a continuous beep.
The laser is now active and the laser hazard area must be secured. The yellow LED next to the
display flashes and the message "Press Footswitch to Fire Laser" appears in the screen header.
When the unit is enabled, pressing the footswitch causes the application fiber to emit
laser light. The laser emits a continuous beep and the LED next to the screen remains lit
during the entire laser emission.
When the unit is in "Ready" mode in the ELVeS® segment mode, laser light is emitted
using the connected application fiber as long as the footswitch is pressed. You can
discontinue laser emission by simply releasing the footswitch. As long as the laser unit is
enabled, the laser can be activated repeatedly.
During laser emission, the message "LASER FIRING" appears on the screen. The treatment time
and the emitted treatment energy are simultaneously added up and displayed on the progress
bar.
Example: You want to emit treatment energy of 80 J per 1 cm of vein when treating a vein that is
50 cm long. You configure the vein length and the quantity of energy on the ELVeS® segment
setting screen. Based on the selected energy per unit of length and the actually emitted treatment
energy, the LEONARDO® automatically calculates the relevant vein segment length and indicates
it on the progress bar.
Caution: Note that the actual energy per vein length emitted in this mode is still determined by
the user. The laser unit only displays the target values as a guide to optimal and homogeneous
The laser output can be configured in the top section of the screen by tapping the white operation
panel.
Tapping the white field next to the energy interval opens another window where you can define
the energy interval by moving the green bar or pressing the or button. When the OK
button is pressed, the settings are accepted.
accepted
Example: You want to emit 140 J for each 1 cm of vein to be treated. You configure an energy
interval of 140 J and also define the required laser power. During treatment, a beep sounds each
time 140 J of treatment energy has been emitted. For example, you will hear a beep after 140 J,
280 J, 420 J, 560 J etc. If you have set power at, say, 10 W, the beep sounds every 14 seconds
and you would have to have moved the laser 1 cm along the vein by the time each beep occurs
in order to achieve the target energy density of 140 J/cm.
The total treatment energy emitted is calculated independently and displayed on the screen.
Once you have checked all parameters again, you can enable the laser unit for laser emission by
tapping the "Ready" button in the upper right of the screen.
When the "Ready" button is activated, the laser switches to the preparation stage. This is
indicated by PREPARATION lighting up in the top right next to SYSTEM STATUS. The laser
emits a continuous beep.
The laser is now active and the laser hazard area must be secured. The yellow LED next to the
display flashes and the message "Press Footswitch to Fire Laser" appears in the screen footer.
When the unit is enabled, pressing the footswitch causes the application fiber to emit
laser light. The laser emits a continuous beep and the LED next to the screen remains lit
during the entire laser emission.
When the unit is in "Ready" mode in the ELVeS® signal mode, laser light is emitted using
the connected application fiber as long as the footswitch is pressed. You can discontinue
laser emission by simply releasing the footswitch. As long as the laser unit is enabled, the
laser can be activated repeatedly.
During laser emission, the message "LASER FIRING" appears on the screen. The treatment time
and the emitted treatment energy are simultaneously added up and displayed on the progress
bar.
Warning: Please note that the actual energy per vein length emitted in this mode is still
determined by the user. The laser unit only displays the target values as a guide to optimal and
homogeneous energy emission.
The user is responsible for comparing these values with the actual emitted energy and distance
covered by the fiber within the vein!
Caution: Note that laser produced skin vaporisation / smoke could contain viable particles.
Caution: Note that the safe distance, i.e. the NOHD (Nominal Ocular Hazard Distance), depends
on the handpiece selected. This distance is specified in the table below:
Handpiece for a beam LEONARDO® Dual 45
diameter of 980nm/30W
0.6mm 4.1m
1.0mm 6.8m
1.5mm 11m
Select the energy density and the applied handpiece by tapping the respective pencil fields
(gearwheel). All values can be selected by moving the green bar or pressing the or
buttons.
If you then change the pulse duration or pulse pause, the laser automatically sets all additional
parameters – in relation to the setted enery density and choosed handpiece (within the available
range of values).
Note: Check that you have received all standard accessories listed here. Check the device and
accessories for any obvious signs of damage, and do not use them should any be found.
Lens
Alternative lens
Model number EN207:2002 EN207: 2010
number (Honeywell)
(Honeywell)
LEONARDO® DUAL
45 950-1010 DL4 950-1010 DLB 6
#xx-21096 -
980nm /30W 1430-1510 DL2 1430-510 DLB 4
1470nm /15W
LEONARDO® 1470
#xx-2219 #xx-21096 1430-1510 DL2 1430-510 DLB 4
1470nm/15W
Attention: Only application systems, fibers and medical probes approved by CeramOptec may
be used. Use of systems that have not been approved may damage the device and will invalidate
the warranty.
Attention: Follow the instructions enclosed with the application fiber to ensure safe handling.
Noncompliance with the manufacturer’s handling recommendations may result in damage to the
application fiber or medical probe and/or injury to the patient or user.
The application fiber or probe must only be cleaned and/or disinfected in accordance with the
manufacturer’s recommendations. Inappropriate use of cleaning or disinfectant products may
result in (sometimes unnoticeable) damage to the fiber, which could cause injury to the patient or
doctor.
Attention: Particular caution is required when handling application systems or medical probes. If
these are knocked against a hard surface or bent excessively, they may be damaged and their
functioning impaired. Fiber tips may be damaged by incorrect handling. Fibers suspected of being
damaged may not be used.
Attention: Safety elements S1 to S9 are part of the laser system. These must always be used to
improve safety when using LEONARDO®. Therefore, check as thoroughly as possible that these
are functioning correctly.
Attention: LEONARDO® contains no parts that are serviceable by the user. Any attempt to
repair, adjust, or modify the system (outside of the procedures outlined in this manual) by
any person not authorized by CeramOptec is strictly FORBIDDEN and will invalidate the
warranty.
The final decision will rest with CeramOptec in all cases.
To avoid risk of electric shock, do not open the housing. During the warranty period, servicing
and maintenance may only be carried out by CeramOptec or by qualified personnel authorized by
CeramOptec.
8.2 Cleaning
Attention: Unplug the device before cleaning.
Clean the housing with a damp cloth and a mild alcohol-based antiseptic detergent or mild
cleaning agent. Chemical cleaning agents, abrasive cleaning agents, and rough cleaning cloths
may damage the surface of the housing and therefore should not be used. Do not spray with
water as it may penetrate the equipment.
Fuses that have blown must only be replaced with fuses of the same type (see the model-specific
specifications in section 9.1.
Should a replacement fuse also blow, stop using the device and contact our service department.
EN 60825-1:2008 EN 60601-2--22:2007
MADE IN GERMANY
CeramOptec GmbH
Siemensstr. 44 * 53121 Bonn
Operation
www.ceramoptec.com/support
Storage
2013
®
Type LEONARDO
Model DUAL 45 SN xxxxx
Input 100-240
240 VAC, 50-60
50 Hz, 450VA Fuse 2 x T4AL / 250V
Figure 59:: Label for devices that emit laser power of two wavelengths
EN 60825-1:2008 EN 60601-2
2-22:2007
MADE IN GERMANY
Operation:
CeramOptec GmbH
Siemensstr. 44 * 53121 Bonn
www.ceramoptec.com/support
Storage:
2013
®
Type LEONARDO
Model 1470 SN xxxxx
Input 100-240
240 VAC, 50-60
50 Hz, 450VA Fuse 2 x T4AL / 250V
Figure 60:: Label for devices that emit laser power of 15W at wavelength of 1470 nm
Warning: To avoid risk of electrical shock, this device must be connected to a power supply with
protective conductor only.
<5% UT <5% UT
(>95% break (0 V)
of UT for 5
seconds
EN 61000-4-8: 2010 3 A/m 3 A/m Magnetic fields at the supply
Magnetic fields at the 10 A/m 10 A/m frequency should conform to the
supply frequency typical values as they occur in the
business or clinic environment.
NOTE 1: For 80 MHz and 800 MHz, the higher frequency range is valid.
NOTE 2: These guidelines may not be applicable for all cases. The propagation
of electromagnetic values is influenced by absorptions and reflections of
buildings, objects and people.
a. The field strength of stationary transmitters such as fixed parts of cellular phones and
mobile radio sets, amateur radio stations, AM and FM radio and television cannot be
determined exactly in theory. A study of the location may be considered to detect the
electromagnetic environment in regard to stationary transmitters. If the measured field
strength at the location where the device is being used exceeds the accordance level
above, the device should be watched to verify proper functioning. If abnormal
performance is observed, additional actions might be necessary such as a modified
orientation or another location of the device.
b. For the frequency range of 150 kHz to 80 MHz, the field strength should be lower than 3
V/m.
For transmitters with a maximum nominal power not mentioned above: To detect the
recommended safety distance, use the equation in the corresponding column. P is the maximum
nominal power of the transmitter in watt (W) according to the specifications of the transmitter
manufacturer.
NOTE 2: These guidelines may not be applicable for all cases. The propagation of
electromagnetic values is influenced by absorptions and reflections of buildings, objects and
people.
10 Service policy
The manufacturer will provide, on request, schematic connection diagrams,
diagrams, lists of replacement
parts, descriptions, configuration instructions and other information to be used by qualified
line maintenance and repair of components of the LEONARDO®
technical personnel for first-line
system that the manufacturer has deemed repairable.
repairabl
When parts are deemed to be irreparable by the manufacturer or when special training and/or
equipment are required to perform the repair or adjustments, the manufacturer reserves the right
to withhold information on the grounds of safety.
If you wish to return the LEONARDO® laser to the manufacturer for servicing, it must be returned
in its original packaging. You must also provide a statement guaranteeing that the device has
been disinfected accordingly.
11 Environmental protection
The LEONARDO® laser must be disposed of in compliance with the applicable regulations.
regulations
The bin symbol on the product or in the manual indicates that this product must not
be disposed of with other waste. CeramOptec has built up a waste management
electroni waste. This also includes free return of your LEONARDO®
system for electronic
laser.
Arrangements for the return of all unusable equipment to CeramOptec for correct
removal and environmentally friendly disposal should be discussed with the local
distributor or dealer of the country in which the equipment was purchased.
Attention: The warranty is invalidated in the cases listed below. The final decision will rest with
CeramOptec in all cases:
• Improper handling or misuse of the laser
• Use of components that do not belong to the laser
• Use of application systems that have not been approved
• Any of the following actions performed by persons not authorized by CeramOptec:
o Installation
o Servicing
o Configuration or adjustment of the laser
Notwithstanding any alternative terms of warranty that may have been specified in the contract of
sale, the conditions of warranty specified here exclude and replace all other verbal warranties or
warranties in writing that were explicitly agreed or tacitly implied.
The conditions of warranty specified here apply exclusively to defective devices. In no case will
CeramOptec accept liability for accidental or indirect losses (including loss of profit), bodily injury,
damages or costs directly or indirectly resulting from use of the LEONARDO® laser system.
Model: nm / W/ µm Operator:
S/N: Location:
Inventory
no.:
Name / Signature:
Date:
Checked:
Name of trained
personnel / Signature:
Person Device
Log entry
responsible / Comments status (Pass Signature
no.:
Date / Fail)
1.5 Connections
2.1 Footswitch
Beam guidance system: Coupling/decoupling/aiming
2.2
beam
2.3 Interlocks
to the output port of the laser. Place the distal end of the application system in the power meter
adapter of your power measurement device. Enable and fire the laser, and record the value.
2 1.6 - 2 - 2.4
5 4- 5 -6
10 8- 10 - 12
15 12 - 15 - 18
5 4- 5 -6
10 8- 10 - 12
15 12 - 15 - 18
20 16 - 20 - 24
30 24 - 30 - 36
5 12 - 15 - 18
10 24 - 30 - 36
15 36 - 45 - 54
®
20 48 - 60 - 72 Only LEONARDO
DUAL 45
®
30 72 - 90 - 108 Only LEONARDO
DUAL 45
®
45 108 - 135 - 162 Only LEONARDO
DUAL 45