CALIBRATION CERTIFICATE

CASE # CAL-0104
CERTIFICATE# PME/CAL/0001
LAB CODE : Q-01
WENOVO PHARMACETICALS
NAME OF
CAMPANY:
SMALL INDUSTRIAL ESTATE SOP # PME-SOP-01
TAXILA.
REF-CERTIFICATE # CL-1160(M-202)2022
LOCATION: Physical Lab. ISO CERTIFICATION:9001-2015

ENVIORMENTAL CONDITIONS
HUMIDITY 55%+ 5% AIR TEMP 25.87C LUX 400 LUX
CONDITIONS
AIR PRESSURE 1013.4HPA WATER TEMP 25.87C VOLTAGE/FREQUENCY 223V- 60Hz

EQUIPMENT DETAILS
EQUIPMENT NAME Analytical Balance SERIAL NO. NIL CAPACITY
EQUIPMENT ID TE214S
WNP/QC/01 MODEL NO. MADE Germany
Sartorius
TRACEABILITY OF MEASUREMENT
THE LABORATORY REFERENCE STANDARDS AND EQUIPMENT USED FOR CALIBRATION ARE TRACEABLE TO
PCSIR LABORTORY COMPLEX PESHAWAR ISO 17025, WHICH IS FURTHER TRACEABLE TO NPSL ISLAMABAD
AND INTERNATIONAL STANDARDS OF MEASUREMENTS IN ACCORDANCE WITH THE METROLOGY RULES.
CALIBRATION STANDARD
Pre-calibrated Dead weight Set CA# CL-182(M-31)2024

CALIBRATION DATA

MEAN
Nominal
S# OBSERVED DEVIATION LIMITS STATUS
Weights (G)
WEIGHTS(G)
1 10.00 10.00 0.00 ±0.1% Satisfactory
2 20.00 20.00 0.00 ±0.1% Satisfactory
3 50.00 50.00 0.00 ±0.1% Satisfactory
4 100.00 100.00 0.00 ±0.1% Satisfactory
5 200.00 200.00 0.00 ±0.1% Satisfactory
6 500.00 500.00 0.00 ±0.1% Satisfactory
REMARKS: On the basis of above tests, equipment has been calibrated & found satisfactory.
In case of any electronic fault/change placement of equipment, calibration will be void & had to call service
person of Re-calibration.

CALIBRATED BY REVIEWED BY APPROVED BY

Calibration Engineer Service Engineer Quality Control Manager

CALIBRATION DATE DUE DATE


JANUARY 10, 2024 JANUARY 10, 2025