Ultrasound Manual
Ultrasound Manual
Thank you for selecting Full Digital Colour Doppler Ultrasonic Diagnostic System.
Prior to use the system, please read this user manual carefully.
In addition, keep this user manual so that you may read it at anytime when necessary.
The manufacturer is responsible for the effects on safety, reliability, and performance of
the product, only if:
Before using the system, read carefully and be sure to completely understand these
conventions in this manual.
I
Symbols and
Description
conventions
DANGER indicates a hazardous situation which, if not avoided, will result in death
or serious injury.
WARNING indicates a potentially hazardous situation, which, if not avoided, may
result in physical injury or death.
CAUTION Indicates a potentially hazardous situation, which, if not avoided, could
result in minor or moderate injury or equipment damage.
Physical buttons or knobs on the device, prompts or menu items on the touch
[Bold]
screen
Warranty
II
The voltage of the power grid is beyond the scope specified by the product.
Irresistible natural disasters.
Replace or use of parts, accessories, and consumables not approved by the
manufacturer or maintain the system by unauthorized personnel.
Failures not arise from the product itself.
Failure to follow any operation instruction of the user manual.
Disclaimer
The manufacturer dose not assume any responsibility for unintended use, mis-operation
by unqualified medical personnel or failure to follow operation instruction of the user
manual.
The manufacturer does not assume any responsibility for any medical disputes arising
from the rental of equipment to third parties or other institutions.
III
Contents
1 Safety ..................................................................................................................... 1
V
3.4.1 Power on ............................................................................................... 18
3.4.2 Shutdown .............................................................................................. 19
3.4.3 Standby ................................................................................................. 19
3.5 Monitor Adjustment ......................................................................................... 20
3.6 Connect / Remove the Probe ......................................................................... 20
3.6.1 Connect the Probe ................................................................................ 21
3.6.2 Remove the Probe ................................................................................ 21
3.7 Foot Switch (Option) ....................................................................................... 21
3.8 Connect the Network ...................................................................................... 22
3.9 Connect Peripheral Device ............................................................................. 22
3.9.1 Connect USB Storage Device .............................................................. 22
3.9.2 Connect Printer ..................................................................................... 23
3.10 Screens ........................................................................................................... 23
4 System Settings...................................................................................................... 25
VI
4.7 Maintenance Preset........................................................................................ 45
4.7.1 System Information ............................................................................... 45
4.7.2 Detailed Information.............................................................................. 45
4.7.3 Option ................................................................................................... 45
4.7.4 Other ..................................................................................................... 45
VII
6.5.3 Image Optimization (PW/CW) .............................................................. 71
6.6 Color M-mode ................................................................................................. 74
8 Measurement........................................................................................................... 84
VIII
8.3.9 Emergency .......................................................................................... 136
IX
11.2.1 DICOM Storage .................................................................................. 156
11.2.2 DICOM Print........................................................................................ 157
11.2.3 DICOM Worklist .................................................................................. 157
11.2.4 DICOM MPPS ..................................................................................... 158
11.2.5 DICOM Storage Commitment............................................................. 159
11.2.6 DICOM Query / Retrieve..................................................................... 159
11.3 Structured Report ......................................................................................... 160
X
1 Safety
1 Safety
1.1.1 Precautions
1
1 Safety
2
1 Safety
The system and accessories are not disinfected and sterilized when
delivering. Before use probe and biopsy bracket, disinfection or
sterilization is required according to the method provided in their
instructions.
Always inspect whether the edge of the probe is too sharp or the
surface is too rough before using the probe. If so, replace the probe
to avoid damaging sensitive tissue.
DO NOT disconnect probe when scanning. Otherwise, system
damage may result.
Normally, there is no risk of burnt when using this system; however,
if the probe is placed in the same part of the patient’s body for a long
time, burnt may result. Under the condition of obtaining effective
images, shorten the inspection time as possible.
Shut down the system before plugging or unplugging the system or
other peripheral equipment (such as printers). Otherwise system
damage may result.
Always use approved, sterile coupling gel. Read the instructions of
the coupling gel carefully before operation.
Remove the coupling gel from the probe thoroughly after each use.
Otherwise, moisture in the coupling gel may affect the safety and
performance of the probe.
3
1 Safety
For patient and personnel safety. Never contact with samples, mixtures
and waste liquids by hand directly. Use protective barriers (gloves and
probe sheaths) and follow sterile procedures when appropriate.
If the sample contacts with skin inadvertently, you must immediately stop
operating and notify your physician.
Some disinfectants have strong acidity or alkalinity. Please use them
carefully to prevent direct contact with hands and clothes. If hands or
clothes are inadvertently exposed to disinfectants, wash them with soap
and flowing water immediately; if disinfectants enter eyes inadvertently,
wash them with plenty of clean water immediately and consult an
ophthalmologist.
Disinfectants, intensive cleaning agent and waste liquids must be
disposed as bio-hazardous material in accordance with local standards.
Consult the relevant disinfectant manufacturer or distributor for details.
Probes used on patients with Creutzfeldt-Jakob disease (CJD) must be
disposed. Otherwise, it may infect the patients and operators. There is no
method to clean, disinfect and sterilize the probes used by
Croytsfelter-Jacob patients currently.
Follow all internal infection control procedures applicable to personnel
and systems.
Disposable biopsy bracket must be disposed of as infectious waste.
Disinfect the reusable biopsy bracket before store it.
4
1 Safety
Labels/symbols Description
Tag includes the following information:
The name and address of the manufacturer
Tag Date of manufacture
Model and Serial Number
Electrical rating
Dangerous voltages
Equipotential terminal
Alternating current
USB port
Ethernet connection
Manufacturer
5
1 Safety
Labels/symbols Description
Date of manufacture
Serial number
Keep dry
Keep facing up
No rolling
6
1 Safety
Labels/symbols Description
1.2.1 ALARA
The guiding principle for the use of diagnostic ultrasound is defined by “ALARA principle”
(As Low As Reasonably Achievable). That is, by keeping ultrasound intensity as low as
possible while obtaining diagnostic images.
The ultrasound intensity depends on the intensity and exposure time of irradiation.
The ultrasound intensity required differs among application modes and patients.
Less ultrasonic power can only obtain low-quality images, which affects the reliability of
the diagnosis; and using overdose ultrasonic power increases the risk of bioeffects.
Therefore, it is operator’s responsibility to reconcile ultrasonic power with diagnostic
image quality. That is, to obtain qualified images while minimizing the ultrasonic bioeffects.
7
1 Safety
1.2.2 Bioeffect
Ultrasound produces two types of bioeffects, thermal and mechanical.
The TI informs the operator about the potential for temperature rise in body tissue. It is an
estimate of temperature rise in body tissue with specific properties. And temperature rise
is influenced by factors such as tissue type, vascularity, operation mode, and others.
There are three types of TI. The system only displays one index at a time. The operator
should choose the appropriate index to display according to the actual application
location:
The soft tissue thermal index (TIS) is suitable for abdominal and cardiac
application. It indicates the potential for temperature rise in homogeneous tissue.
The bone thermal index (TIB) is suitable for obstetric (mid-late pregnancy) or
pediatric (newborn head) applications. It indicates the potential for temperature
rise of bones or at near soft tissue.
The cranial thermal index (TIC) indicates the potential for temperature rise of
bones or at near soft tissue.
The MI informs the operator about the possibility for mechanical bioeffect occurrence. It is
an estimate of damage in body tissue due to cavitation.
The higher the MI value the greater the possibility of mechanical bioeffect result.
NOTE : When gas and soft tissue exist at the same time, special attention
should be paid to setting the MI value to a lower level.
8
1 Safety
patients.
The direct control is to adjust the acoustic power by using [ ][ ] to select [A.Power],
and then rotating the [Value] knob on the control panel.
Refer to “Annex D Acoustic Output Report” for typical values for the acoustic output of
specific application.
The maximum value of acoustic output in any mode should not exceed the limit value of
acoustic output (MI limit value is 1.9, Ispta.3 limit value is 720 mW/cm2).
Indirect control is mainly performed by adjusting image parameters to control the acoustic
power. Parameters that affect acoustic power include: imaging mode, probe and its
frequency, position and number of focus, image depth, and PRF.
The receiver control only affects the receiving mode of the ultrasonic wave, and does not
affect the acoustic output while improving the image quality. The operator can optimize the
image quality by adjusting the gain, TGC, and dynamic range to obtain effective
diagnostic image with the minimum acoustic output.
Receiver control is recommended before direct control and indirect control.
9
2 System Overview
2 System Overview
Obstetric
Cardiac
Urology
Musculoskeletal
Peripheral vascular
2.5 Contraindication
Do not suitable for eye examination
10
2 System Overview
The following table shows the standard packing list. The contents of the packing list
should be based on the actual purchased model.
Table 2 Standard Packing List
No. Description
1 Console * 1
2 Power Adapter * 1
3 User Manual * 1
4 Accessory Package * 1
2.7.1 Console
11
2 System Overview
Battery
12
2 System Overview
Power switch
1 Power on/ off the system.
button
13
2 System Overview
mode.
Move up/
12 Switch menu / Select parameter.
Increase
Move left /
13 Switch menu / Select parameter.
Decrease
Move right /
14 Switch menu / Select parameter.
Increase
Move down /
15 Switch menu / Select parameter.
Decrease
Press this button and then rotate the [Value] knob to
16 BaseLine BaseLine
adjust the baseline position.
Press this button and then rotate the [Value] knob to
17 Depth Depth
adjust the display and scan depth.
Press this button and then rotate the [Value] knob to
18 Angle Angle
adjust the SV angle.
Press this button and then rotate the [Value] knob to
19 Focus Focus
adjust the focus position.
Sample Sample Press this button and then rotate the [Value] knob to
20
Volume volume adjust the SV.
Press this button and then rotate the [Value] knob to
21 Scale Scale
adjust the PRF of CFM/PDI/PW.
Body
26 Body Mark Press to activate body mark function.
Mark
27 Update Update Compound operation keys, used together with other keys;
14
2 System Overview
Save
33 Save Cine Press to save cine to the system quickly.
Cine
Probe/Exam Press to enter the probe/ exam selecting screen and
34 Probe
switch toggles the probe and exam mode.
Image
35 File Press to enter the review screen.
management
36 Save Img Save image Press to save image to the system quickly.
2.7.5 Trackball
Use the trackball and [Set] button to control the screen directly and easily.
Function Description
Use the trackball to move the cursor over the select item, image, or annotation
Click/ Select
you want to select and press the [Set] button.
1) Click on the image or annotation you want to move.
Drag 2) Roll the trackball to drag the image or annotation to the desired location.
3) Press the [Set] button and release trackball.
1) Click on the ROI frame you want to adjust.
Adjust ROI 2) Roll the trackball to adjust the size of ROI frame.
3) Press the [Set] button and release trackball.
15
2 System Overview
1 Video output.
2 VGA port.
3 Network port.
4 Equipotential terminal.
5 DC input port.
6 Power switch.
16
3 Preparing the System
17
3 Preparing the System
If you do not use the system for a long time on battery power, please
shut down the system. Frequent charging and discharging will
impact the service life of batteries.
Replace and properly recycle the battery when it is obviously damaged or exhausted.
Follow the corresponding regulations when disposing the used batteries.
3.4.1 Power on
If an abnormal power interruption occurs during the last system
operation (such as a sudden interruption of AC power), you may
have to wait longer 30~60s than usual to reboot the system.
System can be rebooted 5 minute later of power off. Otherwise
system failure may result.
2. Press the power switch button [ ] located on the upper left section of the
control panel to startup the system and the working status indicator. lights up.
3. A boot screen appears, and the Home screen displays after startup.
18
3 Preparing the System
4. After initialization has been completed, refer to the following items for startup
check. If there is any abnormality, please contact your sales representative or the
customer service department of the manufacturer.
Table 8 Checking List
No. Item
2 Error massage.
When the probe is connected to the system, monitor that the surface temperture of the
4
probe all the time to avoid abnormal heat burn the patient.
5 The buttons and knobs of the control panel are fully functional.
3.4.2 Shutdown
To shutdown the system, perform the following steps:
1. Press the power switch button [ ] located on the upper left section of the
control panel. Click [Shutdown] to perform the shutdown procedure.
2. When the power indicator lights out, the shutdown is completed.
3. Turn off the AC power and unplug the cable cord as needed.
Never attempt to shutdown the system while the system is in an
upgrade or data transfer process.
End the current exam before shutdown the system.
DO NOT unplug the power plug or cut the power supply before the
shutdown procedure is completed. Files damaged and patient data
lost may result.
NOTE : If the system fails to power off normally, cut the power supply or
unplug the power cord to force the system to shut down.
It is not recommended to cut the power supply or to unplug the
power cord directly for doing this may cause damage or loss of
patient data.
If the system is not used for a long time, disconnect the external
power supply and all peripheral equipment after shutdown the
system.
3.4.3 Standby
Without any operation for a period of time, the system will enter standby mode. You can
set the waiting time to enter the standby mode.
19
3 Preparing the System
NOTE : Be careful to prevent crushing your hands when adjusting the monitor.
1. Grab the top and bottom of the monitor with two hands and tilt the monitor up and
down.
2. The tilt range is 0°~90°.
NOTE : When transporting or moving the system, tilt the monitor to the vertical
position and close the console to avoid damage to the screen.
20
3 Preparing the System
for details.
Operation: connect the foot switch to the system via USB port, and step on the foot switch
21
3 Preparing the System
1. Connect the network port of the I/O board and the local LAN port with
network cable
2. Preset the local network.
Setting path: [Preset] → [Network] → [Local TCP/IP], refer to “4.5 Network Preset”
for details.
Place the cursor on the network icon to automatically display the network information.
22
3 Preparing the System
Press [Preset] to enter the Preset screen. Click [Print Preset] to enter print
setting screen.
Enter the printer service name, select the printer type, and click [OK].
4. Add printer.
1) The preset print service is displayed in the print service list; select the
service by pressing [Set] button.
2) Click [Add Printer] and a dialog box pops up. Finish setup as needed.
3) Click [OK] to install the driver.
When using multiple peripherals that are not powered by the
auxiliary output of the system, it must be ensured that the total
leakage current of the peripherals and the system meets the
medical equipment electrical safety standards in relevant areas。
Once the system is connected to the DICOM server and the DICOM print server has been
set up, patient images and reports can be printed remotely. Refer to “11.2.2 DICOM Print”
for details.
3.10 Screens
The system enters the Home screen by default after power on.
23
3 Preparing the System
24
4 System Settings
4 System Settings
4.1 Overview
This section details the access to configuration and service functions for the system.
Customize your system to facilitate operation and clinical needs.
Item Description
Text box Select the text box, enter text and numbers directly.
Click the drop-down box to display all the sub-options under the current
Drop-down box
option.
Save Click to save current settings and return to the Home screen.
Cancel Click to cancel all modifications and return to the Home screen.
In the Preset screen, press [Preset] button again or click [Cancel] to cancel
Exit the Preset all modification and exit the Preset screen.
screen In the Preset screen, click [Save] to save all modification and exit the Preset
screen.
4.2.1 Region
The Region tab is the default page of the Preset screen.
Open the Region tab via [Preset] (press [Preset]) → [System Preset] → [Region]. This
tab is access to related information settings, such as hospital information, system
language, and system time.
25
4 System Settings
26
4 System Settings
4.2.2 General
Open the General tab via [Preset] → [System Preset] → [General].
This tab is access to patient information display settings of the Home screen, including
general measurement parameters, response after exam the exam and standby mode, etc.
Sleep Mode Enable/disable the standby mode and to set the waiting time
Standby
before entering the standby mode.
27
4 System Settings
4.2.3 Image
Open the Image tab via [Preset] → [System Preset] → [Image]. This tab is access to
related settings on save, output, display, and freeze for images and cines.
Center
Set whether to enable the centerline function in B mode.
Line
B
Horizontal
Set whether to enable the horizontal scale function in B mode.
Scale
Auto Set whether to enable the auto invert function in Color/Power
Color/Power
Invert mode.
Auto
Set whether to enable the auto invert function in M/PW/CW mode.
Invert
M/PW/CW
Display
Set the display format of M/PW/CW mode.
Format
28
4 System Settings
Item Description
Set whether to display the bone thermal index (TIB) and the
TI Type cranial thermal index (TIC) or soft tissue thermal index (TIS) in the
image area.
Set status after freeze.
Image Freeze
Options : Cine, Measure, Caliper
settings
Set whether to steer the sampling frame of the Color mode and
the sampling line of the PW mode together under B+Color+PW
Steer
mode.
Options : C&PW(together),C/PW(separately)
4.2.4 Application
Open the application tab via [Preset] → [System Preset] → [Application].
This tab is access to related setting of measurement and annotation.
29
4 System Settings
Item Description
Auto
parameter Set the auto parameter calculation.
calculation
30
4 System Settings
4.2.5 OB
Open the OB tab via [Preset] → [System Preset] → [OB]. This tab is access to related
setting for obstetric calculation and measurement display.
Measurement
Select measurement item for gestation age.
Item
Fetal GA Author Select calculation formula for obstetric measurements.
Fetal weight 2 Set fetal weight 2 for calculation and select calculation
31
4 System Settings
Item Description
4.2.6 Customize
Open the Customize tab via [Preset] → [System Preset] → [Customize]. This tab is
access to function definition for foot switch.
32
4 System Settings
Select the function assigned by the button and the function list displays
Function Type
below.
4.2.7 Admin
The system supports three types of users: the administrator, operater and emergency
user.
Icon User Permission
Access to all patient data in the system and be able to add users,
Administrator delete users, import data, export data, delete data, edit data, store
data, upgrade software / firmware, etc.
Access to all patient data in the system and be able to import data,
Operator
export data, delete data, edit data, and store data.
33
4 System Settings
Emergency Access to view and store patient data only. Data exporting, deleting
user and edition are not allowed.
After enable admin, login is required when exporting data, backing up data,
Enable Admin deleting data, etc.
Admin is not enabled by default.
To add a new user:
1) Click [Add] to pop up the Add New User dialog box.
Add
2) Enter a new user name, enter password, confirm the password, select a
new user type, and click [OK] to complete adding a new user.
To delete an existed user:
Delete
Select the target user in the user list and click [Delete].
To modify the password of current user:
1) Select the target user in the user list and click on [Change Password],
Change
a dialog box pops up.
Password
2) Enter the new password, confirm the new password and click [OK] to
complete the password modification.
If admin is enabled, you can access the data in the system only after you log on the
34
4 System Settings
system.
To log on the system:
1. Open the login dialog box.
To change user:
1. Click the icon at the lower right corner of the screen to pop up the dialog box as
shown below.
35
4 System Settings
2. Click [Logout].
Exam Library Display all available exam modes for current probe.
Exam Selected Display all selected exam modes for current probe.
36
4 System Settings
Item Description
1. Select the user-defined exam mode you want to delete in the Exam Preset
screen.
2. Remove the desired exam mode to the Exam library and click [Delete] to delete
a user-defined exam mode.
37
4 System Settings
1) Select the item(s) you want to add from the left column.
2) Click [>] to add the selected items of the left column to the
Add Measurement Item right column.
Or Click [>>] to add all the selected items of the left column
to the right column.
1) Select the item(s) you want to remove from the rigjt column.
2) Click [<] to remove the selected items from the right column
Remove measurement Items to the left column.
Or Click [<<] to remove all the selected items from the right
column to the left column.
Adjust measurement item Select a selected measurement item and click [Up] / [Down] to
position adjust the position of the measurement items.
Select a selected measurement item and click [Default] to set
Set default measurement
the measurement item as default.
item
Default measurement item is marked with a “√”.
38
4 System Settings
Operation Description
1) Select the item(s) you want to add from the left column.
2) Click [>] to add the selected item of the left column to the
Add Measurement Item right column.
Or click [>>] to add all the selected items of the left column
to the right column.
1) Select the item(s) you want to remove from the right
column.
2) Click [<] to remove the selected item from the right column
Remove measurement Item
to the left column.
Or click [<<] to remove all the selected items from the right
column to the left column.
Adjust measurement item Select a selected measurement item and click [Up] / [Down] to
position adjust the position of the measurement items.
Set default measurement Select a selected measurement item and click [Default] to set
39
4 System Settings
Operation Description
40
4 System Settings
1) Select the package(s) you want to add from the left column.
2) Click [>] to add the selected package from the left column to the
right column.
Add Measurement Or click [>>] to add all the selected packages from the left
Package column to the right column.
3) The measurement package can be activated by rotating the
corresponding knob under the [Exam] button of the touch
screen.
1) Select the package(s) you want to remove from the right
column.
Remove Measurement 2) Click [<] to remove the selected package from the right column
Package to the left column.
Or click [<<] to remove all the selected packages from the r right
column to the left column.
Adjust Measurement Select a selected measurement package and click [Up] / [Down] to
Package Position adjust the measurement package position.
Select a selected measurement package and click [Default] to set
Set Default
the measurement package as default.
Measurement Package
Default measurement package is marked with a “√”
4. Click [OK] to exit the New Measurement Package interface and the current
measurement package will be replaced by the new default measurement
package.
41
4 System Settings
Select a report template in the report template list and click [Default] to set the template
as default.
Default report template is marked with a “√”.
42
4 System Settings
43
4 System Settings
44
4 System Settings
4.7.3 Option
Enter the screen via [Preset] → [About & Maintenance] → [Option]. This screen lists all
the supported options and their installation status in the system.
4.7.4 Other
Enter the screen via [Preset] → [About & Maintenance] → [Other].
This screen is access to import / export for preset data and exam mode, restore factory
45
4 System Settings
46
5 Preparing an Exam
5 Preparing an Exam
Click to erase all patient data and create a new patient record by entering new
New Patient
patient information.
New Exam Click to end current exam and create a new exam for current patient.
Cancel Exam Click to cancel the current exam. Canceled exam cannot be restored.
Save Click to save patient information and back to the home screen.
47
5 Preparing an Exam
Item Description
Cancel Click to cancel patient information input and back to the home screen.
2. Move the cursor to the desired text box and enter the patient basic information.
Exam
Description
Information
DOB (Date of Enter directly in the text box, or click on the calendar icon to select the patient’s
Birth) date of birth in the calendar.
Enter the date of birth; the system will automatically calculate the patient’s age.
Age
Otherwise, manual input is required.
The system auto-synchronizes the serial number of the same patient in the
Accession
worklist.
Use keyboard for information entry. Press < Ctrl > + < Space > to switch
NOTE :
input method between Chinese and English.
3. Click the corresponding tab to select the appropriate exam type, and enter the
patient’s exam information.
CARD (Cardiac), VAS (Vascular), URO (Urology), SMP (Small part), PED (Pediatrics)
Exam
Exam Type Description
Information
Abdomen After entering the height and weight, the system calculates the
BSA (Body BSA according to the formula selected by the user in [System
Surface Area) Preset].To set the BSA formula, refer to “4.2.2 General” for
details.
When perform obstetrical exam, the system calculates the GA
and the EDD based on the entered calculation index; or
calculates the GA and the LMP based on the entered EDD and
Calculation uses this as a basis assessment for fetal physiology scores.
Obstetrics
Index Calculation index:
LMP: Enter the last menstrual period.
DOC: Enter the date of conception.
IVF: Enter the in vitro fertilization date.
48
5 Preparing an Exam
Exam
Exam Type Description
Information
PRV: Enter the previous exam date and GA obtained from the
previous exam.
EDD: Enter the excepted date of delivery.
Select fetus number.
Gestations
The system supports up to 4 fetal measurements.
After entering the height and weight, the system calculates the
BSA (Body BSA according to the formula selected by the user in [System
Cardiology Surface Area) Preset]. To set the BSA formula, refer to “4.2.2 General” for
details.
PPSA
Enter the PPSA coefficient, the default coefficient is 0.12.
coefficients
49
5 Preparing an Exam
Exam
Exam Type Description
Information
Small Parts / /
Pediatrics / /
Diagnostician Enter the name of the diagnostician who is responsible for the exam.
Charge
Enter the charge information.
Information
Charge
Briefly describe the charge information.
Description
5. Click [Save] to save the patient information and return to the Home screen.
NOTE: In case of emergency, you can perform an exam directly without patient
information entry.
When an image or cine is frozen, the system automatically generates a
patient ID.
50
5 Preparing an Exam
Patient Click to enter the Patient Information screen, and view the selected patient
Information information.
Click to enter the Review screen, and review the images of the selected
Review Image
exam.
Review Report Click to enter the Report screen and review the selected report.
New Exam Click to create a new exam for the selected patient.
Resume Exam Click to continue an exam that was interrupted within 24 hours.
Click to export the selected exam to the storage media supported by the
Export Exam
system.
Import Exam Click to import patient information from other media to the system.
Click to send the selected exam to the external storage media, DICOM
Send Exam
server or printer.
51
5 Preparing an Exam
Item Description
Query / Retrieve Click to view patient information stored on the DICOM server.
Select All Exams Click to select all patients and exams for batch operations.
Exit Click to exit Station screen and back to the Home screen.
52
5 Preparing an Exam
53
5 Preparing an Exam
3. Click [OK] to cancel the exam or [Cancel] to resume the current exam.
54
6 Optimizing the Image
B B-mode
6.2 B B-mode
The B-mode imaging uses a multi-beam scanning method to form each scanning line into
a 2D grayscale image, and displays the strength of the tissue echo signal by the intensity
of the brightness.
B-mode imaging is the most basic imaging mode by displaying the anatomical structure of
tissues and organs in real time.
B-mode can also be used together with other modes.
D Depth
G Gain (dB)
55
6 Optimizing the Image
FR Frame Frequency
DR Dynamic Range
6.2.2.1 Gain
The gain is used to adjust the brightness of the B image. You can brighten the image and
observe more echo signals by increasing the gain. However, with increased gain, more
noise has been introduced.this also introduces more noise.
Operation:
Rotate the [Gain] knob on the control panel to adjust the gain value.
6.2.2.2 TGC
The TGC controls compensation for echo attenuation as depth increases by varying the
gain of the received echoes at a specific depth.
The TGC curve is displayed on the left side of the image area when adjusting.
Operation:
Move the TGC slider on the control panel to adjust the brightness of the segmented image
of the corresponding depth.
6.2.2.3 Depth
Depth controls the flied of view. Increase the depth to look at a deeper or larger tissue
structure, decrease the depth to look at tissue structures near the skin line. Increase the
depth tends to decrease the frame rate.
Operation:
Press the [Depth] button on the control panel and then rotate the [Value] knob to adjust
the image depth. Press the [Depth] button again to exit this function.
NOTE : Adjustable range depends on the current probe. The actual depth is
shown in mm in the image parameter area.
6.2.2.4 Steer
Steer control is used to deflect the direction of beam without moving the probe.
Operation:
Press the [Angle] button on the control panel and then rotate the [Value] knob to adjust
the direction of the ultrasonic beam. Press the [Angle] button again to exit this function.
56
6 Optimizing the Image
Select [A.Power] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the acoustic power.
6.2.2.6 Auto-optimization
The system automatically optimizes image parameters based on the tissue characteristics
of the current scan area.
When auto-optimization function is enabled, an “Auto” mark is displayed in the image
parameter area.
Operation:
Press the [Gain] knob on the control panel to enable/ disable the auto optimization
function.
Adjust the focus of the ultrasonic beam to increase the resolution of the particular area
and focus the optimized image.
NOTE: Disable the SCI function before adjusting the focus number.
The phased array probe does not support adjusting the focus
number.
Adjust image quality by controlling the frequency. User selects the appropriate frequency
type and frequency value based on the depth of scanning and tissue characteristics to
achieve optimal imaging results.
The fundamental wave is marked with an “F” while the harmonic wave is marked with an“F
H”.
Operation:
1. Select [Image Quality] by pressing [ ] or [ ] button and then press the [Value]
57
6 Optimizing the Image
Change the display range of the grayscale to adjust the contrast of the image.
Operation:
Select [Dyn Ra.] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the dynamic range.
6.2.2.10 ZClear
Select [ZClear] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the ZClear level.
6.2.2.11 Persistence
Superimpose and average the images of adjacent frames to remove the noise and
enhance the details.
Operation:
Select [Persistence] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the persistence value.
Select [Graymap] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the effect of the grayscale map.
6.2.2.13 TSI
Select the acoustic velocity according to the scanning site to highlight the tissue texture.
Operation:
Select [TSI] by pressing [ ] or [ ] button and then rotate the [Value] knob to select the
58
6 Optimizing the Image
tissue texture.
6.2.2.14 Chrome
Replace grayscale distinction with color distinction to intuitively show grayscale level of
the images.
Operation:
Select [Chrome] by pressing [ ] or [ ] button and then rotate the [Value] knob to
change the color.
Change the scan range without moving the probe or ROI box.
You can get a wider FOV (field of view) by a larger scan range. A larger FOV trends to
decrease the frame rate.
Operation:
Select [Field of View] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the FOV.
Compound imaging is the process of merging frames of different steering angles into a
single frame. This function is used to improve the spatial resolution and signal-to-noise
ratio of the image, making the image subtle and clean.
There are two types of compound imaging: SCI (Spatial Compound Imaging) and FCI
(Frequency Compound Imaging).
SCI (Spatial Compound Imaging)
Operation:
Select [SCI] by pressing [ ] or [ ] button and then press the [Value] knob to to enable /
disable SCI function. Rotate the [Value] knob to adjust the SCI level.
FCI (Frequency Compound Imaging)
FCI is only available for the fundamental frequency.
Operation:
Select [SCI] by pressing [ ] or [ ] button and then press the [Value] knob to to enable /
disable FCI function.
Line density determines the number of the ultrasonic scanning lines of a certain width or
angle.
Operation:
Select [Line Density] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust line density level.
59
6 Optimizing the Image
Select [Rotation] by pressing [ ] or [ ] button and then rotate the [Value] knob to
change the orientation of the image display.
Rotation Options: 0°, 90°, 180°, and 270°.
NOTE : “ ” of the image is consistence with the orientation mark of the probe.
Select [UD Reverse] or [LR Reverse] by pressing [ ] or [ ] button and then press the
[Value] knob to enable / disable image reverse function.
When the extended imaging function is enabled, trapezoidal imaging (for linear array
probes) or extended scanning angle (for convex array and phased array probes) will be
displayed.
Operation:
Select [EFOV] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the EFOV level.
6.3 M-mode
M-Mode is used to determine patterns of motion for objects along the M-line. The most
common use is for viewing motion patterns of the heart.
60
6 Optimizing the Image
5. Move the M-line on B-mode image with the trackball and you can learn the
movement of the anatomy along the M-line in the M-mode image.
6. Optimize the M-mode image. (Refer to “6.3.2 Image Optimization (M)” for
details.)
The parameter area on the right side of the screen provides real-time parameter
information of the M-mode. The meanings of each parameter are as follows:
Parameter M-mode
D Depth
G Gain (dB)
V Velocity
DR Dynamic Range
6.3.2.1 Gain
The gain is used to adjust the brightness of the M-mode image.You can brighten the
image and observe more echo signals by increasing the gain. However, with increased
gain, more noise has been introduced.this also introduces more noise.
Operation:
In M-mode, rotate the [Gain] knob on the control panel to adjust the gain value.
61
6 Optimizing the Image
Change the display range of the grayscale to adjust the contrast of the image.
Operation:
Select [Dyn Ra.] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the dynamic range.
Select [Gray Map] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the gray map.
6.3.2.4 Speed
Control the imaging speed.The speed is displayed in the image menu area.
Operation:
Select [Speed] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the M speed.
6.3.2.5 Chrome
Replace grayscale distinction with color distinction to intuitively show grayscale level of
the images.
Operation:
Select [Chrome] by pressing [ ] or [ ] button and then rotate the [Value] knob to
change the selection color.
Change the display format of B image and M image for easy comparison of images.
Operation:
[System Preset] → [Image] → [M/PW/CW] to set the display format.
62
6 Optimizing the Image
while color below the color bar baseline indicates the flow away from the probe.
The brighter the color, the faster the blood flows; the darker the color, the slower the blood
flows.
To perform a color mode imaging:
1. Observe the anatomical structure, adjust the B-mode image and probe to make
the target area at the center of the B-mode image.
2. Press the [CFM] button on the control panel to enter Color-mode.
3. Adjust the position and size of the ROI box.
When the ROI box displayed a solid line frame, roll the trackball to adjust the position
When the ROI box displayed a dotted line frame, roll the trackball to adjust the size of
Press the [Set] button to toggle between position and size adjustment.
4. Optimize the image. (Refer to “6.4.3 Image Optimization (Color/PDI)” for details )
The parameter area on the right side of the screen provides real-time parameter
information of the Color-mode. The meanings of each parameter are as follows:
Parameter Color-mode
G Gain (dB)
WF Wall Filter
NOTE : PDI mode only provides perfusion information of blood flow, but
cannot indicate the direction and speed of blood flow.
When the ROI box displayed a solid line frame, roll the trackball to adjust the position
63
6 Optimizing the Image
When the ROI box displayed a dotted line frame, roll the trackball to adjust the size of
Press the [Set] button to toggle between position and size adjustment.
G Gain (dB)
WF Wall Filter
6.4.3.1 Gain
In Color/PDI mode,Rotate the [Gain] knob on the control panel to adjust the Color/PDI
gain value.
Calibrate the 0 position of the scale for the best blood flow speed display.
Operation:
Press the [Base Line] button on the control panel and then rotate the [Value] knob to
adjust the image depth. Press the [Base Line] button again to exit this function.
64
6 Optimizing the Image
6.4.3.4 PRF
Adjust the speed range of color flow. The real-time PRF value is displayed in the image
parameter area.
Operation:
Press the [Scale] button on the control panel and then rotate the [Value] knob to adjust
the image depth. Press the [Scale] button again to exit this function.
NOTE : Select appropriate PRF to match the flow velocity. Aliasing may resulted if low
PRF is used for high flow velocity or small signal lost may resulted if high PRF
is used for low flow velocity.
Filter low frequency noise to reduce the impacts of breathing and other activities.
The filtering value is displayed in the image parameter area.
Operation:
Select [Wall Filter] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the filtering value.
6.4.3.6 Smooth
Select [Smooth] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the smooth level.
6.4.3.7 Persistence
Superimpose and average the images of adjacent frames to remove the noise and
enhance the details.
Operation:
Select [Persistence] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the persistence level.
6.4.3.8 Priority
Set the priority to displays blood flow information by color signals or gray signals.
The larger the value, color signals are prior to display; while the smaller the value, gray
signals are prior to display.
Operation:
Select [Priority] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the color priority value.
65
6 Optimizing the Image
Color
Multiple sets of color maps are provided to optimize color flow display. This system
provides 21 different maps for users to choose. The ordinary map is marked with “V” while
2D map is marked with “VV”.
Operation:
Select [Color Map] by pressing [ ] or [ ] button and then rotate the [Value] knob to
choose the color map.
PDI
Provide color image display effect parameters, including energy spectrum and directional
energy spectrum. This system provides 8 different effect maps for users to choose. The
energy maps are P0-3 while the direction energy maps are dP0-3. Directional energy
spectrum also shows the direction information of blood flow.
Operation:
Select [Color Map] by pressing [ ] or [ ] button and then rotate the [Value] knob to
choose the color map.
6.4.3.10 Sensitivity
Reflect the ability to reflect minimal detectable blood flow which is used to adjust the
accuracy of color flow display. The higher the sensitivity, the more sensitive the response
to low-speed blood flow and small blood vessel signals, and the better the image quality.
Operation:
Select [Sensitivity] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the sensitivity.
Select [Flow State] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the flow state.
Change the display range of the grayscale to adjust the contrast of the image.
Operation:
Select [Dyn Ra.] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the dynamic range.
66
6 Optimizing the Image
Line density determines the number of the ultrasonic scanning lines of a certain width or
angle.
Operation:
Select [Line Density] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust line density level.
6.4.3.14 Invert
Select [Line Density] by pressing [ ] or [ ] button and then press the [Value]
knob to enable or disable the invert function.
[System Preset] → [Image] → [Color/Power], check [Auto Invert] to enable or
disable the auto-invert function.
Set the maximum width of the B images the same width as the ROI box.
Operation:
Select [Line Density] by pressing [ ] or [ ] button and then press the [Value] knob to
enable / disable the function.
Select [HR Flow] by pressing [ ] or [ ] button and then press the [Value] knob to
enable / disable the function.
67
6 Optimizing the Image
Roll the trackball left and right to adjust the position of the sampling line.
4. Adjust the sample volume and the cursor angle of blood flow direction.
Press the [Sample Volume] button on the control panel and then rotate the [Value] to
adjust the sample volume. Press the [Sample Volume] button again to exit this
function.
Select [Angle] by pressing [ ] or [ ] button and then press the [Value] knob to
adjust the blood flow direction cursor angle so as to parallel to the blood flow direction.
5. Press the [PW] button on the control panel again to enter the PW-mode.
After entering PW-mode, the image area is divided into two parts: B-mode image
at the top and PW-mode image at the bottom.
68
6 Optimizing the Image
The Y-axis indicates frequency change, and can also indicate the speed of forward or
Press the [Update] button on the control panel to toggle between B-mode and
pre-activated PW-mode.
You can adjust the size and depth of sample volume and the cursor angle of the blood
Select [Volume] by pressing [ ] or [ ] button and then rotate the [Value] knob to
6. Optimize the image. (Refer to “6.5.3 Image Optimization (PW/CW)” for details)
The parameter area on the right side of the screen provides real-time parameter
information of PW-mode. The meanings of each parameter are as follows:
Parameter PW-Mode
G Gain (dB)
WF Wall Filter
SV Sample Volume
69
6 Optimizing the Image
Roll the trackball left and right to adjust the position of the sampling line.
Select [Angle] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the blood flow direction cursor angle so as to parallel to the blood flow direction;
5. Press the [CW] button on the control panel to enter the CW mode.
After entering CW mode, the image area is divided into two parts: B-mode image
at the top and CW mode image at the bottom
The Y-axis indicates frequency change, and can also indicate the speed of forward or
Press the [Update] button on the control panel to toggle between B-mode and
CW-mode.
You can adjust the size and depth of sample volume and the cursor angle of the blood
Select [Volume] by pressing [ ] or [ ] button and then rotate the [Value] knob to
6. Optimize the image. (Refer to “6.5.3 Image Optimization (PW/CW)” for details)
The parameter area on the right side of the screen provides real-time parameter
information of CW mode. The meanings of each parameter are as follows:
Parameter CW Mode
G Gain (dB)
WF Wall Filter
70
6 Optimizing the Image
Parameter CW Mode
6.5.3.1 Gain
6.5.3.2 Baseline
Calibrate the 0 position of the scale for the best blood flow speed display.
Operation:
Press the [Base Line] button on the control panel and then rotate the [Value] knob to
adjust the image depth. Press the [Base Line] button again to exit this function.
Adjust the transmission frequency of the probe in Doppler mode. The higher the frequency,
the higher the resolution and sensitivity, but the penetrating power is reduced. Select
appropriate probe frequency based on the detection depth and tissue texture.
Operation:
6.5.3.4 PRF
NOTE : Select appropriate PRF to match the flow velocity. Aliasing may result if low
PRF is used for high flow velocity or small signal lost may result if high PRF is
used for low flow velocity.
6.5.3.5 SV(PW)
71
6 Optimizing the Image
Operation:
Press the [Sample Volume] button on the control panel and then rotate the [Value] knob
to adjust the SV. Press the [Sample Volume] button again to exit this function.
Change the display range of the grayscale to adjust the contrast of the image.
Operation:
Select [Dyn Ra.] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the dynamic range.
Select [Gray Map] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the gray map.
6.5.3.8 Speed
Control the imaging speed. The smaller the value, the faster the speed is.
Operation:
Select [Speed] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the imaging speed.
6.5.3.9 Chrome
Replace grayscale distinction with color distinction to intuitively show grayscale level of
the images.
Operation:
Select [Chrome] by pressing [ ] or [ ] button and then rotate the [Value] knob to
select the chrome color.
6.5.3.10 Angle
Select [Angle] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the flow cursor angle.
Set the trace area of the Doppler wave in the spectrum map.
Operation:
72
6 Optimizing the Image
tracing area.
Filter low frequency noise to reduce the impacts of breathing and other activities.
Operation:
Select [Wall Filter] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the filtering value.
Select [T/F Resolution] by pressing [ ] or [ ] button and then rotate the [Value] knob
to adjust the T/F res level.
Trace the spectral Doppler wave and perform parameter calculations. The results of the
automatic calculation will be displayed in the measurement results window.
In the real-time state, the automatic spectrum results obtained from the latest or more
cardiac cycles are always displayed; in the frozen and cine replay status, the
corresponding measurement results of the traced area are displayed.
Operation:
Select [Auto Calc] by pressing [ ] or [ ] button and then press the [Value]
knob to enable the auto calculation function.
[System Preset] → [Application] → [Auto Calc] → [Auto Calcu Cycle], set
the cardiac cycles for calculation.
[System Preset] → [Application] → [Auto Calc] → [Auto Calc Para], set the
parameter items for automatic calculation.
6.5.3.15 Invert
Select [Invert] by pressing [ ] or [ ] button and then press the [Value] knob to
enable / disable the invert function.
[System Preset] → [Image] → [M/PW/CW], check [Auto Invert] to enable the
auto-invert function.(Only applicable for linear probe)
Change the display format of B image and PW/CW image to facilitate comparison of
images.
Operation:
73
6 Optimizing the Image
6.5.3.17 Volume
Control and adjust the audio volume produced by the spectral Doppler which helps to
determine the blood flow state and properties more effectively.
Operation:
Select [Volume] by pressing [ ] or [ ] button and then rotate the [Value] knob to adjust
the volume.
Filter out the redundant low frequency signals produced by non-blood flow movement,
e.g.breathing, heartbeat and probe movement.
Operation:
Select [HPRF] by pressing [ ] or [ ] button and then rotate the [Value] knob to to
enable / disable HPRF function.
6.5.3.19 Steer
Steer control is used to deflect the direction of beam without moving the probe.
Operation:
Press the [Angle] button on the control panel and then rotate the [Value] knob to adjust
adjust the direction of the ultrasonic beam. Press the [Angle] button again to exit this
function.
Display the PW image together with B image or with B+Color image at the same time.
Operation:
Select [Multi Sync] by pressing [ ] or [ ] button and then rotate the [Value] knob to to
enable / disable multi sync function.
74
6 Optimizing the Image
In B+M mode, press the [CFM] button to enter the pre-activated CFM mode and then
In B+Color mode, B+Color+PW mode or B+Color+CW mode, double press the [M]
Move the trackball left and right to adjust the position of sampling line, the color ROI
When the ROI box displayed a solid line frame, roll the trackball to adjust the position
When the ROI box displayed a dotted line frame, roll the trackball to adjust the size of
Press the [Set] button to toggle between position and size adjustment.
75
7 Image Review and Management
76
7 Image Review and Management
When the ROI box displayed a solid line frame, roll the trackball to adjust the position
When the ROI box displayed a dotted line frame, roll the trackball to adjust the size of
77
7 Image Review and Management
Press the [Set] button to toggle between position and size adjustment.
3. After ROI adjustment, press the [Value] knbo again on the control panel to zoom
in the image.
4. Press the [Value] knob again to exit image magnification function.
NOTE : Local zooming is only applicable to real-time images.
78
7 Image Review and Management
1. Select [Start Time] by pressing [ ] or [ ] button and then rotate the [Value]
knob to set the start time of the cine replay region.
2. Select [End Time] by pressing [ ] or [ ] button and then rotate the [Value]
knob to set the end time of the cine replay region.
NOTE : The system plays the cine replay region automatically after replay region
setting.
Press the [Save Cine] button on the control panel to save the cine replay
region to the system.
79
7 Image Review and Management
NOTE : When reviewing the cine, image parameters can be adjust for better
observation.
Select [Replay Speed] by pressing [ ] or [ ] button and then rotate the [Value] knob to
adjust the review speed and enable / disable the auto replay.
Method 2
Roll the trackball quickly to activate the automatic replay function. Cines are played
backwards by rolling the trackball to the left quickly and are played forwards by rolling the
trackball to the right quickly.
NOTE : The system plays the cine review region automatically after review region has
been set.
7.5 Annotation
Annotation function provides comment, arrow and body mark functions to the image.
Comment and arrows are available to the ultrasound images in freeze and real-time
status.
NOTE : Users can preset annotation via [System Preset] → [Application]. Refer to
“4.2.4 Application” for details.
7.5.1 Comment
Users can enter text content directly or add preseted comment item to images.
Select [English] by pressing [ ] or [ ] button and then rotate the [Value] knob to
switch text language between Chinese and English.
Select [Exam] by pressing [ ] or [ ] button and then rotate the [Value] knob to set
80
7 Image Review and Management
Select [Font Size] by pressing [ ] or [ ] button and then rotate the [Value] knob to
3. Click comment item you want to add on the right, roll the trackball to move the
comment item to the desired position, press the [Set] button.
4. If there is no applicable comment item, roll the trackball to move the comment
item to the desired position and enter the comment through keyboard, and press
the [Set] button.
5. To add more comments, repeat Step 3 or 4.
6. Press the [Comment] button on the control panel again to exit comment
function.
1. Move the cursor over the comment item to activate the item. Activated comment
item or text is highlight in green.
2. Press the <←> key to delete undesirable characters and enter new comments
directly. After the modification, move the cursor away from the text comment to
complete the modification.
1. Move the cursor to the text comment to be moved, press the [Set] button, the
text comment is highlighted with green box.
2. Roll the trackball to move the text comment to the desire position, press the [Set]
button.
Select [Font Size] by pressing [ ] or [ ] button and then rotate the [Value] knob to set
text size. Or you can change the font size through Preset screen (press [Preset]). Refer to
“4.2.4 Application” for details.
Select [Text Hide] by pressing [ ] or [ ] button and then press the [Value] knob to hide
or display comments.
81
7 Image Review and Management
7.5.2 Arrow
Select [Text Hide] by pressing [ ] or [ ] button and then press the [Value] knob to hide
or display arrows.
1. Move the cursor over the body mark to be moved, press the [Set] button.
2. Move the trackball to the desire position and press the [Set] button to place the
body mark.
82
7 Image Review and Management
Press the [Clean] button on the control panel to delete the body mark.
83
8 Measurement
Before making a measurement of the target area, make sure that the
target area and image to be measured are correct and qualified to
avoid misdiagnosis.
Before performing Doppler color flow image measurement, do not let
the probe emit direction perpendicular to the direction of blood flow,
otherwise incorrect blood flow information may be displayed.
NOTE : During the measurement, if the system has been powered off, all unsaved
data will lose.
Caliper measurement: in the real-time or frozen status, press the [Caliper] button on
Application measurement: in the real-time or frozen status, click [Meas] after the
Roll the trackball to select the measurement item and press the [Set] button.
3. Perform a measurement:
Move the cursor to the target area, press the [Set] button, a measurement cursor
appears. At the same time, the result window displays measurement results and
calculation data.
After measurement, place the cursor on the title bar of the result window to change the
cursor into a hand cursor. Press the [Set] button and roll the trackball at the same time
5. After the measurement is completed, press the [Caliper] button on the control
panel to exit the general / application measurement function.
84
8 Measurement
8.2.1.1 Distance
8.2.1.2 Depth
Measure the distance from the center of sector to the cursor (for surface probe) or the
distance from probe surface to the cursor in the direction of ultrasonic beam (for linear and
convex array probe).
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Depth] in the measurement menu, and the cursor appears.
3. Move the cursor to the desired position and press the [Set] button.
4. Results are displayed in the result window.
Measure the length of a curve. Available measurement method includes Trace and Spline.
Trace Len
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Trace Len] in the measurement menu, and the cursor appears.
3. Move the cursor to the start point of the trace line and press the [Set] button.
4. Roll the trackball to trace the outline of the target and press the [Set] button to fix
the end point of the trace line.
5. Results are displayed in the result window.
Trace Len (Spline)
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Trace Len (Spline)] in the measurement menu, and the cursor appears.
85
8 Measurement
3. Move the cursor to the start point of the target area and press the [Set] button.
4. Move the cursor along the target area to fix the second point and press the [Set]
button.
5. Repeat step 4. A maximum of 12 points can be fixed.
6. Press the [Set] button twice to set the end point of the spline.
7. Results are displayed in the result window.
8.2.1.4 Angle
86
8 Measurement
8.2.1.6 Volume
Measure the volume of the target object. Available measurement method includes 3 Dist,
Ellipse and Ellipse Dist.
3 Dist
To obtain the volume by measuring the vertical section area of the target object.
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Volume] in the measurement menu, and the cursor appears.
3. Measures the length (D1), width (D2), and height (D3) of the target object.
4. The system automatically calculates the volume. The length, width, height, and
volume of the target object are displayed in the result window.
D1, D2, and D3 correspond to the length, width, and height of the target object.
Volume (Ellipse)
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Volume (Ellipse)] in the measurement menu, and the cursor appears.
3. Move the cursor to set the start point of the first axis of the ellipse and press the
[Set] button.
4. Roll the trackball to set the end point of the first axis of the circle and press the
87
8 Measurement
[Set] button.
5. Move the cursor to adjust the second axis of the ellipse, and press the [Set]
button to complete the volume ellipse measurement.
6. The system automatically calculates the volume. Results are displayed in the
result window.
Where, “a” corresponds to the first axis of the ellipse, and “b” corresponds to the
Volume (E+Dist)
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Volume (E+Dist)] in the measurement menu, and the cursor appears.
3. Measure the area of the vertical section as described in “8.2.1.5 Area &
Circumference”.
4. Move the cursor to set the start point of the third axis of the ellipse and press the
[Set] button.
5. Move the cursor to set the end point of the third axis of the ellipse and press the
[Set] button.
6. The system automatically calculates the volume. The area, height, and volume of
the target object are displayed in the result window.
Where, “a” corresponds to the first axis of the ellipse, and “b” corresponds to the
second axis of the ellipse, “m” corresponds to the third axis of the ellipse.
88
8 Measurement
8.2.1.8 Parallel
Measure the distance between five parallel lines on the ultrasound image.
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Parallel] in the measurement menu, and the cursor and two lines
perpendicular to each other appear.
3. Move the cursor to set the start point of the first parallel and press the [Set]
button.
4. Move the cursor along the vertical line and press the [Set] button to set the
second parallel.
5. Repeat step 4 to set the third, fourth, fifth parallel as needed.
6. Results are displayed in the result window.
Measure the lengths of any two lines and calculates their ratio.
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Ratio (D)] in the measurement menu, and the cursor appears.
3. Set two line segments as described in “8.2.1.1 Distance”.
4. The system automatically calculates the ratio, and the results are displayed in
the result window.
Measure the areas of any two enclosed areas and calculates their ratio.
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [Ratio (Trace)], [Ratio (Ellipse)], [Ratio (Spline)] or [Ratio (Cross)] in the
measurement menu, and the cursor appears.
3. Use the area measurement method (choose one of trace, ellipse, spline, cross)
to set two closed areas.
4. The system automatically calculates the ratio, and the results are displayed in
the result window.
8.2.1.11 B Hist
Measures and calculates the gray scale distribution of the ultrasonic echo signals in an
enclosed region.
Available measurement method includes B Hist (Trace), B Hist (Ellipse), B Hist (Spline)
and B Hist (Rectangle).
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
89
8 Measurement
2. Click [Hist (Trace)], [Hist (Ellipse)], [Hist (Spline)] or [Hist (Rectangle)] in the
measurement menu, and the cursor appears.
3. Set up an enclosed area with the area measurement method as described in
“8.2.1.5 Area & Circumference”.
4. The system automatically calculates the gray distribution of the ultrasonic echo
signal in this area and displays it at the top of the screen.
8.2.1.12 B Profile
Measure and calculates the gray distribution of the ultrasonic echo signals across a line.
Perform the following steps:
1. Press the [Caliper] button on the control panel to start measurement.
2. Click [B Profile] in the measurement menu, and the cursor appears.
3. Set a line with the distance measurement method as described in “8.2.1.1
Distance”.
4. The system automatically calculates the gray distribution of the ultrasonic echo
signal on the line and displays it at the top of the screen.
Measure the blood flow through a blood cross section per unit time. Refer to “8.2.3.7 Vol
Flow” for details.
8.2.1.15 IMT
Measure the intima-media thickness of the carotid artery. Refer to “8.3.4.3 Vascular Study
and Calculation” for details.
90
8 Measurement
Measure the distance between two points on the sampling line at a certain time.
Perform the following steps:
1. In M-mode, press the [Caliper] button on the control panel to start measurement.
2. Click [Distance (M)] in the measurement menu, and the cursor appears.
3. Move the cursor to the start point and press the [Set] button.
4. Move the cursor to the end point and press the [Set] button.
5. Results are displayed in the result window.
8.2.2.3 Slope
Measures the depth and time between two points and calculates the slope.
Perform the following steps:
1. In M-mode, press the [Caliper] button on the control panel to start measurement.
2. Click [Slope] in the measurement menu, and the cursor appears.
3. Move the cursor to set the first point and press the [Set] button.
4. Move the cursor to set the second point and press the [Set] button.
5. Results are displayed in the result window.
8.2.2.4 HR (M)
Measures the time of N (N≤7) cardiac cycles and calculates the heart rate in M-mode
image.
Perform the following steps:
1. Set the cardiac cycle.
Select the number of cardiac cycles from the [Cardiac Cycle] drop-down list.
Click [Save] to save current setting and return to the Home screen.
2. In M-mode, press the [Caliper] button on the control panel to start measurement.
3. Click [HR (M)] in the measurement menu, and the cursor appears.
91
8 Measurement
4. Move the cursor to the start point and press the [Set] button.
5. Move the cursor to set the end point in horizontal direction and press the [Set]
button.
6. Results are displayed in the result window.
Figure 31 HR Result
NOTE : During the HR measurement, the number of cardiac cycles between the
start and end points must be consistent with the preset. Otherwise,
misdiagnosis may occur.
8.2.2.5 Velocity
Calculate the average velocity by measuring the distance and time between two points.
1. In M-mode, press the [Caliper] button on the control panel to start measurement.
2. Click [Velocity] in the measurement menu, and the cursor appears.
3. Move the cursor to the desired position and press the [Set] button.
4. Results are displayed in the result window.
8.2.3.2 HR (D)
Measures the time of N (N≤7) cardiac cycles and calculates the heart rate in Doppler
mode image. The measurement procedure is the same as that of HR (M). Refer to
“8.2.2.4 HR (M)” for details.
92
8 Measurement
8.2.3.3 Vel
8.2.3.4 Acceleration
Measure the velocity and time interval between two points on the Doppler image to
calculate the acceleration, differential pressure, speed difference, and spectrum correction
angle.
Perform the following steps:
1. In PW/CW mode, press the [Caliper] button on the control panel to start
measurement.
2. Click [Acceleration] in the measurement menu, and the cursor appears.
3. Move the cursor to the first point and press the [Set] button.
4. Move the cursor to the second point and press the [Set] button.
5. The acceleration, differential pressure, speed difference, and spectrum
correction angle between the two points are displayed in the result window.
8.2.3.5 D Trace
Trace one or several Doppler waveform on the PW mode image to obtain speed and
pressure gradient.
The Doppler spectrum diagram is shown below:
93
8 Measurement
the end position, and press the [Set] button to complete the trace. The
measurement result is displayed in the result window.
The parameters displayed in the result window are described as follows:
Parameter Description
Peak Systolic The maximum velocity of red blood cells passing through the
PS
Velocity sampling volume
End-diastolic
ED The blood flow velocity at the end of the cardiac cycle
velocity
Min-diastolic
MD The minimum velocity of the cardiac cycle
velocity
Tb
TAMAX (cm/ s) T V (t )dt /(Tb
a
Ta )
Pulsatility
PI PI = |( PS(m/s)-ED(m/s)) /TAMAX(m/s) |
Index
Resistive
RI RI = |( PS(m/s)-ED(m/s)) /PS(m/s) |
Index
θ / Correction Angle
94
8 Measurement
Parameter Description
8.2.3.6 PS/ED
Measures the velocity of peak systolic point (PS) and end diastolic (ED) on the Doppler
spectrum.
1. In PW/CW mode, press the [Caliper] button on the control panel to start
measurement.
2. Click [PS/ED] in the measurement menu, and the cursor appears.
3. Move the cursor to the peak systolic velocity on the Doppler spectrum and press
the [Set] button.
4. Move the cursor to the end diastolic velocity on the Doppler spectrum and press
the [Set] button.
5. The RI, S/D and correction between the two points are displayed in the result
window.
Measure the blood flow through a blood cross section per unit time.
Perform the following steps:
1. In PW/CW mode, press the [Caliper] button on the control panel to start
measurement.
2. Click [Vol Flow] in the measurement menu to display the sub-menu of volume
flow.
3. Click [Vas Area] in the measurement menu, and the cursor appears.
4. Move the cursor to the start point of the target area on B-image and press the
[Set] button.
5. Move the cursor to the end point of the target area on B-image and press the
[Set] button.
6. The flow area is displayed in the result window.
NOTE : Steps 4 to 5 can also be performed in B-mode.
7. Click [TAMEAN] or [TAMAX] in the measurement menu, and the cursor
appears.
8. Move the cursor to the start point of the target area on Doppler Spectrum and
press the [Set] button.
9. Move the cursor to the end point of the target area on Doppler Spectrum and
press the [Set] button.
10. The system automatically generates trace lines. Results are displayed in the
result window.
The parameters displayed in the result window are described as follows:
95
8 Measurement
Measures two D velocity and calculates their ratio for blood analysis.
Perform the following steps
1. In PW/CW mode, press the [Caliper] button on the control panel to start
measurement.
2. Click [Ratio (Vel)] in the measurement menu, and the cursor appears.
3. Move the cursor to fix the two points for velocity measurement and press the [Set]
button.
4. Move the cursor, the real time velocity of the position is displayed in the result
window.
5. The velocity ratio of the two points is displayed in the result window.
96
8 Measurement
VAS Applicable for measurement of carotid, cerebral, upper and lower extremities
(Vascular) vessels, etc.
EM
Applicable for measurement of EM general practice.
(Emergency)
8.3.1 Abdomen
The measurement items, calculation items and measurement method for 2D abdomen
measurement are shown in the table below.
Table 24 2D Abdomen Measurement
Items Description Methods or Formula
Liver / /
97
8 Measurement
The measurement items, calculation items and measurement method for Doppler
abdomen measurement are shown in the table below.
98
8 Measurement
8.3.2 Obstetrics
Note: Before performing obstetric measurements, please make sure that the system
date is accurate, otherwise, the GA and EDD will be calculated incorrectly.
The measurement items, calculation items and measurement method for 2D obstetrics
measurement are shown in the table below.
Table 26 2D Obstetrics Measurement
Item Description Methods or Formula
99
8 Measurement
100
8 Measurement
101
8 Measurement
The measurement items, calculation items and measurement method for M obstetrics
measurement are shown in the table below.
Table 27 M Obstetrics Measurement
Item Sub-items Methods or Formula
102
8 Measurement
The measurement items, calculation items and measurement method for Doppler
obstetrics measurement are shown in the table below.
Table 28 Doppler Obstetrics Measurement
Item Sub-items Methods or Formula
103
8 Measurement
This section mainly introduces the Fetal Growth Curve and Fetal Biophysical Profile of OB
reports. For details about report reviewing, printing and exporting etc., please refer to“9
Report”
Fetal Growth Curve
The fetal growth curve is to compare the fetal measurement data with the standard growth
curve to determine whether the fetal development is normal.
Perform the following steps:
1. Press the [Patient] button on the control panel to enter the Patient screen.
2. Click the [Obstetrics] tab, and enter the relevant information as needed.
3. Measure one or more fetal growth parameters.
4. Press the [Report] button on the control panel to enter the obstetric report, and
click [Fetal Growth Curve].
Fetal Biophysical Profile
The fetal biophysical profile is obtained by experiment and measuring the fetal growth
indicators of the fetus to evaluate state of the fetus. Click [Analyze] on the obstetric report
page to display the fetal scores.
The scoring criteria are based on Vintzileos formula, as shown in table below.
Table 29 Fetal Scores
Reactive FHR
<2, or Reactive FHR
FHR ≥15bpm, duration≥15s, 30 minutes
≤15bpm
≥2 times
104
8 Measurement
8.3.3 Cardiology
The measurement items, calculation items and measurement method for 2D cardiology
measurement are shown in the table below.
Table 31 2D Cardiology Measurement
Items Description Methods or Formula
IVC Diam (Insp) Inferior vena cava inspiration See “8.2.1.1 Distance”
105
8 Measurement
Diameter
106
8 Measurement
and Calculation”
The measurement items, calculation items and measurement method for M cardiology
measurement are shown in the table below.
Table 32 M Cardiology Measurement
Items Description Methods or Formula
LA/AO(M) / /
107
8 Measurement
LVET (M) Left Ventricular ejection time See “8.2.2.2 Time (M)”
RVET (M) Right Ventricular ejection time See “8.2.2.2 Time (M)”
ACS (M) Aortic valve Cusp Separation See “8.2.2.1 Distance (M)”
LVET (M) Left Ventricular ejection time See “8.2.2.2 Time (M)”
108
8 Measurement
RVET (M) Right Ventricular ejection time See “8.2.2.2 Time (M)”
The measurement items, calculation items and measurement method for Doppler
cardiology measurement are shown in the table below.
Table 33 Doppler Cardiology Measurement
Items Description Methods or Formula
LVET (Doppler) Left Ventricular Ejection Time See “8.2.3.1 Time (D)”
109
8 Measurement
dP/dt / /
RVET (Doppler) Right Ventricular Ejection Time See “8.2.3.1 Time (D)”
110
8 Measurement
Velocity
111
8 Measurement
Simpson SP
Table 34 Simpson SP
Measurement and
Description Method and Formula
study
𝐿𝑉𝐿𝑑 𝑎𝑝𝑖𝑐𝑎𝑙(𝑐𝑚)
𝐸𝐷𝑉(𝑚𝑙) = 𝜋 ×
20
20
× ∑ 𝑟𝑖2 (𝑐𝑚)
End-diastolic Left Ventricular
EDV(Simp SP) 𝑖=1
Volume
LVLd apical: left Ventricular Long-axis
Length at End-diastole in apical view
𝐿𝑉𝐿𝑠 𝑎𝑝𝑖𝑐𝑎𝑙(𝑐𝑚)
𝐸𝑆𝑉(𝑚𝑙) = 𝜋 ×
20
20
× ∑ 𝑟𝑖2 (𝑐𝑚)
End-systolic Left Ventricular
ESV(Simp SP) 𝑖=1
Volume
LVLs apical: left Ventricular Long-axis
Length at End-systole in apical view
112
8 Measurement
Measurement and
Description Method and Formula
study
Simpson BP
Table 35 Simpson BP
Measurement and
Description Method and Formula
study
End-diastolic Left Ventricular
EDV(Simpson BP) *1
Volume
End-systolic Left Ventricular
ESV(Simpson BP) *2
Volume
EDV Index (Simpson
/ EDV Index=EDV/BSA
BP)
ESV Index (Simpson
/ ESV Index=ESV/BSA
BP)
*1 Means:
20
𝑀𝐴𝑋{𝐿𝑉𝐿𝑑2𝑖 (𝑐𝑚), 𝐿𝑉𝐿𝑑4𝑖 (𝑐𝑚)}⁄
𝐸𝐷𝑉(𝑚𝑙) = 𝜋 × 20 × ∑(𝑟2𝑖 (𝑐𝑚) × 𝑟4𝑖 (𝑐𝑚))
𝑖=1
*2 Means:
113
8 Measurement
20
𝑀𝐴𝑋{𝐿𝑉𝐿𝑠2𝑖 (𝑐𝑚), 𝐿𝑉𝐿𝑠4𝑖 (𝑐𝑚)}⁄
𝐸𝑆𝑉(𝑚𝑙) = 𝜋 × 20 × ∑(𝑟2𝑖 (𝑐𝑚) × 𝑟4𝑖 (𝑐𝑚))
𝑖=1
LV volume (A2C):
20
𝐿𝑉𝐿𝑑2𝑖 (𝑐𝑚) 2
𝐸𝐷𝑉 (2𝑚𝑙) = 𝜋 × × ∑ 𝑟2𝑖 (𝑐𝑚)
20
𝑖=1
20
𝐿𝑉𝐿𝑠2𝑖 (𝑐𝑚) 2
𝐸𝑆𝑉 (2𝑚𝑙) = 𝜋 × × ∑ 𝑟2𝑖 (𝑐𝑚)
20
𝑖=1
LV volume (A4C):
20
𝐿𝑉𝐿𝑑4𝑖 (𝑐𝑚)
𝐸𝐷𝑉 (4𝑚𝑙) = 𝜋 × × ∑ 𝑟4𝑖2 (𝑐𝑚)
20
𝑖=1
20
𝐿𝑉𝐿𝑠4𝑖 (𝑐𝑚)
𝐸𝑆𝑉 (4𝑚𝑙) = 𝜋 × × ∑ 𝑟4𝑖2 (𝑐𝑚)
20
𝑖=1
Note: When using Simpson BP to measure LV function, be sure to keep the apical
four-chamber view and apical two-chamber view perpendicular. Otherwise the
measure result will be incorrect.
Cube
Measurement and
Description Method and Formula
study
End-diastolic Left FoldLine in 2D mode or Parallel method
Diastole
Ventricular Measurement in M mode
End-systolic Left FoldLine in 2D mode or Parallel method
Systole
Ventricular Measurement in M mode
Left Ventricular Internal Distance in 2D/M General
LVIDd
Diameter at End-diastole Measurements
Left Ventricular Internal Distance in 2D/M General
LVIDs
Diameter at End-systole Measurements
Obtained by ECG, input directly or
HR Heart Rate
measure manually
Interventricular Septal Distance in 2D/M General
IVSd
Thickness at End- diastole Measurements
Left Ventricular Posterior
Distance in 2D/M General
LVPWd Wall Thickness at
Measurements
End-diastole
End-diastolic Left 3
EDV(Cube) EDV(ml)= LVIDd(cm)
Ventricular Volume
114
8 Measurement
Measurement and
Description Method and Formula
study
End-systolic Left
ESV(Cube) ESV(ml)= LVIDs(cm)3
Ventricular Volume
Teichholz
Measurement and
Description Method and Formula
study
End-diastolic Left Ventricular FoldLine in 2D mode or Parallel
Diastole
Measurement method in M mode
End-systolic Left Ventricular FoldLine in 2D mode or Parallel
Systole
Measurement method in M mode
Left Ventricular Internal Distance in 2D/M General
LVIDd
Diameter at End-diastole Measurements
Left Ventricular Internal Distance in 2D/M General
LVIDs
Diameter at End-systole Measurements
Obtained by ECG, input directly or
HR Heart Rate
measure manually
Interventricular Septal Distance in 2D/M General
IVSd
Thickness at End-diastole Measurements
Left Ventricular Posterior Wall Distance in 2D/M General
LVPWd
Thickness at End-diastole Measurements
Interventricular Septal Distance in 2D/M General
IVSs
Thickness at End-systole Measurements
Left Ventricular Posterior Wall Distance in 2D/M General
LVPWs
Thickness at End-systole Measurements
End-diastolic Left Ventricular 3
EDV(ml)=(7×(LVIDd(cm)) )/(2.4+LVID
EDV(Teichholz)
Volume d (cm))
End-systolic Left Ventricular 3
ESV(ml)=(7×(LVIDs(cm)) )/(2.4+LVIDs
ESV(Teichholz)
Volume (cm))
115
8 Measurement
Measurement and
Description Method and Formula
study
Gibson
Measurement and
Description Method and Formula
study
End-diastolic Left Ventricular FoldLine in 2D mode or Parallel
Diastole
Measurement method in M mode
End-systolic Left Ventricular FoldLine in 2D mode or Parallel
Systole
Measurement method in M mode
Left Ventricular Internal Distance in 2D/M General
LVIDd
Diameter at End-diastole Measurements
Left Ventricular Internal Distance in 2D/M General
LVIDs
Diameter at End-systole Measurements
Heart Rate Obtained by ECG, input directly or
HR
measure manually
Interventricular Septal Distance in 2D/M General
IVSd
Thickness at End-diastole Measurements
Left Ventricular Posterior Wall Distance in 2D/M General
LVPWd
Thickness at End-diastole Measurements
Interventricular Septal Distance in 2D/M General
IVSs
Thickness at End-systole Measurements
Left Ventricular Posterior Wall Distance in 2D/M General
LVPWs
Thickness at End-systole Measurements
End-diastolic Left Ventricular EDV(ml)=π/6(098×LVIDd(cm)+5.90)×L
EDV(Gibson)
Volume VIDd(cm)²
End-systolic Left Ventricular ESV(ml)=π/6(1.14×LVIDs(cm)+4.18)×
ESV(Gibson)
Volume LVIDs(cm)²
116
8 Measurement
Measurement and
Description Method and Formula
study
LV Mass
Estimate the Index of Left Ventricular Mass (LV Mass-I) by calculating the LV Mass.
LV Mass (Cube)
Measurement and
Description Method and Formula
study
Interventricular Septal Distance in 2D/M General
IVSd
Thickness at End-diastole Measurements
Left Ventricular Internal Distance in 2D/M General
LVIDd
Diameter at End-diastole Measurements
Left Ventricular Posterior
Distance in 2D/M General
LVPWd Wall Thickness at
Measurements
End-diastole
*1 Means:
𝐿𝑉 𝑀𝑎𝑠𝑠(𝑔) = 1.04 × ((𝐿𝑉𝑃𝑊𝑑(𝑐𝑚) + 𝐼𝑉𝑆𝑑(𝑐𝑚) + 𝐿𝑉𝐼𝐷𝑑(𝑐𝑚))3 − 𝐿𝑉𝐼𝑑(𝑐𝑚)3 ) − 13.6
*2 Means:
LV Mass-I (No unit) = LV Mass (g) / Body Surface Area (m2)
LV Mass (A-L)
Measurement and
Description Method and Formula
study
Left Ventricular Epicardial
Distance in 2D/M General
LVAd sax Epi Area at Papillary Muscle level
Measurements
at end-diastole in Short-axis
117
8 Measurement
view
*1 Means:
LV Mass(g) 1.055/6 ( LVAd sax Epi(cm2 ) (LVLd apical(cm) t(cm))- LVAd sax Endo
(cm2 ) LVL(cm));
𝐿𝑉𝐴𝑑𝑠𝑎𝑥𝐸𝑝𝑖(𝑐𝑚2 ) (𝐿𝑉𝐴𝑑𝑠𝑎𝑥𝐸𝑛𝑑𝑜(𝑐𝑚2 )
𝑡(𝑐𝑚) = √( ) − √( )
𝜋 𝜋
*2 Means:
LV Mass-I (No unit) = LV Mass (g) / Body Surface Area (m2)
LV Mass (T-E)
Measurement and
Description Method and Formula
study
Left Ventricular Epicardial Area
LVAd sax Epi at Papillary Muscle level at Area in 2D General Measurements
end-diastole in Short-axis view
Left Ventricular Endocardial
LVAd sax Endo Area at Papillary Muscle level at Area in 2D General Measurements
end-diastole in Short-axis view
Semi-major axis from widest Distance in 2D General
a
minor axis radius to apex Measurements
Truncated semi-major axis
Distance in 2D General
b from widest minor axis radius to
Measurements
mitral annulus plane
*1 Means:
2(𝑎 + 𝑡) 𝑑3 2𝑎 𝑑3
𝐿𝑉 𝑀𝑎𝑠𝑠(𝑔) = 1.05𝜋 × {(𝑏 + 𝑡)2 × [ +𝑑− ] − 𝑏 2
× ( + 𝑑 − )}
3 3(𝑎 + 𝑡)2 3 3𝑎2
118
8 Measurement
𝐿𝑉𝐴𝑑𝑠𝑎𝑥𝐸𝑝𝑖(𝑐𝑚2 ) (𝐿𝑉𝐴𝑑𝑠𝑎𝑥𝐸𝑛𝑑𝑜(𝑐𝑚2 )
𝑡(𝑐𝑚) = √( ) − √( )
𝜋 𝜋
𝐿𝑉𝐴𝑑𝑠𝑎𝑥𝐸𝑛𝑑𝑜(𝑐𝑚2 )
b(cm) = √
𝜋
*2 Means:
LV Mass-I (No unit) = LV Mass (g) / Body Surface Area (m2)
Mitral Valve Area (MVA)
Mitral Valve Area (MVA) can be calculated by two methods: pressure half time (PHT) or
velocity- time integral (VTI).
MVA (VTI)
Measurement and
Description Method and Formula
study
Left Ventricular Outflow Tract Distance in 2D General
LVOT Diam
Diameter Measurements
Left Ventricular Outflow Tract D trace in Doppler General
LVOT VTI
Velocity-Time Integral Measurement
Mitral Valve Velocity-Time D trace in Doppler General
MV VTI
Integral Measurement
*1 Means:
𝜋 × |𝐿𝑉𝑂𝑇 𝑉𝑇𝐼(𝑐𝑚)| × 𝐿𝑉𝑂𝑇 𝐷𝑖𝑎𝑚(𝑐𝑚2 )2
𝑀𝑉𝐴(𝑉𝑇𝐼)(𝑐𝑚2 ) =
4 × |𝑀𝑉 𝑉𝑇𝐼(𝑐𝑚)|
AVA (VTI)
Aortic Valve Area (AVA) can be calculated by velocity-time integral (VTI).
Measurement and
Description Method and Formula
study
Left Ventricular Outflow Tract Distance in 2D General
LVOT Diam
Diameter Measurements
Left Ventricular Outflow Tract D trace in Doppler General
LVOT VTI
Velocity-Time Integral measurements
Aortic Valve Velocity-Time D trace in Doppler General
AV VTI
Integral measurements
*1 Means:
𝜋 × |𝐿𝑉𝑂𝑇 𝑉𝑇𝐼(𝑐𝑚)| × 𝐿𝑉𝑂𝑇 𝐷𝑖𝑎𝑚(𝑐𝑚2 )2
𝐴𝑉𝐴(𝑉𝑇𝐼)(𝑐𝑚2 ) =
4 × |𝐴𝑉 𝑉𝑇𝐼(𝑐𝑚)|
LA Vol
LA Vol (A-L)
119
8 Measurement
Measurement and
Description Method and Formula
study
*1 Means:
8𝜋
𝐿𝐴 𝑉𝑜𝑙(𝐴 − 𝐿)(𝑚𝑙) = 𝐿𝐴𝐴(𝐴4𝐶)(𝑐𝑚2 ) × 𝐿𝐴𝐴(𝐴2𝐶)(𝑐𝑚2 )/𝐿𝐴 𝐷𝑖𝑎𝑚(𝑐𝑚)
3
LA Vol (Simp)
Measurement and
Description Method and Formula
study
Left Atrium Volume at apical
LA Vol(A2C) Same as Simpson SP measurement
2-chamber view
Left Atrium Volume at apical
LA Vol(A4C) Same as Simpson SP measurement
4-chamber view
RA Vol (Simp)
Measurement and
Description Method and Formula
study
Right Atrium Volume at apical
RA Vol(A4C) Same as Simpson SP measurement
4-chamber view
LVIMP
Measurement and
Description Method and Formula
study
Time in M/Doppler General
MV C-O dur Mitral Valve close-open Duration
Measurements
Time in M/Doppler General
LVET Left Ventricular Ejection Time
Measurements
Left Ventricular Index of
LVIMP *1
Myocardial Performance
*1 Means:
𝑀𝑉 𝐶 − 𝑂 𝑑𝑢𝑟(𝑠) − 𝐿𝑉𝐸𝑇(𝑠)
𝐿𝑉𝐼𝑀𝑃(𝑛𝑜 𝑢𝑛𝑖𝑡) =
𝐿𝑉𝐸𝑇(𝑠)
RVSP
Measurement and
Description Method and Formula
study
120
8 Measurement
Measurement and
Description Method and Formula
study
Maximum Velocity Measurements
*1 Means:
𝑇𝑅 𝑃𝐺𝑚𝑎𝑥(𝑚𝑚𝐻𝑔) = 4 × 𝑇𝑅 𝑉𝑚𝑎𝑥 (𝑚/𝑠)2
*2 Means:
𝑇𝑉 𝐶 − 𝑂 𝑑𝑢𝑟(𝑠) − 𝑅𝑉𝐸𝑇(𝑠)
𝑅𝑉𝐼𝑀𝑃(𝑛𝑜 𝑢𝑛𝑖𝑡) =
𝑅𝑉𝐸𝑇(𝑠)
PAEDP
Measurement and
Description Method and Formula
study
Pulmonary Valve Regurgitation D Vel in Doppler General
PR Ved
Velocity at end-Diastole Measurements
*1 Means:
RVSP(mmHg) RAP(mmHg ) 4 (TR V max(m / s))2
RV Tei index(RVIMP)
Measurement and
Description Method and Formula
study
*1 Means:
𝑇𝑉 𝐶 − 𝑂 𝑑𝑢𝑟(𝑠) − 𝑅𝑉𝐸𝑇(𝑠)
𝑅𝑉𝐼𝑀𝑃(𝑛𝑜 𝑢𝑛𝑖𝑡) =
𝑅𝑉𝐸𝑇(𝑠)
121
8 Measurement
LV Tei index(LVIMP)
Measurement and
Description Method and Formula
study
Time in Doppler General
MV C-O dur Mitral Valve close-open Duration
Measurements
Time in Doppler General
LVET Left Ventricular Ejection Time
Measurements
Left Ventricular Index of
LVIMP *1
Myocardial Performance
*1 Means:
𝑀𝑉 𝐶 − 𝑂 𝑑𝑢𝑟(𝑠) − 𝐿𝑉𝐸𝑇(𝑠)
𝐿𝑉𝐼𝑀𝑃(𝑛𝑜 𝑢𝑛𝑖𝑡) =
𝐿𝑉𝐸𝑇(𝑠)
PISA
PISA (Proximal Isovelocity Surface Area) is used in quantitative analysis of the mitral valve
regurgitation (PISA MR), aortic valve regurgitation (PISA AR), tricuspid valve regurgitation
(PISA TR), and pulmonary valve regurgitation (PISA PR) in color mode.
PISA MR
Measurement and
Description Method and Formula
study
*1 Means:
2𝜋𝑀𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝑀𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
𝑀𝑅 𝐹𝑙𝑜𝑤(𝑚𝑙) = × |𝑀𝑅 𝑉𝑇𝐼(𝑐𝑚)|
|𝑀𝑅𝑉 𝑚𝑎𝑥(𝑐𝑚/𝑠)|
*2 Means:
122
8 Measurement
Measurement and
Description Method and Formula
study
*1 Means:
2𝜋𝐴𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝐴𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
AR Flow(ml) = × |𝐴𝑅 𝑉𝑇𝐼(𝑐𝑚)|
|𝐴𝑅𝑉 max(𝑐𝑚/𝑠)|
*2 Means:
AR Flow Rate(ml/s) = 2π AR Rad(𝑐𝑚)2 × 𝐴𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
*3 Means:
𝐴𝑅 𝐹𝑙𝑜𝑤(𝑚𝑙)
AR Fraction (no unit) = × 100%
𝐴𝑉 𝑆𝑉(𝑚𝑙)
*4 Means:
2𝜋𝐴𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝐴𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
AR EROA(cm)2 =
|𝐴𝑅 𝑉𝑚𝑎𝑥(𝑐𝑚/𝑠)|
123
8 Measurement
PISA TR
Measurement and
Description Method and Formula
study
Tricuspid Valve Stenosis
TR Rad PISA measurement
Radius
Tricuspid Valve Regurgitation D Trace in Doppler General
TR VTI
Velocity-Time Integral Measurements
You can select to use top aliasing
Tricuspid Valve Regurgitation
TR Als.Vel velocity or bottom aliasing velocity or
Aliasing Maximum Velocity
input the value directly.
Tricuspid Regurgitation
TR Vmax Obtained from TR VTI measurement
Maximum Velocity
*1 Means:
2𝜋𝑇𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝑇𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
𝑇𝑅 𝐹𝑙𝑜𝑤(𝑚𝑙) = × |𝑇𝑅 𝑉𝑇𝐼(𝑐𝑚)|
|𝑇𝑅𝑉 𝑚𝑎𝑥(𝑐𝑚/𝑠)|
*2 Means:
𝑇𝑅 𝐹𝑙𝑜𝑤 𝑅𝑎𝑡𝑒(𝑚𝑙/𝑠) = 2𝜋 𝑇𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝑇𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
*3 Means:
𝑇𝑅 𝐹𝑙𝑜𝑤(𝑚𝑙)
𝑇𝑅 𝐹𝑟𝑎𝑐𝑡𝑖𝑜𝑛 (𝑛𝑜 𝑢𝑛𝑖𝑡) = × 100%
𝑇𝑉 𝑆𝑉(𝑚𝑙)
*4 Means:
2𝜋𝑇𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝑇𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
𝑇𝑅 𝐸𝑅𝑂𝐴(𝑐𝑚)2 =
|𝑇𝑅 𝑉𝑚𝑎𝑥(𝑐𝑚/𝑠)|
PISA PR
Measurement and
Description Method and Formula
study
124
8 Measurement
Measurement and
Description Method and Formula
study
Pulmonary Valve Stenosis
PR Rad PISA measurement
Radius
Pulmonary Valve Regurgitation D Trace in Doppler General
PR VTI
Velocity-Time Integral Measurements
You can select to use top aliasing
Pulmonary Valve Regurgitation
PR Als.Vel velocity or bottom aliasing velocity
Aliasing Maximum Velocity
or input the value directly.
Pulmonary Regurgitation Obtained from PR VTI
PR Vmax
Maximum Velocity measurement
*1 Means:
2𝜋𝑃𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝑃𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
𝑃𝑅 𝐹𝑙𝑜𝑤(𝑚𝑙) = × |𝑃𝑅 𝑉𝑇𝐼(𝑐𝑚)|
|𝑃𝑅𝑉 𝑚𝑎𝑥(𝑐𝑚/𝑠)|
*2 Means:
𝑃𝑅 𝐹𝑙𝑜𝑤 𝑅𝑎𝑡𝑒(𝑚𝑙/𝑠) = 2𝜋 𝑃𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝑃𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
*3 Means:
𝑃𝑅 𝐹𝑙𝑜𝑤(𝑚𝑙)
𝑃𝑅 𝐹𝑟𝑎𝑐𝑡𝑖𝑜𝑛 (𝑛𝑜 𝑢𝑛𝑖𝑡) = × 100%
𝑃𝑉 𝑆𝑉(𝑚𝑙)
*4 Means:
2𝜋𝑃𝑅 𝑅𝑎𝑑(𝑐𝑚)2 × 𝑃𝑅 𝐴𝑙𝑠. 𝑉𝑒𝑙(𝑐𝑚/𝑠)
𝑃𝑅 𝐸𝑅𝑂𝐴(𝑐𝑚)2 =
|𝑃𝑅 𝑉𝑚𝑎𝑥(𝑐𝑚/𝑠)|
TDI
Measurement and
Description Method and Formula
study
𝐸𝑎(𝑚𝑒𝑑𝑖𝑎𝑙)
Ea/Aa(medial) MV medial E-Vel/ A-Vel 𝐸𝑎 ⁄𝐴𝑎(𝑚𝑒𝑑𝑖𝑎𝑙)(𝑛𝑜 𝑢𝑛𝑖𝑡) =
𝐴𝑎(𝑚𝑒𝑑𝑖𝑎𝑙)
MV medial E-wave
ATa(medial) Obtained from ARa(medial) measurements
Acceleration Time
MV medial E-wave
DTa(medial) Obtained from DRa(medial) measurements
Deceleration Time
𝐸𝑎(𝑙𝑎𝑡𝑒𝑟𝑎𝑙)
Ea/Aa(lateral) MV lateral E-Vel/ A-Vel 𝐸𝑎 ⁄𝐴𝑎(𝑙𝑎𝑡𝑒𝑟𝑎𝑙)(𝑛𝑜 𝑢𝑛𝑖𝑡) =
𝐴𝑎(𝑙𝑎𝑡𝑒𝑟𝑎𝑙)
125
8 Measurement
Measurement and
Description Method and Formula
study
MV lateral E-wave
ATa(lateral) Obtained from ARa(lateral) measurement
Acceleration Time
MV lateral E-wave
DTa(lateral) Obtained from DRa(lateral) measurement
Deceleration Time
8.3.4 Vascular
126
8 Measurement
127
8 Measurement
IMT
IMT (Intima-Media Thickness) refers to the vertical dimension between LI (Lumen-Intima)
and MA (Media-Adventia).
Perform the following steps:
1. In 2D mode , press the [Caliper] button on the control panel to start
measurement.
2. Click [IMT] in the measurement menu and the ROI box appears.
3. Press the [Set] button and move the trackball together to adjust the vertex
position of the ROI box.
4. Roll the trackball to adjust the ROI area and then press the [Set] button. Two
auto trace lines appear in the ROI box. Results are displayed in the result
window.
NOTE: Please freeze the image before starting IMT measurement.
MT measurement is available for linear array image only.
Make sure that you select the right vessel wall (Near/ Far) before
IMT measurement; otherwise wrong calculation may result.
Stenosis D
Measure Normal Diam and Resid Diam, calculates Stenosis D.
𝑆𝑡𝑒𝑛𝑜𝑠𝑖𝑠 𝐷 = (𝐷1 − 𝐷2 )⁄𝐷1 × 100%
Where D1 is Normal Diam and D2 is Resid Diam.
128
8 Measurement
Stenosis A
Measure Normal Area and Resid Area, calculates Stenosis A.
𝑆𝑡𝑒𝑛𝑜𝑠𝑖𝑠 𝐴 = (𝐴1 − 𝐴2 )/𝐴1 × 100%
Where A1 is Normal Area and A2 is Resid Area.
ICA/CCA(PS)
Measure the flow velocity ratio between ICA and CCA to calculate the stenosis.
Measure PS value of ICA and CCA distal by D trace method of Doppler general
measurement, and the system automatically calculates the stenosis.
ABI
Measure the Ankle Systolic Pressure (ASP) and Brachial Systolic Pressure (BSP) on
Doppler image to calculate Ankle Brachial Index (ABI)
Formula: ABI = ASP/BSP
8.3.5 Gynecology
The measurement items, calculation items and measurement method for 2D gynecology
measurement are shown in the table below.
Table 38 2D Gynecology Measurement
Items Description Methods or Formula
UT Vol UT Volume
See “8.3.5.2 Gynecology Study
and Calculation”
Uterus Body /
129
8 Measurement
Follicle Volume /
See “8.3.5.2 Gynecology Study
and Calculation”
Follicle Average Diameter /
Ovary Vol Measure the Ovary L, Ovary H and Ovary W, calculates Ovary Vol
130
8 Measurement
Preset the formula for follicle calculation, refer to “4.2.4 Application” for details.
8.3.6 Urology
The measurement items, calculation items and measurement method for 2D Urology
measurement are shown in the table below.
Table 41 2D Urology Measurements
Items Description Methods or Formula
131
8 Measurement
The measurement items, calculation items and measurement method for Doppler Urology
measurement are shown in the table below.
Table 42 Doppler Urology Measurement
Items Description Methods or Formula
132
8 Measurement
The measurement items, calculation items and measurement method for 2D small parts
measurement are shown in the table below.
Table 44 2D Small Parts Measurement
Items Description Methods or Formula
133
8 Measurement
The measurement items, calculation items and measurement method for Doppler small
parts measurement are shown in the table below.
8.3.8 Orthopedics
134
8 Measurement
and Calculation”
HIP
HIP (Hip Joint Angle) measurement is used in pediatric orthopedics to provide early
diagnosis for infant hip joint dislocation. In this measurement, three straight lines need to
be set in turn, which are the baseline (BL), roof line (RL), and inclination line (IL). As well
as two angles, which are α (the angle between BL and RL) and β (the angle between BL
and IL)
Perform the following steps:
1. Start application measurement.
2. Select [Side] by pressing [ ] or [ ] button and then rotate the [Value] knob to
set the HIP orientation.
3. Click [HIP] from the measurement menu.
4. The baseline (BL) appears. Use the trackball to move the line to the position of
the hip joint. Select [Angle] by pressing [ ] or [ ] button and then rotate the
[Value] knob to adjust the baseline angle and then press the [Set] button to fix
the baseline. The roof line (RL) appears.
5. Repeat step 4 to adjust the roof line (RL) and inclination line (IL). The system
automatically calculates the hip joint α and hip joint β and displays them in the
result window.
6. Enter patient age and dislocation type will be displayed together.
7. Select [Hip (α)] or [Hip (β)] to measure the angles of α and β respectively.
HIP-Graf
The measure method is the same as “HIP”.
Dislocation type can be determined through Graf method, as shown in the following table.
Table 47 Graf Method
Dislocation Criteria
Result
Type α β Patient
135
8 Measurement
d/D
Measure the distance between baseline and bottom line of the osseous acetabular and
maximum width of hip to estimate the hip osseous acetabular coverage.
Perform the following steps:
1. Start application measurement.
2. Select [Side] by pressing [ ] or [ ] button and then rotate the [Value] knob to
select the HIP orientation.
3. Click [d/D] from the measurement menu and the cursor appears.
4. Move the trackball to measure the maximum width of HIP (D) and the distance
between the baseline and bottom line of the osseous acetabular (d). The system
automatically calculates d / D and displays the result in the result window.
8.3.9 Emergency
The emergency measurement package contains commonly used measurement items
under the corresponding mode.
For the measurement method, please refer to the measurement section of each part.
136
9 Report
9 Report
After the measurement is completed, the relevant measurement data will be displayed in
the report.
Previous/ Click [Previous] or [Next] to flip the pages if the report has more than one
Next page
137
9 Report
Item Description
Save Click to save current setting and return to the Home screen
Cancel Click to cancel all modifications and return to the Home screen
138
9 Report
139
9 Report
9.5.1 Load
Click [Load Report] to download existed report.
9.5.2 Export
Click [Export] to open the Export Wizard and set relevant parameters, click [OK] to export
the current report.
140
10 Patient Data Management
10.1 Storage
Image and cine are saved to the system by default. You can also choose to save the
image and cine to a USB storage device.
Export your data to a removable device or DICOM server is available if needed. Refer to
“10.5 Exporting Image/Data” for details.
10.1.1Image Storage
In freeze or real-time status, press the [Save Img] button on the control panel, the system
will automatically save the current image to the current patient database. You can also
enter the review screen to view the image and cine by pressing the [File] button on the
control panel.
The system saves the full-screen image by default (including parameters, comments,
measurement items, etc.), and you can also set to save only the image area. For details,
refer to “4.2.3 Image”.
10.1.2Cine Storage
In the freeze or real-time status, press the [Save Cine] button on the control panel, the
system will automatically save the cine to the current patient database.
10.2.1Current Patient
Image and cine of current patient can be viewed through the Review screen and Station
screen.
1. Press the [File] button on the control panel to enter the Review screen of current
patient.
141
10 Patient Data Management
1. Press the [File] button on the control panel to enter the Review screen and then
click [Station] to enter the Station screen. The current patient is selected by
default in the patient information list, and the clipboard displays the thumbnail
image of the patient.
10.2.2Previous Patient
To view data of previous patient through Station screen:
142
10 Patient Data Management
1. Press the [File] button on the control panel to enter the Review screen and then
click [Station] to enter the Station screen.
2. Search the target patient.
Select the data source and search item, enter the keyword to search the target patient.
Drop down the scroll bar on the right side of the patient information list to locate the
target patient.
3. Select the target patient and press the [Set] button to open the operation menu
and perform operations as needed.
Open the operation menu and click [Patient Info] to enter the Patient screen to check
basic information and exam information of current patient.
10.2.2.2 Image
10.2.2.3 Report
Open the operation menu and click [Review Report] to view the report.
10.3.1Certain Patient
To delete certain patient image/data, please perform the following steps:
1. Press the [File] button on the control panel to enter the Review screen and then
click [Station] to enter the Station screen.
2. Search desired patient
Select the data source and search item, enter the keyword to search the desired
patient.
Roll the trackball over the desired patient, press the [Set] button.
143
10 Patient Data Management
Delete image: move the cursor to the image to be deleted, click the delete icon
to delete the image; or open the operation menu and then select [Review
Delete patient data: open the operation menu and select [Delete Exam], and then click
10.3.2All Patients
To delete all patients’ image/data, please perform the following steps:
1. Press the [File] button on the control panel to enter the Review screen and then
click [Station] to enter the Station screen.
2. Click [Select All] to select all patients.
3. Open the operation menu, select [Delete Exam], and then click [OK] in the
pop-up dialog box to delete the patient data.
2. Scroll the right slider bar to select desire patient image or data.
NOTE : The data in this system is exported to USB storage devices in image
format with different resolution, so the exported image is not the same as
144
10 Patient Data Management
Select the desired patient: search the patient information via the search items in the
patient list.
Select the cine zoom mode, cine compression mode, and cine compression
ratio.
Select [Delete after Exporting], and then click [Delete Exams] to delete patient data.
Select [Delete after Exporting], and then click [Delete Images] to delete patient
image.
Move the cursor over the status bar icon , press the [Set] button, and click [Remove]
in the pop-up menu to safely remove the USB storage device.
145
11 DICOM
11 DICOM
Digital Imaging and Communications in Medicine (DICOM) Standard is an international
normative protocol for communication of medical digital images. DICOM service enables
you to:
Search for patient information archived on the DICOM server (working list) and
copy the patient information to the device for reviewing.
Send patient information and images on the system to the DICOM storage
server.
Send the image saved on the ultrasound system to the DICOM print server for
image printing.
NOTE : The DICOM Server is purchasable. Please confirm the function is purchased
and installed before use.
146
11 DICOM
Item Description
Ping Check network connectivity between the system and DICOM server.
DICOM Service Set the DICOM servers. For details, refer to “11.1.2 Service Preset”.
11.1.2Service Preset
The service preset is mainly used to add or delete DICOM services in a DICOM server, set
the service name, port address, and so on.
DICOM service consists of Storage, Print, Worklist, MPPS, Storage Commitment, and
Query/Retrieve.
11.1.2.1 Storage
Storage server connection properties and image storage format settings can be found in
147
11 DICOM
Enter the AE title. The AE Title here should be consistent with that of
AE Title
the storage server.
Interval Time (Sec) Set the interval time for connection retry.
Refers to the amount of time after which the system will stop trying to
Timeout (Sec)
establish a connection to the service.
Max Frame Rate Set the frame range of transferring cin file into DCM multiframe file.
148
11 DICOM
Item Description
Select a server in the service list, change the parameters and click
Update
[Update] to update the parameters of the server.
Select an item in the server list as the default server, and the default
Default
server is marked with a “Y”.
Verify the network connectivity between the system and storage
Verify
server.
11.1.2.2 Print
Service and print attributes, and print server connection can be found in print service
screen. Print server setting is similar to storage server setting, refer to “11.1.2.1 Storage”
for details.
Enter the print service preset screen via [Preset] → [Network Preset] → [DICOM
Service] → [Print].
149
11 DICOM
Enter the AE title. The AE Title here should be consistent with that of
AE Title
the worklist server.
Interval Time (Sec) Set the interval time for connection retry.
Refers to the amount of time after which the system will stop trying to
Timeout (Sec)
establish a connection to the service.
Trim Set whether to print the trim box around each image of the film.
150
11 DICOM
Item Description
Select a server in the service list, change the parameters and click
Update
[Update] to update the parameters of the server.
Select an item in the server list as the default server, and the default
Default
server is marked with a“Y”.
Verify the network connectivity between the system and the print
Verify
server.
Set the print attribute related parameters based on the actual printer requirements, for
example:
If the printer does not support [Film Size] as 8N*10N, do not set this size to the
printer film size.
As to [Medium Type], Blue Film or Clear Film is usually used for black and white
printing, and Paper is usually used for color printing. For details, please refer to
the printer conformance statement.
Other parameters can be reminded as default. If the printer has special
requirements, please change the parameters according to the requirements of
the printer.
11.1.2.3 Worklist
Worklist server connection properties and relevant settings can be found in the worklist
service preset screen.Worklist server setting is similar to storage server setting, refer to
“11.1.2.1 Storage” for details.
Enter the worklist service preset screen via [Preset] → [Network Preset] → [DICOM
Service] → [Worklist].
151
11 DICOM
Enter the AE title. The AE Title here should be consistent with that of
AE Title
the worklist server.
Interval Time (Sec) Set the interval time for connection retry.
Refers to the amount of time after which the system will stop trying to
Timeout (Sec)
establish a connection to the service.
Select a server in the service list, change the parameters and click
Update
[Update] to update the parameters of the server.
Select an item in the server list as the default server, and the default
Default
server is marked with a “Y”.
Verify the network connectivity between the system and the worklist
Verify
server.
152
11 DICOM
11.1.2.4 MPPS
MPPS server connection properties and relevant settings can be found in the MPPS
service preset screen. MPPS server setting is similar to storage server setting, refer to
“11.1.2.1 Storage” for details.
Enter the MPPS service preset screen via [Preset] → [Network Preset] → [DICOM
Service] → [MPPS].
NOTE : When using MPPS, you need to set the MPPS server as the default server.
Enter the AE title. The AE Title here should be consistent with that of
AE Title
the mpps server.
Interval Time (Sec) Set the interval time for connection retry.
Refers to the amount of time after which the system will stop trying to
Timeout (Sec)
establish a connection to the service.
Select a server in the service list, change the parameters and click
Update
[Update] to update the parameters of the server.
153
11 DICOM
Item Description
Verify the network connectivity between the system and the mpps
Verify
server.
Storage Commitment server connection properties and relevant settings can be found in
the storage commitment service preset screen. Storage commitment server setting is
similar to storage server setting, refer to “11.1.2.1 Storage” for details.
Enter the MPPS service preset screen via [Preset] → [Network Preset] → [DICOM
Service] → [Storage Commitment].
Enter the AE title. The AE Title here should be consistent with that of
AE Title
the storage commitment server.
Port Enter the port number for the storage commitment server.
Interval Time (Sec) Set the interval time for connection retry.
Refers to the amount of time after which the system will stop trying to
Timeout (Sec)
establish a connection to the service.
Associated Storage The associated storage server should be preset before storage
Service commitment, only after the exam is sent out, can storage
154
11 DICOM
Item Description
commitment be created.
Select a server in the service list, change the parameters and click
Update
[Update] to update the parameters of the server.
Select an item in the server list as the default server, and the default
Default
server is marked with a “Y”.
Verify the network connectivity between the system and storage
Verify
commitment server.
Query/Retrieve server connection properties and relevant settings can be found in the
Query/Retrieve service preset screen. Query/Retrieve server setting is similar to storage
server setting, refer to “11.1.2.1 Storage” for details.
Enter the Query/Retrieve service preset screen via [Preset] → [Network Preset] →
[DICOM Service] → [Query/Retrieve].
155
11 DICOM
Item Description
Enter the AE title. The AE Title here should be consistent with that of
AE Title
the query/retrieve server.
Interval Time (Sec) Set the interval time for connection retry.
Refers to the amount of time after which the system will stop trying to
Timeout (Sec)
establish a connection to the service.
Select a server in the service list, change the parameters and click
Update
[Update] to update the parameters of the server.
Select an item in the server list as the default server, and the default
Default
server is marked with a“Y”.
Verify the network connectivity between the system and
Verify
query/retrieve server.
NOTE : Before using DICOM services, please perform verification to ensure that the
system is properly connected to the DICOM server.
11.2.1DICOM Storage
DICOM storage service sends patient image or data to storage server for remote storage.
Perform the following steps:
1. Save images. Refer to “10.1 Storage” for details.
2. Send images/data to the DICOM storage server via following screens:
Press [File] button on the control panel to enter the Review screen. Move the cursor
over the thumbnail you wan to send, click icon on the image or click [Send] to
Press [File] button on the control panel to enter the Review screen, and then click
[Station] to enter the Station screen. Select desire patient, open the operation menu
156
11 DICOM
and click [Send Exam] or move the cursor over the thumbnail you wan to send, click
3. A sending dialog box pops up. Select DICOM and storage server as needed.
4. Click [OK] to send the image/ data to the storage server.
11.2.2DICOM Print
DICOM Print service sends patient image to print server for remote printing. Preset the
print server before using DICOM Print service.
Perform the following steps:
1. Save images. Refer to “10.1 Storage” for details.
2. Send image to DICOM Print Server via following screens
Press [File] button on the control panel to enter the Review screen. Move the cursor
over the thumbnail you wan to send, click icon on the image or click [Send] to
send the selected images.
Press [File] button on the control panel to enter the Review screen, and then click
[Station] to enter the Station screen. Select desire patient, open the operation menu
and click [Send Exam] or move the cursor over the thumbnail you wan to send, click
3. A sending dialog box pops up. Select DICOM and print server as needed.
4. Click [OK] to send the image/ data to the print server.
11.2.3DICOM Worklist
The DICOM Worklist Service is used to query or retrieve patient information stored in the
server, which can be imported into the ultrasound system for viewing and performing new
exam. Preset the DICOM Worklist server before using DICOM Worklist service.
To retrieve Patient information (DICOM), perform the following steps:
1. Press the [Patient] button on the control panel to enter the patient information
screen. Click [Worklist] to display Worklist screen.
157
11 DICOM
Set search condition among patient ID, patient name, accession number,
keyword, scheduled station AE Title, device type, exam date, worklist server
and other information to search for patients.
Click [Query] to display patient information which meets the search conditions
in the lower part of the screen.
[Start Exam]: click to return to the Home screen and start a new exam for the selected
patient.
[Transfer]: click to import the selected patient information into the [Patient
Information] screen for editing. After editing, click [Save] to start a new exam.
[Cancel]: click to exit worklist screen and return to the Patient Information screen.
11.2.4DICOM MPPS
MPPS is used to send patient exam status to the MPPS server, which facilitates the other
systems to obtain the exam progress in time. Preset the DICOM MPPS server before
using DICOM MPPS service.
When MPPS preset is done, the patient exam status will be sent to the MPPS server as
follows:
When start a new exam, the ultrasound system sends an “N-CREATE” message
to the MPPS server.
After the normal image acquisition completes, the ultrasound system sends an
“N-SET COMPLETED” message to the MPPS server.
When the image acquisition is abnormally terminated, the ultrasound system
sends “N-SET DISCONTINUED” to the MPPS server.
158
11 DICOM
Send images automatically to the DICOM storage server and perform storage
commitment when the exam is done.
Set the auto function via [Preset] → [System Preset] → [General] → [Send after
Exam].
When the exam completes, the system prompts whether the exam information was
successfully sent to the DICOM storage server.
5. Click [Query] and the qualified patients are displayed in the patient list.
159
11 DICOM
If the list is more than one page, roll the right slider bar of the list to find the patient
information.
Move the trackball and press the <Ctrl> + [Set] button to select more than one
patients.
Click [Cancel All] to cancel several selected or all patients selected in the list.
6. Click [Retrieve], the system will retrieve the selected patient record to the target
server or local, a progress bar displays and the patient information displays in the
patient list after completion. Or click [Exit] to stop retrieving the patient record.
7. After completing retrieve, click [Exit] to see the retrieved patient records in the
Station screen.
2. After the setting is done, when the image is sent to the DICOM storage server,
the structured report is also sent together.
2. After completing the exam, press the [Patient ] button on the control panel to
enther the Patient Information screen, click [End the Exam].
3. Press [File] button on the control panel to enter the Review screen, and then
click [Station] to enter the Station screen. Select target patient in the patient list.
4. Click [Send Exam] to pop up the sending exam dialog box.
160
11 DICOM
5. Select DICOM and storage server as needed and click [OK] to confirm.
Structured reports cannot be exported separately and can only be exported with patient
data. When export patient data, if there is a structured report, the structured report will be
exported to the external storage device as well.
Perform the following steps:
1. After completing the exam, press the [Patient] button on the control panel to
enther the Patient Information screen, click [End the Exam].
2. Press [File] button on the control panel to enter the Review screen, and then
click [Station] to enter the Station screen.
3. Select target patient and open the operation menu, click [Send Exam] to pop up
the sending exam dialog box.
4. Select [Disk], check [Export Report] as needed and click [OK] to start
exporting.
161
12 Probe and Biopsy
To ensure the performance and safety of the probe and the biopsy bracket, regular
cleaning, disinfection, and maintenance shall be taken.
12.1 Probe
This section mainly introduces the general operation of the probe. For details, please refer
to the instruction manual of each probe.
12.1.1Structure of Probe
The ultrasonic probes suitable for this system are all electronic scanning probes, which
generates high-resolution images by transmitting ultrasonic waves into the human body
and receiving echoes. The system also supports surface probes and intracavitary probe.
The basic structure of the probe is shown in the figure below. The orientation mark on the
probe is consistent with the direction indicated by the probe mark “ ” on the ultrasonic
image.
12.1.2Option Probes
Imaging probes used for exam may affect the amount of radiation in the system.
The following probes have been tested in accordance with the requirements of the CISPR
11 and confirmed to meet the radiation standards of Group 1 Class A products.
162
12 Probe and Biopsy
TC50, TC10, TL40 probes all support B-mode, B + B-mode, 4B-mode, B / M mode, M
mode, Color mode, PD mode, PW mode and B + Color + PW mode; TP16 supports CW
mode in addition to the above modes.
12.1.3Probe Usage
DO NOT use any probe not provided by the manufacturer, otherwise
the host and probe will be damaged. In serious cases, fire and other
accidents may result.
Use the probe with care. Stop use immediately if the probe appears
to be malfunctioning. Otherwise, electric shock and burns may
result.
Carefully inspect the probe sheath for damage or breakage before
using. Do not use a protective cover with holes or cracks to prevent
infection.
DO NOT use mineral oils, oily coupling agents, gels with any type of
detergent or softener, and other unapproved materials to replace
coupling agent supplied or recommended by the manufacturer.
Otherwise, probe damage may result.
1. Check the probe. The probe may have potential damage during use and cleaning.
Therefore, the following checks should be performed on the probe before use to
ensure:
163
12 Probe and Biopsy
No cable damage
2. Connect the probe to the ultrasound system. Refer to “3.6.1 Connect the Probe”
for details.
3. A protective probe sheath should be selected when using an intracavity probe
and the proper sterility method should be used throughout the operating
procedure.
Remove the package, take out and unfold the probe sheath.
Fit the sheath by hand and ensure the cover is free from creases and bubbles.
12.1.4Probe Cleaning
After completing each exam, the probe and reusable parts should be thoroughly
disinfected or sterilized as required.
Prior to clean the probe, disconnect the probe from the ultrasonic
system to avoid electric shock.
When clean the probe, avoid impacting the probe with sharp or hard
objects.
Always wear sterile gloves and to prevent infection.
164
12 Probe and Biopsy
12.1.5Probe Disinfection
After each exam, the probe and reusable parts should be thoroughly disinfected or
sterilized as required.
The disinfection category can be divided into the following two categories based on the
application of the probe.
Table 57 Disinfection Category
low level disinfection Only contact with intact skin Surface probe
Never immerse the probe deeper than the line shows below for sterilization.
165
12 Probe and Biopsy
Hydrogen
ResertXLHLD STERIS 2.0% Wipe/soak <8min
Peroxide
Quaternary
T-spray Ⅱ Pharm. Inc. / Spray / wipe <10min
ammonium salt
Quaternary
T-spray Pharm. Inc. / Spray / wipe <10min
ammonium salt
3. When reach the recommended disinfection time, remove the probe from the
solution or stop spraying or wiping the probe.
4. Rinse the disinfectant on the probe with running water and dry the probe with a
sterile cloth.
NOTE : The recommended disinfectants listed in this manual are recommended
because they are chemically compatible with the materials of the
products to be sterilized, not because of their biological effects. For the
biological effects of disinfectants, please refer to the guidance and
recommendations provided by the disinfectant manufacturer, the US
Food and Drug Administration, and the US Centers for Disease Control.
166
12 Probe and Biopsy
Before and after each exam, the intracavitary probe should be disinfected as required.
Perform the following steps:
1. Rinse the probe. Refer to “12.1.4 Probe Cleaning” for details.
2. Disinfect the probe according to the high-level disinfection method
recommended in the table below.
Table 59 High-level Disinfection
Manner Contact
Disinfectant Manufacturer Active Ingredient Concentration
of time
3. When reach the recommended disinfection time, remove the probe from the
solution or stop spraying or wiping the probe.
4. Rinse the disinfectant on the probe with running water and dry the probe with a
sterile cloth.
NOTE : The recommended disinfectants listed in this manual are recommended
because they are chemically compatible with the materials of the
products to be sterilized, not because of their biological effects. For the
biological effects of disinfectants, please refer to the guidance and
recommendations provided by the disinfectant manufacturer, the US
Food and Drug Administration, and the US Centers for Disease Control.
If the disinfectant recommended in this manual is not selected, prepare
the solution according to the disinfectant manufacturer’s instructions.
12.1.6Probe Maintenance
The probe should be used, stored or transported under the conditions of “Appendix A
Technical Specifications”.
To properly store the probe for transportation, please perform the following steps:
1. Prior to place the probe in the shipping container, make sure the probe is cleaned
and sanitized to avoid contaminating the liner foam of the shipping container.
2. Carefully place the probe in the probe box.
3. Ensure the probe is totally in the box before covering the probe box.
4. Wrap the probe box in a plastic material with air bag (such as foam) and place
the wrapped probe box into the cardboard box.
167
12 Probe and Biopsy
Place the probe in the probe holder when the probe is temporarily idle.
If the probe is not used for a long time, clean and disinfect the probe to ensure
that the probe is completely dry before storing in the probe box.
Prior to place the probe, make sure the probe holder is clean and dry.
Separately store the probe from other medical devices.
Calibrate the guide line is necessary before each biopsy procedure or changing biopsy
angle.
Perform the following steps:
1. Install the needle-guided bracket.
2. Place the head of the probe in a liquid tank with sterile water.
3. Insert the puncture needle along the needle-guided bracket until the tip of the
puncture needle is at the maximum depth of the water.
168
12 Probe and Biopsy
Item Description
6. Select [Biopsy Kit] by pressing the [ ] or [ ] button on the control panel and
then rotate the [Value] knob to select appropriate kit and angle.
7. Select [Calibrate] by pressing the [ ] or [ ] button on the control panel and
then press the [Value] knob to enter the Biopsy Calibration screen.
8. Verify that the needle displayed on the ultrasound image matches the guide line.
If not, adjust the guide line angle and position.
Select [Pos] by pressing the [ ] or [ ] button on the control panel and then rotate
Select [Angle] by pressing the [ ] or [ ] button on the control panel and then rotate
control panel and then press the [Value] knob to return to B-mode.
169
13 System Maintenance
13 System Maintenance
To avoid the long-term exposure to dust and moisture which may result in poor
performance and safety of the system, routine system maintenance should be performed
by the user.
When cleaning the system, pay special attention not to let liquid enter
any part of the system. Otherwise, electric shock may result. If liquid is
accidentally into the system, stop cleaning the system immediately and
shut down the system, contact the manufacturer Customer Service
Department or sales representative.
NOTE : Before cleaning procedure, shut down the system and unplug the
power cord. Otherwise, electric shock may result.
Maintenance shall be taken regularly to ensure the performance of
the system, and ensure that the system is well grounded and meets
the safety requirements.
Except for the basic cleaning and maintenance described in this
section, only personnel authorized by the manufacturer may perform
other maintenance.
170
13 System Maintenance
NOTE : Liquid or other substances penetrating into the electronic components under
the panel may result intermittent failure of the control panel and keyboard.
Visually inspect for potential problems.
13.3 Troubleshooting
Table 61 Troubleshooting
No. Failure Measure
When the above failures occur, and the corresponding measures still cannot solve, please
contact the manufacturer or authorized personnel for further support. In the case of more
difficult problems with individual functions, please contact the manufacturer or authorized
171
13 System Maintenance
personnel.
13.4 Disposal
When the system, probe, needle-guide bracket and other accessories reach the end of
their service period, they shall be disposed according to the regulations on waste
electrical and electronic equipment (WEEE) or local regulations.
Refer to manufacturer or supplier for information on disposal or recycle.
The manufacturer does not assume liability for any damage caused by unauthorized
handling of the system without consulting.
172
Annex
173
Annex
Electronic data
Electronic data interchange
Function interchange
(unidirectional)
(bidirectional)
Clinical images and
Network Used for Software upgrade
patient information
Wired network (10M /
Exchange method 100M / 1000M USB
adaptive), USB
Safety software / /
Monitor 12.1''LCD
174
Annex
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating
normally.
Use of accessories, transducers and cables other than those specified or provided by
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the Ultrasound machine, including cables specified by the manufacturer.
NOTE
The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) this equipment might not offer adequate
175
Annex
The ulyrasound machine is intended for use in the electromagnetic environment specified below. The
2
customer or the user of ultrasound machine should assure that it is used in such an environment.
RF emissions
5 Class A
CISPR 11
The ultrasound machine is suitable for use in all
IEC 61000-3-3
Accessories for Ultrasound System may affect their radiation levels. The accessories
listed in this section have been tested in accordance with international standards for use in
Ultrasound System and have been verified to meet radiation standards. Please use only
the attachments listed in this section.
The user should ensure the electromagnetic compatibility of the Ultrasound System
when attaching accessories to Ultrasound System. Unless otherwise specified, use only
EMC compliant equipment.
176
Annex
1 Power cable 2m No
Warning
This equipment is only used by professional medical staff. The equipment/system may cause radio interference
or disrupt the operation of nearby equipment. Mitigation measures may be necessary, such as recalibrating the
direction, resetting the device or shielding the appropriate site.
The ultrasound machine is intended for use in the electromagnetic environment specified below. The customer or
the user of the ultrasound machine should assure that it is used in such an environment.
177
Annex
Voltage dips, short 0°, 45°, 90°, 135°, 180°, 45°, 90°, 135°, 180°, 225°, typical commercial or hospital
interruptions and 225°, 270°and 315° 270°and 315° environment. If the user of the
voltage variations ultrasound machine requires
on power supply 0 % UT; 1 cycle and 0 % UT; 1 cycle and 70 % continued operation during power
input lines 70 % UT; 25/30 cycles UT; 25/30 cycles Single mains interruptions, it is
178
Annex
The ultrasound machine is intended for use in the electromagnetic environment specified below. The customer or
the user of the ultrasound machine should assure that it is used in such an environment.
Electromagnetic environment -
Immunity test IEC 60601 test level Compliance level
guidance
3.5
d [ ] P
Conducted RF 3 Vrms 3V V1
10 V/m 7
10 V/m d [ ] P 800 MHz to 2.7 GHz
Radiated RF E1
80 MHz to 2.7 GHz
80 MHz to 2.7 GHz
IEC 61000-4-3 where p is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
385MHz-5785MHz and d is the recommended separation
385MHz-5785MHz
Test specifications distance in metres (m).b
Test specifications
for ENCLOSURE
for ENCLOSURE
PORT IMMUNITY to
PORT IMMUNITY to
179
Annex
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection
from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795
MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio
bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz
to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89
MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
b Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the ultrasond machine is used
exceeds the applicable RF compliance level above, the ultrasound machine should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the ultrasound machine.
c Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
180
Annex
The ultrasound machine is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the ultrasound machine can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the ultrasound machine as recommended below,
according to the maximum output power of the communications equipment
100 12 20.00 35 70
For transmitters rated at a maximum output power not listed above the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
181
Annex
Plastic case ○ ○ ○ ○ ○ ○
Metal bracket ○ ○ × × ○ ○
Keyboard silicone ○ ○ ○ ○ ○ ○
Trackball × ○ ○ ○ ○ ○
Cables × ○ ○ ○ ○ ○
PCBA × ○ ○ ○ ○ ○
Hard disk ○ ○ ○ ○ ○ ○
○:Means that the content of the toxic and harmful substance in all homogeneous materials of the part
is below the limit specified in local standard.
×:It means that the content of the toxic and harmful substance in at least one homogeneous material
of the part exceeds the limit requirement specified in local standard.
For products sold since the date of sale, this table shows that products in our supply chain may contain
these substances. Note: all listed components may or may not be included in the products sold.
182
Annex
TC50(IEC60601-2-37:2007)
B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.48 0.15 0.17
Associated pra 2.35
acoustic P 7.34 7.34
parameter min [Pα(zs), Ita,α(zs)]
s zs
zbp
zb
z@max. Ipi,α 2.72
deq (zb)
fawf 2.50 4.00 2.50
Dim of X 3.46 3.46
Aaprt Y 0.90 0.90
Other td 0.812
informatio prr 990
n pr @ max. Ipi 3.21
deq @ max. Ipi
Ipa,α @max. MI 193
183
Annex
PW mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.22 0.42 3.66 2.27
Associated pra 1.93
acoustic P 72.9 72.9
parameters min [Pα(zs), Ita,α(zs)] 22.3
zs 1.10
zbp 2.22
zb 2.35
z@max. Ipi,α 1.47
deq (zb) 0.26
fawf 2.5 2.5 3.5 3.1
Dim of X 13.5 0.77 0.38
Aaprt Y 8.76 0.90 0.90
Other td 1.86
information prr 697
pr @ max. Ipi 2.57
deq @ max. Ipi 0.25
Ipa,α @max. MI 219
Focal X cm 13.0
Length Y cm 5.00
Operating Focus position 2cm 3cm 13cm 2cm
control Display depth 6cm 6cm 21cm 6cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 697 5699 697 5699
Working Pen Pen Res Gen
frequency
SV 0.5mm 0.5mm 0.5mm 0.5mm
184
Annex
B+M mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.27 0.26 1.22 0.91
Associated pra 2.01
acoustic P 25.9 25.9
parameters min [Pα(zs), Ita,α(zs)] 17.1
zs 1.48
zbp 2.98
zb 2.01
z@max. Ipi,α 2.76
deq (zb) 0.31
fawf 2.5 2.0 2.0
Dim of X 10.6 0.58
Aaprt Y 22.5 0.90
Other td 0.76
information prr 1000
pr @ max. Ipi 2.17
deq @ max. Ipi 0.29
Ipa,α @max. MI 201
Focal X 17.0 17.0
Length Y 5.0 5.0
Operating Focus position 3cm 2cm 17cm 2cm
control Display depth 6cm 6cm 21cm 6cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 1000 2000 1000 2000
Working Pen HPen HPen HPen
frequency
Scan Line Mode Lower-density line and minimum scan angle
185
Annex
Color+B/PDI+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.04 0.36 0.47
Associated pra 2.07
acoustic P 20.1 20.1
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb
z@max. Ipi,α 1.91
deq (zb)
fawf 4.0
Dim of X
Aaprt Y
Other td 0.68
information prr 2380
pr @ max. Ipi 2.74
deq @ max. Ipi
Ipa,α @max. MI 287
Focal X cm
Length Y cm
Operating B Focus/Color
control Sampling Gate 2cm 11cm 11cm 11cm
conditions position
Display depth 6cm 21cm 21cm 21cm
Acoustic power 100% 100% 100% 100%
Color PRF(Hz) 2380 2270 2270 3060
B Working
Res Res Gen Gen
frequency
C Working
Res Res Gen Gen
frequency
Color Sampling minimum minimum
Gate Width
Scan Line Mode Lower-density line and minimum scan angle for B-mode and Color
Sensitivity:2 for Color-mode
186
Annex
PW+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.25 0.74 2.84 0.83
Associated pra 1.97
acoustic P 63.0 63.0
parameters min [Pα(zs), Ita,α(zs)] 39.1
zs 1.10
zbp 2.22
zb 3.12
z@max. Ipi,α 1.89
deq (zb) 0.22
fawf 2.5 3.1 3.5
Dim of X 9.29 0.96
Aaprt Y 9.18 0.90
Other td 0.67
information prr 9100
pr @ max. Ipi 2.23
deq @ max. Ipi 0.22
Ipa,α @max. MI 183
Focal X cm 13.0 13.0
Length Y cm 5.00 5.00
Operating B Focus/SV 2cm 4cm 13cm 17cm
control position
conditions Display depth 6cm 21cm 21cm 21cm
Acoustic power 100% 100% 100% 100%
PW PRF(Hz) 690 4476 3756 690
B Working Pen Gen Res Gen
frequency
PW Working Pen Gen Res Gen
frequency
PW SV 0.5mm 0.5mm
Scan Line Mode Lower-density line and minimum scan angle
187
Annex
Color+PW+B mode
TIS TIB
Non-scan
Index label MI MI
Scan Aaprt ≤ Aaprt > 1 Non-scan
1 cm2 cm2
Maximum index value 1.32 0.77 3.00 2.28
Associated pra 2.09
acoustic P 60.6 60.6 16.2
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb 2.35
z@max. Ipi,α 1.45
deq (zb) 0.26
fawf 2.5 2.5 3.5
Dim of X 2.99 0.77
Aaprt Y 6.89 0.90
Other td 1.85
information prr 698
pr @ max. Ipi 2.23
deq @ max. Ipi 0.25
Ipa,α @max. MI 197
Focal X cm 3.00
Length Y cm 5.00
Operating B Focus /PW SV
2cm 3cm 3cm 2cm
control /Color SG position
conditions Display depth 6cm 6cm 6cm 6cm
Acoustic power 100% 100% 100% 100%
PRF(Hz) PW:698 PW:12660 PW:698 PW:15830
C:14370 C:476 C:11020 C:476
B Working
Pen Res Res Res
frequency
PW Working
Pen Pen Res Gen
frequency
Color Working
Pen Res Res Res
frequency
PW SV/Color SG SV=0.5mm SG minimum
Scan Line Mode Lower-density line and minimum scan angle
188
Annex
TC50(IEC61157:2007)
Mode B M B+M PW
Parameter
System settings AP:100% AP:100% AP:100% AP:100%
Focus:13cm Focus:9cm Focus:17cm Focus:13cm
Display:21cm Display:21cm Display:21cm Display:21cm
189
Annex
TP16(IEC60601-2-37:2007)
B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.36 0.41 0.68
Associated pra 1.83
acoustic P 32.9 32.9
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb
z@max. Ipi,α 1.36
deq (zb)
fawf 1.8 2.0 3.0
Dim of X 1.74 1.26
Aaprt Y 1.16 1.16
Other td 1.26
information prr 2590
pr @ max. Ipi 2.03
deq @ max. Ipi
Ipa,α @max. MI 101
Focal X cm 14.0 9.00
Length Y cm 5.00 5.00
Operating Focus position 2cm 14cm 14cm 9cm
control Display depth 20cm 20cm 20cm 20cm
conditions Acoustic power 100% 100% 100% 100%
Focus number 1 1 1 1
Working frequency HPen Pen Pen HRes
Scan Line Mode Lower-density line and minimum scan angle
190
Annex
PW mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.36 1.29 2.97 2.54
Associated pra 2.15
acoustic P 114 114
parameters min [Pα(zs), Ita,α(zs)] 90.9
zs 1.10
zbp 2.04
zb 4.10
z@max. Ipi,α 3.12
deq (zb) 0.37
fawf 2.5 2.5 2.5 2.5
Dim of X 5.89 1.08 1.38
Aaprt Y 9.77 1.16 1.16
Other td 1.38
information prr 699
pr @ max. Ipi 2.86
deq @ max. Ipi 0.34
Ipa,α @max. MI 277
Focal X cm 9.00
Length Y cm 5.00
Operating Focus position 3cm 7cm 9cm 11cm
control Display depth 5cm 20cm 20cm 20cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 699 5699 5699 5699
Working frequency Res Res Res Res
SV 0.5mm 0.5mm 0.5mm 0.5mm
191
Annex
B+M mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.58 0.46 0.79 1.08
Associated pra 2.13
acoustic P 28.6 28.6
parameters min [Pα(zs), Ita,α(zs)] 24.7
zs 1.10
zbp 2.04
zb 3.97
z@max. Ipi,α 1.36
deq (zb) 0.47
fawf 1.8 2.0 3.0
Dim of X 8.85 1.20
Aaprt Y 10.4 1.16
Other td 1.22
information prr 500
pr @ max. Ipi 2.34
deq @ max. Ipi 0.45
Ipa,α @max. MI 135
Focal X cm 9.00 9.00
Length Y cm 5.00 5.00
Operating Focus position 2cm 8cm 9cm 9cm
control Display depth 20cm 20cm 20cm 20cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 500 1000 500 500
Working frequency HPen HPen-Gen HRes HRes
Scan Line Mode Lower-density line and minimum scan angle
192
Annex
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.23 0.61 0.94
Associated pra 2.13
acoustic P 53.3 53.3
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb
z@max. Ipi,α 3.10
deq (zb)
fawf 3.0
Dim of X
Aaprt Y
Other td 1.52
information prr 1190
pr @ max. Ipi 2.78
deq @ max. Ipi
Ipa,α @max. MI 261
Focal X cm
Length Y cm
Operating B Focus/ Color 3cm 10cm 10cm 5cm
control Sampling Gate
conditions position
Display depth 5cm 20cm 20cm 20cm
Acoustic power 100% 100% 100% 100%
Color PRF(Hz) 1190 1200 1200 1133
B Working Gen Res Res Gen
frequency
C Working Gen Res Res Gen
frequency
Color Sampling minimum minimum
Gate Width
Scan Line Mode Lower-density line and minimum scan angle for B-mode and Color
Sensitivity:2 for Color-mode
193
Annex
PW+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.05 1.22 2.75 2.72
Associated pra 1.81
acoustic P 120 120
parameters min [Pα(zs), Ita,α(zs)] 96.8
zs 1.19
zbp 2.04
zb 4.70
z@max. Ipi,α 4.23
deq (zb) 0.42
fawf 3.0 2.5 2.5
Dim of X 8.57 1.26
Aaprt Y 10.3 1.16
Other td 1.41
information prr 699
pr @ max. Ipi 2.67
deq @ max. Ipi 0.40
Ipa,α @max. MI 231
Focal X cm 9.00 9.00
Length Y cm 5.00 5.00
Operating B Focus/SV 7cm 9cm 9cm 9cm
control position
conditions Display depth 20cm 20cm 20cm 20cm
Acoustic power 100% 100% 100% 100%
PW PRF(Hz) 699 2305 2305 2305
B Working Gen Gen Gen Gen
frequency
PW Working Res Res Res Res
frequency
PW SV 0.5mm 0.5mm
Scan Line Mode Lower-density line and minimum scan angle
194
Annex
Color+PW+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ Aaprt > 1 Non-scan
1 cm2 cm2
Maximum index value 1.37 1.40 2.58 2.11
Associated pra 2.74
acoustic P 81.2 81.2 59.4
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb 3.28
z@max. Ipi,α 1.51
deq (zb) 0.44
fawf 4.0 2.5 2.5
Dim of X 4.78 0.84
Aaprt Y 11.0 1.16
Other td 0.47
information prr 7010
pr @ max. Ipi 3.16
deq @ max. Ipi 0.39
Ipa,α @max. MI 409
Focal X cm 3.00
Length Y cm 5.00
Operating B Focus/PW 3cm 3cm 3cm 3cm
control SV/Color SG
conditions position
Display depth 20cm 5cm 5cm 5cm
Acoustic power 100% 100% 100% 100%
PRF(Hz) PW: 699 PW: 8713 PW: 699 PW: 699
C:7077 C:365 C:11454 C:10817
B Working Res Gen Res Gen
frequency
PW Working Gen Res Res Res
frequency
Color Working Res Gen Res Gen
frequency
PW SV/Color SG SV=0.5mm SG minimum
Scan Line Mode Lower-density line and minimum scan angle
195
Annex
CW mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 0.092 0.92 3.23 2.16
Associated pra 0.13
acoustic P 103 103
parameters min [Pα(zs), Ita,α(zs)] 85.3
zs 1.20
zbp 1.78
zb 3.50
z@max. Ipi,α 3.08
deq (zb) 0.49
fawf 2.0 2.0 2.0 2.0
Dim of X 0.96 0.96 6.15
Aaprt Y 1.16 1.16 10.2
Other td
information prr 200000
pr @ max. Ipi 0.16
deq @ max. Ipi 0.44
Ipa,α @max. MI 0.41
Focal X cm 5.00
Length Y cm 5.00
Operating Focus position 5cm 5cm 5cm 5cm
control Display depth 0cm 0cm 0cm 0cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 200000 200000 200000 200000
Working 3.3MHz 3.3MHz 3.3MHz 3.3MHz
frequency
196
Annex
TP16(IEC61157:2007)
Mode B M B+M PW CW
Parameter
197
Annex
TL40(IEC60601-2-37:2007)
B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.29 1.65 0.94
Associated pra 2.88
acoustic P 59.3 59.3
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb
z@max. Ipi,α 1.63
deq (zb)
fawf 5.0 7.5 7.5
Dim of X 2.76 2.76
Aaprt Y 0.70 0.70
Other td 0.258
information prr 4100
pr @ max. Ipi 3.91
deq @ max. Ipi
Ipa,α @max. MI 566
Focal X cm 12.0 12.0
Length Y cm 3.0 3.0
Operating Focus position 1.5cm 12cm 12cm 12cm
control Display depth 15cm 15cm 15cm 15cm
conditions Acoustic power 100% 100% 100% 100%
Focus number 1 1 1 1
Working frequency Pen Gen Pen Gen
Scan Line Mode Lower-density line and minimum scan angle
198
Annex
PW mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.52 0.656 2.55 1.06
Associated pra 3.23
acoustic P 27.9 27.9
parameters min [Pα(zs), Ita,α(zs)] 25.2
zs 1.00
zbp 1.90
zb 0.954
z@max. Ipi,α 0.347
deq (zb) 0.171
fawf 4.5 5.0 5.0 5.0
Dim of X 17.1 0.72 0.36
Aaprt Y 4.30 0.70 0.70
Other td 0.901
information prr 698
pr @ max. Ipi 3.40
deq @ max. Ipi 0.171
Ipa,α @max. MI 358
Focal X cm 12.0
Length Y cm 3.0
Operating Focus position 0.5cm 1cm 12cm 0.5cm
control Display depth 3cm 3cm 15cm 3cm
conditions Acoustic power 100% 100% 100% 100%
Focus number 698 5699 698 5699
Working frequency Pen Res Res Pen
Scan Line Mode 0.5mm 0.5mm 0.5mm 0.5mm
199
Annex
B+M mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.18 1.08 1.08 0.702
Associated pra 2.73
acoustic P 4.68 4.68
parameters min [Pα(zs), Ita,α(zs)] 7.15
zs 1.00
zbp 2.35
zb 1.59
z@max. Ipi,α 1.62
deq (zb) 0.152
fawf 5.3 5.0 6.0
Dim of X 16.0 0.72
Aaprt Y 3.88 0.70
Other td 0.496
information prr 2400
pr @ max. Ipi 3.57
deq @ max. Ipi 0.152
Ipa,α @max. MI 271
Focal X cm 12.0 12.0
Length Y cm 3.00 3.00
Operating Focus position 1.5cm 1.5cm 12cm 12cm
control Display depth 16cm 16cm 16cm 16cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 500 1000 1000 1000
Working frequency HGen Pen Gen Gen
Scan Line Mode Lower-density line and minimum scan angle
200
Annex
Color+B/PDI+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.64 1.58 1.62
Associated pra 3.67
acoustic P 59.8 59.8
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb
z@max. Ipi,α 1.09
deq (zb)
fawf 5.0
Dim of X
Aaprt Y
Other td 0.352
information prr 2700
pr @ max. Ipi 4.41
deq @ max. Ipi
Ipa,α @max. MI 683
Focal X cm
Length Y cm
Operating B Focus/Color 1cm 12cm 12cm 12cm
control Sampling Gate
conditions position
Display depth 3cm 15cm 15cm 15cm
Acoustic power 100% 100% 100% 100%
Color PRF(Hz) 2700 2040 2040 2040
B Working Res Pen Pen Pen
frequency
C Working Res Pen Pen Pen
frequency
Color Sampling minimum minimum
Gate Width
Scan Line Mode Lower-density line and minimum scan angle for B-mode and Color
Sensitivity:2 for Color-mode
201
Annex
PW+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.43 0.348 2.69 0.97
Associated pra 3.19
acoustic P 27.5 27.5
parameters min [Pα(zs), Ita,α(zs)] 12.5
zs 1.00
zbp 1.90
zb 0.954
z@max. Ipi,α 0.345
deq (zb) 0.16
fawf 5.0 6.2 4.5
Dim of X 16.5 0.72
Aaprt Y 2.93 0.70
Other td 0.907
information prr 699
pr @ max. Ipi 3.34
deq @ max. Ipi 0.16
Ipa,α @max. MI 354
Focal X cm 12.0 12.0
Length Y cm 3.00 3.00
Operating B Focus/SV 0.5cm 1cm 12cm 12cm
control position
conditions Display depth 3cm 15cm 15cm 15cm
Acoustic power 100% 100% 100% 100%
PW PRF(Hz) 699 5070 699 699
B Working Pen Res Pen Res
frequency
PW Working Pen Res Pen Res
frequency
PW SV 0.5mm 0.5mm
Scan Line Mode Lower-density line and minimum scan angle
202
Annex
Color+PW+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ Aaprt > 1 Non-scan
1 cm2 cm2
Maximum index value 1.46 1.74 1.97 2.35
Associated pra 3.27
acoustic P 16.2 16.2 58.1
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb 0.954
z@max. Ipi,α 0.341
deq (zb) 0.129
fawf 5.0 5.0 5.0
Dim of X 1.2 0.72
Aaprt Y 6.44 0.70
Other td 0.902
information prr 699
pr @ max. Ipi 3.40
deq @ max. Ipi 0.129
Ipa,α @max. MI 353
Focal X cm 1.00
Length Y cm 3.00
Operating B Focus/ PW SV/ 0.5cm 1cm 1cm 0.5cm
control Color SG
conditions position
Display depth 3cm 3cm 3cm 3cm
Acoustic power 100% 100% 100% 100%
PRF(Hz) PW:699 PW:22995 PW:699 PW:699
C:12470 C:249 C:16799 C:8799
B Working Pen Res Gen Gen
frequency
PW Working Pen Res Res Res
frequency
Color Working Pen Res Gen Gen
frequency
PW SV/Color SG SV=0.5mm SG minimum
Scan Line Mode Lower-density line and minimum scan angle
203
Annex
TL40 (IEC61157:2007)
Mode B M B+M PW
Parameter
System settings AP:100% AP:100% AP:100% AP:100%
Focus:12cm Focus:12cm Focus:12cm Focus:12cm
Display:16cm Display:16cm Display:16cm Display:16cm
204
Annex
TC10(IEC60601-2-37:2007)
B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.30 0.068 0.14
Associated pra 3.19
acoustic P 3.18 3.18
parameters
min [Pα(zs), Ita,α(zs)]
zs
zbp
zb
z@max. Ipi,α 1.12
deq (zb)
fawf 6.0 4.7 4.5
Dim of X 1.30 0.78
Aaprt Y 0.50 0.50
Other td 0.29
information prr 4000
pr @ max. Ipi 3.67
deq @ max. Ipi
Ipa,α @max. MI 276
Focal X cm 12.0 4.5
Length Y cm 3.0 3.0
Operating Focus position 1cm 12cm 12cm 4.5cm
control Display depth 3cm 15cm 15cm 15cm
conditions Acoustic power 100% 100% 100% 100%
Focus number 1 1 1 1
Working frequency Gen HRes HRes Pen
Scan Line Mode Lower-density line and minimum scan angle
205
Annex
PW mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 0.99 1.80 0.61 2.11
Associated pra 2.48
acoustic P 7.83 7.83 7.83
parameters
min [Pα(zs), Ita,α(zs)]
zs
zbp
zb 1.0
z@max. Ipi,α 1.53
deq (zb) 0.207
fawf 6.2 4.6 4.6 4.6
Dim of X 13.4 0.25 0.17
Aaprt Y 4.36 0.50 0.50
Other td 0.69
information prr 679
pr @ max. Ipi 3.53
deq @ max. Ipi 0.198
Ipa,α @max. MI 179
Focal X cm 12.0
Length Y cm 3.0
Operating Focus position 2cm 1cm 12cm 1cm
control Display depth 15cm 3cm 15cm 3cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 679 5699 5300 23990
Working frequency HRes Pen Pen Pen
SV 0.5mm 0.5mm 0.5mm 0.5mm
206
Annex
B+M mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.30 0.475 0.842 1.10
Associated pra 3.07
acoustic P 10.8 10.9 10.9
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb 1.02
z@max. Ipi,α 1.21
deq (zb) 0.218
fawf 5.6 6.2 6.2
Dim of X 13.5 0.25
Aaprt Y 4.77 0.50
Other td 0.339
information prr 1000
pr @ max. Ipi 3.66
deq @ max. Ipi 0.218
Ipa,α @max. MI 220
Focal X cm 12.0 12.0
Length Y cm 3.0 3.0
Operating Focus position 1cm 1cm 12cm 0.5cm
control Display depth 3cm 3cm 15cm 3cm
conditions Acoustic power 100% 100% 100% 100%
PRF(Hz) 1000 2000 500 2000
Working frequency Gen HPen HRes HPen
Scan Line Mode Lower-density line and minimum scan angle
207
Annex
Color+B/PDI+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.45 0.141 0.23
Associated pra 3.08
acoustic P 7.39 7.39
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb
z@max. Ipi,α 1.09
deq (zb)
fawf 4.5
Dim of X
Aaprt Y
Other td 0.933
information prr 1680
pr @ max. Ipi 3.79
deq @ max. Ipi
Ipa,α @max. MI 356
Focal X cm
Length Y cm
Operating B Focus/Color
control Sampling Gate 1cm 12cm 12cm 12cm
conditions position
Display depth 15cm 15cm 15cm 15cm
Acoustic power 100% 100% 100% 100%
Color PRF(Hz) 1680 2290 2290 2290
B Working
Pen Pen Pen Pen
frequency
C Working
Pen Pen Pen Pen
frequency
Color Sampling minimum minimum
Gate Width
Scan Line Mode Lower-density line and minimum scan angle for B-mode and Color
Sensitivity:2 for Color-mode
208
Annex
PW+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ 1 Aaprt > 1 Non-scan
cm2 cm2
Maximum index value 1.36 1.07 1.69 1.64
Associated pra 2.89
acoustic P 22.2 22.2 22.2
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb 1.47
z@max. Ipi,α 1.02
deq (zb) 0.214
fawf 4.5 4.6 4.6
Dim of X 13.4 0.25
Aaprt Y 4.34 0.50
Other td 0.42
information prr 983
pr @ max. Ipi 3.25
deq @ max. Ipi 0.193
Ipa,α @max. MI 311
Focal X cm 12.0 12.0
Length Y cm 3.0 3.0
Operating B Focus/SV 1cm 1cm 12cm 0.5cm
control position
conditions Display depth 15cm 3cm 15cm 3cm
Acoustic power 100% 100% 100% 100%
PW PRF(Hz) 2936 15610 2939 16510
B Working Pen Pen Pen Pen
frequency
PW Working Pen Pen Pen Pen
frequency
PW SV 0.5mm 0.5mm
Scan Line Mode Lower-density line and minimum scan angle
209
Annex
Color+PW+B mode
TIS TIB
Non-scan
Index label MI TIC
Scan Aaprt ≤ Aaprt > 1 Non-scan
1 cm2 cm2
Maximum index value 1.21 0.28 1.52 1.58
Associated pra 2.85
acoustic P 20.7 20.7 11.6
parameters min [Pα(zs), Ita,α(zs)]
zs
zbp
zb 1.38
z@max. Ipi,α 1.04
deq (zb) 0.226
fawf 5.5 4.6 5.6
Dim of X 2.88 0.25
Aaprt Y 4.53 0.50
Other td 0.423
information prr 589
pr @ max. Ipi 3.51
deq @ max. Ipi 0.206
Ipa,α @max. MI 319
Focal X cm 1.00
Length Y cm 3.00
Operating B Focus/PW
control SV/Color SG 1cm 1cm 1cm 0.5cm
conditions position
Display depth 3cm 3cm 3cm 3cm
Acoustic power 100% 100% 100% 100%
PRF(Hz) PW:589 PW:14230 PW:699 PW:14240
C:20557 C:527 C:17957 C:528
B Working
Pen Pen Pen Pen
frequency
PW Working
Pen Pen Res Pen
frequency
Color Working
Pen Pen Pen Pen
frequency
PW SV/Color SG SV=0.5mm SG minimum
Scan Line Mode Lower-density line and minimum scan angle
210
Annex
TC10(IEC61157:2007)
Mode B M B+M PW
Parameter
System settings AP:100% AP:100% AP:100% AP:100%
Focus:12cm Focus:12cm Focus:12cm Focus:12cm
Display:16cm Display:16cm Display:16cm Display:16cm
No. 3510000723
211