Certificate of Calibration
Fluke Nederland B.V.

Certificate Number: SA01118574 Date of Calibration: 06 Dec 2022


Receive Condition: IN TOLERANCE Date of Recalibration: 06 Dec 2023
Return Condition: IN TOLERANCE Place of Calibration: Eindhoven
Manufacturer: FLUKE Temperature within: (23.0 ± 3) °C
Model: DSX-602-NW INT Humidity within: (45 ± 20) %rh
Serial Number: 20400087-20400090
Description: 500 MHZ CABLEANALYZER V2
Procedure: Manual Procedure

Customer: DISCOMP S.R.O.


PLZEN
Customer Asset ID: -
RMA Number: 606297279

All measurements are traceable to national and/or international standards or have been derived by approved ratio
techniques. When possible standards used for this calibration are ISO/IEC 17025 accredited calibrated.

This calibration is performed by a DEKRA certified lab for ISO 9001. This certificate may not be reproduced other than in
full. Calibration certificates without signatures, either electronic or handwritten, are not valid.

Comments:

Issue Date: 06 Dec 2022


F8BBB8F02B75EB75E695E0ACC0F8BBB8F
E2AAA2E035B035B030ADBD25E0E2AAA2E
DC83DDAFD2CD02CD02815B8E1CF237836 Electronically signed
4F4DFDA9D216F536FFE9B2C76EE62D43E
EADD53AB4BDD02CD05E39EE199F0134BE
2DD859ABC066F536F5F6D3A8DBE3324BE Authorized By
B22AA2B87DAD02CDB481156EF8A8F2467
F0EEE0F0B646F533F519A871FC9FE049A W.H.J. van Hulten
888888808008008880888008800800088

Fluke Nederland B.V. E-mail Telephone Rev 221123

Brainport Industries Campus 1, 5657 BX, Eindhoven [email protected] +31 40 267 5300 Page 1 of 8
Certificate of Calibration

Certificate Number: SA01118574

Remarks

- The calibration status found in this certificate on the top of each results page must be interpreted as:

As Found : Data collected before the unit was adjusted and / or repaired
As Left : Data collected after the unit has been adjusted and / or repaired
Found / Left : Data collected without any adjustment and / or repair performed

- The calibration interval (due date) is the responsibility of the end user.

- According to the European norm 'Operation of electrical installations' NEN-EN 50110-1 release 2013 and the Dutch norm
NEN 3140 release 2015 paragraph 5.102.12 through 5.102.16, a safety test is not required. Therefore not performed.

- Temperature conversions (if applicable) are performed according to ISO/IEC 60584:2013 for thermocouples, and ISO/IEC
60751:2022 for resistance temperature devices.

Standards and test-equipment used

Inventory No Model Serial No

WP2992 DSX-CALVERST E000244

en_US Page 2 of 8
DSX Cable Analyzer
Found-Left Report
Model DSX-600 Copper Tester From MAIN
Serial Number 20400087 Test date 6-Dec-22
Page 1 of 6
NEXT
NEXT Artifact SN 5100049
3
2.5
2
(dB)

1.5
1
0.5
0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.330 at 1 MHz in pair 36-45. Worst accuracy at each frequency shown.

CDNEXT
CDNEXT Artifact SN 5100030
2.5

1.5
(dB)

0.5

0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.490 at 31.25 MHz in pair 45-78. Worst accuracy at each frequency shown.

CMRL
CMDMRL Artifact SN 5100016
3.5
3.0
2.5
2.0
(dB)

1.5
1.0
0.5
0.0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.530 at 1 MHz in pair 12. Worst accuracy at each frequency shown.
Measured difference of DSX and reference laboratory equipment added to measurement accuracy of reference laboratory equipment. Worst
accuracy at each frequency shown.
Corresponding measurement accuracy requirement for nominally compliant Level IV or Level 2G/VI field tester.
DSX Cable Analyzer
Found-Left Report
Model DSX-600 Copper Tester From MAIN
Serial Number 20400087 Test date 6-Dec-22
Page 2 of 6
RL
CMDMRL Artifact SN 5100016
3.5
3
2.5
2
(dB)

1.5
1
0.5
0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 1.010 at 350 MHz in pair 78. Worst accuracy at each frequency shown.

TCL
TCL Artifact SN 5100063
1.4
1.2
1.0
0.8
(dB)

0.6
0.4
0.2
0.0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.540 at 31.13 MHz in pair 78. Worst accuracy at each frequency shown.

IL
ILFEXT Artifact SN 3558385
2.5

2.0

1.5
(dB)

1.0

0.5

0.0
0 100 200 300 400 500
Frequency (MHz)

Pass Worst margin: 0.670 at 2.13 MHz in pair 12. Worst accuracy at each frequency shown.

Measured difference of DSX and reference laboratory equipment added to measurement accuracy of reference laboratory equipment. Worst
accuracy at each frequency shown.
Corresponding measurement accuracy requirement for nominally compliant Level IV or Level 2G/VI field tester.
DSX Cable Analyzer
Found-Left Report
Model DSX-600 Copper Tester From MAIN
Serial Number 20400087 Test date 6-Dec-22
FEXT Page 3 of 6
ILFEXT Artifact SN 3558385
3.0
1.2
2.51
2.0
0.8
(dB)

1.5
0.6
1.0
0.4
0.5
0.2
0.00
898.8
0 899100 899.2 200 899.4 300 899.6 400899.8 500
900
Frequency (MHz)

Pass Worst margin: 0.320 at 46.75 MHz in pair 78-45. Worst accuracy at each frequency shown.

Measured difference of DSX and reference laboratory equipment added to measurement accuracy of reference laboratory equipment. Worst
accuracy at each frequency shown.
Corresponding measurement accuracy requirement for nominally compliant Level IV or Level 2G/VI field tester.

Loop Resistance Loop Resistance Artifact SN 3558398


Measured Expected Limit
Resistance on pair 12 0.12 0.00 0.80 Pass
Resistance on pair 36 49.91 49.80 0.60 Pass
Resistance on pair 45 100.06 99.80 1.60 Pass
Resistance on pair 78 453.16 453.00 4.00 Pass

Resistance imbalance Resistance Unbalance Artifact SN 3558426


Measured Expected Limit
Resistance on pair 12 0.12 0.00 0.80 Pass
Resistance on pair 36 25.00 24.90 0.90 Pass
Resistance on pair 45 12.25 12.13 0.90 Pass
Resistance on pair 78 24.21 24.05 0.90 Pass
Resistance imbalance on pair 12 0.00 0.00 0.05 Pass
Resistance imbalance on pair 36 0.01 0.00 0.13 Pass
Resistance imbalance on pair 45 0.33 0.32 0.06 Pass
Resistance imbalance on pair 78 0.85 0.85 0.12 Pass

DSX-8000 only: M_IL and M_FEXT measurements validate the ability of the DSX-8000 to make measurements with DSX-5000
adapters.
M IL Not applicable

M FEXT Not applicable

Test Program TFSTest v2.5.7025


DSX Report Form v3.05 18-May-2017
DSX Cable Analyzer
Found-Left Report
Model DSX-600 Copper Tester From REMOTE
Serial Number 20400090 Test date 6-Dec-22
Page 4 of 6
NEXT
NEXT Artifact SN 5100049
3
2.5
2
(dB)

1.5
1
0.5
0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.530 at 26.88 MHz in pair 12-78. Worst accuracy at each frequency shown.

CDNEXT
CDNEXT Artifact SN 5100030
2.5

1.5
(dB)

0.5

0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.530 at 31.25 MHz in pair 45-78. Worst accuracy at each frequency shown.

CMRL
CMDMRL Artifact SN 5100016
3.5
3.0
2.5
2.0
(dB)

1.5
1.0
0.5
0.0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.420 at 1 MHz in pair 36. Worst accuracy at each frequency shown.
Measured difference of DSX and reference laboratory equipment added to measurement accuracy of reference laboratory equipment. Worst
accuracy at each frequency shown.
Corresponding measurement accuracy requirement for nominally compliant Level IV or Level 2G/VI field tester.
DSX Cable Analyzer
Found-Left Report
Model DSX-600 Copper Tester From REMOTE
Serial Number 20400090 Test date 6-Dec-22
Page 5 of 6
RL
CMDMRL Artifact SN 5100016
3.5
3
2.5
2
(dB)

1.5
1
0.5
0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 1.080 at 350 MHz in pair 78. Worst accuracy at each frequency shown.

TCL
TCL Artifact SN 5100063
1.4
1.2
1.0
0.8
(dB)

0.6
0.4
0.2
0.0
0 100 200 300 400
Frequency (MHz)

Pass Worst margin: 0.550 at 31.25 MHz in pair 78. Worst accuracy at each frequency shown.

IL
ILFEXT Artifact SN 3558385
2.5

2.0

1.5
(dB)

1.0

0.5

0.0
0 100 200 300 400 500
Frequency (MHz)

Pass Worst margin: 0.660 at 2.13 MHz in pair 12. Worst accuracy at each frequency shown.

Measured difference of DSX and reference laboratory equipment added to measurement accuracy of reference laboratory equipment. Worst
accuracy at each frequency shown.
Corresponding measurement accuracy requirement for nominally compliant Level IV or Level 2G/VI field tester.
DSX Cable Analyzer
Found-Left Report
Model DSX-600 Copper Tester From REMOTE
Serial Number 20400090 Test date 6-Dec-22
FEXT Page 6 of 6
ILFEXT Artifact SN 3558385
3.0
1.2
2.51
2.0
0.8
(dB)

1.5
0.6
1.0
0.4
0.5
0.2
0.00
898.8
0 899100 899.2 200 899.4 300 899.6 400899.8 500
900
Frequency (MHz)

Pass Worst margin: 0.340 at 47.5 MHz in pair 78-45. Worst accuracy at each frequency shown.

Measured difference of DSX and reference laboratory equipment added to measurement accuracy of reference laboratory equipment. Worst
accuracy at each frequency shown.
Corresponding measurement accuracy requirement for nominally compliant Level IV or Level 2G/VI field tester.

Loop Resistance Loop Resistance Artifact SN 3558398


Measured Expected Limit
Resistance on pair 12 0.23 0.00 0.80 Pass
Resistance on pair 36 49.98 49.80 0.60 Pass
Resistance on pair 45 100.10 99.80 1.60 Pass
Resistance on pair 78 453.12 453.00 4.00 Pass

Resistance imbalance Resistance Unbalance Artifact SN 3558426


Measured Expected Limit
Resistance on pair 12 0.19 0.00 0.80 Pass
Resistance on pair 36 25.03 24.90 0.90 Pass
Resistance on pair 45 12.33 12.13 0.90 Pass
Resistance on pair 78 24.21 24.05 0.90 Pass
Resistance imbalance on pair 12 0.00 0.00 0.05 Pass
Resistance imbalance on pair 36 0.01 0.00 0.13 Pass
Resistance imbalance on pair 45 0.33 0.32 0.06 Pass
Resistance imbalance on pair 78 0.85 0.85 0.12 Pass

DSX-8000 only: M_IL and M_FEXT measurements validate the ability of the DSX-8000 to make measurements with DSX-5000
adapters.
M IL Not applicable

M FEXT Not applicable

Test Program TFSTest v2.5.7025


DSX Report Form v3.05 18-May-2017