OperatingInstructions LabTowerAFT
OperatingInstructions LabTowerAFT
EC-Declaration of Conformity
in accordance with the EEC machine directive 2006/42/EC, appendix II A
We hereby certify that the following described machine in it's conception and form put by us into
circulation is in accordance with all the relevant essential health and safety requirements of the
EC machinery directive 2006/42/EC as amended and the national laws and regulations adopt-
ing this directive.
This declaration is no longer valid if the machine is modified without our consent.
The agreement with further valid guidelines/regulations following for the product is ex-
plained:
EMC Directive (2004/108/EC)
2
Thermo Scientific Barnstead LabTower AFT
Pure water system
Preface
in deciding to purchase a pure water system of type LabTower AFT you have selected a high-
quality product.
Before you start to install and operate your pure water system, please carefully read the infor-
mation on how it is to be properly installed and operated that is given in these Operating Instruc-
tions.
This is particularly important, as we, the manufacturer, cannot accept liability for any damage
occurring as a result of improper operation of this system, or from use of it for other than the
intended purpose.
Niederelbert, 01.04.2010
3
Thermo Scientific Barnstead LabTower AFT
Pure water system
1. Contents
Preface ....................................................................................................... 3
1. Contents ................................................................................................ 4
2. Explanatory notes on the operating instructions..................................... 6
3. Transport and packaging ....................................................................... 7
3.1 Examination on receipt ......................................................................................... 7
3.2 Complaints ........................................................................................................... 7
3.3 Packaging and return shipment ............................................................................ 7
4
Thermo Scientific Barnstead LabTower AFT
Pure water system
5
Thermo Scientific Barnstead LabTower AFT
Pure water system
EU Mark of Conformity
CSA - admission
The information provided in these operating instructions is only valid for the system which has
the serial number which is to be entered on the front page.
Please enter the serial number* of your LabTower AFT system in the
space provided on the front page.
* Read the serial number of your pure water system from the type plate.
For quick and correct service, please include the following information on all inquiries and re-
placement parts orders which relate to your system:
6
Thermo Scientific Barnstead LabTower AFT
Pure water system
3.2 Complaints
Should the system have been damaged during transport:
- Save the packaging and the outer cardboard carton (for a possible inspection and/or
return shipment of the system).
- Pack the system in a suitable plastic bag or sheet and a strong cardboard box
- so that it is shock-proof.
*The time limit for claims is 6 days after receipt of the goods.
After this deadline, all rights to claims for damages are invalid.
7
Thermo Scientific Barnstead LabTower AFT
Pure water system
4. Safety precautions
For your own safety, please observe the above safety precautions!
Your LabTower AFT system is a modularly constructed, pure water system that
serves exclusively for the purification of tap water.
Do not put the system into operation until you have taken notice of all of the appropriate
information that is given in these Operating Instructions.
Lifting and carrying the pure water system, e.g. to the installation location, should be car-
ried out by two people. To lift it, each takes hold of it under the base plate at two corners.
Note that the manufacturer is freed of all liability for damages that result from improper
operation of the system, or from use of it for other than the intended purpose.
The CE-Mark becomes invalidated should constructional changes be made to the system
or products of other manufacturers be installed in it.
Protect the system from frost. The temperature in the area in which the system is installed
must be at least +5° C and must not exceed +40° C.
Observe all regulations and requirements, including current accident regulations, that are
applicable and appropriate at the installation area, including those for the statics of the
flooring (see weight under „Technical specifications).
The raw water pressure must be at least 2 bar and at most 6 bar, should it be higher, then
an additional pressure reducer must be installed.
DIN EN 1717 requires that water purification systems be equipped with a safety device
that protects against contamination of the drinking water piping.
The installation area must have a drain at floor level with at least DN 50 pipe, otherwise
the manufacturer will not accept any liability for water damage. Should no such drain be
available, then a water watcher (240V only) (article no. 16.0129) must be installed.
After long standstill periods (e.g. holidays), the system must be subjected to rinsing and, if
appropriate, disinfection. Refer to the section on ”Cleaning and disinfection“ for details.
When selecting the installation area and installing the system, make sure that there is
sufficient working area around the system for convenient operation of it.
Never look directly into a switched-on UV-lamp, as UV-light is dangerous to eyesight. The
UV-lamp is only to be replaced by authorized person to do this.
8
Thermo Scientific Barnstead LabTower AFT
Pure water system
5. Extent of delivery
Please check the parts delivered against this list. Contact the manufacturer
should a part be missing.
9
Thermo Scientific Barnstead LabTower AFT
Pure water system
6. Intended use
The LabTower AFT pure water system is a reaction to the continually increasing requirements
that water of pure quality must fulfil, the increasingly strict demands resulting from technological
advances and the need for user-friendly systems and complete solutions.
LabTower AFT systems have been solely and specifically designed to excel in the intended use,
which is to produce sterile filtered pure water free of particles, salts and organic compounds.
To benefit from the long possible service lives of the high-quality purification media, feed the
pure water system with water which has been subjected to an upstream pre-treatment step (re-
verse osmosis, ion exchange or distillation),
10
Thermo Scientific Barnstead LabTower AFT
Pure water system
7. Technical specifications
Dimensions
Height 1500 mm
Width 450 mm
Depth 580 mm
Weight, empty 66 kg
Operational weight (with full tank) 166 kg
Water connections
Raw water inlet Hose, 8 mm od
Concentrate outlet Hose, 8 mm od
Pure water outlet Hose, 8 mm od
11
Thermo Scientific Barnstead LabTower AFT
Pure water system
12
Thermo Scientific Barnstead LabTower AFT
Pure water system
8. Flow chart
F3 Ion exchanger: Removes inorganic ions and traces of dissolved organic sub-
stances.
F5 Sterile vent filter: Prevents the penetration of airborne bacteria and particles >
(option) 0,2 µm.
LIS 100 Float switch: Determines the filling level in the tank.
13
Thermo Scientific Barnstead LabTower AFT
Pure water system
P1 Booster pump: Increases the input pressure to the required operating pres-
sure.
PS200 Pressure switch: Switches booster pump (P2) on when pure water is with-
drawn from tank (B1), and switches the pump off dispensing
valve (V9) is closed.
QIA300 Pure water con- Measuring device for the determination of the conductivity
ductivity measuring (after RO) as parameter for the pure water quality.
cell:
QIA301 Pure water con- Measuring device for the determination of the conductivity
ductivity measuring (after filter cartridge) as parameter for the pure water quality.
cell:
TIA500 Temperature sensor: Measures the temperature for temperature compensation.
V2 Feedwater solenoid valve: Is closed during stand-by and standstills. It prevents water
from flowing into the system when it is not in operation.
V3 Pressure hold valve: Serves to adjust the operating pressure and the WCF-rate.
V4 Rinsing solenoid valve: Opens for cleaning of the membrane prior to pure water-
production, after pure water production and at least every 12
hours.
V7 Check valve: Prevents pure water from flowing back into the tank.
V8 Pressure reducer: For adjustment of the pure water pressure in the system.
V10 Check valve Prevents permeat from flowing back into the device
14
Thermo Scientific Barnstead LabTower AFT
Pure water system
When the ON/OFF key is pressed, the system starts running either in the operating mode or in
the stand-by mode, according to the float switch.
The feed water flows with a maximum pressure of 6 bar into the system.
Feedwater solenoid valve (V2) is closed when the system is at stand-by or at a standstill. This
prevents feedwater from flowing into the system when it is not in operation and so protects stor-
age tank (B1) from overflowing.
Semi-permeable membrane (F2) retains all salts dissolved in the water up to the specified re-
tention quota. The molecular size of the membrane pores is so small, that 99% of bacteria, py-
rogens and particles are also removed from the feedwater.
The pure water flows past pure water probe (QIA300) and into the storage tank.
The retained feedwater constituents are led away with the rest of the concentrate.
Booster pump (P2) pumps the pure water in storage tank (B1) through pressure reducer (V8)
and ion exchanger (F3) and on past the special conductivity measurement probe (QIA301) with
temperature compensation (TIA500) to dispensing valve (V9). The conductivity measuring
probe determines the conductivity and the temperature of the pure water subsequent to UV-
disinfection* (UV1).
The measured values can be called in the microprocessor system control.
* Option
Valve (V3) has been pre-adjusted at the factory. A change in the adjustment
of this valve can result in damage to the reverse osmosis module! Because
of fluctuations in the feedwater temperature and pressure, the adjustment
of the pressure hold valve and the concentrate flow that it governs must be
checked, and readjusted if necessary, on putting the system into operation
and at regular intervals thereafter.
The measured values must be recorded in the maintenance record.
15
Thermo Scientific Barnstead LabTower AFT
Pure water system
Your pure water system is equipped with automatic flushing. Flushing is carried out when the
system is switched on, at the end of each production, and also every 12 hours. For this, rinsing
solenoid valve (V4) is opened and the strong flow of water across reverse osmosis module (F2)
then flushes coarse particles and other contaminants away from the surface of the membranes
and carries them with it to drain.
This automatic flushing has a positive effect on the service life of the reverse osmosis module.
An additional advantage of automatic flushing is that it prevents bacterial growth from occurring
in the reverse osmosis module when the system is at a standstill for a long time. For this rea-
son, we highly recommend that you leave the system switched on over the weekend and during
holiday times, so that the 12 hour flush can effectively guard against bacterial growth.
The following criteria must be taken into consideration when choosing the installation
area.
⇒ The surface that your pure water system is to be stood on must be strong enough
to support it (for weight, see chapter „Technical Specifications").
⇒ A floor drain with waste pipe of DN 50 size (38.5 mm i.d.) is required. Should this not be
available, then a water watcher (article no.: 16.0129) must be installed to protect against
damage from water flooding!
⇒ An electrical socket appropriate to the voltage given on the type plate of the system must
be positioned directly alongside the system. The safety fuse must be appropriate for the
power required (see chapter „Technical Specifications").
⇒ Position the system so that there is no difficulty in separating the device from the electric
mains.
⇒ An R 3/4" R male thread tap water connection which can be shut off must be installed in
the direct vicinity of the system. Easy access to this must be ensured.
16
Thermo Scientific Barnstead LabTower AFT
Pure water system
10. Installation
Before the pure water system can be put into operation, it must be to move to the back
of designated wall with screws.
Note: The distance from the pure water system to the wall is 20 mm.
Use the assembling kit available in wall bracket, the 2 screws 4x40 mm and the 2 dowels
(Ø 6 mm)
Plug the wall bracket into the backside of the pure water attachment, as shown graphi-
cally on the following page.
The wall bracket must shown with the notches (see position “X”) down.
Now position the pure water system at the proposed wall and attach it with the dowels
and screws.
17
Thermo Scientific Barnstead LabTower AFT
Pure water system
18
Thermo Scientific Barnstead LabTower AFT
Pure water system
1) LabTower AFT pure water system 15) Pure water connector/Ion exchanger
2) Front hood, removable 16) Prefiltration
3) Rotary knob for pure water dispens- 17) Sterile filter or fine filter (option)
ing 18) Power supply for booster pump/tank,
4) Sterile capsule filter (option) connector level control
5) Operating unit 19) Power supply 24/48V
6) Front cover, storage tank 20) Printer connection
7) Feedwater inlet 21) connector level control
8) Concentrate outlet 22) Pure water connector, (from storage
9) Pure water outlet tank), Ø8 mm
10) Residual water drainage 23) Permeate connector, Ø8 mm
11) Pressure reducer 24) Raw water connector, Ø8 mm
12) Pressure gauge, pure water pressure 25) Concentrate connector, Ø8 mm
13) Ion exchanger 26) Pure water connector, (to consumer),
14) Feedwater connector/ion exchanger Ø8 mm
19
Thermo Scientific Barnstead LabTower AFT
Pure water system
1. Place the pure water system on the table of the storage tank. Connect the hoses positioned
at the table by putting the connectors in with the appropriate connections (21-25) at the
equipment (see „Views and connectors “, side view).
3. Turn the filter housing of the pre-treatment (16) loosely. Insert the pre-treatment unit con-
tained when assembling (pre-filter + hardness stabilization) into the filterhousing and screw
this again firmly (see chapter „Change of the pre-treatment “).
4. In the case of use of the optional sterile or fine filter you turn the filter housing (17) loosely.
Insert the filter into the filter housing and bolt this on again.
5. Connect the Ø8 – ¾“ female thread feedwater hose supplied with feedwater inlet (7) of the
storage tanks and a water tap that can be turned off. The connecting hose is marked with
„Rawwater“.
6. Connect Ø8 concentrate hose supplied with concentrate outlet (8) of the storage tank and
to the drain provided on-site. The concentrate hose is marked with „Concentrate“.The drain
to the stewer must be max. Are 1 m above the rinsing water connector of the unit.
Caution! Concentrate must be able to flow under free gravity fall to the drain!
7. Place the ion exchanger (13) into the hollow of the storage tank.
8. Press the quick-connector of the raw water hose (14) on connection of your ion exchanger,
marked with „Rawwater“.
9. Press now the quick-connector of the pure water hose (15) on connection ( of your ion ex
changer marked with „Purewater“.
The quick-connector must engage audibly.
Do not get the raw water and pure water hoses mixed up!
10. Plug the 4-pin plug of the level control system (storage tank), marked with „Niveau“, in the
4-pin socket (20) of the pure water systems and screw the milled nut of the plug. tight.
11. Screw the sterile filter capsule (4) into the withdrawal tap G1/4 “.
12. Now fit the hose of the Ø8 mm sterile overflow in the overflow on the back of the tank and
connect it with the drain.
14. Check the feedwater pressure; it must be within the permissible pressure range
(see chapter „Technical specifications”).
Before you now switch the system on, read through the rinsing out
procedure for reverse osmosis membranes packed with preserving
agent in the „Rinsing the membrane“ section!
20
Thermo Scientific Barnstead LabTower AFT
Pure water system
17. Switch the pure water system on with the on/off key on the operating unit.
21
Thermo Scientific Barnstead LabTower AFT
Pure water system
Whenever possible, mount the power pack on the wall to the left or right of the pure
water system where it is freely accessible.
Stick the universal holder which is supplied in the assembly kit to the back of the
power pack as shown in the above Figure.
Stick the universal adapter to a smooth wall surface or screw it to the wall using the
dowels and screws supplied in the assembly kit.
When the universal holder and universal adapter have been fitted, hang the power
pack in.
Plug the connecting cable (appliance cable) in the power pack socket.
Connect the power pack to the pure water system (24/48V DC 4-pin power
22
Thermo Scientific Barnstead LabTower AFT
Pure water system
Open the vent valve on the central closing lid of the ion exchanger (11) by turning it anti-
clockwise.
Open the raw water tap and wait until water emerges from the vent valve.
23
Thermo Scientific Barnstead LabTower AFT
Pure water system
Production 70%
10.0 MΩxcm
18.5°C UV/TC
24
Thermo Scientific Barnstead LabTower AFT
Pure water system
To simplify changing settings, a press on the UV-key allows you to select a certain number in a
numerical value that you wish to change. The arrow-keys can then be used to set a number
from 0-9 at this selected position.
A single press on the menu-key allows the permeate conductivity to be read and the limiting
value for the permeate conductivity to be set. Should this limiting value be exceeded, then the
"Lim.val.permeate" fault message is displayed flashing in the 4th line of the display.
(Measuring point LF 2)
Settings above 150.0 µS/cm result in the limiting value being switched off. The word
"Off".appears in the display.
Permeate
0.083 MΩxcm
Lim.val.permeate
50.0 µS/cm
25
Thermo Scientific Barnstead LabTower AFT
Pure water system
A second press on the menu-key in this menu allows the ultrapure water limiting value to be set.
Should this limiting value be exceeded, then the "Lim.val.ultrapure" fault message is shown
(Measuring point LF 1).
Lim.val.ultrapure
2.0 µS/cm
A third press on the menu-key in this menu allows the operating hours of the UV-lamp and the
reverse osmosis pump to be displayed. The operating hours counter for the UV-lamp registers
the total length of time for which the lamp was switched on. When the maximum operating time
is reached, the "UV-time" fault message is triggered. The limiting value is set in the OEM menu.
The operating hours of the reverse osmosis pump does not have a limiting value.
UV-time
0000 h
RO-time
0000 h
26
Thermo Scientific Barnstead LabTower AFT
Pure water system
A fourth press on the menu-key in this menu brings the operating hours of the pre-treatment
cartridge to display.
The limiting value fort his operating time iss et in the UV menu. The fault message that is dis-
played when the limiting value is exceeded is "Pretreatment".
The operating hours of the pretreatment are counted whenever the reverse osmosis pump is
running.
Pretreatment
0000 h
14.1.5 Cleaning:
A fifth press on the menu-key in this menu allows cleaning to be carried out when there is a
need for it. The cleaning process is triggered by pressing the Enter-key. The pump then starts
and the inlet solenoid valve and the rinsing solenoid valve open for a period of 60 seconds.
During cleaning, no faults or measured values are displayed. When the cleaning process has
finished, the system is in the last operating mode (operating or stand-by).
The remaining cleaning time is displayed during cleaning.
Cleaning ?
Press Enter !
Cleaning
30 sec.
27
Thermo Scientific Barnstead LabTower AFT
Pure water system
14.1.6 Disinfection:
(not active in this version)
Disinfection
Press Enter !
A seventh press on the menu-key in this menu calls the fault storage inquiry. Confirmation of
this with the Enter-key allows the fault storage to be looked through. The display can show two
faults at once, each with time and date. Pressing an arrow key allows previous or following
faults to be displayed.
Pressing the menu-key or the Enter-key returns the system to the last operating mode.
Fault storage
Press Enter!
14.03.04 14.30
Lim.val.permeate
14.03.04 15.30
Pretreatment
28
Thermo Scientific Barnstead LabTower AFT
Pure water system
An eighth press on the menu-key in this menu brings you to the “Code” menu. To prevent unau-
thorized access to the settings in the system control, changes to the settings can only be carried
out when the correct code from the Assignment Table that follows is entered and confirmed with
the Enter-key. The unlocking remains active for 5 minutes. Each access via the code is typed
out by the printer (RS 232), complete with date, time and abbreviated code number (“Code
0001” corresponds to code 150, “Code 0002” to code 250 etc.).
Code
Press Enter !
0000
29
Thermo Scientific Barnstead LabTower AFT
Pure water system
Assignment Table
for code numbers allowing the system to be unlocked
Printer
Code no. Person
output
150 0001
250 0002
350 0003
450 0004
550 0005
650 0006
750 0007
850 0008
950 0009
30
Thermo Scientific Barnstead LabTower AFT
Pure water system
14.2 OEM-menu
Basic settings and limiting values can be changed in this menu.
To make changes in the OEM menu , the system control must previously be unlocked (see
12.1.8).
OEM menu
Press Enter !
IThe maximum temperature to which the system can be exposed can be set in this menu. When
this temperature is exceeded, the "max. Temperature" fault message is triggered. Settings
above 50 °C cause the limiting value to be suppressed and the word „off“ appears in the dis-
play. It is shown in the 4th line of the display.
Basic setting: 50 °C
Setting range: 1 - 50 °C
OEM menu
max. Temperature
35 °C
31
Thermo Scientific Barnstead LabTower AFT
Pure water system
OEM menu
Disinfect. time
30 min.
OEM menu
Recirc. time
15 min.
OEM menu
Rinsing time
0.5 sec.
32
Thermo Scientific Barnstead LabTower AFT
Pure water system
The interval rinse time iss et in this menu. Rinsing is carried out for this time period when the
operating mode is changed, between stand-by and operation and every 12 hours.
OEM menu
Interv. Rinsing time
2 sec.
OEM menu
Day 30 Month 12
Year 2004
Hour 12 min.30
In this menu, the sending interval for transmissions of measured values and fault messages to
the RS 232 interface is set.
OEM menu
Sending interval
8h
33
Thermo Scientific Barnstead LabTower AFT
Pure water system
14.2.8 Language:
An eighth press on the menu-key:
The language in which texts are displayed iss et in this menu. Choice of English, German or
French.
OEM menu
Language
German
In this menu, a choice can be made as to whether the specific electric resistance or the conduc-
tivity is to be displayed.
OEM menu
µS/cm / MΩxcm
MΩxcm
Basic setting: on
Setting range: on, off
OEM menu
Temperature comp.
on
34
Thermo Scientific Barnstead LabTower AFT
Pure water system
With a setting over 100 % for the upper switching point, the display of the tank level is switched
off, so that this setting therefore shows whether an analogue or a digital float switch is installed.
OEM menu
Off: 100 %
On: 70%
The grade of the equipping of the system can be set in this menu to differentiate between UPW
and PW versions.
OEM menu
Equipping grade
TII UV
In this menu, the type and serial number of the system can be entered. They are then printed as
headline on every print-out. The following types of system can be entered:
LabTower RO / LabTower TII / LabTower TII UV / LabTower AFT
OEM menu
Type: LabTower AFT
S.Nr.: 9999/04
35
Thermo Scientific Barnstead LabTower AFT
Pure water system
- Standard message
- Code message
- Fault message
Here a record opf all measured values is printed out in dependence on the sending interval.
Such a complete dats set is also printed out during NONSTOP operation.
Print out:
Whenever a code number is entered in the controls and confirmed with the Enter-key, then the
code input is immediately printed out.
For code identification, see the „Assignment Table for code numbers allowing the system to be
unlocked”.
Print-out:
Should a fault message be shown in the display, e.g. the ultrapure water limiting value, then the
fault message is printed out at the end of the sending interval.
Print-out:
36
Thermo Scientific Barnstead LabTower AFT
Pure water system
37
Thermo Scientific Barnstead LabTower AFT
Pure water system
15. Maintenance
We recommend that you secure a service contract to ensure that the necessary maintenance
work is carried out.
You then have the certainty of a high operational safety and reliability.
The service protocol appended to the service contract serves for certification that maintenance
work specified in the contract has been carried out by authorized service company.
To ensure your system will work reliably for a long time, it must be checked, serviced and cared
for at regular time intervals in accordance with these Operating Instructions!
For this reason, the operating instructions must be readily available to operating and mainte-
nance staff at all times, and be carefully followed!
Any maintenance work which should become necessary during the validity of the guarantee is
only to be carried out by a service professional which is expressly authorized to do such work.
The calibration of the conductivity display is only to be carried out and recorded by customer
service.
Cleaning and disinfection of supply tanks, piping, filter housings etc. Is performed for reasons of
hygiene and has no effect on the technical condition of the system. These components must be
cleaned and disinfected whenever algae or slime are detected inside them or at least once
yearly.
38
Thermo Scientific Barnstead LabTower AFT
Pure water system
Consumable materials are to be replaced at the intervals given in the following Table or when
there is a drop in performance:
Pretreatment 09.4001 F1
Prefilter/Hardness stabilization 06.5204 6 Months
Pretreatment 09.4000 F1
Prefilter 06.5201 6 Months
Hardness stabilization 06.5452 6 Months
Filter cartridge F3 09.4011/09.4012 12 Months
*Please keep in mind that the life of your consumables is directly dependent on the quality of
the feed water and the amount of the water used daily. The interval is contingent on the feedwa-
ter quality so that a shorter one may be necessary.
According to the mode of delivery, the system may be supplied filled with a solution containing a
preservative. Should this be the case, it is important that the permeate be run to drain for at
least 3 – 4 hours when the system is first put into operation and switched to production.
To do this, each time the tank has been filled, open the dispensing tap where your water is
withdrawn and lead the pure water that flows out to drain.
Automatic rinsing lengthens the service life of your reverse osmosis module. Coarse particles
and other contaminants are swept away from the membrane surface during this rinsing process.
The rinsing phases so ensure the longest service and optimal purified water quality.
Leave your pure water system on over the weekend and during
holiday times. Only then can the 12-hourly rinse operate and ensure that
your reverse osmosis module is not subject to bacterial growth during
standstill periods.
Should the performance of your module drop because operation of the system without appro-
priate pretreatment has resulted in blockage of the membrane, it may well be that your module
can be reconditioned.
Reconditioning and disinfection of the module is only to be carried out by authorized service
personnel on-site or by sending the module to the manufacturer of your system, whereby expo-
sure to frost must be avoided.
39
Thermo Scientific Barnstead LabTower AFT
Pure water system
*
Hinweis:
Notice:
2. Open the residual water drainage on the storage tank, see chapter „Views and connectors”,
till the system starts running.
4. Switch the pure water system off with the on/off key on the operating unit.
6. Unscrew the filter housing from the pre-treatment, see chapter „View and connectors“.
Caution! The filter housing is filled with water.
8. Plug the new pre-filter cartridge and hardness stabilizer with the connector together,
see drawing.
40
Thermo Scientific Barnstead LabTower AFT
Pure water system
Before taking off the quick connectors (12, 13) from the ion exchanger (11),
the pressure in this must first be released by switching the system off so that
no water flows out.
3. Take off the quick connectors for raw water (12) and pure water (13) from the ion exchang-
er (11).
4. Remove the exhausted ion exchanger (11) from the storage tank depression and exchange
it for a new ion exchanger.
5. Plug the raw water and pure water quick connects (12, 13) back on the ion exchanger (11).
Do not mix up the the quick connectors! Make sure that they audibly click into posi-
tion!
7. Vent the ion exchanger (see the ”Venting the ion exchanger“ section).
41
Thermo Scientific Barnstead LabTower AFT
Pure water system
Open the cap nut (3) on the pressure tube and remove the RO-membrane (2).
Insert the new RO-membrane in the pressure tube (1) with the two O-rings (4) on the
permeate tube of the RO-module in the direction shown by the arrows.
Inserting the RO-membrane the wrong way round would result in certain
destruction of the membrane!
42
Thermo Scientific Barnstead LabTower AFT
Pure water system
15.6 Disinfection
Disinfection of storage tank
Use MICRO-Chlor Granulate, article no. 09.2202 as disinfectant (Europe only), or Cleaning So-
lution, article no. 50129891 (US-market only).
Please observe the information given in the safety data sheet supplied
with Micro-Chlor disinfectant to avoid possible health hazards!
2. The pure water system must be in the “Stand by” operating mode.
The storage tank must be filled.
3. Switch the system off and unplug the power supply from the pure water system.
5. Open the dispensing valve to bleed the pure water system completely and then close
it again.
7. Remove the four connectors from the pure water system to the storage tank and pull
the 8 pin connector (float switch, pressure pump, (see Chapter 11. Putting the system
into operation).
8. Lift the system from the storage tank and put it aside.
10. Open the lid of the storage tank, pour the contents of a MICRO-Chlor container or the
disinfectant Cleaning Solution in the water-filled tank and then close the lid.
Do not take any water from the dispensing valve or the storage tank during
disinfection.
43
Thermo Scientific Barnstead LabTower AFT
Pure water system
14. Now open the residual drainage of the storage tank and discard all of the water.
Close the residual drainage.
16. Replace the pure water system back to the storage tank and connect the four
connectors for power supply and the 8 pin connector (float switch, pressure
pump) to the storage tank.
19. Fill the storage tank until the “Stand by” operating mode has been reached and
empty the tank as in 11. Repeat this process a second time.
21. The pure water system is now again at your availability to produce pure water.
44
Thermo Scientific Barnstead LabTower AFT
Pure water system
The system is not to be thrown away as household waste but must be properly disposed of. It
can be returned to the manufacturer for safe disposal according to EEC Guideline 2002/96/EC.
We therefore request our customers in Germany and other member States in the European
Economic Area to contact our local service centre or our headquarters:
WEEE-Reg.-no.: DE 12471402
In countries outside of the European Economic Area, please contact your local authorities or
waste disposal company.
45
Thermo Scientific Barnstead LabTower AFT
Pure water system
The system does not start - No power supply - Connect to power supply
System control no longer - Improper operation - Remove line plug for 5 sec
reacts onds
Water flows out - Leaky hose connection - Check and hose connection
and stop the leak
- Feedwater temperature
fluctuates
46
Thermo Scientific Barnstead LabTower AFT
Pure water system
- The limiting value is set too - Check and re-adjust the limit
low ing value setting
Fault message: - The limiting value is set too - Check and re-adjust the limit
"Lim,value.ultrapure" low ing value setting
Fault message: - The operating hours of the - Replace the UV-Lamp (art.-
"UV-time" UV-lamp have been ex no. 09.4002) and set back
ceeded the operating hours counter
Fault message: - Cable break at the measure - Replace the measuring cell
"Measuring cell LF1" ing cell
Fault message: - Cable break at the measure - Replace the measuring cell
"Measuring cell LF2" ing cell
Fault message: - Cable break at the measure - Replace the measuring cell
"Measuring cell Temp." ing cell
47
Thermo Scientific Barnstead LabTower AFT
Pure water system
India
Sales: +91 22 6716 2200
Service: Toll free 1 800 22 8374 or +91 22 6716 2200
Japan
Sales: +81 45 453 9220
Service: +81 45 453 9224
48
Thermo Scientific Barnstead LabTower AFT
Pure water system
Belgium
Sales: +32 53 73 4241
Service: +32 53 73 4241
Finland/Nordic/Baltic countries
Sales: +358 9 329 100
Service: +358 9 329 100
France
Sales: +33 2 2803 2180
Service: +33 825 800 119
Germany:
Postal Address Germany:
Thermo Electron LED GmbH
Robert-Bosch-Straße 1
D – 63505 Langenselbold
Phone
Sales: Toll free 0800 1 536 376
or +49 6184 90 6940
Service: Toll free 0800 1 112110
or +49 6184 90 6940
E-Mail: [email protected]
Italy
Sales: +39 02 95059 341
Service: +39 02 95059 250
Netherlands
Sales: +31 76 579 5555
Service: +31 76 579 5639
Russa/CIS
Sales: +7 812 703 4215
Service: +7 812 703 4215
Spain/Portugal
Sales: +34 93 223 0918
Service: +34 93 223 0918
Switzerland
Sales: +41 44 454 1212
Service: +41 44 454 1212
UK/Ireland
Service: +44 870 609 9203
Sales: +44 870 609 9203
49
Thermo Scientific Barnstead LabTower AFT
Pure water system
If the system or the tank show a fault, then the water supply can be maintained by changing the
3-Way ball tap over to „emergency supply “.
1. Turn the 3-Way ball tap for the emergency supply on the position „Emergency supply”.
2. Adjust the pressure reducer to the permissible pure water pressure of the consumer,
(2 bar).
50
Thermo Scientific Barnstead LabTower AFT
Pure water system
3. Pull out the adjusting knob of the pressure reducer. The pure water pressure is reduced by
turning this knob anti-clockwise and increased by turning it clockwise.
The pressure change can be read off from the manometer.
4. After having adjusted the pure water pressure, press in the adjusting knob of the pressure
reducer to lock it.
The system can only be used for a limited time in emergency mode, i.e. until the capacity of the
ion exchanger is exhausted.
One ion exchanger should always be kept available as reserve.
Please inform Service immediately when you switch your system to emergency supply.
The address to contact should you need service: see chapter „Trouble shooting”.
51
Thermo Scientific Barnstead LabTower AFT
Pure water system
52
Thermo Scientific Barnstead LabTower AFT
Pure water system
We ask for your understanding that the guarantee we give becomes invalid when re-
placement parts, accessories or consumables from other manufacturers are used, as we
have no influence on their appropriateness or quality.
53
Thermo Scientific Barnstead LabTower AFT
Pure water system
20. Accessories
21 Consumables
Pre-treatment
Pre-filter cartridge 5 µm + hardness stabilizer 10" 06.5204
54
Thermo Scientific Barnstead LabTower AFT
Pure water system
55
Thermo Scientific Barnstead LabTower AFT
Pure water system
56
Thermo Scientific Barnstead LabTower AFT
Pure water system
Type of system:
Serial no.:
Year made:
The following points are to be observed to ensure the quality of the system:
57
Thermo Scientific Barnstead LabTower AFT
Pure water system
North America:
USA/Canada +1 866 984 3766 (866-9-THERMO)
Europe:
Austria +43 1 801 40 0, Belgium +32 53 73 42 41, France +33 2 2803 2180,
Germany national toll free 08001-536 376,
Germany international +49 6184 90 6940, Italy +39 02 95059 448,
Netherlands +31 76 579 55 55, Nordic/Baltic/CIS countries +358 9 329 10200,
Russia +7 812 703 42 15, Spain/Portugal +34 93 223 09 18,
Switzerland +41 44 454 12 12, UK/Ireland +44 870 609 9203
Asia:
Australia +61 39757 4300, China +86 21 6865 4588 or +86 10 8419 3588,
India toll free 1800 22 8374, India +91 22 6716 2200,
Japan +81 45 453 9220, New Zealand +64 9 980 6700,
Other Asian countries +852 2885 4613
© 2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fish-
er Scientific Inc. and its subsidiaries. Specifications, terms and pricing
are subject to change. Not all products are available in all countries. Please consult your local sales
representative for details.