Calibration Certificate

Calibration Certificate No.:EISL/Cal-06 Issue Date: 24.10.2021 Page No.: 1 of 1


Date of Calibration: 16.10.2021 Recommended Calibration Due On: 15.10.2022

Calibrated for : Day Apparels Ltd


21, MH Tower,Vogra,Gazipur,Bangladesh

Description of instrument to be calibrated : GSM Cutter


Make: N/P
Model: N/P

Calibration Location & Instrument Receipt Date : At Site (16.10.2021)

Identification of Instrument : N/P


Traceability and its reference certificate :
Description Traceability Certificate No. Calibration Date
Valid upto Calibration Agency
06.06.202
Digital Caliper CE-21-10/069/A27/2702 07.06.2021 CASTRO
2
Traceability: Standards used for calibration are traceable to National Standards through NABL accredited laboratory
Environmental conditions : Temperature: 23°C±2°C Humidity: 50±10%RH
Illumination: 250 to 500 lux, Noise: <60 dB
Principle / Methodology of calibration : As per Work Instruction (WI- 01).

Calibrated By : M. Abdullah

OBSERVATION:
Measurements: (verified against Digital Caliper)

As Found
SL. Measurement d (mm) Calculated Area (cm²) Standard Area(cm²) Uncertainty (cm²)
No.
01 112.67
02 112.66 99.70 100 ±0.050
03 112.67

Area Calculated using the mean diameter

Note: i) DUC: Device under Calibration.


ii) The results in the certificate are valid at the time of measurement under stated conditions.
iii) The certificate if produced for any purpose commercial or otherwise should be reproduced in full. Partial of the certificate or an
abstract thereof has to be got specifically approved from the technical and the quality manager of this job.

This certificate, issued by Eurosia ITC Services Ltd refers only the particular item submitted to calibrate.

…….. End of Certificate ……..