DEF-7000

Defibrillator Monitor
(AED and Manual)

Technical specifications
Display
Display Type: 8 inches High-resolution LCD display.
Sweep Speed: 25mm/sec.
Information: Heart Rate, Lead/Pads, Alarm On/Off, Sp02, AED Functions and Prompts, Alarm Selection
and Limits, Delivered Energy.

Defibrillator
Waveform: Biphasic
Charge Time: Less than 7 seconds with a new fully charged battery.
Energy Display: Monitor display indicates both selected and delivered energy.
Charge Prompt Type: Voice and visual prompts.
Electrode Impedance Measurement Range: 0-250 ohms.

AED Mode
AED Function: Auto analyzes and charges x3 with programmable auto energy level selection, screen
prompts, and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude≥200UV, Ventricular tachycardia with rates≥
140bpm, and QRS complex wave duration≥140ms.
Charge Control: Control on device front panel, press key on paddle.
Prompts: Voice and visual prompts.

1
Manual Mode
Energy Selection: Selectable at 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 Joules.
Synchronized Mode: Synchronizes defibrillator pulse to patient's R-wave. "SYNC" message displayed
on monitor.

ECG Monitoring
Patient Connection: 5-lead ECG cable, or 3-lead ECG cable, paddles.
Lead Selection: Displayed on monitor, paddles, l, II, III, AVR, AVL, AVF, V.
ECG Size: 0.25, 0.5, 1, 1.5, 2, 4 cm/mV display on monitor.
Heart Rate: 20-300BPM.
Heart Rate Alarm: On/Off displayed on monitor, user-selectable.
Smart Alarms: Beeper/Voice prompts indicate shockable rhythm.

SpO2 Module
Measurement Range: 30-100%, +2% between 80%-90%, others + 5%
Alarm Range: User set high limit and low limit
Alarm Accuracy: +10% within setting values
Alarm Time Accuracy: Less than 12 sec

Recorder
Paper: 50mm thermal.
Speed: 12.5mm/sec, 25mm/sec, 50mm/sec. User-selectable 6-second delay.
Printing Method: High-resolution, thermal array print head.
Print-out Modes: Manual or automatic, user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-second recording initiated by alarm activation or defibrillator charge or defibrillator
discharge.

Battery
Type: Rechargeable, Ni-MH battery, 12V.
Operating Time: For a new, fully charged battery: 60 defibrillator discharges at maximum energy, or 3
hours minimum of continuous ECG monitoring.
Additional parameters will affect operating time with different functions.

Option 1
Specification Pacer (Optional)
Type VVI
Pulse Type Rectangular, constant current
Pulse Amplitude 0 to 180 mA±10% or 5 mA (whichever is greater)
Variable from 30ppm to 180ppm±1.5% (increments or decrements
Pacing Rate
by a value of 2ppm)
Multi-Function Electrode
Multipurpose defibrillator/pacing electrodes
(MFE) Pads
Pause Pacing pulse frequency reduced by a factor of 4 when activated
Refractory Period NC (VVI demand do not have this specification)

2
Pulse Width (20+1.5ms)
Output Protection Fully defibrillator protected and isolated

Option 2
Specification EtCO2 (Optional)
Type of sensor By-pass
Technical principle Non-dispersive infrared gas analysis NDIR
Storage condition -40℃ to 70℃, <90℃ RH, non-condensing
Operating conditions 5℃ to 50℃, 10 to 90℃ RH, non-condensing
Ambient pressure 55-115kPa
Power supply 5V±5% (max ripple 200mVp-p)
TDP Typical value 120mA Excursion calibration typical value 280mA
Range 0-19.7% (0-150mmHg, or 0-20Kpa)
Resolution 0.1mmHg
0-40mmHg±2mmHg
41-70mmHg±5% of reading
Accuracy
71-100mmHg±8% of reading
101-150mmHg±10% of reading
Respiratory rate 2-150BPM
Respiratory rate
1%±1BPM
measurement accuracy

Option 3
Specification NIBP (Optional)
Measurement unit mmHg/kPa
Adult: 10~200mmHg/kPa
Measurement Range Pediatric: 10~200mmHg/kPa
Neonatal: 10~135mmHg/kPa
Resolution 1mmHg
Maximum mean error: 5mmHg
Accuracy
Maximum standard deviation: 8mmHg
Type of sensor By-pass
Technical principle Non-dispersive infrared gas analysis NDIR
Storage condition -40℃ to 70℃, <90% RH, non-condensing
Operating conditions 5℃ to 50℃, 10 to 90% RH, non-condensing
Ambient pressure 55-115kPa
Power supply 5V±5% (max ripple 200mVp-p)
TDP Typical value 120mA Excursion calibration typical value 280mA
Range 0-19.7% (0-150mmHg, or 0-20kPa)
Resolution 0.1mmHg
0-40mmHg±2mmHg
Accuracy 41-70mmHg±5% of reading 71-100mmHg±8% of reading
101-150mmHg±10% of reading

3
Respiratory rate 2-150BPM
Respiratory rate
1%±1BPM
measurement accuracy