0% found this document useful (0 votes)
20 views57 pages

CN97010000-530 3502-2000008 V1.2

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
20 views57 pages

CN97010000-530 3502-2000008 V1.2

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 57

Patient Monitor

User Manual
(Part I)

Shenzhen Creative Industry Co., Ltd


User manual for Patient Monitor

This Manual is written and compiled in accordance with the IEC 60601-1 (Medical electrical equipment Part1:
General requirements for safety), and MDD 93/42/EEC. It complies with both international and enterprise
standards and is also approved by State Technological Supervision Bureau. The Manual is written for the current
Patient Monitor.
The Manual describes, in accordance with the Patient Monitor’s features and requirements, main structure,
functions, specifications, correct methods for transportation, installation, usage, repair, maintenance and storage,
etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective chapters for
details.
This manual is applicable to the product series for all our Patient Monitors. It consists of two parts, part I covers
almost the all necessary content, except the operations for user interface, which is included separately in part II.

The Manual is published in English and we have the ultimate right to explain the Manual. No part of this manual
may be photocopied, reproduced or translated into another language without the prior written consent. We reserve
the right to improve and amend it at any time without prior notice. Amendments will however be published in a
new edition of this manual.
This manual is an integral part of the product, it should always be kept close to the equipment so that it can be
obtained conveniently when needed

Version of this manual: V1.2


Date: December 26, 2016
All rights reserved.

Marks in the Manual:

 Warning: must be followed to avoid endangering the operator and the patient.
☞ Note: contains some important information and tips about operations and
application.
 Attention: must be followed to avoid causing damage to the monitor.
Caution:
Federal law restricts this device to sale by or on the order of a physician.

3502-2000008

-I-
User manual for Patient Monitor

Instructions to User
Dear Users,
Thank you very much for purchasing our product. Please read the following very carefully
before using this equipment.
Read these instructions carefully before using this monitor. These instructions describe the
operating procedures to be followed strictly. Failure to follow these instructions can cause
monitoring abnormality, equipment damage and personal injury. The manufacturer is NOT
responsible for the safety, reliability and performance issues and any monitoring
abnormality, personal injury and equipment damage due to user’s negligence of the
operation instructions. The manufacturer’s warranty service does not cover such faults.
 WARNING for PACEMAKER PATIENTS: Although the pacemaker pulse
inhibition function is available in this device, the heart rate meter may
continue to count the pacemaker rate during occurrences of cardiac arrest or
some arrhythmias. Do not rely entirely upon rate meter ALARMS. Keep
pacemaker patients under close surveillance. See this manual for disclosure of
the pacemaker pulse rejection capability of this instrument.

 If uncertain about the accuracy of any measurement, first check the patient’s
vital signs by any alternative means, and then make sure the monitor is
functioning properly.

 The device should be considered an early warning device. As a trend towards


patient deoxygenation is indicated, blood samples should be analyzed by a
laboratory CO-oximeter to completely understand the patient's condition.

 The monitor is not suitable for use in the presence of flammable anesthetic
mixture with air, oxygen or nitrous oxide.

 Monitoring a single person at a time.

 The monitor is defibrillator proof. Verify that the accessories can function
safely and normally and the monitor is grounded properly before conducting
defibrillation.

 Disconnect the monitor and sensors before MRI scanning. Use during MRI
could cause burns or adversely affect the MRI image or the monitor’s
accuracy.

 If you have any doubt to the grounding layout and its performance, you must
use the built-in battery to power the monitor.

 All combinations of equipment must be in compliance with the standard IEC 60601-1.

 Check SpO 2 probe application site periodically (every 30 minutes) to


determine circulation, positioning and skin sensitivity.

 The SpO 2 measurement of this monitor may not work for all testees. If stable
readings cannot be obtained at any time, discontinue use.

 Do not immerse the monitor or its accessories in liquid to clean.

 Do not use accessories other than those provided/recommended by the

-II-
User manual for Patient Monitor

manufacturer.

 Each time the monitor is used, check the alarm limits to ensure that they are
appropriate for the patient being monitored.

 Do not silence the audible alarm if patient safety may be compromised.

 The alarm limit value shall be within the measuring range, or it may disable
the alarm system. Please refer to the related chapter for alarm limit range.

 A HAZARD can exist if different alarm presets are used for the same or
similar equipment in single area.

 The monitor is intended only as an adjunct in patient assessment. It must be


used in conjunction with clinical signs and symptoms.

 When taking the measure of a pediatric or neonates(less than 10 years old) blood pressure,
do NOT operate in the adult mode. The high inflation pressure may cause lesion or even
body putrescence.

 The monitor is prohibited from applying to those who have severe hemorrhagic tendency
or who are with sickle cell disease for they may develop partial bleeding when this
monitor is used to take the blood pressure measurement.

 DO NOT take blood pressure measurement from a limb receiving ongoing transfusion or
intubations or skin lesion area, otherwise, damages may be caused to the limb.

 Continuous use of SpO2 sensor may result in discomfort or pain, especially for those with
microcirculatory problem. It is recommended that the sensor should NOT be applied to
the same place for over two hours, change the measuring site periodically if necessary.
 SpO2 measuring position must be examined more carefully for some special patient. Do
NOT install the SpO2 sensor on the finger with edema or vulnerable tissue.

 To prevent the risk of the short circuit and to ensure the ECG signal quality, the equipment
must be properly grounded.

 Although biocompatibility tests have been performed on all the applied parts, some
exceptional allergic patients may still have anaphylaxis. Do NOT apply to those who have
anaphylaxis.

 All the connecting cables and rubber tubes of the applying parts should be kept away from
the patient's cervix to prevent any possible suffocation of the patient.

 All the parts of the monitor should NOT be replaced at will. If necessary, please use the
components provided by the manufacturer or those that are of the same model and
standards as the accessories along with the monitor which are provided by the same
factory, otherwise, negative effects concerning safety and biocompatibility etc. may be
caused.

 DO NOT stare at the light of SpO2 sensor (infrared is invisible) when switch it on, for the
infrared may do harm to the eye.

 If the monitor falls off accidentally, please do NOT operate it before its safety and technical
indexes have been tested minutely and positive testing results obtained.
 The system might not meet its performance specifications if stored or used outside the
manufacturer’s specified temperature and humidity ranges.

-III-
User manual for Patient Monitor

 Reuse, disassembly, cleaning, disinfecting the single patient use CO2 cannula kits and
on-airway adapters may compromise functionality and system performance leading to a
user or patient hazard. Performance is not guaranteed if an item labeled as single
patient use is reused.

 Electrical Shock Hazard: Always disconnect the CO2 Sensor before cleaning. Do NOT
use if it appears to have been damaged. Refer servicing to qualified service personnel.

 Electrical Shock Hazard:No user serviceable parts inside the CO2 Sensor.
 After the life cycle of the Sidestream CO2 Sensor and its accessories has been met,
disposal should be accomplished following national and/or local requirements.

 Please peruse the relative content about the clinical restrictions and contraindication.

 The accessories which can be used repeatedly should have a thorough cleanness before it is
used to another patient. Please refer to the related chapter for maintenance method.

 When disposing of the monitor and its accessories, the local law should be followed.

 Substitution of a component different from that supplied might result in measurement


error.

-IV-
User manual for Patient Monitor

Table of Contents
Chapter 1 Overview........................................................................................................................................... 1
1.1 Product Name.............................................................................................................................................. 1
1.2 Applications and Scope............................................................................................................................... 1
1.3 Operating Environment............................................................................................................................... 1
1.4 Impact on Environment and Resources....................................................................................................... 1
1.5 Safety........................................................................................................................................................... 1
Chapter 2 Working Theories............................................................................................................................. 3
2.1 Overall Structure......................................................................................................................................... 3
2.2 Composition................................................................................................................................................ 3
2.3 Working Theories........................................................................................................................................ 3
Chapter 3 Installation and Connection............................................................................................................. 5
3.1 Installation................................................................................................................................................... 5
3.1.1 Opening the Box and Check............................................................................................................. 5
3.1.2 Connecting the AC Power Cable...................................................................................................... 5
3.1.3 Starting the Monitor.......................................................................................................................... 6
3.2 Connection................................................................................................................................................... 6
3.2.1 ECG Cable/Lead Wires Connection................................................................................................. 6
3.2.2 Cuff Connection for Blood Pressure Measurement.......................................................................... 8
3.2.3 SpO2 Probe Connection.................................................................................................................. 11
3.2.4 CO2 Sensor Connection.................................................................................................................. 15
3.2.5 TEMP Probe Connection................................................................................................................ 19
3.2.6 Loading Printing Paper................................................................................................................... 19
Chapter 4 Alarm............................................................................................................................................... 21
4.1 Alarm Description..................................................................................................................................... 21
4.1.1 Alarm Condition............................................................................................................................. 21
4.1.2 Alarm Priority................................................................................................................................. 21
4.1.3 Alarm Modes.................................................................................................................................. 22
4.1.4 Alarm Setting.................................................................................................................................. 22
4.2 Alarm Technical Specifications................................................................................................................. 22
Chapter 5 Technical Specifications................................................................................................................. 24
5.1 ECG Monitoring........................................................................................................................................ 24
5.2 RESP Monitoring...................................................................................................................................... 25
5.3 TEMP Monitoring..................................................................................................................................... 25
5.4 NIBP Monitoring....................................................................................................................................... 25
5.5 SpO2 Monitoring....................................................................................................................................... 26
5.6 Pulse Rate Monitoring............................................................................................................................... 26
5.7 CO2 Monitoring......................................................................................................................................... 26
5.8 Data Recording.......................................................................................................................................... 27
5.9 Other Technical Specifications.................................................................................................................. 27
5.10 Classification........................................................................................................................................... 27
5.11 Guidance and manufacturer’s declaration-Electromagnetic compatibility.............................................. 28
Chapter 6 Packaging and Accessories............................................................................................................. 32

-V-
User manual for Patient Monitor

6.1 Packaging.................................................................................................................................................. 32
6.2 Accessories................................................................................................................................................ 32
Chapter 7 Working Principles......................................................................................................................... 33
7.1 Introduction to ECG Measurement............................................................................................................ 33
7.1.1 How to Obtain High Quality ECG and Accurate Heart Rate Value................................................33
7.1.2 Factors affecting ECG signal.......................................................................................................... 33
7.2 Introduction to Blood Pressure Measurement........................................................................................... 34
7.2.1 Blood Pressure Measuring Principle.............................................................................................. 34
7.2.2 Factors affecting NIBP measuring.................................................................................................. 35
7.2.3 Clinical Limitations........................................................................................................................ 36
7.3 Introduction to Oxygen Saturation Measurement...................................................................................... 36
7.3.1 SpO2 Measuring Principle.............................................................................................................. 36
7.3.2 SpO2 Measurement Restrictions (interference reason)................................................................... 37
7.4 Introduction to Respiration Measurement................................................................................................. 37
7.4.1 Respiration Measuring Principle.................................................................................................... 37
7.4.2 Factors affecting respiration monitoring......................................................................................... 38
7.5 Introduction to Temperature Measurement................................................................................................ 38
7.6 Introduction to Capnograph Measurement................................................................................................ 38
7.6.1 CO2 Measuring Principle................................................................................................................ 38
7.6.2 Mainstream vs. Sidestream Sampling............................................................................................. 38
Chapter 8 Troubleshooting.............................................................................................................................. 40
8.1 No Display on the Screen.......................................................................................................................... 40
8.2 Excessive ECG Signal Interference or too Thick Baseline....................................................................... 40
8.3 No Blood Pressure and Pulse Oxygen Measures....................................................................................... 40
8.4 System Alarm............................................................................................................................................ 40
Chapter 9 Maintenance.....................................................................................................................................41
9.1 Service and Examination........................................................................................................................... 41
9.1.1 Daily Examination.......................................................................................................................... 41
9.1.2 Routine Maintenance...................................................................................................................... 41
9.2 Battery Maintenance.................................................................................................................................. 42
9.3 Cleaning and Disinfection of the Device................................................................................................... 42
9.4 Cleaning and Disinfection of Accessories................................................................................................. 43
9.5 Storage....................................................................................................................................................... 43
9.6 Transportation............................................................................................................................................ 43
Chapter 10 Appendix............................................................................................................................................ 44
10.1 Alarm Information................................................................................................................................... 44
10.2 Default Alarming Values and Setup Range.............................................................................................. 45
10.3 Status/Error during NIBP Monitoring..................................................................................................... 47
10.4 Status/Error during CO2 Monitoring....................................................................................................... 48
10.5 Typical Pressures and CO2 Readings at Altitudes.................................................................................... 49
10.6 Accessories List....................................................................................................................................... 49

-VI-
Chapter 1 Overview
1.1 Product Name
Name: Patient Monitor

1.2 Applications and Scope


This Patient Monitor is a multi-functional instrument designed for monitoring the vital physiological signs of
adult and pediatric patients. With the functions of real-time recording and displaying parameters, such as ECG,
heart rate, non-invasive blood pressure, functional oxygen saturation, end-tidal CO2 concentration, respiration
rate, body temperature, and so on, it allows comprehensive analysis of patient’s physiological conditions.
This instrument is applicable for use in hospitals and clinical institutions. The operation should be performed by
qualified professionals only.

1.3 Operating Environment


1. Ambient temperature range: 5°C~40°C
Relative humidity: 30%~80%
Atmospheric pressure: 70kPa ~106.0kPa
Power Voltage: (100-240)VAC
Power frequency: 50Hz/60Hz
2. This equipment should be situated in a place protected against direct sunlight, so as to prevent
overheating inside the equipment.
3. The device should be stored and used within specified temperature, humidity and atmospheric pressure
range, or it may cause damage to the device or inaccurate measurement result.
4.If the device gets wet by accident, the operator should NOT power it on directly until it has been air-dry
enough to avoid any damage to it.
5、Do not use this equipment in an environment with toxic or inflammable gas.
6. This equipment should be placed on a stand or flat platforms, so as to prevent possible shock.
7. Do not use this equipment in combination with any equipment other than those expressly permitted in
the manual.
8. The monitor is defibrillator discharge proof and can be used with electrosurgical unit. But when the
device is used together with defibrillator or electrosurgical equipment, the user (doctor or nurse) should
keep the patient under close surveillance for his/her safety. Refer to the following function description
for specific protective measures or notes.
9. Make sure that the equipotential grounding terminal is grounded correctly.
10. Do not use mobile phone nearby, so as to avoid strong radiant field interference.

1.4 Impact on Environment and Resources


Low

1.5 Safety
a) This device conforms to IEC60601-1, electric safety classification: Class I, with Type BF and CF
User manual for Patient Monitor

applied parts.
b) This device can resist against the discharge of defibrillator and the interference of electrosurgical unit.
c) This device can monitor the patients with pace-maker.
d) DO NOT use this device while the patient is under MRI scanning.

-2-
User manual for Patient Monitor

Chapter 2 Working Theories


2.1 Overall Structure
The overall structure of the monitor is shown as Figure 2.1

Figure 2.1

2.2 Composition
1. The monitor consists of the main units and the corresponding functional components (ECG cables/lead
wires, non-invasive blood pressure cuff, SpO2 probe, temperature transducer, accessories for invasive
blood pressure and CO2 monitoring).
2. The patient monitor has multiple measurement channels including ECG/Heart Rate, NIBP, SpO2/Pulse
Rate, Respiration Rate, Temperature, and CO2 Concentration.
3. The patient monitor has output ports including networking communication and the built-in printer.
4. Basic configuration includes the functions of ECG/Heart Rate, NIBP (systolic, diastolic, mean aterial
pressure & pulse rate), SpO2/Pulse Rate, Respiration Rate, and Temperature.

2.3 Working Theories


This Patient Monitor, which performs physiological parameter measurement through different modules, is a
product with modular design. It consists of several modules: ECG/RESP/TEMP module, NIBP module, SpO2
module, CO2 module (optional), the mother board and other auxiliary boards.
1. The ECG/RESP/TEMP module detects the ECG signal through ECG cable/lead wires via electrodes, it also
measures the respiration by use of the same ECG electrodes. The temperature is measured through the
temperature probe. The heart rate and respiration rate are calculated from its signal waveform respectively.

-3-
User manual for Patient Monitor

2. The SpO2 module detects the plethysmograph data via the SpO2 probe and calculates the pulse rate and
oxygen saturation (SpO2) accordingly.
3. The NIBP module measures the blood pressures including systolic, diastolic, mean arterial pressure and
pulse rate through the pneumatic system and cuff. The cuffs are designed for adult and pediatric patients
respectively, and it can work in different modes to adapt different patient categories.
4. The CO2 module detects CO2 concentration through the sampling line (side stream) or within the airway
(main stream), and calculate the respiration rate, end tidal CO2 concentration (EtCO2), and inspired CO2
concentration (InsCO2).
5. The mother board is the main controlling unit, which also connects to the interface board and key board.
The interface board inter-connects the functional modules with the mother board.

-4-
User manual for Patient Monitor

Chapter 3 Installation and Connection


3.1 Installation
3.1.1 Opening the Box and Check
1. Open the package, take out the monitor accessories from the box carefully and place it in a safe stable and
easy to watch position.
2. Open the user manual to sort the accessories according to the packing list.
 Inspect the accessories for any mechanical damages
 Check all the exposed leads and inserted accessories
 Check whether any risk or abnormity exists in the device and its accessories before using the
monitor. If any abnormity (such as broken cable or crack of the enclosure etc.) is found, stop using
this device.
☞The user can customize the module configuration by choosing necessary modules to meet your own
needs. Therefore, your monitor may not have all the monitoring functions and accessories.
Please contact the local dealer or our company in case of any problems. We will offer the best solution for
your satisfaction.
3.1.2 Connecting the AC Power Cable
1. When powered by AC mains power supply:
 Make sure that the AC power supply is (100-240)VAC, 50Hz/60Hz.
 Use the power cable provided by the manufacturer. Insert one end of it to the AC power input of
the monitor and the other end to the three-pin outlet of the power source with protected-earth.
 It's required that the user should use the 3pin mains power outlet with the protected earth. If there
is any doubt with the mains power outlet for the protected earth, the user should use the built-in
battery and disconnect the power cord to ensure electrical safety.
Note: The optionally provided grounding wire is for functional grounding purpose, which is used to reduce
the power line frequency noise or eliminate the potential difference while connecting the monitor to other
device, so making the system with equipotential grounding.
 After the mains power supply has been interrupted while the power switch of the device remains in the
“on” position and the mains power supply is restored after a period of time that is longer than 30
seconds, the monitor will run by the last settings when restarting the monitor.
 The monitor is applicable to connect to the public mains power network.
Caution: ensure that the monitor is firmly connected to the mains power supply with protected earth.
2. Built-in battery
The following steps should be followed to install the battery:
Step 1: open the battery cover;
Step 2: pull out the battery cable and connect it to the battery pack;
Step 3: push the battery pack into the battery compartment and lock it;
Step 4: close the battery cover.
 Caution: it’s better to recharge the battery after it is used up, and the charging time should be
13~15 hours long.
 Battery life: Provided that a battery is new and fully charged, the minimal working time of the

-5-
User manual for Patient Monitor

monitor with accessories connected is declared in the table below:

Name Battery life

Patient Monitor 120min


 NOTE: It will take 12-15 hours to charge battery from exhaust state to 90% charged.
 The provided battery of the monitor must be recharged after transportation or storage. So if the monitor
is switched on without being connected to the AC power supply, it may not work properly due to
insufficient battery power.

3.1.3 Starting the Monitor


The system performs self-test and enters initial display after switch on the monitor, and the alarm rings to
inform that the user can begin operating it.
 Check all the applicable functions to make sure that the monitor works normally.
 If the built-in battery is applied, please recharge it after using the monitor to ensure sufficient
power storage. It will take at least 8 hours to charge battery from depletion to 90% charge.
 Do not use the monitor to monitor the patient if there are indications of damage or reminders of error.
Please contact the local dealer or our company.
 Start the monitor again 1 minute later after it is switched off.

3.2 Connection
3.2.1 ECG Cable/Lead Wires Connection

ECG measurement is to collect the ECG signal via the ECG electrodes. Electrode connects the patient and the
lead wires and/or ECG cable. The lead wires and/or cable connect to the monitor. The locations of the electrodes
are very important for obtaining accurate ECG signals.
1. Connect the cable to the connector marked with the “ECG" icon on the signal input panel.
2. Select electrodes to be used. Use only one type of electrode on the same patient to avoid variations in
electrical resistance. For ECG monitoring, it is strongly recommended to use silver/silver chloride
electrodes. When dissimilar metals are used for different electrodes, the electrodes may be subject to large
offset potentials due to polarization. Using dissimilar metals may also increase recovery time after
defibrillation.
3. Prepare the electrode sites according to the electrode manufacturer’s instructions.

4. Skin clean
 Clean and scrap skin to ensure low sensor impedance if necessary. Mild soap and Water is
recommended as a skin cleanser.
Note: If alcohol is used as cleanser, it is recommended to have 30-second dry time for a better
connection.

 Scraping the skin gently with a dry wash cloth, gauze, for skin preparation is helpful to remove the
non-conductive skin layer.

The symbol indicates that the cable and accessories are designed to have special protection against
electric shocks and is defibrillator-proof.
The locations of the electrode are in the following Figure:

-6-
User manual for Patient Monitor

Figure 3.5 Electrode Location


Note: If any side-effect such as allergic or itchy reaction is found, remove the electrodes from the patients
immediately.

5. After starting the monitor, if the electrodes become loose or disconnected during monitoring, the system
will display “LEAD OFF” on the screen to alarm the operator.
 It might not display ECG waveform when using ECG cable with 3 lead wires while the setting of
“Cable” is set as “5” in the ECG parameter setup menu. The ECG cable with 5 leads wires should
be used to measure ECG signal at this situation.
6. The ECG leads and their corresponding locations are as follows:

Electrode connection 1 Electrode connection 2


(IEC Standard) (AHA Standard)
Electrode position on body surface
Label on Label on
Color code lead wire Color code lead wire
connection connection

Red R White RA Right Arm: The intersection between the centerline of


the right clavicle and Rib 2

Yellow L Black LA Left Arm: The intersection between the centerline of


the left clavicle and Rib 2

Green F Red LL Left Leg: Left part of the upper abdomen

Black N/RF Green RL Right Leg: Right part of the upper abdomen

White C Brown V Any of the following location (C1-C6 or V1-V6) on


chest

White/red C1 Brown V1 4th Intercostal (IC) space at right border of sternum

White/yellow C2 Brown/yellow V2 4th IC space at left border of sternum

White/green C3 Brown/green V3 Midway between V2 and V4

White/brown C4 Brown/blue V4 5th IC space on left midclavicular line


(blue)

-7-
User manual for Patient Monitor

White/black C5 Brown/red V5 Left anterior axillary line at the horizontal level of V4

White/purple C6 Brown/purple V6 Left midaxillary line at the horizontal level of V4

Safety Instructions for ECG Monitoring


 Patient Monitor can only be equipped with ECG cable and/or lead wires provided by our company;
using ECG cable and/or lead wires supplied by other companies may cause improper performance or
poor protection while using defibrillator.
 Using the same type of qualified and authorized electrodes which should be within its effective life on
the same patient. If any side-effect such as allergic or stimulus skin is found, the measurement should
be stopped at once. It is prohibited to apply the electrode to the patient with lesion and body
putrescence.
 To the patient with pacemaker, due to that this device has been designed to provide function of
pacemaker pulse inhibition for heart beat identification, normally the pacemaker pulse is not counted in
heart rate measurement and calculation, but when the pulse width of the pacemaker pulse is larger than
2ms, the pacemaker pulse inhibition may not be fully effective. In order to reduce this possibility,
observe the ECG waveforms on the screen carefully and do NOT rely entirely on the heart rate display
and alarm system of this monitor when monitoring this kind of patients. Keep pacemaker patient under
close surveillance.
 The improper connection with electrosurgical unit may not only cause burns, but also damage the
monitor or arouse deviations of measurement. You can take some steps to avoid this situation, such as
do NOT use small ECG electrodes, choosing the position which is far away from the estimated
Hertzian waves route, using larger electrosurgical return electrodes and connecting them with the
patient properly.
 Electric parts of electrodes, leads and cable are forbidden to contact any other conductive parts
(including ground).
 Patient Monitor can resist against defibrillator and electrosurgical unit. Readings may be inaccurate for
a short time after or during using defibrillator or electrosurgical unit.
 Transient caused by cable circuitry blocks while monitoring may be similar to the real heartbeat
waveform, as a result resistance heart rate alarm rings. If the electrodes and cable are located in proper
places according to this manual’s instructions and the instructions for using electrode, the chance of
this transient occurrence will be decreased.
 ECG cable and/or lead wires may be damaged while using defibrillator. If the cable and/or lead wires
are used again, please do the functional check firstly.
 When the monitor is inoperable due to overload of ECG or saturation of any part of the amplifier, it will
prompt “Lead off” to remind operator.
 The user should ensure that no predictable hazard will be caused by the summation of leakage currents
when several item of monitor are interconnected.
 When plugging or unplugging the ECG cable, be sure to hold the head of the connector and pull it out.

3.2.2 Cuff Connection for Blood Pressure Measurement

1. Connect the tube with cuff to the connector marked with “NIBP” icon on the signal input panel.
2. Unfold the cuff and wrap it around the patient’s upper arm.
Requirements of the cuff:
1) Appropriate cuff should be selected according to the age and arm circumference of the subject. Its width
should be 2/3 of the length of the upper arm. The inflatable part should be long enough to permit
wrapping appropriately 80% of the limb.

Note: the appropriate cuff should be selected according to the age and arm circumference of the patient.

-8-
User manual for Patient Monitor

2) When putting on the cuff, unfold and wrap it around the upper arm evenly to appropriate tightness.
3) Remember to empty the residual air in the cuff before the measurement is commenced.
4) Locate the cuff in such a way that the artery mark “ ” is at a location where the clearest pulsation of
brachial artery is observed.
5) The cuff should be tightened to a degree where insertion of one finger is allowed.
6) The lower end of the cuff should be 2cm above the elbow joint.

Figure 3.6 Cuff location

The symbol indicates that the cable and accessories are designed to have special protection against
electric shocks, and is defibrillator proof.
 Pressure Accuracy Verification
Pressure Accuracy Verification is a function to inspect the accuracy of pressure measurement by the NIBP
module inside the device. Technician or equipment manager should do pressure accuracy verification every
half year or year in order to check if the pressure measurement still conforms to the requirement of product
performance. If the deviation is beyond the declared specification, it is permitted to return it to factory for
repair or calibration.
Before verification, please connect the monitor to a precision pressure meter as the reference
equipment like a mercury pressure meter

-9-
User manual for Patient Monitor

Figure 3.7 Connection of Pressure Accuracy Verification Fixture

Mode 1: Automatic inflation for the pressure accuracy verification

At this mode, the monitor can activate the inflation, so the pressure will increase automatically until it
exceeds the limit value specified in table A. This pressure limit value depends on the patient type selection
as shown in table A:

Adult 240mmHg

Pediatric 200mmHg

Neonate 120mmHg
Table A
During the inflation, the Monitor will close the deflating valve, and the pressure value will be shown
during the process. If there is no manual deflation operation, the pressure will persist until deflation by
manual operation, so it is necessary to use a manual valve for doing adequate deflation in several steps
to verify the pressure accuracy in the full scale of measurement range.
Mode 2: Manual inflation for the pressure accuracy verification.
At this mode, the pressure should be increased manually by a pumping balloon, and the verification can be
done by applying different pressure value manually. If the increased pressure exceeds the given limit as
shown in table B, the Monitor will deflate automatically because of over-pressure protection.

Adult 300mmHg

Pediatric 240mmHg

Neonate 140mmHg
Table B
 After the verification, do press the button again to return to normal working mode, then
continue other operation, or the NIBP key will be invalid.
 Pressure accuracy verification must be operated by technician or equipment manager. Doctor or
nurse is not allowed to do the verification, it is very dangerous especially when the pressure cuff
is still on patients.
 Air Leakage Check

-10-
User manual for Patient Monitor

In order to avoid significant error of blood pressure measurement or even no measurement result caused by
air leakage in the pneumatic system including the cuff during measuring, it is recommended to check if there
is leak in the pneumatic system as well.
 Please remove the cuff from patient while performing the leakage check.
Safety Instructions for NIBP Measurement
 When taking the blood pressure measurement on a Pediatric or Neonate patient, do NOT
operate in the Adult mode. The high inflation pressure may cause lesion or even body
putrescence. Even though the monitor can identify the cuff type so it will stop inflation and
indicate "wrong cuff" when taking the blood pressure measurement for a pediatric or neonate
in the "adult" patient type setting. The user (doctor or nurse) should pay more attention to
select the correct patient type.
 It is recommended to take the blood pressure measurement manually.
 NIBP monitoring is prohibited to those who have severe hemorrhagic tendency or with sickle
cell disease, or partial bleeding will appear.
 Do NOT bind NIBP cuff on limbs with transfusion tube or intubations or skin lesion area,
otherwise, damages may be caused to the limbs.
 If the patient is moving or suffering tremble, hyperkinesia or arrhythmia, it may cause the
inflation time of inflatable balloon endures longer, which may not only prolong the NIBP
measurement time, but also result in the body wrapped by the cuff is troubled by purpura,
hypoxemia and neuralgia because of the friction.
 Before the measurement is carried out, select an appropriate monitoring mode depending on the
patient type (adult, pediatric or neonate).
 It is prohibited to wrap the cuff to a limb with skin lesion
 DO NOT take blood pressure measurement from a limb receiving ongoing transfusion or
intubations. Because it may damage the limb tissue around the intubation if the transfusion
becomes slower or blocked during the cuff inflation.
 The windpipe which connects the cuff and monitor should be straightway without any tangle.
 When an adult subject is monitored, the machine may fail in giving the blood pressure measure
if the pediatric mode is selected.
 Prior to use of the cuff, empty the cuff until there is no residual air inside it to ensure accurate
measurement.
 Do NOT twist the cuff tube or put heavy things on it.
 When unplugging the cuff, hold the head of the connector and pull it out.
 The NIBP measurement will not be affected when the monitor is connected to the patient on
whom the electrosurgical device such as defibrillator or electrosurgical knife with high frequency
is being used.
 The appearance of arrhythmia results in irregular heart beat which may affect the accuracy of
NIBP measurement data. It is recommended to take the NIBP measurement again at this
situation.
 The Blood pressure measurements determined with this device are equivalent to those obtained
by a trained observer using the cuff/stethoscope auscultatory method, within the limits prescribed
by the American National Standard, Manual, electronic, or automated sphygmomanometers.

The symbol indicates that the cable and accessories are designed to have special protection against
electric shocks, and is defibrillator proof.

3.2.3 SpO2 Probe Connection

Creative reusable SpO2 finger clip sensor is configured by default for general purpose. For further information,

-11-
User manual for Patient Monitor

please contact your local representative.

Creative reusable SpO2 finger clip sensor can be used with a compatible patient monitor (e.g. all models of
Patient Monitors and PC-900A Vital Signs Monitor made by Creative) or pulse oximeters (e.g. PC-66B Handheld
Pulse Oximeter made by Creative).

The sensor is intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO2) and
pulse rate monitoring for adult patients weighing greater than 40kg or pediatric patients weighing between
10~40kg.

SpO2 probe is a kind of very delicate sensor. Please follow the steps and procedures in operating it. Failure to
operate it correctly can cause damage to the SpO2 probe.
Operation procedure:
1. Connect the SpO2 probe to the connector marked with “SpO2” icon on the signal input panel. When
unplugging the probe, be sure to hold the head of the connector and pull it out.

2. Insert one finger (index finger is preferred, but middle or ring finger with proper nail length is possible as well)
into the probe according to the finger mark on the probe, shown as below.

Figure 3.8A Demonstration of Adult/Pediatric SpO2 Finger Clip Sensor


(A) With the upper and lower jaws open, place a finger evenly on the base of the clip. Push the finger tip against
the stop so that it is over the sensor window.
(B) Spread open the rear tabs of the sensor to provide even force over the length of the pads.
(C) The sensor should be oriented in such a way that the cable is positioned along the top of the hand.

-12-
User manual for Patient Monitor

Figure 3.8B Demonstration of Adult SpO2 Finger Rubber Sensor


(A) Hold the sensor with its opening towards the patient’s finger, the sensor should be oriented in such a way
that the sensor side with a finger tip sign is positioned on the top.
(B) Insert the patient’s finger into the sensor until the fingernail tip rests against the stop at the end of the sensor.
Adjust the finger to be placed evenly on the middle base of the sensor. Direct the cable along the top of the
patient’s hand. Apply adhesive tape to secure the cable if necessary.

When selecting a SpO2 probe or sensor, do consider the patient’s category, adequacy of perfusion, availability of
probe site and anticipated monitoring duration. Use only SpO2 probes provided by our company with this monitor.
Read the following table for SpO2 probe information.

SpO2 Probe Patient Category


Adult SpO2 Finger Clip Sensor (reusable)
Adult
Adult SpO2 Finger Rubber Sensor (reusable)
Pediatric SpO2 Finger clip Sensor (reusable) Pediatric
SpO2 Y-type sensor with wrapper (reusable) Neonate
3. If the neonate SpO2 sensor is used, please follow figure to connect.

Y type sensor Sensor adapter: wrapper

Figure 3.8C Neonate SpO2 sensor placement

Some factors may affect the accuracy of saturation measurements. Such factors include: excessive patient motion,
fingernail polish, use of intravascular dyes, excessive light, poorly perfused finger, extreme finger sizes or
improper placement of the sensor.

High ambient light sources such as surgical lights (especially those with a xenon light source), bilirubin lamps,
fluorescent lights, infrared heating lamps, and direct sunlight can interfere with the performance of a SpO2 sensor.

-13-
User manual for Patient Monitor

To prevent interference from ambient light, ensure that the sensor is properly applied, and cover the sensor site
with opaque material if necessary. Failure to take this action in high ambient light conditions may result in
inaccurate measurements.

If patient movement presents a problem, verify that the sensor is properly and securely applied; move the sensor
to a less active site; use an adhesive sensor that tolerates some patient motion; or use a new sensor with fresh
adhesive backing.

For reusable sensors, follow the sensor directions for use, cleaning and reuse. For single-patient use sensors, use a
new sensor for each patient.

Caution: Do not disinfect any SpO2 sensor by irradiation, steaming, or ethylene oxide.
Safety Introductions for SpO2 Monitoring

 Continuous use of fingertip SpO2 sensor may result in discomfort or pain, especially for those
patients with microcirculatory problem. It is recommended that the sensor should NOT be
applied to the same site for over two hours, please inspect the monitoring site every 1~2
hours for skin integrity, and change the measuring site periodically if necessary.
 SpO2 measuring site must be examined more carefully for some special patient. Do NOT place
the SpO2 sensor on the finger with edema or fragile tissue.
 Avoid placing the SpO2 sensor on the same extremity with an arterial catheter, blood pressure
cuff, or intravascular infusion line, otherwise the blood flow could be interrupted by the
cuff or the circulatory condition could make low blood perfusion so that would result in on
pulse found or loss of pulse during SpO2 monitoring and further cause false alarm.
 Do not apply tape to secure the sensor in place or to tape it shut; venous pulsation
may lead to inaccurate oxygen saturation measurements.
 For disposal SpO2 sensor, If the sterile packaging is damaged, do not use it any more.
 Check the SpO2 sensor and cable before use. Do NOT use the damaged SpO2 sensor.
Before each use, surface-clean sensor and cable with a soft gauze pad by saturating it with a solution
such as 70% isopropyl alcohol. If low-level disinfection is required, use a 1:10 bleach solution.
When the temperature of SpO2 sensor is abnormal, do not use it any more.
 Please do not allow the cable to be twisted or bended.
 Please do not use nail polisher or other cosmetic product on the nail.
 The fingernail should be of normal length.
 The SpO2 sensor cannot be immerged into water, liquor or cleanser completely, because the sensor has
no capability to resist the harmful ingress water.
Carefully route cables to reduce the possibility of patient entanglement or strangulation.
Do not use the sensor or other oximetry sensors during MRI scanning.

Our company offers a 6-month warranty against manufacturing defects for the SpO2
sensors mentioned above in its undamaged condition.

If you have any question regarding any of SpO2 sensor instructions, please contact
your local dealer.

3.2.4 CO2 Sensor Connection


3.2.4.1 Sidestream CO2 Sensor Connection

-14-
User manual for Patient Monitor

Demonstration for Sidestream CO2 Sensor Connection


1. Take out the CO2 Sensor and insert the CO2 Sensor Cable into the connector labeled “CO2” on the connector
panel of the monitor;
2. The sample cell of the sampling cannula must be inserted into the sample cell receptacle of the CO2 Sensor. A
“click” will be heard when the sample cell is properly inserted. Then connect to airway tube. After finishing
sensor connection, and make sure that the air input end is exposed to room air and away from all sources of
CO2, including the ventilator, the patient’s breath and your own. Next, turn on the CO2 switch at CO2 Setup
Screen and then wait 2 minutes for the sensor warm-up.
3. Default Tubing Configuration

Adapter and Sampling tube (Single patient use)

Extending airway tube for connecting to sampling tube (Single patient use)

Wye Connector

4. Optional sampling cannula kits

-15-
User manual for Patient Monitor

(1) T connector sampling cannula kits

On-air Connector

Adapter
(2) Nasal Sidestream Cannula Kits

(3) Oral Sidestream Cannula Kits

3.2.4.2 Mainstream CO2 Sensor Connection


Sensor cable

Adapter

CO2 sensor

Demonstration for Mainstream CO2 Sensor Connection


1. Take out the CO2 Sensor and insert the CO2 Sensor Cable into the connector labeled “CO2” on the connector
panel of the monitor;
2. Snap the CO2 sensor onto the airway adapter as shown in Figure 6.9. A “click” will be heard when the airway
adapter is properly inserted.

-16-
User manual for Patient Monitor

3. Position the airway adapter in the patient’s respiratory circuit (as close to the patient as possible) between the
endotracheal tube and the ventilator circuit. Next, turn on the CO2 switch at CO2 Setup Screen and then wait 2
minutes for the sensor warm-up.

Safety Introductions for CO2 Monitoring


 CO2 Sensor is a precision measuring part, please use it correctly and store it
properly;
 Precautions for electrostatic discharge (ESD) and electromagnetic interference
(EMI) to and from other equipment.
 Failure of Operation: If the CO2 Sensor fails to respond as described in this user
manual; DO NOT use it until approved for use by qualified personnel.
 DO NOT position the sensor cables or tubing in any manner that may cause
entanglement or strangulation.
 Support the airway adapter to prevent stress on the ET tube.
 Reuse, disassembly, cleaning, disinfecting the single patient use CO2 airway
adapters may compromise functionality and system performance leading to a user
or patient hazard. Performance is not guaranteed if an item labeled as single patient
use is reused.
 Inspect the sidestream on-airway adapters and sidestream sampling kits for damage
prior to use. DO NOT use the sidestream on-airway adapters and sidestream
sampling kits if they appear to be damaged or broken.
 If the CO2 waveform (Capnogram) appears abnormal, inspect the CO2 airway
adapters and replace if needed.
 Periodically check the CO 2 /Flow sensor and tubing for excessive moisture
or secretion buildup. Do not use them if there is excessive moisture or exterior
condensation.
 Electric Shock Hazard: The CO2 Sensor contains no user serviceable parts.
 Refer service to qualified service personnel. Do not open the sensor cabinet at will,
as electric shock hazard may occur.
 Place the exhaust vent of the CO2 Sensor in drafty ambient and do not let anything
block the exhaust vent.
 Always disconnect the CO2 Sensor before cleaning. Do NOT use if it appears to
have been damaged. Refer servicing to qualified service personnel.

-17-
User manual for Patient Monitor

 DO NOT sterilize or immerse the CO2 Sensor in liquids.


 Replace the sidestream on-airway adapters and sidestream sampling kits if
excessive secretions are observed.
 Do not operate the CO2 Sensor when it is wet or has exterior condensation.
 Monitor the CO2 waveform (Capnogram). If you see changes or abnormal
appearance, check the patient and the sampling line. Replace line if needed.
 DO NOT use device on patients that cannot tolerate the withdrawal of 50 ml/min
+/- 10 ml/min from the airway or patients that cannot tolerate the added dead space
to the airway.
 Do not apply excessive tension to any sensor cable or pneumatic tubing.
 Explosion Hazard: DO NOT use in the presence of flammable anesthetics or other
flammable gasses. Use of the CO2 Sensor in such environment may present an
explosion hazard.
☞ The power voltage over monitor working voltage may cause damage to CO2 sensor.
Likewise, too low power voltage may affect the CO2 measuring accuracy or even
make the CO2 sensor not work.
☞ When changing sampling tube, it is suggested to choose the default sampling tube
with dehumidifying function. The sampling tube without dehumidifying function
may be easily blocked by excessive moisture. (Use life: ordinary sampling tube:
6~12 hours; the sampling tube with dehumidifying function: about 120 hours.)
☞ If the measurement appears abnormity caused by sampling tube block, please
replace it.
☞ The total length of the sampling tube and extending airway tube shouldn’t be longer
than 3 meters, too long may cause measurement abnormity. If using T connector
sampling cannula kits, please insert the sampling tube with the tubes upward to
avoid the affects of excessive moisture;
☞ Altitudes are different in different area, so set the Barometric Pressure setting value
as the ambient barometric pressure.
☞ Use only our company approved accessories.
☞ While using the CO2 sensor, a system leak, that may be caused by an uncuffed
endotracheal tube or a damaged CO2 sensor may significantly affect flow-related
readings. These include flow, volume, pressure and other respiratory parameters.
☞ When stopping CO2 monitor, please disconnected the CO2 sensor from the patient
monitor.
☞ Disposal of the CO2 Sensor and its accessories should comply with national and/or
local requirements.
☞ In the presence of electromagnetic devices (i.e., electrocautery), patient monitoring may be
interrupted due to electromagnetic interference. Electromagnetic fields up to 20 V/m will
not adversely affect system performance.
☞ Nitrous oxide, elevated levels of oxygen, helium and halogenated hydrocarbons can
influence the CO2 measurement.
☞ Excessive moisture in the CO2 may affect the accuracy of the flow measurement.

-18-
User manual for Patient Monitor

3.2.5 TEMP Probe Connection

Patient Monitor has two TEMP probes to measure different body temperature.

Connecting methods:

1. Attach the probes to the patient body firmly;


2. Connect them to “TEMP” on the right panel.
Note: When unplugging the probe, be sure to hold the head of the connector and pull it out.

3.2.6 Loading Printing Paper

This description is for loading paper for the built-in printer.


Operation procedures:
1. Press both “OPEN” notches with force on printer shield with two thumbs to open it.
2. Move the tab of rubber roller lock at the left 90°upwards to unlock it.
3. Cut one end of the paper into triangle, and load the paper from the underside of the rubber roller.
4. Turn the roller clockwise to get the paper rolled, and put the paper roll into the compartment.
5. Pull the paper out of paper slot on the shield.
6. Move the tab of the rubber roller lock 90° downwards to lock it.
7. Put the shield back in position and secure it.
Unloading printing paper
1. Press both “OPEN” notches vertically with force on printer shield with two fingers to open it.
2. Move the tab of roller lock at the left 90° upwards to unlock it.
3. Roll the loading roller anti-clockwise and pull the paper out.
4. Roll the loading roller clockwise to get the paper rolled, and put it into the compartment.
5. Pull the paper out of paper slot on the shield.

Fig.3.9 Fig.3.10

Built-in printer may be used due to the different configuration.


Printer operation instruction:
Power indicator: The green light shows the power is on. While the monitor is out of power, the green light is off.
Error indicator: red light is constant which shows the printer is out of paper, or the printing paper is not installed
properly. When the printering paper is loaded correctly, the red light is off.

-19-
User manual for Patient Monitor

Power Indicator

Error Indicator Open button

Paper cartridge

Figure 3.11 Built-in printer


Loading printing paper:
Step 1: press and hold down the cartridge button to open the paper cartridge;
Step 2: Install the paper to the printer properly, pull the paper out of the printer for 2 cm, as shown in figure 3.12.
Step 3: Close the printer cover along the direction of arrow, as shown in figure 3.12.

Figure 3.12 printing paper

-20-
User manual for Patient Monitor

Chapter 4 Alarm

 Once the patient is changed, please recheck if the monitor can work normally, alarm function works
properly, and alarm setting are appropriate.

4.1 Alarm Description


4.1.1 Alarm Condition
The patient monitor has two alarm modes, respectively for the physiological alarm conditions and technical alarm
conditions, the detailed define as follows:
 physiological alarm conditions: the device would give out alarm when the patient’s physiological
parameter value exceeds the preset limit, such as: Asystole, asphyxia, temperature over-limit and so
on.
 technical alarm conditions: the device would give out alarm when system or sensor makes failures to
cause abnormal monitoring function and inaccurate monitoring results, such as ECG lead off, probe
off , low battery and so on.

4.1.2 Alarm Priority


 The patient monitor has 3 alarm levels: High priority, Medium priority and low priority. In addition,
the monitor has preset alarm level about physiological alarm and technical alarm. Please refer to table
4-1.
Alarm level Alarm Event
Over TEMP1 limit, Over TD limit
Asystole, Unable to detect HR, Over HR limit
Over NIBP PR limit, Over NIBP SYS limit, Over NIBP DIA limit
High Unable to detect SpO2, Over PR limit, Over SpO2 limit,
asphyxia ,Over RR limit
Over EtCO2 limit, Over InsCO2 limit
The battery capacity will exhaust

Lead Off, Probe off, Poor Signal, Electrode Off, High Impedance, Stifle Time Indication, Artefact.
Medium

Low Other abnormal phenomenon


Table 4-1 Alarm priority and alarm event listing
Related to 3 levels alarm modes, medical and nursing staff should have different response to deal with
potential dangerous, the detailed demands as follows:
1. High priority alarm: medical and nursing staff should response immediately.
2. Medium priority alarm: medical and nursing staff should response quickly.
3. Low priority alarm: medical and nursing staff should response as soon as possible.
NOTE: Some models may have only Medium and Low alarm priority because of the different configuration.

-21-
User manual for Patient Monitor

4.1.3 Alarm Modes


1. When an alarm occurs, the patient monitor provides visible and audible alarm indications (which
are shown by three ways: auditory alarm, light alarm and message description or numerical flash
alarm). The detail is defined as follows:
 auditory alarm: are represented by loud speaker in monitor.
 light alarm: are represented by flashing LEDs on the upper side of monitor.
 message description or numerical flash alarm: are represented by waveform on the left display
area or parameters display on the right area of the screen respectively.
2. In order to inform the alarm quickly and accurately to users It is suggested that the distance
between users and device should not exceed 4m. And the condition of clear alarm indication is
when the distance within 1m (no block exists between the device and user)
3. if occurrence of multiple alarm signal at the same time, the monitor will give out alarm which has
much more alarm signals alternative auditory alarm and light alarm. Meanwhile, message
description or numerical flash alarm will be shown respectively in the form of words or numerical
value.
4. The difference between pulse beep sound and alarm sound is shown below:
 Pulse beep sound: frequency is 300Hz, including SpO2 sound and HR beep sound.
 Alarm sound: please refer to Technical Specifications description.

4.1.4 Alarm Setting


1、 Except volume of audible alarm can be adjustable, the other properties of the alarm cannot be adjusted
by the user, such as alarm priority setting, alarm light flashing and so on. In addition, all alarms in this
patient monitor are “non-latched” type, that is to say, when the alarm event disappears, the
corresponding alarm will automatically stop. The alarm volume range is shown as below:
 High: 45dB~80dB (The distance between device front and test instrument is 1m)
 Medium: 45dB~75dB (The distance between device front and test instrument is 1m)
 Low: 45dB~70dB (The distance between device front and test instrument is 1m)
2、 When the icon displays on the screen and its color is red, that means the alarm volume is 0 (alarm
is mute), at this time the user should pay more attention to the patient.
 It is suggested that the users should not change the alarm volume lower than the factory default setting
if close and constant attention could not be paid to the patient, otherwise the negligence of alarm event
might cause irreversible harm to the patient.
 During the alarm silence period, any new alarm event can activate the audible alarm again and the
audible alarm function resumes normal state.
3、 Alarm settings are non-volatile, that means the previous settings will still sustain if the patient monitor
is powered off (by accidental power interrupt or by normal power down) and reboot.

4.2 Alarm Technical Specifications


When an alarm occurs, the monitor responds with visual alarm indications (which are shown by two
ways: alarm indicator and alarm message description) and audible alarm indications.

Visual Alarm Indicators


The visual alarm indications are represented by flashing LEDs on the upper side of monitor, parameters
display on the right area of the screen and the status line displayed on the bottom of the screen respectively.
The LEDs flashing frequency and display colors are specified in the following table :

-22-
User manual for Patient Monitor

Alarm level LED Colors Blinking frequency Duty rate


High priority Red 2Hz 50%
Middle priority Yellow 0.5Hz 50%
Low priority Yellow Continuous lighting
No alarm Green Continuous lighting
Parameters displayed on the screen will be inversed or flashing if corresponding alarm occurs.

Audible Alarm Indications


The audible alarm has different tone pitch and on-off beep patterns for each priority category. These are
summarized in the Table below.
High priority: A chain of 10 beeps with pause intervals in such way: x, x, 2x+td, x, 1s, x, x, 2x+td, x, 1s,
where x (pulse interval)=100ms, td (beep duration)=160ms, 1s=1 second, the beep tone is
400Hz, the pause period between every beeping chain is 5 seconds.
Middle priority: A chain of 3 beeps with 200ms pause interval and 200ms beep duration, the frequency is
500Hz, the pause period between every beeping chain is 10 seconds.
Low priority: A single beep without repeat, the frequency is 500Hz, beep duration is 200ms.
Note: Visual alarm indicators can not be suspended or removed. Audible alarms may be decreased in
volume or silenced.

-23-
User manual for Patient Monitor

Chapter 5 Technical Specifications


5.1 ECG Monitoring
1. Input signals range in amplitude: ±(0.5mVp ~ 5mVp)
2. Heart rate display range: 15 bpm ~ 350 bpm
3. Heart rate display accuracy: ±1% or ±2bpm, whichever is greater.
4. Heart rate averaging: Averages the recent eight beats having RR intervals falling within the acceptable
limits.
5. Heart rate alarm delay time: ≤10s
6. Response time to change in heart rate:
Change from 80bpm to 120bpm: <8 sec
Change from 80bpm to 40bpm: <8 sec
7. Tall T-wave rejection: Rejects all T-wave less than or equal to 120% of 1mV QRS.
8. Pacemaker pulse rejection:
Rejects all pulses of amplitude ±2mV to ±700mV and duration 0.1 to 2 ms without
overshoot;
Rejects all pulses of amplitude ±2mV to ±400mV and duration 0.1 to 2 ms with overshoot.
9. Sensitivity selection:
×1/2, 5mm/mV tolerance: ±5%
×1, 10mm/mV tolerance: ±5%
×2, 20mm/mv tolerance: ±5%
10. Sweeping speed: 12.5mm/s, 25mm/s, 50mm/s tolerance: ±10%
11. ECG noise level: ≤30μVP-P.
12. ECG input loop current: ≤0.1μA
13. Differential input impedance: ≥5MΩ
14. Common-mode rejection ratio (CMRR): ≥89dB
15. Time constant:
Monitoring mode: ≥0.3s Diagnostic mode: ≥3.2s
16. Frequency response:
Monitoring mode: 0.5 Hz~40Hz(+0.4dB
-3.0dB)

Diagnostic mode: 0.05 Hz~75Hz(+0.4dB


-3.0dB)
Additional declarations to conform the particular standard of IEC 60601-2-27 “Medical
electrical equipment – Part 2-27: Particular requirements for the safety, including essential
performance, of electrocardiographic monitoring equipment”

-24-
User manual for Patient Monitor

Direct current for Applied current less than 0.1 microamperes.


respiration, leads-off
sensing, and active noise
suppression

Response to irregular A1 Ventricular bigeminy-80BPM


rhythm A2 Slow alternating ventricular bigeminy-60BPM
A3 Rapid alternating ventricular bigeminy-120BPM
A4 Bidirectional systoles-90BPM

Time to ALARM for Waveform B1, Amplitude Average Time to Alarm


tachycardia 0.5 mV <8 sec
1 mV <8 sec
2mV <8 sec

Waveform B2, Amplitude Average Time to Alarm


1mV <8 sec
2mV <8 sec
4mV <8 sec

5.2 RESP Monitoring


1. RESP rate measuring range: 0rpm~120rpm
2. RESP rate accuracy: ±5% or ±2 rpm, whichever is greater
3. RESP rate alarm limit setting range: 0rpm~120rpm.
4. Alarm tolerance: ±5% or ±2 rpm, whichever is greater

5.3 TEMP Monitoring


1. TEMP measuring range: 25.0oC~45.0 oC
2. TEMP measuring accuracy: ±0.2 oC
3. TEMP responding time: ≤150s

5.4 NIBP Monitoring


1. Measuring method: Oscillometric Technique
2. Pneumatic pressure measuring range: 0 mmHg~300mmHg
3. Accuracy of pressure measurement: ±3 mmHg
4. Cuff inflation time: <10 seconds (typical adult cuff)
5. Measurement time on the average: < 90 seconds
6. Air release time while the measurement is canceled: <2 seconds (typical adult cuff)
7. Initial cuff inflation pressure
Adult: <180 mmHg; Infant: <120 mmHg; Neonate: <90 mmHg

-25-
User manual for Patient Monitor

8. Overpressure protection limit


Adult: 300 mmHg; Infant: 240mmHg; Neonate: 150 mmHg
9. NIBP measurement range:
press(unit) Adult Infant Neonate
SYS mmHg 40~255 40~200 40~135
MAP mmHg 20~215 20~165 20~110
DIA mmHg 10~195 10~150 10~95
10. NIBP accuracy:
Maximum mean difference: ±5 mmHg
Maximum Standard deviation: 8 mmHg
11. Measurement mode: Manual, Auto, STAT

5.5 SpO2 Monitoring


1. Transducer: dual-wavelength LED
Wavelength: Red light: 663 nm, Infrared light: 890 nm.
Maximal optical output power: less than 2mW maximum average
2. SpO2 measuring range: 35%~100%
3. SpO2 measuring accuracy: not greater than 3% for SpO2 range from 70% to 100%
*NOTE: accuracy defined as root-mean-square value of deviation according to ISO 9919
4. Low perfusion performance: the declared accuracy is sustained when the pulse amplitude modulation
ratio is as low as 0.4%

5.6 Pulse Rate Monitoring


1. Pulse rate measuring range: 30bpm~240bpm
2. Pulse rate measurement accuracy: ±2bpm or ±2%, whichever is greater.

5.7 CO2 Monitoring


1. Technology: Infrared absorption method.
2. Mode of Sampling: Sidestream or Mainstream
3. CO2 Response Time:
Sidestream: <3seconds (including transport time and rise time).
Mainstream: <60ms (rise time)
4. Warm-up Time: Not less than two minutes
5. CO2 measurement range: 0~150mmHg
6. CO2 Accuracy: 0~40mmHg ±2mmHg
41~70mmHg ±5% of reading
71~100mmHg ±8% of reading
101~150mmHg ±10% of reading
*NOTE: Gas temperature at 25°C for Sidestream;
Gas temperature at 35°C for Mainstream
7. Flow rate: 50ml/min ±10 ml/min (Sidestream)

-26-
User manual for Patient Monitor

5.8 Data Recording


1. Sensitivity selection tolerance: ±5%
2. Recording speed: 25mm/s
3. Recording speed accuracy: ±10%
4. Hysteresis: ≤0.5mm
5. Frequency response:
Monitoring mode: 0.5~40Hz Diagnostic mode: 0.05~75Hz
6. Time constant:
Monitoring mode: ≥0.3s Diagnostic mode: ≥3.2s

5.9 Other Technical Specifications


1. Power supply: 100~240VAC, 50/60Hz
2. Power consumption: see the nameplate on the monitor
3. Display mode: TFT color LCD
4. Alarming mode: audible & visible alarm
5. Communication: Net port

5.10 Classification

Safety standard: IEC 60601-1


The type of protection against electric shock Class I equipment
The degree of protection against electric shock Type BF, CF applied parts

Electro-Magnetic Compatibility: Group I, Class A

-27-
User manual for Patient Monitor

5.11 Guidance and manufacturer’s declaration-Electromagnetic compatibility


Table 1
Guidance and manufacturer’s declaration-electromagnetic emission-
for all EQUIPMENT AND SYSTEMS

Patient Monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the equipment or system should assure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment-guidance

RF emissions Patient Monitor uses RF energy only for its internal function.
CISPR 11 Group 1 Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.

RF emissions Patient Monitor is suitable for use in all establishments other


CISPR 11 Class A than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
Harmonic emissions
used for domestic purposes.
IEC61000-3-2 Class A

Voltage
fluctuations/flicker
Complies
emissions
IEC61000-3-3

-28-
User manual for Patient Monitor

Table 2
Guidance and manufacturer’s declaration-electromagnetic immunity
for all EQUIPMENT AND SYSTEMS

Patient Monitor is intended for use in the electromagnetic environment specified below. The customer or the user
of the equipment or system should assure that it is used in such an environment.
Electromagnetic environment
Immunity test IEC60601 test level Compliance level
-guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete
discharge(ESD) ±8kV air ±8kV air or ceramic tile. if floors are
IEC61000-4-2 covered with synthetic material,
the relative humidity should be at
least 30%
Electrical fast ±2kV for power ±2kV for power Mains power quality should be
transient/burst Supply lines Supply lines that of a typical commercial or
±1 kV for ±1 kV for hospital environment.
IEC61000-4-4 input/output lines input/output lines

Surge ±1kV line (s) to line(s) ±1kV differential Mains power quality should be
IEC 61000-4-5 ±2kV line(s) to earth mode that of a typical commercial or
±2kV common hospital environment.
mode
Voltage dips, short Mains power quality should be
interruptions and voltage that of a typical commercial or
variations on power hospital environment. If the user
supply input lines of the equipment or system
IEC61000-4-11 requires continued operation
during power mains interruptions,
it is recommended that the
equipment or system be powered
from an uninterruptible power
supply or a battery.

Power Power frequency magnetic fields


frequency(50Hz/60Hz) should be at levels characteristic of
magnetic field a typical location in a typical
3A/m 3A/m
IEC61000-4-8 commercial or hospital
environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

-29-
User manual for Patient Monitor

Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity-for
EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING

Patient Monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of Patient Monitor should assure that it is used in such an electromagnetic environment.
Compliance
IMMUNITY test IEC 60601 test level Electromagnetic environment - guidance
level
Portable and mobile RF communications
equipment should be used no closer to any part of
Patient Monitor, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Conducted RF 3 Vrms 3V
150 kHz to 80 MHz Recommended separation distance
IEC 61000-4-6

Radiated RF 3 V/m 3 V/m


IEC 61000-4-3 80 MHz to 2.5 GHz

Where P is the maximum output power rating of


the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m). b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey ,a
should be less than the compliance level in each
frequency range .b
Interference may occur in the vicinity of
equipment marked with the following symbol.

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, and
electromagnetic site survey should be considered. If the measured field strength in the location in which Patient
Monitor is used exceeds the applicable RF compliance level above, Patient Monitor should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating Patient Monitor.
b: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

-30-
User manual for Patient Monitor

Table 4
Recommended separation distances between portable and mobile RF
communications equipment and The equipment or system-
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING

Patient Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the equipment or system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the equipment or system as recommended below, according to the maximum output power
of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum m
output power of 150kHz to 80MHz 80MHz to 800MHz 80MHz to 2,5GHz
transmitter
W

0,01 0.12 0.12 0.23

0,1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

-31-
User manual for Patient Monitor

Chapter 6 Packaging and Accessories


6.1 Packaging
The product is packed in high quality corrugated cartons with foam inside to protect the apparatus against
damage in the handling process.
Gross weight: Details see the indication on the outer package
For different configuration, there are 3 types of dimension:
Dimension 1: 500(L) ×320(W) ×460(H) mm
Dimension 2: 360(L) ×320(W) ×410(H) mm
Dimension 3: 506(L) ×390(W) ×510(H) mm

6.2 Accessories
(1) ECG lead cable One set
(2) NIBP cuff One set
(3) SpO2 probe One piece
(4) Body temperature probe One piece
(5) Power supply cable One piece
(6) Equipotential grounding wire One piece
(7) Disposable electrode Ten pieces
(8) User Manual One copy
(9) Warranty One copy
(10) Quality certificate One copy
(11) Assembly report Two copies
(12) Dustproof mantle One set
(13) Printing paper (optional) Ten rolls
(14) CO2 accessories (optional)

For Mainstream For Sidestream


Mainstream sensor One set Sidestream Sensor One set
( CAPNOSTAT 5 ) (LoFlo C5)
Airway adapter One piece Sampling Line Kit One set

Note: The accessories are subject to change. Detailed items and quantity see the Packing List.

-32-
User manual for Patient Monitor

Chapter 7 Working Principles


7.1 Introduction to ECG Measurement
7.1.1 How to Obtain High Quality ECG and Accurate Heart Rate Value
The electrocardiogram (ECG or EKG) is primarily a tool for evaluating the electrical events within the heart. The
action potentials of cardiac-muscle cells can be viewed as batteries that cause charge to move throughout the body
fluids. These currents represent the sum of the action potentials occurring simultaneously in many individual cells
and can be detected by recording electrodes at the surface of the skin. The figure below shows the system of the
heart.

First of all, the hospital should be equipped with a 100~250V power supply system with a typical grounding wire.
If big interference in ECG continues, connect one end of the grounding wire provided with this equipment to the
grounding wire on the back panel of this monitor, and the other end to the special grounding wire, water pipe or
radiator.
A common ECG plate electrode used together with this monitor has short shelf life. Generally, the shelf life is
only one month after the package is opened. When outdated plate electrode is used, due to skin’s contact
impedance and big electrode potential, the chance of interference will be increased, and the ECG baseline will
have an unstable inclination. Therefore, always use valid plate electrodes.
7.1.2 Factors affecting ECG signal
 Interference from Electrosurgical Unit;
 Doesn’t filter the interference waveform;
 Poor grounding;
 Electrodes are not placed properly;
 Use expired electrode or use disposable electrode repeatedly;
 The skin placed electrode is unclean or poor contract caused by scurf and hair;
 Electrode long-time used.

-33-
User manual for Patient Monitor

7.2 Introduction to Blood Pressure Measurement


7.2.1 Blood Pressure Measuring Principle
Blood pressure may be measured in an invasive way (whereby the sensor will be inserted into blood vessel
directly) or a non-invasive way. The non-invasive way includes several methodologies, such as the Korotkoff
Sound Method and oscillating method. The Korotkoff Sound Method is used as a conventional way, whereby
stethoscope is used to measure the blood pressure. By the oscillating method, an inflation pump will fill the air,
and release it slowly. A computer will record change of the cuff pressure when the air is released. With this
record, the blood pressure value will be determined. First of all, make sure the signal quality judgment by
computer meets the requirements of accurate calculation (such as sudden limb movement or cuff being hit during
the measurement). If the answer is negative, give up the calculation. If the answer is positive, proceed with
calculation of the blood pressure value.

As change of the blood pressure is recorded by electric sensor, which sensitivity is much higher than that of
human ears, the oscillating method uses different definitions for measurement of diastolic pressure and systolic
pressure from the Korotkoff Sound Method. When the oscillating method is used, the circuit in the measuring
apparatus will separate the amplitude of the cuff pressure from its change with pulsation. The blood pressure at
amplitude of cuff pressure forward reduced according to proper proportion is defined as systolic pressure, while
the blood pressure at amplitude of cuff pressure backward reduced according to proper proportion is defined as
diastolic pressure. The maximum change of pulse pressure occurs at these two points. They are equivalent to the
point with pulse sound and the point without pulse sound respectively in the Korotkoff Sound Method.

When the risk of invasive monitoring method outweighs its advantage of accuracy, non-invasive monitoring
method shall be used.

Comparison between blood pressure measuring methods


To overcome the effect of human hearing variation and air release speed on measurement accuracy when the
conventional Korotkoff Sound Method is used to take measure of blood pressure, people have been dedicated to
study of automatic measurement of blood pressure. By now, automatic blood pressure measuring system based
on the principle of oscillating method is mature. In practice, however, various problems are encountered, such as
why the measures taken by the oscillating method is lower or higher than those taken by Korotkoff Sound
Method? Why the measures are inclined to decline? Why, in some cases, no result is obtained in spite of the
inflation actions? Why the measure values have big discreteness and even abnormal data in some cases? Why the
SpO2 waveforms may disappear suddenly? ...and so on. The following explanations are devised to give the
answers.

The Oscillating method vs. the Korotkoff Sound Method

Blood pressure measurement by the oscillating method and Korotkoff Sound Method has good correlation with
the invasive measurement. Notwithstanding, any of the non-invasive blood pressure measurements has its
one-sidedness when it is compared to the invasive measurement. The oscillating method has its advantages over
the Korotkoff Sound Method in less error, higher reliability and stability. Their differences may be reflected in the
following aspects.
1. The measures by the Korotkoff Sound Method are liable to effect of human factors. For example,
different people may have different sound judging ability, or different reactivity when listening to heart
sound and reading mercury meter. The air release speed and subjectivity may also affect the judgment.
By the oscillating method, the computation is accomplished by the computer, thus relieving the
possibility of effect due to human factor.

-34-
User manual for Patient Monitor

2. With the Korotkoff Sound Method, the measure is taken on the basis of appearance and disappearance of
heart sound. The air release speed and heart rate may have direct effect on the measurement accuracy. It
also has the disadvantages of rapid air release and poor accuracy. In the contrast, with the oscillating
method, the determination is calculated on the basis of cuff pressure oscillatory waveform envelope,
and the air release speed and heart rate has little effect on the measurement accuracy.
3. Statistics show that, when measuring the hypertension, the measure taken by the oscillating method is
likely to be lower than that taken by the Korotkoff Sound Method. When measuring the hypotension, the
measure taken by the oscillating method is likely to be higher than that by the Korotkoff Sound Method.
But, it doesn’t mean the advantages or disadvantages between the oscillating method and the Korotkoff
Sound Method. Comparison with the results taken by more accurate method, let’s say comparison of the
invasive pressure result with the output value by the blood pressure measuring simulator, will show
which method has more accurate results. In addition, higher or lower value should be a statistical concept.
It is recommended those used to adopt the Korotkoff Sound Method use different physiological
calibration for values determined by the oscillating method.
4. The studies have shown that the Korotkoff Sound Method has the worst accuracy when it comes to
measurement of hypotension, while the oscillating method has worse accuracy when it comes to
measurement of controlled hypertension relief.

7.2.2 Factors affecting NIBP measuring


Like common non-invasive blood pressure measurement, improper operation may cause inaccurate or blank result
or misunderstanding of the measuring information when the oscillating method is used to take the measure of
blood pressure. This point needs particular attention of the operators.
1. Requirements of the cuff:
1) Appropriate cuff should be selected according to the age of the subject. For more information, see
Chapter 3.
2) Remember to empty the residual air in the cuff before the measurement is commenced.
3) Locate the cuff in such a way that the artery mark “ ” is at a location where the clearest pulsation of
brachial artery is observed.
4) The cuff should be tightened to a degree where insertion of one finger is allowed.
5) The lower end of the cuff should be 2cm above the elbow joint.
2. The subject should lie on the back so that the cuff and the heart are in a horizontal position and the most
accurate measure is taken. Other postures may lead to inaccurate measurement.
3. Do not speak or move before or during the measurement. Care should be taken so that the cuff will not
be hit or touched by other objects. The air tube which connects the cuff and monitor should be
straightway without any tangle.
4. The measures should be taken at appropriate intervals. Continuous measurement at too short intervals
may lead to pressed arm, reduced blood flow and lower blood pressure, and resulting inaccurate measure
of blood pressure. It is recommended the measure be taken at intervals of more than two minutes.
5. With the oscillating method, when blood pressure is measured, the inflation pressure of the cuff will be
automatically adjusted according to the previous measure. Generally, the initial inflation pressure is
180mmHg (for the adult mode) or 100mmHg (for the pediatric mode) or 80 mmHg (for the neonate
mode) when it is powered on. Following that, 50mmHg (for the adult mode) or 30mmHg (for pediatric
mode) or 10mmHg (for the neonate mode) will be added on the basis of the last measurement of systolic
pressure. In this way, when the blood pressure rises or the subject is changed, the blood pressure meter
may fail in giving the result after the first-time inflation. This monitor will automatically adjust the

-35-
User manual for Patient Monitor

inflation pressure until the measure is taken, after that, up to four measures will be allowed.
6. When an adult subject is monitored, the machine may fail in giving the blood pressure measure if the
Pediatric or neonate is selected.
7. When taking NIBP measurement on pediatric patients, the operator must select correct mode depending
on different patient types (refer to NIBP menu setup) and do NOT operate in the adult mode. The high
inflation pressure for adult is not suitable for pediatric patients.

7.2.3 Clinical Limitations


1. Serious angiospasm, vasoconstriction, or too weak pulse.
2. When extremely low or high heart rate or serious arrhythmia of the subject occurs. Especially auricular
fibrillation will lead to unreliable or impossible measurement.
3. Do not take the measurement when the subject is connected with an artificial heart-lung machine.
4. Do not take the measurement when the subject uses diuresis or vasodilator.
5. When the subject is suffering from major hemorrhage, hypovolemic shock and other conditions with
rapid blood pressure change or when the subject has too low body temperature, the reading will not be
reliable, for reduced peripheral blood flow will lead to reduced arterial pulsation.
6. Subject with hyperadiposis;
In addition, statistics show that 37% people report blood pressure difference of no less than 0.80kPa(6mmHg)
between the left and right arms, and 13% people report difference of no less than 1.47kPa (11mmHg).

Note: Some practitioners may report big discreteness or abnormal value of the blood pressure measures
when the oscillating method is used. As a matter of fact, the so-called “big discreteness” must be a term in
the sense of statistical significance of mass data. Abnormal data may be observed in some individual cases.
It is normal in the scientific experiments. It may be caused by an apparent reason, or by an unknown
factor in some cases. Such individual doubtful experimental data may be identified and eliminated using
the special statistical technique. It is not a part of this manual. The practitioner may eliminate the
apparently unreasonable data according to the experience.
Operation Introduction:
1. Take a measurement in manual mode:
 Enter into the screen of NIBP setting, select “Mode” option and set it as “MANU”, and then press
the NIBP key on the front panel to start measure. If press the NIBP key again, the measurement
will be stopped.

7.3 Introduction to Oxygen Saturation Measurement


7.3.1 SpO2 Measuring Principle
Based on Lamber-Beer law, the light absorbance of a given substance is directly proportional with its density
or concentration. When the light with certain wavelength emits on human tissue, the measured intensity of
light after absorption, reflecting and attenuation in tissue can reflect the structure character of the tissue by
which the light passes. Due to that oxygenated hemoglobin (HbO2) and deoxygenated hemoglobin (Hb) have
different absorption character in the spectrum range from red to infrared light (600nm~1000nm wavelength),
by using these characteristics, SpO2 can be determined. SpO2 measured by this monitor is the functional
oxygen saturation -- a percentage of the hemoglobin that can transport oxygen. In contrast, hemoximeters
report fractional oxygen saturation – a percentage of all measured hemoglobin, including dysfunctional
hemoglobin, such as carboxyhemoglobin or metahemoglobin.

-36-
User manual for Patient Monitor

7.3.2 SpO2 Measurement Restrictions (interference reason)


1. The fingers should be properly placed (see the attached illustration of this instruction manual), or else it
may cause inaccurate measurement result.
2. Make sure that capillary arterial vessel beneath the finger is penetrated through by red and infrared
lights.
3. The SpO2 sensor should not be used at a location or limb tied with arterial or blood pressure cuff or
receiving intravenous injection.
4. Do not fix the SpO2 sensor with adhesive tape, or else it may result in venous pulsation and
consequential inaccurate measurement result of SpO2.
5. Make sure the optical path is free from any optical obstacles like adhesive tape.
6. Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light,
infrared heater, and direct sunlight etc.
7. Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.
8. Please do not use the SpO2 sensor when having the MRI, or burn may be caused by faradism.
9. Always observe the plethysmogram (waveform), which is auto-scaled within the range of 100. The SpO2
reading may be unlikely true when the waveform is not smooth or irregular. If in doubt, rely on your
clinical judgment, rather than the monitor readout
10. A functional tester cannot be used to assess the accuracy of the pulse oximeter monitor or a SpO2 sensor.
However, a functional tester, such as SpO2 simulator can be used to check how accurately a particular
pulse oximeter is reproducing the given calibration curve. Before testing the oximeter by a functional
tester, please firstly ask the manufacturer which calibration curve is used, if necessary, request the
manufacturer for its dedicated calibration curve and download it into the tester.
Clinical Limit
1. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood stream of subject is
required. For a subject with weak pulse due to shock, low ambient/body temperature, major bleeding, or
use of vascular contracting drug, the SpO2 waveform (PLETH) will decrease. In this case, the
measurement will be more sensitive to interference.
2. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green and
acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic
hemoglobin, and some with icterus problem, the SpO2 determination by this monitor may be inaccurate.
3. The drugs such as dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor
blamed for serious error of SpO2 measurements.
4. As the SpO2 value serves as a reference value for judgment of anemic anoxia and toxic anoxia, the
measurement result of some patients with serious anemia may also present as good SpO2 value.
Note: The clinical study for SpO2 measurement accuracy was done on the subjects including Male and
Female with skin color including Medium, light medium, light and dark , No subject was anemic
(Hemoglobin≤10 gm·d1-1) and only healthy non-smoking individuals were tested, their age rang was from
21-49.

7.4 Introduction to Respiration Measurement

7.4.1 Respiration Measuring Principle


The air will be filled into alveolus or be expelled during respiration, and the chest’s volume changes with
this process. Because the conductivity of air is lower than body tissues, the chest’s impedance will be
changed by the inflation. With this specialization, the respiration can be monitored through putting safe

-37-
User manual for Patient Monitor

current into body and measuring the change of voltage between the electrodes. The product will transmit the
high-frequency current whose frequency is much higher than ECG frequency (but with the safe current limit)
to the ECG electrodes (placed at the both sides of chest), which can be detect ECG signal and chest’s
impedance at the same time, and the respiratory rate will be measured through impedance method by the
software. So the additional electrodes for respiratory measurement are unnecessary.

7.4.2 Factors affecting respiration monitoring


 Respiration monitoring doesn’t support monitoring the patient who do much movement, or may lead to
false alarm.

7.5 Introduction to Temperature Measurement


The sensor is thermo-resistor type (25°C 5kΩ) and is supplied with constant micro current. Calculating the
temperature of measured part through measuring the voltage. There is a period responding time, so the
accurate temperature value display after a while. The temperature monitoring can be divided into two
measuring method: measure through body surface temperature (placed in).

Normal value: body surface: 36.5°C ~37°C;

Notes:

 Attach the TEMP transducer to the patient; generally if the TEMP transducer and skin doesn’t contact
closely, the measured value becomes lower, so for those who have requirement for temperature, add a
proper martial to transducer and fix it with adhesive tape to make them contact firmly.
 Especially for pediatric patient, they like sports, pay more attention to the transducer fixing.

7.6 Introduction to Capnograph Measurement


7.6.1 CO2 Measuring Principle
The principle is based on the fact that CO2 molecules absorb infrared light energy of specific wavelengths, with
the amount of energy absorbed being directly related to the CO2 concentration. When an IR light beam is passed
through a gas sample containing CO2, the electronic signal from a photodetector (which measures the remaining
light energy), can be obtained. This signal is then compared to the energy of the IR source, and calibrated to
accurately reflect CO2 concentration in the sample. To calibrated, the photodetector’s response to a known
concentration of CO2 is stored in the monitor’s memory.
The monitor determines CO2 concentration in the breathing gases by measuring the amount of light absorbed by
these gases.EtCO2 is display as a numerical value in millimeters of mercury (mmHg), percent (%), or kilopascals
(kPa). In addition, a CO2 waveform (capnogram) may be displayed which is a valuable clinical tool that can be
used to assess patient airway integrity and proper endotracheal tube placement. Respiration rate is calculated by
measuring the time interval between detected breaths.

7.6.2 Mainstream vs. Sidestream Sampling


Mainstream CO2 sensors are placed at the airway of an intubated patient, allowing the inspired and expired gas to
pass directly across the IR light path. The major advantages of mainstream sensors are fast response time and
elimination of water traps.

-38-
User manual for Patient Monitor

Sidestream CO2 sensors are located away from the airway, requiring a gas sample to be continuously aspirated
from the breathing circuit and transported to the sensor by means of a pump. This type of system is needed for
non-intubated patients.
When using mainstream CO2 sensors, check the window for the patient secretions pooled on periodically.
Because that condition may affect the accuracy of the measurement or even make the sensor not work.
When using sidestream CO2 sensors, there is a water trap or a part of the sampling tube with dehumidifying
function. Please periodically check the flow sensor and tubing for excessive moisture or secretion buildup.

-39-
User manual for Patient Monitor

Chapter 8 Troubleshooting
Note: In case of trouble of this machine in service, follow the instructions below to eliminate the problem first. If
the attempt fails, contact the dealer in your local area or the manufacturer.

 Do NOT open the monitor cabinet without permission

8.1 No Display on the Screen


Shut down the machine and unplug the power cable. Use a universal meter to check if the outlet has proper
voltage, if the power cable is in good condition, and if the power cable is properly connected with this apparatus
or outlet. Remove the fuse from the back cover of this machine, and make sure it is in good condition.If all of
above is in good condition,the display screen may be malfunction.

8.2 Excessive ECG Signal Interference or too Thick Baseline


1. Check if the plate electrodes are properly located, and if valid plate electrodes are used.

2. Check whether the lead wires are properly inserted. If no ECG curve displayed, check if the ECG lead wires
are broken.

3. Make sure the mains outlet has standard grounding wire.

4. Check if the grounding wire of the apparatus properly grounded.

8.3 No Blood Pressure and Pulse Oxygen Measures


1. Check if the blood pressure cuff is properly wrapped around the arm according to the operating instructions,
if the cuff leaks, and if the inlet is closely connected with the NIBP jack on the side panel. Check if the
indicator of the pulse oxygen sensor flashes and if the pulse oxygen probe is properly connected to the SpO2
jack on the side panel.

2. If the problems still exist, please contact the manufacturer.

8.4 System Alarm


1. When the parameter value is higher or lower than the alarm limits, the alarm will ring. Please check whether
the alarm limit value is proper or the condition of the patient.

2. Leads off. Please check the connection of the leads.

3. Probe off. Please check the connection of the probes.

-40-
User manual for Patient Monitor

Chapter 9 Maintenance
In case of trouble of this machine in the service, follow the instructions below to eliminate the problem first.
If the attempt fails, refer to the dealer in your local area or the manufacturer. Refer to the detailed
provisions in contract for the warranty period of the main unit and the accessories of this monitor.

9.1 Service and Examination


9.1.1 Daily Examination
Before using the monitor, the checks below should be carried out:
 Check the monitor for any mechanical damage;
 Inspect the exposed parts and the inserted parts of all the cables and the accessories;
 Examine all the functions of the monitor that are likely to be used for patient monitoring, and ensure
that it is in good working condition;
 Make sure that the monitor is grounded properly.
 Pay close attention to the fluctuation of the local power supply voltage. A power voltage regulator is
recommended when necessary.
In case any indication of damage about the function of the monitor is detected and proven, it is not allowed to
apply it to the patient for any monitoring. Please contact the local dealer or our company, and we are to offer the
best solution as soon as possible for your satisfaction.

9.1.2 Routine Maintenance


At each routinely maintenance or the yearly maintenance, the monitor can be thoroughly inspected by qualified
personnel, including performance and safety examinations. This monitor is designed with life cycle of 5 years. It
is strongly recommended to use the product which is still within its life cycle, or it may cause inaccurate
measurement. In order to ensure its long service life, please pay attention to the maintenance.
 If the hospital fails to carry out a satisfactory maintenance program about the monitor, it may get
disabled and harm the patient’s safety and health.
 In case of ECG cable/lead wires damage or aging, please replace the cable or lead wires.
 If there is any indication of cable and transducer damage or they deteriorate, they are prohibited
from any further use.
 The Monitor is calibrated in the factory before sale, so there is no need to calibrate it during its life
cycle. Any patient simulators should not be used to validate the accuracy of blood pressure and
oxygen saturation measurement, they can only be used as functional testers to verify its precision.
 The accuracy of ECG signal amplification can be verified by the built-in 1mV calibration signal,
please refer to the related chapter in Part 2 of the user manual for detail operation.
 The accuracy of pressure measurement and air leakage in pneumatic system can be verified by
means of the built-in pressure verification function and a precision pressure meter, please refer to
the related chapter in Part 2 of the user manual for detail operation.
 The SpO2 simulator can not be used to verify the SpO2 measuring accuracy, which should be
supported by the clinical study conducted by inducing hypoxia on healthy, non-smoking,
light-to-dark-skinned subjects in an independent research laboratory. However it is necessary for
the user to use SpO2 simulator for routine verification of precision.
 Please note that the specific calibration curve (so called R-curve) should be selected when use of
SpO2 simulator,e.g. for Index 2 series SpO2 simulator from Fluke Biomecidal Corporation, please

-41-
User manual for Patient Monitor

set "Make" to "DownLoadMake: KRK", then the user can use this particular R-curve to test the
SpO2 function. If the SpO2 simulator does not contain "KRK" R-curve, please ask the
manufacturer for helping to download the given R-curve into the SpO2 simulator.
 The adjustable units within the monitor such as potentiometers are not allowed to adjust without
permission to avoid unnecessary failures that affect normal application.
 It is recommended to use the battery once a month to ensure its power capability and long service
life, and recharge it after run out of its power capacity.

9.2 Battery Maintenance


 Please pay attention to the polarity of battery, do NOT insert it into battery compartment with
reversed polarities;
 Do NOT use the batteries manufactured by other companies, if being inserted, the device will
may be damaged;
 In order to avoid damaging the battery, do NOT use other power supply device to charge the
battery;
 After battery ageing phenomenon occurring, do NOT throw the battery into fire to avoid
explosion risk.
 Do not hit or strike it with force;
 Do not use this battery on other devices;
 Do not use this battery below -10°C or above 40°C;
 Dispose of the battery, the local law should be followed.
 In order to maintain battery supply time and prolong battery lifetime, please charge the battery
every one or two months if don’t use battery for a long time. And do charge battery at least 12-15
hours every time. Before connect to AC, do start monitor with battery’s power supply, until
battery power is used up and monitor turn off automatically, then connect monitor to AC and
have it charged for 12-15 hours continuously. The speed of charge will be the same no matter
whether the monitor is working or not. The reason why discharge the battery before charge is to
avoid the decrease of capacity caused by battery’s memory effect. If the monitor won’t be used
for a long time, do have it charged fully before conservation.
 If battery is damaged, please replace with same type and specification battery marked by “CCC”
or “CE” in time, or contact the company directly.

9.3 Cleaning and Disinfection of the Device


 Switch off the monitor and disconnect the power cable before cleaning.
 Kept the monitor from dust.
 It is recommended to clean the outer shell and screen of the monitor to keep it clean. Only non-corrosive
cleanser such as clear water is permitted.
 Wipe the surface of the monitor and transducers with an alcohol impregnated wipe, and dry it with dry
and clean wipe or simply air-dry.
 This monitor can be disinfected, please clean the monitor firstly.
 Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage.
 Clean the exterior of the connector only.
 Dilute the cleanser.

-42-
User manual for Patient Monitor

 Do not use scrub materials.


 Do not let any liquid flow into the shell or any parts of the monitor.
 Do not let the cleanser and disinfectant stay on its surface.
 Do not perform high pressure sterilization to the monitor.
 Do not put any parts of the monitor or its accessories in the liquid.
 Do not pour the disinfectant on its surface while disinfectant.
 If the monitor is accidentally wetted it should be thoroughly dried before use. The rear cover can
be removed by qualified service technician to verify absence of water.
 Never use this machine in an environment with inflammable gas.
 Avoid being hit by lightning. The power cable should be plugged into an outlet with grounding
wire. Do not use an outlet with poor condition. If possible, use power supply system with
regulator.
 It must be used in a clean environment protected against shock. Keep it away from corrosive
substances, explosive substances, high temperature and dampness.
 If it is installed in a cabinet, make sure the installation allows for good ventilation, and easy
maintenance, observation and operation.

9.4 Cleaning and Disinfection of Accessories


It is recommended to clean the accessories (including sensor, leads and plugs) with a piece of gauze which has
been soaked in 75% Alcohol or 70% Isopropanol before use.
 Do not use damaged accessories.
 Accessories can not be entirely immerged into water, liquor or cleanser.
 Do not use radiation, steam or EO to disinfect accessories.
 Do wipe off the remained alcohol or ispropanol on the accessories after disinfection, for good
maintenance can extend the life of accessories.

9.5 Storage
If the equipment will not be used for long period of time, wipe it clean and keep it in the packaging, which shall
be kept in a dry and good ventilation place free from dust and corrosive gases

Storage environment: ambient temperature: -20~60°C


relative humidity: 10%~95%
atmospheric pressure: 53kPa~106kPa

9.6 Transportation
This monitor should be transported by land (vehicle or railway) or air in accordance with the contractual terms.
Do not hit or drop it with force.

-43-
User manual for Patient Monitor

Chapter 10 Appendix

10.1 Alarm Information


Alarm Information Descriptions
Over HR limit
Over RR limit
Over TEMP limit
Over SpO2 limit
The related parameter value exceeds the preset high/low alarm
Over PR limit
limits.
Over NIBP SYS limit
Over NIBP DIA limit
Over NIBP PR limit
Unable to detect HR ECG cable and leads are connected to monitor and patient well,
but HR is unable to be detected. It may caused by inconformity
HR signal.
Unable to detect SpO2 SpO2 probe is connected to monitor and patient well, but SpO2 is
unable to be detected. It may be caused by inconformity SpO2
signal.
The battery capacity will exhaust Low battery voltage
Lead Off The ECG electrodes or cable fell off
Probe Off SpO2 probe fell off

-44-
User manual for Patient Monitor

10.2 Default Alarming Values and Setup Range


The default alarming value:
Mode
Parameter Adult pediatric Neonate
High limit 180 bpm 200 bpm 220 bpm
Heart Rate
Low limit 40 bpm 50 bpm 50 bpm
High limit 40 rpm 50 rpm 60 rpm
Respiration
Low limit 10 rpm 10 rpm 10 rpm
High limit 39℃ 39℃ 39℃
Temperature
Low limit 35℃ 35℃ 35℃
High limit 180 mmHg 130 mmHg 110 mmHg
Systolic
Low limit 60 mmHg 50 mmHg 50 mmHg
High limit 120 mmHg 90 mmHg 90 mmHg
NIBP Diastolic
Low limit 50 mmHg 40 mmHg 30 mmHg
High limit 160 mmHg 110 mmHg 100 mmHg
MAP
Low limit 50 mmHg 40 mmHg 30 mmHg
High limit 100% 100% 100%
SpO2
Low limit 90% 85% 85%
High limit 180 bpm 200 bpm 220 bpm
Pulse Rate
Low limit 40 bpm 50 bpm 50 bpm
High Limit +1.00mV +1.00mV +1.00mV
ST Segment
Low Limit -1.00mV -1.00mV -1.00mV
Temperature Difference Range 2℃ 2℃ 2℃
High limit 200mmHg 160mmHg 140mmHg
SYS
Low limit 10mmHg 10mmHg 10mmHg
Arterial High limit 200mmHg 160mmHg 140mmHg
Pressure DIA
Low limit 10mmHg 10mmHg 10mmHg
High limit 200mmHg 160mmHg 140mmHg
MAP
Low limit 10mmHg 10mmHg 10mmHg
High limit 120mmHg 100mmHg 90mmHg
SYS
Low limit 10mmHg 10mmHg 10mmHg
Pulmonary High limit 120mmHg 100mmHg 90mmHg
Artery DIA
Pressure Low limit 10mmHg 10mmHg 10mmHg
High limit 120mmHg 100mmHg 90mmHg
MAP
Low limit 10mmHg 10mmHg 10mmHg
High limit 30mmHg 30mmHg 30mmHg
SYS
Low limit 0mmHg 0mmHg 0mmHg
Central High limit 30mmHg 30mmHg 30mmHg
Venous DIA
Low limit 0mmHg 0mmHg 0mmHg
Pressure
High limit 30mmHg 30mmHg 30mmHg
MAP
Low limit 0mmHg 0mmHg 0mmHg
Respiration High limit 40 rpm 50 rpm 60 rpm
Rate Low limit 10 rpm 10 rpm 10 rpm
High limit 70 mmHg 70 mmHg 70 mmHg
CO2 EtCO2
Low limit 10 mmHg 10 mmHg 10 mmHg
High limit 10 mmHg 10 mmHg 10 mmHg
InsCO2
Low limit 0 mmHg 0 mmHg 0 mmHg

-45-
User manual for Patient Monitor

The high and low limits setting range:

Parameter Mode Adult Pediatric Neonate


High limit 1~300bpm 1~350bpm 1~350bpm
Heart Rate
Low limit 0~299bpm 0~349bpm 0~349bpm
High limit 1~120rpm 1~150rpm 1~150rpm
Respiration
Low limit 0~119rpm 0~149rpm 0~149rpm
High limit 0.1~50°C 0.1~50°C 0.1~50°C
Temperature
Low limit 0~49. 9°C 0~49. 9°C 0~49. 9°C
High limit 31~270 mmHg 31~200 mmHg 31~135 mmHg
Systolic
Low limit 30~269 mmHg 30~199 mmHg 30~134 mmHg
High limit 11~232 mmHg 11~150 mmHg 11~100 mmHg
Diastolic
Low limit 10~231 mmHg 10~149 mmHg 10~99 mmHg
High limit 21~242 mmHg 21~165 mmHg 21~110 mmHg
Mean
Low limit 20~241 mmHg 20~164 mmHg 20~109 mmHg
High limit 1~100% 1~100% 1~100%
SpO2
Low limit 0~99% 0~99% 0~99%
High limit 1~300bpm 1~350bpm 1~350bpm
Pulse Rate
Low limit 0~299bpm 0~349bpm 0~349bpm
High Limit -2.49Mv~+2.49mV -2.49mV~+2.49mV -2.49mV~+2.49mV
ST Segment
Low Limit -2.49mV~+2.49mV -2.49mV~+2.49mV -2.49mV~+2.49mV
Temperature Difference 0.0~5.0℃ 0.0~5.0℃ 0.0~5.0℃
High limit (1~250)mmHg (1~250)mmHg (1~250)mmHg
Systolic
Low limit (0~249)mmHg (0~249)mmHg (0~249)mmHg
li limit
High i (1~250)mmHg (1~250)mmHg (1~250)mmHg
Arterial
Diastolic
Pressure Low limit (0~249)mmHg (0~249)mmHg (0~249)mmHg
High limit (1~250)mmHg (1~250)mmHg (1~250)mmHg
Mean
Low limit (0~249)mmHg (0~249)mmHg (0~249)mmHg
High limit (1~120)mmHg (1~120)mmHg (1~120)mmHg
Systolic
Low limit (0~119)mmHg (0~119)mmHg (0~119)mmHg
Pulmonary High limit (1~120)mmHg (1~120)mmHg (1~120)mmHg
Artery Diastolic
Pressure Low limit (0~119)mmHg (0~119)mmHg (0~119)mmHg
High limit (1~120)mmHg (1~120)mmHg (1~120)mmHg
Mean
Low limit (0~119)mmHg (0~119)mmHg (0~119)mmHg
High limit (-9~40)mmHg (-9~40)mmHg (-9~40)mmHg
Systolic
Low limit (-10~39)mmHg (-10~39)mmHg (-10~39)mmHg
Central High limit (-9~40)mmHg (-9~40)mmHg (-9~40)mmHg
Venous Diastolic
Pressure Low limit (-10~39)mmHg (-10~39)mmHg (-10~39)mmHg
High limit (-9~40)mmHg (-9~40)mmHg (-9~40)mmHg
Mean
Low limit (-10~39)mmHg (-10~39)mmHg (-10~39)mmHg
Respiration High limit (1~120)rpm (1~150)rpm (1~150)rpm
Rate Low limit (0~119)rpm (0~149)rpm (0~149)rpm
High limit (1~100)mmHg (1~100) mmHg (1~100) mmHg
CO2 EtCO2
Low limit (0~99)mmHg (0~99) mmHg (0~99) mmHg
High limit (1~30)mmHg (1~30) mmHg (1~30) mmHg
InsCO2
Low limit (0~29)mmHg (0~29) mmHg (0~29) mmHg

-46-
User manual for Patient Monitor

10.3 Status/Error during NIBP Monitoring


“Cuff error” —cuff is not wrapped correctly, or is not connected
“Air leak” —Air moving part, tube or the cuff leak air.
“Pressure error” —Unstable cuff pressure or tangled cuff tubing
“Signal weak” —Very weak signal because of the cuff, or the patient has very weak pulse
“Over extent” —The measurement range exceeds 255 mmHg (Pediatric patient over 135 mmHg)
“Over motion” —The repeated measurement due to moving, excessive noise during the stepping inflation
and measuring pressure and pulse, e.g. during patient shaking motion
“Signal overflow” —Blood pressure amplifier overflow due to excessive movement
“Leak in gas run” —Leaking during the pneumatic device testing
“System error” —Abnormal condition of CPU, such as register overflow, divided by zero
“Adult” —The blood pressure measuring now is in adult mode. In this case, it is not allowed to
monitoring Pediatric or neonatal patient. Otherwise, there may be serious danger to the
Pediatric monitored.
“Pediatric” —The blood pressure module is now worked in Pediatric measuring mode.
“PROBE OFF” —SpO2 probe fell off
“LEADS OFF” —The ECG electrodes or cable fell off
“LEARNING” —Learning arrhythmia for 15 seconds
“DEMO” —The monitor is displaying the demo waveforms, which are generated by the monitor itself.

-47-
User manual for Patient Monitor

10.4 Status/Error during CO2 Monitoring


Suggested Message/Response Description
“Sensor Over Temp” The sensor temperature is greater than 40 °C.
Make sure sensor is not exposed to extreme heat (heat
lamp, etc.). If error persists, return sensor to factory for
servicing.
“Sensor Faulty” One of the following conditions exist:
Check that the sensor is properly plugged in. Reinsert Source Current Failure,
or reset the sensor if necessary. If error persists, return EEPROM Checksum Faulty,
sensor to factory for servicing Hardware Error

No Parameter Message Barometric Pressure and/or gas compensations have not been
The host must set the Barometric Pressure and set since power on. For CO2 to be calculated with the stated
compensations to clear this error; no user intervention accuracy, these values should be set whenever the sensor is
should be required. plugged in.

“Module in Sleep Mode” This bit is set when sensor has been placed in sleep mode.
“Zero In Progress “ A Module Zero is currently in progress.
One of the following conditions exist:
“Sensor Warm Up”
Sensor under temperature
This error condition is normal at startup. This error
Temperature not stable
should clear when the warm up is complete.
Source Current unstable
“Check Sampling Line” This error occurs whenever the pneumatic pressure is outside
Check that the sampling line is not occluded or kinked. the expected range.
“Zero Required” One of the following conditions exist:
To clear, check airway adapter and clean if necessary. Zero Required;
If this does not correct the error, perform an adapter Zero Required: Zero Error
zero. If you must adapter zero more than once, a
possible hardware error may exist.
The value being calculated is greater than the upper CO2 limit
“CO2 Out of Range”
(150 mmHg, 20.0 kPa, or 19.7 %). The maximum value output
If error persists, perform a zero. is the upper CO2 limit.
Usually caused when the airway adapter is removed from the
“Check Airway Adapter” sensor or when there is an optical blockage on the windows of
To clear, clean airway adapter if mucus or moisture is the airway adapter. May also be caused by failure to perform
seen. If the adapter is clean, perform a Capnostat zero. sensor zero to when adapter type is changed.

This is prompted if the CO2 sensor is not ready for a Capnostat Zero.
If the “Zero Required” and this massage both prompt message both
The Sensor promptone or more of the following conditions may exist:
not Ready • Breaths detected
• Temperature is not stable
• Source Current unstable
• In sleep mode.

Zero in already progress Normal zero calibration is in already progress.


Zero Fault and Breaths Detected Zero attempted and breaths have been detected in the last 20
seconds.
Zero Ok Zero calibration is successful

-48-
User manual for Patient Monitor

10.5 Typical Pressures and CO2 Readings at Altitudes

Barometric EtCO2 Reading


Altitude
Pressure(mmHg) (%) (mmHg)
0m 760 5 38.0
70m 754 5 37.7
100m 751 5 37.5
200m 743 5 37.1
1500m 641 5 32.0
3000m 537 5 26.8
5000m 420 5 21.0

10.6 Accessories List


Part No. Part Name Remark
15010513 ECG cable
5101-0101310 ECG electrode
15044051 Adult SpO2 Finger clip Sensor
15044041 Pediatric SpO2 Finger clip Sensor Optional
15024402 Adult NIBP cuff(25~35cm)
15021402 Small-sized Pediatric NIBP Cuff Optional
15022402 Middle-sized Pediatric NIBP Cuff Optional
15023402 Large-sized Pediatric NIBP Optional
15084120 Skin TEMP probe
15100420 CO2 Mainstream sensor Optional for mainstream
15100411 Adult airway adapter Optional for mainstream
15100421 Pediatric airway adapter Optional for mainstream
15100410 CO2 Sidestream sensor Optional for sidestream
15100130 Sampling line kit Optional for sidestream
15100214 Extending airway tube Optional for sidestream
15100210 Wye connector Optional for sidestream
2903-0000000 Power cord
900093 Net wire

For more information regarding the accessories, please contact your local sales representative or the
manufacturer.

Note: Part No. is subject to change without prior notice, please refer to the label of parts or package list.

-49-
User manual for Patient Monitor

Shenzhen Creative Industry Co., Ltd.


2/F, Block 3, Nanyou Tian'an Industry Town,
518054 Shenzhen, GD, P.R. China

Tel: +86-755-2643 3514


Fax: +86-755-2643 0930
E-mail: [email protected]
Website: www.creative-sz.com

EC-Representative:
Shanghai International Holding Corp. GmbH (Europe)
Address: Eiffestrasse 80, 20537 Hamburg, Germany

-50-

You might also like