IT-M3900C User Manual-EN
IT-M3900C User Manual-EN
DC Power Supply
IT-M3900C Series User Manual
Warranty
ITECH warrants that the product will be free from defects in material and work-
manship under normal use for a period of one (1) year from the date of delivery
(except those described in the Limitation of Warranty below).
For warranty service or repair, the product must be returned to a service center
designated by ITECH.
• The product returned to ITECH for warranty service must be shipped PRE-
PAID. And ITECH will pay for return of the product to customer.
• If the product is returned to ITECH for warranty service from overseas, all
the freights, duties and other taxes shall be on the account of customer.
Limitation of Warranty
This Warranty will be rendered invalid in case of the following:
• The product model or serial number is altered, deleted, removed or made il-
legible by customer;
Safety Symbols
Working temper- - -
ature of the
machine.
Safety Precautions
The following safety precautions must be observed during all phases of opera-
tion of this instrument. Failure to comply with these precautions or specific warn-
ings elsewhere in this manual will constitute a default under safety standards of
design, manufacture and intended use of the instrument. ITECH assumes no li-
ability for the customer’s failure to comply with these precautions.
Environmental Conditions
The instrument is designed for indoor use and an area with low condensation.
The table below shows the general environmental requirements for the
instrument.
Installation category II
Regulation Tag
Compliance Information
Complies with the essential requirements of the following applicable European
Directives, and carries the CE marking accordingly:
EMC Standard
Reference Standards
IEC 61000-3-2
IEC 61000-3-3
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
2. Connection of the instrument to a test object may produce radiations beyond the specified
limit.
3. Use high-performance shielded interface cable to ensure conformity with the EMC standards
listed above.
Safety Standard
IEC 61010-1:2010+A1:2016
This section provides general and universal terms. For more detailed infor-
mation on the Declaration of Conformity, please contact ITECH personnel.
Content
1 Quick Reference
This Chapter will introduce power-on check steps of this series to ensure normal
start-up and usage under initialization status of the power. Besides, to facilitate
usage, this part also displays the functions of front board, rear board and key-
board keys as well as display functions to a quick view of power appearance,
structure and key usage functions before operation.
♦ Brief Introduction
♦ Front-Panel Overview
♦ Keyboard Introduction
♦ Push-on Knob
♦ Rear Panel Introduction
♦ VFD Indicator Lamps Description
♦ Configuration Menu Function
♦ System Menu Function
♦ Options Introduction
FEATURE
• Power range:+/-12kW
• Wide range of output design, one unit can be used as multiple power
supplies
• Bidirectional energy flow between the DUT and grid, seamless switching
across quadrants
• CC/CV priority
• Adjustable output impedance
• List function
IT-M3902C-800-8 8A 2kW 1U
• As a bidirectional power supply, this series can output and absorb electric en-
ergy. The current and power input capacity is same as the output capacity. For
example, the maximum rated value of the output power of the IT3906C-80-120
model is 6000W, and the maximum rated value of input power is also 6,000W.
1U Model
2U Model
1U Model
2U Model
Keys Description
[On/Off] Turn the power supply output on or off. When lit, indicates
that the output is enabled or on.
Keys Description
. Decimal point
Left / Right The left and right navigation keys are used to adjust the cur-
navigation sor to the specified position or scrolls pages to view menu
keys items.
Composite key [Shift], combined with other keys to realize functions marked
above keys. The detailed functions are listed as follows.
Keys Description
In the value setting interface, rotate the knob clockwise to increase the set value
and anticlockwise to decrease the set value.
The knob can also be used to view menu items. In the menu item display inter-
face, turning the knob clockwise indicates that the next menu item is selected,
and turning the knob anticlockwise indicates that the previous menu item is
selected.
Confirm settings
After completing the value setting or selecting a menu item, pushing the knob
acts like pressing [Enter] key to confirm the operation.
1U Models
2U Models
This interface is used for the parallel connection between the units for the
communication of units in parallel.
9. AC power input terminals (L1, L2, L3, and PE)
1. Press the composite keys [Shift]+[V-set] (Config) on the front panel to enter
the configuration menu.
The VFD shows configuration menu items and each menu item is marked
with a number. The user can use knob to scroll pages to view the other menu
items.
3. After finishing the menu setting, press [Enter] key to save the setting.
The descriptions of configuration menu of the power supply are listed in the ta-
ble below.
Loop response
speed setting:
Speed= High/Low
high speed /
low speed
Loop response
speed setting:
Speed= High/Low
high speed /
low speed
Off Delay Set the delay time to turn off the output.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
At this point, the VFD screen displays the system menu items. Each menu
item has a numbering identifier. The user can use the knob for scrolling dis-
play of other menu items.
3. After the menu items are set, press [Enter] again to save the modified
contents.
Used to control whether the voltage is quickly zeroed after the output is
turned off.
System Rzero • Off: No
• On: Yes
View the system information. For details, see 5.13 View the System Informa-
System Info
tion (System Info).
The accessory has a total of 10 pins, and the details of each pin are as
follows:
Pins Description
8/9/ Terminals for external analog function. For details, see 5.12
10 Analogue Function (Ext-Program) (Optional).
Fiber optic modules and cables are used for data transmission and commu-
nication between parallel units, and are highly resistant to interference. The
fiber optic module and cable are the necessary accessories for the parallel
connection. Different numbers of fiber optic modules and cables are used in
different numbers of parallels.
IT-E168: Used for parallel connection between the units, including one fiber
module and two fiber cables, with lengths of 1.5 meters and 0.3 meters.
It is suitable for motor products, which can prevent the electromotive force
generated during the operation of the motor from damaging the power sup-
ply products.
• Cabinet installation
Upon verification of the shipment, keep the package and relevant contents
thereof in a safe place. When returning the instrument for warranty service or
repair, the specified packing requirements shall be met.
The detailed dimension drawings of this series instrument are as follows (unit:
mm, deviation: ±1 mm):
1U Models
2U Models
To prevent electric shock and damage to the instrument, observe the following
precautions.
• Single-phase utility, the Output DC power limit is, P = phase voltage * maxi-
mum input current * efficiency.
• Three-phase utility, the Output DC power limit is, P = line voltage * maximum
input current * efficiency * 1.732.
Wiring Diagram:
• When the 1U and 2U models are connected to three-phase power, the wir-
ing diagram is shown below.
• When the 1U model is connected to single-phase power (L/N), the wiring di-
agram is shown below.
Operation procedures:
• For 1U/2U model, see the steps below to connect the power cable.
2. Confirm that the power switch is in the OFF position and verify that there
is no dangerous voltage on the connection terminals.
3. Remove the protective cover outside the AC input terminal on the rear
panel.
4. Connect one end of the power cable’s round terminal to the AC power in-
put terminal on the instrument’s rear panel.
6. Refer to the suggestion connection diagram, connect the other end of the
power cable to the required AC distribution box.
Precautions
To prevent electric shock and damage to the instrument, observe the following
precautions.
Test cables are not standard accessories for the instrument. Please select op-
tional red and black test cables for individual sales based on the maximum cur-
rent value. For specifications of test cables and maximum current values, refer
to A.1 Appendix→Specifications of Red and Black Test Lines.
The instrument supports two kinds of wiring methods with the DUT: local meas-
urement and remote measurement (SENSE). The default test mode is local
measurement.
1. Confirm that the power switch is in the OFF position and verify that there is
no dangerous voltage on the connection terminals.
2. Remove the output terminals cover of the power supply.
3. Loosen the screws of the output terminals and connect the red and black
test cables to the output terminals. Re-tighten the screws.
When maximum current that one test cable can withstand fails to meet the
rated current, use multiple pieces of red and black test cables. For example,
the maximum current is 1,200A, then 4 pieces of 360A red and black cables
are required.
4. Thread the red and black test cables through the output terminals cover of
the power supply and install the cover.
5. (Optional) According to the actual situation of DUT, connect the grounding
terminal on the rear panel of the instrument to the DUT to ensure the safe
grounding.
For the location information, see 1.5 Rear Panel Introduction.
6. Connect the other end of the red and black cables to the DUT. The positive
and negative poles must be properly connected and fastened when wiring.
When the DUT consumes large current or the wires are too long, there is a volt-
age drop on the wires between DUT and output terminals of the power supply.
To maximize measurement accuracy, the power supply provides the remote
measurement terminals VS+ and VS- on the rear panel, which can be used to
measure the terminal voltage of the DUT.
When the power supply is used for battery testing in actual applications, the
voltage drop of the wire will lead to voltage inconsistency of both ends and
inconsistency of the cutoff voltage of power supply and the actual voltage of bat-
tery, resulting in inaccurate measurement.
1. Confirm that the power switch is in the OFF position and verify that there is
no dangerous voltage on the connection terminals.
2. Refer to the wiring diagram and connect the Vs+ and Vs- with armored
twisted-pair cables.
To ensure the stability of the system, use armored twisted-pair cables be-
tween the remote sense terminals and the DUT. Pay attention to the posi-
tive and negative poles when wiring, otherwise it will damage the
instrument.
3. Remove the output terminals cover of the power supply.
4. Loosen the screws of the output terminals and connect the red and black
test cables to the output terminals. Re-tighten the screws.
When maximum current that one test cable can withstand fails to meet the
rated current, use multiple pieces of red and black test cables. For example,
the maximum current is 1,200A, then 4 pieces of 360A red and black cables
are required.
5. Thread the red and black test cables through the output terminals cover of
the power supply and install the cover.
6. (Optional) According to the actual situation of DUT, connect the grounding
terminal on the rear panel of the instrument to the DUT to ensure the safe
grounding.
For the location information, see 1.5 Rear Panel Introduction.
7. Connect the other end of the remote sense cables to the DUT.
8. Connect the other end of the red and black cables to the DUT. The positive
and negative poles must be properly connected and fastened when wiring.
9. Power on the instrument and turn on the Sense function of the instrument.
For details, see 5.7 Sense Function (Sense).
When you use the remote interface to send SCPI instructions, if you use the
programming commands that involve modifying the instrument settings, such
as modifying the output voltage value, after completing the communication
connection between the instrument and the host computer, and after the
communication settings are completed, you must execute the SYST:REM
command firstly.
The USB interface is located on the rear panel of the instrument. You can con-
nect the instrument to the computer via a cable with a USB interface on both
ends (USB A-type connector on one end and USB B-type connector on the
other).
The user needs to select the USB interface type in the System menu before us-
ing the USB interface. The following types are optional.
• TMC: USB_TMC interface;
• VCP: Virtual serial port. For Win7 system, you can download IT-M3900 VCP
driver from ITECH official website or contact technical support engineer; for
Win10 system, you do not need to install VCP driver.
The operation steps to change the USB interface type in System Menu are as
follows.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
3. Rotate the knob or press the Left/Right key to select USB, and press
[Enter].
4. Rotate the knob or press the Left/Right key to select the communication
method.
If you select VCP, you also need to set the baud rate and other parame-
ters of the virtual serial port, and the baud rate must be the same as the
configuration on the computer side.
When the user connect PC through LAN interface, the following is required to
use the LAN interface. The LAN interface complies with the LXI standard.
Connect Interface
Use the following steps to quickly connect your instrument to your LAN and con-
figure it. Two typical LAN interface systems are described below: private LAN
and site LAN.
• Connect to the private LAN
• When using one crossover cable to connect PC directly, the gateway address of
the instrument should be consistent with that of the PC, and the IP address
should be at the same network segment with the PC’s IP address.
• When the instrument and computer are connected to the router, an independent
IP address must be assigned for the instrument.
The operation steps to view the LAN interface information in the System Menu
are as follows.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Rotate the knob to select I/O and press [Enter].
3. Press the Left/Right key to select LAN and press [Enter].
4. Press the Left/Right key to select Info and press [Enter].
5. Rotate the knob to view the LAN parameters. See the information in 1.8 Sys-
tem Menu Function for details.
IP-Conf
• IP: This value is the Internet Protocol (IP) address of the instrument. An IP
address is required for all IP and TCP/IP communications with the instru-
ment. An IP Address consists of 4 decimal numbers separated by periods.
Each decimal number ranges from 0 through 255 with no leading zeros (for
example, 169.254.2.20).
• Gateway: This value is the IP Address of the default gateway that allows the
instrument to communicate with systems that are not on the local subnet, as
determined by the subnet mask setting. The same numbering notation ap-
plies as for the IP Address. A value of 0.0.0.0 indicates that no default gate-
way is defined.
• DNS1: This field enters the primary address of the server. Contact your LAN
administrator for server details. The same numbering notation applies as for
the IP Address. A value of 0.0.0.0 indicates that no default server is defined.
• DNS2: This field enters the secondary address of the server. Contact your
LAN administrator for server details. The same numbering notation applies
• Socket Port: This value indicates the port number corresponding to the
service.
The instrument has a built-in Web server for monitoring and controlling the in-
strument via a Web browser in PC. To use the Web server, connect the instru-
ment and PC over LAN interface and enter the instrument's IP address into the
address bar at the top of your PC's Web browser, you can access the front pan-
el control functions including the LAN configuration parameters.
• If you want to remotely control the instrument using the built-in web server, you
must enable the web service. See 2.5.2 LAN Interface for details.
• The format of the address entered in the address bar of the browser is http://
192.168.0.100. The specific IP address is subject to the actual instrument
settings.
The homepage of different models are different. The actual shown page shall
be subject to the connected instrument.
You can select different pages by clicking the buttons shown in the navigation
bar on the left side of the window. The detailed descriptions are as follows.
• Home:Web home interface, displays the model and appearance of the
instrument;
• Information: Displays the serial number of the instrument and more system
information as well as LAN configuration parameters;
• Web Control: Enables the Web control to begin controlling the instrument.
This page allows you to monitor and control the instrument;
• Manual: Go to the ITECH official website and view or download the relevant
documents.
• Upload: Performs a system upgrade.
Click CONNECT to connect the PC with the instrument, then click BROWSE
to select the system upgrade installation package (for example, itech_
3900_P.itech), and then click UPLOAD performs the upgrade operation.
After the upgrade is complete, the instrument needs to be restarted.
The Telnet utility (as well as sockets), is another way to communicate with the
instrument without using I/O libraries or drivers. In all cases, you must first es-
tablish a LAN connection from your computer to the instrument as previously
described.
• Before using this function, you need to configure Socket Port, and
the configuration on the instrument side should be consistent with
the configuration on the PC side.
• The instruments allow any combination of up to six simultaneous
socket and telnet connections to be made.
ITECH instruments have SCPI socket services, which can be used to send and
receive SCPI commands, queries, and query responses. All commands must
be terminated with a newline for the message to be parsed. All query responses
will also be terminated with a newline.
Pins Description
H CAN_H
L CAN_L
GND CAN_GND
CAN Configuration
The user needs to configure the CAN interface parameters in the system menu
before using the remote control. The CAN interface parameters are as follows.
Name Description
Baud rate Select the baud rate from the following options: 5k/10k/20k/
40k/50k/80k/100k/125k/200k/250k/400k/500k/600k/800k/
1000k
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
4. Set the baud rate, address and other parameters, press [Enter].
CAN Troubleshooting
GPIB Configuration
Each device on the GPIB (IEEE-488) interface must have a unique whole num-
ber address between 1 and 30. Your computer’s GPIB interface card address
must not conflict with any instrument on the interface bus. This setting is nonvo-
latile; it will not be changed by *RST.
When you purchase the interface accessory and successfully insert it into the
corresponding position on the rear panel of the instrument, the menu item for
changing the GPIB address appears in the System menu. The specific steps
are as follows:
1. Ensure that the instrument's power switch is off, that is, the instrument is in
Power Off state.
2. Insert the separately purchased GPIB interface card into the card slot on the
rear panel of the instrument.
3. Connect the instrument with the computer via the GPIB cable. After the con-
nection is successful, turn on the power switch of the instrument.
4. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
5. Rotate the knob to select I/O and press [Enter].
6. Press the Left/Right key to select BOARD and press [Enter].
7. Press the Left/Right key to select GPIB and press [Enter].
8. Press the numeric keys to set the GPIB address and press [Enter].
The RS-232 interface shares the same communication card (IT-E177) with the
analog function.
When using the RS-232 interface for communication, connect the pin 1, pin 2,
and pin 3 of the IT-E177 to the PC. The pin description is as follows:
Pins Description
3 DGND, ground
RS–232 Configuration
When you purchase the interface accessory and successfully insert it into the
corresponding position on the rear panel of the instrument, the RS–232 menu
item will appear in the System menu. The specific steps are as follows:
1. Ensure that the instrument's power switch is off, that is, the instrument is in
Power Off state.
2. Insert the separately purchased RS–232 interface card into the card slot on
the rear panel of the instrument.
3. Connect the instrument to the computer via an RS–232 cable. After the con-
nection is successful, turn on the power switch of the instrument.
4. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
RS-232 Troubleshooting
• Check that whether the baud rate of the computer and instrument are the
same;
• Make sure the correct cable and adapter are connected. Note that internal
wiring may not be correct even if the cable has a suitable plug;
• The cable must be connected to the correct serial ports (COM1, COM2, etc)
of PC.
EtherCAT Configuration
When you purchase the interface accessory and successfully insert it into the
corresponding position on the rear panel of the instrument, the menu item for
EtherCAT configuration appears in the System menu. The specific steps are as
follows:
1. Ensure that the instrument's power switch is off, that is, the instrument is in
Power Off state.
2. Insert the separately purchased EtherCAT interface card into the card slot
on the rear panel of the instrument.
3. Connect the instrument with the computer via the network cable. After the
connection is successful, turn on the power switch of the instrument.
4. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
5. Rotate the knob to select I/O and press [Enter].
6. Press the Left/Right key to select BOARD and press [Enter].
7. Press the Left/Right key to select ECAT and press [Enter].
8. After making the communication setting, to ensure normal communication,
you need to restart the instrument.
9. Repeat steps 4~7 above to view the EtherCAT information.
Control code
This series of instruments provides XML configuration files and CANopen con-
trol protocol. For detailed information, please contact ITECH official support.
3 Getting Started
♦ Power-on Self-Test
♦ Set Output Value
♦ Use the Front Panel Menu.
♦ On/Off Control
When you turn the POWER switch on for the first time after purchase, the instru-
ment starts with its factory default settings. Each time thereafter, the instrument
starts according to the setting that you selected as outlined in 5.6 Set the
Power-on State (PowerOn).
Precautions
• Be sure to connect the main power socket to the power outlet with
protective grounding. Do not use terminal board without protective
grounding.
• Ensure that the input electrodes are either insulated or covered us-
ing the safety covers provided, so that no accidental contact with
lethal voltages can occur.
The power switch is located in the lower left corner of the front panel. The power
switch is a button, and pressing once indicates ON and pressing again indicates
OFF.
Self-Test Procedures
If an error occurs during the self-test, an error message is displayed. The follow-
ing table lists the error messages you might see.
Exception Handling
If the instrument cannot start normally, please check and take measures by
reference to steps below.
1. Check whether the power cord is correctly connected and confirm whether
the instrument is powered.
2. Check whether the power in On. The power switch is under “|” On status.
3. Check whether the power voltage matches with the supply voltage. Please
refer to 2.3 Connecting the Power Cord to select proper AC input.
4. After multiple units are connected in parallel and powered on, the error re-
lated to the fiber-optic is indicated. Please refer to the following steps for
troubleshooting.
• FIBER MULTI MASTER: Multiple units (with panel) are connected in par-
allel, please confirm whether multiple Masters are set. Select System→-
Parallel menu to reset. There can only be one single unit as the Master,
and the other single units must be set to Slave. After the setting is com-
pleted, power off and restart each single unit.
• FIBER EXT UNLOCK: Multiple units (with panel) are connected in paral-
lel, check whether the cable connection of the fiber outer ring (TX and
RX) is closed. Then confirm that whether the System→Parallel menu is
set. There can only be one single unit as the Master, and the other single
units must be set to Slave. After the setting is completed, power off and
restart each single unit.
5. If the instrument still does not start, contact ITECH technical support
engineer.
After selecting the operation mode, the interface displays the setting value
under the selected operation mode and the cursor blinks to prompt. The user
can use the following methods to set the values.
• Rotate the knob to set the data in the cursor position. Rotate the knob clock-
wise to increase the set value and anticlockwise to decrease the set value.
Once the data in the cursor position increases to ten, the value will add one
to the front position automatically. and once the data in the cursor position
decreases to zero, the value will minus one from the front position automati-
cally. This provides convenience for the user to set. The knob can works with
the left or right keys. Use the left or right keys to move the cursor position.
After entering the menu interface, the knob can also be used to scroll pa-
ges to view menu items.
• The user can set the parameters related to electrical properties in the
Config menu, including the current operation mode, slope, output delay
time, internal resistance.
• The user can set the function related to instrument system in the System
menu, including the key beeper, Sense function, power-on state, trigger
source, communication method, data logger method, digital I/O function set-
ting, parallel function, system information and so on.
• The user can set the parameters related to instrument protection function in
the Protect menu, including OCP /OVP/OPP /UCP/UVP.
• Function menu includes output list, batterycharging test function and so on.
The user press the complex keys to enter into the menu interface. The menu is
divided according to the function item and the corresponding operation is set in
the lowest level menu item. For example, the key beeper is set in System→-
Beep. Set the Beep item to off or on. Please refer to 1.7 Configuration Menu
Function, 1.8 System Menu Function and the corresponding chapters for de-
tailed menu description.
After entering the menu interface, the screen displays the optional menu. Rotate
the knob to view the menu items. When the number in front of the menu item is
blinking, indicates this item is selected currently. Press [Enter] key to enter the
selected menu item and press [Esc] to exit the menu.
You can press the [On/Off] key on the front panel to control the output status of
the power supply. If the [On/Off] key light is on, indicates that the output is
turned on. If the [On/Off] key light is off, indicates that the output is turned off.
When the output of the power supply is on, the operating status flag (CV/CC/
CW) on the VFD will be illuminated.
It is recommended that you turn on the [On/Off] after the power supply is
connected to the DUT. If the power supply has no output after the output is
turned on, check the voltage and current setting value, set the voltage and
current to a non-zero value, and then turn on the output.
The voltage setting range is from 0V to the maximum rated output voltage.
When [V-set] key is pressed, the key light is lit and the output voltage value can
be set. Press numeric keys or rotate the knob to adjust the value in the voltage
setting area indicated by the cursor. This value takes effect when you press
[Enter].
The range of current settings is between the data ranges defined in the specifi-
cations. When [I-set] key is pressed, the key light is lit and the output current
value can be set. Press numeric keys or rotate the knob to adjust the value in
the current setting area indicated by the cursor. This value takes effect when
you press [Enter].
Operation method
2. In the current input area displayed by the cursor, press [+/-], use the numeric
keys or the adjustment knob to input the current value -10A, and press
[Enter] to make this value effective.
For the bi-directional power supply, you can set the direction of the current
(output / input), that is, control the positive or negative current value by press-
ing the [+/-] key.
The range of power settings is between the data ranges defined in the specifica-
tions. When [P-set] key is pressed, the key light is lit and the output power value
can be set. Press numeric keys or rotate the knob to adjust the value in the
power setting area indicated by the cursor. This value takes effect when you
press [Enter].
Operation method
1. Press [P-set] key, the key light is lit and the output power value can be set.
2. In the power input area displayed by the cursor, press [+/-], use the numeric
keys or the adjustment knob to enter the power value -10W, and press
[Enter] to make this value effective.
For the bi-directional power supply, you can set the direction of the power
(output / input), that is, control the positive or negative power value by press-
ing the [+/-] key.
300V -Imax 3V 1%
The heavy solid line illustrates the locus of possible operating points as a func-
tion of output. As shown by the horizontal portion of the line, the output voltage
remains regulated at its programmed setting as long as the load current remains
within the positive or negative current limit setting. A CV status flag indicates
that the output voltage is being regulated and the output current is within its limit
settings.
Note that when the output current reaches either the positive or negative current
limit, the unit no longer operates in constant voltage mode and the output volt-
age is no longer held constant. Instead, the unit will now regulate the output cur-
rent at its current limit setting.
As shown by the vertical portions of the shaded quadrants, the output voltage
may continue to increase in the positive direction or decrease in the negative di-
rection as current is forced into or pulled out of the unit. When the output voltage
exceeds the over-voltage protection setting, the output will shut down.
CC Priority
300V -Imax 3V 1%
The heavy solid line illustrates the locus of possible operating points as a func-
tion of output. As shown by the vertical portion of the line, the output current re-
mains regulated at its programmed setting as long as the output voltage
remains within its limit setting. A CC (constant current) status flag indicates that
the output current is being regulated and the output voltage is within its limit
settings.
Note that when the output voltage reaches the upper limit, the unit no longer op-
erates in constant current mode and the output current is no longer held con-
stant. Instead, the unit will now regulate the output voltage at its voltage limit
setting.
As shown by the horizontal portion of the shaded quadrants, when the unit is
sinking power, the output current may continue to increase in the negative direc-
tion as more current is forced into the unit. This can happen when the instru-
ment is connected to an external device such as a battery, and its output
voltage is higher than the voltage limit setting of the instrument. Once the cur-
rent exceeds the built-in negative over-current limit, the output will shut down. In
such a case, it is important to set the voltage limit properly in order prevent this
protection shutdown.
How to Set
b. Press the [V-set] key on the front panel to set the voltage upper limit
Vh, and press [V-set] again to set the voltage lower limit Vl.
1. Press the composite keys [Shift]+[V-set] (Config) on the front panel to enter
the configuration menu.
5. Rotate the knob to set the current rise and fall time.
The following takes the setting of 10Ω as an example to introduce the actual
test results.
a. Press the [I-set] key on the front panel to set the output current value Is.
Take –5A as an example.
b. Press the [V-set] key on the front panel to set the voltage upper limit Vh=
20V, and press [V-set] again to set the voltage lower limit Vl=0V.
At this time, 10Ω in sink mode (according to the formula I=U/R, sink current
is 80÷10=8A), due to the limitation of Is=–5A in CC priority mode, the actual
sink current is 5A, and the instrument is working In CC mode. If Sink Res is
set to 20Ω, the sink current is 4A, which is within the limit of Is, so the instru-
ment works in CR mode at this time.
Protection Delay
Protection Prompt
When the instrument enters the protection state, the buzzer sounds (if Beep
menu item is set as default state On), the VFD status indicators Prot and Off
are lit, and [On/Off] is turned off.
• The first row shows the meter voltage and the meter current value.
• The left side of the second row shows the specific protection information
(such as overvoltage protection OVP), the right side shows the duration of
the protection (hour: minute: second) and the number of protection informa-
tion in the message queue and the total number of protection information.
Users can turn the page through the knob to view all the protection
information.
After the protection is cleared, you need to press the [On/Off] key on the front
panel or send the OUTPut ON command to the instrument to reopen [On/Off].
Users can enable the OVP function and set the protection limit Level and pro-
tection delay time Delay. When the voltage (i.e., the Meter value) is greater than
this protection limit and the delay time is exceeded, the power supply will enter
the OVP state.
Possible Cause
• The set protection limit Level is lower than the voltage Meter value.
How to Set
1. Press the composite keys [Protect] on the front panel to enter the protection
menu. (1U Model)
2. Press the composite keys [Shift]+[Recall] (Protect) on the front panel to en-
ter the protection menu. (2U Model)
3. Rotate the knob to select 1. OVP (Off) and press [Enter].
4. Press the left/right key or rotate the knob to select On and press [Enter] to
enter the setting interface.
5. Set the protection limit Level and the delay time Delay in sequence, and
press [Enter] to confirm.
The VFD screen returns to the Protect menu and the OVP setting (take
150V, 1S as an example) is displayed as follows:
PROTECT
1.OVP 150V, 1S
Users can enable the OCP function and set the protection limit Level and pro-
tection delay time Delay. When the current (i.e., the Meter value) is greater than
this protection limit and the delay time is exceeded, the power supply will enter
the OCP state.
Possible Cause
How to Set
1. Press the composite keys [Protect] on the front panel to enter the protection
menu. (1U Model)
2. Press the composite keys [Shift]+[Recall] (Protect) on the front panel to en-
ter the protection menu. (2U Model)
The VFD screen returns to the Protect menu and the OCP setting (take 10A,
1S as an example) is displayed as follows:
PROTECT
2.OCP 10A, 1.000S
Possible Cause
How to Set
1. Press the composite keys [Protect] on the front panel to enter the protection
menu. (1U Model)
2. Press the composite keys [Shift]+[Recall] (Protect) on the front panel to en-
ter the protection menu. (2U Model)
5. Set the protection limit Level and the delay time Delay in sequence, and
press [Enter] to confirm.
The VFD screen returns to the Protect menu and the OPP setting (take
0.15KW, 1S as an example) is displayed as follows:
PROTECT
3.OPP 0.15KW, 1.000S
Users can enable the UCP function and set the instrument warm-up time
Warm-up, protection limit Level and protection delay time Delay. When the cur-
rent (i.e., the Meter value) is lower than this protection limit and the warm-up
time, delay time are exceeded, the power supply will enter the UCP state.
Possible Cause
How to Set
1. Press the composite keys [Protect] on the front panel to enter the protection
menu. (1U Model)
2. Press the composite keys [Shift]+[Recall] (Protect) on the front panel to en-
ter the protection menu. (2U Model)
3. Rotate the knob to select 4. UCP (Off) and press [Enter].
4. Press the left/right key or rotate the knob to select On and press [Enter] to
enter the setting interface.
5. Set the warm-up time Warm-up, protection limit Level and delay time Delay
in sequence, and then press [Enter] to confirm.
The VFD screen returns to the Protect menu and the UCP setting (take 10S,
0.1A, 1S as an example) is displayed as follows:
PROTECT
4.UCP 10S, 0.1A, 1.000S
Possible Cause
• The set protection limit Level is greater than the voltage Meter value.
• The external (AC input) inputs a lower voltage.
How to Set
1. Press the composite keys [Protect] on the front panel to enter the protection
menu. (1U Model)
2. Press the composite keys [Shift]+[Recall] (Protect) on the front panel to en-
ter the protection menu. (2U Model)
4. Press the left/right key or rotate the knob to select On and press [Enter] to
enter the setting interface.
5. Set the warm-up time Warm-up, protection limit Level and delay time Delay
in sequence, and then press [Enter] to confirm.
The VFD screen returns to the Protect menu and the UVP setting (take 10S,
1V, 1S as an example) is displayed as follows:
PROTECT
5.UVP 10S, 1V, 1.000S
When internal temperature of instrument is higher than about 85 °C, the instru-
ment is under temperature protection. At this time, the instrument will automati-
cally be OFF and VFD status indicator Prot lights up and the screen prompts
OTP.
Possible Cause
How to Set
The OTP limit does not need to be set, and the internal device of the instrument
automatically detects and determines whether to enter the OTP state.
When you return power to the instrument, verify that the cooling
fan is running. If not, please contact ITECH Technical Support.
Leaving the instrument powered on with an inoperative cooling
fan may result in damage to the instrument.
When the instrument is in Sense Reverse Protection state, you should check
whether the polarities are connected reversely or not firstly. If yes, you can re-
open the output after the polarities connect correctly.
The voltage difference between output terminal and remote sense terminal of
each model is not the same. When the remote sense terminal is connected re-
versely, the maximum voltage will not exceed the sum of output terminal voltage
and the difference voltage.
The descriptions of list function menu are listed in the table below.
Run Indicates to enter the List run mode, waiting for the trigger to
run the currently selected List file.
USB Open the List file in the external USB flash disk.
File The name of the exported list file. The default suf-
Name fix is .csv, and the file name can only be a string
of numbers.
End Set the running state after the list program is run-
State ning over.
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
5. Set the total number of steps in the List program and press [Enter] to
confirm.
6. Set the current, slope and time width of step 1 in turn, and press [Enter] to
confirm.
7. Set the parameters of step 2 in the same method.
8. Set the number of list repetitions Repeat, and press [Enter] to confirm.
9. Set the End State as Normal, and press [Enter] to confirm.
10. Depending on the test requirement, choose whether to turn on the function
switch that triggers synchronization.
If you choose not to save, press [Esc] to exit the editing interface.
If you need to run the List program on the external USB flash drive, you need to
import the List program from the USB flash drive into the instrument.
The format of the List file in the USB flash drive must be in .csv format and
saved in the root directory of the USB flash drive. You can refer to Export List
Program to customize editing according to the List file template exported to
the USB flash drive.
The procedures are as follows:
1. Insert the USB flash drive into the front panel USB connector.
2. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
Rotate the knob, and the system will automatically read all .csv format files
in the root directory of the USB flash drive at this time. The interface is dis-
played as follows:
XXX.csv
Not-Load Load YY/ZZ
XXX indicates the name of the List file; YY indicates the serial number of the
current List file; ZZ indicates the total number of the List file.
7. Press the Left/Right key to select Load, press [Enter] to complete the im-
port, and wait for the trigger to run this List file.
At this point, the interface returns to the main interface of the system and
LIST RUN is displayed in the lower right corner.
You can select a List file saved inside the instrument to be in the Open state
and wait for subsequent triggers to run. The procedures are as follows:
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
2. Select 1. LIST: Off and press [Enter].
The interface returns to the main interface of the LIST function, which is dis-
played as follows:
FUNCTION LIST
Run Edit Open Export
If you select Run and press [Enter], the instrument will enter LIST mode
and wait for the trigger to run.
Supports exporting the internal List file to an external USB flash drive. The ex-
ported List file is saved in .csv format.
1. Insert the USB flash drive into the front panel USB connector.
2. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
6. Set the name of the List file to be recalled (that is, the file name set in Edit),
and press [Enter].
The interface returns to the main interface of the LIST function, which is dis-
played as follows:
FUNCTION LIST
Run Edit Open Export
Indicates that the selected List file in Open is exported to the USB flash
drive.
You can select a List file to execute according to your needs, so that the power
supply outputs the corresponding waveform sequence. Take the List file in
Internal as an example. The steps are as follows:
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
2. Select 1. LIST: Off and press [Enter].
3. Press the Left/Right key to select Open and press [Enter].
4. Press the Left/Right key to select Internal and press [Enter].
5. Set the name of the List file to be recalled (that is, the file name set in Edit),
and press [Enter].
The interface returns to the main interface of the LIST function, which is dis-
played as follows:
FUNCTION LIST
Run Edit Open Export
6. Press the Left/Right key to select Run and press [Enter].
At this point, the interface returns to the main interface of the system and
LIST RUN is displayed in the lower right corner.
7. Turn on the [On/Off].
8. Based on the selected trigger method, perform the trigger operation.
Take the manual trigger as an example. Press [Shift]+[On/Off](Trigger) on
the front panel to run the selected List file. For details about the List trigger
method, see 5.8 Select Trigger Source (Trig Source).
• The system will adjust the working mode of the power supply according to
the CC or CV priority mode set in the List file. For example, if the current
power supply is in the CV priority mode, and the list file to be run is set to the
CC priority mode, then the power supply will work in the CC priority mode
after the List file is triggered to run.
• After the List file finishes running, the system determines whether to return
to the working mode before the List is run according to the Normal, Last or
Off option you selected.
When the List program is running, if you need to stop running, you can do the
following: Press the composite keys [Shift]+[I-set] (Function) on the front panel
to enter the function menu.
At this point, the interface will prompt whether to stop the running of the Func-
tion, you can press the Left/Right key to select Stop, then the running will be
stopped, and the Function menu item will be displayed, you can re-enter the
Function menu for editing; if Reset is selected, it means to stop the present run-
ning and return to the main interface to wait for the next trigger operation; if
Pause is selected, it means that the present running is suspended, and you can
continue to run by selecting Resume.
This series of instruments comes standard with ARB (arbitrary waveform) func-
tion, which allows the output of user-defined arbitrary complex voltage or current
waveforms. It can be operated in the following ways:
• Import via U disk
The instrument supports importing the edited .csv format file (the template
can be downloaded from ITECH official website or contact ITECH to obtain)
through the USB interface on the front panel, and generating voltage or cur-
rent values after parsing, and then outputting complex voltage or current
waveforms according to the time width specified by the user. The user can
edit a .csv file containing up to 10 million voltage or current points and import
it into the instrument for running to realize arbitrary waveform output or work-
ing condition simulation. The instrument has the ability to quickly parse data.
A .csv file of dozens of megabytes can be imported and parsed within 5S,
making testing more convenient and efficient.
This instrument supports importing the .csv files corresponding to the follow-
ing waveforms:
1. Edit the .csv file corresponding to the template on the PC and save it.
Rotate the knob, and the system will automatically read all .csv format
files in the root directory of the USB flash drive at this time. The interface
is displayed as follows:
XXX.csv
Not-Load Load YY/ZZ
XXX indicates the name of the List file; YY indicates the serial number of
the current List file; ZZ indicates the total number of the List file.
11. Based on the selected trigger method, perform the trigger operation.
• Programming via SCPI instructions
For detailed instructions and parameter introduction, please refer to the in-
structions of "ARB Subsystem" in the Programming Guide.
• Realized by the software IT9000 on PC
This series power supply provides the battery charging/discharging test function
based on its unique bidirectional power supply properties. Suitable for charging/
discharging tests on all types of portable batteries.
according to the
selected test
mode.
This instrument supports battery discharge test in constant current and con-
stant power modes. If the user wants to discharge in constant current mode,
enter the Config menu, select the CC priority, set Vl to 0, set Vh to the rated
voltage value of the instrument, set P- to the rated power value of the instru-
ment, and then set the Battery menu according to the test requirements. If
the user wants to discharge in constant power mode, set P- to the specified
value, and set Discharge V and Discharge I to the rated voltage and current
value of the instrument.
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
After the parameter setting is completed, the interface returns to the Battery
function main interface and displays as follows:
FUNCTION BATTERY
Run Edit
6. Press the Left/Right key to select Run and press [Enter] to run the battery
test.
The trigger method of the battery test is consistent with the trigger method
of the List function, that is, the List trigger source is also effective for the
Battery function.
If you want to stop running during the battery test, you can do the following:
Press the composite keys [Shift]+[I-set] (Function) on the front panel to enter
the function menu.
At this point, the interface will prompt whether to stop the running of the Func-
tion, you can press the Left/Right key to select Stop, then the running will be
stopped, and the Function menu item will be displayed, you can re-enter the
Function menu for editing; if Reset is selected, it means that it will not stop run-
ning, it will return to the main interface, and run Battery again.
This series power supply supports built-in waveforms for user to execute the
test directly. The protocols/standards involved in the built-in waveforms include
the following:
• DIN40839
• ISO16750-2
• ISO21848
• SAEJ1113–11
• LV123
• LV124
• LV148
• ISO21780
Select any of the above five kinds of vehicle waveforms, after completing the
waveform parameter setting (see the subsequent introduction for each parame-
ter), the system will automatically enter the waveform recalling mode and wait
for the triggering operation. At this time, the VFD screen is displayed as the
main interface of the system, and the corresponding waveform name is dis-
played in the lower right corner.
After turning on [On/Off], the output of the vehicle waveform is triggered accord-
ing to the selected trigger method.
The trigger method of the vehicle waveform is consistent with the trigger
method of the List function, that is, the List trigger source is also effective for
the vehicle waveform.
If you want to stop during the vehicle waveform output, you can do the following:
Press the composite keys [Shift]+[I-set] (Function) on the front panel to enter
the function menu.
At this point, the interface will prompt whether to stop the running of the Func-
tion, you can press the Left/Right key to select Stop, then the running will be
stopped, and the Function menu item will be displayed, you can re-enter the
Function menu for editing; if Reset is selected, it means to stop the present run-
ning and return to the main interface to wait for the next trigger operation; if
Pause is selected, it means that the present running is suspended, and you can
continue to run by selecting Resume.
This series power supply has built-in 12V or 24V DIN40839 waveform. This test
verifies the behavior of a DUT during and after cranking. This waveform can re-
produce the voltage curve for automotive power network confirms to DIN40839
standard, thus facilitating quick call by customers.
For automotive startup voltage waveform, the startup voltage can also be set
based on customers’ requirements. In this way, the user can create waveform
between 8V to 32V.
User-de-
V=8.00V fined volt-
age value
1 4.5 60 15 5
2 6 60 2000 5
3 12 60 T 10
1 8V 60 50 10
2 12V 60 2000 5
3 24V 60 T 10
The user can define the startup voltage, ranging from 8V to 32V. When the
waveform program is divided into 8V-16V, the waveform is consistent with
standard 12V; when the waveform program is divided into 16V-32V, the wave-
form is consistent with the standard 24V waveform. The waveform diagram is
shown below.
How to Use
Recall the self-defined DIN waveform operation (taking 12.5V voltage waveform
as an example):
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
2. Use the knob to select 3.Road-Vehicles = Off and press [Enter].
The interface returns to the main interface of the system and displays
DIN40839.
7. According to the selected trigger method (same as the trigger method of the
LIST function), for example, press [Shift]+[On/Off](Trigger) to trigger the
output of the waveform.
Td Pulse width
Un Peak voltage
Td Pulse width
Un Peak voltage
Td Pulse width
Un Peak voltage
Td Pulse width
Un Peak voltage
Us Clamping voltage
This waveform simulates an instantaneous drop in the supply voltage when the
car is started. This test simulates the effect when a conventional fuse element
melts in another circuit.
• 12V system
• 24V system
How to recall this waveform from menu (take 12V system as an example):
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
4. Press the Left/Right key to select Short-Drop and press [Enter] to confirm.
5. Press the Left/Right key to select 12V and press [Enter] to confirm.
The interface returns to the main interface of the system and displays
SHORT-DROP.
6. Press [On/Off] on the front panel to turn on the output.
7. According to the selected trigger method (same as the trigger method of the
LIST function), for example, press [Shift]+[On/Off](Trigger) to trigger the
output of the waveform.
Reset-Test
This test verifies the reset behavior of the DUT at different voltage drops. This
test is applicable to equipment with reset function, e.g. equipment containing mi-
crocontroller. Apply the test pulse simultaneously in figure below to all relevant
inputs (connections) and check the reset behavior of the DUT. Decrease the
supply voltage by 5 % from the minimum supply voltage, Us min, to 0.95 Us
min. Hold this voltage for 5 s. Raise the voltage to Us min. Hold Us min for at
least 10 s and perform a functional test. Then decrease the voltage to 0.95 Us
min. Continue with steps of 5 % of Us min, as shown in figure below, until the
lower value has reached 0 V. Then raise the voltage to Us min again.
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
2. Use the knob to select 3.Road-Vehicles = Off and press [Enter].
3. Press the Left/Right key to select ISO16750-2 and press [Enter] to confirm.
4. Press the Left/Right key to select Reset-Test, and press [Enter] to confirm.
5. Press the numeric keys to set the Usmin, and press [Enter] to confirm.
The interface returns to the main interface of the system and displays
RESET-TEST.
7. According to the selected trigger method (same as the trigger method of the
LIST function), for example, press [Shift]+[On/Off](Trigger) to trigger the
output of the waveform.
Starting Waveform
How to recall this waveform from menu (take 12V system as an example):
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
2. Use the knob to select 3.Road-Vehicles = Off and press [Enter].
3. Press the Left/Right key to select ISO16750-2 and press [Enter] to confirm.
4. Press the Left/Right key to select Starting-Profile, and press [Enter] to
confirm.
5. Press the Left/Right key to select 12V, and press [Enter] to confirm.
6. Press the Left/Right key to select the Level (i.e. 4), and press [Enter] to
confirm.
The interface returns to the main interface of the system and displays
STARTING-PROFILE.
This test is a simulation of load dump transient occurring in the event of a dis-
charged battery being disconnected while the alternator is generating charging
current with other loads remaining on the alternator circuit at this moment.
• The pulse duration of load dump is mainly determined by the time constant
and pulse amplitude of the excitation circuit.
Inside most novel alternator, the amplitude of load dump is decreased by in-
creasing the limiter diode (clamping diode). The load dump may be caused by
cable corrosion, poor cable contact or disconnecting the battery intentionly
when the engine is running.
The pulse shape and parameters for an alternator without centralized load
dump suppression (Test A) are given in follow.
• t: Time
• U: Test voltage
• UA: The supply voltage of the generator in operation: UA = 14V in the 12V
system, UA = 28V in the 24V system. (see ISO 16750-1)
tr(ms) – –
aIf not otherwise agreed, use the upper voltage level with the upper value for
internal resistance or use the lower voltage level with the lower value for in-
ternal resistance.
The pulse shape and parameters for an alternator with centralized load dump
suppression (Test B) are given in follow.
• t: Time
• U: Test voltage
• td: Duration of pulse
• UA: The supply voltage of the generator in operation: UA=14V in the 12V sys-
tem, UA=28V in the 24V system. (see ISO 16750-1)
• US: Peak voltage
• US*: Supply voltage with load dump suppression (i.e. clamping voltage)
tr(ms) – –
aIf not otherwise agreed, use the upper voltage level with the upper value for in-
ternal resistance or use the lower voltage level with the lower value for internal
resistance.
– Nact: The actual alternator speed, unit: round per minute (r/min)
• The pulse is determined by the peak voltage Un, the clamping voltage US,
the internal resistance Ri, and the pulse duration td; in all cases small values
of Un are correlated with small values of Ri and td, and high values of Un with
high values of Ri and td. For the test voltage UA please refer to ISO16750-1.
How to recall this waveform from menu (take Test A 12V as an example):
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
2. Use the knob to select 3.Road-Vehicles = Off and press [Enter].
3. Press the Left/Right key to select ISO16750-2 and press [Enter] to confirm.
4. Press the Left/Right key to select Load-Dump, and press [Enter] to confirm.
5. Press the Left/Right key to select Test A, and press [Enter] to confirm.
6. Press the Left/Right key to select 12V, and press [Enter] to confirm.
The interface returns to the main interface of the system and displays
LOAD-DUMP.
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
3. Press the Left/Right key to select ISO21848, and press [Enter] to confirm.
4. Press the Left/Right key to select Momentary-Drop, and press [Enter] to
confirm.
The interface returns to the main interface of the system and displays
MOMENTARY-DROP.
Detect the function when the DUT is under maximum dynamic Voltage Umax,dyn,
and simulate the maximum dynamic Voltage of high-energy pulse raised from
throw load in 42V electrical system, where the upper limit is the protection volt-
age of throw load.
After turning on [On/Off] and triggering the waveform output, the instrument will
apply a test pulse to the DUT, as shown below:
Simulate the affect from short circuit when fuse element of another circuit is
melt. Detect the function status of the DUT at transient drop of Voltage.
After turning on [On/Off] and triggering the waveform output, when the given
test pulse is applied at all input terminals of the DUT, the rise and fall time be-
tween Ulow and 16V level shall not be longer than 100ms.
• U: Voltage (in V)
Detect the reset performance of the DUT at different Voltage drops. Applicable
for devices with reset function (for example, device installed with one or several
micro controllers) .
After turning on [On/Off] and triggering the waveform output, the instrument will
apply test pulse and detect the reset performance of the DUT, as shown in the
figure below.
The supply voltage drops from Ulow to 0.95Ulow by 5% and keeps for 5s, and
then rise to Ulow and keep for at least 10s for function test. Then, drop Voltage to
0.9Ulow, and so on. As shown in the figure below, drop the voltage from Ulow to
0V by 5% and raise the voltage to Ulow. The Rise and Fall time shall be between
10ms and 1s.
• t: Time (in s)
• Y: Ulow, %
Start-up Characteristics
After turning on [On/Off] and triggering the waveform output, the instrument will
apply the startup characteristic parameters as shown in the figure and table be-
low to related input terminals of the DUT simultaneously.
SAEJ1113-
SAEJ1113-11 Waveform protocol
11
Test–2B Transient from DC motors acting as genera-
tors after ignition switch OFF
T9
T11
T9
T11
Td Test pulse
width
Un Peak voltage
Us Clamping
voltage
Td Test pulse
width
Un Peak voltage
Us Clamping
voltage
Td Test pulse
width
Un Peak voltage
Us Clamping
voltage
Td Test pulse
width
Un Peak voltage
Us Clamping
voltage
Test–2B
VS 10V 20V
Ri ≤0.05Ω ≤0.05Ω
td 0.2–2s 0.2–2s
t1 1ms±50% 1ms±50%
tr 1ms±50% 1ms±50%
t6 1ms±50% 1ms±50%
Test–4
Ri 0Ω to 0.02Ω 0Ω to 0.02Ω
t7 15 to 40ms(1) 50 to 100ms(1)
t8 ≤50ms ≤50ms
(1). The value used should be agreed between the vehicle manufacturer and
the equipment supplier to suit the proposed application.
(2). t11=5 ms is typical of the case when engine starts at the end of the crank-
ing period, while t11=100 ms is typical of the case when the engine does not
start.
(3). t11=10 ms is typical of the case when engine starts at the end of the crank-
ing period, while t11=100 ms is typical of the case when the engine does not
start.
Test–5
For the details, please refer to the Load Dump Dynamic Behavior.
4.6.3.5 LV123
LV123 has been compiled for the electrical characteristics of new energy ve-
hicles for major European car manufacturers. For its high-voltage system parts,
the state standard within its operating voltage range is defined. The relevant pa-
rameters are introduced as follows:
Please pay attention to whether the voltage value defined in the LV123
standard matches the rated voltage of the instrument, otherwise some
waveforms cannot be recalled.
• HV_2b
• HV_3
• HV_2b
• HV_3
• HV_3
Unlimited
Upper-limited
Lower-limited
Highly-limited
How to Use
The following is an example of how to use this function by recalling the LV123
standard regulation Unlimited and HV_1 waveform.
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
5. Set start voltage = 50V, start time = 0.1S, end time = 0.1S and count = 1.
6. Set Save To File = 1, and press [Enter] to confirm.
7. Select Open→Recall File in the main interface of LV123. Set to 1, and press
[Enter] to confirm.
8. Select RUN in the main interface of LV123, and press [Enter] to confirm.
At this time, the information of the recalled waveform is displayed in the low-
er right corner of the VFD screen: R1/HV_1/00001, which is represented as
the interval type of the waveform, the type of the regulation, and how many
times the waveform has been executed cyclically.
10. According to the selected trigger method (same as the trigger method of the
LIST function), for example, press [Shift]+[On/Off](Trigger) to trigger the
output of the waveform.
4.6.3.6 LV124
The built-in curves LV124 can meet general requirements, test conditions and
tests of electrical and electronic components in motor vehicles up to 3.5 t. Re-
lated parameters are as below:
Transient overvoltages may occur in the electric system due to the switching off
of loads and due to short accelerator tip-ins. These overvoltages are simulated
by means of this test. This test may be used for the electrical life test. The test
pulse of E-02 Transient overvoltage is shown in the figure below:
External starting of the vehicle is simulated. The maximum test voltage results
from commercial vehicle systems and their increased power supply voltage.
The test pulse of E-04 Jump start is shown in the figure below:
The slow decrease and increase of the supply voltage is simulated as it occurs
during the slow decharging and charging procedure of the vehicle battery. The
waveform is as follows.
This test simulates the slow decrease of the battery voltage to 0 V and the sud-
den reconnection of the battery voltage e.g. by means of applying a jump start
source. The waveform is as follows.
When starting the engine, the battery voltage drops to a low value for a short pe-
riod and then slightly rises again. Most components are activated directly before
starting for a short period, then deactivated during starting and activated again
after starting when the engine is running. This test serves to verify normal oper-
ation under these conditions. The starting process may be performed under dif-
ferent vehicle starting conditions, cold start and warm start. In order to cover
both cases, two different test sequences are required. A component must al-
ways undergo both sequences.
The behavior of the electric system when intelligent generator controls are used
is simulated.
4.6.3.7 LV148
LV148 covers tests for electric and electronic components in motor vehicles
48V electrical system. The voltage level and definition are shown in the figure
below.
The related parameters of the built-in waveform of the instrument are introduced
as follows:
LV148 LV148
E-01
E-02
E-03
E-04
E-06A
E-06B
E-07
E-08
E-10
E-15
E-16
E-18
How to use
1. Press the composite keys [Shift]+[I-set] (Function) on the front panel to en-
ter the function menu.
3. Use the knob or Left / Right key to select LV148, and press [Enter].
4. Use the knob or Left / Right key to select E-01, and press [Enter].
At this time, the lower right corner of the VFD screen displays the information
of the recalled waveform: LV148–E01.
5. Press [On/Off] on the front panel to turn on the output.
6. According to the selected trigger method (same as the trigger method of the
LIST function), for example, press [Shift]+[On/Off](Trigger) to trigger the
output of the waveform.
4.6.3.8 ISO21780
This topic covers requirements and tests for the electric and electronic
components in road vehicles equipped with an electrical system operating at a
nominal voltage of 48 V DC.
TEST-10 reinitialisation
TEST-01
The purpose of this test is to verify the component functionality in the nominal
voltage range.
TEST-02
The purpose of this test is to verify the functionality of the component in the
upper and lower transitory voltage range.
TEST-03
TEST-04
This test is a simulation of a load dump situation where voltage transients occur
following a sudden reduction in the load current drawn from a generator, motor-
generator or DC/DC converter and whilst the battery is either out of circuit or
incapable of absorbing power.
TEST-05
TEST-06
TEST-07
This test is applicable to all components which not only consume electrical
energy but may also supply electrical energy with no possibility of switching off
this electrical energy supply without negative effects occurring in the vehicle.
This test does not apply to components that supply electrical energy as a
primary function, such as generators. For these components TEST-04 is
applicable.
The purpose of this test is to verify that the component alone complies with the
voltage range specified here and is applicable where systematic
countermeasures at the vehicle level are not used to ensure compliance.
The test emulates a condition where such a component supplies energy into the
vehicle 48 V power supply system which can only be absorbed by the power
network simulation and subsequently leads to an undesired increase in system
voltage.
TEST-08
TEST-10
This test is intended to check the correct reinitialisation of the component after
interruption of onboard power supply. This test shall only be performed on
DUT's whose control logic (or parts thereof) is supplied by 48 V.
TEST-11
How to Use
3. Use the knob or Left / Right key to select ISO21780, and press [Enter].
4. Use the knob or Left / Right key to select TEST-01, and press [Enter].
At this time, the lower right corner of the VFD screen displays the
information of the recalled waveform: ISO21780–T01.
• SCMC: Stand-
ard Crystalline
or Multi-
crystalline
• HEC: High-effi-
ciency
Crystalline
E- • TF: Thin-Film
N50530
• cSi
You can edit the static PV curve and save it inside the instrument (up to 100
curve records can be saved), or you can customize the editing of the PV curve
(up to 100 can be saved).
• Curve
The steps to edit a static curve locally on the instrument are as follows:
– Also you can set the filter speed in SAS→Filter according to the ac-
tual test needs.
– After the above Curve parameters are set completely, you can press
[Esc] until you return to the SAS main interface and select Run to
trigger the current edited Curve file to run.
• User-defined
6. Set the Voc, Imp and the other parameters, and press [Enter].
7. Set the saved file name and press [Enter].
You can select the Static Curve file saved in the instrument to be in the Open
state for later running.
• Curve
• Curve
1. Refer to Select the Static PV Curve to select the Curve file to be run.
2. Set the Vmax (in Config menu) and Filter as the actual test needs.
3. Press the Left/Right key on the Curve interface to select Run and press
[Enter].
At this point, the instrument enters the SAS mode, and the VFD screen is
displayed as the main interface of the system, waiting to trigger the Curve
file to run.
The trigger method of the SAS function is consistent with the trigger
method of the List function, that is, the List trigger source is also effec-
tive for the SAS function.
• User-defined
2. Set the Vmax (in Config menu) and Filter as the actual test needs.
3. Press the Left/Right key on the USER-DEF interface to select Run and
press [Enter].
At this point, the instrument enters the SAS mode, and the VFD screen is
displayed as the main interface of the system, waiting to trigger the
USER-DEF file to run.
The trigger method of the SAS function is consistent with the trigger
method of the List function, that is, the List trigger source is also effec-
tive for the SAS function.
If you want to stop running during the SAS test, you can do the following: Press
the composite keys [Shift]+[I-set] (Function) on the front panel to enter the
function menu.
At this point, the interface will prompt whether to stop the running of the Func-
tion, you can press the Left/Right key to select Stop, then the running will be
stopped, and the Function menu item will be displayed, you can re-enter the
Function menu for editing; if Reset is selected, it means that it will not stop run-
ning, it will return to the main interface, and run SAS again.
This series power supply can simulate battery characteristics in practical appli-
cations based on its unique bidirectional properties and the variable output im-
pedance. You can set battery-related parameters to simulate the charge and
discharge characteristics of the battery to assist with other tests. The menu
items for this function are described below:
Battery
Battery simulation function
Emulator
1. You can contact ITECH Technical Support to obtain a template for the
battery simulation curve file (.csv format).
2. Use the Excel tool to open the template file, edit the relevant parameters,
and save.
1. Save the edited battery simulation curve file to the root directory of the
USB flash disc.
2. Insert the USB flash disc into the USB interface on the front panel of the
instrument.
3. Press the composite keys [Shift]+[I-set] (Function) on the front panel to
enter the function menu.
4. Use knob to select 5.Battery Emulator: Off and press [Enter].
The interface is displayed as follows:
FUNCTION BEMULATOR
User-define Curve
5. Enter the Curve→USB, and press [Enter].
Rotate the knob, and the system will automatically read all .csv format
files in the root directory of the USB flash drive at this time. The interface
is displayed as follows:
XXX.csv
Not-Load Load YY/ZZ
XXX indicates the name of the Battery simulation file; YY indicates the
serial number of the current Battery simulation file; ZZ indicates the total
number of the Battery simulation file.
6. Rotate the knob to select the Battery simulation file.
7. Press the Left/Right key to select Load, and press [Enter].
8. Press [Esc] to return. Press the Left/Right key to select Save, and press
[Enter] to save the Battery simulation file imported from the current USB
flash disc Import to the machine.
• FTP Import
After the battery simulation file is edited, you can select a file to run directly.
After executing Run, you also need to set the initial state of charge (SOC),
which is the ratio of the present battery's capacity divided by the full voltage
capacity.
• Run a user-defined battery simulation file
The instrument will enter the battery simulation test mode and wait for
the function to be triggered.
6. Turn on the [On/Off].
The trigger method of the battery simulation is consistent with the trig-
ger method of the List function, that is, the List trigger source is also
effective for the Battery Emulator.
3. Enter the Curve→RECALL, and set the Recall File to select the file to
run, press [Enter].
The instrument will enter the battery simulation test mode and wait for
the function to be triggered.
6. Turn on the [On/Off].
The trigger method of the battery simulation is consistent with the trig-
ger method of the List function, that is, the List trigger source is also
effective for the Battery Emulator.
After the battery simulation is triggered, the VFD screen displays as follows:
12.000V 2.000A
12.0Ah 12.5%SOC
• The first line shows the voltage and current Meter values in real time.
• The second line shows the present battery capacity and the actual state of
charge (SOC).
As the battery simulation test runs, the SOC value is constantly changing:
• When the SOC rises (simulated charge), the voltage will also rise. When the
voltage reaches full voltage (Full Voltage), the voltage will continue to rise.
If you want to stop testing when the voltage reaches full voltage, turn on the
OVP function and set the OVP level to the full voltage value. See 4.5.1 Set
Over-Voltage Protection (OVP) for details.
• When the SOC drops (simulated discharge), the voltage will also drop.
When the voltage reaches empty voltage (Empty Voltage), the voltage will
continue to drop.
If you want to stop testing when the voltage reaches empty voltage, turn on
the UVP function and set the UVP level to the empty voltage value. See
4.5.5 Set Under-Voltage Protection (UVP) for details.
If you want to stop running during the battery simulation test, you can do the fol-
lowing: Press the composite keys [Shift]+[I-set] (Function) on the front panel to
enter the function menu.
At this point, the interface will prompt whether to stop the running of the Func-
tion, you can press the Left/Right key to select Stop, then the running will be
stopped, and the Function menu item will be displayed, you can re-enter the
Function menu for editing; if Reset is selected, it means that it will not stop run-
ning, it will return to the main interface, and run battery simulation test again.
5 Basic Operation
This chapter describes the basic functions and features of the instrument. Div-
ided into the following sections:
• Local operation mode: Use the front panel keys to control the power supply.
• Remote operation mode: The power supply is connected to PC, and the user
uses the software in PC to control the power supply remotely.
– During remote control operation, the remote symbol “Rmt” is displayed.
All panel keys, except the [On/Off] and [Shift]+[3] (Local) keys, are
locked.
– You can press [Shift]+[3] (Local) to switch the remote control to local
control. The mode modification will not affect the output parameters of
the power supply.
except the [On/Off] and [Shift]+[2] (Lock) keys, are locked. You can repress
[Shift]+[2] (Lock) to unlock the front panel.
Category Parameter
Current setting: Is
2. Press the composite keys [Shift]+[+/-] (Save) to enter the parameter save
interface. (2U Model)
3. Set the storage location.
Enter a number in the "Save data to bank=1" to set the storage location in
non-volatile memory.
4. Press [Enter] to save the parameters.
Enter a number in the “Recall data from bank=1” to set the storage location
in non-volatile memory.
The user can select the following data sources for recording:
• Voltage
Records current and voltage data during the data acquisition period.
1. Press the composite button [Shift]+[1] (Log) on the front panel to enter the
configuration menu of the data logging function.
See the steps in the 5.8 Select Trigger Source (Trig Source) section to set how
the data logging is triggered.
This setting result is only valid for data recording function. When List function
is used, the method for triggering the running of List files needs to be addi-
tionally set (by default, triggered by the panel).
• Before this operation, make sure to connect the USB memory de-
vice to the memory port on the front panel (The USB interface on
the rear panel can only be used for connecting the PC) so that the
recorded data can be stored in the external storage device. Other-
wise, the Data Recording function is not available for usage.
• Trigger immediately
• Manual
Press the composite button [Shift] + [On/Off] (Trigger) on the front panel to
trigger.
• Bus
Triggered by the SCPI instruction, for example, when the instrument re-
ceives the trigger command *TRG, a trigger operation is performed.
• Voltage trigger
When the DC terminals detect that the voltage reaches the trigger voltage
setting value and is within the range of the upper and lower trigger limits, a
data recording operation is triggered.
When the DC terminals detect that the current reaches the trigger current
setting value and is within the range of the upper and lower trigger limits, a
data recording operation is triggered.
Connect to pin 4 of the digital I/O interface (P-IO) and set pin 4 to Ext-Trig→-
Trig-In→Dlog to trigger.
After the data recording function is started, the recorded data is saved in USB
memory device in .csv format. The user can get access to these files for analy-
sis based on needs.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
The first displayed menu item Beep is used to set the beeper status.
• The voltage, current, power setting values displayed on the main interface
when the instrument is powered on.
• Parameter setting values in the Config menu.
• The output state of the power supply, i.e., the [On/Off] key state.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Turn the knob to select the PowerOn and press [Enter].
3. Press the Left / Right key or turn the knob to adjust the value of this
parameter.
• Reset: Default value, indicates when the instrument is powered on, the
instrument will initialize some parameter settings or state.
The affected parameters and the reset information are as shown in the
following table.
Config Mode CV
menu
Speed High
On Delay/Off Delay 0
• Last: Indicates when powered on, the instrument will remain the same
parameter settings and output status as last time you powered off the
instrument.
• Last+Off: Indicates when powered on, the instrument will remain the
same settings as last time you powered off the instrument, but the output
status is Off.
For example, Last is selected, and the voltage value is set to 8V. After the
instrument is powered off and then powered on, the voltage displayed on the
interface is 8V.
This series power supply supports two connection methods: Local measure-
ment and Remote sensing. The remote sensing is used for maximizing meas-
urement accuracy. (Refer to 2.4 Connecting the Device Under Test (DUT))
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Turn the knob to select the Sense and press [Enter].
3. Press the Left / Right key or turn the knob to adjust the value of this
parameter.
• Bus: Indicates the trigger occurs via the communication cables. When the
trigger command *TRG is received, the instrument generates a trigger.
• Voltage (only available in DLogTrig Source): Voltage trigger.
When the DC terminals detect that the voltage reaches the trigger voltage
setting value and is within the range of the upper and lower trigger limits, a
data recording operation is triggered.
– Level: Trigger threshold.
When the DC terminals detect that the current reaches the trigger current
setting value and is within the range of the upper and lower trigger limits, a
data recording operation is triggered.
– Level: Trigger threshold.
The List function and the data logging function have the same steps for select-
ing the trigger source. They need to be set separately in different menu items
(ListTrig Source and DLogTrig Source). Therefore, this section takes the List
trigger source setting as an example to introduce the operation steps.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Turn the knob to select the ListTrig Source and press [Enter].
• When setting the trigger source for data logging, select the
DLogTrig Source menu item.
• The List trigger source set here is also valid for triggering other functions in
the Function menu.
3. Press the Left / Right key or turn the knob to adjust the value of this
parameter.
4. After the parameter settings are complete, press [Enter].
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Turn the knob to select the I/O and press [Enter].
3. Press the Left / Right key or turn the knob to adjust the value of this
parameter.
• The instrument supports the non-standard RS-232, GPIB and EtherCAT in-
terfaces, and the configuration items in the menu are dynamically displayed
according to the interface user selected.
The maximum power of this series power supply is 12 kW. The user can con-
nect power supply in parallel to provide a greater current and power capability.
The following figure shows three units connected in parallel, the fiber optical is
used as connection between master and slaves.
• This series 10V of 2U models can support up to 8 units in parallel, other models
can support up to 16 units in parallel.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
3. Press the Left / Right key or turn the knob to adjust the value of this
parameter.
• Single: Default value, indicates that the instrument is in single mode.
• Master: Indicates that the current single unit is set to master in parallel
mode. When set to Master, you also need to set the number of slaves to
be mounted for the master, so you need to set the value of the Total pa-
rameter. For example, setting Total = 3 means that the total number of
units in the parallel relationship is 3.
In the parallel mode, you only need to operate the master. Synchro-
nous operations on other slaves are implemented by operating the
master.
• Slave: Indicates that the instrument is set to the slave in parallel mode.
• Before connecting the fiber optical, you must ensure that each in-
strument is in single mode (Single).
• Fiber optic cables cannot be flexed or folded. When the cable is too
long and needs to be arranged, gently wrap the cable in a circle and
gently tie it.
The operation steps of parallel function are described below taking 3 instru-
ments (1 Master and 2 Slave) in parallel as an example.
1. Ensure that the power switches of the three units and the main switch of the
AC power distribution box are off.
2. Refer Figure 5–1 Wiring connection diagram to connect three units.
a. Connect the AC input terminals of the three units separately, and connect
them to the AC distribution box.
b. Connect the DC output terminals of the three units in parallel and connect
them to the DUT.
c. Refer to the blue wiring legend in the figure, connect the fiber optical (i.e.,
the fiber outer ring interfaces TX and RX) for fiber-optic communication
between the master and slaves.
3. Turn on the main switch of the AC distribution box and power on each of the
three units.
4. Set three units in parallel mode with one master and two slaves.
a. Press the composite keys [Shift]+[P-set] (System) on the front panel to
enter the system menu.
b. Set Parallel to Master or Slave and press [Enter].
c. After setting the Master, you need to set Total to 3.
For details, see Configure the Menu Item.
5. After the parallel menu of the three units are set, restart the instrument
separately.
After the instrument is restarted, the VFD shows that the instrument is work-
ing in parallel mode.
Pins Introduction
GND Ground terminal, that is, the negative terminal Level signal
corresponding to each of the above 7 pins.
In this chapter, all the pulse signals involved in the digital I/O function are
switched from high level to low level.
• Signal definition
Digital I/O functions involve input and output levels and pulse signals. The in-
put signal is the control signal provided externally to IT-M3900C, the output
signal is the level signal provided externally by IT-M3900C, and the pulse
signal is the edge signal switched between high and low levels.
Typical: 5V
High level signal Range: 1.6V-15V
Current: ≤100mA
Input signal
Typical: 0V
Low level signal Range: -5V-0.8V
Current: ≤100mA
High level signal Voltage level: 5V
Output signal
Low level signal Voltage level: 0V
Level rise slope 10us
Pulse Level fall slope 2us
Width 30us
• Input/Output Function
The IO-1 ~ IO-7 pins are featured default function, the user can setting the
function of pin according to requirement. The Input and Output are the gen-
eral digital I/O function, and the parameter settings and functions of the sev-
en pins are the same.
The IO-1~IO-7 pins provide default functions. Users can realize control ac-
cording to the functions defined. Users can also reset the input or output
properties of the present pin and customize the function use of the pin ac-
cording to their needs.
– When pins 1 to 7 are configured to Output function, either high level
(False) or low level (True) can be Output.
– When pins 1 to 7 are configured to Input function, an external signal can
be Input to this pin, and the instrument can detect the state of the exter-
nal signal.
– Under the default condition (Not-Invert), when the pin (1 to 7) is config-
ured as Output→PWM, the user needs to set the frequency (PWM Freq)
and duty cycle (PWM Duty) values. For example, if the PWM Freq is set
to 100Hz and the PWM Duty is set to 10%, the output waveform is as
follows:
In the above figure, the peak voltage (minimum value) is 5.16V and
the cycle is 10ms. The high level duration is 9ms and the low level du-
ration is 1ms in one cycle.
• Signal Revert
Select Invert or not under the IO Settings menu. If setting to Not-Invert, it
means the default level will be valid. If setting to Invert, it means the valid sig-
nal is reversed. For example, the IO-5 pin is inhibit output by default and the
high level is valid, when select revert Invert, the low level is valid and the in-
strument output is disabled.
Parameter Description
Invert
• Invert: Yes
• Not-Invert: No
How to Use
When pin 1 is set to default Ps-Clear function, pin 1 has bi-directional I/O func-
tion, which can receive pulse signal input from the external instrument and also
can output pulse signal to external instrument. Pulse signal parameter require-
ments are as follows:
• Pulse input: When the instrument is under protection, the instrument will
clear protection after receiving the pulse signal from external input.
4. Build the test environment to enable the instrument to enter the OVP
state.
Parameter Description
Invert
• Invert: Yes
• Not-Invert: No
How to Use
When pin 2 is set to default Ps function, pin 2 will output high or low level based
on whether the instrument is under protection or not. Under normal conditions
(Not under protection), and when pin 2 is under default setting (Not-Invert), pin
2 outputs high level; when the instrument is under protection, pin 2 outputs low
level.
4. Build the test environment to enable the instrument to enter the OVP state.
5. Check the oscilloscope and confirm that pin 2 outputs low level.
Parameter Description
Invert
• Invert: Yes
• Not-Invert: No
How to Use
When pin 3 is set to default Off-Status function, pin 3 will trigger level output
based on the [On/Off] state of the instrument. When it is set to (Not-Invert), the
[On/Off] is turned off, and pin 3 outputs high level; the[On/Off] is turned on, and
pin 3 outputs low level.
4. Check the oscilloscope and confirm that pin 3 outputs low level.
Parameter Description
Invert
• Invert: Yes
• Not-Invert: No
related to AC-
Quire in the Trig-
ger subsystem of
the Instruction
Manual.
How to Use
Taking the triggering of List function as an example, the text below will introduce
how to use pin 4’s default function Ext-Trig.
• Trig–Out
5. Check the oscilloscope and confirm whether pin 4 has following pulse
signal output.
• Trig-In
Parameter Description
Invert
• Invert: Yes
• Not-Invert: No
How to Use
• When pin 5 is set to Inhibit-Living (Not-Invert), pin 5 can control the instru-
ment’s output state based on the level signal from external input.
– Under default conditions (namely, pin 5 is not connected), the input is
high level, and it will not impact the instrument’s output state.
– When the [On/Off] is in On state, input low level to pin 5, and it will im-
pact the output state: The [On/Off] button light is lighted on and VFD still
displays On, but the actual output is 0; when pin 5 receives high level sig-
nal again, the output state is recovered.
If pin 5 is set to Invert, input high level to pin 5, and it will impact the out-
put state.
At this point, the [On/Off] button light is lighted on, and VFD indicator
light on the front panel displays On, while voltage/current Meter value
gradually decreases to 0, the output function is prohibited. In addition,
the VFD screen on the front panel displays INH.
At this point, [On/Off] button light is lighted on, and VFD indicator light on
the front panel displays On, and voltage/current Meter value gradually re-
covers to 10V, and the output function is re-initiated.
• When pin 5 is set to Inhibit-Latch (Not-Invert), pin 5 can control the instru-
ment’s output state based on the pulse signal from external input. The pa-
rameter requirements of this pulse signal are as follows:
– Under default conditions (pin 5 is not connected), it will not effect the in-
strument’s output state.
After confirming that the [On/Off] can be turned on again, the user needs
to manually turn on [On/Off].
At this point, the [On/Off] button light is lighted off, and the VFD indicator
light on the front panel displays Off, and the output function is switched
off. In addition, the VFD screen on the front panel displays Inhibit-Ps.
When the protection state is cleared, manually turn on [On/Off] again.
Parameter Description
Invert
• Invert: Yes
• Not-Invert: No
How to Use
When pin 6 is set to default Sync-On function, pin 6 has bi-directional I/O func-
tion, which can receive pulse signal input from the external instrument and also
can output pulse signal to external instrument. The parameter requirements of
this pulse signal are as follows:
1. Referring to the figure below, connect pin 6 of two instruments to the exter-
nal oscilloscope.
4. Set the voltage to 10V on the front panel of instrument A, and turn on
[On/Off].
At this time, check the oscilloscope. The instrument A’s pin 6 outputs pulse
signal and the instrument B’s output function is synchronously turned on.
Parameter Description
Invert
• Invert: Yes
• Not-Invert: No
How to Use
When pin 7 is set to default Sync-Off function, pin 7 has bi-directional I/O func-
tion, which can receive pulse signal input from the external instrument and also
can output pulse signal to external instrument. The parameter requirements of
this pulse signal are as follows:
• When the instrument’s [On/Off] is in Off state, input pulse signal to pin 7,
and it will not impact the [On/Off] state.
• When the instrument’s [On/Off] is in On state, input pulse signal to pin 7,
and [On/Off] will be changed from On to Off.
Taking two instruments as an example, the operation is as below:
1. Referring to the figure below, connect pin 7 of two instruments to the exter-
nal oscilloscope.
At this time, check the oscilloscope. The instrument A’s pin 7 outputs pulse
signal and the instrument B’s output function is synchronously turned off.
The rear panel of this series of instrument provides an optional interface card in-
tegrating the RS-232 and analog quantity function. The interface card provides
the following capability:
• Remotely control voltage and current values.
• Remotely control voltage and current upper limit values.
• Remotely control voltage and current lower limit values.
• Remote monitoring of output voltage/current measurement.
• RS-232 communication interface, please refer to 2.5.5 RS–232 Interface
(Optional) for detailed description about the interface.
The analog signal bandwidth is less than 100Hz, and the signal bandwidth sup-
ports any waveform. When the programmed signal frequency or amplitude ex-
ceeds the output capacity, the instrument will automatically limit the output
amplitude. The pins description is as below.
When the analog quantity function is optionally equipped, the system menu dis-
plays the analog quantity menu. The analog quantity menu items and parame-
ters are described below:
When the analog quantity control function of this series of instrument is used,
the user needs to set the conversion relationship of the analog quantity ex-
pected value based on requirements. The analog quantity values in each chan-
nel all follow the calculation relationship of y=Mx+b. The user needs to set the
MX (slope coefficient) and MB (offset) values in different channels in the analog
quantity menu. For calculation of the analog quantity value. The user can calcu-
late the M and b values based on the analog requirements in accordance with
the formula below.
Taking the Ch1 programming channel as an example, the user needs to convert
the values of Mx and Mb according to the following formulas, and then set the
two values through the front panel keys or SCPI remote commands.
The principle of parameter setting of Ch1 and Ch2 and Ch3 is the same, so
the description will not be repeated.
Name Description
• CV priority
• CC priority
How to Use
The following takes the CV priority mode as an example to introduce the usage
steps.
2. Calculate the Mx and Mb under the Ch1, Ch2, Ch3 menus according to the
above formulas. The sample data used in this manual is shown in the table
below.
The actual output voltage and current of the instrument will change accord-
ing to the following rules:
• Pin 9, pin 10 monitor the actual output current of the instrument: When
the output current is higher than the upper limit I+ set by pin 9, the instru-
ment outputs the current at the value of I+; When the output current is
lower than the lower limit I- set by pin 10, the instrument outputs the cur-
rent at the value of I-.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Turn the knob to select the System Info and press [Enter].
Parameter Description
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Turn the knob to select the System Reset and press [Enter].
3. Press the Left / Right key or turn the knob to adjust the value of this
parameter.
• No: Default value, indicates the instrument cancels this menu setting.
The parameters affected by System Reset and the parameter values after
reset are shown in the following table.
System Beep On
menu
PowerOn Reset
Sense Off
• IO–3: Off-Status
• IO–4: Trig(In)
• IO–5: INH-Living
• IO–6: Sync-On
• IO–7: Sync-Off
Parallel Single
Config Mode CV
menu
Speed High
On Delay/Off Delay 0
This Chapter introduces in detail how to upgrade the system with the USB inter-
face on the front panel. For the upgrading method of Web, see 2.5.2.1 Using
Web Server.
Before upgrading, please contact ITECH technical staff to obtain the follow-
ing two upgrading files, and put them under the root directory of the USB
flash drive.
• itech_3900_P.itech
• ItechConfig.txt
For example, when there are multiple system upgrading package post-
fixed with .itech in the root directory of your USB flash drive, you need to
use a text editing tool to open the configuration file and specify the up-
grade package name corresponding to the present upgrade operation.
2. The upgrading methods for single unit or units in parallel are slightly differ-
ent: Under the parallel mode, the user can select to upgrade all units or a
single unit only by operating the master. Therefore, the user needs to get ac-
cess to SN numbers of all instruments in the parallel network (To obtain the
unique ID for identifying instrument, see 5.13 View the System Information
(System Info)) for selective upgrading based on the SN number in subse-
quent upgrades.
3. If the upgrade fails, the interface will prompt Update fail. At this point, you
cannot use the instrument. Please contact ITECH technical staff for
treatment.
Upgrade Operation
1. Insert the USB flash drive into the USB interface on the front panel of the
instrument.
2. Power on the instrument. At this point, press Shift for several times till
the instrument can detect the system upgrading files in the USB flash
drive.
After the system upgrading files are detected, the interface displays as
follows:
Update Now?
No Yes
If the USB flash drive is not detected, select Yes to detect again; if No
is selected, it means to exit upgrading, and the system will directly en-
ter into the main interface.
3. Select Yes, and press [Enter]. The system interface displays as follows:
Update Select:00/01
SN: ALL
• System Upgrade under Parallel Mode (Only Need to Operate the Master)
1. Insert the USB flash drive into the USB interface on the front panel of the
instrument.
2. Power on the instrument. At this point, press Shift for several times till
the instrument can detect the system upgrading files in the USB flash
drive.
After the system upgrading files are detected, the interface displays as
follows:
Update Now?
No Yes
– SN1: xxxxxxxxxxxxxxxxxx
6 Technical Specification
This chapter will introduce the main technical parameters of the instrument,
such as rated voltage/current/power and so on. Besides, we will introduce the
working environment and storage temperature.
♦ Main Specification
♦ IT-M3901C-10-170
♦ IT-M3903C-10-340
♦ IT-M3905C-10-510
♦ IT-M3910C-10-1020
♦ IT-M3902C-32-80
♦ IT-M3904C-32-160
♦ IT-M3906C-32-240
♦ IT-M3912C-32-480
♦ IT-M3902C-80-40
♦ IT-M3904C-80-80
♦ IT-M3902C-85-40SAS
♦ IT-M3904C-85-80SAS
♦ IT-M3906C-85-120SAS
♦ IT-M3901C-150-25PV
♦ IT-M3903C-150-50PV
♦ IT-M3906C-80-120
♦ IT-M3912C-80-240
♦ IT-M3902C-300-20
♦ IT-M3904C-300-40
♦ IT-M3906C-300-60
♦ IT-M3912C-300-120
♦ IT-M3902C-500-12
♦ IT-M3904C-500-24
♦ IT-M3906C-500-36
♦ IT-M3912C-500-72
♦ IT-M3902C-800-8
♦ IT-M3904C-800-16
♦ IT-M3906C-800-24
♦ IT-M3912C-800-48
♦ IT-M3906C-1500-12
♦ IT-M3912C-1500-24
♦ Supplemental Characteristics
6.1.1 IT-M3901C-10-170
Parameter IT-M3901C-10-170
Voltage 0~10V
Current -120A~170A
Power -1200W~1700W
Rated value
Resistance in series
(CV priority) 0~0.06Ω
Load resistance
(CC priority) 0.009Ω~3Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.001Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3901C-10-170
Rise Time( no load ) Voltage ≤50ms
Rise Time( full load
Voltage ≤100ms
)
Fall Time( no load ) Voltage ≤100ms
Fall Time( full load ) Voltage ≤50ms
Transient Response Voltage ≤10ms
Time
Voltage ≤0.05% + 0.05%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.002%*I + 0.05%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -125A or 175A
OVP 10.5V
Output protection
Overpower
protection -1224W or 1734W
Remote Sense
Compensation ≤2V
Voltage
External programming voltage -10V to 10V corre-
current program sponds to current -170A to 170A (effective range
-120A to 170A)
Current -170A to 170A corresponds to external
External analog current monitor monitoring voltage -10V to 10V (effective range
(optional) -120A to 170A)
External programming voltage 0 to 10V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 10V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50Hz/60Hz
Max. apparent
power 1.85kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 10kg
6.1.2 IT-M3903C-10-340
Parameter IT-M3903C-10-340
Voltage 0~10V
Current -240A~340A
Power -2400W~3400W
Rated value
Resistance in series (CV
priority) 0~0.03Ω
Voltage 0.001V
Setup Resolution Current 0.01A
Power 1W
Parameter IT-M3903C-10-340
Resistance in series (CV
priority) 0.001Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series (CV
priority) ≤1%FS
Parameter IT-M3903C-10-340
Overpower protection -2448W or 3468W
Remote Sense Com-
pensation Voltage ≤2V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent power 3.7kVA
Max. input current 12.5A
Max. efficiency 92%
Power Factor 0.99
DC component ≤0.2A
Current harmonics ≤3%
Standard: USB/LAN/CAN/IO
Standard Interface Optional: GPIB/Analog&RS232
Command Response Time 0.1ms
Parallel Number ≤16
Working Temperature 0~40℃
Storage Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
6.1.3 IT-M3905C-10-510
Parameter IT-M3905C-10-510
Voltage 0~10V
Current -360A~510A
Power -3600W~5100W
Rated value
Resistance in series (CV
priority) 0~0.02Ω
Voltage 0.001V
Current 0.1A
Power 1W
Setup Resolution
Resistance in series (CV
priority) 0.001Ω
Voltage 0.001V
Read Back
Resolution Current 0.1A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series (CV
priority) ≤1%FS
Parameter IT-M3905C-10-510
Power ≤0.5% + 0.5%FS
Vp-p ≤65mVpp
Voltage ripple*1
Vrms ≤10mV
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load
Voltage ≤50ms
)
Rise Time( full load
Voltage ≤100ms
)
Fall Time( no load ) Voltage ≤50ms
Fall Time( full load ) Voltage ≤100ms
Transient Response Voltage ≤10ms
Time
Voltage ≤0.05% + 0.05%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage 0.0035%*I + 0.05%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -370A or 520A
Output protection OVP 10.5V
Overpower protection -3672W or 5202W
Remote Sense
Compensation ≤2V
Voltage
External programming voltage -10V to
current program 10V corresponds to current -510A to 510A
(effective range -360A to 510A)
Current -510A to 510A corresponds to ex-
External analog current monitor ternal monitoring voltage -10V to 10V (ef-
(optional) fective range -360A to 510A)
External programming voltage 0 to 10V
voltage program
Corresponding voltage 0 to 10V
Voltage 0 to 10V corresponds to external
voltage monitor
monitoring voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 5.55kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 15kg
6.1.4 IT-M3910C-10-1020
Parameter IT-M3910C-10-1020
Voltage 0~10V
Rated value
Current -720A~1020A
Parameter IT-M3910C-10-1020
Power -7200W~10200W
Resistance in series
(CV priority) 0~0.02Ω
Load resistance
(CC priority) 0.00015Ω~0.5Ω
Voltage 0.001V
Current 0.1A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.001Ω
Voltage 0.001V
Read Back
Resolution Current 0.1A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3910C-10-1020
Transient Response
Time
(change from 25% Voltage ≤10ms
to 90% of rated
current)
Voltage ≤0.05% + 0.05%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage 0.0035%*I + 0.05%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -740A or 1040A
OVP 10.5V
Output protection
Overpower
protection -7344W or 10710W
Remote Sense
Compensation ≤2V
Voltage
External programming voltage -10V to 10V corre-
current program sponds to current -1020A to 1020A (effective
range -720A to 1020A)
Current -1020A to 1020A corresponds to external
External analog current monitor monitoring voltage -10V to 10V (effective range
(optional) -720A to 1020A)
External programming voltage 0 to 10V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 10V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
Optional: GPIB/Analog&RS232
Command Re-
sponse Time 0.1ms
Parallel Number ≤8
Working
Temperature 0~40℃
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension ( mm) 767.62mm*483mm*106.9mm
Weight( net) 30kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.5 IT-M3902C-32-80
Parameter IT-M3902C-32-80
Voltage 0~32V
Current -80A~80A
Power -2000W~2000W
Rated value
Resistance in ser-
ies (CV priority) 0~0.2Ω
Load resistance
(CC priority) 0.002Ω~1200Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.001Ω
Parameter IT-M3902C-32-80
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3902C-32-80
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -80A to 80A
Current -80A to 80A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 32V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 32V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 2.25kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
6.1.6 IT-M3904C-32-160
Parameter IT-M3904C-32-160
Voltage 0~32V
Current -160A~160A
Power -4000W~4000W
Rated value
Resistance in series
(CV priority) 0~0.2Ω
Load resistance
(CC priority) 0.006Ω~600Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.001Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3904C-32-160
Power ≤0.5% + 0.5%FS
Vp-p ≤80mVpp
Voltage ripple*1
Vrms ≤30mV
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load
Voltage ≤30ms
)
Rise Time( full load
Voltage ≤60ms
)
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.02% + 0.02%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.02% + 0.02%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -165A or 165A
OVP 33V
Output protection
Overpower
protection -4080W or 4080W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -160A to 160A
Current -160A to 160A corresponds to external
current monitor monitoring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 32V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 32V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 4.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 12.5kg
6.1.7 IT-M3906C-32-240
Parameter IT-M3906C-32-240
Rated value Voltage 0~32V
Parameter IT-M3906C-32-240
Current -240A~240A
Power -6000W~6000W
Resistance in ser-
ies (CV priority) 0~0.2Ω
Load resistance
(CC priority) 0.005Ω~400Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.001Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3906C-32-240
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 25% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.02% + 0.02%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.02% + 0.02%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -250A or 250A
OVP 33V
Output protection
Overpower
protection -6120W or 6120W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -240A to 240A
Current -240A to 240A corresponds to external
current monitor monitoring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 32V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 32V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 6.5kVA
Standard: USB/LAN/CAN/IO
Standard Interface Optional: GPIB/Analog&RS232
Command Re-
sponse Time 0.1ms
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 15kg
6.1.8 IT-M3912C-32-480
Parameter IT-M3912C-32-480
Voltage 0~32V
Current -480A~480A
Power -12000W~12000W
Rated value
Resistance in ser-
ies (CV priority) 0~0.2Ω
Load resistance
(CC priority) 0.0025Ω~200Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.001Ω
Parameter IT-M3912C-32-480
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.05% + 0.05%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3912C-32-480
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -480A to 480A
Current -480A to 480A corresponds to external
current monitor monitoring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 32V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 32V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 767.62mm*483mm*106.9mm
Weight( net) 30kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.9 IT-M3902C-80-40
Parameter IT-M3902C-80-40
Voltage 0~80V
Current -40A~40A
Power -2000W~2000W
Rated value
Resistance in ser-
ies (CV priority) 0~0.3Ω
Load resistance
(CC priority) 0.01Ω~800Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3902C-80-40
Power ≤0.5% + 0.5%FS
Vp-p ≤200mVpp
Voltage ripple*1
Vrms ≤80mV
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load ) Voltage ≤15ms
Rise Time( full load ) Voltage ≤30ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 25% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -42A or 42A
OVP 82V
Output protection
Overpower
protection -2040W or 2040W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -40A to 40A
Current -40A to 40A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 80V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 80V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 2.25kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 10kg
6.1.10 IT-M3904C-80-80
Parameter IT-M3904C-80-80
Voltage 0~80V
Rated value
Current -80A~80A
Parameter IT-M3904C-80-80
Power -4000W~4000W
Resistance in series
(CV priority) 0~0.3Ω
Load resistance
(CC priority) 0.015Ω~1200Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3904C-80-80
Transient Response
Time
(change from 25% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -84A or 84A
OVP 82V
Output protection
Overpower
protection -4080W or 4080W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -80A to 80A
Current -80A to 80A corresponds to external mon-
current monitor itoring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 80V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 80V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 4.5kVA
Optional: GPIB/Analog&RS232
Command Re-
sponse Time 0.1ms
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 12.5kg
6.1.11 IT-M3902C-85-40SAS
Parameter IT-M3902C-85-40SAS
Voltage 0~85V
Current -40A~40A
Power -2000W~2000W
Rated value
Resistance in ser-
ies (CV priority) 0~0.3Ω
Load resistance
(CC priority) 0.03Ω~2400Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Parameter IT-M3902C-85-40SAS
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3902C-85-40SAS
Overpower
protection -2040W or 2040W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage 0V to 10V corre-
current program
sponds to current -40A to 40A
Current -40A to 40A corresponds to external moni-
current monitor toring voltage 0V to 10V
External analog
(optional) External programming voltage 0 to 85V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 85V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*1 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 2.25kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension ( mm) 744.22mm*459mm*56.81mm
Weight( net) 10kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.12 IT-M3904C-85-80SAS
Parameter IT-M3904C-85-80SAS
Voltage 0~85V
Current -80A~80A
Power -4000W~4000W
Rated value
Resistance in series
(CV priority) 0~0.3Ω
Load resistance
(CC priority) 0.015Ω~1200Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3904C-85-80SAS
Voltage ≤0.03% + 0.03%FS
Read Back Accuracy Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Vp-p ≤200mVpp
Voltage ripple
Vrms ≤60mV
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load ) Voltage ≤15ms
Rise Time( full load ) Voltage ≤30ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% to Voltage ≤1ms
90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -84A or 84A
OVP 86V
Output protection
Overpower
protection -4160W or 4160W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage 0V to 10V corre-
current program
sponds to current -80A to 80A
Current -80A to 80A corresponds to external mon-
current monitor itoring voltage 0V to 10V
External analog
(optional) External programming voltage 0 to 85V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 85V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*1 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 4.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension ( mm) 744.22mm*459mm*56.81mm
Weight( net) 12.5kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.13 IT-M3906C-85-120SAS
Parameter IT-M3906C-85-120SAS
Voltage 0~85V
Current -120A~120A
Power -6000W~6000W
Rated value
Resistance in series
(CV priority) 0~0.3Ω
Load resistance
(CC priority) 0.01Ω~800Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3906C-85-120SAS
Current ≤50ppm/℃
Rise Time( no load ) Voltage ≤15ms
Rise Time( full load ) Voltage ≤30ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -125A or 125A
OVP 86V
Output protection
Overpower
protection -6120W or 6120W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -120A to 120A
Current -120A to 120A corresponds to external
current monitor monitoring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 85V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 85V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*1 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 6.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension ( mm) 744.22mm*459mm*56.81mm
Weight( net) 15kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.14 IT-M3901C-150-25PV
Parameter IT-M3901C-150-25PV
Voltage 0~150V
Current -25A~25A
Power -1500W~1500W
Rated value
Resistance in ser-
ies (CV priority) 0~0.5Ω
Load resistance
(CC priority) 0.04Ω~1500Ω
Voltage 0.01V
Setup Resolution
Current 0.001A
Parameter IT-M3901C-150-25PV
Power 1W
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Setup Accuracy Power ≤0.5% + 0.5%FS
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3901C-150-25PV
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -26A or 26A
OVP 160V
Output protection
Overpower
protection -1530W or 1530W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage 0V to 10V corre-
current program
sponds to current -25A to 25A
Current -25A to 25A corresponds to external moni-
current monitor toring voltage 0V to 10V
External analog
(optional) External programming voltage 0 to 150V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 150V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*1 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 1.65kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1600Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension ( mm) 660mm*437mm*43.5mm
Weight( net) 10kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.15 IT-M3903C-150-50PV
Parameter IT-M3903C-150-50PV
Voltage 0~150V
Current -50A~50A
Power -3000W~3000W
Rated value
Resistance in ser-
ies (CV priority) 0~0.5Ω
Load resistance
(CC priority) 0.02Ω~1500Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Setup Accuracy Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Parameter IT-M3903C-150-50PV
Resistance in ser-
ies (CV priority) ≤1%FS
Remote Sense
Compensation ≤5V
Voltage
External analog External programming voltage 0V to 10V corre-
current program
(optional) sponds to current -50A to 50A
Parameter IT-M3903C-150-50PV
Current -50A to 50A corresponds to external moni-
current monitor toring voltage 0V to 10V
External programming voltage 0 to 150V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 150V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*1 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 3.3kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1600Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension ( mm) 660mm*437mm*43.5mm
Weight( net) 12.5kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.16 IT-M3906C-80-120
Parameter IT-M3906C-80-120
Voltage 0~80V
Current -120A~120A
Power -6000W~6000W
Rated value
Resistance in series
(CV priority) 0~0.3Ω
Load resistance
(CC priority) 0.01Ω~800Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3906C-80-120
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load
Voltage ≤15ms
)
Rise Time( full load
Voltage ≤30ms
)
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 25% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -125A or 125A
OVP 82V
Output protection
Overpower
protection -6120W or 6120W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -120A to 120A
Current -120A to 120A corresponds to external
current monitor monitoring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 80V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 80V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 15kg
6.1.17 IT-M3912C-80-240
Parameter IT-M3912C-80-240
Voltage 0~80V
Rated value Current -240A~240A
Power -12000W~12000W
Parameter IT-M3912C-80-240
Resistance in series
(CV priority) 0~0.3Ω
Load resistance
(CC priority) 0.005Ω~400Ω
Voltage 0.001V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.001V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3912C-80-240
(change from 25%
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -250A or 250A
OVP 82V
Output protection
Overpower
protection -12240W or 12240W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -240A to 240A
Current -240A to 240A corresponds to external
current monitor monitoring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 80V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 80V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 300Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 767.62mm*483mm*106.9mm
Weight( net) 30kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.18 IT-M3902C-300-20
Parameter IT-M3902C-300-20
Voltage 0~300V
Current -20A~20A
Power -2000W~2000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.09Ω~9000Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Parameter IT-M3902C-300-20
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3902C-300-20
External programming voltage -10V to 10V corre-
current program
sponds to current -20A to 20A
Current -20A to 20A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 300V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 300V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent power 2.25kVA
Max. input current 12.5Aac
Max. efficiency 94.5%
Power Factor 0.99
DC component ≤0.2A
Current harmonics ≤3%
Standard: USB/LAN/CAN/IO
Standard Interface Optional: GPIB/Analog&RS232
Command Response
0.1ms
Time
Parallel Number ≤16
Working
Temperature 0~40℃
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 800Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 10kg
6.1.19 IT-M3904C-300-40
Parameter IT-M3904C-300-40
Voltage 0~300V
Current -40A~40A
Power -4000W~4000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.065Ω~4500Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3904C-300-40
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load ) Voltage ≤60ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% to Voltage ≤1ms
90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -42A or 42A
OVP 303V
Output protection
Overpower
protection -4080W or 4080W
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 800Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 12.5kg
6.1.20 IT-M3906C-300-60
Parameter IT-M3906C-300-60
Voltage 0~300V
Current -60A~60A
Power -6000W~6000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.05Ω~3000Ω
Parameter IT-M3906C-300-60
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3906C-300-60
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -63A or 63A
OVP 303V
Output protection
Overpower
protection -6120W or 6120W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -60A to 60A
Current -60A to 60A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 300V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 300V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 6.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 800Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 15kg
6.1.21 IT-M3912C-300-120
Parameter IT-M3912C-300-120
Voltage 0~300V
Current -120A~120A
Power -12000W~12000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.1Ω~1500Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.001Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Setup Accuracy
Current ≤0.1% + 0.1%FS
Parameter IT-M3912C-300-120
Power ≤0.5% + 0.5%FS
Resistance in ser-
ies (CV priority) ≤1%FS
Remote Sense
Compensation ≤5V
Voltage
External analog External programming voltage -10V to 10V corre-
current program
(optional) sponds to current -120A to 120A
Parameter IT-M3912C-300-120
Current -120A to 120A corresponds to external
current monitor monitoring voltage -10V to 10V
External programming voltage 0 to 300V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 300V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 800Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 767.62mm*483mm*106.9mm
Weight( net) 30kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.22 IT-M3902C-500-12
Parameter IT-M3902C-500-12
Voltage 0~500V
Current -12A~12A
Power -2000W~2000W
Rated value
Resistance in series
(CV priority) 0~1Ω
Load resistance
(CC priority) 0.3Ω~15000Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3902C-500-12
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load ) Voltage ≤60ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -12.5A or 12.5A
OVP 505V
Output protection
Overpower
protection -2040W or 2040W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -12A to 12A
Current -12A to 12A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 500V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 500V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent power 2.25kVA
Max. input current 12.5Aac
Max. efficiency 94.5%
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1600Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 10kg
6.1.23 IT-M3904C-500-24
Parameter IT-M3904C-500-24
Voltage 0~500V
Current -24A~24A
Power -4000W~4000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.15Ω~7500Ω
Voltage 0.01V
Setup Resolution Current 0.001A
Power 1W
Parameter IT-M3904C-500-24
Resistance in ser-
ies (CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3904C-500-24
OCP -24.6A or 24.6A
OVP 505V
Output protection
Overpower
protection -4080W or 4080W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -24A to 24A
Current -24A to 24A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 500V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 500V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 6.5kVA
Storage
Temperature -10℃~70℃
IP IP20
6.1.24 IT-M3906C-500-36
Parameter IT-M3906C-500-36
Voltage 0~500V
Current -36A~36A
Power -6000W~6000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.1Ω~5000Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Setup Accuracy Power ≤0.5% + 0.5%FS
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3906C-500-36
Load resistance Lower limit: 1/(1/Rset+(1/Rset)*0.05+0.0001)
(CC priority) Upper limit: 1/(1/Rset-(1/Rset)*0.05-0.0001)
Voltage ≤0.03% + 0.03%FS
Read Back Accuracy Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Vp-p ≤500mVpp
Voltage ripple*1
Vrms ≤90mV
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load ) Voltage ≤60ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% to Voltage ≤1ms
90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -37A or 37A
OVP 505V
Output protection
Overpower
protection -6120W or 6120W
Parameter IT-M3906C-500-36
Voltage 0 to 500V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 6.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1600Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 744.22mm*459mm*56.81mm
Weight( net) 15kg
6.1.25 IT-M3912C-500-72
Parameter IT-M3912C-500-72
Voltage 0~500V
Current -72A~72A
Power -12000W~12000W
Rated value
Resistance in series
(CV priority) 0~1Ω
Load resistance
(CC priority) 0.05Ω~2500Ω
Voltage 0.01V
Current 0.01A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.01A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3912C-500-72
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load ) Voltage ≤60ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% to Voltage ≤1ms
90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -74A or 74A
OVP 505V
Output protection
Overpower
protection -12240W or 12240W
Remote Sense
Compensation ≤5V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -72A to 72A
Current -72A to 72A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 500V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 500V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
DC component ≤0.2A
Current harmonics ≤3%
Standard: USB/LAN/CAN/IO
Standard Interface Optional: GPIB/Analog&RS232
Command Re-
sponse Time 0.1ms
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1600Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension(D*W*H) 767.62mm*483mm*106.9mm
Weight( net) 30kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.26 IT-M3902C-800-8
Parameter IT-M3902C-800-8
Voltage 0~800V
Current -8A~8A
Power -2000W~2000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.45Ω~22500Ω
Voltage 0.01V
Current 0.001A
Setup Resolution Power 1W
Resistance in ser-
ies (CV priority) 0.01Ω
Parameter IT-M3902C-800-8
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3902C-800-8
OVP 808V
Overpower
protection -2040W or 2040W
Remote Sense
Compensation ≤8V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -8A to 8A
Current -8A to 8A corresponds to external monitor-
current monitor ing voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 800V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 800V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 2.25kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1600Vdc
6.1.27 IT-M3904C-800-16
Parameter IT-M3904C-800-16
Voltage 0~800V
Current -16A~16A
Power -4000W~4000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.22Ω~11250Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Setup Accuracy Power ≤0.5% + 0.5%FS
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3904C-800-16
Load resistance Lower limit: 1/(1/Rset+(1/Rset)*0.05+0.0001)
(CC priority) Upper limit: 1/(1/Rset-(1/Rset)*0.05-0.0001)
Voltage ≤0.03% + 0.03%FS
Read Back
Accuracy Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Vp-p ≤1600mVpp
Voltage ripple*1
Vrms ≤200mV
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load ) Voltage ≤60ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 25% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -16.8A or 16.8A
OVP 808V
Output protection
Overpower
protection -4080W or 4080W
Remote Sense
Compensation ≤8V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -16A to 16A
External analog Current -16A to 16A corresponds to external moni-
(optional) current monitor toring voltage -10V to 10V
External programming voltage 0 to 800V Corre-
voltage program
sponding voltage 0 to 10V
Parameter IT-M3904C-800-16
Voltage 0 to 800V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 4.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to
1600Vdc
GND
Isolation AC to
3500Vdc
GND
Cooling Air
Dimension ( mm) 744.22mm*459mm*56.81mm
Weight( net) 12.5kg
6.1.28 IT-M3906C-800-24
Parameter IT-M3906C-800-24
Voltage 0~800V
Current -24A~24A
Power -6000W~6000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.15Ω~7500Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3906C-800-24
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load ) Voltage ≤60ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 25% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -25A or 25A
OVP 808V
Output protection
Overpower
protection -6120W or 6120W
Remote Sense
Compensation ≤8V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -24A to 24A
Current -24A to 24A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 800V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 800V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 6.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1600Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension
744.22mm*459mm*56.81mm
(D*W*H)
Weight( net) 15kg
6.1.29 IT-M3912C-800-48
Parameter IT-M3912C-800-48
Voltage 0~800V
Current 0A~48A
Power 0W~12000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.08Ω~3750Ω
Voltage 0.01V
Setup Resolution
Current 0.001A
Parameter IT-M3912C-800-48
Power 1W
Resistance in ser-
ies (CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3912C-800-48
Current ≤0.05% + 0.05%FS
OCP -50A or 50A
OVP 808V
Output protection
Overpower
protection -12240W or 12240W
Remote Sense
Compensation ≤8V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -48A to 48A
Current -48A to 48A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 800V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 800V corresponds to external monitor-
voltage monitor
ing voltage 0 to 10V
Other parameters:
Storage
Temperature -10℃~70℃
IP IP20
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
6.1.30 IT-M3906C-1500-12
Parameter IT-M3906C-1500-12
Voltage 0~1500V
Current -12A~12A
Power -6000W~6000W
Rated value
Resistance in ser-
ies (CV priority) 0~1Ω
Load resistance
(CC priority) 0.5Ω~7500Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in ser-
ies (CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Setup Accuracy Power ≤0.5% + 0.5%FS
Resistance in ser-
ies (CV priority) ≤1%FS
Parameter IT-M3906C-1500-12
Load resistance Lower limit: 1/(1/Rset+(1/Rset)*0.05+0.0001)
(CC priority) Upper limit: 1/(1/Rset-(1/Rset)*0.05-0.0001)
Voltage ≤0.03% + 0.03%FS
Read Back
Accuracy Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Vp-p ≤1500mVpp
Voltage ripple*1
Vrms ≤150mV
Voltage ≤30ppm/℃
Setup Temperature
Coefficient Current ≤50ppm/℃
Voltage ≤30ppm/℃
Read Back Temper-
ature Coefficient Current ≤50ppm/℃
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load ) Voltage ≤60ms
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -12.5A or 12.5A
OVP 1515V
Output protection
Overpower
protection -6120W or 6120W
Remote Sense
Compensation ≤15V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -12A to 12A
External analog Current -12A to 12A corresponds to external moni-
(optional) current monitor toring voltage -10V to 10V
External programming voltage 0 to 1500V Corre-
voltage program
sponding voltage 0 to 10V
Parameter IT-M3906C-1500-12
Voltage 0 to 1500V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
3phase 200V~480V
Grid voltage range
AC Input*2 Single phase 100V~240V
Frequency 50/60Hz
Max. apparent
power 6.5kVA
Storage
Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1800Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension
744.22mm*459mm*56.81mm
(D*W*H)
Weight( net) 15kg
6.1.31 IT-M3912C-1500-24
Parameter IT-M3912C-1500-24
Voltage 0~1500V
Current -24A~24A
Power -12000W~12000W
Rated value
Resistance in series
(CV priority) 0~1Ω
Load resistance
(CC priority) 0.25Ω~3750Ω
Voltage 0.01V
Current 0.001A
Power 1W
Setup Resolution
Resistance in series
(CV priority) 0.01Ω
Load resistance
(CC priority) 0.01Ω
Voltage 0.01V
Read Back
Resolution Current 0.001A
Power 1W
Voltage ≤0.03% + 0.03%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Setup Accuracy
Resistance in series
(CV priority) ≤1%FS
Parameter IT-M3912C-1500-24
Rise Time( no load ) Voltage ≤30ms
Rise Time( full load
Voltage ≤60ms
)
Fall Time( no load ) Voltage ≤30ms
Fall Time( full load ) Voltage ≤15ms
Transient Response
Time
(change from 20% Voltage ≤1ms
to 90% of rated
current)
Voltage ≤0.01% + 0.01%FS
Line Regulation
Current ≤0.03% + 0.03%FS
Voltage ≤0.01% + 0.01%FS
Load Regulation
Current ≤0.05% + 0.05%FS
OCP -25A or 25A
OVP 1515V
Output protection
Overpower
protection -12240W or 12240W
Remote Sense
Compensation ≤15V
Voltage
External programming voltage -10V to 10V corre-
current program
sponds to current -24A to 24A
Current -24A to 24A corresponds to external moni-
current monitor toring voltage -10V to 10V
External analog
(optional) External programming voltage 0 to 1500V Corre-
voltage program
sponding voltage 0 to 10V
Voltage 0 to 1500V corresponds to external moni-
voltage monitor
toring voltage 0 to 10V
Other parameters:
DC component ≤0.2A
Current harmonics ≤3%
Standard: USB/LAN/CAN/IO
Standard Interface Optional: GPIB/Analog&RS232
Command Response Time 0.1ms
Parallel Number ≤16
Working Temperature 0~40℃
Storage Temperature -10℃~70℃
IP IP20
Isolation DC to GND 1800Vdc
Isolation AC to GND 3500Vdc
Cooling Air
Dimension ( mm) 660mm*437mm*87mm
Weight( net) 30kg
*1: At low voltage input, the set power will decrease. For more information,
please contact ITECH.
7 Routine Maintenance
This chapter describes the general maintenance items and maintenance meth-
ods of the device.
♦ Instrument Self-Test
♦ Cleaning and Maintenance
♦ Contact of ITECH Engineers
♦ Return for Repair
• Re-start the instrument. Each time the instrument is powered on, the self-
test is performed. This test assumes that the instrument is running.
Ensure that all test cables connections are removed when self-test is per-
formed. During self-test, errors may be induced by signals present on ex-
ternal wiring, such as long test leads that can act as antennae.
Clean the outside and the front panel screen of the instrument using a soft, lint-
free, cloth slightly dampened with water. Use a brush to clean the dust on the
vent and cooling fans.
When the instrument fails, you should make the following preparations before
returning the instrument to ITECH for repair or contacting engineers.
• Check all the items listed in the Self-inspection of equipment faults and con-
firm whether there are still some problems.
• Collect the SN of the instrument.
For details, see Collect the SN number.
If there are still some problems, carefully read the Warranty and Limitation of
Warranty in the preface of the manual. Confirm that your instrument complies
with warranty service conditions. If after your warranty expires, ITECH offers re-
pair services at competitive prices.
When the instrument fails, make the following checks to ensure that the failure
is in the instrument rather than any external connections. If the instrument fail-
ure can be eliminated via simple inspection, the maintenance cost and time can
be saved.
• Check whether the AC power cord is securely connected to the instrument
and to a AC distribution panel.
• Check whether the front-panel Power On switch has been turned on.
• Check whether self-inspection of the instrument is successful and whether
the specifications and performance are within the indicator ranges.
ITECH will constantly improve the product performance, availability and reliabil-
ity. The service personnel of ITECH will record changes of each instrument. All
relevant information is marked uniquely according to the serial number of each
instrument. The equipment returned for repair must adopt the SN number as the
tracking ID.
When you contact the engineer, the effective SN number of the instrument will
be the effective guarantee for effective service and complete information. You
can obtain the SN number of the instrument by the following ways:
2. Rotate the knob to select System Info and press [Enter] key.
To get the service for your instrument, choose the easiest way to contact the
ITECH engineers. ITECH will arrange for repair or replacement of your instru-
ment or provide warranty and repair cost information (if applicable).
ITECH recommends that you retain the original shipping carton for return ship-
ments and always insure shipments. To ship the unit to ITECH for service or
repair:
1. Download the ITECH maintenance service application from our website,
complete it and place it in the box with the instrument.
2. Place the unit in its original container with appropriate packaging material.
If the original shipping container is unavailable, use a container that will en-
sure at least 10 cm (4 in.) of compressible packaging material around the en-
tire instrument. Use static-free packaging materials.
3. Secure the container with strong tape or metal bands.
A Appendix
♦ Specifications of Red and Black Test Cables
♦ Troubleshooting
♦ Fuse Replacement
AWG 8 10 12 14 16 18 20 22 24 26 28
The
Maxi-
mum 60 40 30 20 13 10 7 5 3.5 2.5 1.7
Current
Value (A)
• AWG (American Wire Gage), it means X wire (marked on the wire). The table
above lists current capacity of single wire at working temperature of 30°C. For
reference only.
• Along with temperature, you must also consider voltage drop when selecting
wire sizes.
Although the instrument will compensate for the voltage in the wires, it is recom-
mended to minimize the voltage drop as much as possible to prevent excessive
power consumption from the instrument and poor dynamic response to load
changes. Larger diameter wire sizes will help minimize wire voltage drops.
Twisting or bundling wires will help reduce transient voltage drops.
A.2 Troubleshooting
The fault information of this series of instruments is described below.
INH LATCH Disable output, turn off the For details, see
output. 5.11.5 IO–5. INH-
Living, Not-Invert.
INH LIVING Disable output, turn on the
output.
FIBER EXT UNLOCK The outer ring of the fiber is Fiber multi-master
not locked. parallel: set the par-
allel in the System
menu in turn. For
details, see 5.10
FIBER INNER UNLOCK The inner ring of the fiber is Please contact
not locked. ITECH authorized
dealer or ITECH
technical support.
If there are no fuse assembly on the instrument rear panel, it means that you
can’t replace the fuse by yourself. Please contact the ITECH engineer on the
condition of the same malfunction.