2-Month Maint
2-Month Maint
WARNING
Probe Wash 3 contains 4.5% potassium hydroxide and 2% sodium hypochlorite.
Avoid contact with skin and eyes. Probe Wash 3 is a corrosive material that can
cause burns. Wear suitable protective clothing, gloves and eye/face protection. In
case of contact with eyes, rinse immediately with plenty of water and seek medical
advice.
1. Prepare 1.5 liters of 5% Probe Wash 3 solution diluted with deionized water.
2. Set the analyzer to Standby.
3. Remove the 6 cuvette segments on the dilution tray (DTT).
a. Unfasten the 2 thumbscrews (1) on each section.
b. Grasp the cuvette section by the tab (2) and lift it from the tray.
c. To remove the cuvettes under the dilution washer (DWUD) and splash cover, turn
the tray by hand until the cuvettes are clear.
NOTE: You will feel resistance when manually moving the tray. This is normal.
Maintenance 139
4. Immerse the cuvette segments in 5% Probe Wash 3 solution.
a. Ensure no air bubbles are in the cuvettes.
b. Allow the cuvettes to soak for at least 10 hours.
5. Wash the cuvettes under running water, then rinse them in deionized water.
6. Drain the water from the cuvettes.
7. Install the cuvette segments on the dilution tray (DTT) and fasten the thumbscrews by
hand.
8. Return the analyzer to the READY mode.
9. At the Operation Panel, select Initialize.
NOTE: Verify that the Operating mode field displays READY before performing any
further actions.
10. At the Operation Panel, select WASH.
11. Perform the daily shutdown wash routine, then verify the operation.
WARNING
Before handling any reagents, read the warnings on page 192
CAUTION
Make a note of the bottle position on the shelf to avoid mixing up the fluid
bottles.
CAUTION
Ensure that the level-sensor connector (2) does not get wet to avoid damaging it.
Maintenance 141