EDSin TEM
EDSin TEM
Contents
• Why EDS in the TEM?
• EDS setup in the TEM
• Recapitulation of the important exp. Parameters
• X ray detector and detection in the TEM?
• Artefacts in the TEM
- above sample
- below
- inside (electron channeling: useful or annoying?)
• Quantification
- Basics, Cliff-Lorimer factors
- Corrections
• Special cases
- Light element analysis
- Trace element analysis (minimum mass fraction)
• How to get best spatial resolution
- beam broadening
- electron beam shaping
- contamination and solutions
Why EDS in the TEM
A different point of view than bulk analysis
Example: EBID Rh small particle analysis
1
6 50 Electron
7
42
probe 3
12 10 11
13 14 8 9
16 15 21 27 17 20 18 Total surface of small
23
24 19 22 30
26 28 3129 25 33 particles: 6700nm2
32
388 36 34 35 37
39 42 43 40
41
Total surface 33000nm2
44
45 49 46 4748
51 52 50 5 -> total Rh coverage
54 56 57 5558 59
62 60 61 20%
63 64 65
68 67
69 70
107 7166 But in Nanoprobe
73 74 72 75
7677 79
78 80
81 analysis Rh
85 84 82 86
83 87 concentration 100%
88 89 90
93 9694
99
9195
97 Or Rh grains no pure
98 92 101 100 102
1033 Rh grains?
106 104 105
109 108 110
1111114113 112 115
Why EDS in the TEM
For 1%
statistical
error count
at least
intensity of
105 counts
i.e. about 1min
Choice of the Dead time
electron beam
blanked as
soon as X ray
detected in
EDS detector
-> 4 fold
increase of
countrate
How increase countrate in TEM
EDS aperture
1) Use thick C2
aperture
2) Use additional
EDS aperture
above sample
Post specimen artefacts
• High angle
scattering
-> electrons generate X rays
far away from beam
centre
• Synchroton
radiation
-> X rays fluoresce far
away from beam centre
-> Minimize by
keeping sample
horizontal
-> possible solution: use sample holder, grids surrounding in Be
C Contamination
Stop contamination!!
- limit organics in sample
preparation
Contamination spot
- O Plasma cleaner
- Ultrasonic bath….
- Clean sample holder
- ……
Peak to background ratio P/B
• Peaks well
defined
• Good P/B ratio
for concentrations
above 0.1%
Quantification
In the TEM:
No absorption correction I ~c
No fluorescence correction
Detector sensitivity
Mass absorption
coefficient
Example: absorption correction
Example: Lineprofile
Ni/Ti multilayers with V-O-N impurities
Ni
Ti
Total
diameter
Average
diameter
Detector
sensitivity
for ultralight
elements low
Ultralight element analysis II
Example O, N, B
1) strong X ray in sample
2) strong X ray in detector
window
3) Weak detector sensitivity
But: Analysis still possible:
Example: Si/Si2 Ti interface
Enhancement of spatial resolution
Deconvolution may
yield subnanometer
resolution
Minimum mass fraction I
Trace elements often hidden in unsufficient
experimental conditions
Si 0.2%Fe