AEP 7.2 Instruction For Use IFU - 011986A
AEP 7.2 Instruction For Use IFU - 011986A
(AEP) System
Copyright © 2013
Natus Medical Incorporated
®
Bio-logic Auditory Evoked Potential
(AEP) System
Copyright © 2013
Natus Medical Incorporated
AEP can be purchased with 1 or 2 channels and can be used with the following transducers: TDH-39
headphones, Bio-logic Insert earphones, bone oscillator, or sound field speakers.
AEP comes standard with three stimulus options, click, toneburst and custom.
Default protocols include ABR (1 & 2 channel), ECochG, AMLR, Forty (40) Hz response,
Other optional features that can expand your AEP capabilities are:
• GraphMaster normative data
• P300
• Electroneuronography (ENOG)
The Bio-logic Navigator Pro EP System can be used for patients of all ages, from children to adults, including
infants and geriatric patients. The use of the Bio-logic EP family of products is to be performed under the
prescription and supervision of a physician or other trained health care professional.
CAUTION
The Navigator Pro is intended for use by a heath care professional or a
technician who is trained and supervised by a health care professional or
accredited organization.
WARNING
Never turn power on or off with a patient connected to the system. Only
qualified personnel should use this device. Please read this section before
installing any of the hardware. Refer to this section when you operate,
transport, store, or re-install the system.
WARNING
All of the disposable supplies from Natus that are coupled to the scalp are
latex-free and comprised of hypoallergenic material. Natus is not
responsible for disposable items purchased from another vendor. Carefully
read and follow the proper use instructions that accompany any disposable
items.
WARNING
Make sure that any platform, table, cart, or other surface used during the
operation, transport, or temporary or permanent storage of the system and
its components is adequate, sturdy, and safe. Natus is not responsible for
any injury or damage that may result from inadequate, poorly constructed,
or unapproved transports, carts, operating surfaces
WARNING
Never use equipment that has parts missing or equipment that might
contain loose parts inside of it (that is, inside an enclosed portion of the
equipment). If you suspect a piece of equipment has missing or loose parts,
make contact with Natus Medical Incorporated.
CAUTION
Simultaneous use of the Navigator Pro and other patient-connected
equipment, such as a cardiac pacemaker or other electrical implants, may
interfere with recording of auditory evoked potentials.
WARNING
Simultaneous use of the optional ENOG electrical stimulator and other
patient-connected equipment, such as a cardiac pacemaker or other
electrical implants, may not be appropriate. Consult the Digitimer operator’s
manual.
CAUTION
The optional ENOG electrical stimulator is appropriate only for
percutaneous stimulation, meaning electrical stimulation through the skin.
Use of needle electrodes for stimulation is NOT recommended. Consult the
Digitimer operator’s manual.
CAUTION
This system is intended for use as a medical instrument only. The
software installed on this system has been verified and validated for
that use, as long as no other unintended or unauthorized software is
installed and running at the same time. Do not install, download, save,
or transfer other programs, software, or data on this equipment without
first consulting Natus Medical Incorporated. Doing so may interfere
with the proper operation of this software.
Natus warrantees the Navigator Pro system for a period of one year and this warranty includes the Natus
supplied computer. Natus’s warranty does not cover compatibility problems that can arise when the customer
adds non-Natus software or hardware to a Natus supplied computer.
Note: Natus Medical is not responsible for resolving incompatibilities between the kit and a
computer that is provided by the customer or authorized Natus dealer representative.
Hardware Installation
WARNING
Each of the colored plastic housings on the cables displays arrows that identify
the top of the connector. This helps align the cable connector male plug to the
female jack on the side of the Navigator Pro system module. Careful alignment
of the cable plug to the jack is important to reduce the chance of bending the
metal pins inside the male connector. You must insert the plugs completely
into the corresponding jacks for the system to operate properly. Never twist
these connectors during insertion or removal. Twisting the connector in the
socket may damage the metal pins and cause communication problems
between the Navigator Pro system module and the computer. It may also
cause other system malfunctions involving stimulus delivery or system
recording.
This figure shows color-coding for the connectors on the Navigator Pro module and cables:
You have a choice of transducers when purchasing your Navigator Pro system. All transducers must be
supplied by Natus Medical to assure that they meet proper specifications. Depending on what you choose,
some of the connector jacks on the sides of the Navigator Pro system will remain unused. For the system to
operate, however, connections must be made to the blue (electrode connection cable), the red (serial cable) or
USB connector, the power supply connector, and the yellow connector (transducer cable) for delivery of the
stimulus into the ear.
Once the cable connections are made into the Navigator Pro module, an additional connection must be made to
enable communication between the Navigator Pro Hardware and the computer. Insert either the 9-pin D
connector (female) end of the serial cable or the USB connector into the computer.
6. Begin Collection
• Select View EEG icon to check the EEG prior to recording. • Allow program to collect to maximum number of sweeps as
defined in protocol.
• Select Record icon to begin averaging, when EEG is
acceptable.
OR
Note: the stimulus always ramps up to the full
intensity and ramps down when averaging is • Select Stop icon to stop collection before maximum number
stopped. of sweeps as defined in protocol is reached.
OR
OR
9. Print Report
• Select Patient, and then Print Report. A print preview of the report will be displayed.
OR
• Click on the printer icon on the tool bar at the top left of the screen
• To select report templates other than the default, use the drop down arrow in the middle of the printing toolbar.
• Select Print icon.
Electrode Connections
.
1-Channel Recordings
• Both connectors for Channel 1 (Input 1 and Input 2) will be used as well as the Common
connector.
• For 1-channel recordings, the inputs for Ch. 2 are not used and should be empty.
• No jumper should be connected to the patient cable.
• When changing ear stimulated, it is not necessary to switch the electrode wires in the patient
cable, when using an ‘Electrode Switching Enabled’ montage. Important: Electrode Switching
must be enabled in the collection protocol. When using electrode switching, the non-test
ear will always automatically be assigned as the common or ground electrode. Refer to the
chart below to insure the electrodes are plugged in correctly.
If the AEP program cannot communicate with the Navigator Pro unit an error message appears:
Select a protocol from the drop down list. You only have access to protocols that your hardware supports
(e.g.: you will not have P300 protocols unless you purchased P300).
CAUTION
Protocol settings, data fields and controls are typically optimized for
the various protocols during installation. Only trained and experienced
users should change these settings. To create a new protocol, we
recommend that you start with the protocol that is most similar to what
you want. Natus assumes no responsibility for any problems that may
arise from improper use.
Toolbars
Channel
Selection Waveform display
& area - Panel B
P300 Waveform
Selection display area -
Closes the Panel A
control
panel
window
The AEP program uses the Main Test Screen to collect data. It gives you all of the information you need to
track the progress of your test or to change test parameters during data collection. The splitter between Panel
A and Panel B can be moved left or right to create a Split screen or a Full screen. Note: Pull down the
“display” menu and select “default display” to return to the default split screen display.
Exceed max: Exceeds the maximum number of sweeps preset in the protocol.
Change stimulated ear: Changes the stimulus from one ear to the other ear.
Increase Intensity: Increases the stimulus intensity by the dB step size specified
in the protocol.
Decrease Intensity: Decreases the stimulus intensity by the dB step size specified
in the protocol.
Show Flat Line - flat lines the beginning of the epoch that is being blocked.
Stimulus Blocking
Appearance Show Response - displays the stimulus artifact in the epoch that is being
blocked.
Specific Scale
Scales all waveforms to a scale defined by the user.
(uV)
Panel B is selected and displays those collections selected by the user for display from the field tree.
After positioning the cursor, add a label to the waveform by clicking on the corresponding label number in the
toolbar, or right click on the waveform, left click on label menu item and select the desired label.
Move a label by placing the cursor to the desired location and remarking the label.
Delete all labels on the waveform by clicking on the icon. You cannot delete only one label at a time.
1-5 Labels 1-5 These numbers represent the labels in the protocol. If you place
A Label Set can your mouse on the number, a tool tip will display the actual label
contain up to 10 (e.g. = 1 = I, 2 = II, 3 = III, 4 = IV and 5 = V)
labels.
Modify label set Changes the label set associated with the highlighted wave(s).
Delete label Deletes all of the labels on the highlighted wave(s). A single label
cannot be deleted.
When a waveform is the currently collecting record, the tag in front of the wave will be highlighted in Yellow.
When data collection has completed, the tag will be highlighted in Blue for the left ear, Red for the right ear,
and Green for a both (bilateral) ears waveform.
Move a waveform by left clicking on it and dragging it up or down, or select the wave and use the or
icons or use the up and down arrow keys on the keyboard.
Multiple waveforms can be selected by holding the Ctrl key and left clicking either on level six (polarity of
stimulus & masking) of the field tree or on the waveforms in the panel.
Waveforms can also be selected and repositioned from the Display drop-down menu.
Note: Choose “default layout” from the Display drop-down menu to resize the panels and data tree to their
original settings.
Right clicking on a specific waveform gives access to the commands below. Note these functions can
also be accessed from the drop-down menus and toolbar buttons.
Lists labels that can be applied to the waveform. The label will be applied at the location of
Label
the active cursor.
Wave info Launches Wave Information dialog to review collection parameters.
List of analysis comments. When applied, an abbreviation appears on the record in the field
Analyze
tree so that users can identify it easily. These analysis comments include:
Defines the waveform as the “Minimum Response Level” (the
Min. Response Level
lowest intensity where a response is observed).
Response Defines the waveform as showing a “Response.”
Threshold Defines a waveform as representing the “Threshold.”
Shows that the waveform exhibited no response to the test
No Response
stimulus.
Launches an Edit Comment display to enter a comment for an individual selected
Comment
waveform. This is displayed in the Records List.
Erases the selected waveform(s) on the active panel. This only erases the waveform from
Erase
the display. It is not a permanent deletion of the waveform.
Marked
Point Launches the Marked Point Information display for the wave under the mouse cursor.
Information
Digital Filter Launches a read-only dialog box displaying the digital filter settings for this waveform as it
Setting is currently displayed on the panel.
Set Display Launches the Set Display Scale dialog box. Scaling will be applied to the selected
Scale waveform(s) on the panel
Show flat line – displays a flat line at the beginning of the epoch that is being blocked.
Blocking
Show response – displays the stimulus artifact/response in the portion of the epoch which is
Appearance
being blocked.
Alt Lists options related to displaying a waveform collected using an alternating polarity
Waveform stimulus.
Displays alternating Rarefaction Displays the
Alternating only
polarity waveforms only. on/off rarefaction waveform.
Baseline
Displays a baseline through the waveform.
On/Off
Erases a selected waveform(s). This only erases the waveform from the display. It is
Erase
not a permanent deletion of the waveform.
If the waveforms have been manually moved to a different location on the panel, refresh
Refresh
will move the waveforms back based on the wave position algorithm in effect.
Highlights all of the waveforms in the panel. Any action that you perform after selecting
Select All
this option affects all of the waveforms.
Additional Wave Analysis Functions – access from Analysis drop-down menu and
toolbar buttons
Gives a list of labels to apply to selected waveforms. The label is applied at the location of
Label
the active cursor.
Delete
Deletes all of the labels on the highlighted wave(s).
Labels
Modify
Changes the label set to be used to mark waveforms.
Label Set
Digital Digitally filters the selected waveform(s). Waves are saved in the database
Filter based on the filters set during collection. Post hoc digital filtering cannot
(optional extend the filter settings wider than the filter settings used at the time of
feature) collection.
Note: To activate these icons, you must highlight more than one waveform. Select
Calculations the first waveform, then hold down the control (Ctrl) key and select a 2nd
waveform. This highlights 2 waveforms.
Weighted Add: Averages 2 or more
waveforms together. Weighting is Unweighted Add: Sums 2 or more waveforms
based on the number of sweeps in each together equally with no weighting.
waveform.
Weighted Subtract: Subtracts the 2nd
waveform highlighted from the 1st Unweighted Subtract: Subtracts the 2nd
waveform highlighted. Weighting is waveform highlighted from the 1st waveform
based on the number of sweeps in each highlighted. No weighting is used.
waveform.
Correlate: Provides a correlation between 2
selected waveforms. Set Cursor 1 and
Save Calculated Waves: Saves the Cursor 2 around the section of waveforms you
calculated waveforms as a permanent want to correlate (e.g. Wave V). Highlight
part of the test database. both waveforms. If the correlation is 0.90,
then the section of the waveform in between
the cursors is 90% similar.
Note: If all the waveforms on the panel are scaled the same, the scaling will be displayed in the top left corner
of the panel. If the waveforms are not scaled the same, no value will appear.
0.50 uV
Auto-specific scaling: Identifies the largest selected waveform and scales all other
selected waveforms on the active panel relative to largest
waveform.
Scaling radio
Specific scale: Scales all selected waveforms to a defined scale.
buttons
Epoch-based scaling: Sets a specific scale based on the epoch (window) size.
Epoch based scaling defaults are setup in the Default.
Display Parameters section.
GraphMaster is a graphic display of marked point latency data (click stimuli only), compared to age-specific
normative data. Shaded areas reflect age specific normative ranges.
As many as four different graphic displays may appear on the GraphMaster View
Graph displays screen.
Note: Use the graphs that match the decibel reference used for collecting the data.
To choose a report other than the default report, click the drop down arrow next to the report name to access
the other reports.
Note: You must choose a report that matches the data on the screen. The Report Templates are
designed to match the label set of the marked points in the protocol. Use of an inappropriate report
template can result in the mislabeling of the marked points in your report.
Note: It is possible to print a report to .PDF format which may be used to interface with electronic
medical records. Natus Medical does not provide this capability as part of our software.
Access the database from the Patients, Database drop-down or from the database toolbar or by clicking the
database button on the Open Patient dialog.
Select a patient by highlighting the patient file in the table and double click on it. This will bring the patient
information into the Open Patient dialog box.
Select a patient and click on the Form View icon. The Form view provides access to additional
information for that patient’s record.
Backup – Creates an XML file of the entire database without deleting files from the database.
2. Setup the default data paths for archive, copy and backup
• In the Database, Table view, change the database location (window on toolbar that reads ‘AEP’ for
the default active database) to the folder where the archived data resides.
• If data is stored on a removable drive, select ‘Other’. The Open browser appears. Locate the path
where the database resides. Select the blpatients.mdb file and click Open.
• Backup creates a backup of newly created or modified patient files that have not been previously
backed up in XML format.
• Select Patient, Database Tools, Backup Databases.
• Verify the source and destination before proceeding with the backup.
• Automatic backups can be scheduled by accessing Setup, Patient, Backup.
Recommendation:
The Protocol Name field gives a list of default protocols.
To create a New Protocol, start with the protocol that is most similar and edit it until it meets your needs.
Click on ‘Save As’ and enter a new name for your protocol so that you still have access to the default protocol.
Makes the currently selected protocol, the default protocol whenever you launch the
Make Default
application.
Removes any protocol you choose from the protocol list. Some protocols cannot be
Delete
deleted.
Labels/Calculation
Displays the Labels/Calculations/Areas for that protocol.
tab
Saves any changes made to a protocol to a new protocol name. The program retains the
Save As
original protocol with its original parameters.
Cancel command Closes the Collection Protocol Setup screen without implementing any changes.
Applies any changes to the protocol you have selected or are creating but leaves the
Apply command dialog box open.. The program uses these changes for subsequent tests in the session.
To save them permanently, use the Save or Save As commands.
Epoch Time (ms) Select the desired epoch time for the recording window in milliseconds (ms).
# Points field Sets the number of points (samples) in the epoch i.e. resolution of the cursor.
Determines the start of the epoch relative to stimulus onset. By entering a non-zero
millisecond (ms) value in the Pre/Post time, you can start the collection before or
Pre/Post Time slide
after the stimulus is delivered. Negative values begin the epoch prior to stimulus
control
onset. Positive values begin the epoch after stimulus onset. An entry of zero (0)
starts the epoch at the onset of the stimulus.
Blocks high amplitude activity in a portion of the epoch from being counted as an
Blocking slide control artifact. Typically used to avoid the situation where a large stimulus artifact will
cause rejection of the entire epoch.
Sets the maximum number of averages you want to include in the recording. The
Maximum # of Averages
recording will automatically terminate when this number is achieved.
Save Impedance Test
Checking this box, saves the impedance test values with the recording.
Values
Note: Refer to the AEP Reference Guide for information on the other parameter settings.
This field contains a list of the transducers that you can use with the AEP system.
Transducer
Transducers include:
Headphones
Insert Earphones
Bone Oscillator
Ear Determines the ear(s) to be stimulated: Left, Right, or Binaural (both).
Determines the intensity (dB level) to begin stimulating and the decibel reference (SPL,
Intensity
nHL, or HL).
Sets a correction value to apply to the latency caused by the delay from acoustic tubes
Insert Delay (ms)
on insert earphones. For all Bio-logic supplied tubes, the correction is 0.8 ms.
Stim Rate (per sec) Sets the number of stimuli presented per second.
Intensity Step Determines the amount by which the intensity increases or decreases.
Determines the stimulus type used for the test. Options are Click, Toneburst, or Custom.
Stimulus Type Note: Custom stimuli can be designed by researchers. For additional information, refer to
the Navigator Pro Reference Guide.
Click Duration (usec) Determines the duration of the click. The default of 100 usec is most commonly used.
Determines the type of Ramping used for the toneburst stimulus. Options are Linear,
Ramp
Hanning, Gaussian and Blackman. Displayed when Toneburst is selected.
Plateau (cycle) Determines the Plateau of the toneburst. Displayed when Toneburst is selected.
Rise/Fall (cycle) Determines the Rise/Fall of the toneburst. Displayed when Toneburst is selected.
2 - Channel
1 - Channel
Channel Enable Note: If Channel 2 is enabled, make sure that Input 2 for both Channel 1 & 2 is
checkboxes set according to how you plan to plug the electrodes into the patient cable. This is
important for automatic loading of collecting waves. Use A1 (left ear) for
Channel 1 Input 2 and A2 (right ear) for Channel 2 Input 2 if you use the
standard 10-20 system of electrode placement.
Determines the amplifier gain. More gain is needed for smaller responses and less
Gain
gain is needed for larger responses. The default value for ABR is 100,000.
Select these checkboxes to activate artifact rejection for each channel. Each gain
setting has a default artifact rejection level. To change the artifact rejection
independently from the gain, type in the value desired.
Artifact Reject
Note: If excessive environmental noise causes every sweep to be rejected, turn off
the Artifact Reject (deselect checkbox) and record the noise to determine the
frequency of the noise.
Low Filter (Hz) Sets the low cutoff filter setting. This is a High Pass filter.
High Filter (Hz) Sets the high cutoff filter setting. This is a Low Pass filter.
Select these checkboxes to apply the notch filter settings for each channel. The notch
filter should only be used when troubleshooting 50 Hz or 60 Hz noise. If noise
Notch Filter
disappears when the notch filter is on, check the electrical outlet and the electrical
environment (other equipment).
Selection is based on the 10-20 Electrode Placement System and the electrode montage
being used.
Input 1
NOTE: Settings in these fields combined with which ear is being stimulated,
determines whether a channel is “Ipsilateral” or “Contralateral”.
Selection is based on the 10-20 Electrode Placement System and the electrode montage
being used.
Input 2
NOTE: Settings in these fields combined with which ear is being stimulated,
determines whether a channel is “Ipsilateral” or “Contralateral”.
For 1-channel protocols, allows for automatic switching (through the program), of
reference and ground electrodes based on the stimulated ear. The user does not have to
Electrode Switching
manually move electrodes. Requires plugging in electrodes in a specific way. Follow
the instructions on the screen.
Sample Tag Displays sample tags for each Channel.
Edit Tag command Select the Edit Tag button to launch the Edit Tag setup display.
Choose the appropriate label set for the protocol or select the closest label set
Label Set Name
and edit it to meet your needs.
Edit Label Set Accesses the Label/Calculation/Ratio Setup dialog box for editing.
Edit Select this command button to edit the fields on the dialog box.
Save Saves the Label Set changes under the current name.
Saves the Label Set changes under a new name. The original label set remains
Save As
with the original values.
Labels,
Displays the label(s), calculation(s), ratio(s) and area(s) defined for the
Calculations, Ratios,
selected protocol.
Areas
Closes the Label/Calculation/Ratio Setup dialog box and implements any
OK command
changes you have saved.
Closes the Label/Calculation/Ratio Setup dialog box without implementing
Cancel command
any changes.
Launches a Facility Name dialog box to enter the Facility name. This will
Edit Facility Name
print near the top of the report(s).
Edit User Defined Launches a User Defined Comment dialog box to enter a fixed comment. This
Comment field can be added to a report template.
Populates the Print Report screen with a specific report template. Use the
Use default template
browse button to select the preferred template.
Displays the data path location for the report templates. The default location
Templates’ path depends on the Windows Operating System. Refer to the Folder Structures
section of the Appendix.
1. Left click the mouse and drag a box around the entire section you want to delete or move. When you
release the mouse button the fields you selected will display surrounded by small boxes around the
perimeter.
To select a field component that is not wanted, hold down the shift button on the keyboard and click
on the unwanted field.
To delete the selected fields, click the delete key on the keyboard.
3. To undo any change that you made in error, click on the Undo button .
60 (50) Hz interference
• Only rarely is 60 or 50 cycle noise the cause of noisy recordings or artifacts. Using a notch filter frequently produces ringing
which increases artifacts. So, to check for true 60 (50) Hz interference, temporarily enable the notch filter. If the problem goes
away, find and correct the source of the noise so that the notch filter may be disabled during actual data collection.
• Turn off fluorescent lights or lights with dimmer switches. Test the outlet to make sure it is grounded.
• In some cases, it will be necessary to install a dedicated electrical outlet or shield the test environment.
Software/Protocol variables
• Shorten the analysis window to 9 msec or less.
• Change the click rate slightly.
• Use a post-stimulation time (+1) or add blocked points.
• Decrease gain on all channels in one or two increments (e.g. Reduce 150,000 to 100,000; reduce 100,000 to 75,000.
• Use alternating polarity.
• Reduce high filter (low pass) to 1500 Hz., OR
• Raise low filter (high pass) to 50, 75 or 100Hz.
• Disable the artifact rejection as a last resort. Check if any useful, readable data can be collected. Interpret with caution as noise
is being averaged into the recording.
Communication Issues (PC to Navigator Pro hardware)
• USB or Serial communication between the device and the host computer may be lost as a result of electrical interference. When
communication is lost, it can be re-established after the interference has ended by clicking on the retry button on the dialog box.
2. Separate the patient cable from the transducer cable and from any power supplies and cables.
3. If possible, avoid clipping the transducer stimulus boxes to the patient. Clip to a pillowcase instead. Do not let the
tubing of the transducers touch the electrodes or rub on the patient’s clothing.
4. Do not place the ground electrode close to the heart (front or back of the patient). Noise can be generated by a
large EKG.
6. Turn off fluorescent light(s) or dimmer switches when operating the equipment. Do not have dimmer set in the
middle position.
7. Turn off/Unplug any equipment in the same room as the system i.e. monitor, computer, typewriter, coffee pot etc.
Of course, temporarily disabling medical monitoring equipment used on the patient must be performed by a
qualified nurse or physician.
8. Only plug the Navigator Pro into the supplied isolation transformer. Do not use outlet strips or extension cords.
Connect the isolation transformer directly to wall outlet.
9. If patient is lying on a chair that plugs in – unplug the chair. If patient is lying on a metal bed, do not have
electrodes touching the metal of the bed.
10. Use a dedicated circuit (line) for the system (i.e. AEP system is only system in outlet.) Verify with an electrician
or Bio-medical dept. If not using dedicated line – other office equipment (i.e. copier) can introduce noise.
11. Assure that your wall outlet is grounded and wired correctly. Verify with an electrician or Bio-medical dept.
12. Do you have a radio station transmission tower nearby? This will create noise.
13. Are there any large devices in adjacent rooms or adjacent floor (e.g. X-ray equip. MRI, refrigerators, air
conditioner – window unit, elevator motors, etc.)
15. If 60 or 50 Hz noise is present (in most window sizes there will be only 1 or 2 cycles present), turn on the notch
filter and see if the noise disappears. See Protocol – Amplifier Tab section.
16. Click on View EEG button on the collection screen. Place the electrodes into a cup of water and walk the cup of
water and electrodes around the room to identify the noise source. Watch the EEG window, as you get closer to the
noise source the EEG will become noisier.
17. If only artifact is being collected, turn off the artifact rejection so that data collection can occur. Collect the noise,
so that you can fax it to Natus to determine the frequency of the noise to assist you with troubleshooting.
18. Move the system to a different room away from the existing location. Move system to a different part of the
building if noise cannot be resolved.
CAUTION
Be careful not to twist the patient electrode cable and transducer
connectors as you remove them. Metal pins inside the connectors can be
bent or broken by twisting, using excessive force, or if they are not
positioned correctly.
2. Disconnect the patient and transducer cables from the Navigator Pro unit.
3. Connect the system loop test cable (part number: 541-TSTCBL) to both the blue and yellow-ringed
sockets on the right side of the Navigator Pro system by matching the connector color to the color
ring around the socket. Make sure the flat side of the molded connector faces the label side of the
box. Be sure to push the connector into the socket as far as it will go.
Yellow Connector
Blue Connector
4. After setting up the system with the loop test cable, follow this process to test the internal components
of the Navigator Pro module:
Click Continue. . .
Click Details. . .
The Loop Test Cable to Patient Cable Adapter (001077) is an adapter cable used to connect the loop test cable
(541-TSTCBL) to the patient connection cables (541-NAVC10). This adapter is used for troubleshooting
purposes only. It attaches to the patient input cable and can be used to eliminate the patent cable as the
problem area. The Loop Test Cable to Patient Cable Adapter has a Blue Mini-Din connector on one end and
five colored 1.5 mm “touch-proof” connectors on the other end.
The cable does not attach to the patient nor should it be used when a patient is connected to the device.
The 1.5 mm touch proof connectors attach to the patient cable. These touch proof connectors are color coded
to match the input connectors on the patient cable. The connections are attached as indicated in the table.
The Patient Cable blue mini-din attaches to the Navigator Pro unit at the patient connector (blue ring). Follow
the section above for performing a system self-diagnostic using the cable.
If the Self Diagnostic test passes using this cable, then you can rule out problems with the patient cable.
For WinXP:
Also, note when browsing in Windows explorer, Documents part shows up as “Shared Documents”,
but when you look in the address bar after selection it will show just Documents.
WARNING
Equipment that supports this program must be designed to meet all applicable
medical equipment standards for patient isolation and electrical safety. Natus
assumes no responsibility for the equipment into which this program is
installed, or for errors and omissions made during the installation process. It is
the installer's responsibility to insure the finished system meets appropriate
equipment standards, such as IEC 60601, BS 5724, and UL2601.
WARNING
Never place powered equipment (that is, equipment that operates with an
electric power source) on any flammable surface. Avoid this whether the
equipment is active or not.
IMPORTANT
This device is intended for use by qualified personnel only. Please read this
section and the remainder of the AEP Instruction for Use Manual before
installing any of the AEP System hardware, Retain this section as a reference
while operating, transporting, storing or re-installing the AEP System
equipment.
Attention
Type BF equipment
Grounding Requirements
The desired installation site must have a grounding wire for electrically connecting the system equipment to a
properly grounded terminal (Earth). The grounded terminal shall be connected to the steel skeleton or under
the main reinforcing steel with any multiple grounding wires grounded at the same point. (One point
grounding.) The shield room or bed must also be grounded at the same grounding terminal as the installation
site.
• A 3-pin power plug requires a 3-pin wall socket with a properly installed grounding pole. Verify
the wall receptacle is correctly grounded before inserting the 3-pin power plug.
• A 2-pin/3-pin adapter for connecting 3-pin power supply plugs into a 2-pin wall socket. The
grounding terminal on the adapter must be fixed with the free end secured to the ground terminal.
• A 2-pin power plug with a grounding terminal for a 2-pin wall socket. The grounding terminal on
the power plug must be secured to the ground terminal.
Environmental Conditions
• Select a room with properly grounded power sources.
• Do not use or store AEP systems and equipment in places where chemicals are stored or where
there is a potential for gas leakage.
• Avoid moisture or contact with water, extreme atmospheric pressure, excessive humidity and
temperature, poorly ventilated areas and dusty, saline or sulfuric air.
• Verify the selected site maintains a relative humidity between 25% and 95% (without
condensation).
• Verify all conditions meet the requirements listed in the Environmental Specifications section of
this manual.
Room Topography
• Place all equipment on an even, level surface. Avoid the potential for mechanical shock or
possible vibrations during setup system operation or when relocating the equipment.
• Verify the iso-transformer and all portable multiple socket-outlets are off the floor and in a dry
location.
Conducted RF 3 Vrms 3V
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the Navigator Pro is
used exceeds the applicable RF compliance level above, the Navigator Pro should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orientating or relocating the Navigator Pro.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Humidity
Operating: 30% to 75% at 40° C non-condensing
Note: These are maximums and should not be considered ‘normal’ operating or storage environments.
Condensation
Recovery Time after condensation to operations specifications: 24 hours
Flammability
UL94V-O
EMC Specifications
Cispr 11 B Conducted and Cispr 11 B Radiated Emissions
EN61000-4-2 Electrostatic Discharge Test
EN61000-4-3 Radiated Susceptibility Test
EN61000-4-4 Transient Susceptibility Test
EN61000-4-5 Surge Susceptibility Test
EN61000-4-6 Conducted Immunity Tests
EN61000-4-11 Voltage Fluctuations Test
EN61000-3-2 & EN61000-3-3 Dips and Flicker Test
RE 101 Magnetic Emissions Test
System Hookups
WARNING
Only use the Power Supply provided with the AEP system.
• Verify the system computer power supply (if present) is plugged into either:
—an iso-transformer from the 520-XFS family of iso-transformers
OR
—a 520-AMPS01 iso-transformer.
• Do not interconnect multiple pieces of equipment without verifying that the sum of all leakage
currents does not exceed the safety limit.
• Do not connect non-medical electrical equipment directly to a wall outlet if the AEP System is using
a multiple, portable socket-outlet with a separating iso-transformer. (The additional equipment may
increase the current amount of leakage and exceed the safety limit.)
• Do not connect equipment, which can potentially provide electromagnetic or other types of
interference. This may cause the AEP System equipment to function incorrectly.
• Verify the equipment is connected to a power line source with the following frequency, voltage and
current capacity.
Frequency: 50 / 60 Hz
Voltage: 100 – 240 VAC
Current Capacity: 15 Amps min.
• Use only Ink Jet printers for AEP System setups that use the AEP System isolation transformer.
Installation Verification
Use the following to verify all equipment is correctly installed before use.
• Verify all equipment and cables are undamaged and in good working condition.
• Verify all equipment and cables are connected according to the instructions in this manual.
• Verify all circuitry with a direct connection to the patient has been checked twice.
• Verify all system or equipment batteries show the correct voltage and are in perfect working
condition.
WARNING
All of the disposable supplies from Natus that are coupled to the scalp are
latex-free and comprised of hypoallergenic material. Natus is not responsible
for disposable items purchased from another vendor. Carefully read and
follow the proper use instructions that accompany any disposable items.
WARNING
Make sure that any platform, table, cart, or other surface used during the
operation, transport, or temporary or permanent storage of the system and its
components is adequate, sturdy, and safe. Natus is not responsible for any
injury or damage that may result from inadequate, poorly constructed, or
unapproved transports, carts, operating surfaces
WARNING
Never use equipment that has parts missing or equipment that might contain
loose parts inside of it (that is, inside an enclosed portion of the equipment). If
you suspect a piece of equipment has missing or loose parts, make contact
with Natus Medical Incorporated.
WARNING
Never place powered equipment (that is, equipment that operates with an
electric power source) on any flammable surface. Avoid this whether the
equipment is active or not.
WARNING
This equipment is not protected from a cardiac defibrillator discharge. Using a
defibrillator while a patient is connected to the system equipment may
permanently damage the equipment.
1. Follow all safety procedures, giving careful and constant attention to the patient and system
equipment.
2. Follow hospital procedures to validate all electrodes are properly applied in order to prevent patient
burns while using electro-surgical equipment.
3. Turn the system power on BEFORE you attach electrodes to the patient. Remove the electrodes from
the patient BEFORE you turn the system power off. This avoids shock hazards from power surges.
4. Attach electrodes to the patient and the system hardware so that the electrode cables can neither
entangle nor strangle the patient.
5. Avoid direct contact between the patient and system equipment at all times.
6. Do not add other equipment that is not protected against ingress of liquids.
7. This equipment is not suitable for use in areas of flammable or anesthetic mixture.
1. The system computer and all other peripheral devices are plugged into the isolation transformer.
2. The portable computer is powered by batteries. It does not plug into a wall outlet. No AC-powered
peripherals except the AEP System module (Navigator Pro) connect to the portable computer.
Cleaning
No special cleaning is required for this device. Ordinary, ‘house-keeping’ type cleaning to prevent the
accumulation of dust or debris is adequate.
Inspection
No special inspection procedures are required for this device.
1. Return all controls to their original positions and turn OFF all system equipment.
Maintenance
AEP System equipment requires professional maintenance and repair service. Never attempt to alter, modify or
repair this equipment on your own.
1. If the AEP System equipment is not functioning correctly, mark all defective parts clearly to prevent
accidental use before the appropriate repairs can be made.
2. When using a defibrillator, either protect the equipment against defibrillator discharge or remove all
patient cables and/or transducers to avoid system damage.
WARNING
Never turn power on or off with a patient connected to the system. Only
qualified personnel should use this device. Please read this section before
installing any of the hardware. Refer to this section when you operate,
transport, store, or re-install the system.
WARNING
Never place powered equipment (that is, equipment that operates with an
electric power source) on any flammable surface. Avoid this whether the
equipment is active or not.
15 or 20 amps
Although 15 amps are acceptable in smaller systems, a 20-amp service is preferable anytime the system in use
has a large amount of features.
BTU Offset
Natus computer systems release a maximum of 2000 BTU per hour.
Magnetic Field
Natus computer systems should operate without any interference in a magnetic field of 0.05 Gauss.
Room Shielding
Although Natus recommends room shielding, it is not absolutely required. Users may elect to shield the room
if the location is close to noise-generating equipment and should provide shielding if there are potential noise
level problems.
Fluorescent Lights
Although the AEP System works in a room with fluorescent lighting, incandescent lighting is preferred. Also,
dimmer switches can produce noise interference unless the switch is set at a full position.
P/N 011986A