Altek 311A Cal Procedure
Altek 311A Cal Procedure
P100904-970 A
Suggested Equipment:
1. Precision DMM with the following specifications:
Resistance Constant Current Source Accuracy
1KΩ 100uA ±0.100Ω
1KΩ 1mA ±0.044Ω
10Ω 10mA ±0.019Ω
Voltage Accuracy
1.00V ±47uV
ENABLING CALIBRATION:
Remove the battery cover and four corner screws. While holding the UUT face down in one hand, carefully
separate the top and bottom housing. Place the unit into Calibration mode by placing the jumper (XCAL) located
on the upper right side of the circuit board into the shorted position (See Figure 1). Verify the UUT is in
Calibration mode by turning on the UUT and viewing the top of the LCD for the word SETUP. Place fresh
batteries into the UUT before any calibration adjustments are made.
CALIBRATION:
NOTE: During calibration, pressing the STORE/RESET button will cause α= to illuminate on the LCD. Then
pressing either °C/°F or 2, 3,4 WIRE will cause the display to blank out and pause, indicating the UUT
received the calibration. While the display is blanked out, do not proceed until the display is restored.
NOTE: Before calibrating the UUT, factory reset the unit by holding the STORE/RESET button while turning
the power on. Continue to hold the STORE/RESET button until the display starts to flash.
311A Field Calibration Procedure DOCUMENT NO. REV.
P100904-970 A
HARDWARE CAL:
1. Connect the 311A output (4-Wire) to the DMM (See Figure 2) and set the DMM to DCV mode.
2. Turn the knob to the Ω position.
3. Press the POWER button to turn the UUT on.
4. Press the SOURCE/READ button until SOURCE is displayed on the UUT LCD.
5. Set the toggle switch to the LO position and adjust pot R260 (See Figure 1) so the DMM reads 0.000mV
±0.005mV.
6. Now place the toggle switch to the HI position and adjust pot R244 (See Figure 1) so the DMM reads
0.000mV ±0.005mV.
NOTE: R260 and R244 are interactive. Thus, it will be necessary to repeat steps 5 and 6 several times.
2-WIRE COMPENSATION:
1. Connect the 311A Output to the Ohmmeter as shown in Figure 4.
2. Set the Ohmmeter to the lowest range with a constant current of 1mA that will measure 500Ω.
3. Check to see that the right side of the UUT LCD displays a 2. This indicates that the excitation current is
1mA.
4. Place the Quik-Chek switch to the HI Position and dial the UUT so the Ohmmeter reads 500.00Ω ± 0.06Ω
and press the STORE RESET button and then the SETUP button. The UUT LCD will display 2-WIRE
when pressing SETUP.
311A Field Calibration Procedure DOCUMENT NO. REV.
P100904-970 A
HARDWARE CHECK:
1. Connect the 311A output (4-Wire) to the DMM (See Figure 2).
2. Turn the knob to select Ni.
CHECK ACCURACY
Turn the UUT off and place the unit into Normal Operating Mode by placing the jumper (XCAL)
located on the upper right side of the circuit board into the open position (See Figure 7).
Reset the unit by holding the STORE/RESET button while turning the power on. Continue to hold the
STORE/RESET button until the display starts to flash.
500
311A Field Calibration Procedure DOCUMENT NO. REV.
P100904-970 A