Standard Operating Procedure: Bio-Chemistry Laboratory
Standard Operating Procedure: Bio-Chemistry Laboratory
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
Press the start/stop button and slowly increase the rpm to the desired speed using
the dial
Once a run is complete, make sure the rotor has COMPLETELY STOPPED
before opening the centrifuge lid by depressing the red stop/start button.
Remove sample vials.
Remember to return the rpm dial back to zero after finishing.
PRECAUTIONS TO BE FOLLOWED
Proper handling of the instrument
Ensure level and stability
Balance centrifuge tubes equally
Ensure use of rubber cushion for glass tubes
Bring speed Knob to off and increase the speed gradually.
Do not open the lid in between the centrifugation cycle
RECORD TO BE MAINTAINED
Laboratory Manual containing the experiments that can be performed with the
equipment
Maintenance Record
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
HoD
17BTCC81-BIOCHEMISTRY LABORATORY
Before use, rinse the electrode with deionized water and blot dry with a soft, clean
paper towel.
Transfer the electrode to the test solution.
Compensate for the temperature if necessary.
Record the pH when the reading is stable (5–20 seconds after insertion of the
electrode into the solution)
Rinse the electrode with deionized water and store according to the manufacturer’s
instructions.
Cleaning and maintenance
Clean the pH meter with a soft, clean, damp paper towel after use. No solvents
should be used.
Replace the electrode filling solution on a regular basis, according to the
manufacturer’s instructions. Record in the logbook (Adjust the speed control knob
to a low stirring rate.
PRECAUTIONS TO BE FOLLOWED
Thoroughly wash the stir bar with distilled water after each application.
Store stir bars in pairs to maintain their magnetic strength and increase their life
span.
RECORD TO BE MAINTAINED
Laboratory Manual containing the experiments that can be performed with the
equipment
Maintenance Record
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
PRECAUTIONS TO BE FOLLOWED
RECORD TO BE MAINTAINED
Laboratory Manual containing the experiments that can be performed with the
equipment
Maintenance Record
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
About 2-3 ml of venous blood collected into EDTA tubes. Specimens should be
transported at room temperature 18 - 26ºC and can be store in the refrigerator of 2
- 8ºC up to 6 hours. If stored in a refrigerator, samples should be returned to room
temperature, for approximately 30 minutes, before analysis.
PROCEDURE:
1.Check operation of the machine, ensuring it is clean and that all required supplies are
present in sufficient quantities.
2. Switch the instrument on by pressing the ON/OFF switch, located on the back of the
instrument.
3. The instrument performs an initialization phase for the internal electronics. Please
wait.
4. Once the initialization phase is complete, the analyzer will automatically run a startup
cycle.
5. If it does not automatically run a startup cycle after the initialization phase is
completed, press "Startup" button in the "Status" area to initiate a startup cycle.
6. Then, the instrument will perform a blank cycle for a reference blank count (an
analysis cycle based on reagents without any blood sample).
7. Check and verify that the reference blank counts do not exceed the following
parameter limits: WBC < 0.3, RBC < 0.02, HGB < 0.3, PLT < 10 then: Press "OK"
button to validate blank results. 8. Perform quality control analysis on 3 levels of control
blood material (low, normal and high) to verify that the instrument is performing within
the specified ranges of the quality control material.
9. Entering patient ID, sample ID, Patient name, etc
10. Follow the indications displayed in the "Sample analysis" dialog box to run the
analysis. a. Mix the sample gently and thoroughly. b. Remove the cap from the sample
tube. c. Place the sample beneath the sampling needle. d. Raise up the tube so that the
sampling needle lowers into the blood and press the manual sample bar. e. The analysis
cycle will begin.
11. When the analysis is completed, the "Sample analysis" dialog box is closed and
results are displayed in the "Result display" menu for print out.
PRECAUTIONS TO BE FOLLOWED
Criteria for rejection haematology specimens 1. When the identification is missing
/inadequate. 2. Insufficient quantity. 3. Inappropriate container 4.Inappropriate
transport/storage. 5. Unknown duration of delay. 6. Clotted sample.
Equipment
RECORD TO BE MAINTAINED
Laboratory Manual containing the experiments that can be performed with the
equipment
Maintenance Record
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
PROCEDURE:
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
PROCEDURE:
1. Fill a clean container with at least 50 ml. of sample water and cover. Set sample aside
to allow sample to equilibrate to air temperature and let gas escape. Avoid contaminants
and analyze as soon as possible.
2) Rinse 2 empty turbidity tubes and caps with a portion of sample. Shake out excess
water.
3) Fill both turbidity tubes to the neck by carefully pouring the sample down the side of
the tube. This will prevent air bubbles from forming. (Similar to pouring soda or beer
into a glass while trying to avoid foam on the top).
4) Cap the tubes and wipe dry with a clean lint free tissue.
5) Carefully invert the turbidity tube twice (do not create air bubbles) just before
inserting the tube into the meter,
6) Open the meter lid. Align the indexing arrow on the tube with the indexing arrow on
the meter. Insert the turbidity tube into the chamber.
7) Close the lid. Push the READ button.The turbidimeter in NTU units will be displayed
within 5 seconds.
8) Record the results from each tube. Take a third reading from a new sample if the
first two readings are significantly different.
9) The meter will turn off automatically one minute after the last button push. To turn
the meter OFF manually, hold the READ button down for at least2 seconds. Release the
button when OFF is displayed.
PRECAUTIONS TO BE FOLLOWED
ER1 (ERROR 1) will be displayed if the battery is low.
ER2 will be displayed when the measured turbidity is very low.
ER3 will be displayed when the light source (bulb) has burned out or if the
turbidity tube is misaligned.
BAT will be displayed when the battery is getting low. The readings are
reliable but the battery needs to be replaced as soon as possible. Carry extra
batteries with the field kit.
RECORD TO BE MAINTAINED
Laboratory Manual containing the experiments that can be performed with the
equipment
Maintenance Record
HoD
17BTCC81 -BIOCHEMISTRY LABORATORY
PROCEDURE:
PRECAUTIONS TO BE FOLLOWED
RECORD TO BE MAINTAINED
Laboratory Manual containing the experiments that can be performed with the
equipment
Maintenance Record
HoD