How To Setup The Dual Detector
How To Setup The Dual Detector
The PC requires two LAN cards except the Internet. LAN Card must support gigabit.
(The files below are included on the CD in the detector box. If you purchased a PC together, it is also
included in the D drive software folder.)
2
Select the detector model to add.
3
Check the added detector model.
4
Open the Config.ini located in the "Venu1717X_2" folder.
5
HostIP=192.168.8.189
6
Make the same changes to the iDetectorconfig.ini file in the “Venu1717X” folder.
7
Stand Detector
Default IP settings:
IP address:
192.168.8.188
Subnet mask:
255.255.255.0
(The detector network speed must be at least 1G. If the speed is 100Mbps, check the LAN cable and network
status.)
8
Table Detector
Default IP settings:
IP address:
192.168.8.189
Subnet mask:
255.255.255.0
(The detector network speed must be at least 1G. If the speed is 100Mbps, check the LAN cable and network
status.)
9
Venu1717X / Venu1717X_2
10
Clear Acq Delay Time => 2200
Select the FreeSync mode and select the Write button to save.
11
Detector Calibration
12
Enter Create Gain page
13
Click "Start" button to start process.
- Take exposure as soon as the Green Progress Bar on the calibration software display.
- After acquired image, display image and display value on the Current Value Text Box. If
the value is close to the Expected Value, the Current Value Text box’s color is changed
GREEN. If the color is changed to GREEN, click the Accept Button. If not, the color is
changed to YELLOW, should change X-ray Exposure condition.
- After clicks accept or not accept, the PREP button is activated. When the PREP button is
activated, work same processing of step
14
- When activated the Generate Button, click generate button after should check the
check box of Download to FPD after generation
15
Download template file dialog will pop up if "Download to FPD after generation" option was checked.
16
Note.
Then click PREP as shown below and the Progress bar, which appears green at the top of the picture screen,
will move for 2 seconds. So X-ray shot should be finished within 2 seconds.
Therefore, before clicking the PREP button, you must click the PREP button to verify that the X-ray Hand
SW is ready for exposure. And press the exposure button of the Hand SW as soon as the green process bar
appears.
17
Generate Defect Template File
You don’t need to change Index no. You just push the button of Download.
18
Grid Calibration (If you are using a grid, you need to calibrate it.)
Fit the detector into the Bucky and Set the grid to the fixed position so that it and use 103 line per inch
for grid.
① Select 1 Create Grid.
19
④ Set the Current Value to around 20000, so that the green color appears.
⑤ Check the message button in the red box as above a and click the generate button.
20
Check as shown above and go to the Acquire tab to check SWGridRM.
21
Open the Config.ini located in the "Venu1717X_2" folder.
Make the same changes to the iDetectorconfig.ini file in the “Venu1717X” folder.
22
Copy and paste the detector calibration file
903-340-26_SMED_SDK_ReleasePackage4.0.34.6582\Tools\iDetector\x64\work_dir\Venu1717X
903-340-26_SMED_SDK_ReleasePackage4.0.34.6582\Tools\iDetector\x64\work_dir\Venu1717X_2
903-340-26_SMED_SDK_ReleasePackage4.0.34.6582\Tools\iDetector\x64\work_dir\Venu1717X_2
C:\ExamVue\work_dir\Venu1717X
C:\ExamVue\work_dir\Venu1717X_2
23
Open the "DeviceConfig.ini" file in the Examvue folder.
24
Open the "DeviceConfig" file in the Examvue folder.
25