R05 MAN REF SAI-Transmitters-Terminals en
R05 MAN REF SAI-Transmitters-Terminals en
Reference Guide
for Transmitters and Terminals
© METTLER TOLEDO 2023
No part of this manual may be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying and recording, for any purpose without the express written
permission of METTLER TOLEDO.
U.S. Government Restricted Rights: This documentation is furnished with Restricted Rights.
Copyright 2023 METTLER TOLEDO. This documentation contains proprietary information of METTLER
TOLEDO. It may not be copied in whole or in part without the express written consent of METTLER
TOLEDO.
COPYRIGHT
METTLER TOLEDO is a registered trademark of Mettler-Toledo, LLC. All other brand or product
®
30587511 | 05 | 02/2023 METTLER TOLEDO SAI Reference Guide for Transmitters and Terminals 1
5.9. Acyclic Commands – IND500x Drum-Filling Application .................. 5-41
6 Frequently Asked Questions .......................................................... 6-1
6.1. What is the easiest way to get my SAI device communicating with
my automation system? ................................................................ 6-1
Contents
6.2. What are the Assembly Instance Values for each device covered in
this manual? ................................................................................ 6-2
6.3. How can I tell if my cyclic command executed successfully? .............. 6-2
6.4. Is it possible to read gross weight and net weight at the same time? ... 6-2
6.5. Bit 15 of my Measuring Block Response is High. What does this
mean? ......................................................................................... 6-3
6.6. I still don't quite understand how SAI works. What other resources
are available?............................................................................... 6-4
A. Modbus RTU/Modbus TCP Commands ........................................... A-1
2 METTLER TOLEDO SAI Reference Guide for Transmitters and Terminals 30587511 | 05 | 02/2023
1 Introduction
This manual is one of several available that explains the Standard Automation Interface (SAI). Each
manual covers SAI in a slightly different way explained in the chart below. The SAI manuals needed
for use with your device can be found on the downloads page of your SAI device at www.mt.com
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 1-1
2 Status Bits
2.1. Device Status Bits
The device status is a composite status word that contains individual bits to indicate the state of
various scale or device-specific binary values. The device status bits are always a part of the
Measuring Block so no matter the SAI block format in use (1 block, 2 block or 8 block), this
information is always available. The 16 bits include the following information:
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Device-specific
Bit Value Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-1
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
Status Bits
IND500x
ACT350
Device-specific
Bit Value Description
1 = Application fault; predictive diagnostics
alarm triggered or command cannot be
executed as requested.
5 Center of zero 1 = Gross weight value is at a value of zero x x x x x x x
+/- one quarter of a weight and measures
verification interval denoted as "e".
6 Motion 1 = Weight is unstable. x x x x x x x x
7 Net Mode 1 = Net weight instead of gross weight is x x x x x x x x
reported.
8 Alternate weight 1 = An alternate weight unit, other than the x
unit primary unit is in use.
9 Device-specific x
bit 1
10 Device-specific
bit 2
11 Device-specific
bit 3
12 Device-specific Refer to section 2.1.3, Device-Specific Bits, to
bit 4 see how different devices utilize these bits.
13 Device-specific
bit 5
14 Device-specific
bit 6
15 Device-specific
bit 7
• Device is powering up
o Device is booting
o Device is in process of conducting power up zero operation
o All operations the device needs to complete before a valid weight value are being
transmitted
2-2 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
• The device indicates an in-progress calibration and adjustment cycle
• While in setup mode no weight transmission is possible
o In ASM (Automatic Setup Mode on Precision Scales): while in the ASM menu, no other
scale information can be transmitted.
• Device is powering down
o Occurs as soon as the device receives shut down or restart signal
o Internal software modules are shut down and correct weight can no longer be guaranteed
Note: This is taken into account only if the device can detect power-down.
In automation indicators (terminals) the over and under capacity limits must be
defined in the configuration of the unit to match the safety requirements for the
device, the structure (or machine), the operators and the environment. The system
safety requirements must be evaluated on a case-by-case basis by qualified
personnel to ensure the structure supporting / holding the scale must be structurally robust (strong)
enough to carry the weight of the scale and any objects placed on/in it including a safety factor that
accounts for any and all worst-case overload scenarios.
When designing control algorithms, it is necessary to evaluate the vessel and scale
capacities independently as a best safety practice. In most cases, the scale capacity
exceeds the vessel’s capacity in weight. This means that scale capacity must never
be used as a method to prevent overfilling of the vessel, or any other object,
because the vessel will reach its maximum capacity limits before the scale capacity
limit. Instead, use an internal comparator that disables the filling mechanism / system at, or before,
the vessel’s maximum capacity limit in weight/volume, and add a compare function in the control
system logic that prevents overfilling of the object on the scale. These precautions must be taken to
avoid injury to personnel and equipment damage.
Note: in safety-relevant systems both Data Ok and Smart5 red must be constantly
evaluated by the control system to verify that the device is delivering valid
measurements. To ensure safe operation DO NOT REMOVE THESE TWO FUNCTIONS
from the sample code, or your programming code
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-3
Device-specific Bit ACT350POWERCELL IND500x
3 —
4 —
Status Bits
5 —
6 —
7 —
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Bit Red Alert Description
2-4 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Bit Red Alert Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-5
2.3. Secondary Scale Status
These status bits are sent by default as Status Group 2 when using either the SAI 2 block or 8 block
formats. This information is not available if using the SAI 1 block format. See section 3 of this
Status Bits
document for all status commands that return the Secondary Scale Status. The 16 bits include the
following information:
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Scale Status
Bit Group Description
0 Unit Bit 1 Unit bits are used to indicate the weight unit. x x x x x x x x
1 Unit Bit 2 Refer to "Unit Bits" table below for more
information.
2 Unit Bit 3
3 Unit Bit 4
4 MinWeigh Error 1 = Scale is below acceptable minimum
weighing range.
5 Range bit 1 Range bits are used to indicate weight x
6 Range bit 2 range or interval based on the values
shown. Refer to the Range Bits table,
below, for more information.
7 In Set Up 1 = Sensor is in setup mode. x x x x x x x x
8 Power Up Zero 1 = Scale has not been able to complete its x x x x x x x
Failure power-up restore / reset of zero.
9 GWP out of 1 = Scale has a GWP out of tolerance
Tolerance error.
10 Smart5 Level 4 Imminent failure according to Smart5 x x x x
definition
Note: Smart5 is not supported for ACT350
devices.
Note: This bit corresponds to "Selected Scale"
for IND500x instead of Smart5 Level 4
11 Smart5 Level 3 Out of specification according to Smart5 x x x
definition
Note: Smart5 is not supported for ACT350
devices.
12 Smart5 Level 2 Predictive Alarm according to Smart5 x x x
definition
Note: Smart5 is not supported for ACT350
devices.
13 LFT Switch On 1 = weights and measures switch is enabled x x x
for transactional weighing
2-6 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Scale Status
Bit Group Description
14 Open Unused
15 Open Unused
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-7
2.4. Alarm Status 2nd Group
These status bits are available when using either the SAI 2 block or 8 block formats. This
information is not available if using the SAI 1 block format. Please note that this group is not sent
Status Bits
by default. A status command that returns this group must be sent. See section 3 of this document
for all status commands that return the Alarm Status 2nd Group. The 16 bits include the following
information:
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Bit Soft Alarm Description
2-8 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Bit Soft Alarm Description
12 Application-defined 6
13 Application-defined 7
14 Application-defined 8
15 Application-defined 9
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-9
2.6.1. Load Cell Group 1
360
Bit Data
ACT
ACT
ACT
ACT
IND
IND
IND
IND
0 Error on attached device 1 x x
Status Bits
2-10 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
format. Please note that this group is not sent by default. A status command that returns this group
must be sent. See section 3 of this document for all status commands that return the Custom
Application Status Group 1. The 16 bits include the following information:
ACT350 POWERCELL
ACT350 DIO
application
ACT350
status bits,
Bit group 1
0 Application-defined 0 — New Data Available 1 = — — Run Run Run Run
triggered weight is available
1 Application-defined 1 — OK (ready for next object) — — Refill Refill Front PE Complete
Finished last weighing
process, can start next
process
2 Application-defined 2 — Up Scale — — Upper Alarm On Scale Pause
Weighing object is coming Limit Entirely
onto the scale
3 Application-defined 3 — On scale — — Lower — Rear PE Dump
Weighing object is on the Limit
scale
4 Application-defined 4 — Weighing start — — Alarm — Complete Refill
Start the weighing or
calibration process
5 Application-defined 5 — Weighing over End the — — — — Ready App Alarm
weighing or calibration
process
6 Application-defined 6 — OK (no error) — — — — Zero State Feed
No error in weighing process
7 Application-defined 7 — Long object — — — — Empty State Fast Feed
Object is too long and cannot
be weighed
8 Application-defined 8 — Short Distance — — — — Alarm —
Distance between objects is
too small and cannot be
weighed
9 Application-defined 9 — Photo Occlusion Front — — — — — Spill
Front light barrier is blocked.
The process should be
stopped and the light barrier
should be cleared
10 Application-defined 10 — Flag of update zero 1 = PLC — — — — — Refill Upper Limit
can send zero command to
clear the zero
11 Application-defined 11 — — — — — — — Refill Lower Limit
12 Application-defined 12 — — — — — — — Over +Tol
13 Application-defined 13 — — — — — — — Under –Tol
14 Application-defined 14 — — — — — — — Jog
15 Application-defined 15 — — — — — — — —
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-11
2.8.1.1. For IND500x
Bit Status Work Mode Work Mode Work Mode Work Mode Mode Mode
0 Custom Bit 1 Running Running Running Running Running Running
Complete:
1 Custom Bit 2 Complete: Fill Complete: Fill Complete: Fill Complete: Fill
Refill
2 Custom Bit 3 Holding Holding Holding Holding Holding Holding
3 Custom Bit 4 Dump Refill Lower Lance
Complete: Complete: Complete:
Custom Bit 5 Raise Lance
Dump Dose Dose
Alarm Alarm Alarm Alarm Alarm Alarm
5 Custom Bit 6
(application) (application) (application) (application) (application) (application)
6 Custom Bit 7 Feed / Jog Feed / Jog Feed / Jog Feed / Jog Feed / Jog Feed / Jog
7 Custom Bit 8 Fast Feed Fast Feed Fast Feed Fast Feed Fast Feed Fast Feed
Coarse Feed Coarse Feed Coarse Feed Coarse Feed Coarse Feed Coarse Feed
8 Custom Bit 9
(Reserved) (Reserved) (Reserved) (Reserved) (Reserved) (Reserved)
9 Custom Bit 10 Start Delay Start Delay Start Delay Start Delay Start Delay Start Delay
After Weigh After Weigh After Weigh After Weigh After Weigh After Weigh
10 Custom Bit 11
Delay Delay Delay Delay Delay Delay
Extend Drip
11 Custom Bit 12
Pan
Over + TOL Over + TOL Over + TOL Over + TOL Over + TOL Over + TOL
12 Custom Bit 13
(Over Zone) (Over Zone) (Over Zone) (Over Zone) (Over Zone) (Over Zone)
Under - TOL Under - TOL Under - TOL Under - TOL Under - TOL Under - TOL
13 Custom Bit 14
(Under Zone) (Under Zone) (Under Zone) (Under Zone) (Under Zone) (Under Zone)
14 Custom Bit 15 TOL OK TOL OK TOL OK TOL OK TOL OK TOL OK
15 Custom Bit 16 Ready Ready Ready Ready Ready Ready
2-12 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
2.8.2. Custom Application Status Group 2
These bits give information relating to the alarm status of an application within the device. For more
information on how to use the custom applications for IND360, please see the specific application
manual for IND360. These status bits are available when using either the SAI 2 block or 8 block
formats. This information is not available if using the SAI 1 block format. Please note that this
group is not sent by default. A status command that returns this group must be sent. Refer to
section 3 of this document for all status commands that return the Custom Application Status
Group 2. The 16 bits include the following information:
Custom
IND360 Tank
IND360 Rate
POWERCELL
ACT350 DIO
IND360 Fill
application
Precision
Dynamic
ACT350
ACT350
ACT350
IND360
Control
status bits,
Vessel
Dose
Bit group 2
0 Application- None None None None Lower Limit Flow Calibration Front PE Starting Weight
defined 0 Alarm Failed Timeout Invalid
1 Application- None None None None Upper Limit Control Rate Rear PE Autotare Fault
defined 1 Alarm Below Lower Timeout
Limit
10 Application- — — — — — — Parameter —
defined 10 Logic Error
11 Application- — — — — — — — —
defined 11
12 Application- — — — — — — — —
defined 12
13 Application- — — — — — — — —
defined 13
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-13
Custom
IND360 Tank
IND360 Rate
POWERCELL
ACT350 DIO
IND360 Fill
application
Precision
Dynamic
ACT350
ACT350
ACT350
IND360
Control
status bits,
Vessel
Dose
Bit group 2
Status Bits
14 Application- — — — — — — — —
defined 14
15 Application- — — — — — — — —
defined 15
Bit Custom Group 2- Alarm Status Advance Auto Filling Drum Filling
0 Custom Bit 1 Starting weight invalid Starting weight invalid
1 Custom Bit 2 Autotare Fault Autotare Fault
2 Custom Bit 3 Process Timeout Process Timeout
3 Custom Bit 4 Initial Feed Timeout Initial Feed Timeout
4 Custom Bit 5 Refill Timeout Reserved
5 Custom Bit 6 Dump Timeout Reserved
6 Custom Bit 7 Parameter Invalid Parameter Invalid
7 Custom Bit 8 Parameter Logic Error Parameter Logic Error
8 Custom Bit 9 Material not enough Reserved
9 Custom Bit 10
10 Custom Bit 11
11 Custom Bit 12
12 Custom Bit 13
13 Custom Bit 14
14 Custom Bit 15
15 Custom Bit 16
Devices which do not support physical I/O may have variables and logic to virtually represent
inputs and outputs within the device. If the device does not support I/O groups, an invalid
command response is sent for any unsupported I/O groups. These status bits are available when
using either the SAI 2 block or 8 block formats. This information is not available if using the SAI 1
block format. Please note that this group is sent by default as status group 3 if no other status
commands have been executed. See section 3 of this document for all status commands that
return the I/O Status Groups. The 16 bits include the following information:
2-14 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
The input and output status bits reflect the state of the associated inputs and outputs, 1 = on, 0 = off.
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 2-15
ACT350 ACT350 ACT350 IND360 IND360 IND360
Bit Data ACT350 DIO POWERCELL Precision Analog POWERCELL Precision IND500x
6 Comparator 7 x x x
x x x
Status Bits
7 Comparator 8
8 Comparator 9
9 Comparator 10
10 Comparator 11
11 Comparator 12
12 Comparator 13
13 Comparator 14
14 Comparator 15
15 Comparator 16
2-16 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
3 SAI Status Block Command
List
By default, Status Block Command = 0 when using SAI 2 block and 8 block formats. Status block
commands are not available when SAI 1 block format is used. If Status Block information other
than the default information is required, simply issue the command from the table below that
returns the necessary data. The most recently executed Status Block Command executed can be
confirmed by checking the Status Block Response Value. The response value will either match the
most recently executed command or will indicate an error executing the command if bit 15 of the
response is high.
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Value Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 3-1
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
SAI Status Block Command List
IND500x
ACT350
Value Description
Word 1 NULL x
22 Word 0 RedAlert
Word 1 Comparator 1 x
Word 2 NULL
3-2 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Value Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 3-3
4 Cyclic Command List for
Measuring Block
The chart below shows which floating-point cyclic commands are supported by specific devices. In
the case of IND360, some cyclic commands are available for specific applications, but not the
base unit. These commands have been separated into the tables found in Sections 4.1 to 4.4.
To issue a cyclic read command, set the Measuring block command equal to the command found
in the read column of the table below. Verify the command was executed successfully by
monitoring the response value of the Measuring block. If the command was executed successfully,
the response value = the command. The result of the read command will be available in the
Measuring block measuring value. If not executed successfully, bit 15 of the response value will be
high. A cyclic read command will continue to update the value being read until a different
Measuring block command is sent.
To issue a cyclic write command, enter the value to be written in the Measuring block command
value and then set the Measuring block command equal to the command found in the write
column of the table below. Verify the command was executed successfully by monitoring the
response value of the Measuring block. If the command was executed successfully, the response
value = the command value. If not executed successfully, bit 15 of the response value will be high.
A cyclic write command will execute one time.
Please note that a cyclic command cannot be executed twice in a row. If required, send a "no-op"
(command = 2000) command between the first and second instances of the command.
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-1
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
Cyclic Command List for Measuring Block
IND500x
ACT350
Read Write Description
4-2 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Read Write Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-3
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
Cyclic Command List for Measuring Block
IND500x
ACT350
Read Write Description
4-4 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
IND500x
ACT350
Read Write Description
N/A 1900 Alarm bit – write value of bit when in test mode x x x x x x x x
N/A 1901 Motion bit – write value of bit when in test mode x x x x x x x x
N/A 1902 Net mode bit – write value of bit when in test mode x x x x x x x x
N/A 1903 Center of zero bit – write value of bit when in test
x x x x x x x x
mode
N/A 1904 Alt weight bit – write value of bit when in test mode x x x x x x x x
N/A 1905 Device bit 1 – write value of bit when in test mode x x x x x x x x
N/A 1906 Device bit 2– write value of bit when in test mode x x x x x x x x
N/A 1907 Device bit 3– write value of bit when in test mode x x x x x x x x
N/A 1908 Device bit 4– write value of bit when in test mode x x x x x x x x
N/A 1909 Device bit 5– write value of bit when in test mode x x x x x x x x
N/A 1910 Device bit 6– write value of bit when in test mode x x x x x x x x
N/A 1911 Device bit 7– write value of bit when in test mode x x x x x x x x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-5
ACT350 POWERCELL
IND360 POWERCELL
ACT350 Precision
IND360 Precision
IND360 Analog
ACT350 DIO
Cyclic Command List for Measuring Block
IND500x
ACT350
Read Write Description
2006 N/A After step failure, skips step and advances to next in
x x x
sequence
IND360 POWERCELL
IND360 Precision
IND360 Analog
4-6 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read Write Description
111 311 Max. Weighing Time – Parameter: Min. Weighing Time – 10000 ms x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-7
IND360 POWERCELL
IND360 Precision
IND360 Analog
Cyclic Command List for Measuring Block
IND360 Precision
IND360 Analog
4-8 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read Write Description
IND360 Precision
IND360 Analog
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-9
IND360 POWERCELL
IND360 Precision
IND360 Analog
Cyclic Command List for Measuring Block
4-10 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read Write Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-11
than the IND360 Rate Control application. More details about the IND360 Rate Control
Application can be found in the IND360 Rate Control Application Manual.
IND360 POWERCELL
Cyclic Command List for Measuring Block
IND360 Precision
IND360 Analog
Read Write Description
101 301 Flow Target – Parameter: 0.1 – Rated Flow x
102 302 Flow Control Mode – Parameter: 0 = Flow Control Mode 1 = x
Fixed Frequency Control Mode 2 = Rapid Cal 3 = Step Cal
103 303 Refill Control Mode – Parameter: 0 = Fixed Frequency 1 = x
Follow Mode 2 = Level Switch Mode
104 304 Upper Limit – Parameter: 0 - Capacity x
105 305 Lower Limit – Parameter: 0 - Capacity x
106 306 Control Rate Upper Limit – Parameter: 0 - 1 x
107 307 Control Rate Lower Limit – Parameter: 0 - 1 x
108 308 Rated Flow – Parameter: 0.1 – 99,999 x
109 309 Control Filter – Parameter: 0 - 9 x
110 310 Flow Filter – Parameter: 0 - 9 x
111 311 Control Tolerance – Parameter: 0 - 1 x
112 312 Flow Tolerance – Parameter: 0 - 1 x
113 313 Test Time – Parameter: 0 – 9,999 Seconds x
114 314 Flow Stability Range – Parameter: 0 - 20 x
115 315 Flow Stability Time – Parameter: 0 - 20 x
116 316 Start Delay Time – Parameter: 0 – 9,999 x
117 317 Refill Delay Time – Parameter: 0 – 9,999 x
118 318 PID Mode – Parameter: 0 = Auto PID 1 = Manual PID x
119 319 Proportional Term (P) – Parameter: 0 – 255 x
120 320 Intergral Term (I) – Parameter: 0 – 255 x
121 321 Derivative Term (D) – Parameter: 0 – 255 x
122 322 Control Rate Factor – Parameter: 0 – 2 x
123 323 Refill Control Rate Factor – Parameter: 0 – 2 x
124 324 Refill Control Rate Target – Parameter: 0 - 1 x
125 325 Control Rate Target – Parameter: 0 - 1 x
126 326 Rapid Cal. Time – Parameter: 10 - 60 x
127 N/A 20% Flow Rate – Parameter: 0 – 99,999 x
128 N/A 40% Flow Rate – Parameter: 0 – 99,999
129 N/A 60% Flow Rate – Parameter: 0 – 99,999 x
130 N/A 80% Flow Rate – Parameter: 0 – 99,999 x
4-12 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read Write Description
131 N/A 100% Flow Rate – Parameter: 0 – 99,999 x
132 N/A Control Rate x
133 333 Refill Time – Parameter: 0 – 9,999 seconds x
134 N/A Flow x
136 N/A Cumulant x
137 N/A Work Cumulant x
138 N/A Test Cumulant x
N/A 339 Control Command – Parameter: 0 = Stop 1 = Run x
N/A 340 Clear Totalization – Parameter: 1 = Clear x
IND500x
Read Write Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-13
IND500x
Read Write Description
155 355 x
2 = % of Target
IND500x
Read Write Description
(-) Tolerance Weight, any weight below Target – Tolerance Weight will be
26 226 x
classified as underfilled – Parameter: 0 - Capacity
(+) Tolerance Weight, any weight above Target + Tolerance Weight will
27 227 x
be classified as overfilled – Parameter: 0 - Capacity
4-14 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND500x
Read Write Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-15
IND500x
Read Write Description
Cyclic Command List for Measuring Block
(-) Tolerance Weight, any weight below Target – Tolerance Weight will be classified
26 226 x
as underfilled – Parameter: – 0 - Capacity
(+) Tolerance Weight, any weight above Target + Tolerance Weight will be
27 227 x
classified as overfilled – Parameter: 0 - Capacity
(-) Tolerance %, any weight below (Target – (Tolerance% * Target / 100)) will be
28 228 x
classified as underfilled – Parameter: 0% – 100%
(+) Tolerance %, any weight above (Target + (Tolerance% * Target / 100)) will be
29 229 x
classified as overfilled – Parameter: 0% – 100%
4-16 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
4.8. Cyclic Commands – IND500x Over Under
Application
The cyclic commands listed below are for use with the IND500x Over/Under application. Please
note that these same commands will not necessarily be supported or may work differently if used
with anything other than the IND500x Over/Under application. Please see the section specific to the
application installed on your device if using something other than the IND500x Over/Under
application. More details about the application can be found in the IND500x Manual.
IND500x
Read Write Description
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 4-17
IND500x
Read Write Description
Cyclic Command List for Measuring Block
4-18 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
5 Acyclic Command List
The chart below shows acyclic commands that can be sent and which devices support specific
commands. In the case of IND360, some commands are supported only for versions of the device
with applications built-in and not the base versions. These application-specific acyclic commands
have been separated into the tables found in Sections 5.1 to 5.4.
To send an acyclic command, use the information for the command corresponding to your
automation protocol (EtherNet/IP, PROFIBUS DP, PROFINET, EtherCAT, CC-Link IE Field Basic). All
devices covered by this manual provide sample code and detailed engineering notes for each
automation protocol with examples of acyclic commands. Sample code and engineering notes can
be found on www.mt.com on your device's downloads page.
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-1
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
Acyclic Command List
ACT350
ACT350
IND360
Index
Gross weight - rounded Gross weight data in defined resolution read Float 32 1 0x14 0x300 0x01 0x01 0, 1 0x2000 0x1000 x x x x x x x x
Gross weight - rounded Gross weight data in defined resolution read Float 32 1 0x15 0x300 0x01 0x02 0, 1 0x2001 0x1002 x x x x x x x x
Tare weight - rounded Tare weight data in defined resolution read Float 32 1 0x16 0x300 0x01 0x03 0, 1 0x2002 0x1004 x x x x x x x x
Net weight - rounded Net weight data in defined resolution read Float 32 1 0x17 0x300 0x01 0x04 0, 1 0x2003 0x1006 x x x x x x x x
Gross weight - internal resolution Gross weight data in internal resolution read Float 32 1 0x18 0x300 0x01 0x05 0, 1 0x2004 0x1008 x x x x x x x x
Tare weight - internal resolution Tare weight data in internal resolution read Float 32 1 0x19 0x300 0x01 0x06 0, 1 0x2005 0x100A x x x x x x x x
Net weight - internal resolution Net weight data in internal resolution read Float 32 1 0x1A 0x300 0x01 0x07 0, 1 0x2006 0x100C x x x x x x x x
Zero adjustment count (Read) Zero Register read/ write Float 32 1 0x20 300 0x01 12 0, 1 0x2007 0x702E x x
Tare procedure status bits Report Tare operation status (used when triggering tare from acyclic interface) –
read UInt 16 1 0x1F 0x300 0x01 0x16 0, 1 0x2008 0x2003 x x x x x x x
Parameter: 0 = tare procedure complete 1 = tare procedure in process
Zero procedure status bits Report Zero operation status (used when triggering zero from acyclic interface) –
read Uint 16 1 0x24 0x300 0x01 0x17 0, 1 0x2009 0x2005 x x x x x x x
Parameter: 0 = zero procedure complete 1 = zero procedure in process
Weight Unit Weight unit (number representing unit from Scale Status Group 2) read/ write Byte, 1 1 0x99 0x300 0x01 0x18 0, 1 0x200A 0x100E x x x x x x x
Tare when stable Tare when within motion limit write Uint 8 1 0x1C 0x300 0x01 0x09 0, 1 0x2010 0x2002 x x x x x x x
Tare immediately Motion not checked, tare executed write Uint 8 1 0x1E 0x300 0x01 0x10 0, 1 0x2011 0x2007 x x x x x x x
Clear tare Motion not checked, clear tare executed write Uint 8 1 0x1D 0x300 0x01 0x11 0, 1 0x2012 0x2006 x x x x x x x x
Zero when stable Zero when within motion limit write Uint 8 1 0x22 0x300 0x01 0x14 0, 1 0x2013 0x2004 x x x x x x x x
Zero immediately Motion not checked, zero executed write Uint 8 1 0x23 0x300 0x01 0x15 0, 1 0x2014 0x2008 x x x x x x x
Preset tare (display unit) Write tare register (Preset Tare) – Parameter: Pre-tare weight (float 32) write Float 32 1 0x1B 0x300 0x01 0x08 0, 1 0x2020 0x2000 x x x x x x x x
Clear Alarm Clear Alarm write Byte, 1 1 0x25 301 0x01 01 0, 1 0x2030 N/A x
Turn all internal & external outputs Forces all outputs OFF
write Byte, 1 1 0x26 0x301 0x01 0x02 0, 1 0x2031 0x9000 x x x x x x x x
OFF
Report scale status group Scale status group according to specification in Section 2.1 of this document read Short, 2 1 0x27 0x302 0x01 0x01 0, 1 0x2040 0xA010 x x x x x x x
Alarm status group Application Specific Errors according to specification in Section 2.4 of this
read Short, 2 1 0x28 0x302 0x01 0x02 0, 1 0x2041 0xA011 x x x x x x x
document
Report RedAlert group RedAlert status according to specification in Section 2.2 of this document read Short, 2 1 0x29 0x302 0x01 0x03 0, 1 0x2042 0xA012 x x x x x x x
Report scale status group Scale Status Group 2 according to specification in Section 2.3 of this document read Short, 2 1 0x2A 0x302 0x01 0x04 0, 1 0x2043 0xA013 x x x x x x x
Model type part 1 Identification (main ID) read String 160 1 0x2B 0x303 0x01 0x01 0, 1 0x2050 N/A x x x x x
Model type part 2 Identification # 2 read String 160 1 0x2C 0x303 0x01 0x02 0, 1 0x2051 N/A x x x x x
Model type part 3 Identification # 3 read String 160 1 0x2D 0x303 0x01 0x03 0, 1 0x2052 N/A x x x x x
Software OS version Software OS Version read String (36 Byte) 1 0x2E 0x303 0x01 0x04 0, 1 0x2053 0x3000 x x x x x x x x
Fieldbus stack version Fieldbus Stack version read String (36 Byte) 1 0x2F 0x303 0x01 0x05 0, 1 0x2054 0x3012 x x x x x x x x
5-2 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
ACT350
ACT350
IND360
Index
Software application version Software Apps Version read String (36 Byte) 1 0x30 0x303 0x01 0x06 0, 1 0x2055 0x3024 x x x x x
SAI version SAI specification version number read String (36 Byte) 1 0x31 0x303 0x01 0x07 0, 1 0x2056 0x3036 x x x x x x x x
Serial number Device main serial number read String (36 Byte) 1 0x33 0x303 0x01 0x08 0, 1 0x2057 0x3048 x x x x
Device identification user configurable ID read String (36 Byte) 1 0x34 0x303 0x01 0x09 0, 1 0x2058 0x305A x x
Start adjustment with external weight Start External Adjustment write Byte, 1 1 0x81 0x410 0x01 0x02 0, 1 0x4002 0x7001 x x
Start customer standard calibration Start User Standard Adjustment write Byte, 1 1 0x82 0x410 0x01 0x03 0, 1 0x4003 0x7002 x x
Start test with internal weight Start Test Function with internal weight write Byte, 1 1 0x84 0x410 0x01 0x05 0, 1 0x4005 0x7004 x x
Start test with external weight Start Test Function with external weight write Byte, 1 1 0x85 0x410 0x01 0x06 0, 1 0x4006 0x7005 x x
Test deviation Adjustment / Test value read Float 32 1 0x87 0x410 0x01 0x08 0, 1 0x4008 0x7007 x x
External adjustment weight Set external adjustment weight and start adjust – Parameter: Weight in unit
write Float 32 1 0x88 0x410 0x01 0x09 0, 1 0x4009 0x7009 x x x x x x x
selected according to Scale Status Group 2
Zero Adjustment Resets the zero (absolute reference) write Byte, 1 1 0x8A 0x410 0x01 0x0B 0, 1 0x400B 0x700C x x x x x
Span Adjustment Value 1 xLow Used in 5 point linearity adjustment read/ write Float 32 1 0x8C 0x410 0x01 0x0C 0, 1 0x400C 0x700D x x x x x
Span Adjustment Value 2 Low Used in 5 point and 4 point linearity adjustment read/ write Float 32 1 0x8D 0x410 0x01 0x0D 0, 1 0x400D 0x700F x x x x x
Span Adjustment Value 3 Middle Used in 5, 4 and 3 point linearity adjustment read/ write Float 32 1 0x8E 0x410 0x01 0x0E 0, 1 0x400E 0x7011 x x x x x
Span Adjustment Value 4 High Used in all forms of span adjustment read/ write Float 32 1 0x8F 0x410 0x01 0x0F 0, 1 0x400F 0x7013 x x x x x
Validate (Confirm) Adjustment Validate Adjustment read/ write Byte, 1 1 0x90 0x410 0x01 0x10 0, 1 0x4010 0x7015 x x x x x
Requested weight Get currently requested external calibration weight during ongoing adjustment or
read Float 32 1 0x91 0x410 0x01 0x11 0, 1 0x4011 0x7016 x x
calibration procedure
External test weight Set external calibration test weight unless default shall be used – Parameter:
write Float 32 1 0x92 0x410 0x01 0x12 0, 1 0x4012 0x7018 x x
Weight in unit selected according to Scale Status Group 2
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-3
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
Acyclic Command List
ACT350
ACT350
IND360
Index
Span Adjustment Counts 1 xLow Used with Span adjustment value to read/write calibration read/ write long, 4 1 0x93 0x410 0x01 0x13 0, 1 0x4013 0x701A x x x x x
Span Adjustment Counts 2 Low Used with Span adjustment value to read/write calibration read/ write long, 4 1 0x94 0x410 0x01 0x14 0, 1 0x4014 0x701C x x x x x
Span Adjustment Counts 3 Middle Used with Span adjustment value to read/write calibration read/ write long, 4 1 0x95 0x410 0x01 0x15 0, 1 0x4015 0x701E x x x x x
Span Adjustment Counts 4 High Used with Span adjustment value to read/write calibration read/ write long, 4 1 0x96 0x410 0x01 0x16 0, 1 0x4016 0x7020 x x x x x
Set number of steps & begin step Set number of steps and begin step calibration
write Byte, 1 1 0x97 0x410 0x01 0x17 0, 1 0x4017 0x7022 x x x x x
calibration
Sets weight value for current step in Set weight value for current step in calibration and starts step
write Float 32 1 0x98 0x410 0x01 0x18 0, 1 0x4018 0x7023 x x x x x
calibration & starts step
Calfree Triggers CalFree calibration start write Byte, 2 1 0x9B 0x410 0x01 0x1A 0, 1 0x401A 0x7025 x x x
Calfree cell capacity Parameter for CalFree read/ write Float 32 1 0x9C 0x410 0x01 0x1B 0, 1 0x401B 0x7026 x x x
Calfree unit Parameter for CalFree read/ write Byte, 1 1 0x9D 0x410 0x01 0x1C 0, 1 0x401C 0x7028 x x x
Calfree cell output Parameter for CalFree read/ write Float 32 1 0x9E 0x410 0x01 0x1D 0, 1 0x401D 0x7029 x x x
CalFree Plus Trigger CalFree Plus calibration start read/ write Byte, 2 1 0x9F 0x410 0x01 0x1E 0, 1 0x401E 0x702B x x
Get temporary weight in step mode Set weight value for current step in calibration and starts step read float, 4 1 0x8B 0x410 0x01 0x1F 0, 1 0x401F N/A x x x
Enable / Disable step control C5 Read/ write Byte, 1 1 0x8B 0x410 0x01 0x1F 0, 1 0x401F 0x7030 x x
Zero adjustment count Resets the zero (absolute reference) read/ write long, 4 1 0xE5 0x410 0x01 0x20 0, 1 0x4020 N/A x x x
Internal adjustment PBK/PFK Internal adjustment (MTCS C9, 1) write Byte, 1 3 0x03 0x410 0x01 0x22 0, 1 0x4022 0x7034 x
Get temporary weight in step mode Get temporary weight in step mode Read Float 32 3 0x04 0x410 0x01 0x23 0, 1 0x4023 0x702C x x
Comparator status group 1 Comparator status group 1 read short, 2 1 0xD0 0x411 0x01 0x01 0, 1 0x4051 0xA014 x x x x x x x
Comparator Status Group 2 Comparator Status Group 2 Read Short, 2 2 0x01 0x411 0x01 0x02 0, 1 0x4052 N/A
Report # of Comparator used Read how many comparators are used read Byte, 1 2 0x03 0x411 0x01 0x04 0, 1 0x4054 N/A x x x x
Write # of Comparator used Write how many comparators are used write Byte, 1 2 0x03 0x411 0x01 0x04 0, 1 0x4054 N/A x x x x
Report/Write Comparator 1 Limit Read/Write value for comparator # 1 read/ write Float 32 2 0x04 0x411 0x01 0x05 0, 1 0x4055 0x8000 x x x x x x x x
Report/Write Comparator 2 Limit Read/Write value for comparator # 2 read/ write Float 32 2 0x05 0x411 0x01 0x06 0, 1 0x4056 0x8002 x x x x x x x x
Report/Write Comparator 3 Limit Read/Write value for comparator # 3 read/ write Float 32 2 0x06 0x411 0x01 0x07 0, 1 0x4057 0x8004 x x x x x x x x
Report/Write Comparator 4 Limit Read/Write value for comparator # 4 read/ write Float 32 2 0x07 0x411 0x01 0x08 0, 1 0x4058 0x8006 x x x x x x x x
Report/Write Comparator 5 Limit Read/Write value for comparator # 5 read/ write Float 32 2 0x08 0x411 0x01 0x09 0, 1 0x4059 0x8008 x x x x x x x x
Report/Write Comparator 6 Limit Read/Write value for comparator # 6 read/ write Float 32 2 0x09 0x411 0x01 0x0A 0, 1 0x405A 0x800A x x x
5-4 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
ACT350
ACT350
IND360
Index
Report/Write Comparator 7 Limit Read/Write value for comparator # 7 read/ write Float 32 2 0x0A 0x411 0x01 0x0B 0, 1 0x405B 0x800C x x x
Report/Write Comparator 8 Limit Read/Write value for comparator # 8 read/ write Float 32 2 0x0B 0x411 0x01 0x0C 0, 1 0x405C 0x800E x x x
Apply Comparator trigger Instructs device to use new comparator values write Byte, 1 2 0x1E 0x411 0x01 0x1F 0, 1 0x406F 0x8060 x x x x x x x
Target – Use Gross Weight Use gross weight as source for target comparison write Short 4 0x08 0x411 1 0x28 0, 1 0x4078 x
Target – Use Net Weight Use net weight as source for target comparison Write Short 4 0x09 0x411 1 0x29 0, 1 0x4079 x
Read / Write Comparator 1 Source Comparator 1 source configuration read/ write Byte, 1 4 0x10 0x411 1 0x30 0, 1 0x4080 0x8040 x x x
Read / Write Comparator 2 Source Comparator 2 source configuration read/ write Byte, 1 4 0x11 0x411 1 0x31 0, 1 0x4081 0x8041 x x x
Read / Write Comparator 3 Source Comparator 3 source configuration read/ write Byte, 1 4 0x12 0x411 1 0x32 0, 1 0x4082 0x8042 x x x
Read / Write Comparator 4 Source Comparator 4 source configuration read/ write Byte, 1 4 0x13 0x411 1 0x33 0, 1 0x4083 0x8043 x x x
Read / Write Comparator 5 Source Comparator 5 source configuration read/ write Byte, 1 4 0x14 0x411 1 0x34 0, 1 0x4084 0x8044 x x x
Read / Write Comparator 6 Source Comparator 6 source configuration read/ write Byte, 1 4 0x15 0x411 1 0x35 0, 1 0x4085 0x8045 x x x
Read / Write Comparator 7 Source Comparator 7 source configuration read/ write Byte, 1 4 0x16 0x411 1 0x36 0, 1 0x4086 0x8046 x x x
Read / Write Comparator 8 Source Comparator 8 source configuration read/ write Byte, 1 4 0x17 0x411 1 0x37 0, 1 0x4087 0x8047 x x x
Read / Write Comparator 1 Active Comparator 1 active read/ write Byte, 1 4 0x20 0x411 1 0x40 0, 1 0x4090 0x8050 x x x
Read / Write Comparator 2 Active Comparator 2 active read/ write Byte, 1 4 0x21 0x411 1 0x41 0, 1 0x4091 0x8051 x x x
Read / Write Comparator 3 Active Comparator 3 active read/ write Byte, 1 4 0x22 0x411 1 0x42 0, 1 0x4092 0x8052 x x x
Read / Write Comparator 4 Active Comparator 4 active read/ write Byte, 1 4 0x23 0x411 1 0x43 0, 1 0x4093 0x8053 x x x
Read / Write Comparator 5 Active Comparator 5 active read/ write Byte, 1 4 0x24 0x411 0x01 0x44 0, 1 0x4094 0x8054 x x x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-5
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
Acyclic Command List
ACT350
ACT350
IND360
Index
Read / Write Comparator 6 Active Comparator 6 active read/ write Byte, 1 4 0x25 0x411 0x01 0x45 0, 1 0x4095 0x8055 x x x
Read / Write Comparator 7 Active Comparator 7 active read/ write Byte, 1 4 0x26 0x411 0x01 0x46 0, 1 0x4096 0x8056 x x x
Read / Write Comparator 8 Active Comparator 8 active read/ write Byte, 1 4 0x27 0x411 0x01 0x47 0, 1 0x4097 0x8057 x x x
Report custom group 1 Custom group 1 status Read Short, 2 2 0x71 0x412 0x01 0x02 0, 1 0x4102 0xA015 x x x
Report custom group 2 Custom group 2 status read Short, 2 2 0x72 0x412 0x01 0x03 0, 1 0x4103 0xA016 x
Voltage monitor channels View voltage monitor channels read Struct 256 1 0xB0 0x413 0x01 0x11 0, 1 0x4161 N/A
Load cycle monitor channels view load cycle monitor channels read Struct 512 1 0xB1 0x413 0x01 0x12 0, 1 0x4162 N/A
Zero deviation Query zero deviation read Float 32 1 0xB2 0x413 0x01 0x13 0, 1 0x4163 N/A
Zero deviation monitor channels view zero deviation monitor channels read Struct 256 1 0xB3 0x413 0x01 0x14 0, 1 0x4164 N/A
Temperature monitor channels view temperature monitor channels read Struct 512 1 0xB4 0x413 0x01 0x15 0, 1 0x4165 N/A
Temperature gradient Query temperature gradient read Struct 96 1 0xB5 0x413 0x01 0x16 0, 1 0x4166 N/A
Temperature gradient channels view temperature gradient channels read Struct 128 1 0xB6 0x413 0x01 0x17 0, 1 0x4167 N/A
Temperature values Query temperature value (multiple channels) read Struct 128 1 0xB7 0x413 0x01 0x18 0, 1 0x4168 N/A x
Internal temperature Query the load cell temperature value read Float 32 1 0xB8 0x413 0x01 0x19 0, 1 0x4169 N/A
Restart device restart device - software restart write Uint 8 1 0xC9 0x413 0x01 0x2A 0, 1 0x417A N/A x
Update CANMaster power diagnosis Send 1 command to update the voltage and current of CANMaster write Byte, 1 1 0xCE 0x413 0x01 0x2E 0, 1 0x417E N/A x
Maximum supply power for LCs Inquire maximum supply voltage for LCs in history in mV read unsigned short, 2 1 0xCF 0x413 0x01 0x2F 0, 1 0x417F N/A x
Maximum supply current for LCs Inquire maximum supply voltage for LCs in history in mA read unsigned short, 2 1 0xEC 0x413 0x01 0x30 0, 1 0x4180 N/A x
Supply power error counts Inquire supply power error counts. Once over current has occurred, the error
read unsigned short, 2 1 0xD1 0x413 0x01 0x31 0, 1 0x4181 N/A x
counts would increase one. Int type, range 0~65535
Supply current error counts Inquire supply current error counts. Once over current has occurred, the error
read unsigned short, 2 1 0xD2 0x413 0x01 0x32 0, 1 0x4182 N/A x
counts would increase one. Int type, range 0~65535
Maximum voltage of CANH Inquire maximum voltage of CANH in mv read short, 2 1 0xD3 0x413 0x01 0x33 0, 1 0x4183 N/A x
Minimum voltage of CANH Inquire minimum voltage of CANH in mv read short, 2 1 0xD4 0x413 0x01 0x34 0, 1 0x4184 N/A x
Maximum voltage of CANL Inquire maximum voltage of CANL in mv read short, 2 1 0xD5 0x413 0x01 0x35 0, 1 0x4185 N/A x
Minimum voltage of CANL Inquire minimum voltage of CANL in mv read short, 2 1 0xD6 0x413 0x01 0x36 0, 1 0x4186 N/A x
Current supply power for LCs Inquire the current supply power for LCs read short, 2 1 0xD7 0x413 0x01 0x37 0, 1 0x4187 N/A x
Current supply current for LCs Inquire the current supply current for LCs read short, 2 1 0xD8 0x413 0x01 0x38 0, 1 0x4188 N/A x
5-6 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
ACT350
ACT350
IND360
Index
Update LCs voltage diagnosis Send 1 to update the current VIN_LC,VIN_ COM,V_ SHIELD voltage of LCs, every
LC need write Byte,1 1 0xD9 0x413 0x01 0x39 0, 1 0x4189 N/A x
one second to update data
VIN_LC of LCs Inquire the current VIN_LC voltage of LCs in mV read long*1 4,56 1 0xDA 0x413 0x01 0x3A 0, 1 0x418A N/A x
Temperature of LCs Inquire the current temperature of every LC in ℃ read long*1 4,56 1 0xDB 0x413 0x01 0x3B 0, 1 0x418B N/A x
VIN_COM of LCs Inquire the current VIN_COM voltage of every LC in mV read long*1 4,56 1 0xDC 0x413 0x01 0x3C 0, 1 0x418C N/A x
V_SHIELD of LCs Inquire the current V_SHIELD voltage of every LC in mV read long*1 4,56 1 0xDD 0x413 0x01 0x3D 0, 1 0x418D N/A x
Update LCs gas diagnosis Send 1 to update the current gas sensor value of LCs, every LC need maximum
write Byte, 1 1 0xDE 0x413 0x01 0x3E 0, 1 0x418E N/A x
six seconds to update data
Gas concentration of LCs Inquire the current gas sensor value of LCs as percentage read long*1 4,56 1 0xDF 0x413 0x01 0x3F 0, 1 0x418F N/A x
Update LCs information Send 1 to update the current capacity, unit & sw version of LCs. Every LC needs
write Byte, 1 1 0xE0 0x413 0x01 0x40 0, 1 0x4190 N/A x
one second to update data
Communication error counts of LCs Inquire the current communication error counts of every LC. If normal at first then
LC communication error occurs, the error counts would increase one. Int type, read long*1 4,56 1 0xE1 0x413 0x01 0x41 0, 1 0x4191 N/A x
range 0~65535
Overload normal range counts of LCs Inquire the current overload normal range counts of every LC. If weight is normal
at first then overload between 101% ~ 150% of LC normal capacity occurs, the read long*1 4,56 1 0xE2 0x413 0x01 0x42 0, 1 0x4192 N/A x
error counts would increase one. Int type, range 0~65535
Overload operate range counts of LCs Inquire the current overload operate range counts of every LC. If weight is normal
at first then overload larger than 150% of LC normal capacity occurs, the error read long*1 4,56 1 0xE3 0x413 0x01 0x43 0, 1 0x4193 N/A x
counts would increase one. Int type, range 0~65535
Temperature beyond normal range Inquire the current temperature beyond normal range counts of every LC. Once first
counts of LCs normal then temperature beyond LC normal range happened, the error counts read long*1 4,56 1 0xE4 0x413 0x01 0x44 0, 1 0x4194 N/A x
would increase one. Int type, range 0~65535
Temperature beyond operate range Inquire the current temperature beyond operate range counts of every LC. If normal
counts of LCs at first then temperature beyond LC operate range occurs, the error counts would read long*1 4,56 1 0xE5 0x413 0x01 0x45 0, 1 0x4195 N/A x
increase one. Int type, range 0~65535
Temperature beyond operate range Inquire the current temperature beyond operate range counts of every LC after
counts of LCs after temperature temperature RunFlat trigger. If normal at first then temperature beyond LC operate
read long*1 4,56 1 0xE6 0x413 0x01 0x46 0, 1 0x4196 N/A x
RunFlat trigger range occurs, the error counts would increase one. Int type, range 0~65535. The
operate ranges differ by LC type (index 736)
PLC communication failure count Cable / PLC / Device failure. Each time fieldbus loses connection, count increases
read unsigned short, 2 1 0xE7 0x413 0x01 0x47 0, 1 0x4197 N/A x
by one
Scale overload count Scale overload count read unsigned short, 4 1 0xE8 0x413 0x01 0x48 0, 1 0x4198 N/A x
Scale calibration count Scale calibration count read unsigned short, 4 1 0xE9 0x413 0x01 0x49 0, 1 0x4199 N/A x
Scale zero command count Scale zero command count read unsigned short, 4 1 0xEA 0x413 0x01 0x4A 0, 1 0x419A N/A x
Scale zero command failed count Scale zero command failed count read unsigned short, 4 1 0xEB 0x413 0x01 0x4B 0, 1 0x419B N/A x
Automatic prefilling Automatic prefilling configuration read/write Struct 32 1 0xEF 0x414 0x01 0x01 0, 1 0x4201 N/A
Material filling duration Material filling duration configuration read/write Uint 16 1 0xF0 0x414 0x01 0x02 0, 1 0x4202 N/A
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-7
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
Acyclic Command List
ACT350
ACT350
IND360
Index
Automatic refilling Automatic refilling configuration read/write Uint 8 1 0xF1 0x414 0x01 0x03 0, 1 0x4203 N/A
Target weight Target weight configuration read/ write Struct 128 1 0xF2 0x414 0x01 0x04 0, 1 0x4204 N/A
Optimization function Optimization function configuration read/ write Struct 24 1 0xF3 0x414 0x01 0x05 0, 1 0x4205 N/A
Weight monitor function Weight monitor function configuration read/ write Struct 480 1 0xF4 0x414 0x01 0x06 0, 1 0x4206 N/A
Time monitor function Time monitor function configuration read/ write Struct 320 1 0xF5 0x414 0x01 0x07 0, 1 0x4207 N/A
Filter stability criteria Filling stability criteria configuration read/ write Struct 64 1 0xF6 0x414 0x01 0x08 0, 1 0x4208 N/A
Filing phase Filling phase configuration read/ write Struct 480 1 0xF7 0x414 0x01 0x09 0, 1 0x4209 N/A
Automatic tare Automatic tare configuration read/ write Struct 192 1 0xF8 0x414 0x01 0x0A 0, 1 0x420A N/A
Digital output function Digital output function configuration read/ write Struct 80 1 0xF9 0x414 0x01 0x0B 0, 1 0x420B N/A
Emptying function Emptying function configuration read/ write Struct 48 1 0xFA 0x414 0x01 0x0C 0, 1 0x420C N/A
Filling statistics Filling statistics read Struct 224 1 0xFB 0x414 0x01 0x0D 0, 1 0x420D N/A
Clear filling statistics Clear filling statistics write Uint 8 1 0xFC 0x414 0x01 0x0E 0, 1 0x420E N/A
Filling application status Filling application status read Uint 16 1 0xFD 0x414 0x01 0x0F 0, 1 0x420F N/A
Report filling status Report filling state read Uint 8 1 0xFE 0x414 0x01 0x10 0, 1 0x4210 N/A
Control filling status control filling write Uint 8 1 0xFF 0x414 0x01 0x11 0, 1 0x4211 N/A
Report/Write low pass poles Report/Write low pass poles Read/write Float 32 1 0x3D 0x415 1 0x02 0, 1 0x4252 N/A
Cut-off frequency for alternate weight Configure / Read Cut-off frequency for alternate weight path
read/ write Float 32 1 0x69 0x415 0x01 0x0B 0, 1 0x4259 N/A
path
Read/Write customer defined Set threshold about the customer overload in divisions
read/ write Short, 2 1 0x6A 0x415 0x01 0x0C 0, 1 0x425C 0xA000 x x x
overload threshold
5-8 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
ACT350
ACT350
IND360
Index
Read/Write customer defined Set threshold about the customer underload in divisions
read/ write Short, 2 1 0x6B 0x415 0x01 0x0D 0, 1 0x425D 0xA001 x x x
underload threshold
Disable weight display 1 = weight display disabled, 0 = weight display enabled read/ write Binary 1 0x71 0x416 0x01 0x02 0, 1 0x4302 N/A
Report LFT State Shows the status of the weights and measures switch position (LFT Y/N) read Binary 1 0x73 0x416 0x01 0x04 0, 1 0x4304 N/A
Display - Energy Saving Mode Time value for the display to turn off "Green MT feature" 0x4002
read/ write Float 32 1 0x78 0x416 0x01 0x09 0, 1 0x4309 x x x x x x
Change Display Resolution (M110) value from -6 to 6 which corresponds with different display resolutions read/ write Uint 8 1 0x7B 0x416 0x01 0x0C 0, 1 0x430C 0x500E x x
Signal output frequency settings reads/sets output frequency signal read/ write Float 32 1 0x7C 0x416 0x01 0x0D 0, 1 0x430D N/A
Reverse weighing mode sets weighing mode for loss in weight read/ write Binary 1 0x7D 0x416 0x01 0x0E 0, 1 0x430E N/A
Disable Keypad Disable keyboard of device to avoid accidental key interactions read/ write Byte, 1 1 0x77 0x416 1 0x08 0, 1 0x4308 0x4005 x x x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-9
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
Acyclic Command List
ACT350
ACT350
IND360
Index
Smallest calculated approvable Smallest calculated approvable interval value
read/ write Float 32 1 0x5A 0x417 0x01 0x15 0, 1 0x4365 N/A
interval value
d, increment Smallest available digit read/ write Float 32 1 0x5B 0x417 0x01 0x16 0, 1 0x4366 0x500F x x x x x
Nmax (Maximal capacity) Scale/Sensor capacity read/ write Float 32 1 0x5C 0x417 0x01 0x17 0, 1 0x4367 0x5011 x x x x x x
Automatic zero tracking Enable / Disable auto zero functionParameter: 0 = Disabled 1 = Enabled read/ write Byte, 1 1 0x63 0x417 0x01 0x1E 0, 1 0x436E 0x5013 x x x x
Report Raw load cell counts Filtered load cell value in counts Read Unsigned Long 1 0x66 0x417 0x01 0x21 0, 1 0x4371 0x1050 x x
Report gross weight from each LC Reports gross weight value from each individual load cell. Placed into 15 element
array of floating point values. The first element is the total gross weight for the
scale. Individual weight values follow in subsequent elements of the array. If not Read Struct(Float 32*15) 2 0xB0 0x417 0x01 0x24 0, 1 0x4374 0x1010 x x
all possible load cells are used, a value of 0 will be reported in all unused array
elements.
Report net weight from each LC Reports net weight value from each individual load cell. Placed into 15 element
array of floating point values. The first element is the total net weight for the scale.
Individual weight values follow in subsequent elements of the array. If not all Read Struct(Float 32*15) 2 0xB1 0x417 0x01 0x25 0, 1 0x4375 0x1030 x x
possible load cells are used, a value of 0 will be reported in all unused array
elements.
Report/Write #1 Input Polarity Determines the polarity of the input at setup. Read/ write Byte, 1 2 0x10 0x418 0x01 0x01 0, 1 0x4401 N/A x x x
Report/Write #1 Input Assignment Application dependent ex. 0=None, 1=Clear Tare, 2 = Tare, 3 = zero 0x9002
Read/ write Byte, 1 2 0x11 0x418 0x01 0x02 0, 1 0x4402 x x x x x x
Report/Write #1 trig mode (edge) Input is triggered on the leading (1) or lagging edge(0) 0x9001
Read/ write Byte, 1 2 0x12 0x418 0x01 0x03 0, 1 0x4403 x x x
Report/Write #2 Input Polarity Determines the polarity of the input at setup. Read/ write Byte, 1 2 0x13 0x418 0x01 0x04 0, 1 0x4404 N/A x x x
Report/Write #2 Input Assignment Application dependent ex. 0=None, 1=Clear Tare, 2 = Tare, 3 = zero 0x9004
Read/ write Byte, 1 2 0x14 0x418 0x01 0x05 0, 1 0x4405 x x x x x x
Report/Write #2 trig mode (edge) Input is triggered on the leading (1) or lagging edge(0) 0x9003
Read/ write Byte, 1 2 0x15 0x418 0x01 0x06 0, 1 0x4406 x x x
Report/Write #3 Input Polarity Determines the polarity of the input at setup. Read/ write Byte, 1 2 0x16 0x418 0x01 0x07 0, 1 0x4407 N/A x x x
Report/Write #3 Input Assignment Application dependent ex. 0=None, 1=Clear Tare, 2 = Tare, 3 = zero 0x9006
Read/ write Byte, 1 2 0x17 0x418 0x01 0x08 0, 1 0x4408 x x x x x x
Report/Write #3 trig mode (edge) Input is triggered on the leading (1) or lagging edge(0) 0x9005
Read/ write Byte, 1 2 0x18 0x418 0x01 0x09 0, 1 0x4409 x x x
Report/Write Output Polarity Read/Write Output polarity Read/ write Byte, 1 2 0x1C 0x418 0x01 0x0D 0, 1 0x440D N/A x x x
Report/Write #1 Output Assignment Value based on application: 0, 1, 2, 3, 4, 5,etc. Read/ write Byte, 1 2 0x1D 0x418 0x01 0x0E 0, 1 0x440E 0x9020 x x x x x x
Report/Write #2 Output Assignment Value based on application: 0, 1, 2, 3, 4, 5,etc. Read/ write Byte, 1 2 0x24 0x418 0x01 0x15 0, 1 0x4415 0x9021 x x x x x x
Report/Write #3 Output Assignment Value based on application: 0, 1, 2, 3, 4, 5,etc. Read/ write Byte, 1 2 0x2B 0x418 0x01 0x1C 0, 1 0x441C 0x9022 x x x x x x
5-10 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
ACT350
ACT350
IND360
Index
Report/Write #4 Output Assignment Value based on application: 0, 1, 2, 3, 4, 5,etc. Read/ write Byte, 1 2 0x32 0x418 0x01 0x23 0, 1 0x4423 0x9023 x x x x x x
Report/Write #5 Output Assignment Value based on application: 0, 1, 2, 3, 4, 5,etc. Read/ write Byte, 1 2 0x39 0x418 0x01 0x2A 0, 1 0x442A 0x9024 x x x x x x
Last dynamic weight value Last dynamic weight value Read Float 32 2 0x90 0x41A 0x01 0x01 0, 1 0x4501 N/A x
Maximum dynamic weight value Max dynamic weight Read Float 32 2 0x93 0x41A 0x01 0x04 0, 1 0x4504 N/A x
Minimum dynamic weight value Min dynamic weight Read Float 32 2 0x94 0x41A 0x01 0x05 0, 1 0x4505 N/A x
Mean dynamic weight value Average dynamic weight Read Float 32 2 0x95 0x41A 0x01 0x06 0, 1 0x4506 N/A x
I/O Group 1 Read current state of I/O Group 1 Read 2 0x3F 0x418 0x01 0x30 0, 1 0x4430 N/A
I/O Group 2 Read current state of I/O Group 2 Read 2 0x40 0x418 0x01 0x31 0, 1 0x4431 N/A
Input setup using structure (DIN & Input setup using structure (DIN & SICS string)
read/ write Struct 2 0x41 0x418 0x01 0x32 0, 1 0x4432 N/A x x
SICS string)
Write Output Signal Manually control output ports of device. High value will force the corresponding
output to turn high.
Write Byte, 1 2 0x42 0x418 0x01 0x33 0, 1 0x4433 N/A x x
e.g. When using little Endian format, 00011101 will cause output 0, 2, 3 and 4
high. Output 1 will be low.
Report Target weight Report Target Weight read Float 32 2 0x5F 0x419 0x01 0x10 0, 1 0x4460 N/A x x
Report/Write #4 trig mode (edge) Input is triggered on the leading (1) or lagging edge (0) 0x9007
read/ write Byte, 1 3 0x22 0x418 0x01 0x43 0, 1 0x4603 x x x
Report/Write #4 input Assignment Application dependent ex. 0=None, 1=Clear Tare, 2 = Tare, 3 = zero, 4… read/ write Byte, 1 3 0x21 0x418 0x01 0x42 0, 1 0x4602 0x9008 x x x
Report/Write #5 trig mode (edge) Input is triggered on the leading (1) or lagging edge (0) read/ write Byte, 1 3 0x25 0x418 0x01 0x46 0, 1 0x4606 0x9009 x x x
Report/Write #5 input Assignment Application dependent ex. 0=None, 1=Clear Tare, 2 = Tare, 3 = zero, 4… read/ write Byte, 1 3 0x24 0x418 0x01 0x45 0, 1 0x4605 0x900A x x x
Report/write #6 output assignment Value based on application: 0, 1, 2, 3, 4, 5, etc. read/ write Byte, 1 3 0x27 0x418 0x01 0x48 0, 1 0x4608 0x9025 x x x
Report/write #7 output assignment Value based on application: 0, 1, 2, 3, 4, 5, etc. read/ write Byte, 1 3 0x2E 0x418 0x01 0x4F 0, 1 0x460F 0x9026 x x x
Report/write #8 output assignment Value based on application: 0, 1, 2, 3, 4, 5, etc. read/ write Byte, 1 3 0x35 0x418 0x01 0x56 0, 1 0x4616 0x9027 x x x
Set Transaction Number Set the transaction number of one Alibi record that will be read read Long 2 0x73 0x412 0x01 0x04 0, 1 0x4104 0xC000 x x x
Read One Alibi Record Read one record from the Alibi log read/ write Struct (56 Byte) 2 0x74 0x412 0x01 0x05 0, 1 0x4105 0xC002 x x x
Report Smart5 Alarm Status Word 0 = Alarm code of latest alarm, Words 1 and 2 reserved read/ write Short*3 1 0xAB 0x413 0x01 0x0C 0, 1 0x415C 0xA017 x x x
Read float32 Test floating point variable – always reads 123.45 – no write permitted read Float 32 1 0x0A 0x30F 0x01 0x01 0, 1 0x5000 N/A x x x x
Write float32 Test floating point variable – no usage in device except for test write Float 32 1 0x0B 0x30F 0x01 0x02 0, 0 0x5001 N/A x x x x
Read uint16 Test integer variable – always reads 9876 read Uint 8 1 0x0C 0x30F 0x01 0x03 0, 1 0x5002 N/A x x x x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-11
EIP Instance Values
PROFINET/EtherCAT
ACT350 Precision
IND360 Precision
PROFIBUS Index
PROFINET slot +
CC Link IE Field
EIP Class Code
IND360 Analog
PROFIBUS slot
EIP Attribute #
Basic Address
POWERCELL
POWERCELL
ACT350 DIO
Description
Read/Write
Command
Data Type
IND500x
Acyclic Command List
ACT350
ACT350
IND360
Index
Write uint16 Test integer variable – no usage in device except for test write Uint 8 1 0x0D 0x30F 0x01 0x04 0, 1 0x5003 N/A x x x x
Read string Test string variable – always read "ABCD" read String 160 1 0x0E 0x30F 0x01 0x05 0, 1 0x5004 N/A x x x x
Write string Test string variable – always read "ABCD" write String 160 1 0x0F 0x30F 0x01 0x06 0, 1 0x5005 N/A x x x x
Read uint32 Test long integer variable – always reads 98765 read Long 1 0x10 0x30F 0x01 0x07 0, 1 0x5006 N/A x x x x
Write uint32 Test long integer variable – no usage in device except for test write Long 1 0x11 0x30F 0x01 0x08 0, 1 0x5007 N/A x x x x
Read uint8 Test Byte variable – always reads 56h read Byte, 1 1 0x12 0x30F 0x01 0x09 0, 1 0x5008 N/A x x x x
Write uint8 Test Byte variable – no usage in device except for test write Byte, 1 1 0x13 0x30F 0x01 0x10 0, 1 0x5009 N/A x x x x
5-12 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
5.1. Acyclic Commands – IND360 Dynamic Application
The acyclic commands listed below are for use with the IND360 Dynamic application. Please note that these same commands will
not necessarily be supported or may work differently if used with anything other than the IND360 Dynamic application. Please refer to
the section specific to the application installed on your device if using something other than the IND360 Dynamic application. More
details about the IND360 Dynamic Application can be found in the IND360 Dynamic Application Manual.
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFINET Slot +
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Photoeye Mode
RW Float 32 3 0xA0 0x41B 0x01 0x01 0, 1 0x4701 0x10000 x
Parameter: 0 = Dual photoeye 1 = Single Photoeye
Installation Position
RW Float 32 3 0xA1 0x41B 0x01 0x02 0, 1 0x4702 0x10002 x
Parameter: 0 = Front 1 = Rear
Multiple Objects
RW Float 32 3 0xA2 0x41B 0x01 0x03 0, 1 0x4703 0x10004 x
Parameter: 0 = False 1 = True
Burr Time
RW Float 32 3 0xA3 0x41B 0x01 0x04 0, 1 0x4704 0x10006 x
Parameter: 1-1000 ms
Interval Time
RW Float 32 3 0xA4 0x41B 0x01 0x05 0, 1 0x4705 0x10008 x
Parameter: 1-1000 ms
Object Length
RW Float 32 3 0xA5 0x41B 0x01 0x06 0, 1 0x4706 0x1000A x
Parameter: < Belt Length
Belt Speed
RW Float 32 3 0xA6 0x41B 0x01 0x07 0, 1 0x4707 0x1000C x
Parameter: 1-10000 rpm
Belt Length
RW Float 32 3 0xA7 0x41B 0x01 0x08 0, 1 0x4708 0x1000E x
Parameter: 1-5000 cm
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-13
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFINET Slot +
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
Read/Write
Acyclic Command List
Data Type
Address
Index
Command/Description
Photoeye Polarity
RW Float 32 3 0xA8 0x41B 0x01 0x09 0, 1 0x4709 0x10010 x
Parameter: 0 = High Level 1 = Low Level
Filter Mode
RW Float 32 3 0xAB 0x41B 0x01 0x0C 0, 1 0x470C 0x10016 x
Parameter: 0 = Automatic 1 = Manual
Filter Parameter
RW Float 32 3 0xAC 0x41B 0x01 0x0D 0, 1 0x470D 0x10018 x
Parameter: 1-1000 ms
Compensation Management
RW Float 32 3 0xAD 0x41B 0x01 0x0E 0, 1 0x470E 0x1001A x
Parameter: 0 = Disable 1 = Enable
Weight 1 – Dynamic Application RW Float 32 3 0xAE 0x41B 0x01 0x0F 0, 1 0x470F 0x1001C x
Weight 2 – Dynamic Application RW Float 32 3 0xAF 0x41B 0x01 0x10 0, 1 0x4710 0x1001E x
Weight 3 – Dynamic Application RW Float 32 3 0xB0 0x41B 0x01 0x11 0, 1 0x4711 0x10020 x
Weight 4 – Dynamic Application RW Float 32 3 0xB1 0x41B 0x01 0x12 0, 1 0x4712 0x10022 x
Weight 5 – Dynamic Application RW Float 32 3 0xB2 0x41B 0x01 0x13 0, 1 0x4713 0x10024 x
5-14 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFINET Slot +
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Object Counts
R Float 32 3 0xBA 0x41B 0x01 0x1B 0, 1 0x471B 0x10034 x
Parameter: 0-5
Total Counts
R Float 32 3 0xBB 0x41B 0x01 0x1C 0, 1 0x471C 0x10036 x
Parameter: 0-99,999,999
Real Weighing Time R Float 32 3 0xBD 0x41B 0x01 0x1E 0, 1 0x471E 0x1003A x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-15
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFINET Slot +
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
Read/Write
Acyclic Command List
Data Type
Address
Index
Command/Description
Clear Statistics
R Float 32 3 0xC6 0x41B 0x01 0x27 0, 1 0x4727 0x1004C x
Parameter: 1 = clear
Zero State
RW Float 32 3 0xC7 0x41B 0x01 0x28 0, 1 0x4728 0x1004E x
Parameter: 0-100 d
Empty State
RW Float 32 3 0xC8 0x41B 0x01 0x29 0, 1 0x4729 0x10050 x
Parameter: 0-1000d
Over Weight
RW Float 32 3 0xC9 0x41B 0x01 0x2A 0, 1 0x472A 0x10052 x
Parameter: 0-capacity
Under Weight
RW Float 32 3 0xCA 0x41B 0x01 0x2B 0, 1 0x472B 0x10054 x
Parameter: 0-capacity
Photoeye Timeout
RW Float 32 3 0xCB 0x41B 0x01 0x2C 0, 1 0x472C 0x10056 x
Parameter: 0-5000 ms
5-16 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFINET Slot +
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Input1 Assignment
Parameter: 0 = None 7 = Run/Stop 8 = Front Photo RW Byte, 1 2 0x11 0x418 0x01 0x02 0, 1 0x4402 0x9002 x
Eye 9 = Rear Photo Eye 10 = Clear Statistics
Input2 Assignment
Parameter: 0 = None 7 = Run/Stop 8 = Front Photo RW Byte, 1 2 0x14 0x418 0x01 0x05 0, 1 0x4405 0x9004 x
Eye 9 = Rear Photo Eye 10 = Clear Statistics
Input3 Assignment
Parameter: 0 = None 7 = Run/Stop 8 = Front Photo RW Byte, 1 2 0x17 0x418 0x01 0x08 0, 1 0x4408 0x9006 x
Eye 9 = Rear Photo Eye 10 = Clear Statistics
Input4 Assignment
Parameter: 0 = None 7 = Run/Stop 8 = Front Photo RW Byte, 1 3 0x21 0x418 0x01 0x42 0, 1 0x4602 0x9008 x
Eye 9 = Rear Photo Eye 10 = Clear Statistics
Input5 Assignment
Parameter: 0 = None 7 = Run/Stop 8 = Front Photo RW Byte, 1 3 0x24 0x418 0x01 0x45 0, 1 0x4605 0x900A x
Eye 9 = Rear Photo Eye 10 = Clear Statistics
Output1 Assignment
Parameter: 0 = None 15 = Run/Stop 16 = Ready 17
RW Byte, 1 2 0x1D 0x418 0x01 0x0E 0, 1 0x440E 0x9020 x
= Complete 18 = Zero State 19 = Empty State 20 =
Alarm 21 = Timed Zero
Output2 Assignment
Parameter: 0 = None 15 = Run/Stop 16 = Ready 17
RW Byte, 1 2 0x24 0x418 0x01 0x15 0, 1 0x4415 0x9021 x
= Complete 18 = Zero State 19 = Empty State 20 =
Alarm 21 = Timed Zero
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-17
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFINET Slot +
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
Read/Write
Acyclic Command List
Data Type
Address
Index
Command/Description
Output3 Assignment
Parameter: 0 = None 15 = Run/Stop 16 = Ready 17
RW Byte, 1 2 0x2B 0x418 0x01 0x1C 0, 1 0x441C 0x9022 x
= Complete 18 = Zero State 19 = Empty State 20 =
Alarm 21 = Timed Zero
Output4 Assignment
Parameter:
0 = None
15 = Run/Stop
16 = Ready
RW Byte, 1 2 0x32 0x418 0x01 0x23 0, 1 0x4423 0x9023 x
17 = Complete
18 = Zero State
19 = Empty State
20 = Alarm
21 = Timed Zero
Output5 Assignment
Parameter: 0 = None 15 = Run/Stop 16 = Ready 17
RW Byte, 1 2 0x39 0x418 0x01 0x2A 0, 1 0x442A 0x9024 x
= Complete 18 = Zero State 19 = Empty State 20 =
Alarm 21 = Timed Zero
Output6 Assignment
Parameter: 0 = None 15 = Run/Stop 16 = Ready 17
RW Byte, 1 3 0x27 0x418 0x01 0x48 0, 1 0x4608 0x9025 x
= Complete 18 = Zero State 19 = Empty State 20 =
Alarm 21 = Timed Zero
Output7 Assignment
Parameter: 0 = None 15 = Run/Stop 16 = Ready 17
RW Byte, 1 3 0x2E 0x418 0x01 0x4F 0, 1 0x460F 0x9026 x
= Complete 18 = Zero State 19 = Empty State 20 =
Alarm 21 = Timed Zero
5-18 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFINET Slot +
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Output8 Assignment
Parameter: 0 = None 15 = Run/Stop 16 = Ready 17
RW Byte, 1 3 0x35 0x418 0x01 0x56 0, 1 0x4616 0x9027 x
= Complete 18 = Zero State 19 = Empty State 20 =
Alarm 21 = Timed Zero
IND360 POWERCELL
EIP Instance Value
IND360 Precision
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
PROFINET Slot
Read/Write
Data Type
+subslot
Address
Index
Command/Description
Target Source
RW Float 32 3 0xA0 0x41B 0x01 0x01 0, 1 0x4701 0x10000 x x x
Parameter: 0 – Tank Capacity
Tank Capacity
RW Float 32 3 0xA1 0x41B 0x01 0x02 0, 1 0x4702 0x10002 x x x
Parameter: 0 – Scale Capacity
Upper Limit
RW Float 32 3 0xA2 0x41B 0x01 0x03 0, 1 0x4703 0x10004 x x x
Parameter: 0 – Tank Capacity
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-19
CC Link IE Field Basic
PROFINET/EtherCAT
IND360 POWERCELL
EIP Instance Value
IND360 Precision
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
PROFINET Slot
Read/Write
Data Type
Acyclic Command List
+subslot
Address
Index
Command/Description
Lower Limit
RW Float 32 3 0xA3 0x41B 0x01 0x04 0, 1 0x4704 0x10006 x x x
Parameter: 0 – Upper Limit
Percentage of tank filled based on capacity R Float 32 3 0xA7 0x41B 0x01 0x08 0, 1 0x4708 0x1000E x x x
Clear Statistics
Parameter: 0 = Disable 1 = Enable (Will W Float 32 3 0xA8 0x41B 0x01 0x09 0, 1 0x4709 0x10010 x x x
automatically disable once complete)
Refill Counts
R Float 32 3 0xAB 0x41B 0x01 0x0C 0, 1 0x470C 0x10016 x x x
Parameter: 0 – 99,999,999
Control Command
Parameter: 0 = Stop W Float 32 3 0xAC 0x41B 0x01 0x0D 0, 1 0x470D 0x10018 x x x
1 = Run
5-20 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
PROFINET/EtherCAT
IND360 POWERCELL
EIP Instance Value
IND360 Precision
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
PROFINET Slot
Read/Write
Data Type
+subslot
Address
Index
Command/Description
Input1 Assignment
Parameter: 0 = None 1 = Tare 2 = Zero 3 = Print 4 =
RW Byte, 1 2 0x11 0x418 0x01 0x02 0, 1 0x4402 0x9002 x x x
Clear Tare 6 = Silence Alarm 7 = Run/Stop 8 = Clear
Statistics
Input2 Assignment
Parameter: 0 = None 1 = Tare 2 = Zero 3 = Print 4 =
RW Byte, 1 2 0x14 0x418 0x01 0x05 0, 1 0x4405 0x9004 x x x
Clear Tare 6 = Silence Alarm 7 = Run/Stop 8 = Clear
Statistics
Input3 Assignment
Parameter: 0 = None 1 = Tare 2 = Zero 3 = Print 4 =
RW Byte, 1 2 0x17 0x418 0x01 0x08 0, 1 0x4408 0x9006 x x x
Clear Tare 6 = Silence Alarm 7 = Run/Stop 8 = Clear
Statistics
Input4 Assignment
Parameter: 0 = None 1 = Tare 2 = Zero 3 = Print 4 =
RW Byte, 1 3 0x21 0x418 0x01 0x42 0, 1 0x4602 0x9008 x x x
Clear Tare 6 = Silence Alarm 7 = Run/Stop 8 = Clear
Statistics
Input5 Assignment
Parameter: 0 = None 1 = Tare 2 = Zero 3 = Print 4 =
RW Byte, 1 3 0x24 0x418 0x01 0x45 0, 1 0x4605 0x900A x x x
Clear Tare 6 = Silence Alarm 7 = Run/Stop 8 = Clear
Statistics
Output1 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 2 0x1D 0x418 0x01 0x0E 0, 1 0x440E 0x9020 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-21
CC Link IE Field Basic
PROFINET/EtherCAT
IND360 POWERCELL
EIP Instance Value
IND360 Precision
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
PROFINET Slot
Read/Write
Data Type
Acyclic Command List
+subslot
Address
Index
Command/Description
Output2 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 2 0x24 0x418 0x01 0x15 0, 1 0x4415 0x9021 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
Output3 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 2 0x2B 0x418 0x01 0x1C 0, 1 0x441C 0x9022 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
Output4 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 2 0x32 0x418 0x01 0x23 0, 1 0x4423 0x9023 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
Output5 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 2 0x39 0x418 0x01 0x2A 0, 1 0x442A 0x9024 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
Output6 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 3 0x27 0x418 0x01 0x48 0, 1 0x4608 0x9025 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
Output7 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 3 0x2E 0x418 0x01 0x4F 0, 1 0x460F 0x9026 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
5-22 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
PROFINET/EtherCAT
IND360 POWERCELL
EIP Instance Value
IND360 Precision
PROFIBUS Index
IND360 Analog
EIP Class Code
PROFIBUS Slot
EIP Attribute #
PROFINET Slot
Read/Write
Data Type
+subslot
Address
Index
Command/Description
Output8 Assignment
Parameter: 0 = None 1 = Center of Zero 2 = Over
RW Byte, 1 3 0x35 0x418 0x01 0x56 0, 1 0x4616 0x9027 x x x
Capacity 3 = Under Zero 4 = Motion 5 = Net 15 =
Upper Limit 16 = Lower Limit 17 = Refill 18 = Alarm
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
t
Read/Write
Data Type
lo
Address
sb
Index
Command/Description us
Work Mode
RW Float 32 3 0xA0 0x41B 0x01 0x01 0, 1 0x4701 0x10000 x x x
Parameter: 0 = Fill Dump 1 = Refill Dose
Feed Speeds
Parameter: 0 = One Speed 1 = Two Speed 2 = RW Float 32 3 0xA1 0x41B 0x01 0x02 0, 1 0x4702 0x10002 x x x
Three Speed
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-23
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Acyclic Command List
Address
Index
Command/Description
Output Type
Parameter: 0 = Concurrent RW Float 32 3 0xA2 0x41B 0x01 0x03 0, 1 0x4703 0x10004 x x x
1 = Independent
Complete Mode
RW Float 32 3 0xA3 0x41B 0x01 0x04 0, 1 0x4704 0x10006 x x x
Parameter: 0 = Weight Mode 1 = Time Mode
Clear Statistics
Parameter: 0 = Disable 1 = Enable (Will W Float 32 3 0xA5 0x41B 0x01 0x06 0, 1 0x4706 0x1000A x x x
automatically disable once complete)
Target Source
RW Float 32 3 0xA6 0x41B 0x01 0x07 0, 1 0x4707 0x1000C x x x
Parameter: 0 = Gross 1 = Net
Target
RW Float 32 3 0xA7 0x41B 0x01 0x08 0, 1 0x4708 0x1000E x x x
Parameter: 0 - Capacity
Spill
RW Float 32 3 0xA8 0x41B 0x01 0x09 0, 1 0x4709 0x10010 x x x
Parameter: 0 - Capacity
Fine Feed
RW Float 32 3 0xA9 0x41B 0x01 0x0A 0, 1 0x470A 0x10012 x x x
Parameter: 0 - Capacity
Fast Feed
RW Float 32 3 0xAA 0x41B 0x01 0x0B 0, 1 0x470B 0x10014 x x x
Parameter: 0 - Capacity
Heel Weight
RW Float 32 3 0xAB 0x41B 0x01 0x0C 0, 1 0x470C 0x10016 x x x
Parameter: 0 - Capacity
+ Tolerance
RW Float 32 3 0xAC 0x41B 0x01 0x0D 0, 1 0x470D 0x10018 x x x
Parameter: 0 - Capacity
5-24 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
- Tolerance
RW Float 32 3 0xAD 0x41B 0x01 0x0E 0, 1 0x470E 0x1001A x x x
Parameter: 0 - Capacity
Upper Limit
RW Float 32 3 0xAE 0x41B 0x01 0x0F 0, 1 0x470F 0x1001C x x x
Parameter: 0 - Capacity
Lower Limit
RW Float 32 3 0xAF 0x41B 0x01 0x10 0, 1 0x4710 0x1001E x x x
Parameter: 0 - Capacity
Inhibit Time
RW Float 32 3 0xB2 0x41B 0x01 0x13 0, 1 0x4713 0x10024 x x x
Parameter: 0 – 9.99 seconds
Stable Time
RW Float 32 3 0xB3 0x41B 0x01 0x14 0, 1 0x4714 0x10026 x x x
Parameter: 0 – 9.99 seconds
Complete Time
RW Float 32 3 0xB4 0x41B 0x01 0x15 0, 1 0x4715 0x10028 x x x
Parameter: 0 – 9.99 seconds
Control Timeout
RW Float 32 3 0xB5 0x41B 0x01 0x16 0, 1 0x4716 0x1002A x x x
Parameter: 0 – 99.99 seconds
Process Timeout
RW Float 32 3 0xB6 0x41B 0x01 0x17 0, 1 0x4717 0x1002C x x x
Parameter: 0 – 99.99 seconds
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-25
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Acyclic Command List
Address
Index
Command/Description
Control Reliability
Parameter: 3 = 93.32%
4 = 99.379% RW Float 32 3 0xBC 0x41B 0x01 0x1D 0, 1 0x471D 0x10038 x x x
5 = 99.977%
6 = 99.9997%
Learning Samples
RW Float 32 3 0xBD 0x41B 0x01 0x1E 0, 1 0x471E 0x1003A x x x
Parameter: 6, 9, 12, 15, 9999
Adjust Factor
RW Float 32 3 0xBE 0x41B 0x01 0x1F 0, 1 0x471F 0x1003C x x x
Parameter: 0.1 – 0.9
Filling Dosing Weight R Float 32 3 0xBF 0x41B 0x01 0x20 0, 1 0x4720 0x1003E x x x
Min Cycle Time R Float 32 3 0xC0 0x41B 0x01 0x21 0, 1 0x4721 0x10040 x x x
Max Cycle Time R Float 32 3 0xC1 0x41B 0x01 0x22 0, 1 0x4722 0x10042 x x x
Total Cycle Weight R Float 32 3 0xC2 0x41B 0x01 0x23 0, 1 0x4723 0x10044 x x x
5-26 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Control Command
W Float 32 3 0xC5 0x41B 0x01 0x26 0, 1 0x4726 0x1004A x x x
Parameter: 0 = Stop 1 = Run 2 = Pause
Dump Timeout
RW Float 32 3 0xC7 0x41B 0x01 0x28 0, 1 0x4728 0x1004E x x x
Parameter: 0 – 99.99 S
Jog Mode Selection RW Float 32 3 0xC9 0x41B 0x01 0x2A 0, 1 0x472A 0x10052 x x x
Jog Pulse Duration Time RW Float 32 3 0xCA 0x41B 0x01 0x2B 0, 1 0x472B 0x10054 x x x
Jog Pause Duration Time RW Float 32 3 0xCB 0x41B 0x01 0x2C 0, 1 0x472C 0x10056 x x x
Jog Max Cycles RW Float 32 3 0xCC 0x41B 0x01 0x2D 0, 1 0x472D 0x10058 x x x
Input1 Assignment
Parameter: 0 = None 6 = Silence Alarm 7 = RW Byte, 1 2 0x11 0x418 0x01 0x02 0, 1 0x4402 0x9002 x x x
Run/Stop 8 = Pause 9 = Clear Statistics
Input2 Assignment
Parameter: 0 = None 6 = Silence Alarm 7 = RW Byte, 1 2 0x14 0x418 0x01 0x05 0, 1 0x4405 0x9004 x x x
Run/Stop 8 = Pause 9 = Clear Statistics
Input3 Assignment
Parameter: 0 = None 6 = Silence Alarm 7 = RW Byte, 1 2 0x17 0x418 0x01 0x08 0, 1 0x4408 0x9006 x x x
Run/Stop 8 = Pause 9 = Clear Statistics
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-27
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Acyclic Command List
Address
Index
Command/Description
Input4 Assignment
Parameter: 0 = None 6 = Silence Alarm 7 = RW Byte, 1 3 0x21 0x418 0x01 0x42 0, 1 0x4602 0x9008 x x x
Run/Stop 8 = Pause 9 = Clear Statistics
Input5 Assignment
Parameter: 0 = None 6 = Silence Alarm 7 = RW Byte, 1 3 0x24 0x418 0x01 0x45 0, 1 0x4605 0x900A x x x
Run/Stop 8 = Pause 9 = Clear Statistics
Output1 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 2 0x1D 0x418 0x01 0x0E 0, 1 0x440E 0x9020 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
Output2 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 2 0x24 0x418 0x01 0x15 0, 1 0x4415 0x9021 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
Output3 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 2 0x2B 0x418 0x01 0x1C 0, 1 0x441C 0x9022 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
Output4 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 2 0x32 0x418 0x01 0x23 0, 1 0x4423 0x9023 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
Output5 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 2 0x39 0x418 0x01 0x2A 0, 1 0x442A 0x9024 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
5-28 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Output6 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 3 0x27 0x418 0x01 0x48 0, 1 0x4608 0x9025 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
Output7 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 3 0x2E 0x418 0x01 0x4F 0, 1 0x460F 0x9026 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
Output8 Assignment
Parameter: 0 = None 15 = Fast Feed 16 = Fine
RW Byte, 1 3 0x35 0x418 0x01 0x56 0, 1 0x4616 0x9027 x x x
Feed 17 = Spill 18 = Complete 19 = Dump 20 =
Refill 21 = Run
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-29
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Acyclic Command List
Address
Index
Command/Description
Flow Target
RW Float 32 3 0xA0 0x41B 0x01 0x01 0, 1 0x4701 0x10000 x
Parameter: 0.1 – Rated Flow
Upper Limit
RW Float 32 3 0xA3 0x41B 0x01 0x04 0, 1 0x4704 0x10006 x
Parameter: 0 - Capacity
Lower Limit
RW Float 32 3 0xA4 0x41B 0x01 0x05 0, 1 0x4705 0x10008 x
Parameter: 0 - Capacity
Rated Flow
RW Float 32 3 0xA7 0x41B 0x01 0x08 0, 1 0x4708 0x1000E x
Parameter: 0.1 – 99,999
Control Filter
RW Float 32 3 0xA8 0x41B 0x01 0x09 0, 1 0x4709 0x10010 x
Parameter: 0 - 9
5-30 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Flow Filter
RW Float 32 3 0xA9 0x41B 0x01 0x0A 0, 1 0x470A 0x10012 x
Parameter: 0 - 9
Control Tolerance
RW Float 32 3 0xAA 0x41B 0x01 0x0B 0, 1 0x470B 0x10014 x
Parameter: 0 - 1
Flow Tolerance
RW Float 32 3 0xAB 0x41B 0x01 0x0C 0, 1 0x470C 0x10016 x
Parameter: 0 - 1
Test Time
RW Float 32 3 0xAC 0x41B 0x01 0x0D 0, 1 0x470D 0x10018 x
Parameter: 0 – 9,999 Seconds
PID Mode
Parameter: 0 = Auto PID 1 = RW Float 32 3 0xB1 0x41B 0x01 0x12 0, 1 0x4712 0x10022 x
Manual PID
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-31
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Acyclic Command List
Address
Index
Command/Description
5-32 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Refill Time
RW Float 32 3 0xC0 0x41B 0x01 0x21 0, 1 0x4721 0x10040 x
Parameter: 0 – 9,999 seconds
Control Command
W Float 32 3 0xC6 0x41B 0x01 0x27 0, 1 0x4727 0x1004C x
Parameter: 0 = Stop 1 = Run
Clear Totalization
W Float 32 3 0xC7 0x41B 0x01 0x28 0, 1 0x4728 0x1004E x
Parameter: 1 = Clear
Input1 Assignment
Parameter: 0 = None 6 = Silence
RW Byte, 1 2 0x11 0x418 0x01 0x02 0, 1 0x4402 0x9002 x
Alarm 7 = Run/Stop 8 = Refill 9
= Clear Statistics
Input2 Assignment
Parameter: 0 = None 6 = Silence
RW Byte, 1 2 0x14 0x418 0x01 0x05 0, 1 0x4405 0x9004 x
Alarm 7 = Run/Stop 8 = Refill 9
= Clear Statistics
Input3 Assignment
Parameter: 0 = None 6 = Silence
RW Byte, 1 2 0x17 0x418 0x01 0x08 0, 1 0x4408 0x9006 x
Alarm 7 = Run/Stop 8 = Refill 9
= Clear Statistics
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-33
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Acyclic Command List
Address
Index
Command/Description
Input4 Assignment
Parameter: 0 = None 6 = Silence
RW Byte, 1 3 0x21 0x418 0x01 0x42 0, 1 0x4602 0x9008 x
Alarm 7 = Run/Stop 8 = Refill 9
= Clear Statistics
Input5 Assignment
Parameter: 0 = None 6 = Silence
RW Byte, 1 3 0x24 0x418 0x01 0x45 0, 1 0x4605 0x900A x
Alarm 7 = Run/Stop 8 = Refill 9
= Clear Statistics
Output1 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 2 0x1D 0x418 0x01 0x0E 0, 1 0x440E 0x9020 x
15 = Run 16 = Refill
Output2 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 2 0x24 0x418 0x01 0x15 0, 1 0x4415 0x9021 x
15 = Run 16 = Refill
Output3 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 2 0x2B 0x418 0x01 0x1C 0, 1 0x441C 0x9022 x
15 = Run 16 = Refill
Output4 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 2 0x32 0x418 0x01 0x23 0, 1 0x4423 0x9023 x
15 = Run 16 = Refill
Output5 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 2 0x39 0x418 0x01 0x2A 0, 1 0x442A 0x9024 x
15 = Run 16 = Refill
5-34 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
CC Link IE Field Basic
IND360 POWERCELL
PROFINET/EtherCAT
EIP Instance Value
IND360 Precision
PROFIBUS Index
PROFINET Slot +
IND360 Analog
PROFIBUS Slot
EIP Attribute #
Read/Write
Data Type
Address
Index
Command/Description
Output6 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 3 0x27 0x418 0x01 0x48 0, 1 0x4608 0x9025 x
15 = Run 16 = Refill
Output7 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 3 0x2E 0x418 0x01 0x4F 0, 1 0x460F 0x9026 x
15 = Run 16 = Refill
Output8 Assignment
Parameter: 0 = None 14 = Alarm RW Byte, 1 3 0x35 0x418 0x01 0x56 0, 1 0x4616 0x9027 x
15 = Run 16 = Refill
PROFINET /
PROFINET
Data Type
EtherCAT
IND500x
Slot +
Index
Command/Description
RW
Report (-) Tolerance Weight. Unit: Active Target unit R Float32 0, 1 0x470E x
Report (+) Tolerance Weight. Unit: Active Target unit R Float32 0, 1 0x470D x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-35
PROFINET /
PROFINET
Data Type
EtherCAT
IND500x
Slot +
Index
Command/Description
RW
Acyclic Command List
Write (-) Tolerance Weight. Unit: Active Target unit W Float32 0, 1 0x470E x
Write (+) Tolerance Weight. Unit: Active Target unit W Float32 0, 1 0x470D x
Read/Write
PROFINET /
PROFINET
Data Type
EtherCAT
IND500x
Slot +
Index
Command/Description
5-36 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
Read/Write
PROFINET /
PROFINET
Data Type
EtherCAT
IND500x
Slot +
Index
Command/Description
Report Fine Feed Weight. Unit: Active Target unit R Float 32 0, 1 0x470A x
Report (-) Tolerance Weight. Unit: Active Target unit R Float 32 0, 1 0x470E x
Report (+) Tolerance Weight. Unit: Active Target unit R Float 32 0, 1 0x470D x
Write Fine Feed Weight. Unit: Active Target unit W Float 32 0, 1 0x470A x
Write (-) Tolerance Weight. Unit: Active Target unit W Float 32 0, 1 0x470E x
Write (+) Tolerance Weight. Unit: Active Target unit W Float 32 0, 1 0x470D x
Control Command:
W Float 32 0, 1 0x4726 x
0-Stop; 1-Run; 2-Pause
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-37
5.7. Acyclic Commands – IND500x Advanced Auto-Filling
Application
Acyclic Command List
The acyclic commands listed below are for use with the IND500x Advanced Auto-Filling Application. Please note that these same
commands will not necessarily be supported or may work differently if used with anything other than IND500x Advanced Auto-Filling
Application. Please refer to the section specific to the application installed on your device if using something other than the IND500x
Advanced Auto-Filling Application. More details about the IND500x Advanced Auto-Filling Application can be found in the IND500x
Manual.
PROFINET /
PROFINET
Data Type
EtherCAT
IND500x
Slot +
Index
Command/Description
RW
Report Weight Unit R Float 32 0, 1 0x4708 x
Report Fine Feed Weight. Unit: Active Target unit R Float 32 0, 1 0x470A x
Report (-) Tolerance Weight. Unit: Active Target unit R Float 32 0, 1 0x470E x
Report (+) Tolerance Weight. Unit: Active Target unit R Float 32 0, 1 0x470D x
Write Fine Feed Weight. Unit: Active Target unit W Float 32 0, 1 0x470A x
Write (-) Tolerance Weight. Unit: Active Target unit W Float 32 0, 1 0x470E x
Write (+) Tolerance Weight. Unit: Active Target unit W Float 32 0, 1 0x470D x
5-38 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
PROFINET /
PROFINET
Data Type
EtherCAT
IND500x
Slot +
Index
Command/Description
RW
Set (+) Tolerance % W Float 32 0, 1 0x472D x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-39
5.8. Acyclic Commands – IND500x Over Under Application
The acyclic commands listed below are for use with the IND500x Over Under Application. Please note that these same commands will
not necessarily be supported or may work differently if used with anything other than IND500x Over Under Application. Please refer to
Acyclic Command List
the section specific to the application installed on your device if using something other than the IND500x Over Under Application.
More details about this Application can be found in the IND500x Manual.
PROFINET /
PROFINET
Data Type
EtherCAT
IND500x
Slot +
Index
Command/Description
RW
Report Target Weight. Unit: Active Target unit R Float32 0, 1 0x4708 x
Report (-) Tolerance Weight. Unit: Active Target unit R Float32 0, 1 0x470E x
Report (+) Tolerance Weight. Unit: Active Target unit R Float33 0, 1 0x470D x
Write (-) Tolerance Weight. Unit: Active Target unit W Float32 0, 1 0x470E x
Write (+) Tolerance Weight. Unit: Active Target unit W Float32 0, 1 0x470D x
5-40 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
5.9. Acyclic Commands – IND500x Drum-Filling Application
The acyclic commands listed below are for use with the IND500x Drum-Filling Application. Please note that these same commands
will not necessarily be supported or may work differently if used with anything other than IND500x Drum-Filling Application. Please
refer to the section specific to the application installed on your device if using something other than the IND500x Drum-Filling
Application. More details about this Application can be found in the IND500x Manual.
PROFINET Slot +
EtherCAT Index
PROFINET /
Data Type
IND500x
Command/Description
RW
Report Target Weight. Unit: Active Target unit R Float32 0, 1 0x4708 x
Report Fine Feed Weight. Unit: Active Target unit R Float32 0, 1 0x470A x
Report (-) Tolerance Weight. Unit: Active Target unit R Float32 0, 1 0x470E x
Report (+) Tolerance Weight. Unit: Active Target unit R Float32 0, 1 0x470D x
Write Fine Feed Weight. Unit: Active Target unit W Float32 0, 1 0x470A x
Write (-) Tolerance Weight. Unit: Active Target unit W Float32 0, 1 0x470E x
Write (+) Tolerance Weight. Unit: Active Target unit W Float32 0, 1 0x470D x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 5-41
PROFINET Slot +
EtherCAT Index
PROFINET /
Data Type
IND500x
Command/Description
Acyclic Command List
RW
Work Mode RW Short 0, 1 0x4701 x
Control Command
0-Stop;1-Run; 2-Pause; 4-Jog; 6-Resume; 7: Silence Alarm; 8: OK Key; W Float32 0, 1 0x4726 x
9: No Key; 13: Manual Accept
5-42 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
6 Frequently Asked Questions
6.1. What is the easiest way to get my SAI device
communicating with my automation system?
METTLER TOLEDO provides various device description files for each SAI device covered in this
manual. These files make integration into the automation system quick and easy. Many of these
device description files are already installed by default in the latest versions of the most popular
automation system configuration programs. If not already installed in your program, available
device description files can be found below.
IND500x www.mt.com/IND500x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 6-1
Automation Device Description METTLER TOLEDO
SAI Device Network File Product Page Downloads Page
If bit 15 of the response is low and the response does not equal the command, it could mean the
command is still being executed. For example, the cyclic command "tare when stable" (command
= 400) will wait for a no-motion before completing the tare. While waiting for a no-motion
condition, the response value may equal 2047 (in-process). Eventually, the command will either
complete successfully (response = 400) or the process will timeout if the no-motion condition is
not met (response bit 15 = 1).
6-2 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
Measuring block cyclic commands and 1 status block cyclic command to be executed at the
same time.
If using the SAI 1 block or 2 block formats, there is only one Measuring block available for cyclic
commands at a time. In order to read gross weight and net weight using these formats, the weight
commands will need to be cycled repeatedly in the automation system.
1000000000000001 -32767 Invalid Command Sent when the device determines that the command
is known but cannot be executed. This might occur
due state restrictions – for example attempting to
zero when the scale is outside of acceptable zero
range.
1000000000000010 -32766 Timeout Command Sent when the valid command that is received by the
device is unable to execute within a pre-determined
time. This might occur for commands that require
stable weight before execution, for example.
1000000000000100 -32764 Unknown Sent when the device does not support this
Command information (for example requesting rate values from
a device that does not provide rate functionality).
1000000000001000 -32760 Invalid Command Sent when a valid write command is received with
Data an invalid argument (for example one that is smaller
or larger than the allowed value).
1000000000010000 -32752 Aborted Command Sent once a second command to cancel the prior
command has been received and processed. This
can only occur if the original command 1) permits
cancellation, 2) has not already completed
successfully, and 3) has not already failed.
1000000000100000 -32736 Step Failed Sent when the device determines that the
Command command's current step has failed. At this point the
control system will need to decide whether to abort
the sequence (command = 2004), to retry the step
(command = 2005), or to skip this step (command
= 2006) and try to perform the next step. Not all
processes will allow all three of these options –
those not permitted will have an invalid command
response when the next step command is sent.
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 6-3
6.6. I still don't quite understand how SAI works.
What other resources are available?
Frequently Asked Questions
One of the best ways to understand SAI is to see an example of it. Sample code is provided for each
SAI device covered in this manual. Sample code is written for Studio 5000 (EtherNet/IP) and TIA
Portal (PROFINET and PROFIBUS DP). The sample code shows how to use cyclic commands such
as zero and tare, scale processes like performing a span adjustment via acyclic commands and
much more. These functions are contained in Add-On Instructions (AOIs) for EtherNet/IP and
Function Blocks for PROFINET and PROFIBUS DP. Even if you do not have access to the software
the sample code was written for, detailed engineering notes are available in the sample code
download that explain how to use the sample code.
This SAI manual is specific to METTLER TOLEDO terminals and transmitters. A more detailed manual
about SAI called 30588288_Rxx_MAN_UG_SAI_EN.pdf is available. This manual details SAI in a
more general sense.
Both the sample code and SAI manual can be found on the downloads page of your SAI device.
Refer to the table in Section 5.1 for a link to the downloads page.
If there are still questions, please contact your local METTLER TOLEDO representative for additional
help.
6-4 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
A. Modbus RTU/Modbus TCP
Commands
The commands in the table below are not SAI commands, but they are PLC commands that closely
mirror some SAI commands so they have been included as an appendix. Please note that these
commands are only supported by IND360.
IND360 supports the Modbus communication protocol. IND360 is able to communicate bi-
directionally with the host system. Modbus commands "03H", "06H" and "10H" are supported.
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
Report Default Value IND360 40001 R Float 32 x x x
Report Rounded Gross Weight IND360 40003 R Float 32 x x x
Report Rounded Tare Weight IND360 40005 R Float 32 x x x
Report Rounded Net Weight IND360 40007 R Float 32 x x x
Report Gross Weight IND360 40009 R Float 32 x x x
Report Tare Weight IND360 40011 R Float 32 x x x
Report Net Weight IND360 40013 R Float 32 x x x
Report Weight Unit IND360 40015 R Float 32 x x x
Report Raw load cell counts IND360 40017 R Float 32 x x
Write Preset Tare Weight IND360 40020 W Float 32 x x x
Tare IND360 40022 W Short x x x
Report Tare operation status IND360 40023 R Short x x x
Zero IND360 40024 W Short x x x
Report zero operation status IND360 40025 R Short x x x
Clear Tare IND360 40026 W Short x x x
Tare Immediate IND360 40027 W Short x x x
Zero Immediate IND360 40028 W Short x x x
Print IND360 40029 W Short x x x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-1
IND360 POWERCELL
IND360 Precision
IND360 Analog
Modbus RTU/Modbus TCP Commands
Read/Write
Data Type
Command Device Address
Report SN IND360 40030 R Short [20] x x x
Report OS version IND360 40040 R Short [20] x x x
Report Software Version IND360 40070 R Short [20] x x x
Device identification IND360 40080 R Short [20] x x x
Read gross weight from each POWERCELL
IND360 40090 R Float 32[1+8] x
LC
Read net weight from each POWERCELL
IND360 40108 R Float 32[1+8] x
LC
GEO code IND360 40200 R/W Float 32 x x
Disable /Enable keyboard
IND360 40202 R/W Short x x x
Parameter: 1 - Disable, 0 - Enable
Report d IND360 40204 R Float 32 x x x
Report "Nmax" IND360 40206 R Float 32 x x x
Cancel adjustment / test (Abort Test
Function / Adjustment) IND360 40401 W Short x x x
Parameter: 1 - exit the calibration
Report Adjustment/Test Status
Parameter: 0 - Success
1 - In process IND360 40402 R Short x x x
2 - Complete with dyanmic
255 - Fail
Set External adjustment weight IND360 40403 W Float 32 x x
Number of linearity ranges
Parameter: 0 - Disable
1 - 3 point IND360 40405 R/W Short x x
2 - 4 point
3 - 5 point
Zero Adjustment IND360 40400 W Short x x
Span Adjustment Value 1 xLow IND360 40407 R/W Float 32 x x
Span Adjustment Value 2 Low IND360 40409 R/W Float 32 x x
Span Adjustment Value 3 Middle IND360 40411 R/W Float 32 x x
Span Adjustment Value 4 High IND360 40413 R/W Float 32 x x
Validate (Confirm) Adjustment
IND360 40415 R/W Short x x
Parameter: 1 - Confirm calibration
Span Adjustment Counts 1 xLow IND360 40416 R/W Long x x
Span Adjustment Counts 2 Low IND360 40418 R/W Long x x
A-2 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
Span Adjustment Counts 3 Middle IND360 40420 R/W Long x x
Span Adjustment Counts 4 High IND360 40422 R/W Long x x
Set number of steps & begin step
calibration
Parameter: 1 - Step 1
2 - Step 2 IND360 40424 write Short x x
3 - Step 3
4 - Step 4
5 - Step 5
Sets weight value for current step in
IND360 40425 write Float 32 x x
calibration & starts step
Calfree
IND360 40427 write Short x
Parameter: 1 - Start Calfree
Calfree cell capacity IND360 40428 R/W Float 32 x
Calfree unit
Parameter: 0 - g
1 - kg IND360 40430 R/W Short x
2 - Ib
3-t
4 - ton
Calfree cell output IND360 40431 R/W Float 32 x
Get temporary weight in step mode IND360 40433 R Float 32 x x
Zero adjustment count IND360 40435 R Long x x
Start adjustment with internal weight IND360 40437 W Short x
Start adjustment with external weight IND360 40438 W Short x
Start customer standard calibration IND360 40439 W Short x
Start test with internal weight IND360 40440 W Short x
Start test with external weight IND360 40441 W Short x
Report test deviation IND360 40442 R Float 32 x
Requested weight IND360 40444 R Float 32 x
External test weight IND360 40446 W Float 32 x
CalFree Plus IND360 40448 R/W Short x
Enable / Disable step control C5 IND360 40449 R/W Short x
Sensitivity adjustment (Triggered) C8 1 -4 IND360 40450 W Short x
Sensitivity adjustment (Display weight) C8
IND360 40451 W Float 32*2 x
7
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-3
IND360 POWERCELL
IND360 Precision
IND360 Analog
Modbus RTU/Modbus TCP Commands
Read/Write
Data Type
Command Device Address
Internal adjustment PBK/PFK IND360 40455 W Short x
Report/Write Comparator 1 Source
Parameter: 0 - None
1- Display weight IND360 40500 R/W Short x x x
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 2 Source
Parameter: 0 - None
1- Display weight IND360 40501 R/W Short x x x
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 3 Source
Parameter: 0 - None
1- Display weight IND360 40502 R/W Short x x x
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 4 Source
Parameter: 0 - None
1- Display weight IND360 40503 R/W Short x x x
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 5 Source
Parameter: 0 - None
1- Display weight IND360 40504 R/W Short x x x
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 6 Source
Parameter: 0 - None
1- Display weight IND360 40505 R/W Short x x x
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 7 Source
Parameter: 0 - None
1- Display weight IND360 40506 R/W Short x x x
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 8 Source
Parameter: 0 - None IND360 40507 R/W Short x x x
1- Display weight
A-4 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
2 - ABS-display weight
3 - Gross weight
Report/Write Comparator 1 Active
Parameter: 0 - “<"
1 - "<="
2 - "=="
3 - ">" IND360 40520 R/W Short x x x
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 2 Active
Parameter: 0 - “<"
1 - "<="
2 - "=="
3 - ">" IND360 40521 R/W Short x x x
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 3 Active
Parameter: 0 - “<"
1 - "<="
2 - "=="
3 - ">" IND360 40522 R/W Short x x x
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 4 Active
Parameter: 0 - “<"
1 - "<="
2 - "=="
3 - ">" IND360 40523 R/W Short x x x
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 5 Active
Parameter: 0 - “<" IND360 40524 R/W Short x x x
1 - "<="
2 - "=="
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-5
IND360 POWERCELL
IND360 Precision
IND360 Analog
Modbus RTU/Modbus TCP Commands
Read/Write
Data Type
Command Device Address
3 - ">"
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 6 Active
Parameter: 0 - “<"
1 - "<="
2 - "=="
3 - ">" IND360 40525 R/W Short x x x
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 7 Active
Parameter: 0 - “<"
1 - "<="
2 - "=="
3 - ">" IND360 40526 R/W Short x x x
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 8 Active
Parameter: 0 - “<"
1 - "<="
2 - "=="
3 - ">" IND360 40527 R/W Short x x x
4 - ">="
5 -"<>"
6 - "_<>_"
7 - ">_<"
Report/Write Comparator 1 Limit IND360 40540 R/W Float 32 x x x
Report/Write Comparator 1 High Limit IND360 40542 R/W Float 32 x x x
Report/Write Comparator 2 Limit IND360 40544 R/W Float 32 x x x
Report/Write Comparator 2 High Limit IND360 40546 R/W Float 32 x x x
Report/Write Comparator 3 Limit IND360 40548 R/W Float 32 x x x
Report/Write Comparator 3 High Limit IND360 40550 R/W Float 32 x x x
Report/Write Comparator 4 Limit IND360 40552 R/W Float 32 x x x
Report/Write Comparator 4 High Limit IND360 40554 R/W Float 32 x x x
A-6 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
Report/Write Comparator 5 Limit IND360 40556 R/W Float 32 x x x
Report/Write Comparator 5 High Limit IND360 40558 R/W Float 32 x x x
Report/Write Comparator 6 Limit IND360 40560 R/W Float 32 x x x
Report/Write Comparator 6 High Limit IND360 40562 R/W Float 32 x x x
Report/Write Comparator 7 Limit IND360 40564 R/W Float 32 x x x
Report/Write Comparator 7 High Limit IND360 40566 R/W Float 32 x x x
Report/Write Comparator 8 Limit IND360 40568 R/W Float 32 x x x
Report/Write Comparator 8 High Limit IND360 40570 R/W Float 32 x x x
Turn all internal & external output OFF
IND360 40700 W Short x x x
Parameter: 1 - off
Report/Write #1 trig mode (edge)
Parameter: 0 - Rising edge IND360 40701 R/W Short x x x
1 - Failing edge
Report/Write #1 input Assignment
Parameter: 0 - None
1 - Tare
2 - Zero
3 - Print IND360 40702 R/W Short x x x
4 - Clear tare
5 - KeypadDisable
6 - SilenceAlarm (Only for APP)
7 - Reserved
8 - KeypadEnable
Report/Write #2 trig mode (edge)
Parameter: 0 - Rising edge IND360 40703 R/W Short x x x
1 - Failing edge
Report/Write #2 input Assignment
Parameter: 0 - None
1 - Tare
2 - Zero
3 - Print IND360 40704 R/W Short x x x
4 - Clear tare
5 - KeypadDisable
6 - SilenceAlarm (Only for APP)
7 - Reserved
8 - KeypadEnable
Report/Write #3 trig mode (edge)
Parameter: 0 - Rising edge IND360 40705 R/W Short x x x
1 - Failing edge
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-7
IND360 POWERCELL
IND360 Precision
IND360 Analog
Modbus RTU/Modbus TCP Commands
Read/Write
Data Type
Command Device Address
Report/Write #3 input Assignment
Parameter: 0 - None
1 - Tare
2 - Zero
3 - Print IND360 40706 R/W Short x x x
4 - Clear tare
5 - KeypadDisable
6 - SilenceAlarm (Only for APP)
7 - Reserved
8 - KeypadEnable
Report/Write #4 trig mode (edge)
Parameter: 0 - Rising edge IND360 40707 R/W Short x x x
1 - Failing edge
Report/Write #4 input Assignment
Parameter: 0 - None
1 - Tare
2 - Zero
3 - Print IND360 40708 R/W Short x x x
4 - Clear tare
5 - KeypadDisable
6 - SilenceAlarm (Only for APP)
7 - Reserved
8 - KeypadEnable
Report/Write #5 trig mode (edge)
Parameter: 0 - Rising edge IND360 40709 R/W Short x x x
1 - Failing edge
Report/Write #5 input Assignment
Parameter: 0 - None
1 - Tare
2 - Zero
3 - Print IND360 40710 R/W Short x x x
4 - Clear tare
5 - KeypadDisable
6 - SilenceAlarm (Only for APP)
7 - Reserved
8 - KeypadEnable
Report/write #1 output assignment
Parameter: 0-None
2-Over capacity IND360 40711 R/W Short x x x
3-Under zero
4-Motion
5-Net
A-8 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
6-Comparator 1
7-Comparator 2
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report/write #2 output assignment
Parameter: 0-None
2-Over capacity
3-Under zero
4-Motion
5-Net
6-Comparator 1
7-Comparator 2 IND360 40712 R/W Short x x x
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report/write #3 output assignment
Parameter: 0-None
2-Over capacity
3-Under zero
4-Motion
5-Net
6-Comparator 1
7-Comparator 2 IND360 40713 R/W Short x x x
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report/write #4 output assignment IND360 40714 R/W Short x x x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-9
IND360 POWERCELL
IND360 Precision
IND360 Analog
Modbus RTU/Modbus TCP Commands
Read/Write
Data Type
Command Device Address
Parameter: 0-None
2-Over capacity
3-Under zero
4-Motion
5-Net
6-Comparator 1
7-Comparator 2
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report/write #5 output assignment
Parameter: 0-None
2-Over capacity
3-Under zero
4-Motion
5-Net
6-Comparator 1
7-Comparator 2 IND360 40715 R/W Short x x x
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report/write #6 output assignment
Parameter: 0-None
2-Over capacity
3-Under zero
4-Motion
5-Net
6-Comparator 1 IND360 40716 R/W Short x x x
7-Comparator 2
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
A-10 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report/write #7 output assignment
Parameter: 0-None
2-Over capacity
3-Under zero
4-Motion
5-Net
6-Comparator 1
7-Comparator 2 IND360 40717 R/W Short x x x
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report/write #8 output assignment
Parameter: 0-None
2-Over capacity
3-Under zero
4-Motion
5-Net
6-Comparator 1
7-Comparator 2
IND360 40718 R/W Short x x x
8-Comparator 3
9-Comparator 4
10-Comparator 5
11-Comparator 6
12-Comparator 7
13-Comparator 8
14-Smart5 red
15-Smart5 orange
Report scale status group
Parameter: bit0 - Set Status
bit1 - out of AD range over/under
bit2 - Reserved
bit3 - Data ok IND360 40800 R Short x x x
bit4 - RedAlert Alarm condition
bit5 - Center of Zero
bit6 - Motion
bit7- Net Mode
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-11
IND360 POWERCELL
IND360 Precision
IND360 Analog
Modbus RTU/Modbus TCP Commands
Read/Write
Data Type
Command Device Address
bit8 - Customer overload
bit9 - Customer underload
bit10 - zero fail at start
Read/Write customer defined overload
threshold IND360 40801 R/W Short x x
Parameter: 0-99 d
Read/Write customer defined underload
threshold IND360 40802 R/W Short x x
Parameter: 0-99 d
Report IO Input Group #1
Parameter: bit0 - Input1
bit1 - Input2 IND360 40803 R Short x x x
bit2 - Input3
bit3 - Input4
bit4 - Input5
Report/Write IO Output Group #1
Parameter: bit0 - Output1
bit1 - Output2
bit2 - Output3
bit3 - Output4 IND360 40805 R/W Short x x x
bit4 - Output5
bit5 - Output6
bit6 - Output7
bit7 - Output8
(not writable when no outputs on device)
Report Alarm status group IND360 40807 R Short x x x
Report RedAlert Group IND360 40808 R Short x x x
Report scale status group 2 IND360 40809 R Short x x x
Report Comparator status group 1 IND360 40810 R Short x x x
Report custom group 1 IND360 40811 R Short x x x
Report custom group 2 IND360 40812 R Short x x x
Re-Initialize the parameters
Parameter: 1- Parameter re-initialize (All IND360 40999 W Short x x x
parameters need to be enabled by this
command)
Target Source IND360 TankVessel 41001 R/W Short x x x
Tank Capacity IND360 TankVessel 41002 R/W Float 32 x x x
Upper Limit IND360 TankVessel 41004 R/W Float 32 x x x
A-12 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
Lower Limit IND360 TankVessel 41006 R/W Float 32 x x x
Lower Limit Alarm IND360 TankVessel 41008 R/W Float 32 x x x
Over Limit Alarm IND360 TankVessel 41010 R/W Float 32 x x x
Current Weight IND360 TankVessel 41012 R Float 32 x x x
Percentage IND360 TankVessel 41014 R Float 32 x x x
Clear Statistics IND360 TankVessel 41016 W Short x x x
Lower Limit Counts IND360 TankVessel 41017 R Long x x x
Upper Limit Counts IND360 TankVessel 41019 R Long x x x
Refill Counts IND360 TankVessel 41021 R Long x x x
Alarm Status IND360 TankVessel 41023 R Short x x x
Run Status IND360 TankVessel 41024 R Short x x x
Control Command IND360 TankVessel 41025 W Short x x x
Photoeye Mode IND360 DYN 42001 R/W Short x
Installation Position IND360 DYN 42002 R/W Short x
Multiple Objects IND360 DYN 42003 R/W Short x
Burr Time IND360 DYN 42004 R/W Short x
Interval Time IND360 DYN 42005 R/W Short x
Object Length IND360 DYN 42006 R/W Short x
Belt Speed IND360 DYN 42007 R/W Short x
Belt length IND360 DYN 42008 R/W Short x
Photoeye Polarity IND360 DYN 42009 R/W Short x
Min. Weighing Time IND360 DYN 42010 R/W Short x
Max. Weighing Time IND360 DYN 42011 R/W Short x
Filter Mode IND360 DYN 42012 R/W Short x
Filter Parameter IND360 DYN 42013 R/W Short x
Compensation Management IND360 DYN 42038 R/W Short x
Weight1 IND360 DYN 42039 R/W Float 32 x
Weight2 IND360 DYN 42041 R/W Float 32 x
Weight3 IND360 DYN 42043 R/W Float 32 x
Weight4 IND360 DYN 42045 R/W Float 32 x
Weight5 IND360 DYN 42047 R/W Float 32 x
Factor1 IND360 DYN 42049 R/W Float 32 x
Factor2 IND360 DYN 42051 R/W Float 32 x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-13
IND360 POWERCELL
IND360 Precision
IND360 Analog
Modbus RTU/Modbus TCP Commands
Read/Write
Data Type
Command Device Address
Factor3 IND360 DYN 42053 R/W Float 32 x
Factor4 IND360 DYN 42055 R/W Float 32 x
Factor5 IND360 DYN 42057 R/W Float 32 x
Completed Signal Time IND360 DYN 42014 R/W Short x
Match Weighing Time IND360 DYN 42015 R/W Short x
Object Counts IND360 DYN 42016 R Short x
Total Counts IND360 DYN 42017 R Long x
Dynamic Weight IND360 DYN 42019 R Float 32 x
Real Weighing Time IND360 DYN 42021 R Short x
Run Status IND360 DYN 42022 R Short x
Alarm Status IND360 DYN 42023 R Short x
Control command IND360 DYN 42025 W Short x
Clear Statistics IND360 DYN 42026 W Short x
Valid Weight Counts IND360 DYN 42034 R Long x
Invalid Weight Counts IND360 DYN 42036 R Long x
Zero state IND360 DYN 42027 R/W Short x
Empty state IND360 DYN 42028 R/W Short x
Over Weight IND360 DYN 42029 R/W Float 32 x
Under Weight IND360 DYN 42031 R/W Float 32 x
Photoeye timeout IND360 DYN 42033 R/W Short x
Dynamic Zero Enable IND360 DYN 42060 R/W Short x
Dynamic Zero Threshold IND360 DYN 42061 R/W Float 32 x
Dynamic Zero Delay IND360 DYN 42063 R/W Short x
Dynamic Zero Interval IND360 DYN 42064 R/W Short x
Work Mode IND360 FillDose 43001 RW Short x
Feed Speeds IND360 FillDose 43002 RW Short x
Output Type IND360 FillDose 43003 RW Short x
Complete Mode IND360 FillDose 43004 RW Short x
Clear Statistics IND360 FillDose 43006 W Short x
Target Source IND360 FillDose 43007 RW Short x
Target IND360 FillDose 43008 RW Float 32 x
Spill IND360 FillDose 43010 RW Float 32 x
Fine Feed IND360 FillDose 43012 RW Float 32 x
A-14 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters 30587511 | 05 | 02/2023
IND360 POWERCELL
IND360 Precision
IND360 Analog
Read/Write
Data Type
Command Device Address
Fast Feed IND360 FillDose 43014 RW Float 32 x
Heel Weight IND360 FillDose 43016 RW Float 32 x
+Tolerance IND360 FillDose 43018 RW Float 32 x
-Tolerance IND360 FillDose 43020 RW Float 32 x
Upper Limit IND360 FillDose 43026 RW Float 32 x
Lower Limit IND360 FillDose 43028 RW Float 32 x
Container Tare Max. IND360 FillDose 43022 RW Float 32 x
Container Tare Min. IND360 FillDose 43024 RW Float 32 x
Inhibit Time IND360 FillDose 43030 RW Float 32 x
Stable Time IND360 FillDose 43032 RW Float 32 x
Complete Time IND360 FillDose 43034 RW Float 32 x
Control Timeout IND360 FillDose 43036 RW Float 32 x
Process Timeout IND360 FillDose 43038 RW Float 32 x
Self Learning Mode IND360 FillDose 43040 RW Short x
Spill Adjust Period IND360 FillDose 43041 RW Short x
Spill Adjust Factor IND360 FillDose 43042 RW Float 32 x
Spill Adjust Range IND360 FillDose 43044 RW Float 32 x
Control Reliability IND360 FillDose 43047 RW Short x
Learning Samples IND360 FillDose 43049 RW Short x
Adjust Factor IND360 FillDose 43050 RW Float 32 x
Filling Dosing Weight IND360 FillDose 43052 R Float 32 x
Min. Cycle Weight IND360 FillDose 43054 R Float 32 x
Max. Cycle Weight IND360 FillDose 43056 R Float 32 x
Total Cycle Weight IND360 FillDose 43058 R Float 32 x
Total Buckets IND360 FillDose 43060 R Long x
Valid Buckets IND360 FillDose 43062 R Long x
Alarm Status IND360 FillDose 43064 R Short x
Run Status IND360 FillDose 43065 R Short x
Control Command IND360 FillDose 43066 W Short x
30587511 | 05 | 02/2023 METTLER TOLEDO Standard Automation Interface Reference Guide, Transmitters A-15
To protect your product’s future:
Congratulations on choosing the quality and
precision of METTLER TOLEDO. Proper use
according to these instructions and regular
calibration and maintenance by our factory-trained
service team ensure dependable and accurate
operation, protecting your investment. Contact us
about a service agreement tailored to your needs
and budget.
www.mt.com
For more information
Mettler-Toledo, LLC
1900 Polaris Parkway
Columbus, OH 43240