Monitoring of the cleaning efficiency of washer-disinfectors according to ISO 15883 Monitoring of the cleaning efficiency of washer disinfectors according to ISO 15883
standardised test soil transparent cover standardised test soil transparent cover
clips clips
simulated joint area simulated joint area
1. For the routine and validation test, TOSI`s are clipped flat on the individual 1. For the routine and validation test, TOSI`s are clipped flat on the individual
instrument trays within the instrument load and the cycle to be tested is switched instrument trays within the instrument load and the cycle to be tested is switched
on. For detailed information regarding the routine and validation test please on. For detailed information regarding the routine and validation test please refer
refer to: The clean monitoring user guide. to: The clean monitoring user guide
2. To avoid any of the TOSI from being displaced by water pressure, make sure all 2. To avoid any of the TOSI from being displaced by the water pressure make sure
TOSI´s are safely clipped onto the wire trays or use the mini-tray if necessary. all TOSI´s are safely clipped onto the wire trays or use the mini-tray if necessary.
3. At the end of the cycle TOSI`s are removed and visually checked for residue. 3. At the end of the cycle TOSI’s are removed and visually checked for residue.
A positive result (0) is indicated by a completely cleaned TOSI. A positive result (0) is indicated by a completely cleaned TOSI.
4. In case of unsatisfactory results, please refer to the troubleshooting guide. 4. In case of unsatisfactory results, please refer to the troubleshooting guide.
5. Results should be recorded in the TOSI log book or according to the department 5. Results may be recorded in the TOSI log book or according to the department
protocol. protocol.
Note: Only completely clean instruments are safe. Residue left on a TOSI is an Note: Only completely clean instruments are safe. Residue left on a TOSI is an
indication of an inefficient cleaning process. indication of an inefficient cleaning process.
STORAGE: Store in closed protective pouch in TOSI box at 2°C - 25°C. STORAGE: Store in closed protective pouch in TOSI box at 2°C - 25°C
Keep away from light, heat and chemicals. Keep away from light, heat and chemicals
Shelf life: see imprint. Shelf life: see imprint
CONTENT: 12 tests. CONTENT: 12 tests