tmp3588122518401173628 Print
tmp3588122518401173628 Print
ELECTROMAGNETIC SPECTRUM
The field of spectroscopy deals with the study of absorption or emission of electromagnetic
radiation. We are familiar with the number of different types of electromagnetic radiations such as
cosmic rays, X-rays ,Ultra violet light (UV), visble light ,infra-red radiations ,radio waves radar
waves etc. All the electromagnetic radiations travel with the same velocity i,e. 3x1010 cm/sec but
differ from one another in the wave length of their waves .The arrangement of all types of
electromagnetic radiations in the order of their increasing wavelength or decreasing frequency is
called the electromagnetic spectrum.
Spectroscopy is the study of the interaction between matter and electromagnetic radiation- radiant
energy that displays the properties of both particles and waves. Several different
spectrophotometric techniques are used to identify compounds. Each employs a different type of
electromagnetic radiation.
Spectroscopy is based on the interaction between light and matter. Spectroscopy is the
measurement and interpretation of electromagnetic radiation absorbed or emitted when the
molecules or atoms or ions of a sample move from one energy state to another energy state.
UV- Visible spectroscopy is a type of absorption spectroscopy in which light of the ultra-violet and
visible region (200-780 nm) is absorbed by the molecule which results in the excitation of the
electrons from the ground state to a higher energy state. Absorption spectroscopy is a quantitative
chemical analysis method based on the Beer-Lambert Law. According to this law, the absorption
of visible light passing through a certain substance is proportional to the concentration of that
substance.
PRINCIPLE OF UV-VISIBLE SPECTROSCOPY
The principle of UV-Visible Spectroscopy is based on the absorption of ultraviolet light or visible
light by chemical compounds, which results in the production of distinct spectra. When the matter
absorbs the light, it undergoes excitation and de-excitation, resulting in the production of a
spectrum.
When matter absorbs ultraviolet radiation, the electrons present in it undergo excitation. This
causes them to jump from a ground state (an energy state with a relatively small amount of energy
associated with it) to an excited state (an energy state with a relatively large amount of energy
associated with it). It is important to note that the difference in the energies of the ground state and
the excited state of the electron is always equal to the amount of ultraviolet radiation or visible
radiation absorbed by it.
There are four possible types of transitions (π–π*, n–π*, σ–σ*, and n–σ*), and they can be ordered
as follows: σ–σ* > n–σ* > π–π* > n–π*. The absorption of ultraviolet light or visible light by a
chemical compound will produce a distinct spectrum that aids in the identification of the
compound.
Instrumentation:
A UV-visible spectrophotometer is an analytical instrument that measures the amount of UV and
visible light that is absorbed by a sample. UV-vis spectrophotometer works by passing a beam of
light through the sample and measuring the amount of light that is absorbed at each wavelength.
The amount of light absorbed is proportional to the concentration of the absorbing compound in
the sample.
The basic components of a spectrometer include: light source (UV and visible), monochromator
(wavelength selector), sample holder, detector, amplifier and computer device.
Light Source
Tungsten filament lamps and Hydrogen-Deuterium lamps are the most widely used and suitable
light sources as they cover the whole UV region. Tungsten filament lamp is used as visible light
source. Tungsten filament lamps are rich in red radiations; more specifically they emit the
radiations of 375 nm, while the intensity of Hydrogen-Deuterium lamps falls below 375 nm.
Detector
Generally, two photocells serve the purpose of the detector in UV spectroscopy. One of the
photocells receives the beam from the sample cell and the second detector receives the beam from
the reference. This results in the generation of alternating currents in the photocells. After the light
has passed through the sample, a detector is used to convert the light into a readable electronic
signal. Generally, detectors are based on photoelectric coatings or semiconductors.
Amplifier
The alternating current generated in the photocells is transferred to the amplifier. The main
purpose of the amplifier is to amplify the signals many times so we can get clear and recordable
signals.
Recording devices
Most of the time amplifier is coupled to a pen recorder which is connected to the computer. The
computer stores all the data generated and produces the spectrum of the desired compound.
V-Vis spectroscopy information can be presented as a graph of absorbance, on the vertical y axis
and wavelength on the horizontal x axis. This graph is typically referred to as an absorption
spectrum an example is shown in the Figure.
Applications of UV Spectroscopy
Detection of Impurities
1. It is one of the best methods for the determination of impurities in organic molecules.
2. Additional peaks can be observed due to impurities in the sample and it can be compared with that
of standard raw material.
3. By also measuring the absorbance at a specific wavelength, the impurities can be detected.
FLAME PHOTOMETRY
Introduction
During 1980s Bowling Barnes, David Richardson, John Berry and Robert Hood developed an
instrument to measure the low concentrations of sodium and potassium in a solution. They named
this instrument as Flame photometer. The principle of flame photometer is based on the
measurement of the emitted light intensity when a metal is introduced into the flame. The
wavelength of the colour gives information about the element and the colour of the flame gives
information about the amount of the element present in the sample. It is also known as flame
emission spectroscopy. Flame photometer can be used to determine the concentration of certain
metal ions like sodium, potassium, lithium, calcium and cesium etc.
Instrumentation
A simple flame photometer consists of the following basic components:
Source of flame: A Burner in the flame photometer is the source of flame. It can be maintained in
at a constant temperature. The temperature of the flame is one of the critical factors in flame
photometry.
Atomizer: It is used to send homogeneous solution into the flame at a balanced rate.
Optical system: The optical system consists of convex mirror and convex lens. The convex mirror
transmits the light emitted from the atoms. Convex mirror also helps to focus the emissions to the
lens. The lens helps to focus the light on a point or slit.
Colour filters: The reflections from the mirror pass through the slit and reach the filters. Filters
will isolate the wavelength to be measured from that of irrelevant emissions.
Photo-detector: The intensity of radiation emitted by the flame is measured by photo detector.
Here the emitted radiation is converted to an electrical signal with the help of photo detector. These
electrical signals are directly proportional to the intensity of light.
Result:
Volume of the test solution = ......................... ml
Amount of Sodium ion in the given test solution = ......................... mg.
POTENTIOMETRY
Electrochemistry is a part of chemistry, which determines electrochemical properties of
substances. Potentiometry is the method to find the concentration of solute in a given solution by
measuring the potential between two electrodes. It involves the measurement of the potential of the
indicator electrode and reference electrode. It gives more accurate and precise results than other
titrations in which different reagents are used as indicators.
Potentiometry Principle
The principle involved in the potentiometry is when the pair of electrodes (reference electrode and
indicator electrode) is placed in the analyte (sample solution) it shows the potential difference by
the addition of the titrant or by the change in the concentration of the ions. The indicator electrode
measures this potential difference. The reference electrode has a potential value and remains stable
when dipped into the sample solution. The salt bridge is used to prevent the interference of the
analyte solution with that of reference solution. Here analyte solution is the solution whose
potential is to be measured. The electromotive force of the complete cell is given by the following
equation-
Ecell = E reference + E indicator + E junction
Where, Ereference is the electrode potential of the reference electrode, Eindicator is electrode potential
of indicator electrode, E junction is the electrode potential at the junction of the liquid.
Instrumentation of potentiometer
Potentiometer consists of two electrodes, whose reduction potential varies when it is inserted in a
test solution. The voltmeter is connected to the electrodes to measure the potential difference
between them. It involves the measurement of potential of a suitable indicator electrode as a
function of the volume of titrant added. The equivalence point of the titration is detected by rapid
changes in cell emf.
Theory of Potentiometry
The electrode potential is developed and given by Nernst equation whenever metal ion immersed
in a solution containing its own ions Mn+.
The value of E can be established by linking the calomel electrode and measuring the EMF of the
resultant cell. Where, E is the potential of the solution; E₀ is the standard electrode potential; n is
the valency of the ions; c is the concentration of the sample solution; 0.0591 is the value obtained
from the RT/F; where R is the gas constant, T is the temperature in Kelvin, F is the Faradays
constant; [Mn+] is the concentration of metal ions present around the electrode.
A suitable reference and an indicator electrode are required. The solution to be titrated is contained
in a beaker and a titrant is added from the burette into a titration vessel. Till end point the amount of
titrant is added successively to the container containing the solution. The potential is noted after
each addition. After each addition a sufficient time should be allowed to reach constant potential
before the next increment. The value of potential or emf is measured by the pH meter or with an
automated recording device. Sudden change in potential in the plots of emf against volume of
titrating solution reveals the end point. A typical potentiometric curve is represented as below:
Advantages of Potentiometry
It requires less quantity of substances and is an inexpensive method.
Turbid, coloured solutions can be titrated.
Mixture of solutions or very dilute solutions can be titrated.
The results are more accurate and reproducible.
Titrations can be carried out without indicators.
Disadvantages of Potentiometry
Calibration should be done on a regular basis.
Sensitive in ionic strength.
Contamination during electrolyte preparation may cause deflection from actual results.
Applications of Potentiometry
It is used in agriculture for the detection of different elements in soils, fertilizers etc.
Paper manufacturing: Atmospheric emissions from the pulp and paper manufacturing industry
can be determined by the potentiometric measurements.
Pharmaceutical industry: To determine the pH of given chemical agents and also in the
detection of end point in potentiometry titration of certain drugs like amoxicillin, propranolol.
Determination of pH of blood for diagnosis of acidosis or alkalosis.
Determination of NO3, NO2 in meat preservatives.
To determine the NaCl content in meat, fish, dairy products and fruit juice.
To determine the Ca content in dairy products, wines and brewing solutions.
CONDUCTOMETRY
Principle
The electrical conductance is in accordance with the ohm′s law which states the strength of current
passing through a conductor (in this case electrolyte solution) is directly proportional to the
potential difference applied across the electrodes and inversely proportional to the resistance
offered by the conductor.
i.e., i=V/R
Where, i= current; V= potential difference; R= resistance (ohms Ω).
During a titration process, one ion is replaced with another and the difference in the ionic
conductivities of these ions directly impacts the overall electrolytic conductivity of the solution.
It can also be observed that the ionic conductance values vary between cations and anions. Finally,
the conductivity is also dependent upon the occurrence of a chemical reaction in the electrolytic
solution.
Instrumentation of Conductometry
It consists of conductivity cell having two platinum electrodes placed at 1 cm apart and each
having a unit area of cross section. The measurement of conductivity is done by using an
instrument called a conductometer. This instrument also measures the resistance of a solution to
the flow of electrical current. The conductivity of a solution is calculated by taking the inverse of
the resistance.
The instrumentation used in conductometry includes the following components:
1. Conductivity Cell: The conductivity cell is a device that measures the conductivity of a
solution. It comprises two electrodes immersed in the solution. The distance between the
electrodes and the area of the electrodes are important factors that affect the accuracy of the
measurement.
2. Conductometer: It is an instrument that measures the conductivity of a solution. It applies a
voltage across the electrodes and measures the current that flows through the solution. The
conductometer also measures the temperature of the solution because the conductivity of a
solution is temperature dependent.
The end-point of the titration process can be determined by means of conductivity measurement.
For a neutralization reaction between an acid and a base, the addition of the base would lower the
conductivity of the solution initially. This is because the H+ ions would be replaced by the cationic
part of the base.
After the equivalence point is reached, the concentration of the ionic entities will increase. This, in
turn, increases the conductance of the solution. Therefore, two straight lines with opposite slopes
will be obtained when the conductance values are plotted graphically. The point where these two
lines intersect is the equivalence point (end point).
Advantages of conductometric titration
This process is very useful in the titrations of very dilute solutions and weak acids.
The end-point of this method of titration is very sharp and accurate when compared to a few other
titration processes.
This type of titration is applicable for solutions that are coloured or turbid.
Conductometric titration has numerous applications in acid-base titrations, redox titrations,
precipitation titrations, and complex titrations.
Disadvantages of conductometric titration
Only a few specific redox titrations can be done with the help of this process. This is because the
conductivity of the solution is masked by relatively high hydronium ion concentration.
The accuracy of conductometric titration is low when the concentrations of the electrolyte are
high, making the titration process unsatisfactory.
Application of conductometry
Quality control in chemical and pharmaceutical manufacturing.
Monitoring water quality and wastewater treatment.
Analyzing acids, bases, and salts in research.
Real-time tracking of chemical reactions.
Measuring ion concentrations in food and drinks.
Diagnosing medical conditions using body fluids.
Studying biomolecule behavior, like proteins and DNA.
Analyzing geological samples for composition.
Determining solution acidity or alkalinity.
SENSORS
Introduction
Sensor is a device that produces an output signal for the purpose of sensing of a physical
phenomenon. This signal must be produced by some type of energy, such as heat, light, motion, or
chemical reaction.
Once a sensor detects one or more of these signals (an input), it converts it into an analog or digital
representation of the input signal that is readable.
Each sensor is based on a transduction principle - conversion of energy from one form to another.
A Typical example is mercury used in a glass thermometer that measures the temperature and
converts it into the expansion of a liquid.
All sensors operate on the basic principle of taking an input and producing a related output. The
steps involved are:
Receptors: The receptor section senses the input phenomena such as temperature, light, or
motion.
Transduction: The input is converted into another form of energy via transduction. For e.g,
thermal energy is converted into electrical energy.
Signal Conditioning: The transduced signal is amplified and processed via filtering and digital
conversion circuits.
Output: The conditioned signal is converted into a human-readable output format like voltage,
current, frequency, digital code, etc.
Feedback: Some sensors also incorporate feedback loops to improve accuracy and compensate
for environmental changes.
In essence, a sensor detects an input phenomenon, transduces it, and generates an output signal
quantifying the input amount or change.
Sensing Principles
Sensors operate on different physical principles depending on the type of input and required
application. Some key sensing principles include:
Mechanical
Sensors using deflection, stress, strain gauge, and changes in frequency or resistance of materials
detect mechanical quantities.
Thermal
Thermal sensors rely on temperature-dependent characteristics of materials to sense heat or cold.
Examples are thermistor and thermocouples.
Electrical
Electric and electrochemical sensors use electrical properties and interactions to detect electric
field, voltage, current, charge, and conductivity changes.
Magnetic
Sensors like magneto resistors, Hall Effect sensors, and compass utilize the magnetic field and
flux density near magnetic materials.
Radiant
Light, infrared, and radiation sensing are based on photoconductivity, photoemission, and
bolometric effects of materials on radiation exposure.
Chemical
1. Thermocouples: Thermocouples consist of two different metal wires joined at one end. When
there is a temperature difference between the junction and the other end, it generates a voltage
proportional to the temperature difference. Thermocouples are robust, versatile, and suitable for
high-temperature applications.
2. Resistance Temperature Detectors (RTDs): RTDs are temperature sensors made of pure metals
or metal alloys whose resistance changes predictably with temperature. The most common RTD
material is platinum, known for its stable and linear resistance-temperature characteristics.
3. Thermistors: Thermistors are temperature-sensitive resistors made from semiconductor
materials. Their resistance decreases exponentially with increasing temperature (NTC
thermistors) or increases exponentially with increasing temperature (PTC thermistors).
Thermistors are highly sensitive and suitable for precise temperature measurement in a limited
temperature range.
4. IC Temperature Sensors: Integrated circuit (IC) temperature sensors utilize the temperature-
dependent voltage or current characteristics of semiconductor devices to measure temperature.
They are compact, low-cost, and suitable for applications where space is limited.
HVAC Systems: Temperature sensors are crucial components in heating, ventilation, and air
conditioning (HVAC) systems for maintaining comfortable indoor temperatures.
Industrial Processes: They are used in industrial processes such as manufacturing, chemical
processing, and food production to monitor and control temperature for quality control and
process optimization.
Medical Devices: Temperature sensors are employed in medical devices such as thermometers,
incubators, and medical imaging equipment for patient monitoring and diagnosis.
Automotive: In vehicles, temperature sensors are utilized for engine temperature monitoring,
climate control systems, and cabin comfort control.
b. OPTOELECTRONIC SENSORS
Optoelectronics is the communication between optics and electronics which includes the study,
design and manufacture of a hardware device that converts electrical energy into light and light
into energy through semiconductors. This device is made from solid crystalline materials which
are lighter than metals and heavier than insulators. Optoelectronics device is basically an
electronic device involving light. This device can be found in many optoelectronics applications
like military services, telecommunications, automatic access control systems and medical
equipments.
1. Photodiode
2. Solar Cells
3. Light Emitting Diodes
4. Optical Fiber
5. Laser Diodes
1. LEDs could become the next generation of lighting and used anywhere like in indication lights,
computer components, medical devices, watches, instrument panels, switches, fiber-optic
communication, consumer electronics, household appliances, traffic signals, automobile brake
lights, 7 segment displays and inactive displays.
2. The solar cells are applicable in rural electrification, telecommunication systems, ocean
navigation aids, and electric power generation in space and remote monitoring and control
systems.
3. Photodiodes are used in many types of circuits and different applications such as cameras,
medical instruments, safety equipments, industries, communication devices and industrial
equipments.
4. Optical fibers are used in telecommunications, sensors, fiber lasers, bio-medicals and in many
other industries.
5. The laser diodes are used in fiber optic communication, optical memories, military applications,
CD players, surgical procedures, Local Area Networks etc.
c. Piezoelectric Sensor
Piezoelectric sensors are electronic components that are able to convert a mechanical or thermal
input into an electrical signal. They use the piezoelectric effect to measure the electrical potential
caused by applying mechanical force to a piezoelectric material. They are based on the principle of
electromechanical energy conversion and primarily measure force, as well as other quantities such
as pressure, acceleration, temperature, and strain by converting the acquired data to an electrical
charge.
Piezoelectric sensors can be found in both analog and digital forms. Analog sensors produce a
continuous signal that varies with the force applied, while digital sensors convert this signal into
readable data for easy analysis.
The piezoelectric sensor can be classified into four types: piezoelectric force sensor, piezoelectric
pressure sensor, piezoelectric acceleration sensor, and polymer pressure sensor.
Piezoelectric sensors finds many applications, from musical instruments like electric guitars to
industrial machines, medical devices, and everyday items like microwave ovens, where they help
detect touch or vibration.
Electrochemical sensors work by reacting with the gas of interest and producing an electrical
signal proportional to the gas concentration. Consisting of two electrodes (a working electrode and
a counter electrode), the sensor operates by allowing charged molecules to pass through a thin
layer of electrolyte. The electrode will either oxidize or reduce the analyte of interest. The current
that is produced from the reaction is monitored and used to calculate important data such as
concentrations from the sample.
e. BIOSENSORS
Definition: A Biosensor is an analytical device which uses enzymes, antibody, tissues that
converts biological response into electrical, thermal or optical signals.
Biosensor consists of two components: Sensing element and transducers. Sensing element may be
either enzymes, antibodies, DNA, tissues or whole cells which then transduces the biochemical
reaction into electrical signals.
Types of Biosensors
The different types of biosensors are classified based on the sensor device as well as the biological
material
1. Electrochemical Biosensor
2. Thermal/Calorimetric Biosensor
3. Magnetic Biosensor
4. Optical Biosensor
5. Piezo-electric Biosensor
6. Resonance Biosensor
Applications of Biosensors
f. GAS SENSORS
Gas sensors are electronic devices that detect and identify different types of gasses. They are
commonly used to detect toxic or explosive gasses and measure gas concentration. Gas sensors are
employed in factories and manufacturing facilities to identify gas leaks, and to detect smoke and
carbon monoxide in homes.
Gas sensors also require a sample and receptor to acquire and analyze measurements; in
semiconductor gas sensors, gas interacts with the metal oxide surface and the subsequent analysis
is made on the basis of whether that gas is reduced or oxidized.