Ic - gc.ca-TE-LP-003Laboratory Practice For The Selection Calibration and Use of Direct Reading Thermometers
Ic - gc.ca-TE-LP-003Laboratory Practice For The Selection Calibration and Use of Direct Reading Thermometers
Version: 7
Record of changes
Table of contents
1 Introduction
1.1 Purpose
The purpose of this document is to provide recommended practice for the selection,
calibration and use of direct reading digital thermometers that will be used for calibration
and inspection work conducted by Measurement Canada (MC) staff or accredited or
1/10
registered organizations. The document also serves as a framework and guideline for the
technical evaluation of organizations providing thermometer calibration data to MC.
1.2 Scope
This recommended practice applies to direct reading electronic digital thermometers used
for calibration and inspection work under the Weights and Measures Act and Regulations
and the Electricity and Gas Inspection Act and Regulations.
To ensure the above tolerances are met, the following uncertainty components are
applied:
Class Application
2/10
Table 1: Classification of digital thermometer
operating conditions
The thermometer must be designed to be stable to within ± 0.1 °C for the duration of the
verification period.
The thermometer measurement variations due to ambient conditions, for the rated
temperature range, must be within ± 0.1 °C. This performance characteristic may be
derived from:
The temperature sensor must be designed and constructed to operate without leakage or
deformation over the expected range of operating pressures, temperatures and
environmental conditions, including the immersion depth of the sensor in liquid when
required.
3/10
diameter is 3 mm (or 1/8 inch) and the recommended lead wire length is approximately 6
feet.
Thermometer probes intended for field applications must be able to withstand thermal
shocks of 50 °C without loss of accuracy.
Thermometers intended for field use must be equipped with a carrying case which is
water resistant, dustproof and dimensionally stable over the range of environmental
conditions expected during the service life of the device.
2.3.6 Construction
The construction must be mechanically and electrically sound. The materials, finish, etc.,
must be of a quality to provide assurance of long life and sustained accuracy.
When power is supplied by a battery, the thermometer must be equipped with a low
power indication.
2.3.8 Markings
The following information must be indelibly marked on, or be capable of being displayed
by, the instrument:
1. manufacturer's name
2. model number
3. serial number
In the case of thermometers with multiple sensors, each sensor must be marked by a
unique identifier.
3 Calibration
Calibration includes the following operations:
4/10
3.1 General requirements for the reference thermometer and calibration
equipment
When used for the calibration and certification of thermometers as measuring apparatus
under the Electricity and Gas Inspection Act and Regulations, the calibrator's reference
thermometer must be traceable to a national reference standard maintained by National
Research Council of Canada.
Measuring range
The reference thermometer must not be used outside its certified measuring range.
Monitoring
The reference thermometer must be monitored for stability with an ice bath, water triple
point cell or other suitable methods with similar accuracy and stability. Monitoring must
be conducted on a monthly basis as a minimum.
Confirm that the thermometer model meets the specification outlined in section 2.
The thermometer is inspected and the inspection results as well as the seal condition are
recorded.
Before beginning the calibration procedure, the thermometer under calibration and the
measurement standard is preconditioned at ambient conditions for at least 12 hours.
Ambient temperatures in the calibration laboratory are maintained within 23 °C ± 5 °C.
The thermometer sensors are sufficiently immersed in the temperature bath to address
immersion errors. If this cannot be achieved, immersion error is considered in the
estimate of the uncertainty.
Temperature of the temperature bath is allowed to reach the set point and stabilized
before readings are taken.
5/10
As Found results, at a minimum of two temperature points, are obtained prior to
adjustments. It is recommended that the ice point be included.
In the case of thermometers that cannot display measured resistance and are adjusted
based on measurement errors, a suitable deviation function is selected for the
thermometer. The deviation function is the equation used to model the expected
thermometer corrections over the calibration range. Often a linear correction deviation
function will be suitable when the thermometer error can be characterized by a slope and
offset. Second and third order equations can also be used for the deviation function where
the thermometer error is not expected to be linear.
Each transducer associated with the thermometer is calibrated at a minimum of six (6)
evenly distributed test points across its measuring range or, as a minimum, the range
given in Table 2. Where a nonlinear deviation function or calibration equation is used to
model the thermometer response, a minimum of three (3) test points per equation
variable is used.
The coefficients of the deviation function or calibration equation, for each sensor, are
determined by the least squares fit.
In the case of thermometers that are to be certified for use with corrections, the deviation
function is used to generate the correction table for the sensor.
The measurement uncertainty of the thermometer calibration results must include the
sources of uncertainty described in sections 3.3.1 to 3.3.5.
The sources of uncertainty for the reference thermometer will include the uncertainty due
to:
6/10
3.3.2 Uncertainty due to the stability and uniformity of the temperature bath
3.3.3 Uncertainty due to the linearity and repeatability of the thermometer under
calibration
This section applies to thermometers that have a resolution ≤ 0.01 °C and are to be used
without a correction table. The maximum bias error, Errorbias, is taken as the maximum
measured error over the measuring range. The maximum bias error must be less than
±0.05 °C.
Where:
7/10
3.3.8 Expanded thermometer measurement uncertainty
The thermometer measurement uncertainty is obtained by the root sum square of the
calibration measurement uncertainty; see section 3.3.6, and the uncertainty contributions
of the relevant uncertainty sources such as the thermometer stability and susceptibility to
environmental conditions.
The thermometer must not be certified if the thermometer indication error, at one or
more calibration points, exceeds ± 0.75 °C.
3.6 Sealing
Thermometers that have physical access to any means of adjustment must be protected by
a seal to indicate any adjustments that can affect the thermometer calibration.
The values of the programmable parameters that can affect measurement accuracy of the
thermometers must be included on the calibration certificate.
8/10
2. name and address of the owner of the instrument
3. a description and unambiguous identification of the instrument
4. date of the calibration
5. range of the calibration
6. estimate of calibration measurement uncertainty
7. traceability of the thermometer standard
8. correction table (as applicable)
9. name and address of the laboratory which performed the calibration
10. any special limitations for the use of the instrument
11. the expiry date of the verification certificate
12. title: calibration certificate
13. certificate number
14. page identifier
15. identification of the calibration method used
16. the signatory's name and signature
17. the value of programmable parameters that can affect measurement accuracy of the
thermometers, such as Callendar-Van Dusen coefficients or other
18. when the thermometer is received with a broken seal, a note to this effect
The thermometer must not be used outside the rated ambient temperature for which it is
certified.
The owner is responsible to ensure the thermometer is suitably rated for the hazardous
location classification as required.
Unless otherwise specified in this document, the thermometer must be used as per the
manufacturer's recommendations.
Temperature measurement errors due to non-ideal location of the probe and dynamic
effects will be taken into consideration when reviewing the suitability of a thermometer
for a measurement application.
9/10
Appendix: Designation of local standard request
Thermometer model
Calibration report/certificate #
Signed by
Accredited laboratory
10/10