Dfu spySCOPE-ds
Dfu spySCOPE-ds
Rx Statement
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.
Prior to use, please refer to all applicable “Instructions for Use” for more information on Intended Use/Indications for Use,
Contraindications, Warnings, Precautions, Potential Adverse Events, and Operator’s Instructions.
REUSE WARNING
For single use only. Do not reuse, reprocess or resterilize. Reuse, reprocessing or resterilization may compromise the structural
integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness or death. Reuse, reprocessing
or resterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including,
but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to
injury, illness or death of the patient.
CONTRAINDICATIONS
Contraindications associated with the use of this device include:
• Patients for whom ERCP is medically contraindicated.
• Contraindications specific to endoscopic pancreatico-biliary duct exploration and cannulation.
WARNINGS
• Do not use the SpyScope DS II Catheter in the presence of uncontained flammable fluids or gases such as detergents,
anesthetics, nitrous oxide (NO), or oxygen. Doing so can result in fire and burns to the operator and patient.
• Do not insert the SpyScope DS II Catheter through the duodenoscope into the duodenum without a clear endoscopic field of
view. Doing so can result in patient injury such as perforation, hemorrhage, or mucous membrane damage.
• Do not perform therapy when an accessory is outside the field of view or force the distal end of the SpyScope DS II Catheter
against the mucosa. Doing so can result in patient injury such as perforation, hemorrhage, or mucous membrane damage.
• Do not use irrigation tubing without a single-use, one-way valve in place to prevent backflow. Doing so can result in
contamination of the device and/or cause patient infection or cross-infection.
• Do not look directly into the light emitted from the SpyScope DS II Catheter. Doing so can result in eye injury.
• The face of the cable remains hot for a period of time after disconnection from the controller. Do not touch the face of the cable
connector immediately after removing it from the controller. Doing so can result in a skin burn.
• If using the Y-port adapter, open the Y-port adapter before back-loading over a guidewire to ensure that guidewire is not
pushed further into the anatomy resulting in perforation.
• The SpyScope DS II Catheter is not intended to be used with RF cutting/coagulation devices.
• Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper
operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating
normally.
ENDO-301306-AC
PRECAUTIONS
• Note: SpyScope DS II Catheter (M00546610) is compatible with Controllers with software revision 2.2 only.
• Note: The working channel of the SpyScope DS II Catheter should be flushed with saline after the use of contrast media. If the
channel is not flushed, accessory devices may not pass through the channel and tip articulation may be reduced.
• Only use the SpyScope DS II Catheter in conjunction with the SpyGlass DS Digital Controller. Connection to other devices
may cause device or property damage or operator injury.
• Excessive bending of the articulation portion of the SpyScope DS II Catheter with the elevator of the duodenoscope can break
or kink the articulation portion. Do not bend the articulation portion excessively with the duodenoscope elevator.
o Note: If breakage or kinking of the SpyScope DS II Catheter is confirmed under X-ray, stop using the SpyScope DS II
Catheter immediately.
• Activating a laser or electrohydraulic lithotripsy (EHL) generator within close proximity of the SpyScope DS II Catheter distal
end can damage the distal end.
• Consult the instructions for use of the laser or EHL manufacturer for the appropriate distance between the laser fiber or EHL
probe and the SpyScope DS II Catheter distal end. At a minimum, ensure the laser fiber or EHL probe is extended at least 2 mm
(0.08 in.) beyond the distal end before actuating the laser or EHL.
• Disconnecting the catheter cable from the controller before removing the insertion portion from the duodenoscope will result in
a loss of visualization. Remove the SpyScope DS II Catheter from the duodenoscope before unplugging the cable.
• Damaging the face of the catheter cable connector can result in no visualization or an unexpected loss of visualization. Handle
the cable with care and inspect the face of the catheter cable connector for damage before use.
• Using a cardiac defibrillator while a SpyScope DS II Catheter remains in a patient can damage the controller. Remove the
SpyScope DS II Catheter before using the defibrillator.
• The SpyScope DS II Catheter should be used with caution in patients with previous gastric or bile duct surgery, or with ductal
strictures. These conditions may prevent passage of the SpyScope DS II Catheter.
• Do not insert a wet connector cable plug into the SpyGlass DS Digital Controller as poor video performance or damage to the
system may result.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the SpyGlass DS II System, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
• The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer
adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as
relocating or re-orienting the equipment
ENDO-301306-AC