Manual s220t8
Manual s220t8
Shop online at
www.huato.cn
E-mail: [email protected]
For latest product manuals:
www.omegamanual.info
S220-T8 Series
Multi-channel Handheld Thermocouple
Temperature Data Logger
www.huato.cn [email protected]
The information contained in this document is believed to be correct, but HUATO accepts no liability for any errors it contains, and reserves
the right to alter specifications without notice.
Table of
S220-T8 Series Contents
Table of Contents
Section ...................................................................................... Page
Section 1 Introduction ........................................................................................ 1
1.1 Introduction ................................................................................. 1
1.2 Features ........................................................................................ 1
1.3 Applications ................................................................................. 1
1.4 Data Logger Model...................................................................... 1
1.5 Temperature Data Logger Appearance ................................... 2
1.6 Temperature Data Logger Screen ............................................. 2
1.7 Button function instructions ....................................................... 3
Section 1 - Introduction
S220-T8 thermocouple temperature data logger is a kind of high precision
instrument, developed and manufactured by HUATO company, which have
pass all strict calibrations and professional testing via high precision instrument.
Support 8 types thermocouples(KJETRSNB), including
thermocouple temperature compensation function and can measure temperature
from -200 to 1800°C.
NOTE:
There are no user serviceable parts inside your unit. Attempting to repair or service your
unit may void your warranty.
1
Introduction 1
T1 T2 ℃
11
T3 T4
1 T5 T6
T7 T8
14
5
2 6
7 12
10
T1 T2 T3 T4 T5 T6 T7 T8
8 13
3
9
4
T1 T2 ℃ 9
1 ℉
2
MAX MIN T3 T4 10
3
4 HOLD
5 T5 T6
6
7
8 T7 T8
2
1 Introduction
Power ON/OFF
1When logger power is off, push this button will turn on the logger.
2When logger power is on, push this button will turn off the logger.
Recording Function
1When logger power is on, press this button, the LCD screen will show “LOG”
a sign which indicates that the logger is recording data.
2Press the button for 3 seconds will put the logger in standby mode.
3The screen will display “FULL” symbol when storage space is full.
Backlight Button
Press this button to open the backlight, and press again to close the light.
3
Installation and Instruction 2
6 3 7
4 8
1. The Windows® operating system cannot handle USB devices being unplugged and plugged back too fast. When
unplugging the logger, wait for about 5 seconds before plugging it in again. If you unplug and plug back a device
too quickly, the computer may stop recognizing any USB devices on that port. If this happens you will have to
restart the computer. This is a Windows® USB problem and is not related to LogPro.
5
Installation and Instruction 2
Calibrate the logger: The logger is factory calibrated to an accuracy given in the device specifications. However, there may be times when you wish to
adjust the calibration of your logger. Logpro provides you with the ability to perform a single point offset calibration. This calibration can be used to
increase the accuracy of the logger for a restricted data range.
2 Installation and Instruction
T1 T2 ℃
T3 T4
T5 T6
T7 T8
Start Logging - When the Logger is on and on stand-by (LCD does not have “LOG”
logo display), press "LOG " button LCD has “LOG” logo display now the logger start
logging and storing data.
T1 T2 ℃ T1 T2 ℃
T3 T4 T3 T4
T5 T6 T5 T6
T7 T8 T7 T8
Stop Logging - When the Logger is on logging, long press “LOG”, LCD display
countdown time, loosen “LOG” button when countdown time finished, then the
instrument switch from logging to standby state.
T1 T2 ℃ T1 T2 ℃ T1 T2 ℃
T3 T4 T3 T4 T3 T4
T5 T6 T5 T6 T5 T6
T7 T8 T7 T8 T7 T8
7
Installation and Instruction 2
Turn Off the Logger - Push the“ON / OFF” Button for short pressing then LCD
display is off.
T1 T2 ℃
T3 T4
T5 T6
T7 T8
LOG Full -The device won't record data while the memory is full, that you have to
download the data and delete it by Logpro software, then device can record again.
℃
℉
Switchable °C or °F - Press the “°C/°F” key shortly When instrument records data in
degrees (the LCD displays the °C symbol), and the instrument switches to Fahrenheit as
the recording unit display data (the LCD displays the °F symbol). It Can Switch back and
forth between °C and °F as you Press this button repeatedly.
T1 T2 ℃ T1 T2 T1 T2 ℃
℉
T3 T4 T3 T4 T3 T4
T5 T6 T5 T6 T5 T6
T7 T8 T7 T8 T7 T8
8
2 Installation and Instruction
12 13
Once the data is downloaded from the logger to the PC, the log graph will be displayed.
Tips:Press and hold the left mouse button to drag a box, when the left mouse button
is released, the graph will be redrawn with the data in the selected rectangle area. Click
right button, and then the graph will be redrawn with all the data in the logs file.
9
Installation and Instruction 2
The data pane lists the data samples collected by the logging device.
The column width of each column is adjustable by using the left mouse button
and dragging the column to the desired width.
10
2 Installation and Instruction
14
LogPro - HS220CW001_20181121_0918.logp
• Double click left mouse button, then the selected file is opened.
• Click right mouse button, a popup menu is shown as below, you can
rename or delete or log file.
• The saved file is saved in the LogPro software’s installation path in the PC’s
hard driver.
15
16
11
Attention 3
Section 3 - Attention
• Plastic shell is ABS, should be retardant, not resistant to acid and
alkali.
• USB power supply (4.8V) cannot let the Data Logger work.