User’s Guide

Shop online at
www.huato.cn
E-mail: [email protected]
For latest product manuals:
www.omegamanual.info

S220-T8 Series
Multi-channel Handheld Thermocouple
Temperature Data Logger
www.huato.cn [email protected]

HUATO HUATO ELECTRIC CO., LTD.


Headquarters: Toll-Free: 0755-29977605
Customer Service: 0755-29977605
Engineering Service: 0755-29977605
Fax: +86-755-29748856
E-mail: [email protected]

For Other Locations Visit www.huato.cn

The information contained in this document is believed to be correct, but HUATO accepts no liability for any errors it contains, and reserves
the right to alter specifications without notice.
Table of
S220-T8 Series Contents

Table of Contents
Section ...................................................................................... Page
Section 1 Introduction ........................................................................................ 1
1.1 Introduction ................................................................................. 1
1.2 Features ........................................................................................ 1
1.3 Applications ................................................................................. 1
1.4 Data Logger Model...................................................................... 1
1.5 Temperature Data Logger Appearance ................................... 2
1.6 Temperature Data Logger Screen ............................................. 2
1.7 Button function instructions ....................................................... 3

Section 2 Installation and Instruction ........................................................... 4


2.1 Software Installation ................................................................... 4
2.1.1 The Requirement of the Computer Hardware ............... 4
2.1.2 Install Driver & Software................................................... 4
2.2 How to Read and Configure the Logger ................................. 5
2.2.1 Read and Configure the Logger's Settings ..................... 5
2.2.2 Setting Parameters Description ........................................ 6
2.3 Turn On/Off Logger & Start/Stop Logging ............................ 7
2.4 Download the Logging Data ...................................................... 9
2.5 View Data Sheet ........................................................................... 10
2.6 Exporting Logs from LogPro ..................................................... 10
2.7 Delete Records in the Logger ..................................................... 11
2.8 Check the Save File in File List .................................................. 11

Section 3 Attention ............................................................................................. 12

Section 4 FAQ ...................................................................................................... 12


4.1 LCD Screen Dim .......................................................................... 12
4.2 Data & Time Error ........................................................................ 12
4.3 Software "Runtime Error" .......................................................... 12
4.4 Check COM Port Number ......................................................... 12
1 Introdu ct ion

Section 1 - Introduction
S220-T8 thermocouple temperature data logger is a kind of high precision
instrument, developed and manufactured by HUATO company, which have
pass all strict calibrations and professional testing via high precision instrument.
Support 8 types thermocouples(KJETRSNB), including
thermocouple temperature compensation function and can measure temperature
from -200 to 1800°C.

Section 1.2 - Features


• LCD screen which can display data from 8 channels simultaneously.
• Switchable °C/°F temperature units.
• Low battery indication.
• MAX, MIN and HOLD mode for all channels.
• Concise appearance, high precision and reliable performance.
• Desktop and wall-mounted.
• Accompanied by powerful software with a concise interface.
• Match professional analysis software.
• Battery can work for 3 months.
( collect one data each minute, record one data 5 minutes)

Section 1.3 - Application


• Monitor temperature in factories, laboratories, storage house, refrigerators.
• Non-contact infrared surface temperature measurements in hard to reach
areas such as heating and air conditioning ducts.

NOTE:

There are no user serviceable parts inside your unit. Attempting to repair or service your
unit may void your warranty.

Section 1.4 - Data Logger Model


Temperature Resolution Accuracy Work Capacity of
Model Environment
Range Records
-200~1800℃ 0.1℃ 0~80%RH
S220-T8 0.8±2‰℃ 86,000
-328~3272℉ 0.1℉ 0~50℃

1
Introduction 1

Section 1.5 - Temperature Data Logger Appearance

T1 T2 ℃
11
T3 T4
1 T5 T6

T7 T8

14
5
2 6
7 12
10
T1 T2 T3 T4 T5 T6 T7 T8

8 13
3
9
4

1. LCD Screen 8. MAX Displaying maximum recorded


2. Power On/Off value for all channels
3. LOGRecording function button 9. MIN Displaying minimum recorded
4. HOLDLock the Value. value for all channels
5. Screen Backlight 10. USB & 9V DC power interface
6. °C/°F Switch 11. Wall-mount hole
7. TYPESwitch for type of Sensor 12. Battery Cover
(KJETRSNB). 13. Back Stand
14 Thermocouple Sensor Interface

Section 1.6 - Temperature Data Logger Screen

T1 T2 ℃ 9
1 ℉
2
MAX MIN T3 T4 10
3
4 HOLD
5 T5 T6
6
7
8 T7 T8

1. Thermocouple Sensor Type KJ 5. USB wire is connected between logger


ETRSNB and computer
2. MAX: Logger running in maximum 6. Battery level
value mode. 7. DC Adapter (9V) connected
3. MIN: Logger running in minimum 8. LOG: logger is recording
value mode. 9. Temperature Unit: °C or °F
4. HOLD: Display is held. 10. Temperature display area

2
1 Introduction

Section 1.7 - Button function instructions

Power ON/OFF
1When logger power is off, push this button will turn on the logger.
2When logger power is on, push this button will turn off the logger.

Recording Function
1When logger power is on, press this button, the LCD screen will show “LOG”
a sign which indicates that the logger is recording data.
2Press the button for 3 seconds will put the logger in standby mode.
3The screen will display “FULL” symbol when storage space is full.

Switchover Button of ° C and °F

Backlight Button
Press this button to open the backlight, and press again to close the light.

Maximum Lock Button


Press this button once LCD shows “MAX” sign, which indicates that the logger is
recording maximum value for all channels; Press again to exit this mode.

Minimum Lock Button


Press this button once LCD shows “MIN” sign, which indicates that the logger is
recording minimum value for all channels; Press again to exit this mode.

Thermocouple Sensor Type Switch Button


Support 8 types of thermocouple sensors: JKETRSNBand press to
switch between different types of sensors.

Measured Values Keeping Button


Press this button once, and the LCD screen will display “HOLD” sigh which
indicates that the logger will keep the just measured value until the button is
pressed again.

3
Installation and Instruction 2

Section 2 - Installation and Instruction


This section will explain how to use the software to upload, read, configure and
synchronize the settings, how to turn on/off the logger & enter/exit the logging
mode, and how to delete and download the logging data.

Section 2.1 - Software Installation

2.1.1 - The Requirement of the Computer Hardware


1. OS: Windows XP/Vista/8/10 (32/64-bit).
2. CPU: 1.6GHz.
3. Physical Memory: 512MB.
4. Hard-drive Space: 4 GB.
5. USB Port: 1.

2.1.2 - Install Driver & Software


Note: "If need to install Driver separately, please install 64-bit driver for Windows
64-bit computer & 32-bit driver for Windows 32-bit computer".
1. Install LogPro Software and Driver together(Default Path: C:\S220-T8 Logger).
2. Installation Steps: When entering the software, click the ‘Install LogPro and Driver’
(as below Pic ), and then click the ‘Next ’(5 times), click the ‘Install,’ click the ‘Finish.’
When entering the LogPro software interface, the installation driver software
window will pop up. Click 'Install,' and the success of installation will be displayed.

LogPro installation page Default installation folder

The successful installed Install driver 4


2 Installation and Instruction

Section 2.2 - How to Read and Configure the Logger


2.2.1 - Read and Configure the Logger's Settings
1. Turn on and Insert the data logger to the computer¹.
2. Start LogPro software on the PC.
3. From the toolbar select Connect.
4. Load & Read the logger’s current settings.
5. Configure the new settings.
6. Sync the settings & time. (The PC time will be synchronized to the logger
as well.) .
7. From the toolbar select disconnect.
8. Sync the Computer's Data & Time to the Data Logger.
9. Unplug the logger from the computer, and then the logger is in Standby mode.

6 3 7

4 8

1. The Windows® operating system cannot handle USB devices being unplugged and plugged back too fast. When
unplugging the logger, wait for about 5 seconds before plugging it in again. If you unplug and plug back a device
too quickly, the computer may stop recognizing any USB devices on that port. If this happens you will have to
restart the computer. This is a Windows® USB problem and is not related to LogPro.

5
Installation and Instruction 2

2.2.2 - Setting Parameters Description


A.System E.Channel Name
Name S220-T8 Name1 1
SN HS220CW001 Name2 2
Type Data Logger Name3 3
Model S220-T8 Name4 4
Firmware H1103 Name5 5
B.Sampling Name6 6
Logging Interval Secs 10 Name7 7
Sampling Interval (Secs) 60 Name8 8
C.Storage
Logs 0
Total 86000
D.Offset
CH1 0.0
CH2 0.0
CH3 0.0
CH4 0.0
CH5 0.0
CH6 0.0
CH7 0.0
CH8 0.0

Setting Parameter Text Description


Name English letters or numbers Name of the logger
Must be the same as the one in the
SN 10 characters
label of logger
sampling Interval (secs) Number from 1 to 240 Sampling frequency in LCD screen
Logging Interval (secs) Number from 2 to 86400 Logging frequency
Logs Number from 0 to 86000 The count of records in the memory
Total Number 86000 The total capacity of the logger’s storage
Calibration mode: For example, pls fill
in 1.5 if the displayed temperature is
higher than the actual temperature of Input positive number to decrease the
Offset 1.5 ; fill in -1.5 if the displayed value.input negative number to increase
temperature is lower than the actual
temperature of 1.5 . If you need to
the value.
adjust the humidity, do the same, the
number field is -100~100.
Channel Name English Letters or Numbers Name of each channel

Calibrate the logger: The logger is factory calibrated to an accuracy given in the device specifications. However, there may be times when you wish to
adjust the calibration of your logger. Logpro provides you with the ability to perform a single point offset calibration. This calibration can be used to
increase the accuracy of the logger for a restricted data range.
2 Installation and Instruction

Section 2.3 - Turn On/Off Logger & Start/Stop Logging Under


Button to Start Mode
Turn On the Logger - Push “ON / OFF” Button for pressing the switch button
momentaril then LCD display is on.

T1 T2 ℃

T3 T4

T5 T6

T7 T8

Start Logging - When the Logger is on and on stand-by (LCD does not have “LOG”
logo display), press "LOG " button LCD has “LOG” logo display now the logger start
logging and storing data.

T1 T2 ℃ T1 T2 ℃

T3 T4 T3 T4

T5 T6 T5 T6

T7 T8 T7 T8

Stop Logging - When the Logger is on logging, long press “LOG”, LCD display
countdown time, loosen “LOG” button when countdown time finished, then the
instrument switch from logging to standby state.

T1 T2 ℃ T1 T2 ℃ T1 T2 ℃

T3 T4 T3 T4 T3 T4

T5 T6 T5 T6 T5 T6

T7 T8 T7 T8 T7 T8

7
Installation and Instruction 2

Turn Off the Logger - Push the“ON / OFF” Button for short pressing then LCD
display is off.

T1 T2 ℃

T3 T4

T5 T6

T7 T8

LOG Full -The device won't record data while the memory is full, that you have to
download the data and delete it by Logpro software, then device can record again.


Switchable °C or °F - Press the “°C/°F” key shortly When instrument records data in
degrees (the LCD displays the °C symbol), and the instrument switches to Fahrenheit as
the recording unit display data (the LCD displays the °F symbol). It Can Switch back and
forth between °C and °F as you Press this button repeatedly.

T1 T2 ℃ T1 T2 T1 T2 ℃

T3 T4 T3 T4 T3 T4

T5 T6 T5 T6 T5 T6

T7 T8 T7 T8 T7 T8

8
2 Installation and Instruction

Section 2.4 - Download the Logging Data


10. Insert the data logger into the computer.
11. Start LogPro software on the PC.
12. From the toolbar select Connect.
13. From the toolbar select Download.

12 13

Once the data is downloaded from the logger to the PC, the log graph will be displayed.

Tips:Press and hold the left mouse button to drag a box, when the left mouse button
is released, the graph will be redrawn with the data in the selected rectangle area. Click
right button, and then the graph will be redrawn with all the data in the logs file.

9
Installation and Instruction 2

Section 2.5 - View Data Sheet


Click “Data List” button on the toolbar, and then the data listing window
is shown below.

The data pane lists the data samples collected by the logging device.
The column width of each column is adjustable by using the left mouse button
and dragging the column to the desired width.

Section 2.6 - Exporting Logs from LogPro

• Export data list to an Excel file.

• Export data list to a PDF file.

• Export graph to a BMP file.

10
2 Installation and Instruction

Section 2.7 - Delete Records in the Logger


14. From the toolbar select Delete.

14
LogPro - HS220CW001_20181121_0918.logp

Logs Graph (SN:HS220CW001)

Section 2.8 - Check the Save File in File list


15. From the toolbar select File List.
16. Click to open the file.

• Double click left mouse button, then the selected file is opened.
• Click right mouse button, a popup menu is shown as below, you can
rename or delete or log file.
• The saved file is saved in the LogPro software’s installation path in the PC’s
hard driver.
15

16

11
Attention 3

Section 3 - Attention
• Plastic shell is ABS, should be retardant, not resistant to acid and
alkali.

• If repair is needed, only authorized technician can do the repair.

• The instrument configuration by 9V block battery is not rechargeable.

• USB power supply (4.8V) cannot let the Data Logger work.

FAQ 4

Section 4.1 - LCD Screen Dim


Insufficient battery or the environment temperature is too low or too high.
Solution: In the case of insufficient battery, please replace the battery. If results
from environment temperature, please immediately take the logger out of the
environment.

Section 4.2 - Data & Time Error


Reason:
• The Battery level is low.
• The Data Logger is not synchronous properties, before start recording.
Solution:
• Replace the battery.
• Please sync the properties, before start recording.
Section 4.3 - Software “Runtime Error”
Reason:
• OS forbid software creating files.
Solution:
• Run the program (software) as an administrator.
• Install software in D or E drive if available.
• A data logger name cannot contain any of the following characters: \ / :
*?“<>|
• Software’s installation path cannot include Chinese character or garbage
character.

Section 4.4 - Check COM Port Number


Press “Win” + “R” in keyboard -> Run “devmgmt.msc” to Open “Device
Manager” in Windows-> Expand “Ports (COM & LPT)” -> “USB-serial CH340
(COM No.)” is the Data Logger.