CT00-000 820 05 01 02
CT00-000 820 05 01 02
ty
u
lo
a
c
d
M
s
n
m
ie
S
Quality Assurance
Image display devices
10351301
07740769
05654269
10045692
08874427
08872017
08379492
08377520
08098951
07734713
07543106
07543015
07407039
07407013
07393114
07114023
07005502
04806118
03815490
03813933
03812950
03810160
03810152
03810137
03810079
03804890
© Siemens AG 2006
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
1. General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Overview of image display devices at the examination or diagnostic site . . . . . . . . . . 6
3. Measurement and test equipmentl used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Phone: Fax:
Tester: Name:________________________________
Date:__________ Signa-
ture:______________
The image display devices meet the requirements set forth in the respective norms and
regulations:
Yes❒ No❒
Remarks:_______________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
Remark: Use n.a. in response to non-applicable fields!
Remarks:_______________________________________________________________
______________________________________________________________________
______________________________________________________________________
* Measuring the veiling luminance (LS) when the monitor is switched off
Calculation, maximum contrast: MC = Lmax / Lmin
Limit value for application category A: MC ≥100
Limit value for application category B: MC ≥40
Remarks: __________________________________________________________
Test criteria:
• For SMPTE test = 5 % - field and 95 % - field visible?
image
• For test image 2 = Upper and lower three grayscale steps visible at equidistant
spacing?
• For other test
images
Monitor 1 Monitor 2
Bar pattern HC LC* (low HC NC*
modulation 100 % contrast) 100 % and/or
and/or 6,2 % 5 %
6,2 % 5 %
Bar pattern detectable Yes No Yes No Yes No Yes No
Horizontally
Center
Vertically
Upper left corner Horizontally
Vertically
Upper right corner Horizontally
Vertically
Lower left corner Horizontally
Vertically
Lower right corner Horizontally
Vertically
Remarks:_______________________________________________________________
_______________________________________________________________________
Monitor 3 Monitor 4
Bar pattern HC NC* HC NC*
modulation 100 % and/or 100 % and/or
6,2 % 5 % 6,2 % 5 %
Bar pattern detectable Yes No Yes No Yes No Yes No
Horizontally
Center
Vertically
Upper left corner Horizontally
Vertically
Upper right corner Horizontally
Vertically
Lower left corner Horizontally
Vertically
Lower right corner Horizontally
Vertically
Remarks:_______________________________________________________________
_______________________________________________________________________
Lcenter - Lcorner
Calculation: Difference in % = _______________ X 100
Lcenter
Remarks:_______________________________________________________________
______________________________________________________________________
_______________________________________________________________________
Requirements: Limit lines have to be fully visible in the images, limit lines have to
have the same length (O= U,L=R). Shifts, turns, and distortions are
not acceptable.
Remarks:_______________________________________________________________
______________________________________________________________________
Requirements: In the diagnostically important area, the test image cannot show one
of the errors described above.
Remarks:_______________________________________________________________
______________________________________________________________________
Physician responsible:
Remarks:_______________________________________________________________
______________________________________________________________________
List of abbreviations
• IDD: Image display device
• SMPTE Society of Motion Picture and Television Engineers
• HC: high contrast
• LC: low contrast
Lcenter
[cd/m2]
Lcorner UL
[cd/m2]
Lcorner UR
[cd/m2]
Lcorner LL (lower left)
[cd/m2]
Lcorner LR
[cd/m2]
Constancy tests 0
The constancy tests are described in the Monitor chapter included in the Operator Man-
ual.