0% found this document useful (0 votes)
42 views114 pages

CT02-025 812 03 03 02

Somatom Emotion New
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
42 views114 pages

CT02-025 812 03 03 02

Somatom Emotion New
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 114

h

tZ
u
lo
a
c
d
M
s
n
m
ie
S

Emotion (2007) / Spirit


CT

Installation Instructions
System

Unpacking/installing components and system wiring

This document is valid for:


SOMATOM Emotion 6 (2007)
SOMATOM Emotion 16 (2007)
SOMATOM Spirit Power
SOMATOM Spirit © Siemens AG 2006
10165977
10165888
10165566
10045692
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: CT02-025.812.03.03.02 English


Replaces: CT02-025.812.03.02.02 Doc. Gen. Date: 10.07
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 2 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Table of Contents 3
0Table of Contents

1 _______ General Information______________________________________________ 6

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Prerequisites for System Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Site Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Gantry Transport Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Abbreviations and Acronyms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Overview of System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
List of Required Installation Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Special Tools for CTDI measurements in accordance with EN 61223-2-6 ( for installations
in Germany and some EU countries) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Special Tools for the DIN & IEC Acceptance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Special Tools for the Monitor ( Germany and some EU countries ) . . . . . . . . . . . . . . . . 21

2 _______ Unpacking Components _________________________________________ 22

General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Gantry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Attaching the Gantry Transport Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Patient Table (PHS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Monitor / Keyboard / Mouse / Control Box / ICS, IRS & IES Towers . . . . . . . . . . . . . . . 29
Console / Container (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Accessories and Phantoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
(Line Connection Cabinet - 400V) optional. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Line Connection Box ( Emotion 6/16 ( 2007 ) ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

3 _______ Installation of Components ______________________________________ 35

General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Gantry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Positioning the Gantry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Removing the Transport Device and the 4 yellow Adapters . . . . . . . . . . . . . . . . . . . 38
Removing the Covers (Spirit) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Removing the Covers (Emotion 2007 / Spirit Power) . . . . . . . . . . . . . . . . . . . . . . . . 42
Removing the Transport Locks and Hand Brake. . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Removing the Transformer Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Leveling the Gantry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Patient Table (PHS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Tools required for installing the Liebig or Hilti Floor Anchors . . . . . . . . . . . . . . . . . . 50
Drilling the Holes for Floor Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Siemens AG CT02-025.812.03.03.02 Page 3 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
4 Table of Contents

Positioning the PHS (Patient Table). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53


Leveling the PHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Leveling the Gap below the vertical Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Installing the Cable Duct and Cable Cover between the Gantry and PHS . . . . . . . . . . . 60
Cable Routing below the Flooring without the Cable Duct . . . . . . . . . . . . . . . . . . . . 62
Line Connection Box (Emotion 6/16 (2007) ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Console / Container / Furniture (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Monitor / Keyboard & Mouse / Control Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
ICS / IRS tower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
UPS (Uninterrupted Power Supply). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Line Voltage Connection Cabinet (400V) /Mains Connection Cabinet(China) (Spirit / Spirit
power) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

4 _______ Special Mounting _______________________________________________ 72

General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Gantry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Bolting the Gantry to the Floor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Line Connection Box (Emotion 6/16 (2007)) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Bolting the LCB to the Floor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Mobile CT (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Bolting the Mobile CT Gantry to the Trailer Floor . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

5 _______ Electrical Connections and System Wiring __________________________ 77

General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Routing Fiber Optic Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Routing Signal and Data Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Cables to be prepared on-site (not in delivery volume) . . . . . . . . . . . . . . . . . . . . . . . 77
Cabling Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
List of System Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Cables for SOMATOM Spirit / Spirit Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Cables for Emotion 6/16 (2007). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Wiring Diagram: Standard Components for SOMATOM Spirit / Spirit Power . . . . . . 82
Wiring Diagram: Standard Components for Emotion 6/16 (2007) . . . . . . . . . . . . . . . 84
Cabling Information for SOMATOM Spirit / Spirit Power: . . . . . . . . . . . . . . . . . . . . . . 85
System Wiring and Cable Connections for SOMATOM Spirit / Spirit Power: . . . . . . . . . 86
Gantry (right stand from front side) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Gantry (right stand from the back): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Patient Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Computers ICS and IRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
ICS Computers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
IRS tower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
UPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 4 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Table of Contents 5

Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
System Wiring and Cable Connections for SOMATOM Emotion (2007) . . . . . . . . . . . 101
Line Connection Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Gantry (right stand from front side) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Gantry (right stand from the back): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Patient Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Computers ICS and IRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
ICS Computers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
IRS tower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
UPS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Line Connection Cabinet (400V) / Mains Connection Cabinet (China) . . . . . . . . . . . . 112

6 _______ Final steps ___________________________________________________ 113

7 _______ Changes to Previous Version ____________________________________ 114

Siemens AG CT02-025.812.03.03.02 Page 5 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
6 General Information
1-
Introduction
1General Information

This document describes the steps required to install the SOMATOM Emotion 6/16
(2007) / Spirit / Spirit Power system.
Every system has undergone intensive testing and adjustment procedures at the factory
and is delivered in its customer-specific configuration.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 6 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 7

Prerequisites for System Installation 1.1

Site Preparation 0

To ensure a smooth and quick system installation process, all site preparation activities
have to be completed prior to system delivery in accordance with the requirements in the
Planning Guide (Tab. 1 / p. 8).
Detailed information and requirements regarding site planning, room preparation, system
transport, and other installation issues are described in the Planning Guides
(Tab. 1 / p. 8). The latest version of the above documents will be distributed via the Sie-
mens Intranet only. These documents are NOT included in the system documentation.
The documents are located at the CS home page on the Intranet and may be accessed
via the hyperwave server:

http://cs.med.siemens.de

The local project manager is responsible for site preparation. The following conditions
have to be met:
• Room construction is finished, installation rooms are cleaned.
• Line power distribution is available.
• Internal line impedance has been checked in accordance with the Planning Guide
(Tab. 1 / p. 8).
• Radiation protection is available (e.g. walls, doors, lead glass window).
• Transport route for system components has been approved (width, height, access).

Gantry Transport Device 0

• The gantry transport device has to be available.

NOTE The gantry transport device is not part of the delivery volume for
all countries. Depending on country-specific regulations, it has to
be ordered separately prior to system delivery. The gantry trans-
port device is identical for the PLUS4, Volume Zoom, Sensation
and Emotion systems as well as Spirit=> Order number: 27 92 146

Siemens AG CT02-025.812.03.03.02 Page 7 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
8 General Information

Documents 1.2

Tab. 1 Print # of available documents

Name of document: Print number of document:


Planning Guide (PG) CT00-000.891.01.xx.02
Common CT
Basic requirements for CT project planning
(English)
Planning Guide (PG) CT02-025.891.01.xx.02
SOMATOM Spirit System
Specific requirements for CT project Plan-
ning
(English)
Planning guide (PG) checklist for electrical CT02-025.891.03.xx.02
installation
System CT02-025.815.02.xx.02
Startup of the CT system

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 8 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 9

Documents delivered with Spirit / Spirit Power / Emotion (2007) system:


• The following documents are delivered with Spirit systems ( for SW VB15 & VB27 ):
Tab. 2 Documents delivered with the Spirit system ( for Software / VB15 & VB27 )

Name of document binder: Print number of document binder:


Technical documents (blue binder) CT02-025.802.10.xx .02 (VB15)
includes: Installation Instructions, Acceptance CT02-025.802.11.xx .02 (VB27)
Test Instructions CT02-025.802.12.xx .02 (VB27)
available in English only
System Owner Manual (white binder) C2-025.629.08.xx (- DHHS)
includes: Customer and site-specific information available in German, French, Span-
ish, Chinese, Japanese, Danish, Ital-
ian, Polish, Portuguese, Greek,
Flemish, Latvian, Bulgarian, and Rus-
sian.

C2-025.629.09.xx (+ DHHS)
special version for USA
Operator Manual ( white binder)
Software: (available in English only) C2-025.626.07.xx
syngo OM (Volume 1)
C2-025.626.08.xx
syngo OM (Volume 2)
C2-025.626.09.xx
syngo OM (Volume 3)

C2-025.620.02.xx
Operator Manual

Hardware: (available in German, English,


French, Japanese and Spanish)
Release Notes C2-025.623.18.xx .

Siemens AG CT02-025.812.03.03.02 Page 9 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
10 General Information

• The following documents are delivered with Spirit & Spirit Power systems ( for SW
VB35 ):
Tab. 3 Documents delivered with Spirit & Spirit power system ( for Software VB35 )

Name of document binder: Print number of document binder:


Technical documents (blue binder) CT02-025.802.12.xx .02
includes: Installation Instructions, Acceptance
Test Instructions
System Owner Manual (white binder) C2-025.629.08.xx (- DHHS)
includes: Customer and site-specific information available in German, French, Span-
ish, Chinese, Japanese, Danish, Ital-
ian, Polish, Portuguese, Greek,
Flemish, Lativan, Bulgaria, Russia.

C2-025.629.09.xx (+ DHHS)
special version for USA
Operator Manual ( white binder)
C2-025.626.23.xx
syngo OM (Volume 1)
C2-025.626.24.xx
syngo OM (Volume 2)
C2-025.626.25.xx
syngo OM (Volume 3)
C2-025.626.26.xx
syngo OM (Volume 4)
C2-025.626.27.xx
syngo OM Security Package

C2-025.620.06.xx
Operator Manual
Release Notes C2-025.623.18.xx

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 10 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 11

• The following documents are delivered with Emotion (2007) systems:


Tab. 4 Documents delivered with Emotion (2007) systems

Name of document binder: Print number of document binder:


Technical documents (blue binder) CT02-025.802.12.xx .02
includes: Installation Instructions, Acceptance available in English only
Test Instructions
System Owner Manual (white binder) C2-025.629.05.xx (- DHHS)
includes: Customer and site-specific information available in German, English, French,
Spanish, Chinese and Japanese

C2-025.629.06.xx (+ DHHS)
special version for USA
Operator Manual ( white binder)
Software: (available in English only) C2-025.626.14.xx
syngo OM (Volume 1)
C2-025.626.15.xx
syngo OM (Volume 2)
C2-025.626.16.xx
syngo OM (Volume 3)

C2-025.620.03.xx
Operator Manual

Hardware: (available in German, English,


French, Japanese and Spanish)
Release Notes C2-025.623.10.xx

Siemens AG CT02-025.812.03.03.02 Page 11 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
12 General Information

Abbreviations and Acronyms 1.3

Tab. 5 Abbreviations and Acronyms

Acronym: Explanation:
CB-Doc Computer-based documentation
CCW Counterclockwise
CW Clockwise
ESD Electrostatically Sensitive Device
HW Hardware
ICS Image Control System
IES Image Evaluation System
IMS Image Measurement System
IRS Image Reconstruction System
PC Personal Computer
ROT Rotation
PDS Power Distribution System
PE Protective Earth
PG Planning Guide
PHS Patient Handling System
SW Software
GPC Gantry Panel Control
UPS Uninterrupted Power Supply
LCB Line Connection Box
LCC/MCC Line Connection Cabinet / Mains Connec-
tion Cabinet (SOMATOM Spirit - used in
China only)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 12 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 13

Overview of System Components 1.4

The SOMATOM Emotion 6/16 (2007) / Spirit / Spirit Power comprise the following com-
ponents:
• Gantry
• PHS (Patient Handling System)
• ICS Tower (Image Control System)
• IRS Tower (Image Reconstruction System)
• IES Tower (Image Evaluation System), (Emotion 2007 only)
• UPS (Uninterrupted Power Supply) for ICS & IRS
• Operating console / control box / monitor / keyboard / mouse
• Spirit / Spirit Power only: LCC / MCC (for the Chinese market only)
• Emotion (2007) only: LCB (line Connection Box)
- With transformer (380-480V Version)
- Without transformer (400V Version)

Siemens AG CT02-025.812.03.03.02 Page 13 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
14 General Information

Fig. 1:

NOTE A router is required for RDIAG.


An ISDN or analog router has to be ordered by the project man-
ager. The router is not included in the CT delivery volume.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 14 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 15

Safety Information 1.5

General 0

Only qualified and trained personnel are allowed to perform CT system installation, ser-
vice, maintenance and quality assurance.
To avoid any risk of injury to persons or damage to the system, the safety instructions
must be read and observed in the:
General Safety Notes (TD00-000.860.01.xx.xx)
and
Product-specific safety notes (CT02-025.860.02.xx.xx) (CT02-025.860.02 / Prod-
uct-specific safety notes)
The product-specific safety notes contain important information regarding:
• General safety information on the gantry, patient table, and laser products.
• Attachment of screws and nuts, adjustment instructions for torque wrench, and use of
Loctite.
• Safety advice (Hazard keys)

Handling Heavy Equipment


Installation personnel have to wear protective clothing (protective shoes, gloves, clothing,
etc.) for personal safety whenever handling heavy equipment.

NOTE Take appropriate precautions to ensure your personal safety


before handling heavy equipment.

ESD Protection
The components are delivered in a pink plastic cover. In addition to providing dust protec-
tion, the plastic foil also protects ESD components against electrostatic discharge. Sev-
eral system components contain electronic parts sensitive to electrostatic discharge
(ESD).

NOTE To provide ESD protection, the pink plastic cover should remain
on the components until access is required.
Always use ESD protective equipment (e.g. wrist band plus
ground strap) before handling electronic components or devices.

Protection from Electrical Shocks or Hazards


Switch off the site-specific line power circuit breaker and remove the main fuses:
• before system start-up.

Siemens AG CT02-025.812.03.03.02 Page 15 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
16 General Information

• before replacing parts or components.


Power is applied to the system for the first time in these installation procedures at system
start-up.

NOTE Do not touch electrical terminals when the system power is on.
Dangerous voltage levels may be present in some components
(e.g. slip ring), even when the system is switched off.

Radiation protection

NOTE All necessary steps for radiation protection have to be taken at all
times prior to releasing radiation at the system.
All installation personnel working on the CT systems have to wear
X-ray film badges for personal dose monitoring at all time.

Automatic procedures

NOTE Do not leave the system unattended when a program is running


with automatic movements or automatic radiation.

Working in the gantry


• Prior to starting work in the gantry:
- Wait until rotation has stopped
- Disable gantry rotation via service switch S5 on MAS D301
- Use the mechanical brake to prevent gantry rotation while working in the gantry.
- Connect the stationary and rotating part of the gantry with a ground cable before
working in the gantry.

NOTE Remove the cable prior to rotating the gantry.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 16 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 17

List of Required Installation Equipment 1.6

Tab. 6 List of Required Installation Equipment

Name of equipment: Delivered Order number Explanation of use:


with unit: or reference:
Standard tool kit & case No Local Siemens Standard installation tools
contains: Office

Metric set of fork


wrenches,
Metric set of Allen
wrenches,
Metric tape measure (3
m),
Set of pliers,
Knife,
Set of screwdrivers (hex,
regular), etc.
Silicone-adhesive Elasto- No n.a. To attach the cable outlet
sil E41; recommended cover at the gantry stand
Protective clothing No Local supply To handle heavy equipment
(shoes, gloves, etc.)
Protractor water level No 28 69 436 or Gantry leveling, acceptance
local supply test
Vacuum cleaner No Local supply To clean after drilling holes,
e.g., when bolting the PHS to
the floor
Gantry transport device No/Yes* 27 92 146 Gantry transport
(*Country-specific regula-
tion)
Hex screws M12X75 Yes 100 49 562 Transport device mounting
(Strength=12.9) -----8pcs 33 42 797
Washers 12----- 8pcs

Drill template Yes 88 62 752 Floor mounting holes for


patient table
Rock drill bits No Local Siemens PHS mounting
(14mm,15mm) Office
File No Local supply To smooth the cut-out in
cover for PHS cable

Siemens AG CT02-025.812.03.03.02 Page 17 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
18 General Information

Name of equipment: Delivered Order number Explanation of use:


with unit: or reference:
Protective strip Yes 20 06 468 To attach the cut-out of the
PHS front cover
Crowbar No Local supply To unpack the boxes
Loctite #221 Yes 20 48 874 To be delivered with PHS
Cable ties No Local Siemens System cabling
Office
Power drill(14mm,15mm) No Local supply To bolt components to the
floor
Metric set of rock drills No Local supply To bolt components to the
floor
Torque wrench 3-20 Nm No 34 24 553 To bolt components to the
for 1/4" sockets floor, replacement of parts
Torque wrench 25-130 No 34 24 561 To install the gantry transport
Nm for 1/2" sockets device and mount compo-
nents
Metric set of Allen sockets No Local supply To install component screws
for torque wrench
Metric set of hex sockets No Local supply To install component screws
for torque wrench

A detailed list of service tools is also available in the system spare parts catalogue, Sie-
mens Med SPC.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 18 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 19

Special Tools for CTDI measurements in accordance with EN


61223-2-6 ( for installations in Germany and some EU countries) 1.7
Tab. 7 Special tools for CTDI measurement

Name of equipment Delivered with unit Order number Explanation of use


CT dosimeter No local supply Constancy test Ger-
includes: dose measure- many and some EU
ment chamber, cable & countries
holder
CTDI body No local supply Constancy test Ger-
(32 cm Plexiglas phantom) many and some EU
countries
CTDI head No local supply Constancy test Ger-
(16 cm Plexiglas phantom) many and some EU
countries

Siemens AG CT02-025.812.03.03.02 Page 19 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
20 General Information

Special Tools for the DIN & IEC Acceptance Test 1.8

Tab. 8 Special tools for the acceptance test

Name of equipment Delivered with unit Order number Explanation of


use
Adapter for low-con- No 81 55 012 Acceptance test
trast phantom (Germany only)
Low-contrast phantom No 96 60 663 Acceptance test
(Germany only)
CT dosimeter No local supply Acceptance test
includes: Dose mea- (Germany only)
surement chamber,
cable & holder
Protractor water level No 28 69 436 Acceptance test
(Germany only)
PHANTOMS Spirit / Spirit Power
Water phantom (20 Yes 72 06 014 Quality test, tuneup
cm)
Wire phantom Yes 88 72 215 Quality test, tuneup
Slice thickness phan- Yes 72 06 121 Quality test,
tom tune-up
PHANTOMS Emotion ( 2007 )
Water phantom (20 Yes 48 06 977 Quality test,
cm) tune-up
Wire phantom Yes 88 72 215 Quality test,
tune-up
Slice thickness phan- Yes 73 96 604 Quality test,
tom tune-up

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 20 of 114 Siemens AG


10.07 SSME CS Medical Solutions
General Information 21

Special Tools for the Monitor ( Germany and some EU countries )1.9
The following additional equipment is required for the monitor test ( only for installations in
Germany and some EU countries ):
Tab. 9 Special tools for the monitor test

Name of equipment Delivered with the Order number Explanation of use


unit
SMfit Spotmeter No See SPC Monitor test

Siemens AG CT02-025.812.03.03.02 Page 21 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
22 Unpacking Components
2-
General Information
2Unpacking Components

• The following unpacking instructions refer to standard sea/air freight packaging with all
components in wooden crates.

NOTE While rigging is in progress, pay attention to the difference in tem-


perature between indoors and outdoors. Keep the system warmed
up according to the temperature gradient (10 k/h.).

Safety 0

NOTE Read and observe the safety information contained in the "Safety
information" section of this document.(Safety Information / p. 15)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 22 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Unpacking Components 23

Gantry 2.1

• Place the gantry container on the floor.


• Check for the shock and tilt indicators. (Fig. 2 / p. 23)(Fig. 3 / p. 23)

Fig. 2: Shock-watch indicator Fig. 3: Tilt-watch indicator

NOTE Check the shock and tilt indicators to see if they have actuated. If
so, notify the project manager accordingly.

• Remove the nails and cardboard washers from the wooden crate.
• Remove the wooden top and sides of the container.
• Check the amount in accordance with the general packing list (located in the fourth
package "Accessories").

Siemens AG CT02-025.812.03.03.02 Page 23 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
24 Unpacking Components

• Spirit systems delivered beginning with Serial No. 31031:


- Remove the two screws and washers attaching the right stand covers. Each screw is
hidden behind a transport adapter plate. However, these two screws can be removed
through the holes on the transport covers. The right stand covers can be removed
before dismounting the transport device so that the gantry can be easily positioned
with respect to the outlet of the cable conduit. (Fig. 4 / p. 24)(Fig. 5 / p. 24)

Fig. 4: Hole in transport adapter plate and mounting Fig. 5: Removing the screw and washer
bracket
Pos. 1 Screw is hidden
Pos. 2 Hole in the adpater plate and mounting bracket

• Remove the 4 yellow mounting brackets.(Fig. 6 / p. 24)


- Two at the left side in front / back.
- Two at the right side in front / back.
• Remove the red transport lock in back of the gantry (Fig. 7 / p. 24) --- Spirit systems
delivered up to serial no. 31349:

Fig. 6: Gantry mounting brackets (front - to the left) Fig. 7: Gantry mounting bracket (back - in the
Pos. 1 Brackets middle)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 24 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Unpacking Components 25

• Remove the red transport lock in back of the gantry. (Fig. 8 / p. 25) --- Emotion (2007),
Spirit Power and Spirit systems delivered beginning with serial no. 31350:

Fig. 8: Gantry mounting bracket (back - in the middle)

Attaching the Gantry Transport Device 0

NOTE Special attachment screws and washers have to be used for


installation of the gantry transport device. They are included in
the delivery volume for the system.
Refer to the information on the warning label attached to the
transport device (Fig. 9 / p. 25).

Fig. 9: Centering and attachment screws


Pos. 1 Warning label
Pos. 2 Centering screw
Pos. 3 Attachment screw

• Turn the centering screws (Fig. 9 / p. 25) into the holes of the transport adapter.
(Fig. 10 / p. 26)(Fig. 11 / p. 26)

Siemens AG CT02-025.812.03.03.02 Page 25 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
26 Unpacking Components

• Adjust the distance between the gantry transport adapters and the transport device with
the centering screws. The distance should be the same on both sides.

Fig. 10: The centering screws in positon at transport Fig. 11: Centering screws in position at
adapter (short ones) transport adapter (long ones)
Pos. 1 Centering screws in holes Pos. 1 Centering screws in holes

• Position the gantry transport device on both sides of the gantry. Insert the attachment
screws (Fig. 9 / p. 25) through the centering screws and tighten them with a torque of
77 Nm ±10%. (Fig. 12 / p. 26)(Fig. 13 / p. 26)
- Be careful to avoid scratching the gantry.

Fig. 12: Positioning the transport device (left side) Fig. 13: Positioning the transport device (right
side)

• Swivel out the wheels of the transport device on both sides to prevent the gantry from
tilting. (Fig. 14 / p. 27)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 26 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Unpacking Components 27

NOTE Transport with the rollers swiveled in is permissible only in nar-


row spaces. As soon as the system has passed through the nar-
row space, the transport rollers must be swiveled out again.

Fig. 14: Swivelling out the wheels of the transport device


Pos. 1 Securing bar

• Lock the securing bars of the transport device.


• Rotate the four transport wheels toward the floor by turning the spindle clockwise. Put a
metal bar through the hole in the upper end of the spindle to facilitate turning.
• Raise the gantry approx. 10-15 mm off the pallet by turning the spindle clockwise.
• Move the gantry next to the pallet and remove the pallet.
• Lower the gantry by turning the spindles counterclockwise until the distance between
the gantry and floor is approx. 15 mm (1/2 inch)

NOTE Keep the shorter up-down adjustment when turning the spindles
one by one.

• Installation of the gantry continues under "Gantry" (Gantry / p. 37).

Siemens AG CT02-025.812.03.03.02 Page 27 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
28 Unpacking Components

Patient Table (PHS) 2.2

• Place the container on the floor.


• Remove the nails from the wooden crate.
• Remove the wooden top of the crate.

Fig. 15: Patient table on transport pallet


Pos. 1 Wooden support
Pos. 2 Drill template
Pos. 3 Ramp
Pos. 4 PHS accessory boxes

• Cut the straps and remove the protective pads.


• Remove the pink ESD foil.
• Check the amount in accordance with the entire packing list.
(The general packing list is in the fourth package "Accessories").
• Remove the parts from underneath the patient table.
- The 2 PHS accessory boxes
- The drill template
- The ramp.

NOTE Do not remove the patient table from the pallet at this point.
The patient table cannot stand alone. It requires a wooden support
under the foot end; otherwise, it will tip over.

• Installation of the patient table continues under (Patient Table (PHS) / p. 50)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 28 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Unpacking Components 29

Monitor / Keyboard / Mouse / Control Box / ICS, IRS & IES Towers2.3
• Place the wooden crate on the floor.
• Remove the nails / screws from the wooden crate .
• Remove the top and side covers of the crate.
• Check the amount in accordance with the entire packing list.
(The general packing list is in the fourth package "Accessories").
• Bring the cardboard boxes into the CT control room.
• Installation of the components continues under (Monitor / Keyboard & Mouse / Control
Box / p. 67)

Siemens AG CT02-025.812.03.03.02 Page 29 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
30 Unpacking Components

Console / Container (Optional) 2.4

• Check the amount in accordance with the entire packing list.


(The general packing list is in the fourth package "Accessories").
• Unpack the furniture.
• Assemble the furniture:
- The console is delivered unassembled as a kit.
- Assemble the furniture/console according to the instructions provided.
- Keep the instructions in a safe place and ready for the system startup process.
• Installation of the components continues under (Console / Container / Furniture
(Optional) / p. 66)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 30 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Unpacking Components 31

Accessories and Phantoms 2.5

• Check the amount in accordance with the overall packing list.


(The general packing list is in the fourth package "Accessories")
• Phantoms
- Unpack all phantoms and store them in a safe place in the examination room (pro-
vided on-site:e.g. service cabinet).
• Patient positioning aids
- Unpack all accessories and store them in a clean and safe location in the examina-
tion room (provided on-site :e.g. cabinet, drawer).

Siemens AG CT02-025.812.03.03.02 Page 31 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
32 Unpacking Components

(Line Connection Cabinet - 400V) optional 2.6

• Unpack the package.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 32 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Unpacking Components 33

Line Connection Box ( Emotion 6/16 ( 2007 ) ) 2.7

• Place the LCB container on the floor.


• Remove packaging:
- Remove the nails and cardboard washers from the wooden crate with a crowbar.
- Remove the wooden covers and side walls of the container.
- Remove the pink ESD foil.
• Open the LCB:
- Remove the front cover of the LCB by pulling it toward the front and upwards.
- Remove the contact cover by removing the six hexagonal screws (Fig. 16 / p. 33)and
the ground wire(Fig. 17 / p. 33). Do not damage the ground wire.

Fig. 16: LCB contact cover Fig. 17: PE cable of LCB contact cover
Pos. 1 Position of screws

• Remove the four mounting screws: Two at the back of the LCB (yellow angles)
(Fig. 18 / p. 33)and two inside the LCB near the front cover(Fig. 19 / p. 33).

Fig. 18: Two mounting screws at the back of the LCB Fig. 19: Two mounting screws inside the LCB
Pos. 1 Position of the two mounting screws Pos. 1 Position of the two mounting screws

Siemens AG CT02-025.812.03.03.02 Page 33 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
34 Unpacking Components

• Move the LCB next to the pallet (handle the LCB only at the consolidated spots, at the
left top and to the right of the LCB) and remove the pallet.(Fig. 20 / p. 34)

Fig. 20: LCB: Consolidated spots


Pos. 1 Consolidated spots

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 34 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 35
3-
General Information
3Installation of Components

• This chapter describes installing components in the examination and control rooms.
• General information regarding the requirements for system transport is described in the
"Planning Guide" (Tab. 1 / p. 8)

Siemens AG CT02-025.812.03.03.02 Page 35 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
36 Installation of Components

Safety Information 3.1

NOTE Read and observe the safety information contained in the "Safety
information" section of this document.(Safety Information / p. 15)

NOTE The minimum distances between the system and the wall have to
be observed in accordance with the planning guide.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 36 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 37

Gantry 3.2

Positioning the Gantry 0

NOTE The gantry should not be transported over uneven flooring.


Impacts to the gantry may damage its more sensitive parts.

• To minimize the risk of tipping, do not raise the gantry by more than 15 mm (~. 1/2
inch) above the floor.
• Move the gantry on the transport device to the final destination.
- Standard transport, wheels swivelled out (Fig. 21 / p. 37)
- Transport through narrow doors/passages , wheels swivelled in (Fig. 22 / p. 37)

Fig. 21: Standard transport with wheels swivelled out Fig. 22: Transport with wheels swivelled in (e.g.
doors)

- Push the gantry only via the spindles of the transport device.

NOTE Do not push or pull on the plastic cover of the gantry.

• Before positioning the gantry in the final position


- Remove the flooring at the gantry location for the 4 adjustment feet in accordance
with the drawings ( site layout) for direct contact between feet and concrete.
- If necessary ( see NOTE), drill holes for bolting the gantry to the floor.

NOTE The gantry does not have to be bolted to the floor.


Some countries (e.g. countries prone to earthquakes) require spe-
cial mountings for components mounted to the floor. (Special
Mounting / p. 72)

• Position the gantry at the installation location in accordance with the drawings (site lay-
out).
• Lower the gantry to the floor and turn the four spindles of the transport device counter-
clockwise.

Siemens AG CT02-025.812.03.03.02 Page 37 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
38 Installation of Components

Removing the Transport Device and the 4 yellow Adapters 0

• Remove the left and right transport device.


- Remove the four attachment screws (Fig. 23 / p. 38) from each transport device.
- Loosen the four centering screws (Fig. 23 / p. 38) from each transport device.
- Remove the left and right transport device from the gantry.

Fig. 23: Centering and attachment screws

Pos. 1 Warning label


Pos. 2 Centering screw
Pos. 3 Attachment screw

NOTE Retain the special attachment screws (Fig. 23 / p. 38) from the gan-
try transport device for future use.

• Remove the 4 yellow adapters.


- Loosen the two screws in the transport device adapter and remove them from the
gantry.(Fig. 24 / p. 38)(Fig. 25 / p. 38)
- Remove the 4 yellow adapters and be careful to avoid scratching the gantry housing.

Fig. 24: Adapter screws (short ones) Fig. 25: Adapter screws (long ones)
Pos. 1 Adapater screws Pos. 1 Adapter screws

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 38 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 39

Removing the Covers (Spirit) 0

• Gantry left stand cover (Fig. 26 / p. 39):


- Remove the screws (2 pcs) at the bottom in front and back of the cover
- Look into the gap between the cover and the gantry and pull the cover up until the
guide pin is released.

Fig. 26: Left stand cover


Pos. 1 Left stand cover
Pos. 2 Position of screws (front and back)

• Gantry right stand & back stand cover:


- Remove the screws ( Pos.1 ) at the back cover of the right stand (Fig. 27 / p. 39)

Fig. 27: Right stand cover


Pos. 1 Position of screws
Pos. 2 Right back cover

Siemens AG CT02-025.812.03.03.02 Page 39 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
40 Installation of Components

- Lift the cover and remove it .


- Remove the upper and lower screws ( Pos.1 ) at the right stand
cover(Fig. 28 / p. 40)(Fig. 29 / p. 40).

Fig. 28: Right stand cover (front) Fig. 29: Right stand cover (back)
Pos. 1 Position of screws Pos. 1 Position of screws
Pos. 2 Right stand cover

- Lift the front cover and remove it.


• Front funnel.

NOTE While removing the gantry funnels, pay special attention to the
window strip. The hooks of the window strip may be broken, and
the surface of it can be scratched very easily.

(Spirit systems delivered up to serial no.31030):


- Remove the screws (4 pcs) at the front funnel(Fig. 30 / p. 41).
- Pull the cover slightly away from the gantry by approx. 15~20cm.
- Unplug the cable connectors ( X5, X6)(Fig. 31 / p. 41).
- Disassemble the transparent window strip in the middle of the funnel.
- Lift the front funnel and remove it.
(Spirit systems delivered beginning with serial no. 31031) :

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 40 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 41

- Remove the screws (4 pcs) at the front funnel(Fig. 30 / p. 41).


- Pull the top of the cover to the front as far as the safety hooks can be hooked into the
guide holes (Fig. 32 / p. 41).
- Unplug the cable connectors ( X5, X6)(Fig. 31 / p. 41).
- Disassemble the transparent window strip in the middle of the funnels.
- Loosen the safety hooks, put them back onto the frame rest, lift the cover out of the
bottom guide, and remove the cover from the gantry.(Fig. 33 / p. 41)

Fig. 30: The front funnel Fig. 31: Cable connectors


Pos. 1 Position of screws Pos. 1 Connector X6
Pos. 2 Connector X5

Fig. 32: Security hooks Fig. 33: Bottom cover guides


Pos. 1 Security hooks Pos. 1 Bottom cover guides

Siemens AG CT02-025.812.03.03.02 Page 41 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
42 Installation of Components

NOTE To protect the collimator box against damage, move the collimator
box to the 12 o’clock position.(Fig. 34 / p. 42)

Fig. 34: Collimator box at 12 o’clock position


Pos. 1 Collimator box

Removing the Covers (Emotion 2007 / Spirit Power) 0

NOTE The figures below are Emotion (2007), but similar to those for Spirit
Power.

• Gantry left stand cover(Fig. 35 / p. 42) :


- Remove the screw ( Pos.2 ) at the bottom in back of the gantry.
- Look into the gap between the cover and the gantry, and pull the cover up until the
guide pin is released.

Fig. 35: Left stand cover


Pos. 1 Left stand cover
Pos. 2 Position of screw

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 42 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 43

• Gantry right stand & back stand cover:


- Remove the screws ( pos.1 ) at the back stand cover(Fig. 36 / p. 43).
- Lift the cover and remove it.
- Remove the upper and lower screws ( Pos.1 ) at the right stand
cover(Fig. 37 / p. 43)(Fig. 38 / p. 43)
- Lift the front cover and remove it.

Fig. 36: The right stand cover


Pos. 1 Position of screws
Pos. 2 Right stand cover

Fig. 37: Right stand cover ( front ) Fig. 38: Right stand cover ( back )
Pos. 1 Position of screws Pos. 1 Position of screws
Pos. 2 Right front cover

• Front funnel.

NOTE While removing the gantry funnels, pay special attention to the
window strip. The hooks of the window strip may be broken, and
the surface of it can be scratched very easily.

Siemens AG CT02-025.812.03.03.02 Page 43 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
44 Installation of Components

- Remove the screws ( Pos. 1) at the front funnel. (Fig. 39 / p. 44)


- Pull the top of the cover to the front as far as the safety hooks can be hooked into the
guide holes (Fig. 41 / p. 44)
- Unplug the cable connectors ( X4, X5, X6 )(Fig. 40 / p. 44).
- Disassemble the transparent window strip in the middle of the funnel.
- Loosen the safety hooks, put them back onto the frame rest, lift the cover out of the
bottom guide, and remove the cover from the gantry.(Fig. 33 / p. 41)

Fig. 39: Front funnel Fig. 40: Cable connectors


Pos. 1 Position of screws Pos. 1 Connector X5
Pos. 2 Connector X6
Pos. 3 Connector X4 (optional)

Fig. 41: Security hooks Fig. 42: Funnel support


Pos. 1 Security hooks Pos. 1 Funnel support

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 44 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 45

NOTE To protect the collimator box against damage, move the collimator
box to the 12 o’clock position

Fig. 43: Collimator box at 12 o’clock position


Pos. 1 Collimator box

Removing the Transport Locks and Hand Brake 0

• The gantry has two red transport locks (left and right gantry stand) and one hand brake
• Removing the transport locks at the left gantry stand
- Remove the 3 Allen screws (Fig. 44 / p. 45)from the metal block at the left gantry
stand.
- Remove the block by pulling it off the tilt mechanism.
• Removing the transport locks at the right gantry stand --- (Spirit systems delivered up to
serial no. 31349 )
- Remove the 3 Allen screws (Fig. 45 / p. 45) from the metal block at the right gantry
stand.
- Remove the block by pulling it off the tilt mechanism.

Fig. 44: Gantry transport lock (left stand)


Pos. 1 Transport lock
Fig. 45: Gantry transport lock (right stand)
Pos. 1 Transport lock

Siemens AG CT02-025.812.03.03.02 Page 45 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
46 Installation of Components

(Emotion ( 2007) Systems and Spirit system delivered beginning with serial no. 31350 :
• Removing the transport locks at the right gantry stand
- Remove the 1 Allen screws (Fig. 46 / p. 46) from the metal block at the right gantry
stand.
- Remove the block by pulling it off the tilt mechanism.

Fig. 46: Gantry transport lock (right stand)


Pos. 1 Transport lock

• Releasing the hand brake at the gantry


- The hand brake is located at the right lower corner of the gantry front.

NOTE Check the rotation by turning the rotation part of the gantry in
clockwise direction only ( CW as seen from the gantry front).

Removing the Transformer Lock 0

NOTE Do not damage the cables during removal.

• Remove the 3 screws and one ground connection nut from PSS, MAS
holder.(Fig. 47 / p. 47)
• Open the PSS, MAS holder.(Fig. 48 / p. 47)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 46 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 47

Fig. 47: PSS, MAS holder (closed) Fig. 48: PSS, MAS holder (open)
Pos. 1 Three screws
Pos. 2 Ground connection

• Remove the screw (Fig. 49 / p. 47)and both nuts (Fig. 50 / p. 47)from the transport
locking plate.
• Remove the transport locking plate.

Fig. 49: Transformer, front Fig. 50: Transformer, back


Pos. 1 Transport retaining screw Pos. 1 Transport retaining nuts

Leveling the Gantry 0

Ensure that all four feet in all four corners(Fig. 51 / p. 48)


(Fig. 52 / p. 48)(Fig. 53 / p. 48)(Fig. 54 / p. 48)will have direct contact with the concrete
floor.

Siemens AG CT02-025.812.03.03.02 Page 47 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
48 Installation of Components

Fig. 51: Adjustable foot in left front gantry stand Fig. 52: Adjustable foot in left back gantry stand

Fig. 53: Adjustable foot in right front gantry stand Fig. 54: Adjustable foot in right back gantry stand

• Position a water level on the gantry frame and check the level in two directions (across
and along(Fig. 55 / p. 48) (Fig. 56 / p. 48) (Fig. 57 / p. 49) (Fig. 58 / p. 49).
• If necessary, correct the leveling of the gantry by adjusting the four feet with an Allen
key.

Fig. 56: Leveling check at the gantry ( Spirit )


Fig. 55: Leveling check at gantry ( Spirit )

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 48 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 49

Fig. 57: Leveling check at the gantry ( Emotion 2007) Fig. 58: Leveling check at the gantry (Emotion 2007 )

• The gap between the gantry and floor can be sealed with silicone.
- NOTE: Silicone is not part of the delivery, it must be provided locally by the project
manager. The color of the silicone can be selected by the customer.
• Installation of the gantry covers continues under (Final steps / p. 113)

Siemens AG CT02-025.812.03.03.02 Page 49 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
50 Installation of Components

Patient Table (PHS) 3.3

General Information 0

NOTE The patient table cannot stand independently. It requires a


wooden support under the foot-end; otherwise, it will tip over.
We recommend having a second person assist when installing
the patient table.

The local Siemens office is responsible for the rock drill bits (14 mm, 15 mm).

Tools required for installing the Liebig or Hilti Floor Anchors 0

The “Liebig self-undercut anchor” or the “Hilti anchor” should be used to anchor the PHS to
the floor.
• Pen (to mark the position of the holes in the floor)
• Power drill ( to drill the holes ), rock drill bits ( size: 14 mm for Liebig and 15 mm for Hilti)
• Vacuum cleaner with a small nozzle ( to clean the holes after drilling )
• Hammer ( to insert the Liebig and the Hilti anchors into the holes )
• Torque wrench ( 25Nm / 50 Nm)
• Allen attachment size 8 for torque wrench ( to secure the anchor in the floor )
• Allen attachment size 6 for torque wrench ( to bolt the PHS to the floor )
• Hex nut sizes 13 and 17 for torque wrench (for the nuts bolting the PHS to the floor)
• Loctite # 221
• Injector for silicone
• Silicone for the strip between the floor and the PHS ( provided by the project manager)
• 8 mm insert for hexagonal head bolt ( 1/4 ))
• Socket adapter 1/4 drive - ( 1/2 )
• Hammer ( to insert the Liebig and the Hilti anchors into the holes )

Drilling the Holes for Floor Mounting 0

NOTE Bolting the PHS to the floor is mandatory.


The installation materials included in the delivery volume are suit-
able for flush or surface mounting (concrete/subfloors). The cus-
tomer is responsible for determining suitable anchoring methods
as well as for procuring installation materials for other floor types.

Drilling template for PHS 8596566 up to serial no. 10500

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 50 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 51

Fig. 59: Drilling template with label


Pos. 1 Outer holes
Pos. 2 Inside holes

Drilling template for PHS 8596566 beginning with serial no. 10501

Fig. 60: Drilling template with label

Siemens AG CT02-025.812.03.03.02 Page 51 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
52 Installation of Components

Pos. 1 Inside holes


Pos. 2 Outer holes
Pos. 3 Alignment mark for adjustment of the table to gantry distance
Pos. 4 Alignment mark for the gantry base
Pos. 5 Alignment mark of the PHS position guide

NOTE There are two kinds of holes on this template. One type is for
Emotion 2003. The other type is for Spirit and Emotion 2007. Make
sure that right type of holes are used as templates for PHS instal-
lation.

Fig. 61: For Emotion 2003 Fig. 62: For Emotion 2007 / Spirit

NOTE The patient table has to be bolted to the floor using 4 screws, one
at each corner.
As an alternative to the inside holes (1), the outer holes (2) may be
used to mount the patient table to the floor, depending on the con-
ditions of the floor at the local site.
Refer to the label on the drill template for drilling the holes and
installing the Liebig or Hilti anchors, or .
The entire length of the Liebig anchor and the lower part (80 mm)
of the Hilti anchor must be inserted into the concrete base. The
anchors may not be inserted into the floor cement.

• Position the drill template with side A (Fig. 59 / p. 51) Pos.4 (Fig. 60 / p. 51)at the exact
center of the gantry base.
• Drill the 4 holes (position 1 or 2 on the drill template) on the floor. Drill holes using a drill
diameter of 14 mm for the Liebig anchor or 15 mm for the Hilti anchor. Refer to the label
on the drill template for the drilling procedure.
• Remove the drill template
• Clean the holes using a vacuum cleaner with a small nozzle.
The following materials and tools are included in the delivery volume (packed in the box
with the patient table covers):
- 4 Liebig self-undercut anchor IL S 14/80 with a 14 mm drill bit.
- 4 Hilti HSL- 3-G M10/100 with a 15 mm drill bit.
- 4 Allen screws M8 x 35 (strength = 8.8)
- 4 Allen screws M8 x 25 (strength = 8.8)
- Large and small washers
- Spacer shims

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 52 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 53

- 4 threaded rods, 170 mm length (strength = 8.8)


- 4 size 10 locking edge washers
- 4 metric hexagon nuts M8
- Hexagon rod 135 mm
• If threaded rods are used, cut them to the proper length. The distance between the floor
cement and the upper end of the threaded rods has to be 45 mm.

Positioning the PHS (Patient Table) 0

NOTE We recommend that a second person assist when installing the


table.

• Accurately position the drill template with side D (Fig. 59 / p. 51) or Pos.3
(Fig. 60 / p. 51)at the gantry base and center it.
• Remove the flooring at the patient table location for the 4 mounting points for direct con-
tact with the concrete flooring. Fill up the flooring height with the metal plates (shims
delivered with the patient table)
• Remove the four mounting screws at the table base that secure the patient table to the
pallet. Two in front (Fig. 63 / p. 53) and two in back (Fig. 64 / p. 53)

Fig. 63: Table mounting screws - front Fig. 64: Table mounting screws - back
Pos. 1 Screws Pos. 1 Screws

• Place the wooden ramp next to the pallet (front end of patient table (Fig. 65 / p. 54)).

Siemens AG CT02-025.812.03.03.02 Page 53 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
54 Installation of Components

• Raise the patient table to maximum position by turning the two hex screws in clockwise
direction on the two transport rollers. (Fig. 66 / p. 54)

Fig. 66: Transport wheel

Fig. 65: Pallet with ramp


Pos. 1 Wooden ramp

• Remove the plastic belt and lift the patient table at the back end of the PHS.

Fig. 67: Handle at the back end of the PHS

Pos. 1 Handle

NOTE Do not lift the patient table by the handle (Fig. 67 / p. 54)

• Carefully move the patient table down the ramp using the two transport wheels. If the
patient table will not be positioned at its final destination, remove the wooden support
from the pallet and put the wooden support under the table for stabilization.

NOTE The patient table cannot stand alone.


It requires a wooden support under the foot-end; otherwise, it will
tip over. However, even with a wooden support, the table remains
unstable. It may tip over by just touching it.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 54 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 55

• Carefully move the patient table to the installation location until the base of the patient
table touches side B of the template or
- The cut-outs on side B of the template are sized to fit the bolts at the base of the
patient table.
- This aligns the patient table's distance and angle to the gantry.

Fig. 68: Positioning the drilling template between the gantry and PHS (PHS part no 8596566.: up

Siemens AG CT02-025.812.03.03.02 Page 55 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
56 Installation of Components
to serial no. 10500)
Pos. 1 Middle of the gantry
Pos. D Side D
Pos. B Side B

Fig. 69: Positioning the drilling template between gantry and PHS ( PHS part no: 8596566
beginning with serial no. 10501 )
• Adjust the patient table to the centered drilling template.
• Ensure that there are no gaps between patient table, drilling template, and gantry.
• Insert the four floor mounting screws plus washers, or the Hilti anchors.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 56 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 57

• Measure the distance at both symmetrical points from the back and front corner of the
patient table base to the corner of the gantry base and align the patient table until the
distance is the same.(Fig. 70 / p. 57) (Fig. 71 / p. 57)

Fig. 70: Measured from the back corner of the PHS Fig. 71: Measured from the front corner of the PHS
Pos. 1 Length A Pos. 1 Length C
Pos. 2 Length B Pos. 2 Length D

Leveling the PHS 0

• Position a water level across the tabletop and ensure it is level (Fig. 72 / p. 57).
• Position the water level along the side of the tabletop and ensure it is level
(Fig. 73 / p. 57)

Fig. 72: Leveling check at the PHS Fig. 73: Leveling check at the PHS

Siemens AG CT02-025.812.03.03.02 Page 57 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
58 Installation of Components

• If necessary, adjust by installing metal plates (shims, (Fig. 74 / p. 58)) at the 4 anchor-
ing positions under the patient table base.

Fig. 74: Position of the shims under the patient table


Pos. 1 Anchor screw

• Use Loctite 221 to secure the screws or nuts used to attach the table to the anchors.
• Tighten the screws or nuts at the Liebig anchors with a torque of 25 Nm. For the nuts in
the Hilti anchors, use a torque of 50 Nm.
• After leveling the patient table, check the positioning again with the drill template and
the gantry.
• Remove the drill template.
• There must be a silicone strip between the patient table base and the floor.
- NOTE: Silicone is not part of the delivery, it must be provided locally by the project
manager. The color of the silicone can be chosen by the customer.

Leveling the Gap below the vertical Drive 0

The gap between the floor and the pit for the table has to be leveled by placing shims
under the drive until the space is filled.(Fig. 75 / p. 59)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 58 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 59

Fig. 75: Opening for leveling the space under the vertical drive
Pos. 1 Opening used for inserting shims

Installation of the PHS covers continues under (Final steps / p. 113)

Siemens AG CT02-025.812.03.03.02 Page 59 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
60 Installation of Components

Installing the Cable Duct and Cable Cover between the Gantry
and PHS 3.4

NOTE The cable duct between the gantry and the patient table has to be
secured to the floor. The installation materials for securing the
cable duct have to be purchased locally.

NOTE • There are two kinds of cable duct sets (2/Fig. 76 / p. 60). An
alternative of the two cable duct sets is delivered with the sys-
tem. For the new set, cables between the gantry and the PHS
are only routed on the flooring.
• The instructions below are not necessary, if the cables are
routed underneath the flooring.
• If the cables are routed on the flooring, the cable cover of the
old type and the front cover of the PHS need to include a
cut-out at the curved section.

• Check the routing of the cables


- On the floor with the cable duct (old type cable duct for example): (Fig. 77 / p. 60))
- Below the floor, without cable duct

Fig. 76: Cable duct sets Fig. 77: Cable routed on the flooring with cable
Pos. 1 Old cable duct duct
Pos. 2 New cable duct

• Prepare the covers


- Break off the curved section (Fig. 78 / p. 61)
- Break off the cable cover at the front cover of the PHS (Fig. 79 / p. 61)
(Fig. 80 / p. 61) (Fig. 81 / p. 61)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 60 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 61

Fig. 78: Cut-out, curved section (gantry) Fig. 79: Front cover of PHS (after cut-out)
Pos. 1 Cable cover
Pos. 2 Curved section
Pos. 3 Cable duct

Fig. 80: Using a wire-cutter for cut-out (front cover) Fig. 81: Using a protective strip to attach to cut-out
(front cover)

• Position the flat part of the cable duct between the gantry and patient table and align it
with the cable outlets at the gantry and PHS.
• Secure the cable duct to the floor using screws with anchors or double-sided adhesive
tape.

Siemens AG CT02-025.812.03.03.02 Page 61 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
62 Installation of Components

• Insert the cables into the cable duct and route them into the patient table.
(Fig. 82 / p. 62)

Fig. 82: Cable routed on the flooring with cable duct before installing the cover
• Install the cover on the cable duct. If necessary, shorten the cover.
• Attach the cable cover (Fig. 78 / p. 61) with ,e.g., using Elastosil E41 silicone adhesive

Cable Routing below the Flooring without the Cable Duct 0

The cable conduit must be provided on-site in the floor.


• Do not cut out the PHS front cover or gantry cable cover.
• Route the cables from the gantry through the on-site cable conduit into the PHS.
• Attach the cable cover using ,e.g., Elastosil E41 silicone adhesive.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 62 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 63

Line Connection Box (Emotion 6/16 (2007) ) 3.5

NOTE Normally, the LCB has to be bolted to the wall, but not to the floor.
However, in countries with special requirements (e.g. in countries
prone to earthquakes), the LCB has to be bolted to the floor as
well.

• If the LCB has to be bolted to the floor (i.e. earthquake regions or Mobile CT):
¹ Also refer to (Special Mounting / p. 72)
• The LCB has to be bolted to the wall with one screw (size M8).
• Level the LCB by turning the pedestals in clockwise or counterclockwise direction.
• If necessary: Open the LCB:
- Remove the front cover of LCB by pulling it toward the front and upwards.
- Remove the contact cover by removing the six hexagonal screws and the PE wire.
• Cable routing from the back:
- Remove the cover plate of the rear cable inlet. (1/Fig. 83 / p. 64)(Fig. 84 / p. 64)
- Remove the right bottom cable holder from (2/Fig. 83 / p. 64) and mount it to the back
cable inlet as shown in (3/Fig. 83 / p. 64)(Fig. 84 / p. 64) (edge protector is delivered
with the LCB).
- Place the LCB in its final position.
- Mark the position of the drill hole through the hole in the LCB back panel.
(Fig. 83 / p. 64)(Fig. 84 / p. 64)

Siemens AG CT02-025.812.03.03.02 Page 63 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
64 Installation of Components

• Cable Routing from the bottom:


- Place the LCB in its final position.
- Mark the position of the drill hole through the hole in the LCB back panel and the
cable inlet from the floor. (Fig. 83 / p. 64)(Fig. 84 / p. 64)

Fig. 83: Installation of LCB


Pos. 1 Cover plate
Pos. 2 Cable holder
Pos. 3 New position for cable holder

Fig. 84: Installation of LCB


Pos. 1 Hole for wall mounting

• Move the LCB to access the drill location.


• Drill the holes for the anchor (M8 DIN 912), or cut-out the floor (if necessary).
• After drilling, clean the holes with a vacuum cleaner that has a small nozzle.
• Insert the anchor (M8) into the hole in the wall.
• Move the LCB back into the installation location and align the hole in the back with the
hole in the wall.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 64 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 65

• Insert the screw (M8) and the required washers into the anchor via the hole.
• Tighten the screw until the bolt head contacts the back panel (tighten carefully to avoid
damaging the LCB back panel).

Siemens AG CT02-025.812.03.03.02 Page 65 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
66 Installation of Components

Console / Container / Furniture (Optional) 3.6

NOTE A console , a container, and/or other furniture may be part of the


delivery volume, if ordered.
The minimum distance between the IMS Container and the wall is
100 mm for cooling purposes. Distances depend on the furniture.

• Position the furniture (when ordered as an option) in the control room at the defined
location as shown in the drawings.
• Remove the front cover of the container in accordance with its instructions:
- The base plate of the container has wheels and slides out toward the front.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 66 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 67

Monitor / Keyboard & Mouse / Control Box 3.7

• Position the monitor, keyboard ,and mouse on the operator desk or


console.(Fig. 85 / p. 69)
• Position the control box (part no.: 03806515 (Fig. 86 / p. 70) or 08598885
(Fig. 87 / p. 70) next to the keyboard.

NOTE For safety reasons, the maximum distance between the control
box and keyboard must be 2 meters.

Siemens AG CT02-025.812.03.03.02 Page 67 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
68 Installation of Components

ICS / IRS tower 3.8

• Position the ICS and IRS under the operator's desk or console.
- The ICS and IRS towers must be positioned next to each other.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 68 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 69

UPS (Uninterrupted Power Supply) 3.9

• Position the UPS.


Place the UPS on top of the ICS tower ( Spirit )
Place the UPS between the ICS and IRS tower ( Emotion 2007)

Fig. 85: Control room - example


Pos. 1 Control box
Pos. 2 Keyboard
Pos. 3 Mouse
Pos. 4 Monitor (17"or 19")
Pos. 5 UPS
Pos. 6 ICS
Pos. 7 IRS (behind the cover)

Siemens AG CT02-025.812.03.03.02 Page 69 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
70 Installation of Components

Fig. 86: Control box (part no.: 03806515) Fig. 87: Control box (part no.: 08598885)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 70 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Installation of Components 71

Line Voltage Connection Cabinet (400V) /Mains Connection Cabi-


net(China) (Spirit / Spirit power) 3.10

• Position the LCC/MCC at the final position in accordance with the on-site layout.
• The LCC/MCC must be bolted to the wall with 4 screws (size M8) or 8 self-tapping
screws .
- The local project manager provides the materials for installation.
• Mark the position of the drill hole through the hole in the back panel
• Drill the holes for the anchor and clean them with a vacuum cleaner.
• Move the LCC/MCC back to the installation location and align the hole in the back with
the hole in the wall.
• Insert and tighten the screws and confirm that they are seated satisfactorily.

Siemens AG CT02-025.812.03.03.02 Page 71 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
72 Special Mounting
4-
General Information
4Special Mounting

Safety 0

NOTE Read and observe the safety information contained in the "Safety
information" section of this document.(Safety Information / p. 15)

NOTE The special component mounting described in this chapter is


mandatory only for installations in countries with specific require-
ments (e.g. countries prone to earthquakes).
The patient table has to be bolted to the floor.

• The procedures described in this chapter are required for installations with country-spe-
cific procedures only. Skip this chapter if it is not applicable.
• The required installation materials (screws, anchors, washers, etc.) have to be pur-
chased locally.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 72 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Special Mounting 73

Gantry 4.1

Bolting the Gantry to the Floor 0

The gantry has to be bolted to the floor using 4 screws (max. size M12) through the feet in
the left and right gantry stand.

Fig. 88: Bolting position in left gantry stand


• Mark the locations of the drill holes through the feet in the gantry stands (Fig. 88 / p. 73)
• Lift the gantry off the floor with the transport device and move it away to access the drill
locations.
• Drill the four holes for the anchors according to your seismic requirements and remove
the dust with the vacuum cleaner.
¹ Drilling depth: Ensure that the anchors are firmly seated in the concrete and not
just in the sub-flooring.
• Insert the anchors into the four holes in the floor.
• Move the gantry back to the installation location and align the holes in the feet with the
holes in the floor.
• Lower the gantry to the floor, ensuring that the holes in the feet and the floor are still
aligned.
• To level the gantry, continue with (Leveling the Gantry / p. 47)
• Insert the screws through the feet into the anchors.
• Tighten all 4 screws with a torque wrench (according to your bolting material used ) to
secure the gantry to the floor.
• Continue with (Removing the Transport Locks and Hand Brake / p. 45)

Siemens AG CT02-025.812.03.03.02 Page 73 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
74 Special Mounting

Line Connection Box (Emotion 6/16 (2007)) 4.2

Bolting the LCB to the Floor 0

• The LCB must be bolted to the floor with 3 screws (size M8 / see (Fig. 89 / p. 74))
through the holes in the bottom of the LCB.

Fig. 89: LCB: holes for floor mounting


Pos. 1 Holes for floor mounting (seen from below)

• Mark the locations of the drill holes through the holes in the bottom of the LCB.
• Move the LCB away to access the drill locations.
• Drill the three holes for the anchors (M8 DIN 912) and remove the dust with a vacuum
cleaner.
¹ Drilling depth: Ensure that the anchors are firmly seated in the concrete and not
just in the sub-flooring.
• Insert the anchors (M8) into the three holes in the floor.
• Move the LCB back to the installation location and align the holes in the bottom of the
LCB with the holes in the floor.
• Insert the M8 screws through the holes into the anchors.
• Tighten the screws carefully to avoid damaging the LCB pedestals / housing.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 74 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Special Mounting 75

Mobile CT (Optional) 4.3

General Information 0

NOTE The trailer manufacturer is responsible for installing / bolting the


metal ground plate in / to the trailer.

• The metal ground plate (refer to the Planning Guide for dimensions) has to be bolted to
the floor of the trailer.
• The metal plate requires threaded holes (refer to the Planning Guide for dimensions).
• The required installation material (screws, anchors, washers, ect.) for the gantry is
included in the delivery volume.

Bolting the Mobile CT Gantry to the Trailer Floor 0

• The gantry has to be bolted to the metal plate with 4 screws (size M12) via the pedes-
tals in the left and right gantry stand. The locations of the 4 holes are shown on the
metal ground plate and cannot be changed.

Fig. 90: Position of additional bracket

• Move the gantry to the installation location and align the holes in the pedestals with the
holes in the metal plate.
• Lower the gantry to the floor, ensuring that the holes in the pedestals and the metal
plate are still aligned.
• Insert the M12 screws through the pedestals into the threaded holes.
• Tighten all 4 screws with a torque wrench to secure the gantry to the floor.
¹ The maximum torque depends on the type of screw being used.
• Continue with (Removing the Transport Device and the 4 yellow Adapters / p. 38). Do
not align the gantry. Fully extend the pedestals (gantry stands have to be level).
• To stabilize the gantry in the trailer, 4 additional brackets are needed for the left and
right gantry stand (refer to (Fig. 90 / p. 75)for an example of the right gantry stand).

Siemens AG CT02-025.812.03.03.02 Page 75 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
76 Special Mounting

• For each bracket, 2 screws have to be attached to the gantry and 2 screws to the metal
plate.
• Tighten all 4 screws with a torque wrench to secure the gantry.
¹ The maximum torque depends on the type of screw being used.
• Continue with (Removing the Transport Locks and Hand Brake / p. 45) and follow the
procedures described.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 76 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 77
5-
General Information
5Electrical Connections and System Wiring

Safety 0

NOTE Read and observe the safety information contained in the "Safety
information" section of this document.(Safety Information / p. 15)

Routing Fiber Optic Cables 0

- Ensure that the cables are not damaged during routing.


- Acceptable minimum bending radius is approx. 80 mm.

Routing Signal and Data Cables 0

NOTE Correct routing of the cables is a prerequisite for reliable system


operation.
Interference on cables may have unusual side effects and cause
problems. Follow instructions carefully.

• Avoid routing system cables next to other equipment cables. The minimum distance to
other equipment cables should be 1 m (3 feet) or more.
• Where possible, have separate cable ducts (shielded) for power and signal cables or
route them at a distance of 5 cm minimum from each other.
• Do not coil up remaining cable length. Route the remaining cable length in a winding or
snaking pattern.

Cables to be prepared on-site (not in delivery volume) 0

Cable/func- Gauge Delivered Comment


tion type length
Power cable Recommendation: In accordance with country-specific regula-
System 5 x 25 mm2 tions and the power consumption of the over-
all system.
Gantry: 5 x 16 mm2
Radiation 2 x 0.5 mm2 for 230 - On-site cable for connecting
release lamp V / 2 A 2 x 1.5 mm2 the floating relay switch or the
for 24 V / 10 A 24 V output voltage.

Siemens AG CT02-025.812.03.03.02 Page 77 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
78 Electrical Connections and System Wiring

Cable/func- Gauge Delivered Comment


tion type length
Emergency In accordance with country-specific regulations. Please refer to
shut-down details in the “Planning Guide”
Door switch 2 x 1.5 mm2 - On-site cable for door switch
Data cable - - On-site cable, e.g., for cam-
era.

Cabling Information 0

Power connection takes place via PDS (Spirit / Spirit Power) or LCB (Emotion 6/16
(2007)).
• All cables have to be routed into the gantry via the cable inlet in the right gantry stand. A
cable inlet is provided in back of the right gantry stand. There is no cable inlet above the
floor in the left gantry stand (only underneath).

NOTE Do not route the power cables into the left gantry stand and
through the cable duct.

• Cables W52 have been pre-installed and connected inside the gantry. The remaining
cable length is placed on top of the right gantry stand.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 78 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 79

Wiring Diagram 5.1

List of System Cables 0

The tables below provide an overview of the different system cables and their correspond-
ing data for Spirit / Spirit Power and Emotion 6/16 (2007). Cables not mentioned in this
table belong to optional system components.
For detailed wiring instructions, refer to (System Wiring and Cable Connections for
SOMATOM Spirit / Spirit Power: / p. 86).

Cables for SOMATOM Spirit / Spirit Power 0

No. Cable type/ func- Item number/gauge Delivered Comment


tion length
W1 Power cable - 2m included in delivery volume
W2 Power cable - 2m included in delivery volume
W3 Power cable - 3m included in delivery volume
W50 Data cable 860 00 04 25 m incl. in delivery volume
W51 Data cable 86 00 012 25 m incl. in delivery volume
W52 Fiber optic cable 86 00 020 30 m incl. in delivery volume,
pre-installed 3)
W53 Data cable - 1m included in delivery volume
W54 Data cable 86 00 962 1.5 m included in delivery volume
W55 Control cable - 3m incl. in delivery volume, for control
box
W56 Keyboard cable - 2m/4m either 2 m (delivered with the key-
board) or 4 m (part of the system)
W57 Mouse cable - 2m/+3m 1.8 m (delivered with the mouse), 3
m for extension (part of the system)
W58 Video cable - 3m included in delivery volume, for
monitor
W59 24V power cable 860 00 38 - incl. in delivery volume,
pre-installed 4)
W60 230V power cable 860 00 46 - incl. in delivery volume,
pre-installed 4)

Siemens AG CT02-025.812.03.03.02 Page 79 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
80 Electrical Connections and System Wiring

No. Cable type/ func- Item number/gauge Delivered Comment


tion length
W61 Ground cable 860 00 53 - incl. in delivery volume,
pre-installed 4)
W62 24V power cable 860 00 61 - incl. in delivery volume,
pre-installed 4)
W63 24V power cable 860 00 79 - incl. in delivery volume,
pre-installed 4)

Cables for Emotion 6/16 (2007) 0

No. Point A Point B Functions/specifications


W1 UPS ICS Supply ICS standard power cord
W2 UPS IRS Supply IRS standard power cord
W3 UPS Monitor Supply monitor, power cord 3 m
W4 Power cable - 3m
W10 LCB UPS PE 10mm2
W11 LCB X 3 UPS inlet connec- Power Supply IMS
tor female 3 x 1.5mm2
W20 UPS UPS housing UPS PE-bridge 6mm2, length
0.3m
W50 Gantry, MAS IRS D31. X 2 CAN, Stop circuit, audio
D301. X 100 24/200mA
W51 Gantry MAS IRS D32 X 1 CAN 24V/200mA
D301,X101 With shield, twisted pairs
W52 Gantry SAS, X2 IRS DMS Data fiber optics
D24, X1
W53 UPS ICS, COM port Serial interface
W54 ICS IRS Supply IMS Ethernet
W55 ICS CIB Board Control Box Control keys, Potentiometers,
X1 Control, Audio
W56 ICS Keyboard Keyboard cable
W57 ICS Mouse Supply IMS
PS/2

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 80 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 81

No. Point A Point B Functions/specifications


W58 ICS Monitor DVI
Delivery part
Included in monitor
W59 Gantry, MAS PHS, Vert_Pos.X3 Vertical position, 24VDC/200mA
D301. X 32
W60 Gantry PSS PHS, Vert_Mot.X1 Vertical motor power 230V
D352, X8 AC/2A
W61 Gantry ground PE_PHS. X5 Gantry-PHS ground cable
bar X399 2.5mm2gnge
W62 Gantry, MAS PHS, Hor_Pow.X4 Horizontal motor power 24V
D301. X33 DC/3A
W63 Gantry, MAS PHS, Vertical brake 24V DC/2A
D301.X35 Vert_brake.X2

Siemens AG CT02-025.812.03.03.02 Page 81 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
82 Electrical Connections and System Wiring

Wiring Diagram: Standard Components for SOMATOM Spirit / Spirit Power 0

Fig. 91: Wiring diagram for SOMATOM Spirit

Footnotes for SOMATOM Spirit / Spirit Power:


No. Cable type/ Item number/gauge Delivered Comment
function length
1) 2-wire cable 2 x 0.5 mm2 for - On-site cable for connecting the
230V / 2A 2 x 1.5 floating relay switch contact in the
mm2 for 24V / 10 A right gantry stand.
2) 2-wire cable 2 x 0.5 mm2 - On-site cable for door switch; con-
nection in the right gantry stand.
3) Cables W50, W51 are in the delivery volume.W52 is pre-installed in the gantry ( internal
cable routing length is 7.20 m).

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 82 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 83

Footnotes for SOMATOM Spirit / Spirit Power:


4) Cables W59 to W63 between the gantry and the patient table are pre-installed in the gan-
try.
5) UPS power - 230V =n.a. 120V is delivered with the UPS vol-
cable 120V = 2m ume.
Other country spec's are prepared
by the local project manager.
6) For UPS 230 V power supplies, an additional permanent PE connection of 2.5 mm 2 has
to be established as well between the PE connection of the UPS (M5 screw) and an
on-site PE connection clamp.(or MCC PE connection for users in China.)The PE cable is
not included in the system delivery volume.It is included in the MCC (P/N:
10049758) delivery volume.
7) On-site power - - In accordance with country-specific
lead regulations and the power con-
sumption of the entire system
8) On-site data - - On-site cable, e.g., for camera.
cable
9) With IMS container (optional) 1300 mm; without IMS container 500 mm.

Siemens AG CT02-025.812.03.03.02 Page 83 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
84 Electrical Connections and System Wiring

Wiring Diagram: Standard Components for Emotion 6/16 (2007) 0

Fig. 92: Cabling for Emotion 6/16 (2007)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 84 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 85

Cabling Information for SOMATOM Spirit / Spirit Power: 0

Fig. 93: Cable inlet


• All cables must be routed into the gantry through the cable inlet in the right gantry stand.
A cable inlet is provided in back or below (Fig. 93 / p. 85) the right gantry stand.

Siemens AG CT02-025.812.03.03.02 Page 85 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
86 Electrical Connections and System Wiring

System Wiring and Cable Connections for SOMATOM Spirit /


Spirit Power: 5.2

• The following pages show all cable connections to be performed at the system during
Spirit installation.
• The table shows where each cable has to be connected, the cable ID, and where the
cable comes from.

Gantry (right stand from front side) 0

(Fig. 94 / p. 86)

Cable connects to: Cable ID: Cable comes from:


X3: L1 / L2 / L3 / N / PE --- (local supply) On-site line power distribu-
tion

Fig. 94: Connections at PDS

Gantry (right stand from the back): 0

NOTE Connect the door switch and radiation warning lamp as shown in
(Fig. 95 / p. 87).

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 86 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 87

Cable connects to: Cable-ID: Cable comes from:


X14 Pins 1-2 *1) --- (local supply) Radiation warning indica-
tor (on-site)
X14 Pins 1-3 *1) --- (local supply) Radiation ready indicator
(on-site)
X13 pins 2-3 *2) --- (local supply) Door switch (on-site)
D301.X100 W50 ICS
D301.X101 W51 IRS
O/E converter *3) W52 (fiber optic cable) IRS

*1) A potential-free contact is supplied at X14


*2) A jumper must be installed if the door switch is not connected.
*3) Cables W52 are preinstalled inside the gantry and coiled up in the right gantry stand.
Route the cables from the right gantry stand to the IRS accordingly.

Fig. 95: D301: door switch and radiation warning lamp connection
Pos. 1 X13 (door switch)
Pos. 2 X14 (radiation warning lamp)
Pos. 3 Shield
Pos. 4 Room door
Pos. 5 Switch
Pos. 6 Common
Pos. 7 X-ray lamp
Pos. 8 Ready lamp

Siemens AG CT02-025.812.03.03.02 Page 87 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
88 Electrical Connections and System Wiring

Fig. 96: D301: X100 and X101 connection


Pos. 1 X100 (W50)
Pos. 2 X101 (W51)
Pos. 3 X14 (radiation warning lamp)
Pos. 4 X13 (door switch)

Patient Table 0

Cable connects to: Cable-ID: Cable comes from:


X1 W60 Gantry (preinstalled)
X2 W63 Gantry (preinstalled)
X3 W59 Gantry (preinstalled)
X4 W62 Gantry (preinstalled)
X5 / PE W61 Gantry (preinstalled)

Note: The cables shields (X1 / 2 / 4) have to be secured to the clip


(Fig. 97 / p. 89)(Fig. 98 / p. 89)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 88 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 89

Fig. 97: PHS cable connections


Pos. 1 X5 PE
Fig. 98:
Pos. 1 Clip for shield

Note: The cables have to be routed between the PHS base and the spring mount.
(Fig. 99 / p. 89)

Fig. 99: PHS cable relocation


Pos. 1 Cable relocation inside the PHS

Computers ICS and IRS 0

General Information 0

The cabling template on the side cover of ICS and IRS contains the label regarding exter-
nal connectors.

ICS Computers 0

- ICS (Navigator) is labeled with "ICS Tower "

Siemens AG CT02-025.812.03.03.02 Page 89 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
90 Electrical Connections and System Wiring

NOTE The PE connector at the Celsius R 610 computer housing is not


required.
The SCSI (LVD/SE) terminator has to be installed (at slot 7) when
working without the Pioneer MOD.

Cabling for ICS (SNR 8879426)

NOTE Required connections can change quickly. The current cabling


schematic is attached to the side cover of the ICS computer
(Fig. 100 / p. 91)

Cable connects to: Cable ID: Cable comes from:


ICS power connector W1 UPS: Power output
D31.X2 W50 Gantry right: D301.X100
D31.X11 W53 UPS
ICS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

D31.X1 W55 Control box


Keyboard port W56 Keyboard
Mouse port W57 Mouse
Monitor port DVI to W58 Monitor
D-SUB

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 90 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 91

Fig. 100: ICS Tower 6

NOTE For 17" monitor (Part No.:07733392)


The ICS graphics board has a DVI-OUT only. An adapter from DVI
TO 15PIN D-SUB is needed for DVI-Port 1 and it is delivered with
the system.
Do not use the delivered DVI - cable!

NOTE For 19" monitor


The DVI to D-SUB adapter is not needed. The monitor DVI cable is
connected to ICS directly.

Cabling for ICS (SNR 8879582)

Siemens AG CT02-025.812.03.03.02 Page 91 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
92 Electrical Connections and System Wiring

Cable connects to: Cable ID: Cable comes from:


ICS power connector W1 UPS: Power output
D31.X2 W50 Gantry right: D301.X100
D31.X11 W53 UPS
ICS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

D31.X1 W55 Control box


Keyboard port W56 Keyboard
Mouse port W57 Mouse
Monitor port DVI to W58 Monitor
D-SUB

Fig. 101: ICS tower 7B


Cabling for ICS (SNR 10143563)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 92 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 93

Cable connects to: Cable ID: Cable comes from:


ICS power connector W1 UPS: Power output
D31.X2 W50 Gantry right: D301.X100
D31.X11 W53 UPS
ICS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

D31.X1 W55 Control box


Keyboard port W56 Keyboard
Mouse port W57 Mouse
Monitor port DVI to W58 Monitor
D-SUB

Fig. 102: ICS tower 8B

IRS tower 0

The IRS computer is labeled with "IRS Tower "


For IRS Tower: Celsius R 610 (SNR: 7733418)

Siemens AG CT02-025.812.03.03.02 Page 93 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
94 Electrical Connections and System Wiring

NOTE The PE connector at the Celsius R 610 computer housing is not


required.

Fig. 103: IRS tower (schematics)


Pos. 1 W2 (UPS)
Pos. 2 W54 (Ethernet)
Pos. 3 W52 (Fiber optic)
Pos. 4 W51

Cable connects to: Cable ID: Cable comes from:


Power connector W2 UPS: Power output
D32, X1 W51 Gantry right: D301, X101
D23, X1 (fiber optic) W52 Gantry SAS, X2
IRS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

For IRS Tower: Tower 7B (SNR: 7734556)

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 94 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 95

Fig. 104: ICS tower 7B (schematics)


Pos. 1 W2 (UPS)
Pos. 2 W54 (Ethernet)
Pos. 3 W52 (fiber optic)
Pos. 4 W51

Cable connects to: Cable ID: Cable comes from:


Power connector W2 UPS: Power output
D32, X1 W51 Gantry right: D301, X101
D23, X1 (fiber optic) W52 Gantry SAS, X2
IRS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

Siemens AG CT02-025.812.03.03.02 Page 95 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
96 Electrical Connections and System Wiring

Fig. 105: ICS tower 8B (schematics)


Pos. 1 W2 (UPS)
Pos. 2 W54 (Ethernet)
Pos. 3 W52 (fiber optic)
Pos. 4 W51

Cable connects to: Cable ID: Cable comes from:


Power connector W2 UPS: Power output
D32, X1 W51 Gantry right: D301, X101
D23, X1 (fiber optic) W52 Gantry SAS, X2
IRS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 96 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 97

UPS 0

Fig. 106: UPS 750VA: 230V Version (P/N:03806739)


Pos. 1 W53 (ICS)
Pos. 2 W3 (Monitor)
Pos. 3 W2 (IRS)
Pos. 4 W1 (ICS)
Pos. 5 UPS battery connector
Pos. 6 PE terminal
Pos. 7 Line power

Siemens AG CT02-025.812.03.03.02 Page 97 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
98 Electrical Connections and System Wiring

Fig. 107: UPS 1500VA:230V Version (part no.: 07733376)


Pos. 1 Serial connection to ICS (W53)
Pos. 2 Connection points for safety cover (Plastic comb)
Pos. 3 PE connector
Pos. 4 Power output
Pos. 5 Power input 230V

Fig. 108: UPS 750VA: 110V Version (P/N:03806747)


Pos. 1 W53 (ICS)
Pos. 2 W3 (Monitor)
Pos. 3 W2 (IRS)
Pos. 4 W1 (ICS)
Pos. 5 UPS battery connector
Pos. 6 PE terminal
Pos. 7 Line power

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 98 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 99

Fig. 109: UPS 1500VA:110V Version (part no.: 07733368)


Pos. 1 Serial connection to ICS (W53)
Pos. 2 Connection points for safety cover (Plastic comb)
Pos. 3 PE connector
Pos. 4 Power output

Cable connects to: Cable ID: Cable comes from:


Line power connector UPS power cord On-site power receptacle
PE terminal PE 2.5 mm2 *1) On-site PE terminal
(local supply)
Power output W1 ICS: power input
Power output W2 IRS: power input
Power output W3 Monitor: power input
Control port W53 ICS: D31.X11

NOTE For APC UPS 750 VA, the UPS battery connector should be con-
nected.

NOTE For APC UPS 750VA: 230V version (P/N:03806739) and 110V ver-
sion (P/N:03806747), the P/N of the W53 cable is 09600954.
For MGE UPS 1500VA: 230V version (P/N:07733376) and 110V ver-
sion (P/N:07733368), the P/N of the W53 cable is 07733079.

Siemens AG CT02-025.812.03.03.02 Page 99 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
100 Electrical Connections and System Wiring

NOTE The part number of cable W53 is different for APC UPS 750VA and
MGE UPS 1500VA. If the APC UPS is replaced by the MGE UPS,
the signal cable (W53) delivered with the MGE UPS is needed!

Monitor 0

Cable connects to: Cable-ID: Cable comes from:


Power connector W3 UPS: Power output
VGA (17" monitor, part no.: 07733392) W58 ICS: Graphic card
DVI (for 19" monitor, part no.:
08675402 or 08598919)
PE cable (for 19” monitor, part no.: n.a. UPS: PE connector
08598919, revision 2 or higher)

NOTE Only for 17" monitors (part no.:0733392)


The access cover of the 17" monitors fits very firmly. For this rea-
son, you may need some force to remove it. Proceed with care,
because the cover damages very easily.

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 100 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 101

System Wiring and Cable Connections for SOMATOM Emotion


(2007) 5.3

• The following pages show all cable connections to be performed at the system during
Emotion (2007) installation.
• The table shows where each cable has to be connected, the cable ID, and where the
cable comes from.

Line Connection Box 0

Fig. 110: Cable Connection in Line Connection Box

NOTE Clamp the LCB transformer in accordance with the incoming volt-
age listed on the label. Also refer to (Adaption to Line Voltage / P. )
Check the output voltage (400V) at X2 prior to connecting the gan-
try power supply cable at the LCB.

Connect the following cables in the LCB:


• Line power to X1 (L1, L2, L3, N, PE / 5 X 25 mm)
• Gantry power supply X2 (L1, L2, L3, PE / 5 X 16 mm or 5 x 25 mm )
• UPS connection to x3 (cable and “WAGO” connector included in delivery volume): con-
nect separate PE wire to X7
• PE cable from IRS to LCB PE terminal

Siemens AG CT02-025.812.03.03.02 Page 101 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
102 Electrical Connections and System Wiring

NOTE Some system parts are still supplied with dangerous voltage lev-
els even when the system is switched off (e.g. slip ring, DPU, DMS,
UPS and parts suppied with UPS power (IRS/ICS/IES), preparation
and/or X-ray warning lamp).
Risk of electric shock!
The switching-off instructions contained in service documents
have to be closely followed.

• To connect the Emergency Shutdown button, refer to the Planning Guide.


• To establish an external on/off switch, remove the jumper between X6.1 and X6.2
(deactivates S1). Route the wire between X6.1 and X6.3
• The connection between the LCB and gantry is established using the 5 pin W12
cable(included in the delivery volume). Connection points: X100 in the gantry and X8
(“WAGO connector”) in the LCB.
• Connect the NO contact of the “Door Switch” at X8.4 and X8.5.
• Service outlet X10 remains under voltage (230V) when the system contactor is off.
• External “Radiation Release” and “Ready” lamps: The lamps are provided with volt-
age from an external source (max. 230V, 2A).
- The “Radiation Release” lamp is connected at X9.1 / X9.2
- The “Ready” lamp is connected at X9.3/ X9.4.

NOTE When using external voltage (on-site voltage supply), the cable
has to be labeled accordingly.

• External “Radiation Release” and “Ready” lamps (24V):


to connect them, remove jumpers X8.3-7 (Ready) and X8.2-8 (Radiation Release).
Relays K3 and K4 (contact X9) are now inactive.

X8.2 X8.6 X8.3


24 V for “Radiation 0V 24 V for “Ready”
Release”

NOTE If LCB without transformer (400V Version): Connector X12 (trans-


former temperature switch) has to be jumpered for this version of
the LCB.

• Check the setting for residual current switch :

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 102 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 103

Gantry (right stand from front side) 0

Cable connects to: Cable ID: Cable comes from:


X3: L1 / L2 / L3 / N / PE --- (LCB supply) LCB power distribution
PDS. X 100 W12 LCB. X 8

Fig. 111: Connections at the PDS

Gantry (right stand from the back): 0

NOTE Connect the door switch and radiation warning lamp as shown in
(Fig. 95 / p. 87).

Siemens AG CT02-025.812.03.03.02 Page 103 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
104 Electrical Connections and System Wiring

Cable connects to: Cable-ID: Cable comes from:


X14 Pins 1-2 *1) --- (local supply) Radiation warning indica-
tor (on-site)
X14 Pins 1-3 *1) --- (local supply) Radiation ready indicator
(on-site)
X13 pins 2-3 *2) --- (local supply) Door switch (on-site)
D301.X100 W50 ICS
D301.X101 W51 IRS
O/E converter *3) W52 (fiber optic cable) IRS

*1) A potential-free contact is supplied at X14


*2) A jumper must be installed if the door switch is not connected.
*3) Cables W52 are preinstalled inside the gantry and coiled up in the right gantry stand.
Route the cables from the right gantry stand to the IRS accordingly.

Fig. 112: D301: door switch and radiation warning lamp connection

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 104 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 105

Patient Table 0

Cable connects to: Cable-ID: Cable comes from:


X1 W60 Gantry (preinstalled)
X2 W63 Gantry (preinstalled)
X3 W59 Gantry (preinstalled)
X4 W62 Gantry (preinstalled)
X5 / PE W61 Gantry (preinstalled)

Note: The cables shields (X1 / 2 / 4) have to be secured to the clip


(Fig. 113 / p. 105)(Fig. 114 / p. 105)

Fig. 113: PHS cable connections


Pos. 1 X5 PE
Fig. 114:
Pos. 1 Clip for shield

Note: The cables have to be routed between the PHS base and the spring mount.
(Fig. 115 / p. 105)

Fig. 115: PHS cable relocation


Pos. 1 Cable relocation inside the PHS

Siemens AG CT02-025.812.03.03.02 Page 105 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
106 Electrical Connections and System Wiring

Computers ICS and IRS 0

General Information 0

The cabling template on the side cover of ICS and IRS contains the label regarding exter-
nal connectors.

ICS Computers 0

- ICS (Navigator) is labeled with "ICS Tower "


Cabling for ICS (SNR 10143563)

NOTE Required connections can change quickly. The current cabling


schematic is attached to the side cover of the ICS computer
(Fig. 116 / p. 107)

Cable connects to: Cable ID: Cable comes from:


ICS power connector W1 UPS: Power output
D31.X2 W50 Gantry right: D301.X100
D31.X11 W53 UPS
ICS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

D31.X1 W55 Control box


Keyboard port W56 Keyboard
Mouse port W57 Mouse
Monitor port DVI W58 Monitor

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 106 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 107

Fig. 116: ICS tower 8B

NOTE For 19" monitor (part no.:08598919)


The DVI to D-SUB adapter is not needed. The monitor DVI cable is
connected to ICS directly.

IRS tower 0

The IRS computer is labeled with "IRS Tower "


IRS Tower 8F: (SNR: 08879806) (Only for Emotion 6(2007))

Siemens AG CT02-025.812.03.03.02 Page 107 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
108 Electrical Connections and System Wiring

Fig. 117: ICS tower 8F (schematics)


Pos. 1 W2 (UPS)
Pos. 2 W54 (Ethernet)
Pos. 3 W51
Pos. 4 W52 (fiber optic)

Cable connects to: Cable ID: Cable comes from:


Power connector W2 UPS: Power output
D32, X1 W51 Gantry right: D301, X101
D23, X1 (fiber optic) W52 Gantry SAS, X2
IRS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

IRS Tower 8G: (SNR: 08879814) (Only for Emotion 16(2007))

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 108 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 109

Fig. 118: ICS tower 8G (schematics)


Pos. 1 W2 (UPS)
Pos. 2 W54 (Ethernet)
Pos. 3 W51
Pos. 4 W52 (fiber optic)

Cable connects to: Cable ID: Cable comes from:


Power connector W2 UPS: Power output
D32, X1 W51 Gantry right: D301, X101
D23, X1 (fiber optic) W52 Gantry SAS, X2
IRS.DATA W54 IRS.DATA
On-board Ethernet On-board Ethernet

Siemens AG CT02-025.812.03.03.02 Page 109 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
110 Electrical Connections and System Wiring

UPS 0

Fig. 119: MGE Pulsar M2200va (part no.: 077350990)


Pos. 1 Serial connection to ICS (W53)
Pos. 2 Power output
Pos. 3 Mains

Cable connects to: Cable ID: Cable comes from:


Line power connector UPS power cord On-site power receptacle
PE terminal PE 2.5 mm2 *1) On-site PE terminal
(local supply)
Power output W1 ICS: power input
Power output W2 IRS: power input
Power output W3 Monitor: power input
Control port W53 ICS: D31.X11

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 110 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Electrical Connections and System Wiring 111

Monitor 0

Cable connects to: Cable-ID: Cable comes from:


Power connector W3 UPS: Power output
DVI (19" monitor, part no.: 08598919) W58 ICS: Graphic card
PE cable (for 19” monitor,part no.: n.a. UPS: PE connection
08598919, revision 2 or higher)

For revision 2 or higher level monitor, there is one added 5 V power supply at the back
socket.

Fig. 120: 5V DC supply in the back socket (only for revision 2 or higher)

Siemens AG CT02-025.812.03.03.02 Page 111 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
112 Electrical Connections and System Wiring

Line Connection Cabinet (400V) / Mains Connection Cabinet


(China) 5.4

For detailed information on cable connections, refer to the Planning Guide: "Specific
requirements for CT Project Planning"

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 112 of 114 Siemens AG


10.07 SSME CS Medical Solutions
Final steps 113
6Final steps 6-
After all chapters of this document are completed, the mechanical installation of the
SOMATOM Spirit / Spirit Power or SOMATOM Emotion 6/16 (2007) is finished, including
system wiring. Continue the installation with document CT02-025.815.02.xx "Startup of
the CT system".

NOTE Do not leave the system components open for an extended


uncontrolled period of time.
If the system installation cannot be continued with start-up due to
unfinished room preparations (e.g. missing power), the installer
should close all covers (gantry, patient table and computer cabi-
net) according to chapter 3 "Installing all covers" of document
CT02-025.815.02.xx"Startup of the CT system", to prevent ESD
damage by unskilled persons or contamination.

Siemens AG CT02-025.812.03.03.02 Page 113 of 114 Emotion (2007) / Spirit


Medical Solutions 10.07 SSME CS
114 Changes to Previous Version
7Changes to Previous Version 7-

Chap- Change Reason


ter/page
All Initial Version n.a.
All V2: Add SOMATOM Spirit Power n.a.
Chapter 01 V3: Add Doc list for VB35 on Spirit, ECR510264
Chapter 03 Add description for new cable duct between ECR513333
gantry and PHS

Chapter 05 Add two notes for UPS signal cable (W53) ECR513468

Emotion (2007) / Spirit CT02-025.812.03.03.02 Page 114 of 114 Siemens AG


10.07 SSME CS Medical Solutions

You might also like