Cygnus DIVE
Cygnus DIVE
CYGNUS
DIVE
ULTRASONIC THICKNESS GAUGE
IDEAL FOR
USE IN
WWW.CYGNUS-INSTRUMENTS.COM
CYGNUS DIVE
•
CYGNUS DIVE Multiple-Echo mode for accurate, through-coat
measurements as specified by Classification Societies
KEY FEATURES • Single-Echo and Echo-Echo modes for heavily corroded
metals with a thin or no coating
• Deep-Coat function to ignore coatings up to 20mm
thick
• A-scan display to assist with measurement verification
• Optional Data Logging feature with AutoLog - storing
up to 5000 measurements including A-scan data
without pressing any button
• MSI (Measurement Stability Indicator) verifies stable,
reliable readings
• Rechargeable lithium-ion battery giving up to 10
hours continuous measurement
• Optional remote display HelmetViewTM, Topside
Repeater and CygLink surface display software
Call our team today on +44 (0) 1305 265 533 for expert product advice
CYGNUS DIVE SPECIFICATION
Feature Description
2.8” quarter VGA colour LCD (320 x 240 pixels). Large clear thickness measurement
(15mm high numbers), viewable from all angles.
Display
A-scan display with automatic X axis. Battery level, signal strength, probe type,
velocity. Measurement mode and units indication
Single 3.6V Li-ion 8.2 W battery. 10 hours continuous measurement
Battery
Low battery warning ‘alert’ message
Multiple-Echo using 3 echoes to ignore coatings up to 20mm thick.
Measuring Modes Echo-Echo using 2 echoes to ignore coatings up 1mm thick.
Single-Echo using 1 echo
In Multiple-Echo mode, allows measurement to be made through thicker coatings of
Deep Coat
suitable materials of up to 20mm thick
±0.1 mm or 0.1% of thickness measurement, whichever is greatest, when calibrated in
Accuracy
accordance with Cygnus Instruments calibration procedure
Probe Options Single crystal probes and Twin crystal probes
Double outer jacket in tough PU. Coloured yellow for easy locations underwater.
Probe Cables
Coiled for ease of use. Fischer S105 series connectors.
Measurement 1 - 250 mm (0.039 - 10 in.) depending on selected probe and configuration, material
Range in Steel and temperature
Measurement Multiple-Echo with single crystal probes - 0.1 or 0.05 mm
Resolution Single-Echo and Echo-Echo with twin crystal probes - 0.1 or 0.01 mm
Measurement
mm or inches
Units
Probe Zero Fully automatic probe zeroing for all probes types
V-Path
Automatic V-Path correction for all twin crystal probes
Correction
Velocity Range 2000 to 9000 m/s in 1 m/s steps
Pulser Twin channel 70 V spike pulser
10 MHz bandwidth, 120 dB range, automatic TCG.
Receiver/Amplifier
60 MHz measurement time-base
Data Logging One-handed automatic logging of stable measurements. Capacity for up to 5000
(optional) points including 640 point A-scan data
Data Output RS-485 single pair, half duplex for surface connection
Standards Designed for EN 15317
Compliance CE, UKCA, RoHS
Warranty 3 years on gauge and 6 months on probe
ISS7 03/21
All information provided is subject to change without prior notice.