LMB Biosealer CR 4 User's Manual and Technical Reference
LMB Biosealer CR 4 User's Manual and Technical Reference
including
Bench Unit
Manual Handle
Ergonomic Handle MSH-III
User’s Manual
and
Technical reference
Rev 2012-11-09
CR 4 AA User’s Manual
1. Important
This user’s manual is written for the person responsible for the operation of Biosealer CR 4.
The operational methods and routines are developed and tested to ensure a reliable, safe and
effective operation of CR 4. It is important that the operator has studied and understood the
contents of this manual before using CR 4.
Throughout this manual, CR4 means the same thing as CR4 AA.
• This instrument is a sealing equipment using radio frequency for welding and the
electrical emission at the operating frequency 40.68 MHz is high. Make sure that other
instruments and equipment near the sealing unit can withstand this emission.
• Never touch the electrodes with your fingers during the sealing process! This may cause
burn damages!
• Ensure that the PVC tube is dry on the surface.
CR4 contains electronic components which are classified as dangerous for the environment
and the equipment must be left to a recycling station when disposed.
2. Warranty
2.1. Ljungberg & Kögel AB, hereby guarantees the original buyer that Biosealer CR 4 is
manufactured in a professional and quality manner, and will be free from all faults
during a period of one year from the date of delivery from L&K.
2.2. The warranty includes equipment or components that prove to have faults during the
warranty period. L&K will without cost for the customer, repair or replace the
equipment that is faulty.
2.3. The warranty is not valid if the equipment has been repaired by anyone else than
qualified personnel, approved by L&K.
2.4. The warranty is not valid if the equipment has been changed in any way that
according to L&K:s opinion, affects the reliability or stability of the instrument.
2.5. The warranty is not valid when the serial number has been changed, crossed over or
been removed, or if the fault has been caused by misuse or abnormal use.
2.6. In these cases L&K or L&K:s representative will inform the customer about the
decision, and if wished by the client will repair the equipment for normal rate. An
estimated price can be given on request.
2 (20)
CR 4 AA User’s Manual
3. Table of contents
1. Important .............................................................................................................................................. 2
2. Warranty ............................................................................................................................................... 2
3. Table of contents .................................................................................................................................. 3
4. Properties .............................................................................................................................................. 4
5. Description ........................................................................................................................................... 4
5.1. Power Unit .................................................................................................................................... 4
5.2. Bench Unit .................................................................................................................................... 5
5.3. Handle Unit. ................................................................................................................................. 5
5.4. Ergonomic Sealing Handle (MSH-III)......................................................................................... 6
6. Operation with the Bench Unit ............................................................................................................. 7
6.1. Connection of Bench Unit ............................................................................................................ 7
6.2. Sealing .......................................................................................................................................... 7
7. Operation with the Handle Unit ............................................................................................................ 8
7.1. Connection of Handle Unit ........................................................................................................... 8
7.2. Sealing .......................................................................................................................................... 8
8. Operation with the Ergonomic Handle Unit ......................................................................................... 9
8.1. Connection of Ergonomic Handle Unit MSH-III......................................................................... 9
8.2. Sealing .......................................................................................................................................... 9
9. Adjustments ........................................................................................................................................ 10
9.1. Change to manual timer .............................................................................................................. 10
10. Maintenance ................................................................................................................................... 10
10.1. Cleaning the electrodes. .......................................................................................................... 10
10.1.1. Bench Unit .......................................................................................................................... 10
10.1.2. Handle unit ......................................................................................................................... 11
10.1.3. Ergonomic Handle MSH-III ............................................................................................... 11
11. Technical description...................................................................................................................... 12
12. Trouble shooting ............................................................................................................................. 13
13. Technical data................................................................................................................................. 14
13.1. Power Unit .............................................................................................................................. 14
13.2. Bench Unit .............................................................................................................................. 14
13.3. Handle Unit ............................................................................................................................ 14
13.4. Ergonomic Handle Unit .......................................................................................................... 14
14. Spare parts list Power Unit ............................................................................................................. 15
15. Spare parts list Handle Unit ............................................................................................................ 16
16. Spare parts list Bench Unit ............................................................................................................. 17
17. Spare parts list Ergonomic Sealing Handle MSH-III ...................................................................... 18
MDD Declaration ....................................................................................................................................... 19
Rev history
3 (20)
CR 4 AA User’s Manual
4. Properties
4.1. Model for continuous use. Biosealer CR 4 is built for sealing PVC-tubes, especially
blood bags tubes or sets for plasmaferes. CR 4 has a powerful HF-unit (high
frequency), which makes it suitable both for routine procedures at donation rooms and
repeated operations at preparation rooms without overheating.
4.2. Bench Unit and Handle Unit. The Power Unit can either be connected to a handle
Unit and a Bench Unit or two Handle Units. Bench Unit automatically seals when a
tube is in position. The Handle Unit is easy to move and therefore it is easy to reach
the part of the tube which shall be sealed. Both units has a 1,9 m cable. The electrodes
of the Handle Units are well protected by a plastic cover and are easy to clean.
4.3. Wide and safe seal. The high frequency HF gives a wide and safe seal of about 3mm
with a marking in the middle for making it easier to pull the tubes apart.
5. Description
4 (20)
CR 4 AA User’s Manual
STATUS: BI-colour LED indicating power on (fixed green), sealing (blinking orange)
or some error (blinking red).
TEMP ALARM: Red LED if HF generator is overheated.
Indication
lamp for HF
on
Electrodes
Protection
cover
5 (20)
CR 4 AA User’s Manual
(1) Sealing indication. Is blinking blue during sealing process and turns to green when
ready.
(2) Electrodes. Transfers the high frequency energy to the PVC tube. The tube to be sealed
must be placed between the electrodes.
(3) Protection cover. Protects electrodes from damage and also prevent unintentional
finger contact with electrodes.
(4) Coaxial cable
(1)
(2)
(3)
(4)
6 (20)
CR 4 AA User’s Manual
• Always turn the power off before you connect or disconnect a cable or both cables.
• Always use the coax cable that is delivered with the equipment. It is not allowed to extend
the cable. If you need a longer cable, then order a longer cable, see spare parts list.
6.2. Sealing
Note !
Do not pull the PVC-tube when sealing.
Note !
When sealing the PVC-tube on two or more places the distance between the seals must be
more than 50 mm. Shorter distance may cause leakage.
7 (20)
CR 4 AA User’s Manual
Always turn the power off before you connect or disconnect the coax cable.
7.2. Sealing
Note !
When sealing the PVC-tube on two or more places the distance between the seals must be
more than 50 mm. Shorter distance may cause leakage.
8 (20)
CR 4 AA User’s Manual
Always turn the power off before you connect or disconnect the coax cable.
8.1.1. Connect the Handle Unit to the HF-contact on the Power Unit with the coax
cable.
8.2. Sealing
Note !
When sealing the PVC-tube on two or more places the distance between the seals must be
more than 50 mm. Shorter distance may cause leakage.
9 (20)
CR 4 AA User’s Manual
9. Adjustments
If you have hard sealed tubes that needs longer timer i.e. more time for the high frequency,
you can disconnect the automatic state by turning the potentiometer TP2 on the timer PC
board clockwise with a small screwdriver. Turning it more than 20% turns on the manual
timer and the timer will be set from 2 to 10 sec. (potentiometer 20-100%). To be able to locate
TP2 you have to lift off the cover by unscrewing the 4 screws on the side.
TP2
10. Maintenance
10 (20)
CR 4 AA User’s Manual
11 (20)
CR 4 AA User’s Manual
Note!
Due to regulations of maximum emission (radiation) at other frequencies than 40,68 MHz the
HFG-01 module is soldered up and shall be regarded as a component. It can not and it is not
allowed to be adjusted or repaired by anyone else than the manufacturer. In the event of
malfunction of HFG-01 it must be sent to the manufacturer for repair.
The voltage that occurs during sealing may cause burn damages if someone touches the hot
electrode during the sealing process. Therefore it is absolutely necessary to turn off the power
before cleaning or service of the Handle Unit or Bench Unit.
12 (20)
CR 4 AA User’s Manual
Do not adjust the electrodes by yourself. They can be damaged by erroneous performed
adjustment.
Sealing does not start and the diode ”Status” • Test another handle.
is not blinking when the handle is pressed. • Change the coax cable
The diode ”Status” is blinking orange but not • Test another handle.
the lamp on the handle or the Bench Unit. • Test another Bench Unit
• Change the coax cable
• Change the HF-generator. (Only for service
personnel).
The diode ”Temp Alarm” is lit and sealing • The HF-generator is overheated. Let it cool
does not start. down until the alarm is turned off.
The diode ”Status” flashes red and sealing • The tube is damp, dry it.
does not start.
Sparks appear by the electrodes during • This can happen if the electrodes are
sealing. overheated due to too many seals in a short
time. Let the electrodes cool down.
Sparks appear by the electrodes during • If there has been a leakage and flashes have
sealing. created carbon particles on the electrodes,
these particles must be removed
completely before sealing can continue.
Sparks appear by the electrodes during • It can indicate that the electrodes are not
sealing. parallel. Contact service personnel.
The seal is done but with bad quality, leakage • Check that the coax cable is the right type,
may appear. the length must be 1.9 m for the Bench
Unit and the Manual Handle. The
Ergonomic Handle must have the coiled
cable. Note! The cables must not be
shortened or lengthened.
Short-circuit is shown on display or Status is • The electrodes are getting hot and the tubes
blinking red during a seal, blue lamp is turned melts easier which may cause the
off. electrodes to shorten during sealing. Let
the electrodes cool down by pausing for
10-15 minutes.
13 (20)
CR 4 AA User’s Manual
14 (20)
CR 4 AA User’s Manual
12
11 1
2
3
10
4
15
8
5
13
6
7
9-50401-00 Bottom plate (chassis) 9-50408-00 24V power contact incl cable
9-50402-00 Rubber feet x 4 (Not in 9-50409-00 Bench unit internal signal
picture) cable
9-50403-00 Power supply 200W 9-50410-00 HF generator incl. timer
board
9-50404-00 20-pole FFC cable 9-50411-00 Mains inlet
9-50405-00 Front board CR5 9-50412-00 Mains internal cable
9-50406-00 Signal contact bench unit 9-50413-00 Ferrite core φ 29mm
9-50407-00 Internal coax cable 9-50414-00 Cover (Not in picture)
9-50415-00 2-pole input power contact
15 (20)
CR 4 AA User’s Manual
15
1
14
10
11
5 4
2 16
8 3 13
9-32401-00 Cover
9-32402-00 BNC chassis female contact
9-32403-00 Switch mom close
9-32404-00 Internal coax cable
9-32405-00 Coil incl. delrin core and dim glowing lamp
9-32406-00 Dim glowing lamp
9-32408-00 Plate spring
9-32409-00 Handle
9-32410-00 Chassis
9-32411-00 Pin
9-32413-00 Rubber tile
9-32414-00 Electrode house complete
9-32415-00 Electrode cover
9-32416-19 Coax cable 1.9m
9-32416-42 Coax cable 4.2m
9-32416-91 Coax cable 9.1m
16 (20)
CR 4 AA User’s Manual
5 12 19 7
6
1
8
11
21 9
10
14 18
13
15 16 22 17
4 3
20
17 (20)
CR 4 AA User’s Manual
15
13
11
1 12 4
14
9
10
6
8
5
2 7
3 17 18
16
19
18 (20)
MDD Declaration
DECLARATION OF CONFORMITY
according to the
Medical Devices Directive 93/42/EEC
and
RoHS directive, 2011/65/EU
STANDARDS TO WHICH CONFORMITY IS DECLARED:
EN 60 601-1 Safety
EN 60 601-1-2 EMC
Including: EN 55011 Kl B, EN 61000-4-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5
Marketing: Ljungberg&Kögel AB
Address: Box 1032, 251 10 Helsingborg
Telephone: +46 042-139860
Telefax: +46 042-132181
I, the undersigned, hereby declare that the equipment specified above conforms to the above
Directive and Standards.
/ Thommy Lundström
Position/title: President
CR 4 AA User’s Manual