EasyView QA Specification
EasyView QA Specification
SPECIFICATION
1 Glossary
Administrator General term: A person having special access, or other, rights in
a system.
EasyView Administrator A person having special operational rights in EasyView.
EasyView policies Settings in EasyView governing administrator rights, login and
signing functionality and responses.
Login to EasyView EasyView administrators must log in to EasyView Administrator
accounts
Login to System Owner defined log in rules can be in effect at system (OS) level.
Migrate record Move EasyView database to other system.
Operator A person having access rights as a “Windows user” but not as an
EasyView administrator..
Owner A person having all necessary rights to install and maintain a
system.
Project An EasyView term for a pre-programmed set of commands and
parameters that govern the acquisition of data and how it is
displayed.
System policy Settings in the OS, governing user rights and login procedures.
User A Windows term for a person having the right to use a windows
account. Access rights are defined by the owner.
Windows Administrator A person with unrestricted rights in Windows.
2 Software categorisation
EasyView QA is a COTS (IEEE term: Commercial Off-the-shelf Software) and GAMP
Category 3 software.
I.e. “it is defined by a market-driven need, its fitness for use has been demonstrated by a
broad spectrum of commercial user”.
3 Computer requirements
EasyView QA requires a reasonably fast computer. The following is a list of recommended
hardware and properties.
1. Processor speed > 200MHz
2. RAM > 64MB
3. Available disk space > 32MB
4. Graphics accelerator 4MB
5. Monitor resolution > 1280 x 1024
4 OS requirements
To take full advantage of the safety and data integrity safeguard features incorporated in
EasyView QA, it is mandatory that it be installed in a MS Windows XP, Windows 2000 or
NT4 environment.
5 Installation
1. Installation must be performed in Windows by a person who has Windows
administrator privileges and “owner rights”.
The system should normally be run with reduced access rights e.g. user accounts.
2. It is the responsibility of the “owner” to ensure that correct and sufficient safety
barriers are in force. I.e. operators should use operator accounts with no more rights
than absolutely necessary.
N.B. that none of these operations affect data numerically. No data can be added, deleted or
altered.
7 Passwords
1. All operations, beyond those allowed the operator, are password protected.
2. Passwords must have a length not shorter than 7 characters.
3. Passwords must be assigned to all EasyView administrators.
4. All passwords can be altered by administrators with the right to create administrator
accounts.
5. Passwords can be programmed to expire after a set period.
6. An EasyView QA administrator can at all times change his own password.
Policy Action/Argument
1 Approve documents using electronic signature Select either.
Approve documents using handwritten signature.
2 Automatic Inactivity log out Activate/no of
minutes
3 Automatic disabling of account after failed login attempts Activate/no of
attempts
4 Notify e-mail recipient of disabled account Activate/E-mail
address
5 All logins are recorded in the Audit trail Activate
6 All failed login attempts are recorded in the Audit trail Activate
Password expiry is set in each account: never or after a programmed number of days
10 Electronic records
EasyView QA creates, manages and “handles” measurement and other data as recordings or
documents. These documents meet the specifications of electronic records in FDA “21 CFR
part 11”.
The EasyView documents contain measurement data, meta data, audit trail, name of the
responsible administrator and, where applicable, electronic signatures.
The Administrator responsible for recorder settings will be noted as responsible for the
document regardless of who actually created it. See Projects 16
All EasyView QA documents are kept in a secure, password protected, database.
1. Retention period The “owner” must ascertain that the storage medium meets his
demands.
2. Record migration Record migration in the form of databases must be handled at the
“owner” level where it is the responsibility of the owner to ensure
system integrity and security.
3. Recording import Only copies of EasyView secure documents can be “imported”.
They must be “exported” by an EasyView QA and have intact
encryption and contain the audit trail and signatures where
applicable.
Conventional import to EasyView QA is not possible
4. Recording export Documents can be copied to other EasyView QA systems. These
copies are encrypted using the ECB and contain the Audit trail
and applicable signatures.
It is normally not possible to export data to non-EasyView QA
systems. Customised secure non-EasyView export plug-ins can
be made available.
5. Secure records EasyView QA will only accept data via its built in logger drivers.
Recordings from a non-QA EasyView will not open in EasyView
QA.
When a document is opened it is “checked out”. Other users will
only be able to open a write-protected copy of the document. If,
for some reason, this document is not subsequently closed
correctly, it will remain “checked out” until the same user opens
it and closes it again.
11 Document integrity
Each EasyView QA document is identified by OS account; Responsible Administrator; Title;
Date and Time. It is also assigned a 38 character long globally unique identifier.
All EasyView QA documents are kept in a secure, password protected, ODBC compliant
database. They are stored in the form of tamper-proof data packages. These data packages can
only be accessed by EasyView QA.
12 Electronic signatures
1. Electronic signatures in EasyView QA are of two kinds: Reviewed and Approved.
2. Only Reviewed documents can be Approved.
3. Any modification forces a new Review signature before Approval is made possible.
4. An Approved document is locked and cannot be modified.
5. Electronic signatures can only be entered by administrators with signature (Approve or
Review) privileges.
6. Administrators with signature privileges must supply their user name and password
when signing.
7. Electronic signatures are shown in the record, in electronic or printed form, with
meaning (Approve or Review), full name and date, time and time zone of the signing.
8. When signing, EasyView QA informs the signer that the signature is legally binding.
9. All signings are noted in the Record Audit Trail.
10. Unsuccessful attempts at signing are, if so ordered , noted in the Audit trail.
11. A programmable number of unsuccessful attempts may disable the account. A
notifying e-mail can be sent as a result of such a disabling.
13 Handwritten signatures
The EasyView QA policy settings allow “hybrid” configuration. I. e. the documents are
electronic but approval signatures will be handwritten.
Approval will in these cases result in the issue of a printed form.
1. The “Electronic Record Approval” form can only be issued from “Reviewed”
documents.
2. Any modification forces a new Review entry before Approval is made possible.
3. Issuing the Approval form locks the document.
4. Only Administrators with approval rights can issue the Approval forms.
5. Administrators with signature privileges must supply their user name and password
when approving.
6. The Approval form shows full name and date, time and time zone of the signer.
7. Issue of the Approval form is noted in the Record Audit Trail.
8. Unsuccessful attempts at issuing the approval form are noted in the Audit trail.
9. A programmable number of unsuccessful attempts may disable the account. A notifying
e-mail can be sent as a result of such a disabling.
14 Audit Trails
EasyView QA maintains two types of Audit trails: the (global) System Audit Trail and the
Record Audit trail.
The System Audit Trail records all EasyView QA account transactions:
Account creation, modification of Account rights, disabling of accounts.
The Record Audit Trail records all the actions that affect the recording. These are:
Change channel label, Change transform, Change transform unit, Change channel time skew,
Change start time, Edit formulas, Approve document, Review document.
1. All entries in the audit trails include the, at the time logged in, user/administrator.
2. All entries are time stamped.
3. Time and date formats follow those specified in the Windows control panel.
4. Audit trails are not editable by the user.
5. Audit trails are at all times viewable and printable
6. All entries are sequentially added.
15 Record printouts
The electronic record, in the form of an EasyView QA document, can in full or selected parts
be printed on paper. All pages of a printout contain the 38 character Document identifier;
Responsible Administrator; Signatures with reason, date, time and time zone.
Selectable parts are:
1. Active diagram tab
2. Active info-table tab
3. Selected plug-ins. (See main manual.)
4. Record Audit Trail
16 Projects
EasyView QA projects (see main manual) can only be created by EasyView QA
Administrators. It is possible for an EasyView QA administrator to allow operators to Start
and/or Stop and/or Offload a recording. It is up to the project responsible to judge whether
projects should thus be made available to Operators.
Within a dedicated single role system, it is only possible to fetch off-line data using the same
project that initiated the collection of data.
The Administrator should be aware of the possibility to inadvertently use the wrong project in
multi-role, multi-logger, multi-project environments and thus have as few projects open to the
operator as possible.