6 Module 7 Quality Assurance Asessment
6 Module 7 Quality Assurance Asessment
Clinical Chemistry
1:00-3:00PM MWF | 9:00:10:00AM TTHS | NENITA EUSTAQUIO
MODULE 7 → Material is usually run at the beginning of each shift;
after an instrument is serviced; when reagent lots are
TOPIC OUTLINE changed; after calibration, and when patient results
1 Quality Assurance/Assessment seem inappropriate
a. Quality Control → Quantitative tests should include a minimum of one
b. Laboratory Quality Control control with a target value in the healthy person
c. Types of Quality Control reference interval and a second control with a target
d. Pre-Analytical and Analytical Causes of Error value that would be seen in a sick patient
e. Post-Analytical Causes of Error - Examples include sodium controls of 140 and
2 Statistics 115 mEq/L and glucose controls of 75 and 225
a. Statistical Analysis of Lab Test Data mg/dL.
b. Measures of Central Tendency - If three control levels are run, an abnormally
i. Mean low patient range should be included
ii. Median → Must be integrated within routine workload and
iii. Mode analyzed by personnel who are running the tests
c. Precision and Accuracy → Ongoing evaluation of results to correct for
d. Systematic Error unacceptable results
i. Causes of Systematic Error → Used to access employee competency
e. Random Error
i. Causes of Random Error TYPES OF QUALITY CONTROL
f. Bias INTERNAL
g. Analytical Measurement Range → Daily establishment of reference ranges
h. Clinical Reportable Range → Validation of a new reagent lot and/or shipment
i. Statistical Formula → Following instrument repair
j. Standard Deviation and Probability EXTERNAL
k. Statistical Formulas → Determination of laboratory testing performance by
l. Establishment of a QC System means of intralaboratory comparisons
m. Collecting Data for Quality Control → CAP, CLIA, The Joint Commission requirement
n. Shift, Trend, and Out of Control
o. System Flags PRE-ANALYTICAL AND ANALYTICAL CAUSES OF ERROR
i. Delta Check
ii. Critical Values and Read Back of Results
QUALITY ASSURANCE/ASSESSMENT
→ An all-inclusive / comprehensive system monitoring the
accuracy of test results where all steps before, during and
after the testing process are considered
→ Includes:
✓ Pre-analytic
✓ Analytic
✓ Post analytic factors
→ Essentials include commitment to quality, facilities, POST-ANALYTICAL CAUSES OF ERROR
resources, competent staff, and reliable procedures, → Incorrect reference values
methods and instrumentation → Physician not notified of a panic or critical value
→ Provides a structure for achieving lab and hospital quality → Incorrect interpretation of lab results by physician
goals → Incorrect data entry of lab results
MFJR 1
Clinical Chemistry
1:00-3:00PM MWF | 9:00:10:00AM TTHS | NENITA EUSTAQUIO
1 Reference ranges for patient’s (normal and → Systematic error of an analytic system is predictable
abnormal) and causes shifts or trends on control charts that are
2 Acceptable ranges for control specimens (“in” and consistently low or high
“out” of control) CAUSES OF SYSTEMATIC ERROR
1 → Change in reagent or calibrator lot numbers
MEASURES OF CENTRAL TENDENCY 2 → Wrong calibrator values
MEAN (x) 3 → Improperly prepared reagents
→ The mathematical average of a group of numbers, 4 → Deterioration of reagents or calibrators
determined by adding a group of numbers (events) and 5 → Inappropriate storage of reagents or calibrators
dividing the result by the number of events 6 → Variation in sample or reagent volumes due to
MEDIAN pipettor misalignments
→ Determined as the ‘middle’ of a group of numbers 7 → Variation in temperature or reaction chambers
that have been arranged in sequential order 8 → Deterioration of photometric light source
→ That is to say, there are an equal number of numbers 9 → Variation in procedure between technologists
on either side of the ‘middle’ number
→ In an odd # of observations, it is the middle RANDOM ERROR
observation → Imprecision of the test system causing a scatter or
→ In an even # of observations, average the two middle spread of control values around the mean
values CAUSES OF RANDOM ERROR
MODE 1 → Air bubbles in reagent
→ The number that appears most frequently in a group 2 → Improperly mixed reagents
of numbers 3 → Reagent lines, sampling, or reagent syringes
→ There can be more than mode, or none at all 4 → Improperly fitting pipette tips
5 → Clogged or imprecise pipetter
PRECISION 6 → Fluctuations in power supply
→ The degree of fluctuation in the measurements is
indicative of the precision of the assay
BIAS
→ Precision-refers to the ability to get the same (but not
→ The amount by which an analysis varies from the
necessarily ‘true’) result time after time
correct result
ACCURACY
→ Example:
→ The closeness of measurements to the true value is - If the Expected Value is 50 units, and the result
indicative of the accuracy of the assay of an analysis is 47, the bias is 3 units.
→ An accurate result is one that is the ‘true’ result
ANALYTICAL MEASUREMENT RANGE
→ Range of analyte values that a method can directly
measure on the specimen without any dilution,
concentration or other pretreatment
STATISTICAL FORMULA
SYSEMATIC ERROR
→ Systematic change in the test system resulting in a
displacement of the mean from the original value
MFJR 2
Clinical Chemistry
1:00-3:00PM MWF | 9:00:10:00AM TTHS | NENITA EUSTAQUIO
MFJR 3
Clinical Chemistry
1:00-3:00PM MWF | 9:00:10:00AM TTHS | NENITA EUSTAQUIO
SHIFT
→ QC data results are distributed on one side of the
mean for 6-7 consecutive days
TREND
→ Consistent increase or decrease of QC data points
over a period of 6-7 days
OUT OF CONTROL
→ Means that there is too much dispersion in your result
compared with the rest of the results
→ This suggests that something is wrong with the
process that generated that observation
→ Patient test results cannot be reported to physicians
when there is something wrong with the testing process
that is generating inaccurate reports
→ Remember … No information is better than wrong
information
WHAT IF YOUR CONTROL SPECIMEN IS OUT OF
CONTROL?
Things that can go wrong Corrective action
Instrument malfunction Identify malfunction and fix
Reagents: preparation, New reagents
contamination, volume
Tech error Identify error and repeat
test
Control specimen is old Use new control
or prepared improperly
SYSTEM FLAGS
DELTA CHECK
→ Comparison of individual patient results throughout
the day or week with computer detection of changes
from earlier individual patient results
→ Helpful to identify pre-analytical errors
CRITICAL VALUES AND READ BACK OF RESULTS
→ Values that indicate a life-threatening situation for
the patient
→ Require notification of the value to nurse or physician
→ Nurse or physician must “read back” the results to the
technician
MFJR 4