621 Chromatography
621 Chromatography
〈 621 〉 CHROMATOGRAPHY
To view the Notice from the Expert Committee that posted in conjunction with this accelerated revision, please click
https://www.uspnf.com/rb-gc- 621 -20230929.
Change to read:
(The sections System Sensitivity and Peak Symmetry will become official on ▲May 1, 2025▲ (RB 1-Oct-2023) as indicated.)
INTRODUCTION
Chromatographic separation techniques are multistage separation procedures in which the components of a sample are distributed between
two phases, one of which is stationary while the other is mobile. The stationary phase may be a solid or a liquid supported on a solid or a gel.
The stationary phase may be packed in a column, spread as a layer, or distributed as a film, etc. The mobile phase may be gaseous or liquid or
supercritical fluid. The separation may be based on adsorption, mass distribution (partition), ion exchange, etc., or may be based on differences
in the physicochemical properties of the molecules such as size, mass, volume, etc.
Portions of the present general chapter text that are national USP–NF text, and therefore not part of the harmonized text, are marked with
symbols ( ♦♦ ) to specify this fact.
♦ This chapter describes general procedures, definitions, and calculations of common parameters and generally applicable requirements for
system suitability. AL
The types of chromatography useful in qualitative and quantitative analysis employed in USP procedures are column, gas, paper, thin-layer
(including high-performance thin-layer chromatography), and pressurized liquid chromatography (commonly called high-pressure or high-
performance liquid chromatography).
GENERAL PROCEDURES
CI
This section describes the basic procedures used when a chromatographic method is described in a monograph. The following procedures
are followed unless otherwise indicated in the individual monograph.
Paper Chromatography
FI
stationary phase
The stationary phase is a sheet of paper of suitable texture and thickness. Development may be ascending, in which the solvent is carried up
OF
the paper by capillary forces, or descending, in which the solvent flow is also assisted by gravitational force. The orientation of paper grain with
respect to solvent flow is to be kept constant in a series of chromatograms. (The machine direction is usually designated by the
manufacturer.)
apparatus
The essential equipment for paper chromatography consists of a vapor-tight chamber with inlets for addition of solvent and a rack of
corrosion-resistant material about 5 cm shorter than the inside height of the chamber. The rack serves as a support for solvent troughs and for
antisiphon rods that, in turn, hold up the chromatographic sheets. The bottom of the chamber is covered with the prescribed solvent system or
mobile phase. Saturation of the chamber with solvent vapor is facilitated by lining the inside walls with paper wetted with the prescribed
solvent system.
spotting
The substance or substances analyzed are dissolved in a suitable solvent. Convenient volumes delivered from suitable micropipets of the
resulting solution, normally containing 1–20 µg of the compound, are placed in 6- to 10-mm spots not less than 3 cm apart.
Thin-Layer Chromatography
stationary phase
The stationary phase is a relatively thin, uniform layer of dry, finely powdered material applied to a glass, plastic, or metal sheet or plate
(typically called the plate). The stationary phase of thin-layer chromatography (TLC) plates has an average particle size of 10–15 µm, and that
of high-performance TLC (HPTLC) plates has an average particle size of 5 µm. Commercial plates with a preadsorbent zone can be used if
they are specified in a monograph. Sample applied to the preadsorbent region develops into sharp, narrow bands at the preadsorbent–sorbent
interface. The separations achieved may be based on adsorption, partition, or a combination of both effects, depending on the particular type
of stationary phase.
apparatus
A chromatographic chamber made of inert, transparent material and having the following specifications is used: a flat-bottom or twin trough,
AL
a tightly fitted lid, and a size suitable for the plates. The chamber is lined on at least one wall with filter paper. Sufficient mobile phase or
developing solvent is added to the chamber so that, after impregnation of the filter paper, a depth appropriate to the dimensions of the plate
used is available. The chromatographic chamber is closed and allowed to equilibrate. [Note—Unless otherwise indicated, the chromatographic
separations are performed in a saturated chamber.]
CI
detection/visualization
An ultraviolet (UV) light source suitable for observations under short- (254 nm) and long- (365 nm) wavelength UV light and a variety of spray
reagents to make spots visible are often used.
FI
spotting
Solutions are spotted on the surface of the stationary phase (plate) at the prescribed volume in sufficiently small portions to obtain circular
spots of 2–5 mm in diameter (1–2 mm on HPTLC plates) or bands of 10–20 mm × 1–2 mm (5–10 mm × 0.5–1 mm on HPTLC plates) at an
OF
appropriate distance from the lower edge and sides of the plate. [Note—During development, the application position must be at least 5 mm
(TLC) or 3 mm (HPTLC) above the level of the mobile phase.] The solutions are applied on a line parallel to the lower edge of the plate with an
interval of at least 10 mm (5 mm on HPTLC plates) between the centers of spots, or 4 mm (2 mm on HPTLC plates) between the edges of
bands, then allowed to dry.
procedure
1. Place the plate in the chamber, ensuring that the spots or bands are above the surface of the mobile phase.
2. Close the chamber.
3. Allow the mobile phase to ascend the plate until the solvent front has traveled three-quarters of the length of the plate, or the distance
prescribed in the monograph.
4. Remove the plate, mark the solvent front with a pencil, and allow to dry.
5. Visualize the chromatograms as prescribed.
6. Determine the chromatographic retardation factor (RF) values for the principal spots or zones.
7. Presumptive identification can be made by observation of spots or zones of identical RF value and about equal magnitude obtained,
respectively, with an unknown and a standard chromatographed on the same plate. A visual comparison of the size or intensity of the
spots or zones may serve for semiquantitative estimation. Quantitative measurements are possible by means of densitometry
(absorbance or fluorescence measurements).
Column Chromatography
solid support
Purified siliceous earth is used for normal-phase separation. Silanized chromatographic siliceous earth is used for reverse-phase partition
chromatography.
stationary phase
The solid support is modified by the addition of a stationary phase specified in the individual monograph. If a mixture of liquids is used as
the stationary phase, mix the liquids before the introduction of the solid support.
mobile phase
The mobile phase is specified in the individual monograph. If the stationary phase is an aqueous solution, equilibrate with water. If the
stationary phase is a polar organic fluid, equilibrate with that fluid.
apparatus
Unless otherwise specified in the individual monograph, the chromatographic tube is about 22 mm in inside diameter and 200–300 mm in
length. Attached to it is a delivery tube, without stopcock, about 4 mm in inside diameter and about 50 mm in length.
Apparatus preparation: Pack a pledget of fine glass wool in the base of the tube. Combine the specified volume of stationary phase and the
specified amount of solid support to produce a homogeneous, fluffy mixture. Transfer this mixture to the chromatographic tube, and tamp
using gentle pressure to obtain a uniform mass. If the specified amount of solid support is more than 3 g, transfer the mixture to the column in
portions of approximately 2 g, and tamp each portion. If the assay or test requires a multisegment column with a different stationary phase
specified for each segment, tamp after the addition of each segment, and add each succeeding segment directly to the previous one. Pack a
pledget of fine glass wool above the completed column packing. [Note—The mobile phase should flow through a properly packed column as a
moderate stream or, if reverse-phase chromatography is applied, as a slow trickle.]
If a solution of the analyte is incorporated into the stationary phase, complete the quantitative transfer to the chromatographic tube by
scrubbing the beaker used for the preparation of the test mixture with a mixture of about 1 g of solid support and several drops of the solvent
used to prepare the sample solution before adding the final portion of glass wool.
procedure
1. Transfer the mobile phase to the column space above the column packing, and allow it to flow through the column under the influence
of gravity.
2. Rinse the tip of the chromatographic column with about 1 mL of mobile phase before each change in composition of mobile phase and
after completion of the elution.
AL
3. If the analyte is introduced into the column as a solution in the mobile phase, allow it to pass completely into the column packing, then
add mobile phase in several small portions, allowing each to drain completely, before adding the bulk of the mobile phase.
4. Where the procedure indicates the use of multiple chromatographic columns mounted in series and the addition of mobile phase in
divided portions is specified, allow each portion to drain completely through each column, and rinse the tip of each with mobile phase
CI
before the addition of each succeeding portion.
packed column GC
OF
The liquid stationary phase is deposited on a finely divided, inert solid support, such as diatomaceous earth, porous polymer, or graphitized
carbon, which is packed into a column that is typically 2–4 mm in internal diameter and 1–3 m in length.
capillary column GC
In capillary columns, which contain no packed solid support, the liquid stationary phase is deposited on the inner surface of the column and
may be chemically bonded to it.
apparatus
A gas chromatograph consists of a carrier gas source, injection port, column, detector, and recording device. The injection port, column, and
detector are temperature controlled and may be varied as part of the analysis. The typical carrier gas is helium, nitrogen, or hydrogen,
depending on the column and detector in use. The type of detector used depends on the nature of the compounds analyzed and is specified in
the individual monograph. Detector output is recorded as a function of time, and the instrument response, measured as peak area or peak
height, is a function of the amount present.
temperature program
The length and quality of a GC separation can be controlled by altering the temperature of the chromatographic column. When a
temperature program is necessary, the individual monograph indicates the conditions in table format. The table indicates the initial
temperature, rate of temperature change (ramp), final temperature, and hold time at the final temperature.
procedure
1. Equilibrate the column, injector, and detector with flowing carrier gas until a constant signal is received.
2. Inject a sample through the injector septum, or use an autosampler.
3. Begin the temperature program.
4. Record the chromatogram.
5. Analyze as indicated in the monograph.
Liquid Chromatography
The term "liquid chromatography" (LC), as used in the compendia, is synonymous with high-pressure liquid chromatography and high-
performance liquid chromatography. LC is a separation technique based on a solid stationary phase and a liquid mobile phase.
stationary phase
Separations are achieved by partition, adsorption, or ion-exchange processes, depending on the type of stationary phase used. The most
commonly used stationary phases are modified silica or polymeric beads. The beads are modified by the addition of long-chain hydrocarbons.
The specific type of packing needed to complete an analysis is indicated by the “L” designation in the individual monograph (see also the
section Chromatographic Columns). The size of the beads is often described in the monograph as well. Changes in the packing type and size
are covered in the System Suitability section of this chapter.
chromatographic column
The term "column" includes stainless steel, lined stainless steel, and polymeric columns, packed with a stationary phase. The length and
inner diameter of the column affects the separation, and therefore typical column dimensions are included in the individual monograph.
Changes to column dimensions are discussed in the System Suitability section of this chapter. Compendial monographs do not include the
name of appropriate columns; this omission avoids the appearance of endorsement of a vendor’s product and natural changes in the
marketplace. See the section Chromatographic Columns for more information.
In LC procedures, a guard column may be used with the following requirements, unless otherwise is indicated in the individual monograph:
(a) the length of the guard column must be NMT 15% of the length of the analytical column, (b) the inner diameter must be the same or
AL
smaller than that of the analytical column, and (c) the packing material should be the same as the analytical column (e.g., silica) and contain
the same bonded phase (e.g., C18). In any case, all system suitability requirements specified in the official procedure must be met with the
guard column installed.
mobile phase
CI
The mobile phase is a solvent or a mixture of solvents, as defined in the individual monograph.
apparatus
A liquid chromatograph consists of a reservoir containing the mobile phase, a pump to force the mobile phase through the system at high
FI
pressure, an injector to introduce the sample into the mobile phase, a chromatographic column, a detector, and a data collection device.
gradient elution
OF
The technique of continuously changing the solvent composition during the chromatographic run is called gradient elution or solvent
programming. The gradient elution profile is presented in the individual monograph as a gradient table, which lists the time and proportional
composition of the mobile phase at the stated time.
procedure
1. Equilibrate the column and detector with mobile phase at the specified flow rate until a constant signal is received.
2. Inject a sample through the injector, or use an autosampler.
3. Begin the gradient program.
4. Record the chromatogram.
5. Analyze as directed in the monograph.
CHROMATOGRAPHIC COLUMNS
A complete list of packings (L), phases (G), and supports (S) used in USP–NF tests and assays is located in USP–NF, Reagents, Indicators,
and Solutions—Chromatographic Columns. This list is intended to be a convenient reference for the chromatographer in identifying the pertinent
chromatographic column specified in the individual monograph. ♦
DEFINITIONS
The system suitability and acceptance criteria in monographs have been set using parameters as defined below. With some equipment,
certain parameters, such as the signal-to-noise ratio and resolution, can be calculated using software provided by the manufacturer. It is the
responsibility of the user to ensure that the calculation methods used in the software are equivalent to the requirements of the US
Pharmacopeia and to make any necessary corrections if this is not the case.
Chromatogram: A graphical or other representation of detector response, effluent concentration, or other quantity used as a measure of effluent
concentration versus time or volume. Idealized chromatograms are represented as a sequence of Gaussian peaks on a baseline (Figure 1).
Figure 1.
VM = hold-up volume
tM = hold-up time
Distribution constant (K0): In size-exclusion chromatography, the elution characteristics of a component in a particular column may be given
by the distribution constant (also referred to as distribution coefficient), which is calculated using the following equation:
FI
tR− t0
K0 =
tt −t0
tR = retention time
OF
Dwell volume (D) (also referred to as VD): The dwell volume (also known as gradient delay volume) is the volume between the point at which
the eluents meet and the inlet of the column. It can be determined using the following procedure.
column: Replace the chromatographic column by an appropriate capillary tubing (e.g., 1 m × 0.12 mm).
Table 1
20–30 0 100
D = tD × F
Figure 2.
[Note—where applicable, this measurement is performed with the autosampler in the "inject" position so as to include the injection loop
AL
volume in the dwell volume.]
Hold-up time (tM): Time required for elution of an unretained component (see Figure 1, baseline scale being in minutes or seconds).
In size-exclusion chromatography, the term retention time of an unretained compound (t0) is used.
CI
Hold-up volume (VM): Volume of the mobile phase required for elution of an unretained component. It may be calculated from the hold-up time
and the flow rate (F) in milliliters per minute using the following equation:
VM = t M × F
FI
Peak: Portion of a chromatogram recording the detector response when a single component (or two or more unresolved components) is eluted
OF
Peak-to-Valley Ratio (p/v): The peak-to-valley ratio may be employed as a system suitability criterion when baseline separation between two
peaks is not achieved (see Figure 3).
Figure 3.
Hp
p/v =
Hv
Hp
Hv
= height above the extrapolated baseline of the minor peak
AL
= height above the extrapolated baseline at the lowest point of the curve separating the minor and major peaks
Plate height (H) (height equivalent to one theoretical plate [HETP]): Ratio of the column length (L), in micrometers, to the plate number
CI
(N):
L
H =
N
Plate number (N) (number of theoretical plates): A number indicative of column performance (column efficiency) can only be calculated
FI
from data obtained under either isothermal, isocratic, or isodense conditions, depending on the technique, as the plate number, using the
following equation, the values of tR and Wh being expressed in the same units:
OF
2
tR
N = 5.54 ( )
Wh
The plate number varies with the component as well as with the column, the column temperature, the mobile phase, and the retention time.
Reduced plate height (h): Ratio of the plate height (H), in micrometers, to the particle diameter (dp) in micrometers:
H
h =
dp
Relative retardation (Rrel): The relative retardation, used in thin-layer chromatography, is calculated as the ratio of the distances travelled by the
spot of the compound of interest and a reference compound (Figure 4).
Figure 4.
Rr el = b/c
r =
AL
Relative retention (r): Relative retention is calculated as an estimate using the following equation:
tRi −tM
tRst −tM
tM = hold-up time
♦or (RRT)♦
FI
tRi
rG =
tRst
Unless otherwise indicated, values for relative retention stated in monographs correspond to unadjusted relative retention.
1.18(tR2 −tR1 )
RS =
Wh1 +Wh2
In quantitative thin-layer chromatography, using densitometry, the migration distances are used instead of retention times and the resolution
between peaks of two components may be calculated using the following equation:
1.18a(RF 2 −RF 1 )
RS =
Wh1 +Wh2
RF 2 > RF 1
Retardation factor (RF): The retardation factor, used in thin-layer chromatography, is the ratio of the distance from the point of application to
the center of the spot and the distance simultaneously travelled by the solvent front from the point of application (Figure 4).
b
RF =
a
Retention factor (k): The retention factor (also known as mass distribution ratio (Dm) or capacity factor (k′)) is defined as:
The retention factor of a component may be determined from the chromatogram using the following equation:
tR −tM
k =
tM
tR = retention time
tM = hold-up time
Retention time (tR): Time elapsed between the injection of the sample and the appearance of the maximum peak response of the eluted
AL
sample zone (Figure 1, baseline scale being in minutes or seconds).
Volume of the mobile phase required for elution of a component. It may be calculated from the retention time and the flow rate (F), in
CI
milliliters per minute, using the following equation:
VR = tR × F
Retention time of an unretained compound (t0): In size-exclusion chromatography, retention time of a component whose molecules are
FI
Figure 5.
Retention volume of an unretained compound (V0): In size-exclusion chromatography, retention volume of a component whose molecules
are larger than the largest gel pores. It may be calculated from the retention time of an unretained compound and the flow rate (F), in milliliters per
minute, using the following equation:
V0 = t0 × F
Separation factor (α): Relative retention calculated for two adjacent peaks (by convention, the value of the separation factor is always >1):
α = k2 /k1
Signal-to-noise ratio (S/N): The short-term noise influences the precision and accuracy of quantitation. The signal-to-noise ratio is calculated
using the following equation:
2H
S /N =
h
H = height of the peak (Figure 6) corresponding to the component concerned, in the chromatogram obtained with the prescribed reference so‐
lution, measured from the maximum of the peak to the extrapolated baseline of the signal observed over a distance of at least 5 times
the width at half-height
h = range of the noise in a chromatogram obtained after injection of a blank (Figure 7), observed over a distance of at least 5 times the width
at half-height of the peak in the chromatogram obtained with the prescribed reference solution and, if possible, situated equally around
the place where this peak would be found
AL
CI
FI
OF
Figure 7.
Symmetry factor (AS): The symmetry factor of a peak (also known as the asymmetry factor or tailing factor) (Figure 8) is calculated using the
following equation:
W0.05
As =
2d
d = distance between the perpendicular dropped from the peak maximum and the leading edge of the peak at one-twentieth of the peak
height
An As value of 1.0 signifies symmetry. When As > 1.0, the peak is tailing. When As < 1.0, the peak is fronting.
Figure 8.
System repeatability: The repeatability of response is expressed as an estimated percentage relative standard deviation (%RSD) of a
consecutive series of measurements for not fewer than three injections or applications of a reference solution, and is calculated using the
following equation.
−−−−−−
2
100 Σ(yi −ȳ )
%R S D = √
ȳ n−1
yi
ȳ
AL
= individual values expressed as peak area, peak height, or ratio of areas by the internal standardization method
Total mobile phase volume (Vt): In size-exclusion chromatography, retention volume of a component whose molecules are smaller than the
FI
smallest gel pores. It may be calculated from the total mobile phase time and the flow rate (F), in milliliters/minute, using the following equation:
Vt = tt × F
OF
Change to read:
SYSTEM SUITABILITY
[Note—This section only covers liquid chromatography and gas chromatography]
The various components of the equipment employed must be qualified and be capable of achieving the performance required to conduct the
test or assay. The system suitability tests represent an integral part of the analytical procedure and are used to ensure adequate performance
of the chromatographic system. Column plate number, retention factor (mass distribution ratio), system repeatability, signal-to-noise, symmetry
factor, and resolution/peak-to-valley ratio are the parameters that may be employed in assessing the performance of the chromatographic
system. When it is indicated in the individual monograph, in case of complex chromatographic profiles (e.g., for biotechnological and biological
products) visual comparison of the profiles can be used as a system suitability test. Factors that may affect the chromatographic behavior
include:
Composition and temperature of the mobile phase
Ionic strength and pH of the aqueous component, of the mobile phase
Flow rate, column dimensions, column temperature, and pressure
Stationary phase characteristics including, type of chromatographic support (particle-based or monolithic), particle or pore size, and
porosity, and specific surface area
Reverse phase and other surface modification of the stationary phases, the extent of chemical modification (as expressed by end-
capping, carbon loading, etc.)
Retention times and relative retentions may be provided in monographs for information purposes only, unless otherwise stated in the
monograph. There are no acceptance criteria applied to relative retentions.
Compliance with the system suitability criteria is required throughout the chromatographic procedure. No sample analysis is acceptable
unless the suitability of the system has been demonstrated.
The following requirements are to be fulfilled, in addition to any other system suitability criteria stated in the monograph. When specific
requirements are stated in the monograph, they supersede the requirements mentioned in this chapter.
System Repeatability
♦ When a relative standard deviation requirement is specified in an individual monograph, if the requirement is 2.0 or less the calculation is
based on data from five replicate injections of the analyte, if the requirement is more than 2.0% data from six replicate injections are used. ♦
In an assay of an active substance or an excipient, where the target value is 100% for a pure substance, and a system repeatability
requirement is not specified, the maximum permitted relative standard deviation (%RSDmax) for the defined limits is calculated for a series (n = 3
to 6) of injections of the reference solution. The maximum permitted relative standard deviation of the peak response does not exceed the
appropriate value given in Table 2.
K B √n
%R S Dmax =
t90%,n−1
B = upper limit given in the definition of the individual monograph minus 100%
t90%,n = Student’s t at the 90% probability level (double sided) with n−1 degrees of freedom
–1
B (%)
3
AL
4
▲
System Sensitivity
The signal-to-noise ratio is used to define the system sensitivity. The limit of quantitation (corresponding to a signal-to-noise ratio of 10) is
equal to or less than the reporting threshold.
Peak Symmetry
Unless otherwise stated, in a test or assay, the symmetry factor (tailing factor) of the peak used for quantification is 0.8–1.8.▲ (Official 1-May-
2025)
Thin-Layer Chromatography
Composition of the mobile phase: the amount of the minor solvent components may be adjusted by ±30% relative or ±2% absolute, whichever
is the larger; no other component is altered by more than 10% absolute. A minor component comprises less than or equal to (100/n)%, n being
the total number of components of the mobile phase. For a minor component at 10% of the mobile phase, a 30% relative adjustment allows a
range of 7%–13% whereas a 2% absolute adjustment allows a range of 8%–12%, the relative value therefore being the larger; for a minor
component at 5% of the mobile phase, a 30% relative adjustment allows a range of 3.5%–6.5% whereas a 2% absolute adjustment allows a
range of 3%–7%, the absolute value being the larger in this case.
pH of the aqueous component of the mobile phase: ±0.2 pH units, unless otherwise prescribed
Concentration of salts in the buffer component of a mobile phase: ± 10%
Application volume: 10%–20% of the prescribed volume if using fine particle size plates (2–10 µm)
2 2
F2 = F1 × [(d c × d p1 )/ (d c × d p2 )]
2 1
When a change is made from ≥3-µm to <3-µm particles in isocratic separations, an additional increase in linear velocity (by adjusting the
flow rate) may be justified, provided that the column performance does not drop by more than 20%. Similarly, when a change is made from <3-
µm to ≥3-µm particles, an additional reduction of linear velocity (flow rate) may be justified to avoid reduction in column performance by more
than 20%.
After an adjustment due to a change in column dimensions, an additional change in flow rate of ±50% is permitted.
Column temperature: ± 10°C, where the operating temperature is specified, unless otherwise prescribed
Further adjustments in analytical procedure conditions (mobile phase, temperature, pH, etc.) may be required, within the permitted ranges
described under System Suitability and Adjustment of Chromatographic Conditions in this chapter.
mobile phase
Composition: The amount of the minor components of the mobile phase can be adjusted by ±30% relative. However, the change in any
component cannot exceed ±10% absolute. A minor component comprises less or equal than (100/n) %, n being the total number of
components of the mobile phase:
♦ Examples of adjustments for binary and ternary mixtures are given below.
binary mixtures
Specified ratio of 50:50: 30% of 50 is 15% absolute, but this exceeds the maximum permitted change of ±10% absolute in either component.
Therefore, the mobile phase ratio may be adjusted only within the range of 40:60–60:40.
Specified ratio of 2:98: 30% of 2 is 0.6% absolute. Therefore, the maximum allowed adjustment is within the range of 1.4: 98.6–2.6: 97.4.
ternary mixtures
Specified ratio of 70:25:5: For the second component, 30% of 25 is 7.5% absolute. Therefore, the second component may be adjusted within
the range of 32.5%–17.5% absolute. For the third component, 30% of 5 is 1.5% absolute. In all cases, a sufficient quantity of the first
component is used to give a total of 100%. Therefore, mixture ranges of 62.5: 32.5: 5 to 77.5: 17.5: 5 or 68.5: 25: 6.5 to 71.5: 25: 3.5 would meet
the requirement. ♦
pH of the aqueous component of the mobile phase: ±0.2 pH units, unless otherwise prescribed
Concentration of salts in the buffer component of a mobile phase: ±10%
Flow rate: In absence of a change in column dimensions, an adjustment of the flow rate by ±50% is permitted
detector wavelength
No adjustment permitted.
injection volume
When the column dimensions are changed, the following equation may be used for adjusting the injection volume:
2 2
Vi nj2 = Vi nj1 (L2 dc2 ) / (L1 dc1 )
L2
dc1
= new column length (mm)
of the peak response(s) to be determined. An increase is permitted provided, in particular, linearity and resolution of the peak(s) to be
determined remain satisfactory.
Adjustment of chromatographic conditions for gradient systems requires greater caution than for isocratic systems.
the prescribed column ♦ for all the peaks used to determine the system suitability parameters ♦ . These changes are acceptable provided
system suitability criteria are fulfilled, and selectivity and elution order of the specified impurities to be controlled are demonstrated to be
equivalent.
Internal diameter: In absence of a change in particle size and/or length, the internal diameter of the column may be adjusted.
Caution is necessary when the adjustment results in smaller peak volume resulting from smaller particle size or smaller internal column
diameter, a situation which may require adjustments to minimize extra-column band broadening by factors such as instrument connections,
detector cell volume and sampling rate, and injection volume.
When the particle size is changed, the flow rate requires adjustment, because smaller-particle columns will require higher linear velocities
for the same performance (as measured by reduced plate height). Flow rate is adjusted for both the change in column diameter and particle
size using the following equation:
2 2
F2 = F1 × [(d c2 × d p1 )/ (d c1 × d p2 )]
F1 = flow rate indicated in the monograph (mL/min)
A change in column dimensions, and thus in column volume, impacts the gradient volume which controls selectivity. Gradients are adjusted
to the column volume by changing the gradient volume in proportion to the column volume. This applies to every gradient segment volume.
Since the gradient volume is the gradient time, tG, multiplied by the flow rate, F, the gradient time for each gradient segment needs to be
adjusted to maintain a constant ratio of the gradient volume to the column volume (expressed as L × dc2). Thus, the new gradient time, tG2 can
be calculated from the original gradient time, tG1, the flow rate(s), and the column dimensions as follows:
2 2
tG2 = tG1 × (F1 /F2 ) [(L2 × d c2 )/ (L1 × d c1 )]
Thus, the change in conditions for gradient elution requires three steps:
1. Adjust the column length and particle size according to L/dp.
2. Adjust the flow rate for changes in particle size and column diameter.
3. Adjust the gradient time of each segment for changes in column length, diameter, and flow rate. The example below illustrates this process.
Table 3
Gradient conditions — — —
Time Time
B (%) (min) (min)
30 0 0 —
30 3 (3 × 0.4) = 1.2 —
Column temperature: ±5° C, where the operating temperature is specified, unless otherwise prescribed
Further adjustments in procedure conditions (mobile phase, temperature, pH, etc.) may be required, within the permitted ranges described
under System Suitability and Adjustment of Chromatographic Conditions in this chapter.
mobile phase
Composition/gradient: Adjustments of the composition of the mobile phase and the gradient are acceptable provided that:
The system suitability criteria are fulfilled.
The principal peak(s) elute(s) within ±15% of the retention time(s) obtained with the original conditions; this requirement does not apply
when the column dimensions are changed.
The composition of the mobile phase and the gradient are such that the first peaks are sufficiently retained and the last peaks are
eluted.
pH of the aqueous component of the mobile phase: ±0.2 pH units, unless otherwise prescribed
Concentration of salts in the buffer component of a mobile phase: ±10%
Where compliance with the system suitability criteria cannot be achieved, it is often preferable to consider the dwell volume or to change the
column.
dwell volume
The configuration of the equipment employed may significantly alter the resolution, retention time, and relative retentions described. Should
this occur, it may be due to a change in dwell volume. Monographs preferably include an isocratic step before the start of the gradient program
so that an adaptation can be made to the gradient time points to take account of differences in dwell volume between the system used for
analytical procedure development and that actually used. It is the user’s responsibility to adapt the length of the isocratic step to the analytical
equipment used. If the dwell volume used during the elaboration of the monograph is given in the monograph, the time points (t min) stated in
the gradient table may be replaced by adapted time points (tc min), calculated using the following equation:
(D−D0 )
tc = t −
F
The isocratic step introduced for this purpose may be omitted if validation data for application of the analytical procedure without this step
is available. AL
Detector wavelength: No adjustment permitted
Injection volume: When the column dimensions are changed, the following equation may be used for adjusting the injection volume:
2 2
Vi nj2 = Vi nj1 (L2 d c )/ (L1 d c )
2 1
CI
Vinj1 = injection volume indicated in the monograph (µL)
This equation may not be applicable to changes from TPP columns to SPP columns.
Even in the absence of any column dimension change, the injection volume may be varied provided system suitability criteria remain within
their established acceptability limits. When the injection volume is decreased, special attention is given to (limit of) detection and repeatability
of the peak response(s) to be determined. An increase is permitted provided, in particular, linearity and resolution of the peak(s) to be
determined remain satisfactory.
Gas Chromatography
column parameters
Stationary phase
Particle size: Maximum reduction of 50%; no increase permitted (packed columns)
Film thickness: −50% to +100% (capillary columns)
Column dimensions
Length: −70% to +100%
Internal diameter: ±50%
Column temperature: ±10%
Temperature program: Adjustment of temperatures is permitted as stated above; adjustment of ramp rates and hold times of up to ±20% is
permitted.
Flow rate: ±50%
The above changes are acceptable provided system suitability criteria are fulfilled and selectivity and elution order of the specified impurities
to be controlled are demonstrated to be equivalent.
Injection volume and split ratio: May be varied provided system suitability criteria remain within their established acceptability limits. When
the injection volume is decreased, or the split ratio is increased special attention is given to (limit of) detection and repeatability of the peak
response(s) to be determined. An increase in injection volume or decrease in split ratio is permitted provided, in particular, linearity and
resolution of the peak(s) to be determined remain satisfactory.
Injection port temperature and transfer-line temperature in static headspace conditions: ±10° C, provided no decomposition or condensation
occurs
QUANTITATION
The following quantitation approaches may be used in general texts or monographs.
Using a calibration function: Standard solutions with several graded amounts of a reference standard of the compound to be analyzed are
prepared in a range that has been demonstrated to give a linear response, and a fixed volume of these standard solutions is injected. With the
chromatograms obtained, a calibration function is prepared by plotting the peak areas or peak heights on the ordinate against the amount of
reference standard on the abscissa. The calibration function is generally obtained by linear regression. Then, a sample solution is prepared
according to the procedure specified in the individual monograph. The chromatography is performed under the same operating conditions as
for the preparation of the calibration function, the peak area or peak height of the compound to be analyzed is measured, and the amount of the
compound is read out or calculated from the calibration function.
Using one-point calibration: In an individual monograph, generally one of the standard solutions with a concentration within the linear range of
the calibration function and a sample solution with a concentration close to that of the standard solution are prepared, and the chromatography
is performed under fixed conditions to obtain the amount of the component by comparing the responses obtained. In this method, all
procedures, such as the injection procedure, must be carried out under constant conditions.
Using a calibration function: In the internal standard method, a stable compound is chosen as an internal standard which shows a retention
time close to that of the compound to be analyzed, and whose peak is well separated from all other peaks in the chromatogram. Several
AL
standard solutions containing a fixed amount of the internal standard and graded amounts of a reference standard of the compound to be
analyzed are prepared. Based on the chromatograms obtained by injection of a fixed volume of individual standard solutions, the ratio of peak
area or peak height of the reference standard to that of the internal standard is calculated. A calibration function by plotting these ratios on the
ordinate against the amount of the reference standard or the ratio of the amount of reference standard to that of the internal standard on the
abscissa is prepared. The calibration function is generally obtained by linear regression. Then, a sample solution containing the internal
CI
standard in the same amount as in the standard solutions used for the preparation of the calibration function is prepared according to the
procedure specified in the individual monograph. The chromatography is performed under the same operating conditions as for the preparation
of the calibration function. The ratio of the peak area or peak height of the compound to be analyzed to that of the internal standard is
FI
calculated, and the amount of the compound is read out or calculated from the calibration function.
Using one-point calibration: In an individual monograph, generally one of the standard solutions with a concentration within the linear range of
the calibration function and a sample solution with a concentration close to that of the standard solution, both containing a fixed amount of the
OF
internal standard, are prepared, and the chromatography is performed under fixed conditions to determine the amount of the compound to be
analyzed by comparing the ratios obtained.
Normalization procedure: Provided linearity of the peaks has been demonstrated, individual monographs may prescribe that the percentage
content of a component of the substance to be examined is calculated by determining the area of the corresponding peak as a percentage of
the total area of all the peaks, excluding those due to solvents or reagents or arising from the mobile phase or the sample matrix, and those at
or below the disregard limit or reporting threshold.
OTHER CONSIDERATIONS
Detector response: ♦ The detector sensitivity is the signal output per unit concentration or unit mass of a substance (also known as
response factor) in the mobile phase entering the detector. The relative detector response factor, commonly referred to as relative response
factor, expresses the sensitivity of a detector for a given substance relative to a standard substance.
In tests for related substances, any correction factors indicated in the monograph are applied. ♦
Interfering peaks: Peaks due to solvents and reagents or arising from the mobile phase or the sample matrix are disregarded.
Measurement of peaks: Integration of the peak area of any impurity that is not completely separated from the principal peak is preferably
performed by tangential skim (Figure 9).
Figure 9.
reporting threshold
When the related substances test prescribes a limit for the total of impurities or a quantitative determination of an impurity, it is important to
choose an appropriate reporting threshold and appropriate conditions for the integration of the peak areas.
In such tests the reporting threshold, i.e., the limit above which a peak is reported, is defined generally 0.05%.
AL
Auxiliary Information - Please check for your question in the FAQs before contacting USP.
Page Information:
USP43-NF38 - 6853
USP42-NF37 - 6781
USP41-NF36 - 6363