Fundamentals of Thermometry - Part 4
Fundamentals of Thermometry - Part 4
FUNJ3AMENTAL.S OF THERMOMETRY
PART IV
by Henry E. Sostmann
ABSTRACT
1: FIXED RESISTORS
(The use of the QHE ohm in nations which are members of the Treaty of
the Meter became effective on January 1, 1990, replacing at NIST the
NBS ohm of 1948. The relationship is:
1 Q(NBS-48) = 0.99999831(NIST-90)
62
of the deposited metal and the substrate. These resistors, however, have
proven to be very valuable as highly precise circuit elements.
Various alloys are available for the bulk wire resistance element. Per-
haps the most important of these is manganin, an alloy developed by
Edward Weston in 1889. The composition is 84% copper, 12% manganese
and 4X nickel. The resistivity at 20’C is about 49.3 #J/cm (290 Q per
circular-mil-foot). The temperature coefficient of resistance is about
KLOOOO15 Q/Q/-C over the normal laboratory range of 15’ to 35-C,
although it varies from lot to lot for reasons which are not well
understood or predictable. It can be adjusted somewhat by heat-
treatment in fabrication. For precise work, the temperature of the
resistor must be noted and a correction made. The calibration certificate
of the specific resistor should include the measured coefficients of the
correction equation, which assumes standardisation at 25-C.
fit&fJq
I
Ei;
. m
10 20 30 40 50 10 20 30 40 50
‘C ‘C
FIG. 1 FIG. 2
The ends of the winding are brazed to a pair of short heavy copper
leads, the free ends of which are connected to massive terminals.
Externally, four connections are proved, a potential connection and a
current connection to each end of the resistance winding, so that the
resistor may be connected four-terminal.
A four-terminal resistor
Pl P2 schematic is shown in Fig. 5.
The value of the four-terminal
resistor is the potential dif-
ference between the potential
terminals divided by the cur-
R3 R4 rent through the current ter-
minals, so that external cir-
cuitry is not a part of the
measured resistance. The po-
tential and current terminals
FIGURE 5 are interchangeable. For both
Thomas and NBS-type stan-
A FOUR-TERMINAL RESISTOR. RS is the dard resistors, the potential
resistance to be measured. Rl, R2, R3, terminals are usually binding
R4 are external resistances; e.g., lead posts and the current ter-
resistances. Cl, C2 are current
minals are connected into the
terminals, PI P2 potential terminals.
For an SPRT these are conventionally circuit using massive mer-
labeled c, C, t, T respectively. cury-wetted amalgam contacts.
Rosa resistors are filled with a bland neutral oil, which assists in
dissipating and transferring heat. They are equipped with a central
well, into which a thermometer is placed to measure, with very close ap-
proximation, the coil temperature. A typical manufacturers’ specification
for stability is 20 ppm per year, which is, in my experience, highly con-
servative for a unit which is handled carefully.
67
T = (L - CR2)/R
and may also be expressed as the time required for the current to
reach l/s of its final value after a fixed voltage is impressed. Obviously,
for a-c bridge service, the time constant must be short enough to allow
the full value to be achieved within 1 ppm or 0.1 ppm.
68
NATIONAL PHYSICAL LABORATORY
Tcddington Middlesex TWll OLW England
Certificate of Calibration
STANDARD RESISTOR
PREVIOUS
CERTIFICATE; NOIF2
MFASUREMFNTS :
The resistor wa+ immersed in an oil-bath controlled at II temperature Of
20.000 t 0.005'c for at least 2 hours prior to and during the meesurement
and was measured in a 4-terminal configuration. The power dissipated in
the resistor was less thnn 1 mW.
Resist@Jxe Uncertainty
1 2
Confidence Level Arithmetic Sum
fit least yjx of Contributions
9.999 917 ohm * 0.25 ppm * 0.43 ppln
The& uncertainties refer only to the measured value and do not carry any
implication regarding the stability of the instrument.
R!z.sULTS:
Uncertainty
1 2
Test frequency Rf/Rdc Confidence Level Arithmetic Sum
IIt &least 95% of Contributions
These uncertainties refer only to the measured value and do not carry any
implication regarding the stability of the instrument.
resistor design are that (a) resistance wires with 1OW Siebeck
coefficients should be chosen (b) the physical design should exert every
effort to maintain the alloy-to-copper and the winding itself in an
isothermal condition.
With careful consideration of all the above factors, Wilkins has produced
resistors whose dc and ac resistances are the same within 0.1 ppm to
frequencies as high as 1.6 kHz.
(At high temperatures other a-c effects have been noted., Studies of the
best insulations used for the former6 on which SPRT windings are sup-
ported indicate that the degradation of the insulating properties of
quarts and sapphire is marked. I have measured insulation resistance in
a thermometer of 120 MP per square at 960-C, which drops to about 20
MP per square at 11OO’C. 20 MP of random shunt resistance cannot be
ignored, and this effect is a prominent reason why ITS-90 terminates
the SPRT range at the silver freezing point (961’C) instead of the gold
freezing point .(1064’C).) Such a shunt effect would be, of course, as
undesirable with dc as with ac, but much of the effect seems to be due
to a long time-constant polarization, so that resistance rises over a time
which may be measurable in minutes. Thus the recovery of most ~of the
insulation resistance would take place with a direct current impressed,
but might not within the reversal time of an alternating current.
72
An elementary (Wheatstone)
bridge is shown schematically
in Fig. 7, where A and B are
fixed resistors (often of
identical value, in which case
the bridge is said to be
“equal-arm”, a resistance
decade S, and an unknown
resistance to be measured, X.
At balance, when the unknown
resistance is balanced by the
decade resistance (assuming A
= B), the potential drop from
Junction 1 to Junction 2 is
equal to the potential drop
from Junction 1 to Junction 4,
so that i,bA = ixsX, and iabB
= rxsS. These equalities can
FIGURE 7 be expressed as
A WHEATETONE BRIDGE
A/B = X/S
which is the equation of balance for the Wheatstone bridge. At this bal-
ance, no current flows in the galvanometer arm G.
The equalities of the balance condition include the entire bridge circuit,
including any external lead wires which extend to the resistance, X,
which is the subject of the measurement. These external resistances may
be negligible, for example is X is very large in comparison with Rll and
R12, but in an elementary bridge, they are always a component of the
measurement in series with X. In industrial resistance thermometry, lead
resistances can be partially compensated, or almost completely compen-
sated under specific restrictions, (e.g., by inserting a dummy loop of
lead wire in the opposite side of the bridge), but must be eliminated for
precise and standards-quality resistance thermometry. Figs. 8(a) and
8(b) show two methods for making such connections. Obviously, the
degree of compensation depends upon the other resistances in the
circuit; for example, compensation is more closely. approximated if the
bridge is constructed with equal ratio arms.
Sensor
FIG 8A FIG 8B
(Anyone visiting NIST will be interested to have a look into the Museum,
which is rather well concealed behind a door to the right of the main
74
The bridge is always used with equal ratio arms of moderate individual
resistance, e.g., 500 Q or 3000 Q. An small slide wire is provided which
is used to balance these resistances exactly, by interchanging them.
Once the arms are equal, the zero resistance value can be determined
precisely. Commutators are supplied for these adjustments.
Since the bridge ratio is 1:1, if the C and T leads are equal in resis-
tance, they will cancel each other, and the net resistance added to the
X arm of the bridge is only the resistance of the thermometer coil. At
the level of precision required of this measurement, leads C and T can-
not be assumed to have equal resistance. A reversal of leads C and T
FIGURE 10
The complete circuit of the Leeds and Northrup G-3 Mueller Bridge. The
least dial is 10 pQ. Bridge arm resistances of 500 or 3000 P may be chosen.
Note that the x0.1 Q decade is subtractive.
R2 + T = RT t C, or R2 = RT t C - T
(RI + R2)/2 = RT t T - C t RT + C - T
= ;RT/2 = RT
6R = R - RS/(R t S) = R2/(R t S)
(10 - n)6R;
S q [(R2/(10 - n)6RJ - R
FIGURE 11
Rs
R6 R7 R8 R9 R10 RI1
Resistors Rl through Rll are the same for all decades. They shunt Rs,
adding one increment of resistance per dial step. The values shown are one
of a number of possible sets of value.
Decade Rl
R2 = 5.6 Q x 0.1 pi; 50.4 - J50.5
R3 = 1.0 Q X 0.01 50.4 - 45.04
R4 = 9.0 Q X 0.001 llo:504 50.4 - JO.504
R5 = 12.0 Q x 0.0001 JO.0504 50.4 - JO.0504
R6 = 16.8 Q
RI = 25.2 Q Note that with all dials set at 0, the value of the
R8 = 42.0 Q decade string is not zero. This non-zero resistance
R9 = 84.0 Q is balanced by an equal and trimmable resistor in
RlO = 252.0 Q the opposing arm of the bridge.
Rll = m
If the ratio R2/6R is made the same for all decades, then identical sets
of fixed coils per step may be used for all decades. The resistances
which must be put in series with the Rl, when Rs2 = 50.4, are shown in
Fig. 11
78
,- j
FIGURE 12
The circuit schematic of the Guildline Model 9975 Direct Current Comparator.
The ac carrier of the F-17 bridge is fixed at 1-i line frequency; for 50
Hz supplies at 75 Hz, and for 60 Hz supplies at SO Hz. This choice al-
lows maximum rejection of noise from the line and from active circuit
components, while providing maximum detector bandwidth. The more so-
phisticated F-18 operates at * and 1% line frequency (25 Hz and 75 Hz
for 50 Hz supplies and 30 Hz and 90 Hz for 60 Hz supplies) selectable.
This feature allows the estimate of any effects on the measurement due
tn carrier frequency, and extrapolation to dc where necessary.
A block diagram of the F-17 and F-18 bridges is shown in Fig. 13. A
stable ac signal is produced by a carrier generator current source. This
drives current through the standard resistor and the unknown resistor,
which are connected in series. The voltage generated across R, is used
as the reference signal to excite the input windings of a multistage in-
ductive divider. The inductive divider’s secondary output is compared
with the voltage appearing across the unknown resistor Rx, in this case,
the SPRT, by the detector circuitry. The inductive divider acts as a
precision ratio transformer. Its taps are adjusted to balance (that is, to
bring to zero) the output to the detector circuit. At balance, the voltage
from the inductive divider is exactly equal and opposite to that appear-
ing across Rx. The output of the inductive divider is also a precise ra-
F17 and FIB Schematic
Detector Waveform
Generator
Referknce
Switched
current
Source
and Ratio
Transformer
,
Unknwm
Resistor
Lke-
Amplifie?
S!?hed
Gain
Amplifier
q
Phase
Sensitive
---L ’
Low
Pass
-
Detector Filter
FIGURE 13
83
tio of the voltage across Rs. Since the current flowing through Rs and
R, is identical, the ratio set on the inductive divider is equal to the
ratio Rx/R,.
The nominal accuracy of the F-17 bridge is lppm, and of the F-18,
O.lppm. The F-18 is optimised specifically for SPRT values of 100, 25.5,
2.5 and 0.25 Q. The F-18 has 7: decades, and reads resistance ratios
over the range 0.000 000 0 and 1.299 999 9. A generous selection of op-
erating sensitivities and of thermometer currents (from 0.1 mA to 50 mA
rms, (F-18) including a (2 multiplier) are supplied.
IMPORTANT REFERENCES
In this text, I have relied, as I have for many years, upon two impor-
tant references on fixed resistors and resistance measurement. These
me:
Forest K. Harris, Electrical Measurements, John Wiley & Sons, New York,
Chapman and Hall, London, 1956
FOOTNOTES
(4) [H. Tinsley Ltd, Standards House, Croyden, Surrey, CR0 4RR,
England]
(6) Leeds and Northrup Models 8067 (G-l), 8069B (G-2) and 8071B (G-3)
Mueller Bridges, respectively. [Leeds and Northup Company, North Wales,
Pennsylvania USA]
(9) Models F-17 and F-18 [Automatic Systems Laboratories Ltd., 28 Blun-
dells Road, Bradville, Milton K‘2YlES MK13 7HF, England]